Search Results
Found 102 results
510(k) Data Aggregation
(106 days)
LPN
Bausch + Lomb Preservative Free Lubricating and Rewetting Drops are indicated for use to lubricate and rewet soft (hydrophilic) contact lenses, including silicone hydrogel as well as silicone acrylate (SA) and fluorosilicone acrylate (FSA) rigid gas permeable (RGP) contact lenses during wear. The product may be used with daily or extended wear, and disposable lenses.
Bausch + Lomb Preservative Free Lubricating and Rewetting Drops is a sterile, preservative free, buffered, isotonic, aqueous solution that contains boric acid, sodium borate, potassium chloride, erythritol, poloxamine, glycerol and hyaluronan for lubricating and rewetting soft (hydrophilic) contact lenses including silicone hydrogel contact lenses and rigid gas permeable contact lenses during wear.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
The document describes the device, Bausch + Lomb Preservative Free Lubricating and Rewetting Drops, as being substantially equivalent to the predicate device, OPTI-FREE Replenish Rewetting Drops. The "acceptance criteria" are implicitly defined by demonstrating non-inferiority to the predicate device in terms of safety and effectiveness.
Here's a table summarizing the implicit acceptance criteria (non-inferiority to the predicate) and the reported device performance, as evidenced by the clinical study:
Acceptance Criteria (Implicit: Non-inferiority to Predicate Device) | Reported Device Performance (Bausch + Lomb Preservative Free Lubricating and Rewetting Drops) |
---|---|
Safety: | |
Non-inferiority to predicate for slit lamp findings > Grade 2 | Achieved. The Bausch + Lomb Preservative Free Lubricating and Rewetting Drops group demonstrated non-inferiority to the OPTI-FREE Replenish Rewetting Drops group Over All Follow-up Visits for slit lamp findings greater than Grade 2. |
Acceptable adverse event profile | Minimal. Five ocular treatment-emergent adverse events (TEAEs) in three eyes of two subjects (eye irritation (2), eye pruritus (2), eye complication associated with ocular discomfort with device (contact lens) (1)). All resolved with no action required. No discontinuations due to AEs. No corneal infiltrates or changes to pre-existing corneal scars. No notable differences in lens wear/performance parameters or visual acuities compared to control. |
Effectiveness: | |
Non-inferiority to predicate for overall comfort | Achieved. Treatment differences indicated that the Bausch + Lomb Preservative Free Lubricating and Rewetting Drops is non-inferior to the OPTI-FREE Replenish Rewetting Drops solution for overall comfort averaged over all follow-up visits. |
Non-inferiority to predicate for dryness | Achieved. Treatment differences indicated that the Bausch + Lomb Preservative Free Lubricating and Rewetting Drops is non-inferior to the OPTI-FREE Replenish Rewetting Drops solution for dryness averaged over all follow-up visits. |
Non-inferiority to predicate for optimal (none/light) lens deposits | Achieved. Treatment differences indicated that the Bausch + Lomb Preservative Free Lubricating and Rewetting Drops is non-inferior to the OPTI-FREE Replenish Rewetting Drops solution for optimal (none or light) lens deposits at all follow-up visits. |
Acceptable performance for additional effectiveness assessments | Achieved. No notable differences between the two treatment groups with regard to additional effectiveness assessments of symptoms/complaints, worn lens characteristics, and dispensed lens characteristics. |
Compatibility with various lens types | Demonstrated. Compatible with soft contact lenses (including silicone hydrogel), rigid gas permeable, silicone acrylate (SA), and fluorosilicone acrylate (FSA) lenses, as evidenced by lens compatibility studies and inclusion of these lens types in the clinical study. |
Study Details:
2. Sample size used for the test set and the data provenance:
- Sample Size (Clinical Study):
- Enrolled: 369 subjects
- Completed: 365 subjects
- Test Group (Bausch + Lomb Preservative Free): 188 subjects
- Control Group (OPTI-FREE Replenish): 181 subjects
- Data Provenance: The study was a "multicenter, randomized, masked, parallel, bilateral study conducted in the USA." This indicates prospective data collection from multiple sites within the contiguous United States.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
The document does not specify the number of experts used or their qualifications for establishing ground truth for the clinical study's endpoints. The "ground truth" for safety and effectiveness endpoints (e.g., slit lamp findings, comfort, dryness, lens deposits) would typically be assessed by the clinicians (e.g., optometrists or ophthalmologists) participating in the multicenter study.
4. Adjudication method for the test set:
The document does not specify an explicit adjudication method (e.g., 2+1, 3+1). Since it was a "masked" study, assessors were likely blinded to the treatment assignment. The primary endpoints were compared statistically between the two groups. It's implied that the findings from the clinicians at each site contributed to the data for analysis.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, an MRMC comparative effectiveness study was not done. This study is for contact lens rewetting drops, not an AI-powered diagnostic device. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. This device is a medical product (rewetting drops) and not an algorithm or AI system.
7. The type of ground truth used:
For the clinical study, the ground truth was based on clinical assessments and subjective patient reporting.
- Clinical Assessments: E.g., slit lamp findings (graded by clinicians).
- Subjective Patient Reporting: E.g., overall comfort and dryness (self-reported by subjects).
- Objective Observations: E.g., lens deposits (likely observed and graded by clinicians).
8. The sample size for the training set:
Not applicable. This is a medical product (rewetting drops), not an AI algorithm that requires a training set.
9. How the ground truth for the training set was established:
Not applicable. As above, there is no training set for this type of device.
Ask a specific question about this device
(163 days)
LPN
Aqua Naina Plus Sterile Saline Solution: The Aqua Naina Plus Sterile Saline Solution for single use is indicated for rinsing soft (hydrophilic) and rigid gas permeable (hard) contact lenses prior to lens insertion. This solution may be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses. The Aqua Naina Plus Sterile Saline Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner.
Aqua Naina Sterile Saline Solution: The Aqua Naina Sterile Saline Solution is indicated for rinsing soft (hydrophilic) and rigid gas permeable (hard) contact lenses prior to lens insertion and following proper lens disinfection as recommended by the eye care practitioner. The Aqua Naina Sterile Saline Solution may be used as needed throughout the day to rinse contact lenses and contact lens cases.
Aqua Naina Plus Sterile Saline Solution is a sterile, preservative-free, buffered, isotonic, clear (colorless) aqueous solution containing isotonic solution of sodium chloride, boric acid and sodium borate in purified water USP. Aqua Naina Plus Sterile Saline Solution has pH of 7.00 to 7.40 and Osmolality of 290 to 320 mOsmol/kg. The Sterile Saline Solution is packed in single dose 1 fl oz (30 ml) high density polyethylene (HDPE) container with a tamper evident seal. Aqua Naina Plus Sterile Saline Solution is for single use only.
Aqua Naina Sterile Saline Solution is a sterile, preservative-free, buffered, isotonic, clear (colorless) aqueous solution containing isotonic solution of sodium chloride, boric acid and sodium borate in purified water USP. Aqua Naina Sterile Saline Solution has pH of 7.00 to 7.40 and Osmolality of 290 to 320 mOsmol/kg. The Sterile Saline Solution is packed in a multidose 4 fl oz (118ml) high density polyethylene (HDPE) container with a tamper evident seal.
The provided text is a 510(k) summary for a medical device (Aqua Naina Sterile Saline Solution and Aqua Naina Plus Sterile Saline Solution). This document outlines the device's characteristics and its substantial equivalence to existing predicate devices, but it does not contain information about acceptance criteria or specific performance studies as would be typically found for AI/ML-based diagnostic devices.
The relevant section, "VII. PERFORMANCE DATA," explicitly states:
"The package volume change was validated using the same methods identified in K193441. No other performance data (clinical or non-clinical) are required for this special 510(k) submission."
This indicates that for this specific type of device (saline solution), comprehensive performance studies with acceptance criteria, sample sizes, ground truth establishment, expert reviews, and MRMC studies, as typically requested for AI/ML medical devices, were not required by the FDA for this submission. The submission relies on establishing substantial equivalence to previously cleared predicate devices based on chemical composition, intended use, and production methods, rather than demonstrating performance against specific clinical metrics.
Therefore, I cannot provide the requested information in the format of acceptance criteria and study details because such data is not present in the provided document.
Ask a specific question about this device
(241 days)
LPN
Bausch + Lomb ABT12 multi-purpose solution is indicated for use in the daily cleaning, conditioning, removing protein deposits, rinsing, chemical (not heat) disinfection, and storage of soft (hydrophilic) contact lenses, including silicone hydrogel contact lenses, as recommended by your eye care practitioner.
Bausch + Lomb ABT12 multi-purpose solution is a sterile isotonic aqueous solution [containing polyaminopropyl biguanide (0.0005%), polyquaternium (0.00015%), and alexidine (0.00025%)] for disinfecting, cleaning, conditioning, rinsing, protein removal, and storing soft (hydrophilic) and silicone hydrogel contact lenses. The sterile solution is packaged in a plastic bottle with a tamper evident seal and labeled with a lot number and expiration date.
Based on the provided text, the device in question is a multi-purpose contact lens solution (ABT12 multi-purpose solution), not an AI-powered medical device for image analysis. Therefore, many of the typical acceptance criteria and study details related to AI performance (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this type of product and approval process.
The document describes the regulatory approval of a contact lens solution (Bausch + Lomb ABT12 multi-purpose solution) for which the primary claims are related to its safety and effectiveness in cleaning, disinfecting, and storing contact lenses. The "acceptance criteria" here are based on demonstrating substantial equivalence to a predicate device (COMPLETE multi-purpose solution Easy Rub Formula) through a combination of preclinical (biocompatibility, microbiology, lens compatibility, cleaning efficacy) and clinical studies.
Here's a breakdown of the requested information, adapted to the context of this non-AI medical device:
Acceptance Criteria and Study for Bausch + Lomb ABT12 Multi-Purpose Solution
As the device is a contact lens solution, not an AI-powered diagnostic tool, the typical "acceptance criteria" table for AI performance metrics (e.g., sensitivity, specificity, AUC) is not applicable. Instead, the acceptance criteria are based on demonstrating substantial equivalence to a predicate device in terms of safety and effectiveness.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for a contact lens solution primarily revolve around meeting established safety and effectiveness standards, often demonstrated through non-inferiority to a legally marketed predicate device.
Criteria Category | Specific Acceptance Criteria (Demonstrated via Non-Inferiority to Predicate) | Reported Device Performance (ABT12 Multi-Purpose Solution) |
---|---|---|
Clinical Safety | Non-inferiority in ophthalmological findings (e.g., slit lamp findings) compared to predicate. Absence of serious adverse events (SAEs) or device-related adverse events (AEs). No significant changes to corneal health. | The ABT12 group demonstrated non-inferiority to the COMPLETE group Over All Follow-up Visits for slit lamp findings greater than Grade 2. No serious adverse events, adverse device effects, or significant non-serious adverse events were reported in the ABT12 multi-purpose solution group. No eyes were discontinued due to an AE. None of the AEs were considered related to the study solution. No corneal infiltrates were reported, and no changes to pre-existing corneal scars. Visual acuities were similar between groups. Conclusion: Safe for use. |
Clinical Effectiveness | Non-inferiority in user-reported comfort and dryness. Non-inferiority in lens deposit prevention/removal. | For the three primary effectiveness endpoints (overall comfort averaged over all follow-up visits; dryness averaged over all follow-up visits; and optimal [{none or light} lens deposits at all follow-up visits), the treatment differences indicated that the ABT12 solution is non-inferior to the COMPLETE solution. No notable differences were found regarding symptoms/complaints, worn lens characteristics, and dispensed lens characteristics. Conclusion: Effective for use. |
Biocompatibility | No significant cytotoxicity, ocular irritation, oral toxicity, or sensitization. Compatibility with various lens materials. | Studies demonstrated the biocompatibility of ABT12 multi-purpose solution, including tests on the solution itself, with maximum preservative concentrations, and with extracts from different lens materials. Leveraged existing clearance for packaging components. Predicate K014202 was used for side-by-side comparison in relevant studies. |
Microbiological Efficacy | Meets or exceeds criteria for disinfection and preservative efficacy against relevant microorganisms (e.g., as per EN ISO 14729:2001 and EN ISO 14730:2014). | The solution exceeds the criteria for disinfection and preservative efficacy. |
Lens Compatibility | Compatible with soft contact lenses, including silicone hydrogel contact lenses. | Studies demonstrated compatibility with soft contact lenses, including silicone hydrogel contact lenses. |
Cleaning Efficacy | Demonstrates effective cleaning properties (e.g., protein removal). | In-vitro laboratory studies demonstrated cleaning properties with artificially protein-deposited contact lenses. Surfactant concentrations are well above the Critical Micelle Concentration (CMC). |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Clinical Study - Test Set): A total of 252 subjects were enrolled. 240 subjects completed the study. These subjects were randomized into two groups:
- ABT12 multi-purpose solution (Test): 127 subjects
- COMPLETE multi-purpose solution (Control/Predicate): 125 subjects
- Data Provenance: The clinical study was a multicenter study conducted in the USA.
- Retrospective or Prospective: This was a prospective, randomized, masked, parallel, bilateral clinical study.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This question is not applicable in the context of this device. The "ground truth" for a contact lens solution's performance is established through direct clinical observations (e.g., slit lamp findings by ophthalmologists/optometrists) and patient-reported outcomes, as well as laboratory testing results, rather than expert consensus on image interpretation. The study was masked, implying that the clinicians assessing outcomes were blinded to the treatment assignment.
4. Adjudication Method for the Test Set
This concept (adjudication of interpretations) is not applicable to this type of study. Clinical outcomes were measured directly by investigators in a masked fashion and patient-reported outcomes were collected.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC study was not done. MRMC studies are typically used to evaluate the impact of a diagnostic tool (like AI) on physician performance in interpreting medical images. This study focused on the safety and effectiveness of a physical product (contact lens solution) in a clinical trial setting.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. There is no algorithm for this device. The "standalone performance" of the contact lens solution is demonstrated through the preclinical studies (biocompatibility, microbiology, cleaning efficacy, lens compatibility) which evaluate the product's intrinsic properties irrespective of human handling errors (though proper usage instructions are provided).
7. The Type of Ground Truth Used
The ground truth was established through a combination of:
- Clinical Measurements: Objective clinical assessments by masked investigators (e.g., slit lamp findings, visual acuities).
- Patient-Reported Outcomes: Subjective data gathered directly from subjects regarding comfort, dryness, and other symptoms.
- Laboratory Testing: In-vitro and ex-vivo studies to assess intrinsic properties like microbiological efficacy, cleaning efficacy, and biocompatibility.
- Comparison to Predicate: The "truth" of effectiveness and safety was established by demonstrating non-inferiority to a known, legally marketed predicate device.
8. The Sample Size for the Training Set
This question is not applicable as this is not an AI/machine learning device.
9. How the Ground Truth for the Training Set was Established
This question is not applicable as this is not an AI/machine learning device.
Ask a specific question about this device
(138 days)
LPN
OTE MPS 045 Multi-Purpose Lens Care Solution is indicated for daily cleaning, chemical (not heat) disinfection, conditioning, rinsing and storage of hydrogel and silicone hydrogel soft (hydrophilic) contact lenses as recommended by your eye care practitioner.
OTE MPS 045 Multi-Purpose Lens Care Solution is a sterile, buffered aqueous solution that contains sodium hyaluronate, poloxamer 407, boric acid, sodium borate, propylene glycol, and is preserved with Polyaminopropyl Biguanide Hydrochloride 0.00002%, Polyquaternium-2 0.00043%, and edetate disodium.
The provided text describes a 510(k) premarket notification for the "OTE MPS 045 Multi-Purpose Lens Care Solution." The document focuses on demonstrating substantial equivalence to a predicate device, BioTrue® Multi-Purpose Solution, rather than providing explicit acceptance criteria and performance against those criteria in a typical AI/device study format.
However, I can extract the relevant information from the "Clinical Data" section to answer your request in the closest possible way, interpreting "acceptance criteria" as the primary endpoints of the clinical study and "reported device performance" as the study findings for those endpoints.
Here's the breakdown based on the provided text:
1. Table of Acceptance Criteria and the Reported Device Performance
For the clinical study, the "acceptance criteria" can be interpreted as the successful demonstration of non-inferiority for the specified primary endpoints.
Category | Acceptance Criteria (Study Endpoints) | Reported Device Performance (Study Findings) |
---|---|---|
Primary Safety | Non-Inferiority of slit lamp findings | The safety endpoint was supported by the study findings. |
Non-Inferiority of limbal hyperemia | The safety endpoint was supported by the study findings. | |
Non-Inferiority of bulbar hyperemia | The safety endpoint was supported by the study findings. | |
Non-Inferiority of corneal staining | The safety endpoint was supported by the study findings. | |
Primary Effectiveness | Non-Inferiority of comfort | The efficacy endpoint was supported by the study findings. |
Non-Inferiority of visual acuity | The efficacy endpoint was supported by the study findings. | |
Non-Inferiority of surface wettability (compared to BioTrue) | OTE MPS 045 Multi-Purpose Lens Care Solution was non-inferior to BioTrue with respect to wettability. | |
Non-Inferiority of lens deposits (compared to BioTrue) | OTE MPS 045 Multi-Purpose Lens Care Solution was non-inferior to BioTrue with respect to film deposits. |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: 199 subjects.
- Data Provenance: The study was a multi-center study, but specific country of origin is not mentioned. It was a prospective study, as it involved subjects using the device for a duration of 30 days.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not specify the number of experts, nor their qualifications, used to establish ground truth for the clinical study endpoints. It mentions "Investigator" as one of the masked parties, suggesting that clinical investigators (likely ophthalmologists or optometrists) assessed the endpoints, but no specific number or detailed qualifications are provided.
4. Adjudication method for the test set
The document does not describe an adjudication method. It was a "double-masked (Investigator and Subject)" study, implying that the investigator assessments would form the basis of the outcome.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, this was not an MRMC study related to AI. This was a clinical trial comparing a new lens care solution to a predicate device.
- Effect size of human readers improve with AI: Not applicable, as this study does not involve AI assistance for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
No, this was a clinical study involving human subjects and a medical device (lens care solution), not an algorithm or AI.
7. The type of ground truth used
The ground truth was established through direct clinical observations and assessments by investigators (implied to be eye care practitioners) and subject-reported outcomes during the 30-day clinical study. This would encompass clinical findings (slit lamp, hyperemia, corneal staining) and subjective reports (comfort, visual acuity).
8. The sample size for the training set
Not applicable. The provided document describes a clinical study for a traditional medical device (lens care solution), not a machine learning model, so there is no "training set."
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of device and study.
Ask a specific question about this device
(167 days)
LPN
Aqua Naina Sterile Saline Solution is indicated only for rinsing soft contact lenses after cleaning and disinfection before use.
Aqua Naina Sterile Saline Solution is a sterile, preservative-free, buffered, isotonic, clear colorless aqueous solution containing isotonic solution of sodium chloride, boric acid and sodium borate in purified water USP. Aqua Naina Sterile Saline Solution has pH of 6.30 to 7.80 and Osmolality of 280 to 320 mOsmol/kg The Sterile Saline Solution is packed in 4 fl oz (118ml) high density polyethylene (HDPE) container with a tamper evident seal.
The provided text describes the 510(k) Premarket Notification for Aqua Naina Sterile Saline Solution and its substantial equivalence to predicate devices. It focuses on the chemical composition, intended use, and various safety tests (sterility, biocompatibility, ocular irritation, cytotoxicity).
However, this document does not describe acceptance criteria, performance data, or study details relevant to a machine learning or AI device. The device in question is a sterile saline solution, which is a chemical product, not a medical imaging or AI diagnostic device. Therefore, the requested information about test sets, data provenance, expert consensus, MRMC studies, or training sets is not applicable to this submission.
The "study that proves the device meets the acceptance criteria" in this context refers to a series of chemical and biological tests conducted on the saline solution to demonstrate its safety and performance, as outlined in section (7) "PERFORMANCE STSTANDARDS APPLIED".
Here's a breakdown of what is provided and why it doesn't fit the typical AI/ML device study format:
-
Device: Aqua Naina Sterile Saline Solution (a chemical product for contact lens care).
-
Acceptance Criteria & Reported Performance: These are presented implicitly through the successful completion of various tests against established standards for saline solutions.
- Sterility: USP
- Bacteriostasis: USP
- Ocular irritation: Tested per ISO 10993-10:2010
- Cytotoxicity (Direct contact test & MEM Elution test): Tested per ISO 10993-5:2009
- Biocompatibility: Tested per ISO 10993-1:2009
- pH: 6.30 to 7.80 (reported in description)
- Osmolality: 280 to 320 mOsmol/kg (reported in description)
The document states: "Results of all testing demonstrate the solution is non-toxic, is comparable to other currently marketed soft contact lens solution and is substantially equivalent to legally marketed predicates." This is the reported device performance meeting the acceptance criteria set by these standards.
-
Sample Size for Test Set and Data Provenance: Not applicable. This refers to chemical and biological samples, not image datasets. The tests are laboratory-based, not based on patient data in the sense of AI/ML.
-
Number of Experts, Qualifications, Adjudication Method: Not applicable. These relate to human interpretation of diagnostic data, which is not relevant for testing a saline solution.
-
MRMC Comparative Effectiveness Study: Not applicable. This is for evaluating AI assistance in diagnostic tasks.
-
Standalone Performance (algorithm only): Not applicable, as there is no algorithm.
-
Type of Ground Truth: Not applicable in the AI/ML sense. Ground truth here is established by laboratory measurements and biological assays, comparing results against defined pass/fail criteria from recognized standards (e.g., USP, ISO).
-
Sample Size for Training Set & How Ground Truth for Training Set was Established: Not applicable. There is no machine learning model involved, therefore no training set. The "training" of the product is its formulation and manufacturing process, and its "ground truth" is its chemical and physical properties verified through standard laboratory testing.
In summary, the provided document details the regulatory clearance of a contact lens saline solution, not an AI/ML-based medical device. Therefore, the questions posed regarding AI/ML device acceptance criteria and study methodology cannot be answered from this information.
Ask a specific question about this device
(53 days)
LPN
PuriLens Plus Preservative-Free Saline is indicated for use following proper lens disinfection as recommended by the eye care practitioner. PuriLens Plus Preservative-Free Saline is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses.
The PuriLens Plus Preservative Free Saline is a sterile, buffered saline solution for rinsing and storing contact lenses. The solution is buffered to roughly physiological pH and has an osmolarity similar to that of human tears. The product is sold over-the-counter in various packaging configurations.
The current generation of the product, Purilens Saline Solution, was cleared under K002319 in 2000, and has been legally marketed in the US since that time. This 510(k) application is to expand the indications for use to include use with rigid gas permeable lenses. There is no change in the device design, formulation, or manufacturing. The labeling will be revised with the new indications.
This submission describes a 510(k) for a contact lens solution, PuriLens Plus Preservative Free Saline, and argues for its substantial equivalence to a predicate device based on similar formulation, manufacturing, and packaging, and expands its indications for use.
Since this is a submission for substantial equivalence based on a predicate device and there have been no changes to the device design, formulation, or manufacturing, no performance data or studies are presented. Therefore, I cannot provide information on acceptance criteria, specific study details, sample sizes, ground truth establishment, or expert involvement as these were not required or reported in this 510(k) summary.
The summary states: "Because the subject is identical to the predicate in terms of formulation, manufacturing, and packaging, no bench-testing is required to demonstrate substantial equivalence." (Page 5)
As such, for all the requested points regarding acceptance criteria, performance data, sample sizes, ground truth, expert involvement, and comparative effectiveness studies, the answer is that this information is not applicable or not provided in this 510(k) summary due to the nature of the submission (substantial equivalence based on an identical product with expanded indications, requiring no new bench testing).
Ask a specific question about this device
(234 days)
LPN
Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution is indicated for use in cleaning, daily protein removal, disinfection, and storing of soft (hydrophilic) contact lenses (including silicone hydrogel lenses) and rigid gas permeable (fluoro silicone acrylate and silicone acrylate) contact lenses, as recommended by your eye care professional.
The Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution system consists of: Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution and the special lens case. The lens case consists of a transparent cup and a lens holder-lens case cap assembly with a neutralizer catalyst disk. Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution and the special lens case must always be used together.
The preservative free, aqueous Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution contains: hydrogen peroxide 3%, phosphonic acid compound (stabilizer), citric acid, polyethylene glycol, sodium chloride, sodium hydroxide, and phosphate (buffer system).
The provided document describes the FDA 510(k) clearance for Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution. Here's a breakdown of the requested information based on the text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state numerical "acceptance criteria" for performance metrics in a table. Instead, it states that tests were conducted according to established guidance and protocols, and the results "demonstrated" effectiveness or compatibility. The general acceptance criterion for most tests appears to be "similar to the predicate device" or "effective/compatible."
Test Category | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Lens Compatibility | Compatible with various commercially available hydrogel, silicone hydrogel, and RGP lenses. | "Results demonstrated that the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution was compatible with the test contact lenses." |
Neutralization Profile | Similar neutralization profile to the predicate device; neutralizer catalyst disk stability. | "The Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution showed a similar neutralization profile to the predicate device. In addition, stability of the neutralizer catalyst disk was confirmed for 90-cycles treatments." |
Disinfecting Efficacy | Harmful microorganisms reduced to an ISO level at specified time points. | "Results demonstrated that harmful microorganisms were effectively reduced to an ISO level at 6 and 24 hours and 7 days after neutralization." |
Preservative Efficacy | Ensure 3-month open bottle discard date. | "Preservative Efficacy test was conducted to ensure 3-month open bottle discard date of the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution." |
Cleaning Efficacy | Similar performance to the predicate device in protein and lipid removal. | "Results suggested that both solutions performed similarly in the removal of lens deposition." |
Toxicology/Biocompatibility | Non-toxic and non-irritating for the neutralized solution, lens case, and catalyst. | "Results demonstrated that the neutralized solution, the lens case materials and the neutralizer catalyst were non-toxic and nonirritating under the experimental conditions." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the exact sample sizes (e.g., number of lenses, microorganisms, or test runs) for each non-clinical test. It mentions testing with:
- Lens Compatibility: Six commercially available hydrogel/silicone hydrogel soft contact lenses and two commercially available RGP lenses.
- The data provenance (country of origin, retrospective/prospective) and specific details about the laboratories conducting these non-clinical tests are not provided in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This device is a cleaning and disinfecting solution, not an AI or diagnostic imaging device that requires human expert review for "ground truth." The "ground truth" for its performance is established through objective laboratory tests and scientific protocols that measure chemical properties, microbiological reduction, and material compatibility. Therefore, the concept of "experts establishing ground truth" in the way it applies to AI diagnostic studies is not relevant here.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or AI diagnostic evaluations where human readers interpret data. This is a non-clinical, laboratory-based performance study for a chemical solution. Therefore, such adjudication methods are not applicable and not mentioned.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done. This is a chemical cleaning solution, not an AI-assisted diagnostic device, so comparison with human readers is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to an algorithm's performance. As the device is a chemical solution, this concept is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the performance of this device is established through:
- Scientific Protocols and Laboratory Measurements: For lens compatibility, neutralization profiles (residual hydrogen peroxide levels), disinfecting efficacy (microbial reduction counts against ISO standards), preservative efficacy, cleaning efficacy (quantification of protein/lipid removal), and toxicology (in-vitro and in-vivo biocompatibility tests based on GLP regulations). This is essentially objective, measurable data generated by laboratory instruments and techniques following validated methods.
8. The sample size for the training set
This device is not an AI algorithm requiring a "training set." Therefore, the concept of a training set sample size is not applicable. The performance studies are validation tests, not machine learning model training.
9. How the ground truth for the training set was established
As there is no training set for an AI algorithm, this question is not applicable.
Ask a specific question about this device
(146 days)
LPN
REFRESH® RELIEVA™ FOR CONTACTS is indicated to lubricate and rewet soft and rigid gas permeable contact lenses, to help relieve dryness, discomfort and irritation that may be associated with lens wear and to cushion lenses by placing a drop on the lens prior to the application on the eye.
REFRESH® RELIEVA™ FOR CONTACTS is a sterile, buffered, isotonic, preserved solution. This aqueous formulation includes carboxymethylcellulose sodium, glycerin, boric acid, calcium chloride dihydrate, erythritol, magnesium chloride hexahydrate, potassium chloride, purified water, sodium borate decahydrate, sodium citrate dihydrate, sodium hyaluronate and the preservative PURITE® (stabilized oxychloro complex 0.01%). The solution may also contain hydrochloric acid and/or sodium hydroxide to adjust pH. All ingredients except for Purite are compendial grade (USP/NF/Ph Eur). The product is supplied in a teal colored, low density polyethylene (LDPE) multi-dose bottle and tip with a teal colored high impact polystyrene (HIPS) cap. The bottles are further packaged in paper cartons with a patient information insert.
This document is a 510(k) Summary for the REFRESH® RELIEVA™ FOR CONTACTS device. It describes the device, compares it to predicate devices, and summarizes testing performed to support its substantial equivalence. The document is a regulatory submission for a medical device and does not inherently describe a study that uses AI, nor does it provide detailed acceptance criteria related to AI performance.
Therefore, many of the requested sections (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set size) are not applicable as this document pertains to a medical device (contact lens solution), not an AI/ML powered device.
However, I can extract the information relevant to the performance of the device as described in this document.
Device Name: REFRESH® RELIEVA™ FOR CONTACTS
Device Description: A sterile, buffered, isotonic, preserved solution containing carboxymethylcellulose sodium, glycerin, boric acid, calcium chloride dihydrate, erythritol, magnesium chloride hexahydrate, potassium chloride, purified water, sodium borate decahydrate, sodium citrate dihydrate, sodium hyaluronate and the preservative PURITE® (stabilized oxychloro complex 0.01%).
Indications for Use: To lubricate and rewet soft and rigid gas permeable contact lenses, to help relieve dryness, discomfort and irritation that may be associated with lens wear and to cushion lenses by placing a drop on the lens prior to the application on the eye.
Predicate Devices:
- REFRESH® Contacts Lubricating and Rewetting Drops (K992028)
- REFRESH OPTIVE CONTACTS Lubricating and Rewetting Drops (K083812)
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative "acceptance criteria" for performance in the typical sense of numerical thresholds for sensitivity, specificity, or similar metrics for an AI/ML device. Instead, the "acceptance criteria" are implied by the requirement for "substantial equivalence" to predicate devices, which the studies aimed to demonstrate. The performance is reported in terms of safety and efficacy.
Acceptance Criteria (Implied by Substantial Equivalence Goal) | Reported Device Performance |
---|---|
Stability: Chemical, physical, and microbiological stability to support the claimed shelf-life and preservative effectiveness. | Demonstrated: Product showed chemical and physical stability for a 24-month shelf-life. Microbiology testing confirmed sterility maintenance and preservative effectiveness throughout the shelf-life. |
Biocompatibility: Safety and effectiveness according to FDA guidance for contact lens care products. | Demonstrated: In-vitro and in-vivo studies, conducted per FDA guidance (Premarket Notification [510(k)] Guidance Document for Contact Lens Care Products dated 1 May 1997), indicated physical, chemical, and microbiological properties are substantially equivalent to predicate devices. |
Clinical Performance: Safety and efficacy for lubricating, rewetting, and cushioning contact lenses, comparable to marketed predicate devices. | Demonstrated: A 3-month clinical study showed substantial equivalence to REFRESH CONTACTS® for lubricating and rewetting during lens wear, and cushioning upon lens insertion. |
Technological Characteristics: No new significant safety or effectiveness concerns compared to predicate devices. | Demonstrated: Components are not novel, similar concentrations to cleared products, no additional significant safety or effectiveness concerns identified. The intended use is identical to predicate devices. |
2. Sample Size Used for the Test Set and the Data Provenance
-
Sample Size for Test Set:
- Clinical Study: The document states "A 3-month clinical study was conducted... in contact lens wearers." However, the exact number of participants (sample size) in this clinical study is not provided in the available text.
- Biocompatibility/Stability: Not applicable as these are bench/lab tests, not human test sets.
-
Data Provenance: Not explicitly stated (e.g., country of origin). The clinical study was conducted to evaluate the device against a marketed predicate, implying a controlled clinical trial setting. The studies are prospective as they were specifically performed to support the 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This section is not applicable as the described studies are for a contact lens solution, not an AI/ML powered device where expert readers typically establish ground truth. Clinical endpoints in this context would likely be based on participant reporting of symptoms and potentially objective ophthalmological assessments, not "expert ground truth" for image interpretation.
4. Adjudication Method for the Test Set
This section is not applicable for the reasons stated above. Clinical study endpoints would be collected and analyzed, but "adjudication" in the sense of resolving discrepancies between expert interpretations of data (as typically seen in AI/ML performance studies) is not relevant here.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is not applicable. This document describes the safety and effectiveness testing for a contact lens comfort solution, not an AI/ML diagnostic or assistive device. There are no "human readers" interpreting images with or without AI assistance in this context.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This section is not applicable. The device is a physical product (contact lens solution), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the clinical study, the "ground truth" or primary endpoints would likely be based on:
- Patient-reported outcomes (PROs): Subjective relief of dryness, discomfort, and irritation; comfort upon lens insertion.
- Clinical assessments: Potentially objective measures of ocular surface health or lens-eye interaction, as assessed by an optometrist or ophthalmologist, though specific measures are not detailed in this summary.
- The study aimed to demonstrate substantial equivalence in safety and efficacy compared to a predicate device, meaning its performance was judged against the established acceptable performance of the predicate.
For biocompatibility, the ground truth involves adherence to and successful completion of established test methods (e.g., cytotoxicity, irritation, sensitization) as per FDA guidance.
For stability, the ground truth involves chemical, physical, and microbiological specifications being met over time.
8. The sample size for the training set
This section is not applicable. The device is a contact lens solution, not an AI/ML model that requires a training set.
9. How the ground truth for the training set was established
This section is not applicable for the reasons stated above.
Ask a specific question about this device
(28 days)
LPN
For storage of soft (hydrophilic), and rigid gas permeable (RGP) hard contact lenses during disinfection with buffered 3% hydrogen peroxide contact lens solution.
The device is a specialized lens case consisting of a transparent cup with a connected unit of screw cap, lens holders (baskets) and platinum-coated neutralization disc. The platinum catalyst coating and shape of the plastic disc are designed to effectively neutralize hydrogen peroxide in the lens care system to harmless water and oxygen gas.
This looks like a 510(k) summary for a medical device with a minor modification (change in silicone material of a gasket). Generally, devices with such minor changes do not contain extensive studies as they are not typically required for demonstrating substantial equivalence. The document confirms this in Section VII, "Performance Data," stating that the device was evaluated in "biocompatibility, physicochemical and design functionality tests."
Given the information provided in the 510(k) summary:
- No extensive clinical study was conducted or is reported for the device. The primary purpose of this 510(k) (K191795) is to demonstrate substantial equivalence to a legally marketed predicate device after a change in a gasket material.
- The "performance data" mentioned refers to nonclinical testing, which is generally not a clinical trial involving human subjects or AI algorithm performance as you might expect in more complex device submissions.
- The information you are asking for (e.g., sample size for training set, MRMC study, number of experts for ground truth) is typical for AI/ML-driven devices or devices requiring extensive clinical validation. This device (AOCup Lens Case with AODisc) is a contact lens case, which falls under a different regulatory pathway and testing requirements.
Therefore, I cannot provide the detailed information you requested about acceptance criteria and study particulars in the format you specified, as those types of studies and criteria are not described in this 510(k) submission for this specific device. The provided text outlines a regulatory submission for a contact lens case after a material change, not an AI or imaging-based diagnostic device.
However, I can interpret the "Performance Data" section to infer general acceptance criteria and the "study" (nonclinical testing) that proves the device meets them:
General Acceptance Criteria and "Study" (Nonclinical Testing) for AOCup Lens Case with AODisc (K191795)
The primary acceptance criterion for this 510(k) submission is to demonstrate that the modified device (with the new silicone gasket material) maintains the same safety and effectiveness as the predicate device. This is achieved through nonclinical testing to ensure the material change does not negatively impact the device's intended function or biocompatibility.
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria Category | Specific Criteria (Inferred) | Reported Device Performance (Inferred) |
---|---|---|
Biocompatibility | The new gasket material must be biocompatible (e.g., non-cytotoxic, non-irritating, non-sensitizing) and not leach harmful substances. | "The AOCup Lens Case with AODisc and the changed gasket was evaluated in biocompatibility... tests." |
(Implicitly, these tests were successful as the submission concludes substantial equivalence and safety.) | ||
Physicochemical Properties | The new gasket material must maintain appropriate physical and chemical properties (e.g., durability, chemical stability) for its function as a seal in the lens case, particularly in contact with hydrogen peroxide solution. | "The AOCup Lens Case with AODisc and the changed gasket was evaluated in... physicochemical... tests." |
(Implicitly, these tests were successful, indicating the material change did not compromise the physical or chemical integrity of the gasket or its interaction with the solution.) | ||
Design Functionality | The device, with the new gasket, must continue to function as intended (e.g., seal properly, prevent leaks, accommodate lenses, allow neutralization). | "The AOCup Lens Case with AODisc and the changed gasket was evaluated in... design functionality tests." |
"Dimensions and technical function of the gasket are unchanged." | ||
"The modifications of gasket material does not change any indications for use nor the basic technical principle of the device functions." | ||
(Implicitly, these tests were successful, confirming the gasket still performs its sealing role and the overall lens case maintains its designed function for storage and disinfection.) | ||
Substantial Equivalence (Overall) | The modified device must be as safe and effective as the predicate device, and the material change must not raise new questions of safety or effectiveness. | "Successful results of all nonclinical testing supported the substantial equivalence and therefore safety and efficacy of the modified AOCup Lens Case with AODisc to the existing product for its intended use." |
Regarding the other points in your request:
- 2. Sample sized used for the test set and the data provenance: Not applicable. These were nonclinical (bench) tests, not involving a "test set" in the context of clinical data or AI algorithms.
- 3. Number of experts used to establish the ground truth... and qualifications: Not applicable. Ground truth for nonclinical tests is based on established testing standards and specifications, not expert interpretation of data.
- 4. Adjudication method: Not applicable.
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No. This is not an imaging or diagnostic device that would require such a study.
- 6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No. This device does not involve an algorithm.
- 7. The type of ground truth used: For biocompatibility, physicochemical, and design functionality tests, the "ground truth" would be defined by the established industry standards, test methods (e.g., ISO standards for biocompatibility), and internal product specifications for the performance of the gasket and the lens case.
- 8. The sample size for the training set: Not applicable. There is no training set for this type of device.
- 9. How the ground truth for the training set was established: Not applicable.
In summary, the provided document describes a regulatory submission for a physical medical device (contact lens case) with a minor material change. The "study" refers to nonclinical laboratory and bench testing rather than clinical trials or AI performance evaluations, therefore, most of your requested information is not relevant to this specific type of device and submission.
Ask a specific question about this device
(54 days)
LPN
Bausch + Lomb Sensitive Eyes Plus Salinon is a sterile, preserved solution for rinsing soft (hydrophilic) and gas permeable contact lenses after cleaning and before lens insertion with chemical (not heat) and hydrogen peroxide systems.
Bausch + Lomb Sensitive Eyes Plus Saline Solution is a sterile, isotonic, buffered solution that contains boric acid, sodium borate, potassium chloride, sodium chloride, preserved with polyaminopropyl biguanide (0.00003%) and edetate disodium (0.025%). The sterile solution is packaged in a plastic bottle with a tamper evident seal and labeled with a lot number and expiration date.
This document describes the premarket notification (510(k)) for the Bausch + Lomb Sensitive Eyes Plus Saline Solution, which is a sterile, preserved solution for rinsing contact lenses. The study mainly focuses on demonstrating substantial equivalence to predicate devices and expanding its indications for use to include gas permeable contact lenses.
Here's an analysis of the provided text in the context of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present acceptance criteria in a quantitative table format with corresponding device performance metrics for a diagnostic device. Instead, it refers to standards and general requirements for contact lens care products. The "acceptance criteria" are implied by adherence to these standards and the demonstration of safety and effectiveness.
Here's a summary of the performance claims based on the described studies:
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Microbiology: Antimicrobial Preservative Efficacy for Discard Dating | |
(Meets requirements of ISO 14730, Annex C Procedure II for a 90-day discard date) | Sensitive Eyes Plus Saline Solution meets the requirements of ISO 14730, Annex C Procedure II for the designation of a 90-day discard date. The test articles were evaluated initially and again following simulated use for the proposed discard date of 90 days, demonstrating efficacy. |
Microbiology: Regimen Evaluation (Modified Regimen) | |
(Meets FDA performance criteria for regimen evaluation as described in the May 1, 1997 Guidance for Industry, Pre-market Notification (510(k)) Guidance Document for Contact Lens Care Products when used in a 10-second rub, 5-second rinse regimen, and ISO Standard 14729: 2001/Amd. 1:2010 (E)) | A modified regimen procedure (rinse with Sensitive Eyes Plus Saline Solution in place of tap water) for Boston Conditioning Solution and Boston Advance Conditioning Solution demonstrated that the modified regimen with Sensitive Eyes Plus Saline Solution meets FDA performance criteria (1997 Guidance) and ISO Standard 14729:2001/Amd. 1:2010 (E) when used in a 10-second rub, 5-second rinse (per lens side) regimen. |
Biocompatibility | |
(No new testing required as per previous clearance for use and no new indications that necessitated new testing) | Biocompatibility tests were deemed unnecessary for this application because the Sensitive Eyes Plus Saline Solution was previously cleared for use, and the expansion of the indication (to include gas permeable lenses) did not trigger the need for new testing. This implies that the product's biocompatibility was already established and met previous criteria. |
Lens Compatibility | |
(Meets ISO 11981 for rigid gas permeable contact lenses) | Results of lens compatibility studies per ISO 11981 demonstrate Bausch + Lomb Sensitive Eyes Plus Saline Solution is compatible with rigid gas permeable contact lenses. |
Clinical Data / Overall Safety and Efficacy | |
(Established safety and efficacy, substantial equivalence to predicate) | Clinical studies were deemed unnecessary for this application as the Sensitive Eyes Plus Saline Solution has been commercially available for over 25 years with "well demonstrated safety and efficacy." The cumulative results of laboratory testing sponsored by Bausch + Lomb demonstrate substantial equivalence in safety, effectiveness, and performance when used with gas permeable contact lenses to the current product indicated for soft contact lenses. |
2. Sample size used for the test set and the data provenance
- The document does not specify the sample sizes used for the microbiology, biocompatibility, or lens compatibility test sets.
- Data provenance: The studies were performed by or sponsored by Bausch + Lomb. The specific country of origin for the data is not mentioned but appears to be part of a submission to the U.S. FDA, indicating the data would be accepted under U.S. regulatory standards. These are preclinical (laboratory) studies, not typically categorized as retrospective or prospective in the same way clinical trials are.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This information is not provided in the document. The studies are primarily laboratory-based (microbiology, lens compatibility) and a declaration of no new biocompatibility testing needed due to prior clearance. Expert consensus for "ground truth" as typically understood in diagnostic device evaluations is not applicable here.
4. Adjudication method for the test set
- This information is not provided and is not applicable for the type of laboratory and preclinical studies presented. Adjudication methods are typically used in clinical studies involving human observers or subjective assessments.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. This device is a saline solution and not an AI-powered diagnostic device. Therefore, a study involving human readers and AI assistance is not relevant or described.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- No, a standalone algorithm performance study was not done. This device is a physical product (saline solution), not an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the microbiology studies, the "ground truth" would be defined by the established laboratory methods and criteria in ISO 14730 and the FDA Guidance document. This involves direct measurement of antimicrobial efficacy against specified microorganisms.
- For lens compatibility, the "ground truth" is defined by the objective standards and methods outlined in ISO 11981, which assess physical characteristics of lenses after exposure to the solution.
- For biocompatibility, the "ground truth" was established by previous testing and regulatory clearance of the device/components, meaning it met established safety standards.
- The absence of clinical studies for this submission means no direct "outcomes data" or "pathology" was generated for new approval, relying instead on historical data and preclinical testing.
8. The sample size for the training set
- This is not applicable as this submission is for a medical device (saline solution), not an AI/machine learning model that requires a "training set."
9. How the ground truth for the training set was established
- This is not applicable for the same reason as point 8.
Ask a specific question about this device
Page 1 of 11