Intended Use

The Cordis Vanguard Steerable Guidewires are designed for use in angiographic procedures to introduce and position catheters and interventional devices within the peripheral vasculature.

Device Description

The Cordis Vanguard Steerable Guidewires have a multi-tapered stainless steel corewire with a soft, flexible, highly radiopaque platinum alloy coil over the tip. A steering / torquing device may be attached to aid in steering/torquing the guidewire.

AI/ML Overview

This document is a 510(k) premarket notification for the Cordis Vanguard Steerable Guidewire, submitted in 1997. It is a regulatory submission for a medical device and, as such, does not contain information about a study proving the device meets acceptance criteria in the way a clinical trial or performance study for a new AI/software device would.

The document discusses "substantial equivalence" to predicate devices, which is a regulatory pathway for class II devices. The "performance" mentioned refers to the device's functional characteristics compared to existing devices, not a study against predefined acceptance criteria for a novel technology.

Therefore, the requested information cannot be extracted from this document as it pertains to a different type of regulatory submission and device evaluation. Specifically, the following points cannot be answered from the provided text:

  1. A table of acceptance criteria and the reported device performance: This document does not establish specific performance acceptance criteria or report performance data against them. It states the device is "similar in design, construction, indication for use and performance characteristics to other commercially available steerable guidewires."
  2. Sample size used for the test set and the data provenance: Not applicable, as no such "test set" for performance evaluation (e.g., diagnostic accuracy) is mentioned.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  4. Adjudication method for the test set: Not applicable.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable.
  6. If a standalone performance (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used: Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

The document focuses on:

  • Biocompatibility testing: "All appropriate Biocompatibility tests were successfully performed on the materials used for the Cordis Vanguard Steerable Guidewires per ISO 10993-1." (This is a type of acceptance criterion, but not performance in the sense of diagnostic accuracy or effect size.)
  • Substantial Equivalence: The primary claim is that the device is substantially equivalent to existing predicate devices (TAD II Guide Wire System and Wholey Hi-Torque Guide Wire System K861765) in design, construction, indication for use, and performance characteristics.

In summary, this document is a regulatory submission demonstrating substantial equivalence of a physical medical device, not a performance study for a software/AI algorithm against specific acceptance criteria.

{0}------------------------------------------------

00-00008
K974560

MAY | | 1998

Image /page/0/Picture/2 description: The image shows the logo for Cordis, a Johnson & Johnson company. The word "Cordis" is written in a bold, sans-serif font, with a thick line underneath. Below the line, the words "a Johnson & Johnson company" are written in a smaller, serif font. The logo is simple and professional, and it clearly identifies the company.

510(k) Summary of Safety and Effectiveness

Date: December 4, 1997

General Information

Manufacturers AddressContact
Cordis Corporation14201 NW 60th AvenueMiami Lakes, Florida 33014Dennis Griffin
Proprietary NameCommon/Usual Name
Cordis Vanguard SteerableGuidewireCatheter Guide Wire
ClassificationCatheter Guide Wire
Class II, 21 CFR 870.1330, 74DQX
PredicateTAD II Guide Wire System 510K Number Unknown
DevicesWholey Hi-Torque Guide Wire System K861765
management of the program of the promote of Propertial of Propertial Andrews of Children and Children and Children andად იდლისიის მაკულტული სამხრეთვე დღება 1930 წლების მონაწილეოდა სოციალი სამარტი დღელები მამ P-ზილეთი სათარი მარტი სამხრეთით სამარტი მო

Performance standards have not been established by the FDA under section 514 of the Food Performance and Drug Cosmetic Act. Standards

The Cordis Vanguard Guidewire is intended for use in angiographic procedures to introduce Indications for and position catheters and interventional devices within the peripheral vasculature. Use

ので、その他の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の

The Cordis Vanguard Steerable Guidewires have a multi-tapered stainless steel corewire Device with a soft, flexible, highly radiopaque platinum alloy coil over the tip. A steering / Description torquing device may be attached to aid in steering/torquing the guidewire.

Continued on next page

PREMARKET NOTIFICATION Cordis Corporation Vanguard Steerable Guidewire

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the word "Cordis" in a bold, sans-serif font. There is a horizontal line underneath the word. Below the line, the text "a Johnson & Johnson company" is written in a smaller, cursive font. The text is black and the background is white.

510(k) Summary of Safety and Effectiveness, Continued

Biocompatibility

All appropriate Biocompatibility tests were successfully performed on the materials used for the Cordis Vanguard Steerable Guidewires per ISO 10993-1.

Summary of Substantial Equivalence

I

The Cordis Vanguard Steerable Guidewire is similar in design, construction, indication for use and performance characteristics to other commercially available steerable guidewires.

Any statement of substantial equivalence to another product is required by 21 CFR 807.87, and relates only to whether the present product can be marketed without prior reclassification or clinical approval. The present submission is therefore not related to the coverage of any patent, and is not to be interpreted as an admission or used as evidence in a patent infringement lawsuit. As the Commissioner of the FDA has stated, "... a determination of substantial equivalence under the federal Food, Drug, and Cosmetic Act relates to the fact that the product can lawfully be marketed without pre-market approval or reclassification. This determination is not intended to have any bearing whatever on the resolution of patent infringement suits." 42 Fed. Reg. 42,520, et seg. (1977).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY | | 1998

Mr. Dennis S. Griffin Manager, Regualtory and Clinical Affairs Cordis Corporation 14201 N.W. 60th Avenue Miami Lakes, FL 33014

Re : K974560 Cordis Vanguard Steerable Guidewires Trade Name: Requlatory Class: II Product Code: DQX Dated: March 30, 1998 Received: March 31, 1998

Dear Mr. Griffin:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obliqation you might have under sections 531 through 542 of the Act

{3}------------------------------------------------

Page 2 - Mr. Dennis S. Griffin

for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

sincerely yours,

sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

510(K) Number (if known): [To be assigned by the FDA]
Device Name:Vanguard Steerable Guidewire
Indications for Use:The Cordis Vanguard Steerable Guidewires are designed for use inangiographic procedures to introduce and position catheters andinterventional devices within the peripheral vasculature.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Th. A. R.

(Division Sign-Off) (Division Sign-On)
Division of Cardiovascular, Res Division of ourological Devices 510(k) Number

ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ Prescription Use (Per 21 CFR 801.109)

OR

ﺔ ﺍﻟﻤﺘﺤﺪﺓ

Over-The Counter Use _________________________________________________________________________________________________________________________________________________________

PREMARKET NOTIFICATION Cordis Corporation Vanguard Steerable Guidewire

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.