K Number
K972113

Validate with FDA (Live)

Date Cleared
1997-09-02

(89 days)

Product Code
Regulation Number
870.3535
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BSC/CA IABC's are indicated for use in patients with the following conditions: Refractory power failure. Cardiogenic shock. Unstable refractory angina. Histodynamically significant mechanical complications secondary to acute MI: Ventricular septal defect. Mitral valve regurgitation. Papillary muscle rupture. Cardiac support for high risk general surgical and coronary angiography/ angioplasty patients. Septic shock. The intended use of the 30 cc and 40 cc SUB-9 remain identical to that of the currently marketed IABC's. All of the devices are intended to provide temporary circulatory ntireded that 3. The or in athrough controlled mechanical displacement of a volume of blood in the aorta. The mechanical action of the IAB catheter therapy lowers the cardiac workload by two reans : Systolic unloading, as noted by a reduction in the patient's systolic pressure, l . which provides reduced myocardial oxygen consumptions (MVO,). Diastolic augmentation which provides an increase in the mean ann aortic pressure and leads to an improvement in systemic and coronary arterial perfusion. Balloon pump therapy is achieved by inserting an intra-aortic balloon catheter into the descending thoracic aorta via the common femoral artery. Balloon inflation is timed to occur during diastole, beginning with the aortic valve closure. The balloon remains inflated until the onser of left ventricular ejection or systole, then rapidly deflates, reducing the aortic pressure, which in turn reduces the afterload.

Device Description

The proposed 30 cc and 40 cc SUB-9 Stainless Steel and Nitinol consists of a polyurethane blend symmetrical balloon at the distal end of a polyurethane-covered nylon shaft. The balloons are coated with a thin layer of silicone fluid. A Nitinol central lumen runs throughout the length of the 30 cc and 40 cc SUB-9 Nitinol catheters and terminates at the distal tip. A Stainless Steel central lumen runs throughout the length of the 30 cc and 40 cc SUB-9 Stainless Steel catheters and terminates at the distal Each central lumen may be used to pass the devices over their tip. guidewire. The balloon is supplied prewrapped for insertion utilizing either a hemostasis sheath (provided in each IAB insertion kit) or sheathless.

AI/ML Overview

The provided 510(k) summary for the Boston Scientific Corporation/Cardiac Assist (BSC/CA) SUB-9 Intra-Aortic Balloon Catheter describes the device and its substantial equivalence to predicate devices. However, this document does not contain the detailed acceptance criteria, specific device performance metrics, or a formal study report with the requested sample sizes, ground truth establishment, or multi-reader multi-case study information that would typically be found in a comprehensive clinical or non-clinical study report.

The document focuses on demonstrating substantial equivalence through a series of non-clinical tests comparing the new SUB-9 device to its predicate devices. The "acceptance criteria" in this context are implicitly that the SUB-9 design performs substantially equivalently to, or better than, the predicate devices in various mechanical and functional aspects.

Here's an attempt to extract and infer the information based on the provided text, with clear indications where information is not available in the document:


Acceptance Criteria and Device Performance for K912113 SUB-9 Intra-Aortic Balloon Catheter

This 510(k) summary focuses on demonstrating substantial equivalence of the SUB-9 Intra-Aortic Balloon Catheter to its predicate devices through non-clinical testing. The "acceptance criteria" are implicitly met when the SUB-9 device performs comparably to, or better than, the predicate devices in the described tests.

1. Table of Acceptance Criteria and Reported Device Performance

Test CategoryImplicit Acceptance CriteriaReported Device Performance and Remarks
1. Sheathed Insertion TestForce required for insertion is substantially equivalent to predicate devices.The force required to insert the SUB-9 designs (30 and 40 cc, Stainless Steel and Nitinol) was demonstrated to be substantially equivalent to the force required to insert the predicate NICATH™ and Sensation™ designs.
2. Sheathless Insertion TestAbility to pass through restrictions comparable to, or better than, predicate devices.The tightest restriction that the SUB-9 design can pass through was found to be smaller than that required for the predicate BSC/CA IAB's, the Datascope 40 cc 9.5 F Percor-Stat, and the Kontron/Arrow 40 cc 9 F IAB. (SUB-9 performed better)
3. Maximum Pumping Rate Limit TestMaximum pumping rate at which the balloon fully inflates/deflates is comparable to, or better than, predicate devices.The maximum achievable pumping rate limits for the SUB-9 designs were found to be higher than that for the predicate NICATH™ and Models 940 and 930 on the BSC/CA 3001. It is inferred that SUB-9 will perform faster on any IABP it is labeled for use on. (SUB-9 performed better).
4. Reliability/Integrity TestReliability and integrity comparable to predicate devices (ability to cycle at least 3.6 million cycles).The reliability of the SUB-9 was found comparable to all BSC/CA predicate devices, all of which were able to cycle for at least 3.6 million cycles. Maximum pumping rate limit post-reliability was identical to pre-reliability rate for 40cc SUB-9. Inflation and deflation times post-reliability were comparable to pre-reliability times.
5. Aneurization and Burst Pressure TestAneurization and burst pressures are substantially equivalent to predicate devices, and external IABP output pressures are below these limits.Demonstrated substantially equivalent aneurization and burst pressures as compared to the predicate devices. The SUB-9 balloons will not aneurize or burst when used on labeled IABP's, as IABP's cannot provide the volume/pressure required. Peak output pressures of all indicated IABP's are below the pressures required to aneurize and burst the balloons.
6. Transmembrane Pressure and Volume Measurement TestTransmembrane pressure at a given inflated volume is substantially equivalent to predicate devices.Since the balloons on SUB-9 are the same as predicate Models 940 and 930 and 30/40 cc NICATH™, they will have the same transmembrane pressure at the same inflated volume. The volume of the 30 and 40 cc SUB-9 were found substantially equivalent to the predicate Models 940 and 930, demonstrating equal transmembrane pressures.
7. Trackability TestAbility to track over guidewire into position in aorta without incident.All SUB-9 test samples (30 cc and 40 cc; Stainless Steel and Nitinol) were able to track their guidewire into position in the aorta without incident. No catheter kinks, guidewire kinks, or problems inserting/removing were encountered.
8. Kink Resistance TestMean kink radius substantially equivalent to predicate devices.The mean kink radius of the SUB-9 Stainless Steel and Nitinol were demonstrated to be substantially equivalent to the predicate NICATH™, Sensation™, and Models 940 and 930.
9. Initial Performance, Insertion/Removal Integrity TestProper opening per Directions for Use, and integrity maintained post-insertion/removal.All 30 cc and 40 cc SUB-9 Nitinol and Stainless Steel test samples opened as they should per the Directions for Use. The integrity of the SUB-9 design post insertion and removal was demonstrated.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document repeatedly refers to "the SUB-9 designs (30 and 40 cc and Stainless Steel and Nitinol)" or "SUB-9 test samples (30 cc and 40 cc; Stainless Steel and Nitinol)". However, specific numerical sample sizes (e.g., "n=X devices were tested for each configuration") are not provided.
  • Data Provenance: All tests described are "Nonclinical Tests" performed by Boston Scientific Corporation/Cardiac Assist. The origin of the data is internal laboratory testing and physical characterization. The document states, "No clinical testing was performed by Boston Scientific Corporation/Cardiac Assist in support of this premarket notification." Therefore, the data is retrospective non-clinical bench testing data, not from human subjects, and there is no country of origin for human data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • Not applicable. The "ground truth" in this context refers to the physical and functional characteristics of the devices (e.g., force required for insertion, pumping rate, burst pressure). These are objective measurements determined by laboratory equipment and procedures, not by expert interpretation. Therefore, experts were not used to establish a ground truth in the sense of clinical interpretations.

4. Adjudication Method for the Test Set

  • Not applicable. As the ground truth is based on objective physical measurements, there is no need for expert adjudication. The tests likely follow validated laboratory protocols.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study was done. This device is an Intra-Aortic Balloon Catheter, which is a physical medical device, not an AI or imaging diagnostic tool that would typically involve human readers or AI assistance in interpretation. The document explicitly states: "No clinical testing was performed by Boston Scientific Corporation/Cardiac Assist in support of this premarket notification."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a hardware medical device, not a software algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The "ground truth" for the non-clinical tests is established by objective physical measurements and engineering specifications. For example, burst pressure is measured until the balloon bursts, pumping rate until full inflation/deflation cannot be achieved, and insertion force is quantified using mechanical testing equipment. Comparison is made against the performance of legally marketed predicate devices, which serve as the reference standard for "substantially equivalent" performance.

8. The Sample Size for the Training Set

  • Not applicable. Since this is a hardware device and not an AI or machine learning model, there is no "training set."

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no training set for an AI/ML model, there is no ground truth established for it.

{0}------------------------------------------------

K912113

510(k) SUMMARY

Pursuant to Section 513(i)(3)(A) of the Federal Food, Drug and Cosmetic Act, Boston Scientific Corporation/Cardiac Assist (BSC/CA) is required to submit within this premarket notification either an " ... adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." BSC/CA chooses to submit a summary of information regarding safety and effectiveness.

GENERAL INFORMATION A.

Submitter's Name:Boston Scientific CorporationOne Boston Scientific PlaceNatick, MA 01760-1537
----------------------------------------------------------------------------------------------------------

Contact Person: Leo Basta Director, Regulatory Affairs and Clinical Research

Date of Preparation: 30 May 1997

B. DEVICE INFORMATION

Device Generic Name: Intra-Aortic Balloon Catheter

Device Trade Name:30 cc SUB-9 Nitinol
40 cc SUB-9 Nitinol
30 cc SUB-9 Stainless Steel
40 cc SUB-9 Stainless Steel

Percutaneous Intra-Aortic Balloon Catheter Classification Name:

C. PREDICATE DEVICE INFORMATION:

The following devices are referenced in this premarket notification as predicate devices for the 30 cc and 40 cc SUB-9 Nitinol and Stainless Steel, subject of this submission:

K963187: Modified Labeling of BSC/CA IAB's.

K954431: Modified Model 940 and Model 930 for use on BSC/CA 3001, Datascope Systems 90 and 97, St. Jude/Aries 700, Kontron KAAT and Bard H-8000.

K952221: 30 and 40 cc Sensation™, Model 940 and Model 930 for use on Bard H-8000.

K943919: 40 cc Sensation™ for use on Datascope System 97 and St. Jude/Aries 700.

{1}------------------------------------------------

K940298: Model 940 for use on BSC/CA 3001, Datascope Systems 90 and 97, Kontron KAAT and St. Jude/Aries 700.

K936232: Model 930 and 30 cc Sensation™ for use on BSC/CA 3001. Datascope Systems 90 and 97, Kontron KAAT and St. Jude/Aries 700.

FDA has concurred with the substantial equivalence of the above referenced premarket notifications. All of these devices are currently legally marketed.

PROPOSED DEVICE INFORMATION D.

This premarket notification proposes the following changes to the current legally marketed NICATH™, Sensation™ and Models 940 and 930:

  1. The catheter shaft outer diameter is 8.75 F for the proposed SUB-9 (30 cc and 40 cc volumes and Stainless Steel and Nitinol central lumens) as compared to 9 F for the predicate 30 cc and 40 cc NICATH™ and Models 940 and 930, and 9.5 F for the Sensation™,

The central lumen material of the proposed 30 cc and 40 cc SUB-9 2. Nitinol is the same as the predicate NICATH™, except that the central lumen is slightly smaller allowing a smaller folded balloon diameter.

The central lumen material of the proposed 30 cc and 40 cc SUB-9 အဲ Stainless Steel is the same as the predicate Sensation™, except that the central lumen is slightly smaller allowing a smaller folded balloon diameter.

  1. The smaller catheter shaft of the SUB-9 design (8.75 F vs. 9 F) allows the balloons to be folded to a smaller folded balloon diameter (0.121 inch vs. 0.129 inch) as compared to the predicate devices. This smaller folded balloon diameter allows the use of a 9 F introducer with the SUB-9 design as compared to a 10 F introducer for the predicate devices.

DEVICE DESCRIPTION ui

The proposed 30 cc and 40 cc SUB-9 Stainless Steel and Nitinol consists of a polyurethane blend symmetrical balloon at the distal end of a polyurethane-covered nylon shaft. The balloons are coated with a thin layer of silicone fluid. A Nitinol central lumen runs throughout the length of the 30 cc and 40 cc SUB-9 Nitinol catheters and terminates at the distal tip. A Stainless Steel central lumen runs throughout the length of the 30 cc and 40 cc SUB-9 Stainless Steel catheters and terminates at the distal Each central lumen may be used to pass the devices over their tip. guidewire. The balloon is supplied prewrapped for insertion utilizing either a hemostasis sheath (provided in each IAB insertion kit) or sheathless.

ட் INDICATIONS FOR USE

The indications for use for the proposed SUB-9 IAB's is identical to that of the currently marketed predicate devices. The indications are as follows:

  • · Refractory power failure.
  • · Cardiogenic shock.

{2}------------------------------------------------

  • · Unstable refractory angina.
  • · Impending or extending myocardial infarction (MI),

Hemodynamically significant mechanical complications secondary to acute MI:

  • · Ventricular septal defect.
  • · Mitral valve regurgitation.
  • · Papillary muscle rupture.
  • · Angiography/Angioplasty patients.
  • · Septal shock.

G. TECHNOLOGICAL CHARACTERISTICS

The 30 cc and 40 cc SUB-9 Stainless Steel and Nitinol design is identical to the predicate NICATHTM, uses the same balloons as the predicate NICATH™ and Models 940 and 930 and uses the same central lumen materials as the predicate NICATH™ and Sensation™. The slightly smaller catheter shaft of the SUB-9 design does not effect it's ability to perform equivalently to the predicate devices. Test data and information demonstrates that the use of the 30 cc and 40 cc SUB-9 Stainless Steel and Nitinol is substantially equivalent to the performance of the predicate devices on the BSC/CA 3001, Datascope Systems 90 and 97, St. Jude/Aries 700, Kontron KAAT and Bard H-8000.

ાનં. NONCLINICAL TESTS

  1. Sheathed Insertion Test:

The force required to insert the SUB-9 designs (30 and 40 cc and Stainless Steel and Nitinol) was demonstrated to be substantially equivalent to the force required to insert the predicate NICATH™ and Sensation™ designs.

    1. Sheathless Insertion Test:
      The tightest restriction that the SUB-9 design can pass through was found to be smaller than that required for the predicate BSC/CA IAB's. the Datascope 40 cc 9.5 F Percor-Stat and the Kontron/Arrow 40 cc 9 F IAB.

3. Maximum Pumping Rate Limit Test:

The maximum pumping rate limit, defined as the maximum pumping rate at which the balloon is able to inflate and deflate fully (greater than or equal to 90% of it's nominal volume), was measured for the 30 cc and 40 cc SUB-9. The maximum achievable pumping rate limits for the SUB-9 designs were found to be higher than that for the predicate NICATH™ and Models 940 and 930 on the BSC/CA 3001. Since the SUB-9 performed faster on the BSC/CA 3001 it will perform faster on any IABP it is labeled for use on.

{3}------------------------------------------------

    1. Reliability/Integrity Test:
      The reliability of the SUB-9 was found comparable to all BSC/CA predicate devices, all of which were able to cycle for at least 3.6 million cycles. The maximum pumping rate limit post-reliability was found identical to the prereliability rate for the 40 cc SUB-9. Inflation and deflation times postreliability were also found comparable to pre-reliability times.
    1. Aneurization and Burst Pressure Test:
      The aneurization and burst pressure tests demonstrated substantially equivalent aneurization and burst pressures as compared to the predicate devices. The SUB-9 balloons will not aneurize or burst when used on the IABP's labeled for use on since the IABP's are unable to provide the volume required to aneurize the balloons then continue on to burst the balloons. The peak output pressures of all the IABP's indicated for use on is below the pressures required to aneurize and burst the balloons.
    1. Transmembrane Pressure and Volume Measurement Test:
      Since the balloons used on the SUB-9 are the same as the predicate Models 940 and 930 and 30 and 40 cc NICATH™, they will have the same transmembrane pressure at the same inflated volume. The volume of the 30 and 40 cc SUB-9 were found substantially equivalent to the predicate Models 940 and 930, demonstrating equal transmembrane pressures.
    1. Trackability Test:
      The SUB-9 test samples (30 cc and 40 cc; Stainless Steel and Nitinol) were inserted over their guidewire, into position, and withdrawn to demonstrate that the catheters can perform with their insertion accessories. All the SUB-9, as well as predicate devices, were able to track their guidewire into position in the aorta without incident. There were no catheter kinks, no guidewire kinks, nor any problems encountered inserting or removing the catheters through their introducer.
    1. Kink Resistance Test:
      The mean kink radius of the SUB-9 Stainless Steel and Nitinol were demonstrated to be substantially equivalent to the predicate NICATH™, Sensation™ and Models 940 and 930.
    1. Initial Performance, Insertion/Removal Integrity Test:
      All of the 30 cc and 40 cc SUB-9 Nitinol and Stainless Steel test samples opened as they should per the Directions for Use. The intearity of the SUB-9 design post insertion and removal was demonstrated.

Additional nonclinical tests performed for previous premarket notifications and submitted as reference information within this submission include abrasion testing and biocompatibility testing.

{4}------------------------------------------------

CLINICAL TESTS 1.

testing was performed by Boston Scientific No clinical Corporation/Cardiac Assist in support of this premarket notification.

STERILIZATION AND PACKAGING J.

There are no changes to the packaging and sterilization of the SUB-9 IAB's as compared to the predicate devices. The catheters are placed in a plastic tray and sealed into a Tyvek/Mylar pouches and sterilized using 100% Ethylene Oxide gas. Ethylene oxide gas residuals and bacterial endotoxin levels are monitored for compliance with maximum release limits.

POTENTIAL COMPLICATIONS K.

Potential complications associated with the use of intra-aortic balloon catheters, in general, appear in the device Directions for Use and are reproduced below:

  • · Leg ischemia.
  • Femoral, aortic or illiac dissection.
  • · Arterial injury.
  • · Renal artery occlusion.
  • · Arterial rupture.
  • · Arterial perforation.
  • · Hypotension.
  • · Distal embolization.
  • Death.
  • · Vascular thrombosis.
  • · Short-term hemodynamic deterioration.
  • · Hemorrhage.
  • · Arteriovenous fistula formation.

ட் CONCLUSIONS

Based on the functional and performance data and information submitted in this premarket notification, Boston Scientific Corporation/Cardiac Assist believes that the 30 cc and 40 cc SUB-9 Nitinol and Stainless Steel are substantially equivalent to the predicate devices, NICATHTM, Sensation™ and Models 940 and 930 for use on the BSC/CA 3001, Datascope Systems 90 and 97. St. Jude/Aries 700, Kontron KAAT and Bard H-8000.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856

Mr. Leo Basta Director, Regulatory Affairs and Clinical Research Boston Scientific Corporation One Boston Scientific Place Natick, Massachusetts__01760-1537.____________________________________________________________________________________________________________________________________________

Re: K972113 Cardiac Assist Sub-9 Intra-Aortic Balloon Catheters Requlatory Class: III (Three) Product Code: 74 DSP Dated: June 3, 1997 Received: June 5, 1997

Dear Mr. Basta:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requiation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

SEP - 2 1997

{6}------------------------------------------------

Page 2 - Mr. Leo Basta

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

FROM : CSA

Page 1 of

K972113 510(k) Number lif known):

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

The BSC/CA IABC's are indicated for use in patients with the following conditions: * Refractory power failure.

    • Cardiogenic shock.
    • Unstable refractory angina.
    • Histodynamically significant mechanical complications secondary to acute MI:
        • Ventricular septal defect.
        • Mitral valve regurgitation.
        • Papillary muscle rupture.
    • Cardiac support for high risk general surgical and coronary angiography/ angioplasty patients.
    • Septic shock.

The intended use of the 30 cc and 40 cc SUB-9 remain identical to that of the currently marketed IABC's. All of the devices are intended to provide temporary circulatory ntireded that 3. The or in athrough controlled mechanical displacement of a volume of blood in the aorta.

The mechanical action of the IAB catheter therapy lowers the cardiac workload by two reans :

  • Systolic unloading, as noted by a reduction in the patient's systolic pressure, l . which provides reduced myocardial oxygen consumptions (MVO,).
    IPLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED!

Concurrance of CDRH, Office of Device Evaluation (ODE)

Thum Mye kn. BLL
(Division Sign Off)

Division of Cardiovascular, Respiratory, and Neurological Devices

510(k) Number K97211

Prescription Use (Per 21 CFR 801.109)

OR over-

Over-The-Counter Usa__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

{8}------------------------------------------------

K972113 510(k) Number:

Page 2 of 2

30 cc and 40 cc SUB-9 Device Name:

Indications for Use (continued):

    1. Diastolic augmentation which provides an increase in the mean ann aortic pressure and leads to an improvement in systemic and coronary arterial perfusion.
      Balloon pump therapy is achieved by inserting an intra-aortic balloon catheter into the descending thoracic aorta via the common femoral artery. Balloon inflation is timed to occur during diastole, beginning with the aortic valve closure. The balloon remains inflated until the onser of left ventricular ejection or systole, then rapidly deflates, reducing the aortic pressure, which in turn reduces the afterload.

§ 870.3535 Intra-aortic balloon and control system.

(a)
Identification. An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.(b)
Classification. (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.