K Number
K963187

Validate with FDA (Live)

Date Cleared
1996-11-04

(82 days)

Product Code
Regulation Number
870.3535
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The indications for use for the Cardiac Assist IAB's are identical to that of the currently marketed devices. The indications are as follows:

    • Refractory power failure.
    • Cardiogenic shock.
    • Unstable refractory angina.
    • Impending or extending myocardial infarction (MI).
    • Hemodynamically significant mechanical complications secondary to acute MI:
        • Ventricular septal defect.
        • Mitral valve regurgitation.
        • Papillary muscle rupture.
    • Angiography/angioplasty patients.
    • Septal shock.
Device Description

The Models 940 and 930, 30 cc and 40 cc Sensation™ and NICATH™ 40 cc and 30 cc IAB's consist of a polyurethane blend balloon at the distal end of a polyurethane-covered nylon shaft. The balloons are coated with a thin layer of Medical Grade silicone fluid. A central lumen runs throughout the length of the catheters and terminates at the distal tip. This central lumen may be used to pass the device over a guidewire. The balloon is supplied pre-wrapped for insertion.

AI/ML Overview

The manufacturer, Boston Scientific Corporation / Cardiac Assist (BSC/CA), is seeking to label its Intra-Aortic Balloon Catheters (IABs) – Models 940 and 930, 30 cc and 40 cc Sensation™, and NICATH™ 40 cc and 30 cc – for an optional sheathless method of insertion. The claim is that these devices are substantially equivalent to existing predicate devices (Datascope 9.5 F 40 cc PERCOR STAT-DL and Kontron 9 F 40 cc IAB for Sheathless Insertion, and for some tests, Arrow 9 F 40 cc).

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Test PerformedPredicate Device Performance / ReferenceBSC/CA Device PerformanceOutcome
Insertion Test (Ability to be inserted through a restriction without damage)Datascope and Arrow IABs required larger openings to pass without failure.Able to be inserted through a tighter restriction without failure compared to predicate devices.Meets / Exceeds
Kink Resistance Test (Smallest bend radius without failure)Datascope mean: 0.24 inch Arrow mean: 0.23 inchMean radius ranged from 0.124 to 0.54 inch.Substantially Equivalent
Trackability Test (Ability to track guidewire into position and be removed without damage)Datascope and Arrow IABs were able to track their guidewire into proper position without incident.All Cardiac Assist IABs were able to track their guidewire into proper position and were removed without incident. No kinks in catheter or guidewire.Substantially Equivalent
Dimensional Comparison (Difference between folded balloon diameter and catheter shaft diameter)Datascope: 0.009 inch Arrow: 0.029 inchRanged from 0.009 to 0.015 inch.Substantially Equivalent

2. Sample Size Used for the Test Set and Data Provenance

The provided summary does not specify the sample size (number of devices tested) for the nonclinical tests (Insertion, Kink Resistance, Trackability, Dimensional Comparison). The data provenance is nonclinical (laboratory bench testing) conducted by Boston Scientific Corporation / Cardiac Assist. No country of origin is specified for the data, but it can be assumed to be from the manufacturer's testing facilities. The data is prospective as it was generated specifically to support this premarket notification.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable. The ground truth for the nonclinical tests was established by direct measurement and observation of device performance under controlled laboratory conditions, not by expert consensus or interpretation.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods are typically used for clinical studies involving human interpretation or subjective assessments. The nonclinical tests described involve objective measurements of physical device characteristics and performance.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was conducted. This device is a medical device (Intra-Aortic Balloon Catheter) and not an AI-powered diagnostic or assistive technology for human readers. Therefore, an MRMC study comparing human reader performance with and without AI assistance is not relevant to this submission.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical medical instrument, not an algorithm. Therefore, "standalone" performance in the context of AI algorithms is not relevant.

7. The Type of Ground Truth Used

The ground truth for the nonclinical tests was based on objective measurements and direct observation of physical and mechanical performance under controlled laboratory conditions. For example, in the insertion test, the "ground truth" was whether the device could pass through a restriction without damage, measured mechanically. In the kink resistance test, it was the smallest radius without failure, measured directly.

8. The Sample Size for the Training Set

This information is not applicable. This is a traditional medical device, not an AI/ML-based device that requires a training set.

9. How the Ground Truth for the Training Set was Established

This information is not applicable as there is no training set for this device.

{0}------------------------------------------------

NOV - 4 1986

K963187

Image /page/0/Figure/2 description: The image shows a text box with the words "510(K) SUMMARY" written in bold, black letters. The text is centered within the box. The box itself is a light gray color.

Pursuant to Section 513(i)(3)(A) of the Federal Food, Drug and Cosmetic Act, Boston Scientific Corporation / Cardiac Assist (BSC/CA) is required to submit within this Premarket Notification either an "...adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." BSC/CA chooses to submit a summary of information regarding safety and effectiveness.

A. GENERAL INFORMATION

Submitter's Name:Boston Scientific CorporationOne Boston Scientific PlaceNatick, MA 01760-1537
Contact Person:Leo BastaDirector, Regulatory Affairs and Clinical Research
Date of Preparation:01 August 1996

B. DEVICE INFORMATION

Device Generic Name: Intra-Aortic Balloon Catheters

Device Trade Names: Model 940 Model 930 30 cc Sensation™ 40 cc Sensation™ NICATH™ 40 CC NICATH™ 30 cc

Classification Name: Percutaneous Intra-Aortic Balloon Catheter

C. PREDICATE DEVICE INFORMATION

The following devices are referenced in this premarket notification as predicate devices for the sheathless insertion of BSC/CA IAB's:

K910997: Datascope 9.5 F 40 cc PERCOR STAT-DL.

K892222: Kontron 9 F 40 cc IAB for Sheathless Insertion.

These devices are currently legally marketed for sheathless insertion.

{1}------------------------------------------------

PROPOSED DEVICES INFORMATION D.

This premarket notification proposes to allow the labeling of the BSC/CA Models 940 and 930, 30 cc and 40 cc Sensation™ and NICATH™ 40 cc and 30 cc IAB's for an optional sheathless method of insertion. The sheathless insertion technique is proposed to be offerred in the Directions for Use along with the current instructions for a sheathed insertion.

E. DEVICE DESCRIPTIONS

The Models 940 and 930, 30 cc and 40 cc Sensation™ and NICATH™ 40 cc and 30 cc IAB's consist of a polyurethane blend balloon at the distal end of a polyurethane-covered nylon shaft. The balloons are coated with a thin layer of Medical Grade silicone fluid. A central lumen runs throughout the length of the catheters and terminates at the distal tip. This central lumen may be used to pass the device over a guidewire. The balloon is supplied pre-wrapped for insertion.

F. INDICATIONS FOR USE

The indications for use for the Cardiac Assist IAB's are identical to that of the currently marketed devices. The indications are as follows:

    • Refractory power failure.
    • Cardiogenic shock.
    • Unstable refractory angina.
    • Impending or extending myocardial infarction (MI).
    • Hemodynamically significant mechanical complications secondary to acute MI:
        • Ventricular septal defect.
        • Mitral valve regurgitation.
        • Papillary muscle rupture.
    • Angiography/angioplasty patients.
    • Septal shock.

G. TECHNOLOGICAL CHARACTERISTICS

The Cardiac Assist IAB's proposed to be labeled for optional sheathless insertion are Identical to the current legally marketed IAB's. Test data and information demonstrates that the use of the Cardiac Assist IAB's is substantially equivalent to the performance of the predicate devices (Datascope 9.5 F 40 cc PERCOR STAT-DL and Arrow 9 F 40 cc) when tested for sheathless insertion.

H. NONCLINICAL TESTS

The following tests were performed:

    1. Insertion Test:
      A test was designed and manufactured with an adjustable restriction to simulate a sheathless insertion. Cardiac Assist. Datascope and Arrow IAB's were inserted through the fixture at smaller and smaller increments to ascertain the tightest restriction the device could be passed through without damaging the IAB or its guidewire.

{2}------------------------------------------------

In each case the Cardiac Assist IAB's were able to be inserted through a tighter restriction without failure as compared to the Datascope and Arrow IAB's. The competitor catheters required larger openings to be able to pass through without failure.

  1. Kink Resistance Test:

The Cardiac Assist. Datascope and Arrow IAB's were tested for their resistance to kinking by bending the catheter samples around a radius and measuring the smallest step on the fixture that the catheter was able to be bent around without failure. The Cardiac Assist mean radius ranged from 0.124 to 0.54 inch while the Datascope mean was 0.24 inch and the Arrow mean was 0.23 inch.

The Cardiac Assist IAB's have been demonstrated to be substantially equivalent in kink resistance to the Datascope and Arrow IAB's.

  1. Trackability Test:

The Cardiac Assist, Datascope and Arrow IAB's were inserted over their guidewire into position and removed without damaging the balloon or guidewire. All of the IAB's were able to track their quidewire into proper position in the aorta without incident. There were no catheter kinks, guidewire kinks, and no problems or difficulties encountered during insertion and removal demonstrating substantial equivalence of the Cardiac Assist IAB's to Datascope and Arrow.

    1. Dimensional Comparison:
      A dimensional comparison was made between the Cardiac Assist and Datascope and Arrow IAB's catheter shaft diameter, folded balloon diameter, and the difference between the folded balloon diameter and catheter shaft diamter. Results indicated that the difference between the folded balloon diameter and catheter shaft diamter of the Cardiac Assist IAB's ranges from 0.009 to 0.015 inch while the Datascope is 0.009 inch and the Arrow is 0.029 inch.

The difference between the folded balloon diameter and catheter shaft diameter of the Cardiac Assist IAB's is substantially equivalent to that of the Datascope and Arrow IAB's.

CLINICAL TESTS 1.

Based on conversations with FDA in April 1996, it was agreed that no formal clinical data was required to support this premarket notification.

J. STERILIZATION AND PACKAGING

There are no changes to the packaging and sterilization of the Cardiac Assist IAB's. The catheters are placed in plastic trays and sealed into Tyvek/Mylar pouches and are sterilized using ethylene oxide gas. Ethylene oxide gas residuals and bacterial endotoxin levels are monitored for compliance with maximum release limits.

{3}------------------------------------------------

POTENTIAL COMPLICATIONS K.

Potential complications associated with the use of intra-aortic balloon catheters in general appear in the devices' Directions for Use and are reproduced below:

    • Leg ischemia.
    • Femoral, aortic or illiac dissection.
    • Arterial injury.
    • Renal artery occlusion.
    • Arterial rupture.
    • Hypotension.
    • Distal embolization.
    • Death.
    • Vascular thrombosis.
    • Short-term hemodynamic deterioration.
    • Hemorrhage.
    • Arteriovenous fistula formation.

لــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ CONCLUSIONS

Based on the performance data and information submitted in this premarket notification, Boston Scientific Corporation / Cardiac Assist believes that the Cardiac Assist Models 940 and 930, 30 cc and 40 cc Sensation™ and NICATH™ 40 cc and 30 cc IAB's are substantially equivalent to the predicate devices, Datascope 9.5 F 40 cc PERCOR STAT-DL and Arrow 9 F 40 cc IAB's, with regard to sheathless insertion of the devices.

§ 870.3535 Intra-aortic balloon and control system.

(a)
Identification. An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.(b)
Classification. (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.