K Number
K971706

Validate with FDA (Live)

Date Cleared
1997-07-21

(74 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nobel Biocare's TiAdapt™ Abutment System is a set of screw retained preparable abutments that are secured to an endosseous implant and are intended to function as an anchor to which prosthetic devices, such as artificial teeth, may be attached using dental cement to restore a patient's chewing function.

Device Description

The Nobel Biocare TiAdapt™ Abutment System is comprised of preparable titanium abutments which are secured to the Brånemark System® dental implants with a gold abutment screw. This system functions as an anchor to which a single crown can be cemented. When more than one preparable abutment is used, a multi-unit bridge can be cemented to the abutments. The TiAdapt Abutment System also includes a counter torque device and adaptors, temporary cylinders, impression copings and machine Uni-grip screwdrivers.

AI/ML Overview

This document is a 510(k) Pre-market Notification for the Nobel Biocare TiAdapt™ Abutment System. It outlines the safety and effectiveness of the device by comparing it to predicate products. However, it does NOT contain information on acceptance criteria or a study proving the device meets those criteria, nor does it provide details about any specific performance studies with quantitative results.

The document primarily focuses on:

  • Device Description: The components of the TiAdapt™ Abutment System.
  • Intended Use: Functioning as an anchor for prosthetic devices to restore chewing function.
  • Predicate Products: Listing similar Nobel Biocare components already on the market.
  • Technological Characteristics: Stating that the characteristics are "identical" to predicate products.
  • Summary Of Safety And Effectiveness Concerns: Discussing potential problems encountered with similar abutment systems (e.g., screw fractures, loosening, improper seating). This section lists potential problems as a general safety overview, not as a study proving a specific performance metric was met or evaluating the TiAdapt's performance against these concerns.
  • FDA Clearance: The letter from the FDA affirming "substantial equivalence" to legally marketed predicate devices.

Therefore, I cannot provide a table of acceptance criteria or details of a study based on the provided text, as this information is not present in the document. The document describes a regulatory submission for pre-market notification, which establishes substantial equivalence, but not a performance study with specific acceptance criteria and results.

{0}------------------------------------------------

JUL 21 1997

Pre-market Notification Page -3-�������������������������������

510(k) Summary of Safety and Effectiveness VII.

A. Name and Address

This Summary of Safety and Effectiveness is being submitted by Nobel Biocare USA, Inc., 777 Oakmont Lane, Suite 100, Westmont, IL 60559. The telephone number is: (630) 654-9100, extension 2002, and the contact person will be Betsy Brown, the Vice President, Regulatory Affairs.

B. Name of the Device

This device system consists of various preparable titanium abutments, abutment screws, a counter torque device with handle, adaptors, machine Uni-grip screwdrivers with autoclavable box, temporary cylinders, temporary tubes, laboratory screws and impression copings.

C. The Predicate Product

The predicate products used in this Pre-market Notification are similar components marketed by Nobel Biocare including the CeraOne Abutments (K910611), Ceramic Abutment (K913255), CeraOne™ Abutment Screw (K961737), Guide Pin (K911969), Abutment Lab-Screw (K925769), CerAdapt™, Machine Counter Torque Device (K953775), Estheticone™ Temporary Cylinder (K925766), Miruscone Temporary Cylinder (K962403), Temporary Tube (K926766), Titanium Impression Coping (K952448), Plastic Impression Coping (K910611), CeraOne™ Screwdrivers (K922805) and Fixture Guide Set Tray (K944963).

D. Description of Device

The Nobel Biocare TiAdapt™ Abutment System is comprised of preparable titanium abutments which are secured to the Brånemark System® dental implants with a gold abutment screw. This system functions as an anchor to which a single crown can be cemented. When more than one preparable abutment is used, a multi-unit bridge can be cemented to the abutments. The TiAdapt Abutment System also includes a counter torque device and adaptors, temporary cylinders, impression copings and machine Uni-grip screwdrivers.

{1}------------------------------------------------

Pre-market Notification Page -4-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

E. Intended Use of the Device

Nobel Biocare's TiAdapt™ Abutment System is a set of screw retained preparable abutments that are secured to an endosseous implant and are intended to function as an anchor to which prosthetic devices, such as artificial teeth, may be attached using dental cement to restore a patient's chewing function.

F. Comparison of Technological Characteristics

The technological characteristics between the components of the TiAdapt Abutment System and the corresponding predicate products found in the Brånemark System® are identical.

Summary Of Safety And Effectiveness Concerns

Safety and effectiveness problems that have been encountered with similar Abutment Systems used with endosseous implants that are currently being marketed include:

    1. occasional fractures of the screw attaching the abutment to the fixtures usually due to functional overload from masticastory forces:
    1. screws working loose (usually because torque force below recommended values was applied when the abutment was attached to the fixture) may lead to the formation of granulation tissue at the level of the fixture and abutment connection which may, in turn, result in infection:
    1. improper initial seating of abutments resulting in gingival inflammation and fistulae formation, both conditions resolve when proper seating of the abutment is accomplished.
    1. occasional fracture of the abutment screw which is usually caused by poorly designed and/or fabricated restorations that creates overloads or cause's metal fatigue.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem featuring a stylized depiction of an eagle or bird-like figure with outstretched wings. The emblem is rendered in a blue color, contrasting with the white background of the seal.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP 1 7 2010

Ms. Phuong Nguyen Regulatory Affairs Manager Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Lind, California 92887

Re: K971706

Trade/Device Name: TiAdapt Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: July 28, 2010 Received: July 29, 2010

Dear Mr. Nguyen:

This letter corrects our substantially equivalent letter of July 29, 2010.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Nguyen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Supa Viner

Anthony D. Watson, B.S., M.S., M.B.A.
Director
Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: TiAdapt™ Abutment System_

Indications For Use:

Nobel Biocare's TiAdapt Abutment System is a set of screw retained preparable abutments which are secured to an endosseous implant and is intended to function as an anchor to which prosthetic devices, such as artificial teeth, may be attached using dental cement to restore a patient's chewing function.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)Field W. Shippen
Division of Dental, Infection Control,
and General Hospital Devices
510(k) NumberK971706

Prescription Use. (Per 21 CFR 801.109)

OR

Over-The-Counter Use_

(Optional Format 1-2-96)

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)