K Number
K970725

Validate with FDA (Live)

Manufacturer
Date Cleared
1997-05-09

(70 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Opti-Plast® XL 5.5 F PTA Catheter is recommended for use in the Percutaneous Transluminal Angioplasty of the Iliac and Femoral vessels. This catheter is not for use in coronary arteries. The device is indicated for single use only and indicated not to be resterilized nor reused.

Device Description

The Vas-Cath Opti-Plast® XL 5.5 F PTA Catheter is a dual lumen catheter with a balloon mounted on its distal end. One lumen accomodates the insertion guidewire and the other provides a channel for inflation and deflation of the balloon with contrast media. The catheters were designed to be used in conjuction with a 0.035 inch diameter guidewire. There are two radiopaque marker bands secured on the catheter body to indicate the position of the balloon within the vasculature. The balloon inflates to the labelled diameter and length at a 5 atmosphere operating pressure.

AI/ML Overview

Here's an analysis of the provided text regarding the Opti-Plast® XL 5.5F PTA Catheter, addressing the requested information:

Analysis of Acceptance Criteria and Study for Opti-Plast® XL 5.5F PTA Catheter

Based on the provided 510(k) summary, the device is a medical catheter and the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are related to its physical and functional performance, demonstrating substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria here are implicitly defined by the characteristics of the predicate device (Opti-Plast® PTA Catheter K933483) and industry standards for such devices. The "reported device performance" refers to the results of the bench testing.

FeatureAcceptance Criteria (Predicate Device K933483)Reported Device Performance (Opti-Plast® XL 5.5F PTA Catheter)
Material/Design:
Balloon MaterialNylonNylon
Catheter Shaft MaterialNylonNylon
RadiopaqueYesYes
Shaft ConfigurationDouble lumenDouble lumen
Marker BandsYes, TantalumYes, Tantalum
Tip Length5 mm5 mm
Tip Inner Diameter0.037 inch0.037 inch
Guidewire Capability0.035 inch diameter0.035 inch diameter
Functional/Dimensional:
Inflated Balloon Diameters10 mm (closest size for comparison)12 mm
Inflated Balloon Length4 cm4 cm
Recommended Operating Pressure6 atmospheres (closest size for comparison)5 atmospheres
Rated Burst Pressure8 atmospheres8 atmospheres
Shaft Diameter5.5 French5.5 French
Shaft Length75, 100, 120 cm75, 100, 120 cm
Tested Characteristics (Bench Testing indicates substantial equivalence):
Balloon Burst StrengthPerformance comparable to K933483Met standards, substantially equivalent to K933483
Balloon DistensibilityPerformance comparable to K933483Met standards, substantially equivalent to K933483
Balloon Inflation/Deflation TimePerformance comparable to K933483Met standards, substantially equivalent to K933483
Balloon FatiguePerformance comparable to K933483Met standards, substantially equivalent to K933483
Joints and Material TestPerformance comparable to K933483Met standards, substantially equivalent to K933483
Catheter Shaft and Balloon ProfilePerformance comparable to K933483Met standards, substantially equivalent to K933483
Introducer Sheath CompatibilityPerformance comparable to K933483Met standards, substantially equivalent to K933483
Catheter Flow/Maximum PressurePerformance comparable to K933483Met standards, substantially equivalent to K933483
Trackability/DeflatabilityPerformance comparable to K933483Met standards, substantially equivalent to K933483
Tip TorquePerformance comparable to K933483Met standards, substantially equivalent to K933483

Note: For the dimensional characteristics (e.g., balloon diameter, operating pressure), the Opti-Plast® XL is different from the specific 10mm predicate balloon but is being compared for its safety and effectiveness, likely falling within acceptable variations for its intended use or being equivalent to the Medi-tech® XXL™ Balloon Dilatation Catheter (K952063) for its indications. The key assertion is that the overall performance based on the bench tests is substantially equivalent.

2. Sample Size for the Test Set and Data Provenance

The provided text only explicitly mentions "Bench testing." This implies the "test set" consisted of physical samples of the Opti-Plast® XL 5.5F PTA Catheter undergoing various laboratory tests.

  • Sample Size: The document does not specify the exact number of catheters tested for each characteristic.
  • Data Provenance: The data provenance is laboratory-based bench testing of the manufactured device. It is not related to patient data or human subjects.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This question is not applicable to this type of device submission. For a medical device like a catheter, "ground truth" is established through engineering specifications, material standards, and performance metrics defined by regulatory bodies and industry best practices. There are no human experts determining a "ground truth" in the way one would for diagnostic imaging.

4. Adjudication Method for the Test Set

This question is not applicable as the "test set" involves physical device testing against predefined engineering and performance standards, not human interpretation or consensus. No adjudication method is described or implied.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of outputs (e.g., medical images, AI algorithms). The Opti-Plast® XL 5.5F PTA Catheter is an interventional device; its performance is assessed through bench testing, not human reading of cases.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone (algorithm-only) performance study was not done. This question is relevant for AI or software as a medical device (SaMD). The Opti-Plast® XL 5.5F PTA Catheter is a physical medical device.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is based on:

  • Engineering specifications and design parameters: The device is designed to meet specific dimensions, material properties, and functional characteristics.
  • Industry standards: Performance characteristics (e.g., burst pressure, fatigue) are likely evaluated against recognized ASTM or ISO standards for angioplasty catheters.
  • Predicate device performance: The performance of the Opti-Plast® XL 5.5F PTA Catheter in bench tests is compared against the established performance of the legally marketed predicate device (Opti-Plast® PTA Catheter K933483) to demonstrate substantial equivalence.

8. The Sample Size for the Training Set

This question is not applicable. There is no "training set" in the context of a physical medical device submission like this. Training sets are used for machine learning algorithms.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as no training set exists for this type of device.

{0}------------------------------------------------

K970725

MAY -9 1997

510(k) SUMMARY STATEMENT of SAFETY AND EFFECTIVENESS

FOR THE

OPTI-PLAST® XL 5.5F PTA CATHETER

A.Submitter Information:Vas-Cath Incorporated
Address:2380 Tedlo Street
Mississauga, Ontario L5Z 3V3
Canada
Telephone Number:(905) 281-7745
Fax Number:(905) 848-6638
Contact Person:Dolores McGirr
Date of Preparation:January 24, 1997 :
B.Device Name:Opti-Plast® XL 5.5 F PTA Catheter
XL (Xtra-Large)
Common/Usual Name:Peripheral Angioplasty Catheter
Classification Name:Surgical Vessel Dilator
C.Predicate Device :Opti-Plast® PTA Catheter (K933483)
Trade Name:Opti-Plast
Predicate Device for the Indications for Use: Medi-tech® XXL™ Balloon Dilatation

Catheter (K952063)

D. · Device Description:

The Vas-Cath Opti-Plast® XL 5.5 F PTA Catheter is a dual lumen catheter with a balloon mounted on its distal end. One lumen accomodates the insertion guidewire and the other provides a channel for inflation and deflation of the balloon with contrast media. The catheters were designed to be used in conjuction with a 0.035 inch diameter guidewire. There are two radiopaque marker bands secured on the catheter body to indicate the position of the balloon within the vasculature. The balloon inflates to the labelled diameter and length at a 5 atmosphere operating pressure.

E. Indicated Use:

The Opti-Plast® XL 5.5 F PTA Catheter is recommended for use in the Percutaneous Transluminal Angioplasty of the Iliac and Femoral vessels. This catheter is not for use in coronary arteries. The device is indicated for single use only and indicated not to be resterilized nor reused.

  • F. Technological Characteristics Summary: Summarized in Table 6.1.
    Opti-Plast® XL PTA Catheter 510(k) Summary Page(o

2380 Tedlo Street Mississauga, Ontario Canada, LSA 3V3

Tel. (905) 848-5800 Fax (905) 848-6638

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "vascoath." in a stylized font. Below the word is the word "INCORPORATED" in a smaller, sans-serif font. The text is black on a white background. The dot after the word "vascoath" is larger than the letters in the word "INCORPORATED."

FeaturesProposed Vas-CathOpti-Plast® XL 5.5 FPTA CatheterVas-Cath Opti-Plast®PTA Catheter(K933483)Medi-tech" XXLTMBalloon DilatationCatheter (K952063)
Indicated UsePTA of Iliac andFemoral vesselsPTA of vessels of theperipheral vascularsystemPTA of the Iliac andFemoral arteries in theperipheral vasculature
Balloon MaterialNylonNylonNA
Inflated BalloonDiameters12 mm10 mm*12, 14 mm
Inflated BalloonLength4 cm4 cm*2, 4, 6 cm
RecommendedOperating Pressure5 atmospheres6 atmospheres*NA
Rated Burst Pressure8 atmospheres8 atmospheres*8 atmospheres
Catheter Shaft MaterialNylonNylonNA
RadiopaqueYesYesNA
Shaft Diameter5.5 French5.5 French5.8 French
Shaft Length75, 100, 120 cm75, 100, 120 cm75, 120cm
Shaft configurationDouble lumenDouble lumenDouble lumen
Marker BandsYes, TantalumYes, TantalumYes
Tip Length5 mm5 mm5 mm
Tip Inner Diameter0.037 inch0.037 inch0.037 inch
Guidewire Capability0.035 inch diameter0.035 inch diameter0.035 inch diameter

TABLE 6.1 : Technological Characteristics Summary

NA- Not available from promotional literature

  • Note that there are many other sizes of balloons in this product line. However, the comparison here is made with the 10 mm x 4 cm balloon since this is the closest size to the proposed catheter.

G. Performance Data

Bench testing was conducted for the following characteristics: balloon burst strength, balloon distensibility, balloon inflation/deflation time, balloon fatigue, joints and material test, catheter shaft and balloon profile, introducer sheath compatibility, catheter flow/maximum pressure, trackability/ deflatability, and tip torque. The test results indicate that the proposed Opti-Plast XL 5.5 F PTA Catheter is substantially equivalent to the current Opti-Plast® PTA Catheter marketed under 510(k) concurrence K933483. There are no new safety or effectiveness issues.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).