Intended Use

The Thyroid Calibrator is intended to be used on the Dimension® RxL clinical chemistry system with the heterogeneous immunoassay module to calibrate the FT4 and TSH methods.

Device Description

The Thyroid Calibrator is a liquid bovine albumin based product. The Level 1 calibrator contains no detectable thyroxine or TSH. Levels 2 through 5 contain thyroxine and human TSH. The kit consists of ten vials; two at each of five levels.

AI/ML Overview

The provided text describes a Thyroid Calibrator, a liquid bovine albumin-based product designed for calibrating FT4 and TSH methods on the Dimension® RxL clinical chemistry system. The document is a "Summary of Safety and Effectiveness Information" for a 510(k) submission, indicating a comparison to a predicate device for substantial equivalence.

However, the document does not contain information related to acceptance criteria or a study proving device performance in the way requested by the prompt. This document is a summary of safety and effectiveness information and focuses on comparing the new device to a predicate device to establish substantial equivalence, not on detailed performance validation studies with acceptance criteria, sample sizes for test sets, ground truth establishment, or clinical effectiveness studies.

Specifically, the requested information elements are not available in the provided text:

  1. A table of acceptance criteria and the reported device performance: This information is not present. The document compares features of the new calibrator to predicate calibrators but does not define or report performance against specific acceptance criteria for accuracy, precision, or other analytical characteristics.
  2. Sample sized used for the test set and the data provenance: Not mentioned.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
  4. Adjudication method: Not mentioned.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable, as this is a calibrator, not a diagnostic imaging or interpretive device that would typically involve human readers.
  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable for a calibrator.
  7. The type of ground truth used: Not mentioned.
  8. The sample size for the training set: Not mentioned.
  9. How the ground truth for the training set was established: Not mentioned.

The document's purpose is to argue for substantial equivalence based on intended use, analytes, matrix, form, volume, and levels, not to report on detailed performance studies against pre-defined acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "DADE" in large, bold, black letters. The letters are sans-serif and evenly spaced. The word is presented in a straightforward, unadorned manner, with a clear and legible typeface.

Image /page/0/Picture/1 description: The image shows a sequence of numbers, specifically 'K970389'. The numbers are written in a bold, handwritten style, with thick strokes. Below the numbers, there is a dark, rectangular block that obscures any underlying content.

FEB 1 8 1997

Interoffice Memorandum

Summary of Safety and Effectiveness Information

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter's Name:Cathy P. CraftDade International Inc.P.O. Box 6101Newark, DE 19714-6101
Date of Preparation:1/29/97
Name of Product:Thyroid Calibrator
FDA Classification Name:Calibrator
Predicate Device:aca® plus FT4 Calibrator and aca® plusTSH Calibrator

Device Description: The Thyroid Calibrator is a liquid bovine albumin based product. The Level 1 calibrator contains no detectable thyroxine or TSH. Levels 2 through 5 contain thyroxine and human TSH. The kit consists of ten vials; two at each of five levels.

Intended use: The Thyroid Calibrator is intended to be used on the Dimension® RxL clinical chemistry system with the heterogeneous immunoassay module to calibrate the FT4 and TSH methods.

4

0000000

{1}------------------------------------------------

Comparison to Predicate Device:

ItemThyroid Calibratoraca® plusFT4 Calibratoraca® plusTSH Calibrator
Intended UseCalibratorCalibratorCalibrator
Analytesthyroxine, TSHthyroxineTSH
Matrixbovine albuminhuman serumhorse serum
Formliquidlyophilizedlyophilized
Volume2.0 mL per vial3.0 mL per vial,reconstituted5.0 mL per vial,reconstituted
Levels5 levels3 levels3 levels

Comments on Substantial

Equivalence: Both the Thyroid Calibrator and the aca8 plus FT4 Calibrator are intended to be used as callbrators for free thyroxine methods. Both the Thyroid Calibrator and the aca® plus TSH Calibrator are intended to be used as calibrators for thyroid stimulating hormone methods.

Conclusion: The Thyroid Calibrator is substantially equivalent to the aca® plus FT4 Calibrator and the acad plus TSH Calibrator for their respective analytes based on the comparison discussed above.

$\pi, \mathcal{R}e|x$

Cathy P. Craft Regulatory Affairs and Compliance Manager Date: 1/29/97

000009

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.