K Number
K962572

Validate with FDA (Live)

Date Cleared
1996-12-17

(169 days)

Product Code
Regulation Number
870.1330
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SCIMED ChoICE PT Vision PTCA Guide Wires are intended to facilitate the placement and exchange of balloon dilatation catheters or other therapeutic devices during PTCA or other intravascular interventional procedures. The ChoICE PT Vision PTCA Guide Wires are not intended for use in the cerebral vasculature. The devices are provided sterile and intended for one procedure only.

Device Description

The ChoICE PT Vision PTCA Guide Wires are steerable guide wires available in a nominal diameter of 0.014 inches and nominal lengths of 182 and 300 centimeters. The available tip flexibilities will be Floppy, Intermediate, Extra Support, Standard, Super Support, and Super Support II. The distal two centimeters of all models are straight shapeable or available in a pre-formed J-Tip.

The tapered, distal stainless steel core wire is surrounded by a polyurethane sleeve and is coated with ICE™ Hydrophilic Coating. The Floppy, Intermediate, Extra Support, and Standard guide wires will have the distal 38 centimeters surrounded by the polyurethane sleeve and are available in radiopaque lengths of 2 and 38 cm. The Super Support and Super Support II guide wires will have the distal 10 centimeters surrounded by the polyurethane sleeve, and are available in radiopaque lengths of 2 and 10 cm. The proximal wire is coated with polytetrafluoroethylene (PTFE)

AI/ML Overview

The provided text describes a 510(k) submission for the SCIMED ChoICE PT Vision PTCA Guide Wires. This submission focuses on establishing substantial equivalence to previously marketed devices rather than presenting a study proving a device meets specific acceptance criteria in the context of AI/ML or diagnostic performance. Therefore, many of the requested categories for a study proving device performance against acceptance criteria are not applicable.

Here's an analysis based on the provided text, highlighting what is available and what is not:

1. A table of acceptance criteria and the reported device performance

The document states: "Test results verified that the ChoICE PT Vision PTCA Guide Wires met all of the minimum requirements and are adequate for their intended use." However, it does not provide specific numerical acceptance criteria or the reported performance values for each test. Instead, it lists the types of tests performed.

Acceptance Criteria CategoryReported Device Performance
Tip TensileNot specified, but "met all minimum requirements"
Tip TorsionNot specified, but "met all minimum requirements"
Combined LoadNot specified, but "met all minimum requirements"
Tip FlexibilityNot specified, but "met all minimum requirements"
J-Tip Curve RetentionNot specified, but "met all minimum requirements"
Polyurethane Shear StrengthNot specified, but "met all minimum requirements"
Torque ResponseNot specified, but "met all minimum requirements"
PTCA Catheter CompatibilityNot specified, but "met all minimum requirements"

2. Sample size used for the test set and the data provenance

The document does not specify the sample size for the "test set" in terms of the number of guide wires tested for each non-clinical test. The data provenance is "in vitro testing and evaluation" conducted by SCIMED Life Systems, Inc. The country of origin of the data is not explicitly stated but can be inferred to be the USA, given the company's address (Maple Grove, Minnesota).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as this is a submission for a physical medical device (guide wire) and the non-clinical tests described are mechanical/physical performance tests, not diagnostic or AI/ML performance tests requiring expert interpretation of results to establish ground truth.

4. Adjudication method for the test set

This information is not applicable for the same reason as point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This is a physical medical device (guide wire), not an AI/ML diagnostic device, and no human reader studies are relevant or mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable as it's a physical device, not an algorithm.

7. The type of ground truth used

The "ground truth" here would be the established engineering/design specifications and performance thresholds for the physical/mechanical properties of the guide wire. The text indicates that the device "met all of the minimum requirements," implying comparison to these specifications. It is based on engineering and physical measurements rather than expert consensus, pathology, or outcomes data in a clinical sense.

8. The sample size for the training set

This information is not applicable. There is no "training set" in the context of a physical guide wire's non-clinical performance evaluation as described.

9. How the ground truth for the training set was established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

K962572

510(k) Notification SCIMED ChoICE PT Vision PTCA Guide Wires

Image /page/0/Picture/2 description: The image shows the logo for SCIMED, which is a medical device company. The logo consists of the word "SCIMED" in bold, black letters, with a partial circle to the left of the "S". Below the logo is the text "Boston Scientific Corporation" in a smaller font size. The logo is simple and professional, and it is likely used on the company's products and marketing materials.

SCIMED Life Systems, Inc. One SCIMED Place Maple Grove, MN 55311-1566 612.494.1700

DEC 1 7 1996

SECTION THREE-- SUMMARY OF SAFETY AND EFFECTIVENESS

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS (Pursuant To Section 12 of the SAFE MEDICAL DEVICES ACT of 1990) Prepared June 28, 1996

I. General Provisions:

Submitter's Name and Address:

SCIMED Life Systems, Inc. One SCIMED Place Maple Grove, Minnesota 55311

Contact Person:

Connie J. Del Toro (612) 494-2656

Classification Name:

Catheter Guide Wire

PTCA Guide Wire

Common or Usual Name:

Proprietary Name:

SCIMED ChoICE PT Vision PTCA Guide Wires

Name of Predicate Devices: II.

510(k) Reference No.Predicate Device
K943192 & K961015SCIMED ChoICE PTCA Guide Wire
K945129 & K961015SCIMED ChoICE Plus PTCA Guide Wire
K950113 & K961015SCIMED ChoICE Exchange PTCA Guide Wire
K941895 & K960657SCIMED ChoICE PT PTCA Guide Wire
K950216 & K960657SCIMED ChoICE PT Plus PTCA Guide Wire
K950266 & K960657SCIMED CholCE PT Exchange PTCA Guide Wire
K960563SCIMED Sceptor and Sceptor Exchange PTCA Guide Wire(Alternative Black PTFE Coating)

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for Scimed, a Boston Scientific Corporation. The logo is in black and white, with the word "SCIMED" in large, bold letters. Below the word "SCIMED" is the text "Boston Scientific Corporation" in a smaller font. The logo is simple and professional, and it is likely used on the company's products and marketing materials.

SCIMED Life Systems, Inc One SCIMED Place Maple Grove, MN 55311-1566 612.494.1700

III. Device Description:

The ChoICE PT Vision PTCA Guide Wires are steerable guide wires available in a nominal diameter of 0.014 inches and nominal lengths of 182 and 300 centimeters. The available tip flexibilities will be Floppy, Intermediate, Extra Support, Standard, Super Support, and Super Support II. The distal two centimeters of all models are straight shapeable or available in a pre-formed J-Tip.

The tapered, distal stainless steel core wire is surrounded by a polyurethane sleeve and is coated with ICE™ Hydrophilic Coating. The Floppy, Intermediate, Extra Support, and Standard guide wires will have the distal 38 centimeters surrounded by the polyurethane sleeve and are available in radiopaque lengths of 2 and 38 cm. The Super Support and Super Support II guide wires will have the distal 10 centimeters surrounded by the polyurethane sleeve, and are available in radiopaque lengths of 2 and 10 cm. The proximal wire is coated with polytetrafluoroethylene (PTFE)

IV. Intended Use:

The SCIMED ChoICE PT Vision PTCA Guide Wires are intended to facilitate the placement and exchange of balloon dilatation catheters or other therapeutic devices during PTCA or other intravascular interventional procedures. The ChoICE PT Vision PTCA Guide Wires are not intended for use in the cerebral vasculature. The devices are provided sterile and intended for one procedure only.

V. Summary of Technological Characteristics:

The ChoICE PT Vision PTCA Guide Wires utilize the same materials and methods of construction as the currently marketed ChoICE and ChoICE PT family of Guide Wires.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for SCIMED, which is a company that is part of Boston Scientific Corporation. The logo is in black and white and features the word "SCIMED" in a bold, sans-serif font. The words "Boston Scientific Corporation" are written in a smaller font below the logo. The logo is simple and professional, and it is likely used on the company's website, marketing materials, and products.

SCIMED Life Systems, Inc. One SCIMED Place Maple Grove, MN 55311-1566 612·494·1700

Non-Clinical Test Summary: VI.

Testing and evaluation of the guide wires included tip tensile, tip torsion, combined load, and tip flexibility. Additionally, J-Tip curve retention, polyurethane shear strength, torque response, and PTCA catheter compatibility were performed. Test results verified that the ChoICE PT Vision PTCA Guide Wires met all of the minimum requirements and are adequate for their intended use. The ChoICE PT Vision PTCA Guide Wires are considered to be substantially equivalent to guide wires currently marketed by SCIMED (ChoICE, and ChoICE PT Guide Wires) based on a comparison of intended use, design and the results of in vitro testing and evaluation.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.