Intended Use

The proposed test reagent (N Latex CRP mono Reagent) is an in vitro diagnostic reagent intended to be used together with the Behring Nephelometer Systems in the quantitative determination of C-reactive protein in human serum.

Device Description

The proposed test reagent (N Latex CRP mono Reagent) is an in vitro diagnostic reagent intended to be used together with the Behring Nephelometer Systems in the quantitative defermination of C-reactive protein in human serum.

In the proposed product polystyrene particles coated with mouse monoclonal Y antibodies to C-reactive protein are agglutinated when mixed with samples containing C-reactive proteins. The intensity of the resulting scattered light in the nephelometer is dependent upon the C-reactive protein content so that, by comparison to standards of known concentration the C-reactive protein content of a sample can be determined.

AI/ML Overview

The provided text describes a 510(k) summary for the N Latex CRP mono Reagent, an in vitro diagnostic device. Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined "acceptance criteria" in terms of specific performance thresholds for correlation or precision that the device needed to meet. Instead, it reports the performance characteristics observed in studies. The legal standard for 510(k) is "substantial equivalence" to a legally marketed predicate device, rather than meeting specific performance criteria.

Performance MetricReported Device Performance (N Latex CRP mono Reagent)
Correlation
Correlation Coefficient0.98 (vs. N Latex CRP)
Y-intercept1.43 (vs. N Latex CRP)
Slope0.95 (vs. N Latex CRP)
Precision
Intra-assay (n=20)4.0% CV (at ~15 mg/L), 2.3% CV (at ~25 mg/L), 4.4% CV (at ~60 mg/L)
Inter-assay (n=10)2.6% to 5.7% CV (across ~10, 15, 25, 45, 60 mg/L)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 71 serum samples were used for the correlation study. The precision studies used samples with various CRP concentrations, tested 20 times for intra-assay and 10 times for inter-assay precision.
  • Data Provenance: Not specified. The document does not indicate the country of origin of the samples or whether the data was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable as the device is an in vitro diagnostic reagent measuring C-reactive protein (CRP) levels. Ground truth is established by the measurement itself, not by expert interpretation of images or clinical assessments in the way it would be for a diagnostic imaging AI. The "ground truth" for the correlation study would implicitly be the results obtained by the predicate device (N Latex CRP).

4. Adjudication Method for the Test Set

This information is not applicable for this type of in vitro diagnostic device, as there's no interpretation by multiple readers requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

This information is not applicable as the device is an in vitro diagnostic reagent, not an AI for image analysis or clinical decision support that would involve human readers.

6. Standalone (Algorithm Only) Performance Study

The reported performance characteristics (correlation and precision) are essentially standalone performance evaluations of the N Latex CRP mono Reagent itself, compared to a predicate device or against itself under various conditions. There is no human-in-the-loop component for the measurement of CRP by this device.

7. Type of Ground Truth Used

For the correlation study, the "ground truth" was implicitly established by the measurements obtained using the legally marketed predicate device, N Latex CRP. For precision studies, the measurements themselves define the performance.

8. Sample Size for the Training Set

This information is not applicable. This device is a reagent for a specific biochemical measurement, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The "training" of the device is inherent in its chemical formulation and calibration process.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the reasons stated in point 8. The device's calibration refers to a "lot independent standard" rather than a training set with established ground truth.

{0}------------------------------------------------

K962523

JUL 10 1998

Attachment 1

510(k) Summary of Safety and Effectiveness for N Latex CRP mono Reagent

1. Manufacturer Name, Adress, phone number, contact name and date of preparation:

  • Manufacturer: Behrinawerke AG. Postfach 1140 35001 Marburg Germany
  • Distributor: Behring Diagnostics Inc. 151 University Avenue Westwood, MA 02090 617 - 320 - 3023 Contact name: Kathleen Dray-Lyons

date of preparation: May 29, 1996

2. Device name/Classification:

In vitro diagnostic reagents for the quantitative determination of C-reactive protein (CRP) Class II (866.5270).

3. ldentification of the legally marketed device to which the submitter claims equivalence:

The Behringwerke N Latex CRP (K860894)

4. Proposed Device Description:

The proposed test reagent (N Latex CRP mono Reagent) is an in vitro diagnostic reagent intended to be used together with the Behring Nephelometer Systems in the quantitative defermination of C-reactive protein in human serum.

In the proposed product polystyrene particles coated with mouse monoclonal Y antibodies to C-reactive protein are agglutinated when mixed with samples containing C-reactive proteins. The intensity of the resulting scattered light in the nephelometer is dependent upon the C-reactive protein content so that, by comparison to standards of known concentration the C-reactive protein content of a sample can be determined.

{1}------------------------------------------------

5. Proposed Device Intended Use:

The proposed test reagent (N Latex CRP mono Reagent) is an in vitro diagnostic reagent intended to be used together with the Behring Nephelometer Systems in the quantitative determination of C-reactive protein in human serum.

6. Medical device to which equivalence is claimed and comparison information:

The N Latex CRP mono Reagent is substantially equivalent in intended use and results obtained to N Latex CRP. The N CRP Reagent, like the proposed product is intended to be used for the quantitative determination of C-reactive protein in human serum by particle enhanced nephelometry

The N Latex CRP mono Reagent differs from the N Latex CRP in that the N Latex CRP mono Reagent is calibrated using a lot independent standard, whereas the N CRP Reagent is calibrated using a lot dependent standard (included in the kit). Also. the N Latex CRP mono Reagent is a liquid reagent consisting of polystyrene particles coated with mouse monoclonal antibodies to CRP, where as the N CRP Reagent is a lyphilized reagent containing polystyrene particles which are coated with antibodies of the y-fraction from a specific rabbit anti-human-CRP serum. Additionally, the N Latex CRP mono Reagent is designed to measure CRP concentrations within an overall range of approx. 0.175 - 1100 mg/l , whereas, the N Latex CRP measuring range is approx. 0.625 - 800 mg/l. A copy of the N Latex CRP Reagent package insert is included in this submission.

7. Proposed Device Performance Characteristics:

Correlation

Results of comparative studies using the N Latex CRP mono Reagent and the N Latex CRP test kit for 71 serum samples qave a correlation coefficient of 0.98 and a y-intersept of 1.43, and a slope of 0.95.

Precision

N Latex CRP mono Reagent was used to measure 3 different CRP concentrations (approx. 15, 25 and 60 mg/l) and yielded coefficients of variation of 4.0%, 2.3% and 4.4% for the intra-assay precision (n = 20).

Five CRP concentrations (approx. 10, 15, 25, 45 and 60 mg/l) were used to determine the inter-assay reproducibility (n = 10) and here the coefficients of variation ranged from 2,6 % to 5.7 %.

§ 866.5270 C-reactive protein immunological test system.

(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).