K Number
K962373

Validate with FDA (Live)

Date Cleared
1996-07-11

(21 days)

Product Code
Regulation Number
866.5270
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Accuracy control of the quantitative nephelometric and turbidimetric determination of ASL and CRP in the human serum using N Latex ASL and N Latex CRP mono Reagents or Turbiquant ASL and Turbiquant CRP respectively.

In addition the product also serves as a accuracy control of the quantitative nephelometric determination of rheumatoid factors, and ADNase B* (only in N/T Rheumatology Control SL/2) in human serum using N Latex RF and N Latex ADNase B Reagent respectively.

Device Description

N/T Rheumatology Controls SL (bi-level control; N/T Rheumatology Control SL/1 and N/T Rheumatology Control SL/2) is a multi-constituent control intended for use as a quality control material to monitor the accuracy of selected rheumatic and inflammatory disease with the Behring Nephelometer (BN), Behring Nephelometer 100 (BN-100) and the Behring TurbiTimeSystem® (TTS).

AI/ML Overview

This is a quality control device for laboratory tests, not a device that involves human physiological measurements or AI. Therefore, most of the requested information (e.g., expert-established ground truth, MRMC studies, effect size of human improvement with AI, sample size for training sets) is not applicable to this type of device.

Here's the relevant information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text describes performance characteristics related to accuracy. While explicit "acceptance criteria" for the device are not stated as numerical thresholds (e.g., "recovery must be between X% and Y%"), the study demonstrates that the device met the intended accuracy by showing a range of recovery within a generally acceptable range for quality control materials.

Acceptance Criteria (Implied)Reported Device Performance
Accuracy as measured by percent recovery from assigned values.Percent recoveries compared to assigned values ranged from 92.1% to 107.2%.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: "Multiple vials of the controls" were used. "Two lots of each control were tested." "Three lots of N Latex CRP mono reagent were assayed in replicates of five."
  • Data Provenance: Not specified, but generally, quality control materials are tested in laboratory settings. This would be considered a prospective lab study.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. For a quality control material, the "ground truth" (assigned values) is typically established by the manufacturer through rigorous calibration and assay development against reference standards, not by clinical experts for diagnostic interpretation.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods are used when there is subjective interpretation of data (e.g., medical images). Here, the comparison is to pre-established assigned values.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC study was not done. This type of study is relevant for diagnostic devices where human readers interpret patient cases, often with and without AI assistance. This device is a quality control material.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Not applicable. This is a quality control material, not an algorithm or AI system. Its performance is evaluated by its ability to accurately reflect known concentrations when used in conjunction with a diagnostic test.

7. The Type of Ground Truth Used

The "ground truth" for this device is the assigned values for the constituents in the control material. These values are established by the manufacturer using reference methods and standards during the product development and manufacturing process.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable.

{0}------------------------------------------------

JI | 1996

K962373

0000000

510(k) Summary of Safety and Effectiveness for Rheumatology Controls SL1/2 N/T

  • Manufactures Name, Address, Telephone, and contact person, date of 1 . preparation :
ManufactureBehringwerke AGPostfach 114035001 MarburgGermany
DistributorBehring Diagnostics Inc.151 University AvenueWestwood, MA 02090617-320-3000Attn: Kathleen Dray-Lyons

Preparation date: June 19, 1996

2 . Device Name/ Classification:

N/T Rheumatology Control SL1/2: Quality Control Material (assaved) class I (862.1660) Classification Number:

3 . Identification of the legally marketed device:

N/T CRP, RF, and ASL Control Serum

4 . Proposed Device Description:

N/T Rheumatology Controls SL (bi-level control; N/T Rheumatology Control SL/1 and N/T Rheumatology Control SL/2) is a multi-constituent control intended for use as a quality control material to monitor the accuracy of selected rheumatic and inflammatory disease with the Behring Nephelometer (BN), Behring Nephelometer 100 (BN-100) and the Behring TurbiTimeSystem® (TTS) .

5. Proposed Device Intended Use:

ﻤﺴﺘﻘﻠﺔ

Accuracy control of the quantitative nephelometric and turbidimetric determination of ASL and CRP in the human serum using N Latex ASL and N Latex CRP mono Reagents or Turbiquant ASL and Turbiquant CRP respectively.

In addition the product also serves as a accuracy control of the quantitative nephelometric determination of rheumatoid factors, and ADNase B* (only in N/T Rheumatology Control SL/2) in human serum using N Latex RF and N Latex ADNase B Reagent respectively.

{1}------------------------------------------------

Medical device to which equivalence is claimed and comparison 6 . . information:

The N/T Rheumatology Controls SL are substantially equivalent in intended use to the individual N/T CRP Control Serum. N/T RF Control Serum and N/T ASL Control Serum. All products are in vitro diagnostic reagents intended for use as a quality control material to monitor the accuracy of certain immunology procedures. The N/T Rheumatology Controls SL, however, eliminates using three separate controls since all three constituents are included in one vial.

The N/T Rheumatology Controls SL differ from the N/T CRP Control Serum. N/T RF Control Serum and N/T ASL Control Serum in that the N/T Rheumatology Controls SL are liquid while the individual controls are lyophilized and the N/T Rheumatology Controls SL are bi-level while the individual controls are single-level.

7. Proposed Device Performance Characteristics:

Accuracy studies using multiple vials of the controls were run on three Behring Nephelometers. Two lots of each control were tested. Three lots of N Latex CRP mono reagent were assayed in replicates of five on the Behring Nephelometers. The mean values and the percent recovery from the assigned values were calculated for each parameter. The percent recoveries compared to assigned values ranged from 92.1 to 107.2%

§ 866.5270 C-reactive protein immunological test system.

(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).