BRAUN PERCUTANEOUS INTRODUCER SET

K961745 · B.Braun Medical, Inc. · DYB · Sep 17, 1996 · Cardiovascular

Device Facts

Record IDK961745
Device NameBRAUN PERCUTANEOUS INTRODUCER SET
ApplicantB.Braun Medical, Inc.
Product CodeDYB · Cardiovascular
Decision DateSep 17, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The Braun Percutaneous Introducer Kits are designed to facilitate percutaneous introduction of devices into the vasculature.

Device Story

Braun Percutaneous Introducer Kits consist of introducer sheath, dilator, guidewire, and introducer needle. Used by clinicians to facilitate percutaneous vascular access for subsequent device placement. Sheath/dilator assembly inserted into vasculature; provides conduit for catheter or lead delivery. Benefits include simplified vascular access and reduced trauma during device introduction.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Materials: Teflon FEP or HDPE sheath, HDPE dilator. Dimensions: 3-9 Fr, 7-24 cm length. Mechanical device; no energy source or software. Non-active.

Indications for Use

Indicated for percutaneous introduction of medical devices into the vasculature in patients requiring vascular access.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K961745 SEP 17 1996 # II. SUMMARY OF SAFETY AND EFFECTIVENESS Applicant Name & Address: B. Braun Medical Inc., 824 Twelfth Avenue Bethlehem, PA 18018 (610)691-5400 Contact: Mark Alsberge, Regulatory Affairs Manager Date Prepared: 05/01/96 Common or Usual Name: Braun Percutaneous Introducer Set Device Classification Name: Catheter Introducers & vessel dilators ## Substantial Equivalence: K810460 K894446 K894343 and/or K884494 K771982 and/or K770561 K780126 K850820 K932733 Burron Percutaneous Introducer Set Angeion Hemostasis Introducer Set Venous Introducer "888" Percutaneous Catheter Introducer PCI Kit Arterial/Venous Percutaneous Catheter Introducer Valveless/Portless Sheath Introducer B. Braun Medical, Inc. B. Braun Medical, Inc. Daig USCI Argon Universal Medical Instrument(UMI) Cordis ## Device Description: Braun Percutaneous Introducer Kits consist of an introducer sheath packaged with a corresponding dilator. A guidewire and introducer needle may also be packaged with the sheath and dilator. ## Statement of Intended Use: The Braun Percutaneous Introducer Kits are designed to facilitate percutaneous introduction of devices into the vasculature. ## Summary/Comparison of Technological Characteristics: | | Braun PCI | K810460 | K894446 | Daig | USCI | Argon | UMI | Cordis | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Fr sizes | 3 - 9 | 6-8 | 4-15 | 4 - 9 | 4 - 9 | 4 - 9 | 4 - 8 | 6 - 8 | | Lengths | 7 - 24 cm | not specified in 510(k) | 12 - 24 cm | 8.5 - 12 cm | 8 - 12 cm | unknown | 14 cm | 11 cm | | Matls | Teflon FEP or HDPE sheath with HDPE dilator | Teflon® FEP sheath & dilator | Teflon® FEP sheath and HDPE dilator | unknown - "proprietary materials" | Teflon | unknown | white teflonsheath with polyethylene dilator | unknown | | Sideport | no | yes | yes | no | no | no | no | no | | Gasket | no | no | yes | no | no | no | no | no | | Use | percutaneous introduction of devices into the vasculature | percutaneous introduction of devices into the vasculature | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood. | Designed for introduction of electrophysiology and other catheters into the venous system | percutaneous introduction of catheters into the vasculature (either venous or arterial) | percutaneous introduction of balloon catheters, closed end catheters and temporary pacing leads into the vasculature. | percutaneous introduction of catheters into the vasculature (either venous or arterial) | percutaneous introduction of catheters into the vasculature |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...