Intended Use

The Selector Angiographic Catheter is intended to deliver contrast medium during angiography as well as to provide a conduit for various diagnostic and therapeutic procedures. Selector Angiographic Catheters are intended for single-patient use only.

Device Description

The Selector Angiographic Catheter consists of a radiopaque shaft with an attached molded female luer lock hub. The shaft of the catheter includes a flexible section and an optional soft distal tip. The Selector Angiographic catheter may vary with respect to length, diameter, curve style, and the number of side holes present. Selector Angiographic Catheters can be introduced into the selected blood vessel by either cutdown or percutaneous entry techniques.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Selector Angiographic Catheter based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance CriteriaReported Device Performance
Tensile StrengthRequired criteriaDevice meets criteria
Dynamic PressureRequired criteriaDevice meets criteria
Flow RateRequired criteriaDevice meets criteria
DiametersRequired criteriaDevice meets criteria
RadiopacityRequired criteriaDevice meets criteria
BiocompatibilityRequired criteriaDevice meets criteria

2. Sample Size for Test Set and Data Provenance:

The document does not explicitly state the sample sizes used for the non-clinical performance testing. It also does not specify the country of origin for the data or whether the study was retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

This information is not applicable. The device is a physical medical device (catheter), and its performance is evaluated through physical and material property testing, not through expert review of data like an AI algorithm would be.

4. Adjudication Method for the Test Set:

This information is not applicable, as the evaluation involves non-clinical performance testing of a physical device, not subjective expert assessment requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done. This type of study is typically conducted for AI/software devices to assess the impact of AI assistance on human readers' performance. This document describes a physical medical device.

6. Standalone Performance Study:

Yes, a standalone performance study was done in the sense that the device's physical and material properties were tested independently against pre-defined performance criteria. The description "non-clinical performance testing to provide data supporting the safety and effectiveness of the device" indicates a standalone evaluation of the device itself.

7. Type of Ground Truth Used:

The "ground truth" for this device's performance is based on engineering specifications and established performance criteria for medical devices that ensure safety and effectiveness. This includes measurements of tensile strength, dynamic pressure, flow rate, diameters, radiopacity, and biocompatibility, as determined by laboratory testing.

8. Sample Size for the Training Set:

This information is not applicable. The device is a physical medical device, not an AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable, as there is no training set for a physical medical device.

{0}------------------------------------------------

PREMARKET NOTIFICATION

K960801

VII: 510(K) SUMMARY

Manufacturer:NAMIC USA CorporationGlens Falls, New York 12801MAR 2 9 1996
www.rammannen.comment.com------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Contact Person:Mary Meagher RubinRegulatory Affairs Specialist
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Telephone Number (518) 798-0067Facsimile Number (518) 798-5475
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Date Prepared:February 26, 1996
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Trade Name:Selector Angiographic Catheter
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Common Name:Angiographic Catheter
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Classification Name:Catheter, AngiographicCatheter, CardiovascularCatheter, Intravascular, Diagnostic
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Predicate Device:Selector Angiographic Catheter (K922552)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Intended Use:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------The Selector Angiographic Catheter is intended to deliver contrast medium duringangiography as well as to provide a conduit for various diagnostic and therapeuticprocedures. Selector Angiographic Catheters are intended for single-patient use only.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Product Description:
The Selector Angiographic Catheter consists of a radiopaque shaft with an attachedmolded female luer lock hub. The shaft of the catheter includes a flexible section and anoptional soft distal tip. The Selector Angiographic catheter may vary with respect to
length, diameter, curve style, and the number of side holes present.
Selector Angiographic Catheters can be introduced into the selected blood vessel byeither cutdown or percutaneous entry techniques.
Selector™Angiographic CatheterNAMIC® USA CORPORATIONPage 4

{1}------------------------------------------------

VII. 510(K) Summary (Continued).

Comparison to Predicate Device:

The Selector Angiographic Catheter is equivalent to the currently marketed Selector Angiographic Catheter. The proposed new Selector Angiographic Catheter will be offered in a larger French size (diameter) than the currently marketed device.

The Selector Angiographic Catheter is manufactured in the same manner as the predicate device. Packaging and sterilization for the Selector Angiographic Catheter are the same as for other NAMIC devices.

Performance Testing:

The new Selector Angiographic Catheter has been subjected to non-clinical performance testing to provide data supporting the safety and effectiveness of the device. The test protocol included verification of tensile, dynamic pressure testing, flow rate, diameters, and radiopacity. The resultant data demonstrates that the device meets the required performance criteria.

Biocompatibility:

Biocompatibility testing has been performed on the currently marketed Selector Angiographic Catheter (K922552). The material composition of the new Selector Angiographic Catheter is the same as currently marketed NAMIC products.

§ 870.1200 Diagnostic intravascular catheter.

(a)
Identification. An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.(b)
Classification. Class II (performance standards).