K Number
K955895

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-08-30

(248 days)

Product Code
Regulation Number
866.3220
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The E. histolytica Test can be used to detect adhesin (also referred to as galactose-inhibitable lectin) produced by strains of E. histolytica. It does not cross-react with the adhesin from E. dispar (formerly known as nonpathogenic E. histolytica). The test can be used to detect the adhesin in fecal specimens from persons suspected of having amebiasis.

Device Description

The kit, which includes ready-to-use reagents, contains microtiter wells coated with polyclonal antibody, positive control reagent, monoclonal-antibody conjugate, diluent, two component substrate, wash solution, and intensifier. The microtiter wells coated with polyclonal antibody "capture" the adhesin and the monoclonal antibodyconjugate serves as the "detecting" antibody. The polyclonal antibody used to coat the wells is prepared from hyperimmune antiserum developed in rabbits. The monoclonal antibody used to prepare the conjugate is prepared from mouse ascites fluid. The E. histolytica Test is to be used in an ELISA format.

AI/ML Overview

Here's an analysis of the provided text to extract the requested information about device acceptance criteria and the study that proves the device meets those criteria:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Correlation with zymodeme analysis of > 93%> 93% correlation with zymodeme analysis

Study Details

  1. Sample size used for the test set and the data provenance: The document does not explicitly state the exact sample size. It mentions "stool specimens in areas where amebiasis is endemic," implying a prospective or at least recently collected retrospective set from endemic regions globally.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: The text does not specify the number of experts or their qualifications. The ground truth method (zymodeme analysis) is described as "available only in a select number of clinical laboratories around the world," suggesting specialized expertise is involved in performing this analysis, but not in a consensus reading of results.

  3. Adjudication method for the test set: Not applicable, as the ground truth (zymodeme analysis) is presented as a definitive biochemical method, not requiring expert adjudication in the traditional sense of image or clinical interpretation.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an ELISA diagnostic test, not an AI-assisted diagnostic tool or an imaging modality that would involve human readers.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, the device's performance stands alone, comparing its ELISA results directly against zymodeme analysis. It is an in vitro diagnostic test, not requiring human interpretation of complex outputs.

  6. The type of ground truth used: Zymodeme analysis of Entamoeba isolates cultured from stool specimens. This is described as the "gold standard" for distinguishing between pathogenic E. histolytica and non-pathogenic E. dispar.

  7. The sample size for the training set: Not explicitly stated. The document describes the "clinical evaluations" and comparison with zymodeme analysis, implying a single evaluation set rather than a distinct training set. For an ELISA kit, "training" would typically refer to assay development and optimization, rather than machine learning training.

  8. How the ground truth for the training set was established: Not applicable in the context of an ELISA kit with machine learning "training" as described above. The ground truth for the performance evaluation (test set) was established by zymodeme analysis.

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS

1. Name of Manufacturer

K955895

TechLab, Inc. Corporate Research Center 1861 Pratt Drive Blacksburg, VA 24060

AUG 30 1996

2. Establishment Registration

Federal ID # 54-1527427 Initial Registration of Medical Device Establishment, #1122855

3. Trade Name

E. histolytica Test

4. Common Name

E. histolytica ELISA

5. Class of Device

This device is classified in Class I.

6. Performance Standards

No performance standards have been developed for this device under 514 of the Food, Drug, and Cosmetic Act.

7. Safety and Effectiveness

The E. histolytica Test can be used to detect adhesin (also referred to as galactose-inhibitable lectin) produced by strains of E. histolytica. It does not cross-react with the adhesin from E. dispar (formerly known as nonpathogenic E. histolytica). The test can be used to detect the adhesin in fecal specimens from persons suspected of having amebiasis. The kit, which includes ready-to-use reagents, contains microtiter wells coated with polyclonal antibody, positive control reagent, monoclonal-antibody conjugate, diluent, two component substrate, wash solution, and intensifier. The microtiter wells coated with polyclonal antibody "capture" the adhesin and the monoclonal antibodyconjugate serves as the "detecting" antibody. The polyclonal antibody used to coat the wells is prepared from hyperimmune antiserum developed in rabbits. The monoclonal antibody used to prepare the conjugate is prepared from mouse ascites fluid.

{1}------------------------------------------------

The E. histolytica Test is to be used in an ELISA format and is substantially equivalent to culturing and zymodeme analysis that are used in some clinical laboratories as diagnostic aids. Culturing is used to obtain the isolate and zymodeme analysis is used to examine the enzyme profile of the isolate. Zymodeme analysis must be used to determine if the isolated Entamoeba strain is pathogenic. The major disadvantages are that culturing and zymodeme are time-consuming and laborintensive. Only a few clinical laboratories around the world are capable of performing this type of analysis. The E. histolytica Test offers a major advantage to clinical laboratories because it is rapid. easy-to-perform, and it is the first test to offer clinical labs a simpler and easier alternative format for the specific detection of pathogenic E. histolytica.

The E. histolytica Test is different from two other ELISAs currently on the market in the U.S. These other ELISAs, TechLab's Entamoeba Test and the Alexon ProSpecT Entamoeba histolytica Test, detect both E. histolytica (formerly referred to as pathogenic E. histolytica) and E. dispar (formerly referred to as nonpathogenic E. histolytica). They do not distinguish between E. histolytica and E. dispar. The E. histolytica Test is specific for the adhesin of E. histolytica and it does not cross-react with the adhesin of E. dispar. Although all of these tests serve as diagnostic aids for amebiasis, the E. histolytica Test offers the advantage of being specific for pathogenic strains.

The E. histolytica Test was used to analyze stool specimens in areas where amebiasis is endemic, and the results were compared with zymodeme analysis of Entamoeba isolates cultured from these specimens. It is important to remember, however, that culture, without the aid of zymodeme analysis, does not distinguish between E. histolytica and E. dispar. Zymodeme analysis, which is available only in a select number of clinical laboratories around the world, is the only method that distinguishes these species. For the purpose of our studies, zymodeme analysis represents the "gold standard". The results of our clinical evaluations show that the E. histolytica Test exhibits a correlation of >93% when compared with zymodeme analysis, and demonstrate that the test is useful for the detection of pathogenic E. histolytica in stool specimens.

§ 866.3220

Entamoeba histolytica serological reagents.(a)
Identification. Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies toEntamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identifyEntamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasiteEntamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.