Intended Use

Function: to relax muscle spasm

Device Description

powered muscle stimulator

AI/ML Overview

Here's an analysis of the provided text regarding the XH-1 Neuromuscular Stimulator's acceptance criteria and study, structured as requested:

Acceptance Criteria and Device Performance for XH-1 Neuromuscular Stimulator

This submission (K950550) asserts substantial equivalence to existing devices rather than defining specific performance-based acceptance criteria for the XH-1 itself. The "study" presented is a comparison to a predicate device (Myopulse) and other 510(k)-cleared devices to demonstrate similar technical features.

1. Table of Acceptance Criteria and Reported Device Performance

No.ItemAcceptance Criteria (Predicate/Comparable Device Feature)Reported Device Performance (XH-1)
1Frequency0.9 Hz (Myopulse); Adjustable to 1 Hz (K850013, K893878, K893879)0.9 Hz
2Biphasic Wave without Direct CurrentBiphasic wave without direct current (Myopulse)Equivalent
3Wave FormSimilar (Myopulse)Similar
4Pulse Peak Output Voltage & Current (10,000 ohms load)+90V, -115V; +9mAmp, -11.5mAmp (Myopulse)+110V, -110V; +11mAmp, -11mAmp
5Output Frequency FixedFixed (Myopulse)Equivalent
6Output Pulse-Width Fixed under Constant Resistant LoadFixed (Myopulse)Equivalent
7Interpulse Rest TimeApproximately 1.1 second (Myopulse); Adjustable to 1-second off time (K893878, K893879, K926510)Approximately 1.1 second
8Continuous Output at 0.9 HzContinuous output at 0.9 Hz (Myopulse); Adjustable to continuous output (K840346, K926510)Continuous output
9Average Output Current (10,000 ohms load)0.013 mAmp (Myopulse)0.18 mAmp
10Electrical Consumption / Battery LifeOne 9V battery, 350 hours operation (Myopulse)Two AA batteries, 3V, 700 hours operation
11Electrode Coupling MediumSponge moistened with water (Myopulse)Paper towel moistened with water
12Function: to relax muscle spasmTo relax muscle spasm (Myopulse)Equivalent
13History of Clinical UsageSince 1968 to present (Myopulse)Since 1987 to present
14Adverse EffectsEquivalent (Myopulse)Equivalent
15Proprietary Name (e.g., Neuromuscular Stimulator)Equivalent proprietary names (K840346, K926510)Equivalent

Note: For items marked "Equivalent" or "Similar" in the original table, the "Acceptance Criteria" column reflects the predicate device's characteristic that the XH-1 claims equivalence to. For quantifiable metrics, the predicate's value is listed as the implicit "acceptance criterion." Specific 510(k) numbers are cited where indicated in the second comparison chart.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not applicable. The submission does not describe a clinical "test set" in the traditional sense, but rather a comparison of technical specifications. No human subjects or patient data are presented.
  • Data Provenance: The data provenance is from the technical specifications of the XH-1 device itself, and the comparison is made against the technical specifications of the Myopulse Neuromuscular Stimulator (a preamendment device) and other 510(k) cleared devices (K840346, K850013, K893878, K893879, K926510). This is not retrospective or prospective clinical data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not applicable. No "ground truth" for a test set was established by experts in this submission. The comparison is based on published technical specifications and claims of equivalence made by the submitter (Dr. Laiyin Yan).
  • Qualifications of Experts: Dr. Laiyin Yan (alias Fa Lin) from the California Medical Research Institute is the signatory of the submission. The submission references "the Federal Engineering department associated with the National Electric Code and according to the State of California OSHA Consultation Group" as sources for the "Equivalent*" notation for average output current, suggesting consultation with these bodies. However, these are not "experts establishing ground truth for a test set" in the context of clinical performance.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. There was no test set or clinical data requiring adjudication. The comparison is a direct technical feature-by-feature evaluation against predicate devices.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No. This submission pertains to a neuromuscular stimulator and does not involve imaging or AI assistance for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable. The XH-1 is a physical medical device (neuromuscular stimulator), not an algorithm or AI system. Its "performance" is based on its electrical output characteristics.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The "ground truth" in this context refers to the established technical specifications and performance characteristics of the predicate device (Myopulse) and other cleared 510(k) neuromuscular stimulators. These are official device clearances and product specifications rather than expert consensus, pathology, or outcomes data. The claim of "equivalence" is the primary "truth" being established based on these technical comparisons.

8. The Sample Size for the Training Set

  • Sample Size: Not applicable. There is no concept of a "training set" for this type of device submission. The device's design and manufacturing are presumably based on established engineering principles for neuromuscular stimulation.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth Establishment: Not applicable. As there is no training set, there is no ground truth for a training set to be established.

{0}------------------------------------------------

K950550


"This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92."

SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION OF SUBSTANTIAL EQUIVALENCE ON THE MAIN TECHNICAL FEATURES BETWEEN THE XH-1 NEUROMUSCULAR STIMULATOR AND THE MYOPULSE NEUROMUSCULAR STIMULATOR

No.ItemXH-1, apoweredmusclestimulatorMyopulse, apoweredmusclestimulatorLevel ofSafety andEffectiveness
1Frequency 0.9 HzEquivalentEquivalentEquivalent
2Biphasic wave withoutdirect currentEquivalentEquivalentEquivalent
3Wave FormSimilarSimilarEquivalent
4Pulse peak outputvoltage and current,10,000 ohms load.+110volts,-110volts.+11mAmp,-11mAmp.+90volts-115volts.+9mAmp,-11.5mAmp.Equivalent
5Output frequency fixed.EquivalentEquivalentEquivalent
6Output pulse-widthfixed under a constantresistant load.EquivalentEquivalentEquivalent
7Interpulse rest ofapproximately 1.1second.EquivalentEquivalentEquivalent
8@ 0.9 Hz frequency,continuous outputEquivalentEquivalentEquivalent
9Average output current,10,000 ohms load.0.18mAmp0.013 mAmpEquivalent*
10Frugal on electricalconsumptionTwo AAbatteries,3volts,canoperatefor700 hours.One 9voltbatter. canoperatefor350 hours.Equivalent
11Electrode CouplingMediumPapertowelmoistenedwith water.Spongemoistenedwith water.Equivalent
12Function: to relaxmuscle spasmEquivalentEquivalentEquivalent
13History of clinicalusageSince 1987to presentSince 1968to presentEquivalent
14Adverse effects areEquivalentEquivalentEquivalent

according to the Federal Engineering department associated with the National Electric Code and according to the State of California OSHA Consultation Group.

The 14 main technical features in the summary prove that the XH-1 Neuromuscular Stimulator is substantially equivalent, in terms of safety and effectiveness, to the Myopulse Neuromuscular Stimulator, which is a preamendment device marketed prior to May 28, 1976.

text
Fa Lin
Dr. Laiyin Yan alias Fa Lin
CALIFORNIA MEDICAL RESEARCH INSTITUTE
January 17, 1995

ਤੇ

The requested image does not appear to contain any text.

{1}------------------------------------------------

COMPARISON CHART OF THE SIMILARITIES OF MAIN TECHNICAL FEATURES BETWEEN THE XH-1 NEUROMUSCULAR STIMULARDE AND OTHER ELECTRICAL MUSCLE STIMULATORS WITH 510(K) RUMBER

No.ItemXH-1Other Products
1NeuromuscularStimulator asproprietarynameEquivalentThose with equivalentproprietary names:K840346K926510
2Frequency of 1 Hz0.9 HzThose that can be adjustedto 1 Hz output:K850013K893878K893879
3Interpulse restingtime ofapproximately 1second1.1 secondThose that can be adjustedto 1-second off time:K893878K893879K926510
4Pulses withcontinuous outputAt 0.9 Hz,continuousoutputThose that can be adjustedto continuous output:K840346K926510

fa Lin

Dr. Laiyin Yan CALIFORNIA MEDICAL RESEARCH INSTITUTE U.S.A. January, 1995

4

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).