K Number
K203564

Validate with FDA (Live)

Date Cleared
2021-12-22

(380 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For In Vitro Diagnostic Use.

The Immunalysis SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 30 ng/mL in neat oral fluid collected by Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of oxycodone in human oral fluid with clinical analyzers. This assay is calibrated against oxycodone.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

Device Description

The SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay is an in vitro test to detect the presence of oxycodone in human oral fluid samples collected by Quantisal II Oral Fluid Collection Device.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
Precision (Qualitative Mode) - QuantisalConsistently identify negative samples below cutoff and positive samples above cutoff, with mixed results around cutoff.At -100%, -75%, -50%, -25% of cutoff (0, 7.5, 15, 22.5 ng/mL): All 60 determinations were Negative. At Cutoff (30 ng/mL): 31 Negative / 29 Positive. At +25%, +50%, +75%, +100% of cutoff (37.5, 45, 52.5, 60 ng/mL): All 60 determinations were Positive.
Precision (Semi-Quantitative Mode) - QuantisalMean concentrations should be close to expected values, with consistent negative/positive calls.At -100% (0 ng/mL): Mean Conc. -0.2 ng/mL, 60 Negative.At -50% (15 ng/mL): Mean Conc. 16.7 ng/mL, 60 Negative.At -25% (22.5 ng/mL): Mean Conc. 24.3 ng/mL, 60 Negative.At Cutoff (30 ng/mL): Mean Conc. 32.3 ng/mL, 16 Negative / 44 Positive.At +25% (37.5 ng/mL): Mean Conc. 42.1 ng/mL, 60 Positive.At +50% (45 ng/mL): Mean Conc. 53.3 ng/mL, 60 Positive.At +75% (52.5 ng/mL): Mean Conc. 64.3 ng/mL, 60 Positive.At +100% (60 ng/mL): Mean Conc. 76.3 ng/mL, 60 Positive.
Precision (Qualitative Mode) - Quantisal II Pad AConsistently identify negative samples below cutoff and positive samples above cutoff, with mixed results around cutoff.At -100%, -75%, -50%, -25% of cutoff (0, 7.5, 15, 22.5 ng/mL): All 60 determinations were Negative. At Cutoff (30 ng/mL): 31 Negative / 29 Positive. At +25%, +50%, +75%, +100% of cutoff (37.5, 45, 52.5, 60 ng/mL): All 60 determinations were Positive.
Precision (Semi-Quantitative Mode) - Quantisal II Pad AMean concentrations should be close to expected values, with consistent negative/positive calls.At -100% (0 ng/mL): Mean Conc. 0.5 ng/mL, 60 Negative.At -75% (7.5 ng/mL): Mean Conc. 7.3 ng/mL, 60 Negative.At -50% (15 ng/mL): Mean Conc. 14.7 ng/mL, 60 Negative.At -25% (22.5 ng/mL): Mean Conc. 21.7 ng/mL, 60 Negative.At Cutoff (30 ng/mL): Mean Conc. 29.9 ng/mL, 36 Negative / 24 Positive.At +25% (37.5 ng/mL): Mean Conc. 42.4 ng/mL, 60 Positive.At +50% (45 ng/mL): Mean Conc. 50.9 ng/mL, 60 Positive.At +75% (52.5 ng/mL): Mean Conc. 57.3 ng/mL, 60 Positive.At +100% (60 ng/mL): Mean Conc. 66.5 ng/mL, 60 Positive.
Precision (Qualitative Mode) - Quantisal II Pad BConsistently identify negative samples below cutoff and positive samples above cutoff, with mixed results around cutoff.At -100%, -75%, -50%, -25% of cutoff (0, 7.5, 15, 22.5 ng/mL): All 60 determinations were Negative. At Cutoff (30 ng/mL): 28 Negative / 32 Positive. At +25%, +50%, +75%, +100% of cutoff (37.5, 45, 52.5, 60 ng/mL): All 60 determinations were Positive.
Precision (Semi-Quantitative Mode) - Quantisal II Pad BMean concentrations should be close to expected values, with consistent negative/positive calls.At -100% (0 ng/mL): Mean Conc. 0.3 ng/mL, 60 Negative.At -75% (7.5 ng/mL): Mean Conc. 7.0 ng/mL, 60 Negative.At -50% (15 ng/mL): Mean Conc. 14.9 ng/mL, 60 Negative.At -25% (22.5 ng/mL): Mean Conc. 22.7 ng/mL, 60 Negative.At Cutoff (30 ng/mL): Mean Conc. 30.8 ng/mL, 28 Negative / 32 Positive.At +25% (37.5 ng/mL): Mean Conc. 42.9 ng/mL, 60 Positive.At +50% (45 ng/mL): Mean Conc. 49.4 ng/mL, 60 Positive.At +75% (52.5 ng/mL): Mean Conc. 58.3 ng/mL, 60 Positive.At +100% (60 ng/mL): Mean Conc. 67.1 ng/mL, 60 Positive.
Specificity and Cross-ReactivityMinimal or no cross-reactivity with structurally similar compounds at high concentrations, and significant cross-reactivity with known related compounds.No/Minimal Cross-Reactivity (<0.08% at 40,000 ng/mL): 6-acetylcodeine, 6-acetylmorphine, Buprenorphine, Codeine, Desomorphine, Dihydrocodeine, Ethylmorphine, Fentanyl, Heroin, Hydrocodone, Hydromorphone, Levorphanol, Meperidine, Morphine, Morphine-3-B-D-glucuronide, Morphine-6-B-D-glucuronide, Naltrexone, Norbuprenorphine, Norcodeine, Normorphine, Tapentadol, Tramadol.Significant Cross-Reactivity: Naloxone (0.5% at 6,000 ng/mL), Noroxycodone (0.6% at 4,750 ng/mL), Noroxymorphone (0.2% at 12,500 ng/mL), Oxymorphone (85.7% at 35 ng/mL), Oxymorphone-3-β-D-glucuronide (5.5% at 550 ng/mL).
Interference - Structurally Unrelated CompoundsNo interference observed for a wide range of common compounds.No interference observed with 89 common drugs and substances at concentrations up to 40,000 ng/mL (20,000 ng/mL for Chlorpromazine). (See Table 9 for complete list).
Interference - Endogenous CompoundsNo interference observed for common endogenous substances.No interference observed with common endogenous compounds at tested concentrations (e.g., Ascorbic Acid: 3 mg/mL, Bilirubin: 0.15 mg/mL, Hemoglobin: 3 mg/mL). (See Table 10).
Interference - Exogenous Compounds (Oral Fluid Samples)No interference observed for common exogenous substances and orally used products.No interference observed with common exogenous compounds (e.g., Acetaminophen: 0.1 mg/mL, Ibuprofen: 0.1 mg/mL, Alcohol: 6% v/v, Caffeine: 0.1 mg/mL) (See Table 11). No interference observed with orally used products (e.g., Teeth Whitener, Cigarette, Hard Candy, Chewing Gum, Hydrogen Peroxide, Sugar, Cough Syrup). (See Table 12).
Interference - pHNo interference observed across a broad pH range.No interference observed for pH values ranging from 3.0 to 11.0.
Linearity/RecoveryLinear range confirmed with acceptable drug recovery percentage.Linear range confirmed to be 10-110 ng/mL with drug recovery percentage from 90.6% to 111.6% across different collection devices. (See Tables 13, 14, 15).
Oxycodone Stability in Oral FluidStability for a specified duration at recommended storage conditions.Oral fluid samples containing oxycodone are stable for up to 12 months at 2°C - 8°C in Quantisal II Oral Fluid Collection Device. Data for 10-day stability at ambient temperature (8°C - 25°C) in Quantisal II was referenced from K183048 and K200801.
Calibration DurationMaintains performance over a specified calibration interval.Test results met acceptance criteria up to 24 days. Recommended frequency of calibration is 14 days.
Method ComparisonGreater than 95% agreement with confirmatory method (LC-MS/MS).Achieved 100% agreement (40/40 positive and 40/40 negative) in both qualitative and semi-quantitative modes for Quantisal, Quantisal II "A", and Quantisal II "B" when compared to LC-MS/MS. (See Tables 16, 17, 18).

2. Sample size used for the test set and the data provenance:

  • Precision Studies:

    • For Quantisal: 60 determinations (15 days, 2 runs/day, 2 collection devices/run, 1 replicate/device).
    • For Quantisal II Pad A: 60 determinations (same as above).
    • For Quantisal II Pad B: 60 determinations (same as above).
    • An additional 20-day study was performed (implied to be similar sample size) to demonstrate repeatability across 3 reagent lots.
    • Data Provenance: The samples were "spiked into drug free negative oral fluid." This indicates a laboratory-prepared, retrospective approach using controlled samples. The document does not specify the country of origin, but the company (Immunalysis Corporation) is based in Pomona, California, USA.
  • Specificity and Cross-Reactivity: Compounds were "spiked into drug free pooled oral fluid." This is also a retrospective, laboratory-controlled study. Sample sizes are not explicitly stated for individual compounds but implied multiple tests per compound.

  • Interference (Structurally Unrelated, Endogenous, Exogenous, pH): Potential interferents were "spiked into drug free oral fluid containing oxycodone at ±25% of the cutoff." Orally used products were tested by collecting oral fluid "from volunteers after use of the substances." This combines prospective (volunteers) and retrospective (spiking in lab) approaches. Sample sizes for interferent tests are not explicitly stated for each compound but involve multiple tests per condition.

  • Linearity/Recovery: "Additional pools were made by serially diluting the high concentration specimen with drug free oral fluid... Each pool was collected by Quantisal and Quantisal II oral fluid collection devices and tested in triplicate." This is a retrospective, laboratory-controlled study.

  • Oxycodone Stability in Oral Fluid: "Drug free negative oral fluid spiked with oxycodone... were collected and stored... tested by LC-MS/MS at each time point." This is a retrospective, laboratory-controlled study.

  • Calibration Duration: "Drug free negative oral fluid spiked with oxycodone at ±25% of the cutoff were tested... at time points up to 14 days and in semi-quantitative at time points up to 25 days." This is a retrospective, laboratory-controlled study.

  • Method Comparison: "Eighty (80) deidentified, unaltered clinical oral fluid samples collected by Quantisal II Oral Fluid Collection Devices were obtained from clinical research facilities." This is a retrospective study using real-world clinical samples. Data Provenance: "clinical research facilities" is general, so specific country of origin is not noted.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The document does not mention any human experts establishing ground truth for the test set in the traditional sense of a diagnostic interpretation study.
  • The ground truth for the "Method Comparison" study was established by Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS), which is an analytical gold standard, not human expert consensus.
  • For the "Precision," "Specificity," "Interference," "Linearity/Recovery," "Oxycodone Stability," and "Calibration Duration" studies, the "ground truth" was established by the precise spiking concentrations of oxycodone and other compounds into known drug-free oral fluid, and confirmation by mass spectrometry (LC-MS/MS) where mentioned (e.g., for precision studies).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • No adjudication method involving human experts is described, as the studies primarily rely on analytical measurements and comparisons to a gold standard (LC-MS/MS or known spiked concentrations).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC comparative effectiveness study was done. This device is an automated in vitro diagnostic assay for detecting a substance, not an AI-assisted diagnostic imaging or interpretation tool for human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Yes, this entire submission describes the performance of a standalone algorithm/device (the SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay) without human-in-the-loop performance influencing the assay results themselves. The assay is intended for use with "clinical analyzers." Human involvement is in collecting samples, running the assay, and interpreting the final quantitative or qualitative output, but not in the diagnostic decision-making of the assay itself during the performance evaluation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • The primary type of ground truth used is analytical gold standard, specifically Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS), which is considered the definitive method for drug identification and quantification in forensic and clinical toxicology.
  • For spiked samples (in precision, specificity, interference, linearity, stability, and calibration studies), the ground truth was the known, precisely prepared concentration of the analyte.

8. The sample size for the training set:

  • The document does not explicitly describe a "training set" in the context of machine learning or AI models. This device is a homogeneous enzyme immunoassay, which is a biochemical reaction-based test, not a software algorithm that undergoes traditional machine learning training. Therefore, this question is not directly applicable. The data presented are performance characterization studies.

9. How the ground truth for the training set was established:

  • Since there is no "training set" described for a machine learning model, this question is not applicable. The assay's "learning" or development process would involve optimizing reagent formulations and reaction conditions, not training on labeled data in the AI sense.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

December 22, 2021

Immunalysis Corporation Elina Arroyo Manager Regulatory Affairs 829 Towne Center Drive Pomona, California 91767

Re: K203564

Trade/Device Name: SEFRIATM Oxycodone Oral Fluid Enzyme Immunoassay Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: DJG Dated: October 7, 2021 Received: October 8, 2021

Dear Elina Arroyo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiologial Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K203564

Device Name SEFRIATM Oxycodone Oral Fluid Enzyme Immunoassay

Indications for Use (Describe) For In Vitro Diagnostic Use.

The Immunalysis SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 30 ng/mL in neat oral fluid collected by Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of oxycodone in human oral fluid with clinical analyzers. This assay is calibrated against oxycodone.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red arrow. The arrow is pointing to the right. The word is written in all capital letters and is centered on the arrow.

510(K) SUMMARY

510(k) Number: K203564

A. GENERAL INFORMATION

Applicant Name:Immunalysis Corporation
829 Towne Center Drive
Pomona, CA 91767
Establishment # 2020952
Company Contact:Elina Arroyo, Manager Regulatory AffairsImmunalysis Corporation829 Towne Center Drive, Pomona, CA 91767 USA(224) 361-7080elina.arroyo@abbott.com
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Date Prepared: October 7, 2021

B. DEVICE IDENTIFICATION

Trade or Proprietary Names: SEFRIATM Oxycodone Oral Fluid Enzyme Immunoassay

Common Name: Oxycodone Oral Fluid Enzyme Immunoassay

C. REGULATORY INFORMATION

Device Classification Name:Enzyme Immunoassay, Opiates
Product Codes:DJG
Regulatory Class:Class II
Classification Regulation:862.3650
Panel:Toxicology (91)
Predicate Device:Thermo Scientific CEDIA Opiate OFT Assay [K101754]

{4}------------------------------------------------

D. DEVICE DESCRIPTION

The SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay is an in vitro test to detect the presence of oxycodone in human oral fluid samples collected by Quantisal II Oral Fluid Collection Device.

Oxycodone was developed in Germany in 1916, intended to be a better medication than other opioids such as codeine and morphine but unfortunately has significant potential for abuse and addiction. It is an opioid analgesic, most often prescribed to control moderate to severe pain; one in 16 surgical patients becomes a long-term user. Overprescribing opioids after surgery is common, and the lack of multidisciplinary procedure-specific guidelines contributes to the wide variation in opioid prescribing practice.[1] Oxycodone is present in formulations combined with other pain-relieving drugs like acetaminophen (Percocet®) or as extended release tablets (OxyContin®). The disposition of oxycodone in oral fluid has been well documented showing the main drug found after intake is the parent compound. After a single dose of 20 mg oxycodone, concentrations as high as 200 ng/mL may be detected in oral fluid.[1] Oral fluid has been shown to be a useful matrix for the analysis of oxycodone in patients undergoing addiction treatment.13,14]

E. INTENDED USE

For In Vitro Diagnostic Use.

The Immunalysis SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 30 ng/mL in neat oral fluid collected by Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of oxycodone in human oral fluid with clinical analyzers. This assay is calibrated against oxycodone.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

1 Overton HN, Hanna MN, Bruhn WE, Hutfless S, Bicket MC, Makary MA. Opioid-prescribing guidelines for common surgical procedures: An Expert Panel Consensus. J Am Coll Surg. 2018 Oct;227(4):411-418.

2 Cone EJ, DePriest AZ, Heltsley R, Black DL, Mitchell JM, LoDico C, Flegel R. Prescription Opioids. III. Disposition of oxycodone in oral fluid and blood following controlled single-dose administration. J Anal Toxicol. 2015;39(3):192-202.

3 Kunkel F, Fey E, Borg D, Stripp R, Getto C. Assessment of the use of oral fluid as a matrix for drug monitoring in patients undergoing treatment for opioid addiction. J Opioid Manag. 2015:11(5):435-42.

4 Conermann T, Gosalia AR, Kabazie AJ, Moore C, Miller K, Fetsch M, Irvan D. Utility of oral fluid in compliance monitoring of opioid medications. Pain Physician. 2014;17(1):63-70.

{5}------------------------------------------------

F. COMPARISON WITH PREDICATE

The selected predicate device is Thermo Scientific CEDIA Opiate OFT Assay K101754.

AttributeCandidate DeviceSEFRIA Oxycodone Oral FluidEnzyme ImmunoassayPredicate DeviceThermo Scientific CEDIA OpiateOFT Assay [K101754]
Similarities
Test PrincipleIdenticalHomogeneous enzymeimmunoassay
Assay MaterialsIdenticalantibody reagent, drug conjugatereagent
Cutoff LevelIdentical30 ng/mL in Neat Oral Fluid
UserEnvironmentIdenticalFor use in laboratories
Sample MatrixIdenticalHuman oral fluid
Reagent StorageIdentical2-8°C until expiration date
InstrumentationIdenticalAutomated clinical chemistryanalyzer
MassSpectrometryConfirmationIdenticalRequired for preliminary positiveanalytical results
Differences
Intended UseQualitative and semi-quantitativeanalysis of oxycodone in humanoral fluid collected by Quantisal orQuantisal II Oral Fluid CollectionDeviceQualitative determination of opiatesin human oral fluid collected byOral-Eze™ Saliva CollectionSystem
CalibratedAgainstOxycodoneMorphine
SampleCollectionDeviceOral fluid is collected with theQuantisal and Quantisal II OralFluid Collection Device. Sample isstored in a plastic tube containingpreservative buffer with snap cap.Oral fluid is collected with theOral-Eze™ Saliva CollectionSystem. Sample is stored in aplastic tube containing preservativebuffer with snap cap.

G. PERFORMANCE CHARACTERISTICS

The following laboratory performance studies were performed to determine substantial equivalence of the SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay to the predicate device. Assay performance was established using the Beckman Coulter AU480 chemistry analyzer.

1. Precision

Precision study was performed over 15 days, 2 runs per day with 2 collection devices per run (N=60),

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered and in all capital letters. The background is a solid red color and the letters are a bright white color.

one replicate per collection device on 1 lot of reagent and 1 lot of Quantisal and 1 lot of Quantisal II oral fluid collection devices. Drug free negative oral fluid was spiked to concentrations of assay cutoff and ±25%, ±50%, ±75%, ±100% of the cutoff and was collected using the collection devices. The spiked concentrations were confirmed by mass spectrometry (LC-MS/MS) before collection. The study established the repeatability of the testing system, including assay and oral fluid collection device. Test results in qualitative and semi-quantitative modes are presented in Tables 1 to 6.

An additional 20-day study was performed on 3 lots of assay reagent to demonstrate the repeatability across multiple reagent lots.

Concentration(ng/mL)% of Cutoff# of DeterminationsResult
0-100%6060 Negative
7.5-75%6060 Negative
15-50%6060 Negative
22.5-25%6060 Negative
30Cutoff6031 Neg/29 Pos
37.5+25%6060 Positive
45+50%6060 Positive
52.5+75%6060 Positive
60+100%6060 Positive

Table 1. Precision – Qualitative - Quantisal

Table 2. Precision - Semi-Quantitative - Quantisal

Concentration(ng/mL)% ofCutoff# ofDeterminationsMean Conc.(ng/mL)Result
0-100%60-0.260 Negative
15-50%6016.760 Negative
22.5-25%6024.360 Negative
30Cutoff6032.316 Neg/44 Pos
37.5+25%6042.160 Positive
45+50%6053.360 Positive
52.5+75%6064.360 Positive
60+100%6076.360 Positive
Table 3. Precision - Qualitative - Quantisal II Pad A
---------------------------------------------------------------
Concentration(ng/mL)% of Cutoff# of DeterminationsResult
0-100%6060 Negative
7.5-75%6060 Negative
15-50%6060 Negative

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, elongated hexagon. The word is centered in the hexagon. The hexagon is oriented horizontally.

Concentration(ng/mL)% of Cutoff# of DeterminationsResult
22.5-25%6060 Negative
30Cutoff6031 Neg/29 Pos
37.5+25%6060 Positive
45+50%6060 Positive
52.5+75%6060 Positive
60+100%6060 Positive

Table 4. Precision - Semi-Quantitative – Quantisal II Pad A

Concentration(ng/mL)% ofCutoff# ofDeterminationsMean Conc.(ng/mL)Result
0-100%600.560 Negative
7.5-75%607.360 Negative
15-50%6014.760 Negative
22.5-25%6021.760 Negative
30Cutoff6029.936 Neg/24 Pos
37.5+25%6042.460 Positive
45+50%6050.960 Positive
52.5+75%6057.360 Positive
60+100%6066.560 Positive

Table 5. Precision - Qualitative - Quantisal II Pad B

Concentration(ng/mL)% of Cutoff# of DeterminationsResult
0-100%6060 Negative
7.5-75%6060 Negative
15-50%6060 Negative
22.5-25%6060 Negative
30Cutoff6028 Neg/32 Pos
37.5+25%6060 Positive
45+50%6060 Positive
52.5+75%6060 Positive
60+100%6060 Positive

Table 6. Precision - Semi-Quantitative - Quantisal II Pad B

Concentration(ng/mL)% ofCutoff# ofDeterminationsMean Conc.(ng/mL)Result
0-100%600.360 Negative
7.5-75%607.060 Negative

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, horizontally-oriented, hexagon-like shape. The word is written in all capital letters and is centered on the red shape. The red shape is a bright, saturated red color.

Concentration(ng/mL)% ofCutoff# ofDeterminationsMean Conc.(ng/mL)Result
15-50%6014.960 Negative
22.5-25%6022.760 Negative
30Cutoff6030.828 Neg/32 Pos
37.5+25%6042.960 Positive
45+50%6049.460 Positive
52.5+75%6058.360 Positive
60+100%6067.160 Positive

2. Specificity and Cross-Reactivity

Structurally and functionally similar compounds were spiked into drug free pooled oral fluid at levels that will yield a result that is equivalent to the cutoff, if cross reacting. The study verified the cross reactivity of the oxycodone assay to related drugs and drug metabolites, in both the qualitative and semiquantitative modes. Cross-reactivity test results in qualitative mode are presented in Table 7. Crossreactivity test results in semi-quantitative mode are presented in Table 8.

CompoundCompoundConc. (ng/mL)OxycodoneEquivalentConc. (ng/mL)ResultCross-Reactivity (%)
6-acetylcodeine40,000<30NEG<0.08
6-acetylmorphine40,000<30NEG<0.08
Buprenorphine40,000<30NEG<0.08
Codeine40,000<30NEG<0.08
Desomorphine40,000<30NEG<0.08
Dihydrocodeine40,000<30NEG<0.08
Ethylmorphine40,000<30NEG<0.08
Fentanyl40,000<30NEG<0.08
Heroin40,000<30NEG<0.08
Hydrocodone40,000<30NEG<0.08
Hydromorphone40,000<30NEG<0.08
Levorphanol40,000<30NEG<0.08
Meperidine40,000<30NEG<0.08
Morphine40,000<30NEG<0.08
Morphine-3-B-D-glucuronide40,000<30NEG<0.08
Morphine-6-B-D-glucuronide40,000<30NEG<0.08
Naloxone6,00030POS0.5
Naltrexone40,000<30NEG<0.08
Norbuprenorphine40,000<30NEG<0.08
Norcodeine40,000<30NEG<0.08
Normorphine40,000<30NEG<0.08
Noroxycodone4,75030POS0.6
Noroxymorphone12,50030POS0.2

Table 7. Cross-Reactivity - Qualitative

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, elongated hexagon. The word is centered in the shape. The shape is oriented horizontally.

CompoundCompoundConc. (ng/mL)OxycodoneEquivalentConc. (ng/mL)ResultCross-Reactivity (%)
Oxymorphone3530POS85.7
Oxymorphone-3-β-D-glucuronide55030POS5.5
Tapentadol40,000<30NEG<0.08
Tramadol40,000<30NEG<0.08
CompoundCompoundConc. (ng/mL)OxycodoneEquivalentConc. (ng/mL)MeanValue(ng/mL)ResultCross-Reactivity(%)
6-acetylcodeine40,000<303.1NEG<0.08
6-acetylmorphine40,000<302.1NEG<0.08
Buprenorphine40,000<301.4NEG<0.08
Codeine40,000<303.1NEG<0.08
Desomorphine40,000<303.2NEG<0.08
Dihydrocodeine40,000<304.0NEG<0.08
Ethylmorphine40,000<304.9NEG<0.08
Fentanyl40,000<302.1NEG<0.08
Heroin40,000<301.9NEG<0.08
Hydrocodone40,000<304.7NEG<0.08
Hydromorphone40,000<304.1NEG<0.08
Levorphanol40,000<302.9NEG<0.08
Meperidine40,000<301.8NEG<0.08
Morphine40,000<304.0NEG<0.08
Morphine-3-β-D-glucuronide40,000<301.3NEG<0.08
Morphine-6-β-D-glucuronide40,000<301.2NEG<0.08
Naloxone6,0003032.5POS0.5
Naltrexone40,000<3019.8NEG<0.08
Norbuprenorphine40,000<301.3NEG<0.08
Norcodeine40,000<301.4NEG<0.08
Normorphine40,000<301.4NEG<0.08
Noroxycodone4,7503032.6POS0.6
Noroxymorphone12,5003033.0POS0.2
Oxymorphone353031.0POS85.7
Oxymorphone-3-β-D-glucuronide5503030.3POS5.5
Tapentadol40,000<301.6NEG<0.08
Tramadol40,000<301.8NEG<0.08

Table 8. Cross-Reactivity - Semi-Quantitative

3. Interference - Structurally Unrelated Compounds

Structurally unrelated compounds were evaluated in qualitative and semi-quantitative modes by spiking the potential interferent into drug free oral fluid containing oxycodone at ±25% of the cutoff. At the levels tested, there was no interference with structurally unrelated compounds. The concentration levels of structurally unrelated compounds are presented in Table 9.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, horizontally-oriented, hexagon-like shape. The word is centered in the shape. The shape is a bright red color.

CompoundConc. Tested (ng/mL)
4-Bromo-2,5,Dimethoxyphenethylamine40,000
Alprazolam40,000
7-Aminoclonazepam40,000
7-Aminoflunitrazepam40,000
7-Aminonitrazepam40,000
Amitriptyline40,000
S-(+) Amphetamine40,000
Benzylpiperazine40,000
Bromazepam40,000
Bupropion40,000
Butabarbital40,000
Butalbital40,000
Cannabidiol40,000
Cannabinol40,000
Carbamazepine40,000
Carisoprodol40,000
Chlordiazepoxide40,000
Chlorpromazine20,000
Clobazam40,000
Clomipramine40,000
Clonazepam40,000
Clozapine40,000
Cocaine40,000
Cotinine40,000
Cyclobenzaprine40,000
Demoxepam40,000
Desalkylflurazepam40,000
Desipramine40,000
Diazepam40,000
Digoxin40,000
Dehydronorketamine40,000
Delta-9-THC40,000
Diphenhydramine40,000
Dextromethorphan40,000
Doxepin40,000
Ecgonine40,000
Ecgonine Methyl Ester40,000
CompoundConc. Tested (ng/mL)
EDDP40,000
EMDP40,000
1R,2S(-)-Ephedrine40,000
1S,2R(+)-Ephedrine40,000
Ethyl-β-D-Glucuronide40,000
Fenfluramine40,000
Flunitrazepam40,000
Fluoxetine40,000
Flurazepam40,000
Haloperidol40,000
11-hydroxy-delta-9-THC40,000
Imipramine40,000
Ketamine40,000
Lamotrigine40,000
Lidocaine40,000
Lorazepam40,000
Lorazepam Glucuronide40,000
Lormetazepam40,000
LSD40,000
Maprotiline40,000
MDA40,000
MDEA40,000
MDMA40,000
Meprobamate40,000
S(+)-Methamphetamine40,000
Methadone40,000
Methaqualone40,000
Methoxetamine40,000
Methylone40,000
Methylphenidate40,000
Midazolam40,000
N-desmethyltapentadol40,000
N-desmethyl tramadol40,000
N-desmethyl venlafaxine40,000
Nalorphine40,000
Nitrazepam40,000
11-nor-9 carboxy THC40,000
Nordiazepam40,000
Norketamine40,000
CompoundConc. Tested (ng/mL)
Norpropoxyphene40,000
Norpseudoephedrine40,000
Nortriptyline40,000
O-desmethyl tramadol40,000
O-desmethyl venlafaxine40,000
Olanzapine40,000
Oxazepam40,000
Pentazocine40,000
Pentobarbital40,000
Phencyclidine40,000
Phenobarbital40,000
Phentermine40,000
Phenylephrine40,000
Phenytoin40,000
Phenylpropanolamine40,000
PMA40,000
Prazepam40,000
Propranolol40,000
Propoxyphene40,000
Protriptyline40,000
R,R(-)-Pseudoephedrine40,000
S,S(+)-Pseudoephedrine40,000
Ritalinic Acid40,000
Salicylic Acid40,000
Secobarbital40,000
Sertraline40,000
Sufentanil40,000
Temazepam40,000
Theophylline40,000
Thioridazine40,000
Trazadone40,000
Triazolam40,000
Trifluoromethylphenyl-piperazine40,000
Trimipramine40,000
Venlafaxine40,000
Verapamil40,000
Zolpidem Tartrate40,000

Table 9. Non-Interfering Structurally Unrelated Compounds

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, hexagon-like background. The word is written in all capital letters and is centered on the background. In the bottom right corner of the image, there is a small "TM" symbol.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red hexagon. The letters are bold and sans-serif. In the bottom right corner of the hexagon, there is a small "TM" symbol.

{13}------------------------------------------------

4. Interference - Endogenous Compounds and Exogenous Compounds

Endogenous compounds and exogenous compounds were evaluated in qualitative and semi-quantitative modes by spiking the potential interferent into drug free oral fluid containing oxycodone at ±25% of the cutoff. Additional orally used products were tested by collecting oral fluid using Quantisal II Oral Fluid Collection Devices from volunteers after use of the substances. At the levels tested, there was no interference observed with endogenous compounds, exogenous compounds and orally used compounds. Endogenous compounds and exogenous compounds are presented in Tables 10 and 11. Orally used compounds are presented in Table 12.

CompoundConcentration Tested
Ascorbic Acid3 mg/mL
Bilirubin0.15 mg/mL
Cholesterol0.45 mg/mL
γ-Globulin0.8 mg/mL
Hemoglobin3 mg/mL
Human Serum Albumin15 mg/mL
IgA1 mg/mL
IgG1 mg/mL
IgM0.5 mg/mL
Salivary-α-amylase1000 U/mL

Table 10. Non-interfering Endogenous Compounds

Table 11. Non-interfering Exogenous Compounds

CompoundConcentration Tested
Acetaminophen0.1 mg/mL
Acetylsalicylic Acid0.1 mg/mL
Baking Soda0.6% v/v
Denture Adhesive0.6% w/v
Ibuprofen0.1 mg/mL
Alcohol (Ethanol)6% v/v
Caffeine0.1 mg/mL
Cough Syrup6% v/v
Coffee6% v/v
Cranberry Juice6% v/v
Hydrogen Peroxide (3% OTC)0.5% v/v
Milk1% v/v
Mouthwash6% v/v
Naproxen0.1 mg/mL
Orange Juice6% v/v
Soft Drink (Pepsi)6% v/v
Sodium Chloride18 mg/mL
Sugar50 mg/mL
Tea6% v/v
Toothpaste6% v/v

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The background is shaped like a stretched hexagon. The word is written in all capital letters and is centered in the image. The image is simple and clear.

CompoundConcentration Tested
Teeth Whitener2 strips
Cigarette1 cigarette
Hard Candy1 piece
Chewing Gum1 piece
Hydrogen Peroxide (3% OTC)Neat (2 min. mouth rinse)
Sugar2 Teaspoons
Cough Syrup2 Teaspoons
Table 12. Non-interfering Orally Used Exogenous Products

5. Interference - pH

To evaluate potential interference from the effect of oral fluid pH, device performance in the qualitative and semi-quantitative modes was tested using a range of oral fluid pH values (3.0, 4.0, 5.0, 6.0, 7.0, 8.0, 9.0. 10.0 and 11.0). All test samples were prepared in drug free oral fluid containing oxycodone at ±25% of the cutoff. At the pH levels tested, there was no interference observed for each test mode.

6. Linearity/Recovery

Assay linearity was evaluated in the semi-quantitative mode by spiking a drug free oral fluid pool with a high concentration of oxycodone. Additional pools were made by serially diluting the high concentration specimen with drug free oral fluid to achieve concentrations ranging from 10 ng/mL The 0 ng/mL specimen was made from drug free oral fluid. Each pool was collected by Quantisal and Quantisal II oral fluid collection devices and tested in triplicate to calculate the mean concentration values that were used to calculate drug recovery. Linearity test results in semi-quantitative mode are presented in Tables 13 to 15. The study confirmed the linear range to be 10-100 ng/mL with a drug recovery percentage of 90.6% to 111.6% across the collection devices.

Expected ConcentrationMean ConcentrationRecovery (%)
(ng/mL)(ng/mL)
01.8N/A
109.695.7
2021.0104.8
3033.2110.6
4040.7101.8
5055.8111.6
6060.5100.8
7075.8108.2
8080.9101.2
9095.6106.2
100106.7106.7
110112.4102.2

Table 13. Linearity/Recovery - Quantisal

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image shows the word "IMMUNALYSIS" in white text on a red background. The background is shaped like a horizontal arrow pointing to the right. The text is in all caps and appears to be a title or heading.

Expected ConcentrationMean ConcentrationRecovery (%)
(ng/mL)(ng/mL)
0-1.5N/A
1010.2102.0
2020.4102.2
3029.899.2
4042.0104.9
5050.2100.3
6057.195.2
7069.499.1
8081.9102.3
9087.997.7
100107.7107.7
110109.799.7

Table 14. Linearity/Recovery - Quantisal II "A"

Table 15. Linearity/Recovery - Quantisal II "B"

Expected Concentration(ng/mL)Mean Concentration(ng/mL)Recovery (%)
0-1.2N/A
1010.9108.7
2019.497.0
3027.290.6
4043.0107.5
5046.793.5
6059.999.8
7074.3106.2
8085.4106.8
9090.8100.9
100105.4105.4
110117.7107.0

7. Oxycodone Stability in Oral Fluid

Drug free negative oral fluid spiked with oxycodone at +50% of the 30 ng/mL cutoff were collected and stored in Quantisal and Quantisal II Oral Fluid Collection Devices at 2°C - 8°C, tested by LC-MS/MS at each time point and compared to the baseline concentration results indicate that oral fluid samples containing oxycodone are stable for up to 12 months stored in Quantisal II Oral Fluid Collection Device at 2°C - 8°C.

Data to support 10-day storage in Quantisal II Oral Fluid Collection Device at ambient temperature 8°C - 25°C were reported in K183048 and K200801.

8. Calibration Duration

Drug free negative oral fluid spiked with oxycodone at ±25% of the cutoff were tested in qualitative

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red arrow. The arrow is pointing to the right. The text is bolded and in all caps.

mode at time points up to 14 days and in semi-quantitative at time points up to 25 days. At the initial time point, a two-point calibration curve was established in qualitative mode and multi-points calibration curve was established in semi-quantitative mode. These calibrations were used through the duration of the study. The test results met acceptance criteria at each time point up to 24 days. The recommended frequency of calibration is 14 days.

9. Method Comparison

Eighty (80) deidentified, unaltered clinical oral fluid samples collected by Ouantisal II Oral Fluid Collection Devices were obtained from clinical research facilities, analyzed for oxycodone at assay cutoff with the SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay in both qualitative and semi-quantitative modes and compared to Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) results. The instruments used were the Beckman Coulter AU480 chemistry analyzer and an Agilent 6430 Liquid Chromatography-Tandem Mass Spectrometry. The data demonstrate that the design goal of greater than 95% agreement was achieved. Method comparison test results in qualitative and semi-quantitative modes are presented from Tables 16 to 18.

LC-MS/MS Oxycodone Concentration
ImmunoassayResult< 15 ng/mL(less than -50%cutoff)15 – 29 ng/mL(between -50%cutoff and cutoff)30 – 45 ng/mL(between cutoff and+50% cutoff)> 45 ng/mL(greater than +50%cutoff)Agreement (%)
Qual.Positive00535100% (40/40)
Negative36400100% (40/40)
Semi-Quant.Positive00535100% (40/40)
Negative36400100% (40/40)

Table 16. Method Comparison - Quantisal

Table 17. Method Comparison - Quantisal II "A"

ImmunoassayResultLC-MS/MS Oxycodone Concentration
< 15 ng/mL(less than-50%cutoff)15 – 29 ng/mL(between -50%cutoff andcutoff)30 – 45 ng/mL(betweencutoff and+50% cutoff)> 45 ng/mL(greaterthan +50%cutoff)Agreement (%)
Qual.Positive00535100% (40/40)
Negative36400100% (40/40)
Semi-Quant.Positive00535100% (40/40)
Negative36400100% (40/40)

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, hexagon-shaped background. The word is written in all capital letters and is centered on the background. The background is a bright red color and has a slight gradient effect. The image is simple and clean, with a focus on the word "IMMUNALYSIS".

LC-MS/MS Oxycodone Concentration
ImmunoassayResult< 15 ng/mL(less than-50%cutoff)15 – 29 ng/mL(between -50%cutoff andcutoff)30 - 45 ng/mL(betweencutoff and+50% cutoff)> 45 ng/mL(greaterthan +50%cutoff)Agreement (%)
Positive00535100% (40/40)
Qual.Negative36400100% (40/40)
Semi-Positive00535100% (40/40)
Quant.Negative36400100% (40/40)

Table 18. Method Comparison – Quantisal II "B"

H. CONCLUSION

The information provided in this pre-market notification demonstrates that the SEFRIA Oxycodone Oral Fluid Enzyme Immunoassay is substantially equivalent to the legally marketed predicate device for its intended use.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).