K Number
K160459

Validate with FDA (Live)

Date Cleared
2016-04-01

(42 days)

Product Code
Regulation Number
866.3990
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FilmArray Gastrointestinal (GI) Panel is a qualitative multiplexed in vitro diagnostic test intended for use with FilmArray systems. The FilmArray GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/Shigella pathotypes), parasites, and viruses are identified using the FilmArray GI Panel:

  • Campylobacter (C. jejuni/C. coli/C. upsaliensis)
  • Clostridium difficile (C. difficile) toxin A/B
  • Plesiomonas shigelloides
  • Salmonella
  • Vibrio (V. parahaemolyticus/V. vulnificus/V. cholerae) including specific identification of Vibrio cholerae
  • Yersinia enterocolitica
  • Enteroaggregative Escherichia coli (EAEC)
  • Enteropathogenic Escherichia coli (EPEC)
  • Enterotoxigenic Escherichia coli (ETEC) lt/st
  • Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of the E. coli 0157 serogroup within STEC)
  • Shigella/Enteroinvasive Escherichia coli (EIEC)
  • Cryptosporidium
  • Cyclospora cayetanensis
  • Entamoeba histolytica
  • Giardia lamblia (also known as G. intestinalis and G. duodenalis)
  • Adenovirus F 40/41
  • Astrovirus
  • Norovirus GI/GII
  • Rotavirus A
  • Sapovirus (Genogroups I, II, IV, and V)

The FilmArray GI Panel is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out co-infection with organisms not included in the FilmArray GI Panel. The agent detected may not be the definite cause of the disease.

Concomitant culture is necessary for organism recovery and further typing of bacterial agents.

This device is not intended to monitor or guide treatment for C. difficile infection.

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for E. coli O157, Plesiomonas shigelloides, Yersinia enterocolitica, Astrovirus, and Rotavirus A were established primarily with retrospective clinical specimens.

Performance characteristics for Entamoeba histolytica, and Vibrio (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) were established primarily using contrived clinical specimens.

Negative FilmArray GI Panel results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or noninfectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.

Device Description

The FilmArray Gastrointestinal (GI) Panel is a multiplex nucleic acid test designed to be used with FilmArray systems. The FilmArray GI pouch contains freeze-dried reagents to perform nucleic acid purification and nested, multiplex PCR with DNA melt analysis. The FilmArray Gastrointestinal (GI) Panel simultaneously conducts 22 tests for the identification of GI pathogens from stool specimens collected in Cary Blair transport medium (Table 1). Results from the FilmArray GI Panel test are available within about one hour.

A test is initiated by loading Hydration Solution into one port of the FilmArray pouch and a stool sample (in Cary Blair transport medium) mixed with the provided Sample Buffer into the other port of the FilmArray GI pouch and placing it in the FilmArray Instrument. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and sample/Sample Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArray Software guides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

The FilmArray instrument contains a coordinated system of inflatable bladders and seal points. which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

Nucleic acid extraction occurs within the FilmArray pouch using mechanical and chemical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, a nested multiplex PCR is executed in two stages. During the first stage, a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction is performed. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green® Plus, BioFire Defense, LLC). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in each well of the array. At the conclusion of the 2rd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the array captures fluorescent images of the PCR2 reactions and software interprets the data.

The FilmArray software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the FilmArray Gastrointestinal Panel (GI) for use with FilmArray Torch, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria (e.g., a specific percentage for sensitivity or specificity) for the FilmArray GI Panel. However, the study aims to demonstrate reproducible detection at LoD (Limit of Detection) for each analyte and agreement with expected negative results.

Acceptance Criteria (Implied)Reported Device Performance (FilmArray Torch)
Reproducible detection at LoD (1x LoD)≥99.0% detection for most analytes (at 1x LoD)
Agreement with expected negative results>99.0% for each analyte

Note: For Giardia lamblia, the performance was <95% detection at LoD, but this was attributed to lower than expected organism levels in the prepared samples, as comparison testing on original FilmArray systems also yielded <95%.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 96 total replicates per analyte. (Each analyte was tested 32 replicates per system on 3 systems).
  • Data Provenance: The samples were "contrived samples containing each FilmArray GI analyte at low positive levels (1x LoD)". This indicates that the data was generated experimentally in a controlled environment, rather than being derived from real patient samples (retrospective or prospective). The country of origin is not specified but implied to be the US given the submission to the FDA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. The ground truth for the contrived samples was established by the known concentration of the analytes (low positive levels, 1x LoD) that were spiked into the samples. Expert review for ground truth is typically relevant for studies involving real patient data with diagnostic uncertainty.

4. Adjudication Method for the Test Set

Not applicable. As the test set consisted of contrived samples with known concentrations, there was no need for an adjudication method. The expected outcome for each sample was predetermined (detected or not detected).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This study focuses on the analytical performance of the device (FilmArray GI Panel on FilmArray Torch) compared to the predicate device, not on human reader performance with or without AI assistance. The device is an in vitro diagnostic test, not an imaging or interpretation aid.

6. If a Standalone (algorithm only without human-in-the-loop performance) was done

Yes, this study is inherently a standalone performance evaluation of the FilmArray GI Panel on the FilmArray Torch instrument. The device is fully automated, performing nucleic acid purification, PCR, and result interpretation without human intervention in the result generation process. "Automated test interpretation and report generation. User cannot access raw data" is explicitly stated.

7. The Type of Ground Truth Used

The ground truth used for this specific study was known concentrations of analytes in contrived samples. This means the samples were engineered to contain a specific target (or not) at a specific concentration (1x LoD).

8. The Sample Size for the Training Set

The document does not explicitly mention a separate "training set" or its size for the FilmArray GI Panel on FilmArray Torch. This study is focused on validating the performance of the existing FilmArray GI Panel (reagent kit unchanged) when used with a new instrument system (FilmArray Torch). The original FilmArray GI Panel (predicate device K143005) would have undergone its own development and validation, which would likely have involved training sets, but that information is not provided in this specific document.

9. How the Ground Truth for the Training Set Was Established

Not applicable in this document, as a training set for the FilmArray GI Panel on FilmArray Torch is not described. If a training set was used for the development of the original FilmArray GI Panel, its ground truth would have been established through methods like:

  • Clinical culture: For bacterial and fungal pathogens.
  • PCR or sequencing: For viruses and parasites, often considered the gold standard for nucleic acid detection.
  • Electron microscopy or immunoassay: For some viral targets.
  • Expert consensus: In cases of diagnostic ambiguity, though less common for molecular diagnostics where a positive/negative result is expected from a reference method.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around a stylized caduceus symbol. The caduceus is depicted with three intertwined figures, representing health and well-being. The text "HEALTH & HUMAN SERVICES - USA" is visible around the top portion of the seal.

April 1, 2016

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

BIOFIRE DIAGNOSTICS, LLC KRISTEN KANACK VICE PRESIDENT, REGULATED PRODUCTS & CLINICAL AFFAIRS 390 WAKARA WAY SALT LAKE CITY UT 84108

Re: K160459

Trade/Device Name: FilmArray Gastrointestinal Panel (GI) for use with FilmArray Torch Regulation Number: 21 CFR 866.3990 Regulation Name: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assay Regulatory Class: II Product Code: PCH Dated: February 18, 2016 Received: February 19, 2016

Dear Dr. Kanack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Steven R. Gitterman -S

For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160459

Device Name FilmArray Gastrointestinal Panel (GI)

Indications for Use (Describe)

The FilmArray Gastrointestinal (GI) Panel is a qualitative multiplexed in vitro diagnostic test intended for use with FilmArray systems. The FilmArray GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/Shigella pathotypes), parasites, and viruses are identified using the FilmArray GI Panel:

  • · Campylobacter (C. jejuni/C. coli/C. upsaliensis)
  • · Clostridium difficile (C. difficile) toxin A/B
  • Plesiomonas shigelloides
  • · Salmonella
  • · Vibrio (V. parahaemolyticus/V. cholerae) including specific identification of Vibrio cholerae
  • · Yersinia enterocolitica
  • · Enteroaggregative Escherichia coli (EAEC)
  • · Enteropathogenic Escherichia coli (EPEC)
  • · Enterotoxigenic Escherichia coli (ETEC) lt/st
  • · Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of the E. coli 0157 serogroup within STEC)
  • · Shigella/Enteroinvasive Escherichia coli (EIEC)
  • · Cryptosporidium
  • · Cyclospora cayetanensis
  • Entamoeba histolytica
  • · Giardia lamblia (also known as G. intestinalis and G. duodenalis)
  • · Adenovirus F 40/41
  • Astrovirus
  • Norovirus GI/GII
  • Rotavirus A
  • · Sapovirus (Genogroups I, II, IV, and V)

The FilmArray GI Panel is indicated as an aid in the diagnosis of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out co-infection with organisms not included in the FilmArray GI Panel. The agent detected may not be the definite cause of the disease.

Concomitant culture is necessary for organism recovery and further typing of bacterial agents.

This device is not intended to monitor or guide treatment for C. difficile infection.

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study,

performance characteristics for E. coli 0157, Plesiomonas shigelloides, Yersinia enterocolitica, Astrovirus, and Rotavirus A were established primarily with retrospective clinical specimens.

Performance characteristics for Entamoeba histolytica, and Vibrio (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) were established primarily using contrived clinical specimens.

Negative FilmArray GI Panel results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute

{3}------------------------------------------------

gastroenteritis in the context of outbreaks.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Special 510(k) Summary BioFire Diagnostics, LLC

FilmArray Gastrointestinal (GI) Panel for use with FilmArray Torch

Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitted by:

BioFire Diagnostics, LLC 390 Wakara Way Salt Lake City, UT 84108

Contact:

Kristen J. Kanack, Ph.D. Telephone: 801-736-6354, ext. 330 Fax: 801-588-0507 Email: Kristen.kanack@biofiredx.com

Date Submitted:

February 18, 2016

Trade Name: FilmArray Gastrointestinal (GI) Panel

Classification Name:

Gastrointestinal microorganism multiplex nucleic acid-based assay (21 CFR 866.3990)

Predicate Device: K143005 – FilmArray Gastrointestinal (GI) Panel

Intended Use:

The FilmArray Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems. The FilmArray GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/Shigella pathotypes), parasites, and viruses are identified using the FilmArray GI Panel:

  • Campylobacter (C. jejuni/C. coli/C. upsaliensis) ●
  • Clostridium difficile (C. difficile) toxin A/B
  • Plesiomonas shigelloides
  • Salmonella
  • . Vibrio (V. parahaemolyticus/V. vulnificus/V. cholerae) including specific identification of Vibrio cholerae

BioFire Diagnostics, LLC Special 510(k) FilmArray Torch GI Panel

{5}------------------------------------------------

  • Yersinia enterocolitica ●
  • Enteroaggregative Escherichia coli (EAEC)
  • Enteropathogenic Escherichia coli (EPEC) ●
  • Enterotoxigenic Escherichia coli (ETEC) lt/st
  • Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific ● identification of the E. coli 0157 serogroup within STEC)
  • Shigella/Enteroinvasive Escherichia coli (EIEC) ●
  • Cryptosporidium
  • Cyclospora cayetanensis
  • Entamoeba histolytica
  • Giardia lamblia (also known as G. intestinalis and G. duodenalis)
  • Adenovirus F 40/41 ●
  • Astrovirus
  • Norovirus GI/GII
  • Rotavirus A
  • Sapovirus (Genogroups I, II, IV, and V) ●

The FilmArray GI Panel is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out co-infection with organisms not included in the FilmArray GI Panel. The agent detected may not be the definite cause of the disease.

Concomitant culture is necessary for organism recovery and further typing of bacterial agents.

This device is not intended to monitor or guide treatment for C. difficile infection.

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for E. coli O157, Plesiomonas shigelloides, Yersinia enterocolitica, Astrovirus, and Rotavirus A were established primarily with retrospective clinical specimens.

Performance characteristics for Entamoeba histolytica, and Vibrio (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) were established primarily using contrived clinical specimens.

Negative FilmArray GI Panel results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or noninfectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.

Device Description:

The FilmArray Gastrointestinal (GI) Panel is a multiplex nucleic acid test designed to be used with FilmArray systems. The FilmArray `GI pouch contains freeze-dried reagents to perform

BioFire Diagnostics, LLC Special 510(k) FilmArray Torch GI Panel

{6}------------------------------------------------

nucleic acid purification and nested, multiplex PCR with DNA melt analysis. The FilmArray Gastrointestinal (GI) Panel simultaneously conducts 22 tests for the identification of GI pathogens from stool specimens collected in Cary Blair transport medium (Table 1). Results from the FilmArray GI Panel test are available within about one hour.

Table 1. Bacteria, Viruses, Diarrheagenic E. coli/Shigella, and Parasites Detected by the FilmArray GI Panel
BacteriaViruses
Campylobacter (C. jejuni/C. coli/C. upsaliensis)Adenovirus F 40/41
Clostridium difficile (toxin A/B)Astrovirus
Plesiomonas shigelloidesNorovirus GI/GII
SalmonellaRotavirus A
Vibrio(V. parahaemolyticus/V. vulnificus/V. cholerae)Sapovirus (Genogroups I, II, IV, and V)
Vibrio cholerae
Yersinia enterocolitica
Diarrheagenic E. coli/ShigellaParasites
Enteroaggregative E. coli (EAEC)Cryptosporidium
Enteropathogenic E. coli (EPEC)Cyclospora cayetanensis
Enterotoxigenic E. coli (ETEC) lt/stEntamoeba histolytica
Shiga toxin-producing E. coli (STEC) stx1/stx2Giardia lamblia
E. coli O157
Shigella/Enteroinvasive E. coli (EIEC)

A test is initiated by loading Hydration Solution into one port of the FilmArray pouch and a stool sample (in Cary Blair transport medium) mixed with the provided Sample Buffer into the other port of the FilmArray GI pouch and placing it in the FilmArray Instrument. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and sample/Sample Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArray Software guides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

The FilmArray instrument contains a coordinated system of inflatable bladders and seal points. which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

Nucleic acid extraction occurs within the FilmArray pouch using mechanical and chemical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, a nested multiplex PCR is executed in two stages. During the first stage, a single, large volume, highly multiplexed reverse transcription

{7}------------------------------------------------

PCR (rt-PCR) reaction is performed. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green® Plus, BioFire Defense, LLC). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in each well of the array. At the conclusion of the 2rd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the array captures fluorescent images of the PCR2 reactions and software interprets the data.

The FilmArray software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

Device Comparison:

The purpose of this submission is to add FilmArray Torch as an additional instrument system for use with the FilmArray Gastrointestinal (GI) Panel. There have been no changes to the previously cleared FilmArray GI Panel reagent kit itself. In order to increase throughput capacity, the FilmArray 2.0 was modified to develop the FilmArray Torch by organizing densely packaged instruments (now called FilmArray Torch Modules) into a "tower" configuration having the computer with touchscreen interface incorporated into the system base. Pouches are now inserted horizontally rather than from the top. This configuration allows the FilmArray Torch Modules to be stacked directly on top of each other to minimize system footprint.

The following table compares the FilmArray GI Panel for use with the FilmArray Torch to the previously cleared FilmArray GI Panel for Use with the FilmArray 2.0 (K143005) The table outlines the similarities and differences between the two systems.

ElementModified Device:FilmArray Gastrointestinal (GI) Panelfor use with the FilmArray TorchPredicate:FilmArray Gastrointestinal (GI) Panel(K143005)
OrganismsDetectedCampylobacter (C. jejuni/C. coli/C. upsaliensis), Clostridium difficile (C. difficile) toxin A/B, Plesiomonas shigelloides, Salmonella, Vibrio (V. parahaemolyticus/V. vulnificus/V. cholerae) including specific identification of Vibrio cholera, Yersinia enterocolitica, Enteroaggregative Escherichia coli (EAEC), Enteropathogenic Escherichia coli (EPEC), Enterotoxigenic Escherichia coli (ETEC) lt/st, Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of the E. coli O157 serogroup within STEC), Shigella/Enteroinvasive Escherichia coliSame
ElementModified Device:FilmArray Gastrointestinal (GI) Panelfor use with the FilmArray TorchPredicate:FilmArray Gastrointestinal (GI) Panel(K143005)
(EIEC),Cryptosporidium, Cyclosporacayetanensis, Entamoeba histolytica,Giardia lamblia (also known as G.intestinalis and G. duodenalis), AdenovirusF 40/41, Astrovirus, Norovirus GI/GII,Rotavirus A, and Sapovirus (Genogroups I,II, IV, and V).
AnalyteRNA/DNASame
Specimen TypesHuman stool in Cary Blair transportmediumSame
TechnologicalPrinciplesNested multiplex RT-PCR followed by highresolution melting analysis to confirmidentity of amplified product.Same
InstrumentationSingle instrument FilmArray System,FilmArray 2.0 System, orFilmArray Torch SystemSingle instrument FilmArray System orFilmArray 2.0 System
Instrument-SoftwareCommunicationCommunication for multiple FilmArrayTorch Modules travels via Ethernetcable/port.Same (multiple instruments)
Time to resultAbout 1 hourSame
TestInterpretationAutomated test interpretation and reportgeneration. User cannot access raw data.Same
ReagentHydration andSample LoadingFilmArray Injection Vial-based loadingprocedureSame
SamplePreparationMethodSample Processing is automated in theFilmArray GI pouch.Same
Reagent StorageReagents are stored at room temperature.Same
ControlsTwo controls are included in each reagentpouch to control for sample processing andboth stages of PCR and melt analysis.Same
UserComplexityModerate/LowSame

Table 2. Comparison of the FilmArray Gastrointestinal (GI) Panel for use with the FilmArray Torch to the current FilmArray Gastrointestinal (GI) Panel.

{8}------------------------------------------------

Performance Characteristics of FilmArray GI Panel on FilmArray Torch

Contrived samples containing each FilmArray GI analyte at low positive levels (1x LoD) were tested on three complete FilmArray Torch systems (12 FilmArray Torch Modules per system) over four days (32 replicates per analyte per system) for 96 total replicates per analyte. Reproducible detection at LoD was confirmed for each FilmArray GI Panel analyte with the expected Detected results for ≥99.0% of samples tested on FilmArray Torch systems. One exception was for Giardia lamblia, which was detected in <95% of replicate samples; comparison testing of these samples on original FilmArray systems was also <95%, indicating

{9}------------------------------------------------

lower than expected organism levels in the prepared samples. Agreement with expected negative results (Not Detected or N/A) on the Torch systems was >99.0% for each analyte.

Conclusion:

The intended use and fundamental scientific technology of the FilmArray GI Panel used with the modified device, FilmArray Torch, is unchanged from use of the legally marketed FilmArray GI Panel on FilmArray and FilmArray 2.0 systems. Non-clinical validation studies have established that the performance characteristics of FilmArray GI, including reproducibility, are substantially equivalent on FilmArray, FilmArray 2.0, and FilmArray Torch. These data demonstrate that the device performs as well as the predicate device.

§ 866.3990 Gastrointestinal microorganism multiplex nucleic acid-based assay.

(a)
Identification. A gastrointestinal microorganism multiplex nucleic acid-based assay is a qualitativein vitro diagnostic device intended to simultaneously detect and identify multiple gastrointestinal microbial nucleic acids extracted from human stool specimens. The device detects specific nucleic acid sequences for organism identification as well as for determining the presence of toxin genes. The detection and identification of a specific gastrointestinal microbial nucleic acid from individuals exhibiting signs and symptoms of gastrointestinal infection aids in the diagnosis of gastrointestinal infection when used in conjunction with clinical evaluation and other laboratory findings. A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.(b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assays for Detection and Identification of Microorganisms and Toxin Genes from Human Stool Specimens.” For availability of the guideline document, see § 866.1(e).