K Number
K251902

Validate with FDA (Live)

Date Cleared
2025-09-17

(89 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Vitaform Procedural Masks are intended to be worn to protect both the patient and the healthcare worker from transfer of microorganisms, body fluids, and particulate material. They are single-use and intended for use in infection control practices to reduce potential exposure to blood and body fluids.

Device Description

Vitaform Procedural Mask (henceforth referred to as "Vitaform")

The Vitaform Procedural Mask is a single-use, three-layered, fish-shaped surgical mask with ear loops and an aluminum nose piece. The inner and outer layers are made of spunbond polypropylene, and the middle filter layer consists of meltblown polypropylene. The ear loops are made of PET and Spandex material, and worn around the ears to keep the mask close to the face. The nose piece is a flexible aluminum strip that is fitted over the nose so that the mask conforms better to the user's face.

The Vitaform Procedural Mask is provided in the color blue. It is non-sterile and intended to be a single-use, disposable medical device.

Vitaform Procedural Mask with Shield (henceforth referred to as "Vitaform-FS")

The Vitaform Procedural Mask with Shield is a single-use, fish-shaped surgical mask with ear loops and an aluminum nose piece. The inner and outer layers are made of spunbond polypropylene, and the middle filter layer consists of meltblown polypropylene. The insertion layer that provides structural support is made of thermal-bonded polypropylene. The ear loops are made of PET and Spandex material, and is worn around the ears to keep the mask close to the face. The nose piece is a flexible aluminum strip that is fitted over the nose so that the mask conforms better to the user's face. The mask also contains a face shield (FS) made from a polyethylene terephthalate film, and an anti-reflective flap. The face shield is welded to the upper half of the mask to cover the upper part of the face.

The Vitaform Procedural Mask with Shield is provided in the color blue. It is non-sterile and intended to be a single-use, disposable medical device.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Vitacore Industries Inc.

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

September 17, 2025

Vitacore Industries Inc.
Yuxuan Fan
R&D Project Manager
1615 Kebet Way
Port Coquitlam, BC V3C 5W9
Canada

Re: K251902
Trade/Device Name: Vitaform Procedural Mask - Blue (Vitaform Blue); Vitaform Procedural Mask with Shield (Vitaform-FS)
Regulation Number: 21 CFR 878.4040
Regulation Name: Surgical Apparel
Regulatory Class: Class II
Product Code: FXX
Dated: June 20, 2025
Received: June 20, 2025

Dear Yuxuan Fan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not

Page 2

K251902 - Yuxuan Fan
Page 2

required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Page 3

K251902 - Yuxuan Fan
Page 3

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

ALLAN GUAN -S

For
Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251902

Device Name: Vitaform Procedural Mask - Blue (Vitaform Blue); Vitaform Procedural Mask with Shield (Vitaform-FS)

Indications for Use (Describe):
Vitaform Procedural Masks are intended to be worn to protect both the patient and the healthcare worker from transfer of microorganisms, body fluids, and particulate material. They are single-use and intended for use in infection control practices to reduce potential exposure to blood and body fluids.

Type of Use (Select one or both, as applicable):
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number"

FORM FDA 3881 (8/23)
Page 1 of 1

Page 5

510(k) Summary

K251902

Submitter

Sponsor and Correspondent Information

Submitted By: Vitacore Industries Inc.
Contact person: Yuxuan Fan
Email: yuxuan.f@vitacore.com
Address: Vitacore Industries Inc.
1615 Kebet Way, Port Coquitlam BC,
V3C 5W9 Canada
Phone: (780) 710-6872
Date Prepared: September 17th, 2025

Device

Device Information

Device Name: Vitaform Procedural Mask - Blue (Vitaform Blue); Vitaform Procedural Mask with Shield (Vitaform-FS)
Device Classification Name: Surgical Apparel
Product Code: FXX
Regulation Number: 21 CFR 878.4040
Classifications: Class II, 21 CFR 878.4040 - Surgical Apparel

Predicate device

Primary Predicate: Magnum Surgical Mask, K220670
Additional Predicate: 3M™ High Fluid-Resistant Procedure Mask with Face Shield (Model# 1840FS), K191355

Device Description

Vitaform Procedural Mask (henceforth referred to as "Vitaform")

The Vitaform Procedural Mask is a single-use, three-layered, fish-shaped surgical mask with ear loops and an aluminum nose piece. The inner and outer layers are made of spunbond polypropylene, and the middle filter layer consists of meltblown polypropylene. The ear loops are made of PET and Spandex material, and worn around the ears to keep the mask close to the face. The nose piece is a flexible aluminum strip that is fitted over the nose so that the mask conforms better to the user's face.

The Vitaform Procedural Mask is provided in the color blue. It is non-sterile and intended to be a single-use, disposable medical device.

Vitaform Procedural Mask with Shield (henceforth referred to as "Vitaform-FS")

The Vitaform Procedural Mask with Shield is a single-use, fish-shaped surgical mask with ear loops and an aluminum nose piece. The inner and outer layers are made of spunbond polypropylene, and

Page 6

the middle filter layer consists of meltblown polypropylene. The insertion layer that provides structural support is made of thermal-bonded polypropylene. The ear loops are made of PET and Spandex material, and is worn around the ears to keep the mask close to the face. The nose piece is a flexible aluminum strip that is fitted over the nose so that the mask conforms better to the user's face. The mask also contains a face shield (FS) made from a polyethylene terephthalate film, and an anti-reflective flap. The face shield is welded to the upper half of the mask to cover the upper part of the face.

The Vitaform Procedural Mask with Shield is provided in the color blue. It is non-sterile and intended to be a single-use, disposable medical device.

Indications for Use

Vitaform Procedural Masks are intended to be worn to protect both the patient and the healthcare worker from transfer of microorganisms, body fluids, and particulate material. They are single-use and intended for use in infection control practices to reduce potential exposure to blood and body fluids.

Comparison of the Technological Characteristics with the Predicates

The table below compares the construction, technology, design, claims, and intended use of the proposed (Vitaform and Vitaform-FS) and predicate devices.

Table 1 - Comparison of Proposed (Vitaform and Vitaform-FS) and Predicate Devices

DeviceProposed Devices Vitaform, Vitaform-FSPrimary PredicateAdditional PredicateResults
510K #K251902K220670K191355Different
Product CodeFXXFXXFXXSame
Product OwnerVitacore Industries Inc.Magnum Health and Safety Pvt Ltd3M Health CareDifferent
Intended UseVitaform Procedural Masks are intended to be worn to protect both the patient and the healthcare worker from transfer of microorganisms, body fluids, and particulate material. They are single-use and intended for use in infection control practices to reduce potential exposure to blood and body fluids.When properly worn, the surgical face masks are intended to protect both patient and healthcare workers from transfer of microorganisms, body fluids, and particulate material. The face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This device is disposable,3M™ High Fluid-Resistant Procedure Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. This is a single use, disposable device,Similar

Page 7

non-sterile and for single use only.provided non-sterile.
Regulation NumberClass II, 21 CFR 878.4040Class II, 21 CFR 878.4040Class II, 21 CFR 878.4040
Mask StyleVitaform: Fish-shaped mask with earloop Vitaform-FS: Fish-shaped mask with earloop and visor attachmentFlat - PleatedFlat - Pleated
Multiple LayersYesYesYes
Mask Color(s)Vitaform: Blue Vitaform-FS: BlueBlueGreen (Outer)

Materials and Specifications

| Outer layer | Vitaform Blue, Vitaform-FS: Blue spunbond polypropylene | Spunbond polypropylene | Polypropylene Spunbond | Same |
| Insertion | Vitaform-FS only: Polypropylene Thermal-bonded | N/A | Polypropylene Spunbond | Different |
| Filter layer | Polypropylene Meltblown | Meltblown filter media | Polypropylene Meltblown | Similar |
| Inner layer | Polypropylene Spunbond | Spunbond polypropylene | Polypropylene Thermal-bonded | Different |
| Earloop | PET/Spandex | Knitted elastic | Spandex elastic cord (polyurethane core with polyethylene terephthalate / nylon cover) | Similar |
| Nose piece | Aluminum | PVC coated aluminum wire | Polyethylene coated steel wire | Different |
| Anti-Glare layer | Vitaform-FS only: Polypropylene Spunbond | N/A | N/A | Different |
| Face Shield | Vitaform-FS only: Polyethylene Terephthalate Film | N/A | Polyethylene Terephthalate Film | Similar |
| Length | 215±15MM/8.5"±0.6" | 172±3MM/6.9"±0.1" | 175±5MM/6.9" ±0.2" | Different |
| Width | 85±2MM/3.3"±0.08" | 95±3MM/3.7"±0.1" | 89±7.6MM/3.5" ±0.3" | Different |
| ASTM 2100 Level | Level 3 | Level 2 and 3 | Level 3 | Similar |
| ASTM F2299 Particulate filtration efficiency | At least 98% | At least 98% | At least 98% | Same |
| ASTM F2101 Bacterial filtration | At least 98% | At least 98% | At least 98% | Same |

Page 8

efficiency
Differential pressureLess than 6.0 mmH20/cm²Less than 6.0 mmH20/cm²Less than 6.0 mmH20/cm²Same
ASTM F1862 Fluid resistancePass at 160 mmHgLevel 2: Pass at 120 mmHg Level 3: Pass at 160 mmHgPass at 160 mmHgSimilar
16 CFR 1610 FlammabilityClass IClass IClass ISame
SterilityNon-sterileNon-sterileNon-sterileSame
Prescription vs. OTCOTCOTCOTCSame
UseSingle use, disposableSingle use, disposableSingle use, disposableSame
Biocompatibility CytotoxicityUnder the conditions of the study, non-cytotoxicUnder the conditions of the study, non-cytotoxicUnder the conditions of the study, non-cytotoxicSame
Biocompatibility IrritationUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritantSame
Biocompatibility SensitizationUnder the conditions of the study, non-sensitizingUnder the conditions of the study, non-sensitizingUnder the conditions of the study, non-sensitizingSame
Shelf LifePredetermined Change Control Plan for future three year shelf lifeNot specifiedNot specifiedDifferent

Test results for all the subject devices were similar to results for the ASTM F2100 Level 3 predicates, and exceed the test results for the ASTM F2100 Level 2 variant of the primary predicate. While the subject device is a fish-shape design and the predicates are flat pleated, the devices meet the same performance criteria.

Summary of Non-Clinical Testing

Non-clinical tests were performed on the Vitaform and Vitaform-FS to verify that the proposed device met all design specifications as the Magnum surgical mask (K220670). The test results showed that the proposed device complies with the standards and requirements set out in the Guidance for Industry and FDA staff: Surgical Masks – Premarket Notification [510K] Submission issued in 2004.

Performance Tests

  • ASTM F1862, Standard Test Method for Resistance of Medical Face Masks To Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume At A Known Velocity)
  • EN 14683, Medical Face Masks–Requirements and Test Methods
  • ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus
  • ASTM F2299, Standard test method for determining the initial efficiency of materials used in medical face masks to penetration by particulate using latex spheres

Page 9

  • 16 CFR 1610, Standard for the Flammability of clothing textiles
  • ISO 10993-5: 2009 Biological Evaluation of Medical Devices–Part 5: Tests For In Vitro Cytotoxicity
  • ISO 10993-10: 2010 Biological Evaluation of Medical Devices–Part 10: Tests For Irritation And Skin Sensitization
  • ISO 10993-23:2021 Biological Evaluation of Medical Devices, Part 23: Tests for Irritation

Table 2 –Performance Testing results.

Performance TestsPurposeCriteriaResult
Particulate Filtration Efficiency as per ASTM F2299To determine the particulate filtration efficiencyPass ≥ 98%Pass
Bacterial Filtration Efficiency as per ASTM F2101To determine the bacterial filtration efficiencyPass ≥ 98%Pass
Blood Penetration Resistance as per ASTM F1862To determine the resistance to synthetic blood penetrationPass at 160 mm HgPass
Differential Pressure as per EN 14683 / MIL M36954CTo determine the resistance to airflowPass at less than 6.0 mm H₂O/cm²Pass
Flammability as per 16 CFR 1610To evaluate flammabilityClass IPass
Biocompatibility Cytotoxicity as per ISO 10993-5To evaluate cytotoxic potentialUnder the conditions of the study, non-cytotoxicPass
Biocompatibility Sensitization as per ISO 10993-10To evaluate sensitization potentialUnder the conditions of the study, not a sensitizerPass
Biocompatibility Irritation as per ISO 10993-23To evaluate the irritation potentialUnder the conditions of the study, not an irritantPass

Predetermined Change Control Plan for addition of three-year shelf life to label

| Verify Particulate Filtration Efficiency as per ASTM F3502 after three-years real time aging | To verify the particulate filtration efficiency after three years real time aging | Pass ≥ 85% | Not yet conducted |
| Verify Bacterial Filtration Efficiency as per ASTM F2101 after three-years real time aging | To verify the bacterial filtration efficiency after three years real time aging | Pass ≥ 98% | Not yet conducted |
| Verify Blood Penetration Resistance as per | To verify the resistance to synthetic blood penetration after three years real time aging | Pass at 160 mm Hg | Not yet conducted |

Page 10

ASTM F1862 after three-years real time aging
Verify Differential Pressure as per EN 14683 / MIL M36954CTo verify the resistance to airflow after three years real time agingPass at less than 6.0 mm H₂O/cm²Not yet conducted

Summary of Clinical Testing

No clinical testing was used in support of this submission.

Conclusion

The conclusion drawn from the tests performed demonstrates that the Vitaform Procedural Mask - Blue (Vitaform Blue); Vitaform Procedural Mask with Shield (Vitaform-FS) are as safe, as effective, and perform as well as or better than the legally marketed predicate devices, Magnum Surgical Mask (K220670) and 3M™ High Fluid-Resistant Procedure Mask with Face Shield (K191355).

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.