K Number
K191355

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-09-11

(113 days)

Product Code
Regulation Number
878.4040
Age Range
0 - 21
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 3M™ High Fluid-Resistant Surgical Mask and 3M™ High Fluid-Resistant Procedure Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. The face mask is single use, disposable device, provided non-sterile.

Device Description

The 3M™ High Fluid-Resistant Surgical Mask and 3M™ High Fluid-Resistant Procedure Mask are composed of four-layers and are flat-pleated. The mask materials consist of an outer cover web (polypropylene spunbond, green), insertion layer (polypropylene, spunbond, white), filter web (polypropylene melt-blown, white) and inner cover web (polypropylene thermal-bonded, white). Each mask contains tie strings or ear loops to secure the users' mouth and face and includes a malleable nosepiece to provide a firm fit over the nose. The mask may also contain a face shield (FS) made from a polyethylene terephthalate film, which includes an antireflection strip. The face shield is adhered to the top edge of the mask to cover the upper part of the face. The mask is a single use, disposable device, provided non-sterile.

AI/ML Overview

This document is a 510(k) Summary for the 3M™ High Fluid-Resistant Surgical Mask and 3M™ High Fluid-Resistant Procedure Mask. It describes the device, its intended use, and a comparison to a predicate device to demonstrate substantial equivalence.

Here's an analysis of the provided information concerning acceptance criteria and supporting studies:

1. Table of Acceptance Criteria and Reported Device Performance:

The document includes a detailed "Technological Characteristics Product Barrier Specifications Per ASTM F2100 – Meets Level 3" table. This table serves as the primary source for acceptance criteria and the reported performance of the subject device.

Acceptance Criterion (Test)Acceptance Limit (Requirement)Subject Device Performance (3M™ High Fluid-Resistant Surgical Mask & 3M™ High Fluid-Resistant Procedure Mask)Predicate Device Performance (Kimberly-Clark* KC300 Surgical Mask (K131879))
Particulate Filtration Efficiency (PFE) (ASTM F2299)≥98% @ 0.1 micron32/32 Passed at ≥98% @ 0.1 micronPass at ≥98% @ 0.1 micron
Fluid Resistance (ASTM F1862)160mm Hg32/32 Passed at 160mm HgFluid Resistant 160mm Hg
Bacterial Filtration Efficiency (BFE) (ASTM F2101)≥98%31/32 Passed at ≥98%Pass at ≥98%
Differential Pressure (MIL-M36954C)<5 mmH2O/cm²32/32 Passed at <5 mmH2O/cm²Pass at <5 mmH2O/cm²
Flammability (CFR 16 1610)Class 1 (≥3 Seconds burn time)5/5 Passed ≥3 Seconds burn time - Class 1Class 1
BiocompatibilityNon-cytotoxic, Non-sensitizing, Non-irritating (ISO 10993)Non-cytotoxic, Non-sensitizing, Non-irritatingNon-cytotoxic, Non-sensitizing, Non-irritating

2. Sample Size Used for the Test Set and Data Provenance:

The document provides the following sample sizes for the performance tests:

  • PFE, Fluid Resistance, BFE, Differential Pressure: "32/32 Passed" indicates a sample size of 32 units tested for each of these criteria.
  • Flammability: "5/5 Passed" indicates a sample size of 5 units tested.
  • Biocompatibility: No specific sample size is mentioned in the table; however, the standards ISO10993-1, ISO10993-5, and ISO10993-10 are referenced, which would involve in vitro and potentially in vivo testing with relevant samples.

The document does not explicitly state the provenance of the data (e.g., country of origin) or whether the studies were retrospective or prospective. Given the context of a 510(k) submission, these would typically be internal testing results conducted by the manufacturer or a contracted lab to demonstrate compliance with recognized standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:

The concept of "ground truth" and "experts" is not applicable in this context. This document pertains to the evaluation of a medical device (surgical/procedure masks) based on
physical, chemical, and biological performance characteristics as defined by established industry standards (ASTM, MIL-M, CFR, ISO). The "ground truth" for these tests is the quantitative measurement of the device's properties against predefined thresholds in the standards, not expert subjective assessment.

4. Adjudication Method for the Test Set:

Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies or studies involving human readers/interpreters to resolve discrepancies in diagnoses or interpretations. These methods are not relevant for the type of objective physical and biological performance testing described for surgical masks. The results are quantitative and pass/fail based on the specific test methods and criteria.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study was done. MRMC studies are relevant for evaluating the performance of diagnostic devices or AI algorithms where human readers interpret medical images or data. This document describes the testing of a passive barrier device (surgical masks), not a diagnostic or AI-driven device.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

No standalone (algorithm only) study was done. This concept is applicable to AI/ML software as a medical device (SaMD) where the algorithm performs a task independently. The device in question is a physical surgical mask and does not involve any algorithms or AI.

7. Type of Ground Truth Used:

As explained in point 3, the "ground truth" for this device's performance evaluation is based on established objective measurements against pre-defined thresholds specified in recognized international and national standards. These standards dictate the test methods and acceptance criteria for:

  • Particulate Filtration Efficiency (ASTM F2299)
  • Fluid Resistance (ASTM F1862)
  • Bacterial Filtration Efficiency (ASTM F2101)
  • Differential Pressure (MIL-M-36954C)
  • Flammability (16 CFR Part 1610)
  • Biocompatibility (ISO 10993 series)

The results indicate whether the device "Passed" or "Failed" these objective criteria in comparison to the predicate device.

8. Sample Size for the Training Set:

The concept of a "training set" is relevant for machine learning algorithms. This document describes the testing and comparison of a physical medical device (surgical mask). Therefore, there is no training set used in the context of this submission.

9. How the Ground Truth for the Training Set Was Established:

Since there is no training set for this type of device submission, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of an eagle, while the text on the right is written in a bold, sans-serif font. The word "FDA" is in a blue box.

September 11, 2019

3M Health Care Angie Draper Sr. Regulatory Affairs Associate 3M Center, 2510 Conway Ave., Building 275-5W-06 St. Paul, Minnesota 55144

Re: K191355

Trade/Device Name: 3M™ High Fluid-Resistant Surgical Mask and 3M™ High Fluid-Resistant Procedure Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 19, 2019 Received: August 20, 2019

Dear Angie Draper:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K191355

Device Name 3MTM High Fluid-Resistant Surgical Mask 3MTM High Fluid-Resistant Procedure Mask

Indications for Use (Describe)

The 3M™ High Fluid-Resistant Surgical Mask and 3M™ High Fluid-Resistant Procedure Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. The face mask is single use, disposable device, provided non-sterile.

  • 1835 3MTM High Fluid-Resistant Surgical Mask
  • 1835FS 3M™ High Fluid-Resistant Surgical Mask with Face Shield
  • 1840 3MTM High Fluid-Resistant Procedure Mask
  • 1840FS 3M™ High Fluid-Resistant Procedure Mask with Face Shield
Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

K191355

3M™ High Fluid-Resistant Surgical Mask3M™ High Fluid-Resistant Procedure Mask
Sponsor Information:3M Health Care2510 Conway Ave3M Center, Building 275-5W-06 St.Paul, MN 55144
Contact Person:Angie DraperTitle: Sr. Regulatory Affairs AssociatePhone Number: (651) 733-1179Fax Number: (651) 737-5320 email:amdraper01@mmm.com
Date of Summary:April 30, 2019
Common Name:Surgical Mask
Classification Name:Surgical Apparel
Proprietary Name:3M™ High Fluid-Resistant Surgical Mask3M™ High Fluid-Resistant Procedure Mask
Review Panel:General and Plastic Surgery
Product Code:FXX
Device Classification:Class II per (21 CFR §878.4040)
Predicate Device:Kimberly-Clark* KC300 Surgical Mask (K131879)
Intended Use:The 3M™ High Fluid-Resistant Surgical Mask and 3M™ HighFluid-Resistant Procedure Mask is intended to be worn to protect boththe patient and healthcare personnel from transfer of microorganisms,body fluids, and particulate material. These face masks are intended foruse in infection control practices to reduce potential exposure to bloodand body fluids. The face mask is single use, disposable device,provided non-sterile.

{4}------------------------------------------------

Available Model Numbers

18353M™ High Fluid-Resistant Surgical Mask
1835FS3M™ High Fluid-Resistant Surgical Mask with Face Shield
18403M™ High Fluid-Resistant Procedure Mask
1845FS3M™ High Fluid-Resistant Procedure Mask with Face Shield

Device Description:

The 3M™ High Fluid-Resistant Surgical Mask and 3M™ High Fluid-Resistant Procedure Mask are composed of four-layers and are flat-pleated. The mask materials consist of an outer cover web (polypropylene spunbond, green), insertion layer (polypropylene, spunbond, white), filter web (polypropylene melt-blown, white) and inner cover web (polypropylene thermal-bonded, white). Each mask contains tie strings or ear loops to secure the users' mouth and face and includes a malleable nosepiece to provide a firm fit over the nose. The mask may also contain a face shield (FS) made from a polyethylene terephthalate film, which includes an antireflection strip. The face shield is adhered to the top edge of the mask to cover the upper part of the face. The mask is a single use, disposable device, provided non-sterile.

This device is not made from Natural Rubber Latex.

Technological Characteristics:

The 3M™ High Fluid-Resistant Surgical Mask and 3M™ High Fluid-Resistant Procedure Mask are compared with the predicate device (KC300 Surgical Mask (K131879)). The results are shown below in the Technological Characteristics Comparison Table:

{5}------------------------------------------------

Item(S)Subject Device (K191355)3M™ High Fluid-Resistant Surgical MaskASTM Level 3Subject Device (K191355)3M™ High Fluid-ResistantProcedure Mask K191355ASTM Level 3PredicateDevice(K131879)KC300 Face MaskASTM Level 3Comparison
Intended Use/Indicationsfor Use3M™ High Fluid- ResistantSurgical Mask is intendedto be worn to protect boththe patient and healthcarepersonnel from transfer ofmicroorganisms, bodyfluids, and particulatematerial. These face masksare intended for use ininfection control practicesto reduce potential exposureto blood and body fluids.This is a single use,disposable device, providednon-sterile.3M™ High Fluid- ResistantProcedure Mask is intended tobe worn to protect both thepatient and healthcarepersonnel from transfer ofmicroorganisms, body fluids,and particulate material.These face masks are intendedfor use in infection controlpractices to reduce potentialexposure to blood and bodyfluids. This is a single use,disposable device, providednon-sterile.The Kimberly-ClarkKC300 Surgical Mask isintended to be worn toprotect both the patientand healthcare personnelfrom transfer ofmicroorganisms, bodyfluids, and particulatematerial. These facemasks are intended foruse in infection controlpractices to reduce thepotential exposure toblood and body fluids.This is a single use,disposable device,Same
Materials
Outer CoverWebPolypropylene Spunbond,greenPolypropylene Spunbond,greenPolyethylene/Polyesterwith pink and blue inkprintDifferent
InsertionPolypropylene Spunbond,whitePolypropylene Spunbond,whiteUnknownNot known
Filter Web(Middle)Polypropylene Meltblown,whitePolypropylene Meltblown,whitePolypropylene Spunbondand Polypropylene Melt-blownDifferent
Inner CoverWebPolypropylene Thermal-bonded, whitePolypropylene Thermal-bonded, whitePolyethylene/PolyesterDifferent
Nose WirePolyethylene Coated SteelWirePolyethylene Coated SteelWireUnknownNot known
Edge wrapPolypropylene Spunbond,white orPolyethyleneTerephthalate, whitePolypropylene Spunbond,white or PolyethyleneTerephthalate, whitePolyester Spunlace orPolypropylene SpunbondDifferent
Ear LoopsNot ApplicableSpandex elastic cord(polyurethane core withpolyethylene terephthalate/nylon cover)Polyester/Lycra KnittedDifferent
Tie StringsPolypropylene Spunbond,White or PolyethyleneTerephthalate, whiteNot ApplicablePolypropylene SpunbondDifferent
Design Features
ColorsGreen (Outer)Green (Outer)Multiple (Outer)Different
StyleFlat - PleatedFlat - PleatedFlat - PleatedSame

{6}------------------------------------------------

MultipleLayersYesYesYesSame
Single UseYesYesYesSame
Sterility
SterileNon-SterileNon-SterileNon-SterileSame
Dimensions
Length6.9" ± 0.2"6.9" ± 0.2"6.5" ± 0.75"Different
Width3.5" ± 0.3"3.5" ± 0.3"4" ± 0.75"Different
Technological Characteristics Product Barrier Specifications Per ASTM F2100 – Meets Level 3
ParticulateFiltrationEfficiency(PFE)32/32 Passed at ≥98% @ 0.1micronASTM F229932/32 Passed at ≥98% @ 0.1micronASTM F2299Pass at ≥98% @ 0.1micronASTM F2299Same
FluidResistance32/32 Passed at 160mm HgASTM F186232/32 Passed at 160mm HgASTM F1862Fluid Resistant 160mm HgASTM F1862Same
BacterialFiltrationEfficiency(BFE)31/32 Passed at ≥98%ASTM F210131/32 Passed at ≥98%ASTM F2101Pass at ≥98%ASTM F2101Same
DifferentialPressure32/32 Passed at <5 mmH2O/cm²MIL-M36954C32/32 Passed at <5 mmH2O/cm²MIL-M36954CPass at <5 mmH2O/cm²MIL-M36954CSame
Flammability5/5 Passed ≥3 Seconds burntime - Class 1CFR 16 16105/5 Passed ≥3 Seconds burntime - Class 1CFR 16 1610Class 1CFR 16 1610Same
Biocompatibility
ResultsNon-cytotoxic, Non-sensitizing, Non-irritatingNon-cytotoxic, Non-sensitizing, Non-irritatingNon-cytotoxic, Non-sensitizing, Non- irritatingSame

The following standards have been met for the 3M™ High Fluid-Resistant Surgical Mask and 3M™ High Fluid-Resistant Procedure Mask

3M High Fluid-Resistant Procedure Mask
ASTM F2100Standard Specification for Performance of Materials Used in Medical Face Masks
ASTM F1862Standard Test Method for Resistance of Medical Face Masks to Penetration bySynthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
ASTM F2299Standard Test Method for Determining the Initial Efficiency of Materials Used inMedical Face Masks to Penetration by Particulates Using Latex Spheres
ASTM F2101Standard Method for Evaluating the Bacterial Filtration Efficiency (BFE) ofMedical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
MIL-M- 36954CMilitary Specification, Mask, Surgical, Disposable
16 CFR Part 1610Standard for the Flammability of Clothing
ISO10993-1Biological evaluation of medical devices - Part 1: Evaluation and testing within arisk management process
ISO10993-5Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityof medical devices
ISO10993-10Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin

{7}------------------------------------------------

Conclusion:

The conclusion drawn from the nonclinical tests demonstrates that the subject devices in 510(K) submission K191355, the 3M™ High Fluid-Resistant Surgical Mask and 3M™ High Fluid-Resistant Procedure Mask are as safe, as effective, and performs as well as or better than the legally marketed predicate device K131879.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.