AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Merit Hydrophilic Guide Wire is intended to be used in the peripheral vascular system to facilitate the placement of devices during diagnostic and interventional procedures. These guide wires are not intended for Percutaneous Transluminal Coronary Angioplasty use or for use in the neurovasculature.

Device Description

The Splashwire Hydrophilic guide wire consists of a nitinol core wire with a ground tapered distal section. A polyurethane jacket which contains tungsten for radiopacity is applied over the core wire and a hydrophilic coating is applied over the polyurethane jacket. The surface of the jacket is uniform with both the distal and proximal ends fully coated. The wire is placed inside a multiple loop flush dispenser, also referred to as a hoop. The dispenser has a flush port which facilitates solution flushing through the hoop to hydrate the guide wire. The wire is placed so that the distal end of the wire comes out of the outer portion of the hoop. A J-straightener is placed on the other end of the hoop to introduce the wire into the catheter. The wire is sold sterile and is a single use device. The subject Splashwire Hydrophilic Guide Wire Line Extensions and the predicate Splashwire Hydrophilic Guide Wire share the device characteristics described above, with the subject guide wire incorporating the following additional modifications:

J-Tip Guide Wires:
The J Tip devices incorporate a distal tip formed into a J shape. These guide wires are provided with a 1.5 mm J Tip.

Long Taper Guide Wires:
The Long Taper Guide Wires incorporate an extended core wire taper length providing enhanced flexibility.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the Splashwire Hydrophilic Guide Wire (K251181) outlines the regulatory approval process and demonstrates substantial equivalence to a predicate device (K201595). However, it does not contain the specific information required to address your request regarding acceptance criteria and the detailed study that proves the device meets those criteria, particularly for an AI-enabled medical device.

The document describes a traditional medical device (a guide wire) and the non-clinical bench testing performed to ensure its safety and performance. This typically involves physical and mechanical property tests. It does not refer to AI/ML software, algorithms, or any studies involving human readers or ground truth expert consensus in the way you've outlined.

Therefore, I cannot provide a table of acceptance criteria and reported device performance for an AI-enabled device, nor details on sample sizes, expert ground truth establishment, or multi-reader multi-case studies, as this information is not present in the provided text.

Based on the provided document, the device is a physical medical device (a guide wire), not an AI-enabled one. The acceptance criteria and "study" described are standard bench tests for physical properties, not performance metrics for an AI algorithm.

Here's a breakdown of what can be inferred from the document regarding the device's performance, but it does not align with the parameters of an AI/ML study:

1. Table of Acceptance Criteria and Reported Device Performance:

The document states: "Testing was conducted on the subject Splashwire Hydrophilic Guide Wire in accordance with protocols based on requirements outlined in guidance's and industry standards and these were shown to meet the acceptance criteria that were determined to demonstrate substantial equivalence." And "All test results were comparable to the predicate Splashwire Hydrophilic Guide Wire and the subject Splashwire Hydrophilic Guide Wire met the predetermined acceptance criteria."

While specific numerical acceptance criteria and performance data are not provided in this summary, the document indicates that the following categories of non-clinical bench tests were successfully completed and met their predetermined acceptance criteria:

Acceptance Criteria Category (Test Type)Reported Device Performance (Implied)
Dimensional VerificationMet predetermined specifications (comparable to predicate)
Finished Wire SurfaceMet predetermined specifications (comparable to predicate)
Tip Tensile StrengthMet predetermined specifications (comparable to predicate)
Tip StiffnessMet predetermined specifications (comparable to predicate)
Tip FlexibilityMet predetermined specifications (comparable to predicate)
Prolapse TestMet predetermined specifications (comparable to predicate)
Torque StrengthMet predetermined specifications (comparable to predicate)
TorqueabilityMet predetermined specifications (comparable to predicate)
Coating IntegrityMet predetermined specifications (comparable to predicate)
Catheter CompatibilityMet predetermined specifications (comparable to predicate)
Fracture TestMet predetermined specifications (comparable to predicate)
Flexing TestMet predetermined specifications (comparable to predicate)
Simulated Use TestingMet predetermined specifications (comparable to predicate)
Kink ResistanceMet predetermined specifications (comparable to predicate)
Ancillary Device CompatibilityMet predetermined specifications (comparable to predicate)

2. Sample Size Used for the Test Set and Data Provenance:

This information is not available in the provided document. For a physical medical device, "test set" would refer to the number of units tested for each characteristic. The provenance of the data is not applicable in the context of clinical images or patient data; it would refer to the samples of the manufactured guide wire used for the bench tests. These are likely internal lab tests.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This information is not applicable to this type of device and study. "Ground truth" in this context would be derived from the physical properties of the device measured by laboratory instruments, not expert human interpretation of data.

4. Adjudication Method for the Test Set:

This information is not applicable as there are no human interpretations or adjudications involved in the physical bench tests.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

This information is not applicable. MRMC studies are relevant for software that aids human interpretation (e.g., AI for radiology). This device is a physical guide wire.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable. This describes the performance of a physical device, not an algorithm.

7. The Type of Ground Truth Used:

The "ground truth" for this device is based on measurements of its physical and mechanical properties according to established industry standards and FDA guidance for guide wires (e.g., ISO standards, FDA guidance on guidewire performance). It's not expert consensus, pathology, or outcomes data.

8. The Sample Size for the Training Set:

This information is not applicable. This device is not an AI/ML algorithm that requires a training set. The "design controls" mentioned refer to the engineering and manufacturing process, not data training.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable as there is no training set for a physical device.

In summary, the provided document details the 510(k) clearance for a physical medical device (Premarket Notification: K251181) and its substantial equivalence based on standard bench testing. It does not provide information relevant to the assessment of an AI-enabled medical device.

FDA 510(k) Clearance Letter - Splashwire Hydrophilic Guide Wire

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.08.00
Silver Spring, MD 20993
www.fda.gov

Merit Medical Ireland Ltd.,
Shane Costello
Principal Regulatory Affairs Specialist
Parkmore Business Park West
Galway,
Ireland

Re: K251181
Trade/Device Name: Splashwire Hydrophilic Guide Wire
Regulation Number: 21 CFR 870.1330
Regulation Name: Catheter Guide Wire
Regulatory Class: Class II
Product Code: DQX
Dated: April 16, 2025
Received: July 30, 2025

Dear Shane Costello:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K251181 - Shane Costello Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Page 3

K251181 - Shane Costello Page 3

Jenny R. Katsnelson -S Digitally signed by Jenny R. Katsnelson -S Date: 2025.08.29 14:35:27 -04'00'

for Lydia Glaw
Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251181

Device Name: Splashwire Hydrophilic Guide Wire

Indications for Use (Describe):
The Merit Hydrophilic Guide Wire is intended to be used in the peripheral vascular system to facilitate the placement of devices during diagnostic and interventional procedures. These guide wires are not intended for Percutaneous Transluminal Coronary Angioplasty use or for use in the neurovasculature.

Type of Use (Select one or both, as applicable):
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Merit Medical Systems, Inc.
Splashwire™ Hydrophilic Guide Wire
Special Premarket Notification 510(k)

510(k) Summary

K251181

General Provisions
Submitter Name:Merit Medical Systems, Inc.
Address:1600 West Merit ParkwaySouth Jordan, UT 84095
Telephone Number:(+353) 91 703700
Fax Number:(+353) 91 680104
Contact Person:Shane Costello
Registration Number:1721504
Correspondent Name:Merit Medical Ireland Ltd.
Address:Parkmore Business ParkParkmore, Galway, Ireland
Telephone Number:(+353) 91 703700
Fax Number:(+353) 91 680104
Contact Person:Shane Costello
Date of Preparation:16th April 2025
Registration Number:9616662

K251181 Page 1 of 6

Page 6

Merit Medical Systems, Inc.
Splashwire™ Hydrophilic Guide Wire
Special Premarket Notification 510(k)

510(k) Summary

K251181

Subject Device
Trade Name:Splashwire Hydrophilic Guide Wire
Class:II
Product code:DQX
Classification Name:Wire, Guide, Catheter
Regulation Number:21 CFR 870.1330
Regulation Medical Specialty:Cardiovascular
Manufacturer:Merit Medical Systems Inc.
Predicate Device
Trade Name:Splashwire Hydrophilic Guide Wire
Class:II
Product code:DQX
Classification Name:Wire, Guide, Catheter
Regulation Number:21 CFR 870.1330
Regulation Medical Specialty:Cardiovascular
Premarket Notification:K201595
Manufacturer:Merit Medical Systems Inc.

K251181 Page 2 of 6

Page 7

Merit Medical Systems, Inc.
Splashwire™ Hydrophilic Guide Wire
Special Premarket Notification 510(k)

510(k) Summary

Device Description:
The Splashwire Hydrophilic guide wire consists of a nitinol core wire with a ground tapered distal section. A polyurethane jacket which contains tungsten for radiopacity is applied over the core wire and a hydrophilic coating is applied over the polyurethane jacket. The surface of the jacket is uniform with both the distal and proximal ends fully coated. The wire is placed inside a multiple loop flush dispenser, also referred to as a hoop. The dispenser has a flush port which facilitates solution flushing through the hoop to hydrate the guide wire. The wire is placed so that the distal end of the wire comes out of the outer portion of the hoop. A J-straightener is placed on the other end of the hoop to introduce the wire into the catheter. The wire is sold sterile and is a single use device. The subject Splashwire Hydrophilic Guide Wire Line Extensions and the predicate Splashwire Hydrophilic Guide Wire share the device characteristics described above, with the subject guide wire incorporating the following additional modifications:

J-Tip Guide Wires:
The J Tip devices incorporate a distal tip formed into a J shape. These guide wires are provided with a 1.5 mm J Tip.

Long Taper Guide Wires:
The Long Taper Guide Wires incorporate an extended core wire taper length providing enhanced flexibility.

Indications for Use:
The Merit Hydrophilic Guide Wire is intended to be used in the peripheral vascular system to facilitate the placement of devices during diagnostic and interventional procedures.

These guide wires are not intended for Percutaneous Transluminal Coronary Angioplasty use or for use in the neurovasculature.

K251181 Page 3 of 6

Page 8

Merit Medical Systems, Inc.
Splashwire™ Hydrophilic Guide Wire
Special Premarket Notification 510(k)

510(k) Summary

Comparison to Predicate:
The technological characteristics of the subject Splashwire Hydrophilic Guide Wire Guide Wire are substantially equivalent to those of the predicate Splashwire Hydrophilic Guide Wire Guide Wire [K201595].

Both the subject and predicate Splashwire Hydrophilic guide wires consist of a nitinol core wire with a ground tapered distal section. A polyurethane jacket which contains tungsten for radiopacity is applied over the core wire and a hydrophilic coating is applied over the polyurethane jacket. The surface of the jacket is uniform with both the distal and proximal ends fully coated. The wire is placed inside a multiple loop flush dispenser, also referred to as a hoop. The dispenser has a flush port which facilitates solution flushing through the hoop to hydrate the guide wire. The wire is placed so that the distal end of the wire comes out of the outer portion of the hoop. A J-straightener is placed on the other end of the hoop to introduce the wire into the catheter. The wire is sold sterile and is a single use device. The subject Splashwire Hydrophilic Guide Wire Line Extensions and the predicate Splashwire Hydrophilic Guide Wire share the device characteristics described above, with the subject guide wire incorporating the following additional modifications:

J-Tip Guide Wires:
The J Tip devices incorporate a distal tip formed into a J shape. These guide wires are provided with a 1.5 mm J Tip.

Long Taper Guide Wires:
The Long Taper Guide Wires incorporate an extended core wire taper length providing enhanced flexibility.

The fundamental technology and operating principles of the subject and the predicate are the same. The subject and predicate devices have the same indications for use.

Safety & Performance Tests:
No performance standards have been established under section 514 of the Food, Drug and Cosmetic Act for these devices. Testing was conducted on the subject Splashwire Hydrophilic Guide Wire in accordance with protocols based on requirements outlined in guidance's and industry standards and these were shown to meet the acceptance criteria that were determined to demonstrate substantial equivalence.

K251181 Page 4 of 6

Page 9

Merit Medical Systems, Inc.
Splashwire™ Hydrophilic Guide Wire
Special Premarket Notification 510(k)

510(k) Summary

Where appropriate, the tests were based on the requirements of the following documents:

  • FDA Guidance - "Coronary, Peripheral, and Neurovascular Guidewires Performance Tests and Recommended Labelling" – October 2019
  • FDA Guidance – "Intravascular Catheters, Wires, and Delivery Systems with Lubricious Coatings - Labeling Considerations" – October 2019
  • FDA Guidance – "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile" – January 2016
  • FDA Guidance – "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" – September 2020
  • ISO 10993-1:2018, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process
  • ISO 10993-18:2020, Biological evaluation of medical devices – Part 18: Chemical characterization of materials
  • TIR28:2016, Product adoption and process equivalence for ethylene oxide sterilization

The following is a list of all testing that was successfully completed:

Performance Non-Clinical Testing-Bench:

  • Dimensional Verification
  • Finished Wire Surface
  • Tip Tensile Strength
  • Tip Stiffness
  • Tip Flexibility
  • Prolapse Test
  • Torque Strength
  • Torqueability
  • Coating Integrity
  • Catheter Compatibility
  • Fracture Test

K251181 Page 5 of 6

Page 10

Merit Medical Systems, Inc.
Splashwire™ Hydrophilic Guide Wire
Special Premarket Notification 510(k)

510(k) Summary

  • Flexing Test
  • Simulated Use Testing
  • Kink Resistance
  • Ancillary Device Compatibility

All test results were comparable to the predicate Splashwire Hydrophilic Guide Wire and the subject Splashwire Hydrophilic Guide Wire met the predetermined acceptance criteria. This demonstrated that the subject device is substantially equivalent to the predicate device.

Summary of Substantial Equivalence:
Based on the comparisons noted, the subject Splashwire Hydrophilic Guide Wire meets the requirements that are considered essential for its intended use and is substantively equivalent to the Predicate Device, the Splashwire Hydrophilic Guide Wire [K201595].

K251181 Page 6 of 6

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.