(117 days)
The Daxor BVA-200 is an automated system that is used to measure/calculate the red cell mass (mL), plasma volume (mL) and total blood volume (mL), along with the related deviations from ideal values by amount (mL) and percentage (%) in adults. In addition, the Normalized Hematocrit (%) and Albumin Transudation Rate (%/min) are calculated. It is an in vitro medical device composed of a microprocessor, software, touchscreen, and gamma counter and accessory convenience kit.
The Daxor BVA-200 is intended to calculate human blood volumes by the method of tracer diffusion (Indicator dilution technique) with I-131 as the tracer after injection of I-131 Human Serum Albumin. The Daxor BVA-200 provides a Quantitative Assessment of total blood and plasma volumes using an automated system.
Data inputs to the software come from the measured characteristics of patient venous whole blood samples collected in K3EDTA vacutainer tubes (hematocrit and tracer concentration) and tracer calibration standards. The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program. The package also calculates the patient expected (or ideal) blood volume from physical parameters. Hyper- or hypovolemia, and associated red cell volumes are reported, with statistics showing the quality of the results.
For in vitro diagnostic use in a Clinical Laboratory setting and operated by laboratory technicians.
Rx use only.
The Daxor BVA-200 is an automated system that is used to calculate the red cell mass, plasma volume and total blood volume. It is an in vitro medical device composed of a microprocessor, software, touchscreen, and gamma counter. The accessory convenience kit includes single-use whole blood cartridges and protective sleeves.
The Daxor BVA-200 is designed to calculate human blood volume, using the method of tracer dilution, utilizing tagged serum albumin (I-131, resulting in "I-HSA"). Data inputs to the software come from the measured characteristics of patient blood samples (hematocrit and tracer concentration) and tracer calibration standards. The package also calculates the patient expected (or ideal) blood volume from physical parameters. Hyper- or hypovolemia, and associated red cell volumes, are reported, with statistics showing the quality of the results.
The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program.
The BVA-200 has a touchscreen for operator interaction, and provides clear instructions and prompts for the steps necessary for performing the test.
Here's a breakdown of the acceptance criteria and study information for the Daxor Blood Volume Analyzer (200), based on the provided FDA 510(k) clearance letter.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in a tabulated format. However, it provides performance metrics from a clinical comparison study and precision, linearity, carryover, and stability studies. The implicit acceptance criteria are that the device performance should be substantially equivalent to the predicate device and demonstrate acceptable precision and linearity.
Implicit Acceptance Criteria and Reported Device Performance for BVA-200 (vs. BVA-100)
| Measure | Acceptance Criteria (Implicit, based on substantial equivalence) | Reported Device Performance (vs. Predicate BVA-100) |
|---|---|---|
| Clinical Comparison (Regression Analysis) | Strong correlation, slope near 1, intercept near 0, narrow 95% CI covering 1 for slope and 0 for intercept. | |
| TBV (Total Blood Volume) | - | Passing-Bablok Slope: 1.031 (95% CI: 1.014 to 1.049), Intercept: -138 (-235 to -59), Pearson r: 0.99Deming Slope: 1.033 (95% CI: 1.017 to 1.050), Intercept: -153 (-241 to -69), Pearson r: 0.99 |
| RCV (Red Cell Volume) | - | Passing-Bablok Slope: 1.016 (95% CI: 1.002 to 1.030), Intercept: -25 (-50 to 2), Pearson r: 0.99Deming Slope: 1.024 (95% CI: 1.008 to 1.040), Intercept: -37 (-68 to -10), Pearson r: 0.99 |
| PV (Plasma Volume) | - | Passing-Bablok Slope: 1.033 (95% CI: 1.016 to 1.051), Intercept: -99 (-155 to -46), Pearson r: 0.99Deming Slope: 1.029 (95% CI: 1.014 to 1.046), Intercept: -83 (-138 to -32), Pearson r: 0.99 |
| nHct (Normalized Hematocrit) | - | Passing-Bablok Slope: 1.007 (95% CI: 0.982 to 1.032), Intercept: -0.19 (-1.18 to 0.77), Pearson r: 0.98Deming Slope: 1.022 (95% CI: 0.999 to 1.047), Intercept: -0.76 (-1.71 to 0.21), Pearson r: 0.98 |
| Precision (Total Precision %CV) | (Implicitly) Low %CV, indicating high precision. | All total precision %CV results were below 3.6%. |
| Linearity | (Implicitly) Strong correlation with expected values, low bias, and CV for repeatability ≤5%, max deviation from linearity ≤8%. | Global Slope near 1 (range 1.00-1.02), Mean Bias Percent low (range 0.37-0.41).For contrived samples (2000-18000 ml): CV for repeatability ≤5%, Max deviation from linearity ≤8%. |
| Carryover | No carryover observed. | No carryover observed. |
| Sample Stability | Impervious to temperature/humidity, decay matches I-131, withstands shipping, non-level operation up to 15 degrees, settling does not affect readings, stability for 4 days (refrigerated/room temp). | All stated conditions were met. "No significant differences were observed, nor was there any significant drift of measurements observed." |
| Handling (Cartridge drop test) | Performance criteria still met after drop. | None of the average measured parameters differed from pre-drop by more than 2.7%. |
| Handling (Shipping & Handling) | Resistance to breakage and radio-emission stability. Sample and device integrity maintained. | None of the average measured parameters differed from pre-ship by more than 3.1%. |
| Handling (Tilt position) | Stable measurement at non-level positioning up to 15 degrees. | None of the average measured parameters differed from pre-ship by more than 3.1%. |
| Interference (Hemolysis) | Results not affected by hemolysis. | Partial hemolysis: -0.72% ± 0.77% difference from baseline. Gross hemolysis: -0.14% ± 2.16% difference from baseline. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 319 unique, independent, comparable measurements.
- Data Provenance:
- Blood derived from patients undergoing blood volume measurement with the BVA-100 (predicate device) as part of their clinical treatment or as part of a research study.
- The document implies the data is retrospective for the clinical comparison, as it used blood derived from patients undergoing measurement with the predicate device.
- For precision, linearity, sample stability, handling, and interference studies, contrived samples and venous whole blood samples from male and female volunteers were used. No specific country of origin is mentioned.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of experts to establish ground truth for the test set in the context of a reader study or image interpretation. The ground truth for the clinical comparison study was established by the predicate device (Daxor BVA-100) measurements.
4. Adjudication Method for the Test Set
Not applicable. The ground truth for the clinical comparison was based on measurements from the predicate device, not expert adjudication of subjective assessments.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This device is a quantitative measurement system, not an AI software intended for interpretation by human readers. The clinical study focused on comparing the performance of the new device (BVA-200) to its predicate (BVA-100) on quantitative measurements.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the studies performed focused on the standalone performance of the BVA-200. The device is an automated system that measures and calculates blood volumes. The performance data presented (clinical comparison, precision, linearity, stability, etc.) represents the device's output independently, without human interpretation as part of a "human-in-the-loop" workflow being evaluated.
7. The Type of Ground Truth Used
The ground truth for the primary clinical comparison study was the measurements obtained from the predicate device, the Daxor BVA-100. For other studies (precision, linearity, stability, etc.), the ground truth was based on known properties of contrived samples or established physical/biological principles (e.g., known decay rate of I-131).
8. The Sample Size for the Training Set
The document does not provide information on a training set sample size. This suggests that the BVA-200, similar to its predicate BVA-100, is likely based on established physiological and radiological principles (indicator dilution technique, I-131 decay) rather than a machine learning model that requires a distinct "training set." The software performs calculations based on these principles and measured inputs, not through a learned model from a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as a "training set" in the context of machine learning is not implied or described in the document for this device.
FDA 510(k) Clearance Letter - Blood Volume Analyzer (200)
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.00
August 4, 2025
Daxor Corporation
Jonathan Feldschuh
Chief Scientific Officer
107 Meco Lane
Oak Ridge, Tennessee 37830
Re: K251087
Trade/Device Name: Blood Volume Analyzer (200)
Regulation Number: 21 CFR 864.5950
Regulation Name: Blood Volume Measuring Device
Regulatory Class: Class II
Product Code: JWO
Dated: July 3, 2025
Received: July 3, 2025
Dear Jonathan Feldschuh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act.
Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K251087 - Jonathan Feldschuh Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K251087 - Jonathan Feldschuh Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Takeesha Taylor-bell -S
Takeesha Taylor-Bell
Deputy Director
Division of Immunology and Hematology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
Indications for Use
510(k) Number (if known): K251087
Device Name: Blood Volume Analyzer (200)
Indications for Use (Describe)
The Daxor BVA-200 is an automated system that is used to measure/calculate the red cell mass (mL), plasma volume (mL) and total blood volume (mL), along with the related deviations from ideal values by amount (mL) and percentage (%) in adults. In addition, the Normalized Hematocrit (%) and Albumin Transudation Rate (%/min) are calculated. It is an in vitro medical device composed of a microprocessor, software, touchscreen, and gamma counter and accessory convenience kit.
The Daxor BVA-200 is intended to calculate human blood volumes by the method of tracer diffusion (Indicator dilution technique) with I-131 as the tracer after injection of I-131 Human Serum Albumin. The Daxor BVA-200 provides a Quantitative Assessment of total blood and plasma volumes using an automated system.
Data inputs to the software come from the measured characteristics of patient venous whole blood samples collected in K3EDTA vacutainer tubes (hematocrit and tracer concentration) and tracer calibration standards. The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program. The package also calculates the patient expected (or ideal) blood volume from physical parameters. Hyper- or hypovolemia, and associated red cell volumes are reported, with statistics showing the quality of the results.
For in vitro diagnostic use in a Clinical Laboratory setting and operated by laboratory technicians.
Rx use only.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
Page 5
510(k) Summary
I. SUBMITTER
Daxor Corporation
107 Meco Lane
Oak Ridge, TN 37803
Phone: 865-425-0555
Fax: 865-425-0551
Contact Person: Jonathan Feldschuh
Date Prepared: July 30, 2025
II. DEVICE
Name of Device: Blood Volume Analyzer
Common or Usual Name: BVA-200
Regulatory Class: Class 2
Product Code: JWO
III. PREDICATE DEVICE
Daxor BVA-100, K964406
This predicate has not been subject to a design-related recall
IV. DEVICE DESCRIPTION
The Daxor BVA-200 is an automated system that is used to calculate the red cell mass, plasma volume and total blood volume. It is an in vitro medical device composed of a microprocessor, software, touchscreen, and gamma counter. The accessory convenience kit includes single-use whole blood cartridges and protective sleeves.
The Daxor BVA-200 is designed to calculate human blood volume, using the method of tracer dilution, utilizing tagged serum albumin (I-131, resulting in "I-HSA"). Data inputs to the software come from the measured characteristics of patient blood samples (hematocrit and tracer concentration) and tracer calibration standards. The package also calculates the patient expected (or ideal) blood volume from physical parameters. Hyper- or hypovolemia, and associated red cell volumes, are reported, with statistics showing the quality of the results.
The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program.
Page 6
The BVA-200 has a touchscreen for operator interaction, and provides clear instructions and prompts for the steps necessary for performing the test.
V. INDICATIONS FOR USE
The Daxor BVA-200® is an automated system that is used to measure/calculate the red cell mass (mL), plasma volume (mL) and total blood volume (mL), along with the related deviations from ideal values by amount (mL) and percentage (%) in adults. In addition, the Normalized Hematocrit (%) and Albumin Transudation Rate (%/min) are calculated. It is an in vitro medical device composed of a microprocessor, software, touchscreen, and gamma counter and accessory convenience kit.
The Daxor BVA-200 is intended to calculate human blood volumes by the method of tracer diffusion (Indicator dilution technique) with I-131 as the tracer after injection of I-131 Human Serum Albumin. The Daxor BVA-200 provides a Quantitative Assessment of total blood and plasma volumes using an automated system.
Data inputs to the software come from the measured characteristics of patient venous whole blood samples collected in K3EDTA vacutainer tubes (hematocrit and tracer concentration) and tracer calibration standards. The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program. The package also calculates the patient expected (or ideal) blood volume from physical parameters. Hyper- or hypovolemia, and associated red cell volumes are reported, with statistics showing the quality of the results.
For in vitro diagnostic use in a Clinical Laboratory setting and operated by laboratory technicians.
Rx use only.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The proposed BVA-200 differs from the BVA-100 in that the BVA-200 can evaluate and calculate blood volume based on whole blood rather than requiring plasma fractionation (centrifugation and pipetting); incorporates a Mil Spec rugged tablet computer that uses a touch screen display for patient data entry in place of the 1990s era desktop computer, separate keyboard and monitor; has enhanced automated internal Quality Control; can be battery powered as well as main powered; and can be loaded and carried in a custom portable Pelican case.
The gamma count data for each sample are generated automatically by the BVA-200 when the operator inserts the labeled sleeve containing the filled custom Whole Blood Container into the custom aperture, enters the time of blood draw
Page 7
(or confirms the time prompted by the device if the operator has confirmed the time is contemporaneous with the device's internal clock), and activates the gamma counter on a screen prompt.
Both devices are capable of detecting I-131 concentrations quantitatively based on the known rate of gamma emissions from radioactive decay which is tightly concentration dependent. The source of I-131 in the original injectate is, after dilution, present in the quantities needed and thus will readily register against background for both detectors, with the number of counts scaling linearly with activity and the error scaling as the square root of activity. This is the counting statistical behavior required for the indicator dilution technique mathematics which is identical between the BVA-100 and BVA-200 or, indeed, the labeling of the approved radio-iodinated albumin (BLA 017837). If the time of blood draw differs from current, both devices will accommodate the extended time span by 'discounting' the gamma count rate by the known rate of decay from actual time of blood draw to actual time of gamma concentration measurement.
The BVA-100 predicate device is designed to automatically count a series of precision pipetted aliquots of patient plasma, which are placed in the carousel mechanism, and introduced into the detector well as the carousel moves under computer control. The time of measurement of radiation will be slightly different from the time of specimen withdrawal. Since both times are entered, the marginal decrease in radiation release based on decay of the tracer can be calculated and the algorithm can adjust for this time spread. The BVA-200 provides the same capability of delayed measurement at up to 48 hours.
Device & Predicate Device(s):
| Device & Predicate Device(s): | Device K251087 | Predicate K964406 |
|---|---|---|
| Device Trade Name | Blood Volume Analyzer (200) | Automated Multi-Point Blood Volume Analyzer |
General Device Characteristic Similarities
| Device K251087 | Predicate K964406 | |
|---|---|---|
| Intended Use/Indications For Use | The Daxor BVA-200® is an automated system that is used to measure/calculate the red cell mass (mL), plasma volume (mL) and total blood volume (mL), along with the related deviations from ideal values by amount (mL) and percentage (%) in adults. In addition, the Normalized Hematocrit (%) and Albumin Transudation Rate (%/min) are calculated. It is an in vitro medical device composed of a | The Daxor BVA-100 is a software package designed to calculate human blood volume, using the method of tracer dilution, utilizing tagged serum albumin (a commonly used tag is I-131, resulting in "I-HSA"). Data inputs to the software come from the measured characteristics of patient blood samples (hematocrit and tracer concentration) and tracer calibration standards. The package |
Page 8
| Device & Predicate Device(s): | Device K251087 | Predicate K964406 |
|---|---|---|
| microprocessor, software, touchscreen, and gamma counter and accessory convenience kit.The Daxor BVA-200 is intended to calculate human blood volumes by the method of tracer diffusion (Indicator dilution technique) with I-131 as the tracer after injection of I-131 Human Serum Albumin. The Daxor BVA-200 provides a Quantitative Assessment of total blood and plasma volumes using an automated system.Data inputs to the software come from the measured characteristics of patient venous whole blood samples collected in K3EDTA vacutainer tubes (hematocrit and tracer concentration) and tracer calibration standards. The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program. The package also calculates the patient expected (or ideal) blood volume from physical parameters. Hyper- or hypovolemia, and associated red cell volumes are reported, with statistics showing the quality of the results.For in vitro diagnostic use in a Clinical Laboratory setting and operated by laboratory technicians.Rx use only. | also calculates the patient expected (or ideal) blood volume from physical parameters.Hyper- or hypovolemia, and associated red cell volumes, are reported, with statistics showing the quality of the results.The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program. |
| Device & Predicate Device(s): | Device K251087 | Predicate K964406 |
|---|---|---|
| microprocessor, software, touchscreen, and gamma counter and accessory convenience kit. | also calculates the patient expected (or ideal) blood volume from physical parameters. | |
| The Daxor BVA-200 is intended to calculate human blood volumes by the method of tracer diffusion (Indicator dilution technique) with I-131 as the tracer after injection of I-131 Human Serum Albumin. The Daxor BVA-200 provides a Quantitative Assessment of total blood and plasma volumes using an automated system. | Hyper- or hypovolemia, and associated red cell volumes, are reported, with statistics showing the quality of the results. | |
| Data inputs to the software come from the measured characteristics of patient venous whole blood samples collected in K3EDTA vacutainer tubes (hematocrit and tracer concentration) and tracer calibration standards. The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program. The package also calculates the patient expected (or ideal) blood volume from physical parameters. Hyper- or hypovolemia, and associated red cell volumes are reported, with statistics showing the quality of the results. | The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program. | |
| For in vitro diagnostic use in a Clinical Laboratory setting and operated by laboratory technicians.Rx use only. |
Page 9
| Device & Predicate Device(s): | Device K251087 | Predicate K964406 |
|---|---|---|
| Principle of Operation | Indicator Dilution method | Same |
| Operator | Nuclear Medical Technician (NucMed Tech) | Same |
| Tracer | I-131 labeled Human Serum Albumin | Same |
| Radiation emitted by device | None (for radiation emitted by disposable samples placed in device). Gamma ("Geiger") counters do not emit radiation but only detect radiation. Further, the shielding around the detector itself is designed to isolate the gamma counter from ambient or external radiation to permit counting samples of low activity (improve signal to noise ratio). | Same |
| Values Reported | Total Blood Volume (TBV), Red Cell Volume (RCV), Plasma Volume (PV), along with patient-specific norms and deviations | Same |
| Scintillation Counter | Scintillates on bombardment with gamma rays | Same |
| Patient exposure to radiation | I-131 dose is limited to maximum activity of 25 microcuries (μCi). The EDE from a 25μCi dose is 101 rem, which is roughly the same as the annual exposure from naturally occurring sources of radiation (excluding radon) at sea level. | Same |
| Operator exposure to radiation | Operator (Nuclear Medical technician) is exposed to very small fraction (approximately 1/1000) of patient radiation during procedure, | Same |
Page 10
| Device & Predicate Device(s): | Device K251087 | Predicate K964406 |
|---|---|---|
| based on time handling dose and in proximity to patient. Nuclear Medical technicians are monitored for exposure to radiation. | ||
| Quality Control | Integrated Procedures | Same |
| Report Available | Onscreen Printed via attached printer Downloaded as file via thumb drive | Same |
| Total Blood Volume | The BVA-200 software will report the total blood volume with results that range from 1,700 mL to 14,500 mL as a function of the patient's true blood volume calculated by radio diffusion. | Same |
General Device Characteristic Differences
| Device K251087 | Predicate K964406 | |
|---|---|---|
| Sample Type | Whole Blood | Plasma |
| Phlebotomy volume required | 3 mL blood | 6 mL blood |
| Sample size | 1.8 mL | 1 mL |
| Shielding | Tungsten | Lead |
| Gamma Detection | Smaller Crystal size and material- counts fewer emissions, linearly corrected mathematically | Larger Crystal size and material - counts more emissions |
| Crystal Material | CsI advantages increased moisture resistance and plastic deformation rather than cracking | NaI |
| Photomultiplier | Photomultiplier silicon | Photomultiplier tube |
| Mitigation of Radiation Contamination | Samples are introduced into detector well inside disposable plastic sleeves. | Samples are placed in plastic drop tubes of sample changer. If drop |
Page 11
| Device & Predicate Device(s): | Device K251087 | Predicate K964406 |
|---|---|---|
| tubes become contaminated, they can be cleaned or swapped out. | ||
| Samples Introduced to gamma counter | Directly by operator, in response to software prompts. | Aliquots are placed in a carousel according to designated pattern indicated on device, and carousel sequentially delivers them to the gamma counter well. |
| User Interaction | Integrated Touchscreen (6" diagonal) | Attached Windows based computer & keyboard |
VII. PERFORMANCE DATA
Performance of the BVA-200 was validated in a clinical comparison study with the predicate BVA-100 device, using blood derived from patients undergoing blood volume measurement with the BVA-100 as part of their clinical treatment or as part of a research study. A total of 319 unique, independent, comparable measurements were recorded. This study demonstrated robust and clinically meaningful substantial equivalence. This study also demonstrated that the effects of matrix in whole blood (K3EDTA anticoagulant) are no different with the BVA-200 than with the BVA-100, which is cleared for marketing and analysis of blood volumes using the anticoagulated plasma fraction.
The following table shows the results for Passing-Bablok and Deming Regressions for the Instrument vs Comparator (BVA-200 vs. BVA-100):
| Measure | Regression | N | Slope (95% CI) | Intercept (95% CI) | Pearson correlation coefficient |
|---|---|---|---|---|---|
| TBV | Passing-Bablok | 319 | 1.031 (1.014 to 1.049) | -138 (-235 to -59) | 0.99 |
| TBV | Deming | 319 | 1.033 (1.017 to 1.050) | -153 (-241 to -69) | 0.99 |
| RCV | Passing-Bablok | 319 | 1.016 (1.002 to 1.030) | -25 (-50 to 2) | 0.99 |
Page 12
| Measure | Regression | N | Slope (95% CI) | Intercept (95% CI) | Pearson correlation coefficient |
|---|---|---|---|---|---|
| RCV | Deming | 319 | 1.024 (1.008 to 1.040) | -37 (-68 to -10) | 0.99 |
| PV | Passing-Bablok | 319 | 1.033 (1.016 to 1.051) | -99 (-155 to -46) | 0.99 |
| PV | Deming | 319 | 1.029 (1.014 to 1.046) | -83 (-138 to -32) | 0.99 |
| nHct | Passing-Bablok | 319 | 1.007 (0.982 to 1.032) | -0.19 (-1.18 to 0.77) | 0.98 |
| nHct | Deming | 319 | 1.022 (0.999 to 1.047) | -0.76 (-1.71 to 0.21) | 0.98 |
This study also confirmed the reference range of the ideal (patient normal) values for the calculated volumes (which are based on calculations detailed in reference 9):
| Reference Status | Normal | Has Medical Condition |
|---|---|---|
| N | 22 | 41 |
| Age | 36.7 ± 16.5 | 53.8 ± 16.8 |
| Sex | Female = 59% Male = 41% | Female = 24% Male = 76% |
| Inpatient / Outpatient | Outpatient Subject = 100% | Outpatient Subject = 54% Inpatient Subject = 46% |
| Hct | 41.3 ± 3.1 | 39.5 ± 5.91 |
| Deviation from Ideal Weight (%) | 26.9% ± 33.7% | 40.2% ± 30.5% |
| Ideal TBV | 4860 ± 655 | 5430 ± 848 |
| Ideal RCV | 1850 ± 328 | 2150 ± 397 |
| Ideal PV | 3010 ± 352 | 3280 ± 466 |
| BVA200_TBV | 4950 ± 808 | 5770 ± 1420 |
| BVA200_RCV | 1850 ± 357 | 2030 ± 507 |
| BVA200_PV | 3110 ± 497 | 3730 ± 1070 |
| BVA200_SD | 0.0196 ± 0.0107 | 0.0154 ± 0.00641 |
| BVA200_Slope (%/min) | 0.205% ± 0.070% | 0.197% ± 0.076% |
Page 13
| Reference Status | Normal | Has Medical Condition |
|---|---|---|
| BVA200_TBV Deviation from patient Ideal (%) | +1.75% ± 7.77% | +5.72% ± 15.8% |
| BVA200_RCV Deviation from patient Ideal (%) | -0.13% ± 8.29% | -5.40% ± 15.3% |
| BVA200_PV Deviation from patient Ideal (%) | +3.07% ± 10.5% | +13.1% ± 23.2% |
The following table shows the results of the reference range study, along with the categories listed in the BVA report. The study confirmed that the reference range is accurate.
| Value | Normal Range | Normal Females (N=13) | Normal Males (N=9) |
|---|---|---|---|
| TBV | -8 to +8 | 0.529 ± 0.09 | 3.52 ± 5.58 |
| RCV | -10 to +10 | 0.95 ± 8.39 | -1.69 ± 8.39 |
| PV | -8 to +8 | 0.291 ± 11.9 | 7.07 ± 6.78 |
| ATR | 0 to 0.25 | 0.179 ± 0.0608 | 0.242 ± 0.0692 |
| nHct | 37 - 41 (females) 40 - 46 (males) | 39.0 ± 4.53 | 44.2 ± 3.74 |
| Value | Normal | Mild | Moderate | Severe | Extreme |
|---|---|---|---|---|---|
| TBV (absolute dev %) | 0 to 8 | >8 to 16 | >16 to 24 | >24 to 32 | >32 |
| RCV (absolute dev %) | 0 to 10 | >10 to 20 | >20 to 30 | >30 to 40 | >40 |
| PV (absolute dev %) | 0 to 8 | >8 to 16 | >16 to 24 | >24 to 32 | >32 |
| Value | Normal | Elevated | High | Very High |
|---|---|---|---|---|
| ATR (%/min) | 0 to 0.25 | 0.25 to 0.40 | 0.40 to 0.50 | > 0.50 |
Page 14
Precision
Venous whole blood samples were obtained from male and female volunteers. Samples were spiked with various amounts of radio-iodinated albumin using Volumex to duplicate representative high (30 microCuries/mL radio-diluted in 2727mL of blood volume) and low (25 microCuries/mL diluted in 6250 mL of blood volume) ranges encountered by the BVA-100. Reference samples without radio-iodinated albumin were also prepared. A set of samples were diluted with plasma to reduce the hematocrit by 10% – thus a range of baseline hematocrits from approximately 40% to approximately 50% is expected plus a set with a range from 30% to 40%. Indeed, while testing of radiation equipment is more typically done with check samples without blood contents, the addition of blood contents permits a thorough performance evaluation.
Testing was conducted to assess relative precision in gamma radiation, TBV, PV, RCV, and nHCT in both the BVA-100 and BVA-200. Comparison of measurements taken at different times, under different conditions, and by different operators where the same sample was run in different settings demonstrated no substantial effect on results.
The following tables show the precision results for the BVA-200. All total precision percentage coefficient of variation results were below 3.6%.
BVA-200 TBV
| Sample | N | Mean Value | Repeatability (Within-Run) ||| Between-Run ||| Between-Day ||| Between-Operator ||| Between-Instrument ||| Between-Calibrator ||| Reproducibility Total Precision |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| | | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV |
| Sample (HIGH_Vol_1000) | 64 | 989 | 16.7 | 1.69 | | 0.1 | 0.01 | | 14.3 | 1.45 | | 2.0 | 0.20 | | 4.2 | 0.42 | | 3.3 | 0.33 | | 22.71 | 2.30 |
| Sample (HIGH_Vol_15000) | 64 | 14,322 | 379.6 | 2.65 | | 14.2 | 0.10 | | 186.6 | 1.30 | | 42.7 | 0.30 | | 93.7 | 0.65 | | 101.2 | 0.71 | | 447.17 | 3.12 |
| Sample (HIGH_Vol_2500) | 64 | 2,443 | 51.6 | 2.11 | | 4.6 | 0.19 | | 36.6 | 1.50 | | 6.9 | 0.28 | | 12.3 | 0.50 | | 20.1 | 0.82 | | 68.02 | 2.78 |
| Sample (HIGH_Vol_5000) | 64 | 4,967 | 120.6 | 2.43 | | 6.6 | 0.13 | | 101.1 | 2.04 | | 42.4 | 0.85 | | 40.0 | 0.81 | | 10.5 | 0.21 | | 168.28 | 3.39 |
| Sample (HIGH_Vol_7500) | 64 | 7,392 | 173.3 | 2.34 | | 1.7 | 0.02 | | 106.0 | 1.43 | | 23.8 | 0.32 | | 52.6 | 0.71 | | 60.6 | 0.82 | | 219.72 | 2.97 |
| Sample (LOW_Vol_1000) | 64 | 1,108 | 13.8 | 1.25 | | 1.3 | 0.12 | | 14.1 | 1.27 | | 4.3 | 0.39 | | 2.8 | 0.25 | | 3.7 | 0.33 | | 20.76 | 1.87 |
| Sample (LOW_Vol_15000) | 64 | 16,133 | 353.9 | 2.19 | | 50.1 | 0.31 | | 79.9 | 0.50 | | 107.9 | 0.67 | | 115.2 | 0.71 | | 81.9 | 0.51 | | 407.14 | 2.52 |
Page 15
| Sample | N | Mean Value | Repeatability (Within-Run) ||| Between-Run ||| Between-Day ||| Between-Operator ||| Between-Instrument ||| Between-Calibrator ||| Reproducibility Total Precision |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| | | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV |
| Sample (LOW_Vol_2500) | 64 | 2,747 | 68.6 | 2.50 | | 2.1 | 0.08 | | 31.3 | 1.14 | | 15.5 | 0.56 | | 1.7 | 0.06 | | 22.4 | 0.82 | | 80.22 | 2.92 |
| Sample (LOW_Vol_5000) | 64 | 5,419 | 104.6 | 1.93 | | 8.0 | 0.15 | | 84.1 | 1.55 | | 17.7 | 0.33 | | 33.8 | 0.62 | | 26.2 | 0.48 | | 142.20 | 2.62 |
| Sample (LOW_Vol_7500) | 64 | 6,925 | 151.0 | 2.18 | | 1.0 | 0.01 | | 84.1 | 1.21 | | 63.7 | 0.92 | | 7.3 | 0.11 | | 47.6 | 0.69 | | 190.40 | 2.75 |
BVA-200 RCV
| Sample | N | Mean Value | Repeatability (Within-Run) ||| Between-Run ||| Between-Day ||| Between-Operator ||| Between-Instrument ||| Between-Calibrator ||| Reproducibility Total Precision |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| | | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV |
| Sample (HIGH_Vol_1000) | 64 | 536 | 9.0 | 1.68 | | 0.0 | 0.00 | | 7.8 | 1.46 | | 1.1 | 0.21 | | 2.3 | 0.43 | | 1.9 | 0.35 | | 12.33 | 2.30 |
| Sample (HIGH_Vol_15000) | 64 | 7,984 | 211.6 | 2.65 | | 7.9 | 0.10 | | 104.0 | 1.30 | | 23.8 | 0.30 | | 52.2 | 0.65 | | 56.4 | 0.71 | | 249.25 | 3.12 |
| Sample (HIGH_Vol_2500) | 64 | 1,343 | 28.4 | 2.11 | | 2.6 | 0.19 | | 20.1 | 1.50 | | 3.8 | 0.28 | | 6.7 | 0.50 | | 11.2 | 0.83 | | 37.44 | 2.79 |
| Sample (HIGH_Vol_5000) | 64 | 2,763 | 67.1 | 2.43 | | 3.7 | 0.13 | | 56.3 | 2.04 | | 23.7 | 0.86 | | 22.2 | 0.80 | | 5.8 | 0.21 | | 93.67 | 3.39 |
| Sample (HIGH_Vol_7500) | 64 | 4,121 | 96.6 | 2.34 | | 0.9 | 0.02 | | 59.1 | 1.43 | | 13.2 | 0.32 | | 29.3 | 0.71 | | 33.8 | 0.82 | | 122.48 | 2.97 |
| Sample (LOW_Vol_1000) | 64 | 186 | 2.3 | 1.24 | | 0.2 | 0.11 | | 2.4 | 1.29 | | 0.7 | 0.38 | | 0.4 | 0.22 | | 0.6 | 0.32 | | 3.48 | 1.87 |
| Sample (LOW_Vol_15000) | 64 | 2,834 | 62.2 | 2.19 | | 8.8 | 0.31 | | 14.0 | 0.49 | | 19.0 | 0.67 | | 20.2 | 0.71 | | 14.5 | 0.51 | | 71.57 | 2.53 |
| Sample (LOW_Vol_2500) | 64 | 464 | 11.6 | 2.50 | | 0.3 | 0.06 | | 5.3 | 1.14 | | 2.6 | 0.56 | | 0.3 | 0.06 | | 3.8 | 0.82 | | 13.57 | 2.92 |
| Sample (LOW_Vol_5000) | 64 | 928 | 17.9 | 1.93 | | 1.4 | 0.15 | | 14.4 | 1.55 | | 3.0 | 0.32 | | 5.8 | 0.62 | | 4.4 | 0.47 | | 24.33 | 2.62 |
Page 16
| Sample | N | Mean Value | Repeatability (Within-Run) ||| Between-Run ||| Between-Day ||| Between-Operator ||| Between-Instrument ||| Between-Calibrator ||| Reproducibility Total Precision |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Sample (LOW_Vol_7500) | 64 | 1,146 | 25.0 | 2.18 | | 0.2 | 0.02 | | 13.9 | 1.21 | | 10.5 | 0.92 | | 1.2 | 0.10 | | 7.9 | 0.69 | | 31.50 | 2.75 |
BVA-200 PV
| Sample | N | Mean Value | Repeatability (Within-Run) ||| Between-Run ||| Between-Day ||| Between-Operator ||| Between-Instrument ||| Between-Calibrator ||| Reproducibility Total Precision |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| | | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV |
| Sample (HIGH_Vol_1000) | 64 | 453 | 7.6 | 1.68 | | 0.0 | 0.00 | | 6.6 | 1.46 | | 0.9 | 0.20 | | 1.9 | 0.42 | | 1.5 | 0.33 | | 10.39 | 2.29 |
| Sample (HIGH_Vol_15000) | 64 | 6,339 | 168.0 | 2.65 | | 6.3 | 0.10 | | 82.5 | 1.30 | | 18.9 | 0.30 | | 41.4 | 0.65 | | 44.8 | 0.71 | | 197.86 | 3.12 |
| Sample (HIGH_Vol_2500) | 64 | 1,101 | 23.2 | 2.11 | | 2.0 | 0.18 | | 16.5 | 1.50 | | 3.1 | 0.28 | | 5.5 | 0.50 | | 9.1 | 0.83 | | 30.61 | 2.78 |
| Sample (HIGH_Vol_5000) | 64 | 2,204 | 53.5 | 2.43 | | 3.0 | 0.14 | | 44.9 | 2.04 | | 18.9 | 0.86 | | 17.7 | 0.80 | | 4.7 | 0.21 | | 74.70 | 3.39 |
| Sample (HIGH_Vol_7500) | 64 | 3,271 | 76.7 | 2.34 | | 0.8 | 0.02 | | 47.0 | 1.44 | | 10.5 | 0.32 | | 23.3 | 0.71 | | 26.8 | 0.82 | | 97.28 | 2.97 |
| Sample (LOW_Vol_1000) | 64 | 922 | 11.5 | 1.25 | | 1.1 | 0.12 | | 11.7 | 1.27 | | 3.6 | 0.39 | | 2.3 | 0.25 | | 3.0 | 0.33 | | 17.25 | 1.87 |
| Sample (LOW_Vol_15000) | 64 | 13,299 | 291.7 | 2.19 | | 41.3 | 0.31 | | 65.8 | 0.49 | | 89.0 | 0.67 | | 95.0 | 0.71 | | 67.6 | 0.51 | | 335.62 | 2.52 |
| Sample (LOW_Vol_2500) | 64 | 2,283 | 57.1 | 2.50 | | 1.7 | 0.07 | | 26.0 | 1.14 | | 12.8 | 0.56 | | 1.4 | 0.06 | | 18.6 | 0.81 | | 66.72 | 2.92 |
| Sample (LOW_Vol_5000) | 64 | 4,492 | 86.6 | 1.93 | | 6.7 | 0.15 | | 69.7 | 1.55 | | 14.7 | 0.33 | | 28.1 | 0.63 | | 21.6 | 0.48 | | 117.79 | 2.62 |
| Sample (LOW_Vol_7500) | 64 | 5,778 | 126.0 | 2.18 | | 0.8 | 0.01 | | 70.2 | 1.21 | | 53.2 | 0.92 | | 6.1 | 0.11 | | 39.8 | 0.69 | | 158.92 | 2.75 |
Page 17
BVA-200 nHct
| Sample | N | Mean Value | Repeatability (Within-Run) ||| Between-Run ||| Between-Day ||| Between-Operator ||| Between-Instrument ||| Between-Calibrator ||| Reproducibility Total Precision |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| | | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV | | SD | %CV |
| Sample (HIGH_Vol_1000) | 64 | 12 | 0.2 | 1.67 | | 0.0 | 0.00 | | 0.2 | 1.67 | | 0.0 | 0.00 | | 0.1 | 0.83 | | 0.0 | 0.00 | | 0.30 | 2.50 |
| Sample (HIGH_Vol_15000) | 64 | 175 | 4.7 | 2.69 | | 0.2 | 0.11 | | 2.3 | 1.31 | | 0.5 | 0.29 | | 1.2 | 0.69 | | 1.2 | 0.69 | | 5.53 | 3.16 |
| Sample (HIGH_Vol_2500) | 64 | 29 | 0.6 | 2.07 | | 0.1 | 0.34 | | 0.4 | 1.38 | | 0.1 | 0.34 | | 0.2 | 0.69 | | 0.2 | 0.69 | | 0.79 | 2.72 |
| Sample (HIGH_Vol_5000) | 64 | 61 | 1.5 | 2.46 | | 0.1 | 0.16 | | 1.2 | 1.97 | | 0.5 | 0.82 | | 0.5 | 0.82 | | 0.1 | 0.16 | | 2.05 | 3.36 |
| Sample (HIGH_Vol_7500) | 64 | 90 | 2.1 | 2.33 | | 0.0 | 0.00 | | 1.3 | 1.44 | | 0.3 | 0.33 | | 0.6 | 0.67 | | 0.7 | 0.78 | | 2.65 | 2.95 |
| Sample (LOW_Vol_1000) | 64 | 4 | 0.1 | 2.50 | | 0.0 | 0.00 | | 0.1 | 2.50 | | 0.0 | 0.00 | | 0.0 | 0.00 | | 0.0 | 0.00 | | 0.14 | 3.54 |
| Sample (LOW_Vol_15000) | 64 | 62 | 1.4 | 2.26 | | 0.2 | 0.32 | | 0.3 | 0.48 | | 0.4 | 0.65 | | 0.4 | 0.65 | | 0.3 | 0.48 | | 1.58 | 2.55 |
| Sample (LOW_Vol_2500) | 64 | 10 | 0.3 | 3.00 | | 0.0 | 0.00 | | 0.1 | 1.00 | | 0.1 | 1.00 | | 0.0 | 0.00 | | 0.1 | 1.00 | | 0.35 | 3.46 |
| Sample (LOW_Vol_5000) | 64 | 20 | 0.4 | 2.00 | | 0.0 | 0.00 | | 0.3 | 1.50 | | 0.1 | 0.50 | | 0.1 | 0.50 | | 0.1 | 0.50 | | 0.53 | 2.65 |
| Sample (LOW_Vol_7500) | 64 | 25 | 0.6 | 2.40 | | 0.0 | 0.00 | | 0.3 | 1.20 | | 0.2 | 0.80 | | 0.0 | 0.00 | | 0.2 | 0.80 | | 0.73 | 2.91 |
Linearity
Multiple samples were measured in both the BVA-100 and BVA-200 to assess both linearity and bias at the high and low measurements. The following results were obtained from the data in the method comparison study:
| N | Mean Expected | Mean Observed | SD | CV Percent | Global Slope | Mean Bias Percent | CI Lower | CI Upper | |
|---|---|---|---|---|---|---|---|---|---|
| TBV | 319 | 5323 | 5348 | 170 | 3.24 | 1.02 | 0.41 | 5088 | 5608 |
| RCV | 319 | 1905 | 1912 | 56 | 3.00 | 1.01 | 0.38 | 1828 | 1996 |
| PV | 319 | 3417 | 3433 | 105 | 3.11 | 1.02 | 0.37 | 3275 | 3590 |
| nHCT | 319 | 40.14 | 40.27 | 1.22 | 3.00 | 1.00 | 0.37 | 38.43 | 42.11 |
Page 18
An additional study using contrived samples was performed to verify linearity and absence of bias at very high and low volume measurements (2000 ml to 18000 ml). The coefficient of variation (CV) for repeatability was ≤5% for all measured parameters across the tested concentration range. The maximum deviation from linearity was ≤8% for all measured parameters across the tested concentration range, as determined by comparison to a linear regression model. In conclusion, there is no clinically meaningful deviation from linearity.
Carryover
Analysis of the impact of a preceding high level spiked sample on the values of a following low level spiked sample were conducted for the BVA-200 analyses of radiation emission. Sufficient high-level samples immediately followed by low-level samples by the same operator and with the same equipment were utilized to assess the potential for carryover. There was no carryover observed.
Be sure that blood is not spilled in the sample port. Clean as necessary. QC checks will demonstrate contamination if any. There was no carryover observed in testing.
Sample Stability
The spiked venous whole blood stability, settling, FEDEX shipping study, and non-level positioning demonstrate that the gamma emission from a fixed amount of blood is impervious to temperature and humidity; that the decrease in radiation emission over time corresponds to the known decrease in radiation from the isotope in question (I-131); that the WBCC can withstand shipping; that the BVA-200 operates in a non-level position up to 15 degrees in any direction; and that settling does not affect gamma emissions read by the gamma counter scintillator inside the BVA-200.
Stability testing was performed on whole blood cartridges that were first measured, then allowed to settle in either refrigerated or room temperature conditions for a period up to four days. Samples were recounted at 18, 24, 48, and 120 hours in both agitated and settled states. Contrived samples were used that covered a wide range of possible Hct and volume values. No significant differences were observed, nor was there any significant drift of measurements observed. The study performed validates the measurement of cartridges up to 4 days of stability at refrigerated (4ºC/ 40ºF) and room temperature (22ºC/ 72ºF).
Handling Testing
A cartridge drop test was performed. The results validated that the cartridges still met all performance criteria after cartridges were dropped from a height of 48 inches onto a hard surface. None of the average measured parameters of the test differed from the pre-drop measurements by more than 2.7%.
Page 19
A simulated shipping & handling study was performed to assess resistance to breakage and radio-emission stability in a standardized shipping stress testing (Study ISTA-3A). This includes testing for temperature, humidity, vibration, and shock. As with the settling study, this study demonstrated that, once filled, the whole blood cartridge emission rate performs as expected for the volume and concentration of radioisotope. The testing validated sample integrity after shipping blood in a vacutainer tube, sample integrity after shipping blood in a cartridge, sample integrity after shipping a cartridge from a shipped vacutainer, and device integrity after shipping a device and a shipping cartridge. None of the average measured parameters of the test differed from the pre-ship measurements by more than 3.1%.
Tilt position – Testing established that the measurement of the analyte concentration (I-131 gamma emission) from the venous whole blood sample is stable notwithstanding non-level positioning at 15 degrees in any direction. None of the average measured parameters of the test differed from the pre-ship measurements by more than 3.1%.
Interference
A hemolysis interference study was performed. Hemolysis is a common interferant in medical diagnostic tests involving measuring analytes in blood. A study was performed with contrived samples to compare the BVA-200 reported parameters across three hemolysis levels. Baseline samples were verified to have no hemolysis. Partial hemolysis was induced by subjecting blood samples to a partial freezing cycle at -20C. Gross hemolysis was induced by subjecting the partially hemolyzed blood samples to an additional complete freezing cycle (until completely frozen) at -20C. Level of hemolysis was verified by three independent methods (colorimetric matching, quantification of fluorescent absorption at 540 nm, and Hct measurement). Ten independent samples were used, with volume values measured after partial hemolysis and gross hemolysis compared to the baseline (no hemolysis). Results for all parameters (TBV, PV, RCV, and nHct) were not affected by hemolysis: the difference between the partial hemolysis values and the no hemolysis values were - 0.72% ± 0.77%, expressed as a percentage of the baseline no hemolysis values; and gross hemolysis values were - 0.14% ± 2.16%, expressed as a percentage of the baseline no hemolysis values.
VIII. CONCLUSIONS
The clinical data, performance, and non-clinical data support the safety of the device and the substantial equivalence to the predicate for the same intended use. Hardware and software verification and validation testing demonstrates that the BVA-200 device should perform as intended in the specified use conditions.
N/A