K Number
K251019

Validate with FDA (Live)

Date Cleared
2025-12-22

(264 days)

Product Code
Regulation Number
876.5010
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The catheter is designed for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and others fluids.

Device Description

The BT-PD1-SERIES-G / BT-PD1-SERIES(MN)-G / BT-PD1-SERIES-G(+FSC) / BT-PDS-SERIES-G / BT-PDS-SERIES(MN)-G / BT-PDS-SERIES(B)-RB-G Percutaneous Drainage Catheter with hydrophilic coating is a percutaneous drainage catheter used for drainage of abscess and fluid collections. The catheter is made from a soft, biocompatible plastic, a material that is radiopaque for X-rays. The distal end of catheter contains a pigtail or mini-pigtail close loop and drainage holes.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - BIOTEQ Drainage Catheter Set

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.02

December 22, 2025

Bioteque Corporation
Stella Hsu
RA Specialist
5F-6, No. 23, Sec.1, Chang'an E. Rd.,
Zhongshan Dist
Taipei, 104
Taiwan

Re: K251019
Trade/Device Name: BIOTEQ Drainage Catheter Set: BT-PD1-SERIES and BT-PDS-SERIES
Regulation Number: 21 CFR 876.5010
Regulation Name: Biliary Catheter And Accessories
Regulatory Class: Class II
Product Code: FGE, LJE, GBO
Dated: November 20, 2025
Received: November 20, 2025

Dear Stella Hsu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251019 - Stella Hsu Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251019 - Stella Hsu Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

ANTHONY LEE -S

Anthony Lee, Ph.D., MBA
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026

Indications for Use

See PRA Statement below.

510(k) Number (if known)
K251019

Device Name
BIOTEQ Drainage Catheter Set: BT-PD1-SERIES-G and BT-PDS-SERIES-G

Indications for Use (Describe)
The catheter is designed for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and others fluids.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1

Page 5

Bioteque Corporation
BIOTEQ Drainage Catheter Set

510(k) Summary

1. Type of submission: Traditional
Date of Summary: December 19, 2025

2. Submitter: Bioteque Corporation

Address5F-6, No. 23, Sec. 1, Chang'an E. Rd., Zhongshan Dist. Taipei City, 104 Taiwan
Phone+886-2-2571-0269 ext. 1348
Fax+886-2-2536-1967
ContactStella Hsu (stellahsu@bioteq.com.tw)

3. Identification of the Device:
Device Name: BIOTEQ Drainage Catheter Set
Classification Product Code: FGE
Subsequent Product Code: LJE, GBO
Regulation Number: 876.5010
Regulation Description: Biliary catheter and accessories; Introduction/drainage catheter and accessories.
Review Panel: Gastroenterology/Urology; General & Plastic Surgery
Device Class: II
Basis for the Submission: New device

4. Identification of the Predicate Device I:
Predicate Device Name: BIOTEQ Drainage Catheter Set (Seldinger Type) Model name: BT-PDS-series
Manufacturer: Bioteque Corporation
Classification Product Code: FGE
Subsequent Product Code: LJE, GBO
Regulation number: 876.5010
Device Class: II
510(k) number: K210419

5. Identification of the Predicate Device II:
Predicate Device Name: BIOTEQ Drainage Catheter Set (One Step Type) Model name: BT-PD1-series
Manufacturer: BIOTEQUE CORPORATION
Classification Product Code: FGE

K251019 page 1 of 10

Page 6

Bioteque Corporation
BIOTEQ Drainage Catheter Set

Subsequent Product Code: LJE, GBO
Regulation number: 876.5010
Device Class: II
510(k) number: K200103

6. Device Description

The BT-PD1-SERIES-G / BT-PD1-SERIES(MN)-G / BT-PD1-SERIES-G(+FSC) / BT-PDS-SERIES-G / BT-PDS-SERIES(MN)-G / BT-PDS-SERIES(B)-RB-G Percutaneous Drainage Catheter with hydrophilic coating is a percutaneous drainage catheter used for drainage of abscess and fluid collections. The catheter is made from a soft, biocompatible plastic, a material that is radiopaque for X-rays. The distal end of catheter contains a pigtail or mini-pigtail close loop and drainage holes.

7. Indications for Use

The catheter is designed for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and others fluids.

8. Performance Data - Non-clinical Testing

The following performance data were provided in support of the substantial equivalence determination.

(1) Sterilization validation

The ethylene oxide (EO) sterilization and related validation testing were conducted in accordance and complied with ISO 11135, ISO 10993-7, ISO 11737-1, ISO 11737-2, USP, and EN 556-1.

(2) Shelf-life

The shelf-life testing was conducted and complied with ISO 20697, ISO 80369-7, 80369-20, ASTM F1140/F1140M, ISO11070, ISO9626, ISO 2859-1.

(3) Biocompatibility

The biocompatibility evaluation was conducted, and the endpoint testing included the following tests:

  • In vitro Cytotoxicity test
  • Skin Sensitization Study in Guinea Pigs
  • Intracutaneous Irritation Study in White Rabbits
  • Acute Systemic Toxicity Study in Mice

K251019 page 2 of 10

Page 7

Bioteque Corporation
BIOTEQ Drainage Catheter Set

  • Pyrogen test in white rabbit
  • A Dual Route sub chronic systemic toxicity study in rats
  • Salmonella reverse mutation test
  • In vitro mammalian chromosomal aberration test
  • Rodent Micronucleus Test in Peripheral Blood
  • Muscle Implantation study in white rabbits

(4) Performance

The performance testing was conducted and complied with ISO 20697, ISO 80369-7, ISO 80369-20.

9. Performance Data - Clinical Testing

No clinical test data was used to support the decision of substantial equivalence.

10. Substantial equivalence comparison

BIOTEQ Drainage Catheter Set submitted in 510(k) files is substantially equivalent in intended use, safety and performance to the FDA cleared BIOTEQ Drainage Catheter Set (Seldinger Type) (K210419) and BIOTEQ Drainage Catheter Set (One Step Type) (K200103). Differences between the devices cited in this section do not raise any new issue of substantial equivalence.

K251019 page 3 of 10

Page 8

Bioteque Corporation
BIOTEQ Drainage Catheter Set

ItemSubject devicePredicate deviceSubstantial equivalence determination
Proprietary nameBIOTEQ Drainage Catheter SetBIOTEQ Drainage Catheter Set
SeriesBT-PDS-series-GBT-PDS-series-
510(k) No.K251019K210419-
Intended UseThe BIOTEQ Drainage Catheter Set (Seldinger Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids.The BIOTEQ Drainage Catheter Set (Seldinger Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids.Equivalent Both the devices are used for percutaneous drainage in gastroenterology and urology.
Type of UsePrescription UsePrescription UseSame
Intended patient populationAdults and recommending clinician should choose an appropriate catheter size for pediatric use.Adults and recommending clinician should choose an appropriate catheter size for pediatric useSame
Catheter Shaft MaterialTPUTPUSame
Coating MaterialHydromer AQUA 65JLHydromer AQUA 65JLSame
Catheter colorGreenBlueDifferent
Distal ConfigurationString Locking Pigtail, Non-String Locking PigtailString Locking Pigtail, Non-String Locking PigtailSame

K251019 page 4 of 10

Page 9

Bioteque Corporation
BIOTEQ Drainage Catheter Set

ItemSubject devicePredicate deviceSubstantial equivalence determination
Proprietary nameBIOTEQ Drainage Catheter SetBIOTEQ Drainage Catheter Set
Distal ShapePigtail, Mini-closed PigtailPigtail, Closed-PigtailEquivalent Both the devices have pigtail shape.
Distal Hydrophilic CoatingYesYesSame
Shaft Depth Printing MarkersYesYesSame
Proximal Hub AssemblyHub (for String Lock Pigtail), F.L.L. AdapterHub (for String Lock Pigtail), F.L.L. AdapterSame
Size5 Fr (Non-String Lock), 6, 7, 8, 10, 12, 14, 16 Fr8, 10, 12, 14 FrDifferent but do not raise new issues of SE
Usable Length20, 25, 30, 35, 40, 45, 50 cm40, 45, 50 cmDifferent but do not raise new issues of SE
Included Insert AccessoryMetal Stiffening Cannula Flexible (plastic) Stiffening Cannula Wire cap, Suture Wire, Curve Straightener, Radiopaque bandMetal Stiffening Cannula Flexible (plastic) Stiffening Cannula Wire cap, Suture Wire, Curve Straightener, Radiopaque bandEquivalent The accessory of subject device is equivalent to the predicate device
PackagingTyvek/Mylar (PET/LDPE) pouchTyvek/Mylar (PET/LDPE) pouchSame

K251019 page 5 of 10

Page 10

Bioteque Corporation
BIOTEQ Drainage Catheter Set

ItemSubject devicePredicate deviceSubstantial equivalence determination
Proprietary nameBIOTEQ Drainage Catheter SetBIOTEQ Drainage Catheter Set
Sterilization MethodEthylene OxideEthylene OxideSame

K251019 page 6 of 10

Page 11

Bioteque Corporation
BIOTEQ Drainage Catheter Set

ItemSubject devicePredicate deviceSubstantial equivalence determination
Proprietary nameBIOTEQ Drainage Catheter Set (One step Type)BIOTEQ Drainage Catheter Set (One Step Type)
SeriesBT-PD1-series-GBT-PD1-series-
510(k) No.K251019K200103-
Intended UseThe BIOTEQ Drainage Catheter Set (One step Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids.The BIOTEQ Drainage Catheter Set (One Step Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids.Equivalent Both the devices are used for percutaneous drainage in gastroenterology and urology.
Type of UsePrescription UsePrescription UseSame
Intended patient populationAdults and recommending clinician should choose an appropriate catheter size for pediatric useAdults and recommending clinician should choose an appropriate catheter size for pediatric useSame
Catheter Shaft MaterialTPUTPUSame
Coating MaterialHydromer AQUA 65JLHydromer AQUA 65JLSame
Catheter colorGreenBlueDifferent
Distal ConfigurationString Locking Pigtail, Non-String Locking PigtailString Locking Pigtail, Non-String Locking PigtailSame

K251019 page 7 of 10

Page 12

Bioteque Corporation
BIOTEQ Drainage Catheter Set

ItemSubject devicePredicate deviceSubstantial equivalence determination
Proprietary nameBIOTEQ Drainage Catheter Set (One step Type)BIOTEQ Drainage Catheter Set (One Step Type)
Distal ShapePigtail, Mini-closed PigtailPigtail, Closed-Pigtail, Mini-Pigtail, Mini-closed Pigtail, J shapeEquivalent All the devices have pigtail and mini-closed pigtail shape.
Distal Hydrophilic CoatingYesYesSame
Shaft Depth Printing MarkersYesYesSame
Proximal Hub AssemblyHub (for String Lock Pigtail), F.L.L. AdapterHub (for String Lock Pigtail), F.L.L. AdapterSame
Size5 Fr (Non-String Lock), 6, 7, 8, 10, 12, 14, 16 Fr5 Fr (Non-String Lock), 6, 7, 8, 10, 12, 14, 16 FrSame
Useable Length20, 25, 30, 35, 40, 45, 50 cm20, 25, 30, 35, 40, 45, 50 cmSame
Included Insert AccessoryTrocar needle, Trocar stylet Flexible (plastic) Stiffening Cannula Wire cap, Suture Wire, Curve StraightenerTrocar needle, Trocar stylet Flexible (plastic) Stiffening Cannula Wire cap, Suture Wire, Curve StraightenerSame
PackagingTyvek/Mylar (PET/LDPE) pouchTyvek/Mylar (PET/LDPE) pouchSame

K251019 page 8 of 10

Page 13

Bioteque Corporation
BIOTEQ Drainage Catheter Set

ItemSubject devicePredicate deviceSubstantial equivalence determination
Proprietary nameBIOTEQ Drainage Catheter Set (One step Type)BIOTEQ Drainage Catheter Set (One Step Type)
Sterilization MethodEthylene OxideEthylene OxideSame

K251019 page 9 of 10

Page 14

Bioteque Corporation
BIOTEQ Drainage Catheter Set

11. Conclusion

The data supports the safety and performance of the subject device and demonstrates that the subject device is substantially equivalent to the predicate device.

K251019 page 1 of 10

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.