K Number
K251014

Validate with FDA (Live)

Device Name
APRO 45 Catheter
Manufacturer
Date Cleared
2025-05-01

(29 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The APRO 45 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 45 Catheter is also indicated for use as a conduit for retrieval devices.

Device Description

The APRO 45 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 45 Catheter is designed to facilitate the insertion and guidance of interventional devices into peripheral and neuro vasculature. Using standard catheterization techniques under fluoroscopic guidance, the APRO 45 Catheter is introduced through a guide catheter or guide sheath and over a guidewire into the target vasculature. The distal segment of the catheter shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal tip of the catheter for visualization under fluoroscopy. The APRO 45 Catheter is provided with an Introducer Sheath. The APRO 45 Catheter contains animal derived materials made from tallow derivatives.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the APRO 45 Catheter focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance and biocompatibility testing. It does not present information related to an AI/ML-driven medical device, a multi-reader multi-case (MRMC) study, or acceptance criteria for algorithmic performance.

Therefore, based solely on the provided text, it's not possible to fulfill most of the requested information points, as they pertain to the evaluation of AI/ML diagnostic or assistive devices. The document describes a physical medical device (catheter) and its non-clinical testing.

However, I will extract and present the acceptance criteria and performance information that is available from the provided text, primarily from the "Performance Data" section.


Acceptance Criteria and Device Performance for APRO 45 Catheter

The APRO 45 Catheter is a physical medical device, not an AI/ML driven diagnostic or assistive device. The acceptance criteria and performance data provided in the FDA 510(k) summary pertain to the physical and biological characteristics of the catheter, supporting its substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Visual and Dimensional CharacteristicsCatheter meets the visual and dimensional specifications.The APRO 45 Catheter met the acceptance criteria.
ParticulateCatheter meets the acceptance criteria. Subject device was evaluated with a predicate device under the same test conditions.The APRO 45 Catheter met the acceptance criteria.
Kink ResistanceCatheter shaft shall not kink at clinically relevant radii.The APRO 45 Catheter met the acceptance criteria.
Hub Air LeakageCatheter does not leak air into hub assembly with methods specified in ISO 10555-1, Annex D.The APRO 45 Catheter met the acceptance criteria.
Hub CompatibilityCatheter meets the requirements specified in ISO 80369-7.The APRO 45 Catheter met the acceptance criteria.
Torque StrengthCatheter must withstand the minimum required number of rotations without breakage and without kinking compared to legally marketed devices.The APRO 45 Catheter met the acceptance criteria.
Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 45 Catheter met the acceptance criteria.
Liquid LeakageCatheter must withstand pressure with methods specified in ISO 10555-1, Annex C.The APRO 45 Catheter met the acceptance criteria.
Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 45 Catheter met the acceptance criteria.
Hub and Shaft Tensile StrengthCatheter hub and shaft must meet tensile strength specification.The APRO 45 Catheter met the acceptance criteria.
Tip Tensile StrengthCatheter tip must meet tip tensile strength specification.The APRO 45 Catheter met the acceptance criteria.
3-Point BendCatheter 3-Point Bend force must be acceptable. Forces were compared to the predicate.The APRO 45 Catheter met the acceptance criteria.
Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to the predicate.The APRO 45 Catheter met the acceptance criteria.
Simulated UseWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model and during simulated clot retrieval, the Catheter must meet functionality specifications including compatibility with a stent retriever.The APRO 45 Catheter met the acceptance criteria.
Corrosion ResistanceCatheter shall show no signs of corrosion when tested in accordance with ISO 10555-1, Annex A.The APRO 45 Catheter met the acceptance criteria.
Sterilization (Ethylene Oxide)Verified to a sterility assurance level of 1 x 10⁻⁶ in accordance with ISO 11135.Met acceptance criteria.
Shelf-Life (Aging Studies)Subject device and packaging remain functional for the labeled expiration date; packaging integrity, seal strength, and device functionality met acceptance criteria.Met acceptance criteria. (1-year shelf life established)
Biocompatibility: SensitizationDid not elicit a sensitization response.Non-sensitizing
Biocompatibility: Irritation/Intracutaneous ReactivityNo evidence of irritation.Non-irritant
Biocompatibility: CytotoxicityDid not elicit a cytotoxic response at 24 hours and 48 hours.Non-cytotoxic
Biocompatibility: Hemolysis - IndirectNo significant differences between the test article extract and negative control.Non-hemolytic
Biocompatibility: Hemolysis - DirectNo differences between the hemolytic index of the test article and negative control.Non-hemolytic
Biocompatibility: Thrombogenicity (PTT)Average clotting time not significantly different from vehicle and negative control.Non-thrombogenic
Biocompatibility: Thrombogenicity (Platelet Leukocyte Count)Test article performed similar to comparator and negative controls.Met acceptance criteria (performed similar)
Biocompatibility: Thrombogenicity (Comparative Surface and Geometry Assessment)Similar surface morphology as the predicate device, no differences in roughness or defects.Subject device has similar surface morphology as predicate.
Biocompatibility: SC5b9 Complement ActivationSc5b9 concentration statistically less than positive control and not statistically higher than negative control.Acceptable
Biocompatibility: Acute Systemic ToxicityNo weight loss, mortality, or evidence of systemic toxicity.Non-toxic
Biocompatibility: Material-Mediated PyrogenicityTotal rise in temperature of < 0.5ºC; determined to be nonpyrogenic.Non-pyrogenic

Regarding the requested points specifically for AI/ML device evaluations:

Due to the nature of the device (a physical catheter) described in the provided FDA 510(k) clearance letter and summary, the following information points are not applicable and are not found in the document:

  1. Sample size used for the test set and the data provenance: This relates to data for algorithmic evaluation. Not applicable for a physical catheter.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for a physical catheter.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for a physical catheter.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted device.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable, as this is not an algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for a physical device is established through performance testing (bench, animal, biocompatibility) against engineering specifications and regulatory standards.
  7. The sample size for the training set: Not applicable, as this is not an AI/ML device requiring a training set.
  8. How the ground truth for the training set was established: Not applicable, as this is not an AI/ML device.

Summary of Study Type:

The study conducted to support the APRO 45 Catheter's 510(k) clearance was primarily non-clinical bench testing, sterilization validation, shelf-life studies, and biocompatibility testing. Animal testing and human clinical data were noted as "not deemed necessary" for this particular submission based on the substantial equivalence argument to the predicate device.

FDA 510(k) Clearance Letter - APRO 45 Catheter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 1, 2025

Alembic, LLC
Stina Almaleh
Regulatory and Quality Engineer
627 National Avenue
Mountain View, California 94043

Re: K251014
Trade/Device Name: APRO 45 Catheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: DQY, QJP
Dated: April 1, 2025
Received: April 2, 2025

Dear Stina Almaleh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251014 - Stina Almaleh
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251014 - Stina Almaleh
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan -S

Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K251014

Device Name
APRO 45 Catheter

Indications for Use (Describe)
The APRO 45 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 45 Catheter is also indicated for use as a conduit for retrieval devices.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

APRO 45 Catheter
510(k) Number: K251014
Page 1 of 5

This 510(k) Summary is provided in accordance with the requirements of 21 CFR §807.92.

1) Submitter information

Submitter: Alembic, LLC
627 National Ave.
Mountain View, CA 94043

Contact: Stina Almaleh
Regulatory and Quality Engineer
Email: salmaleh@alembicllc.com
Telephone Number: (650) 388-5080

Date Prepared: April 29, 2025

2) Device Name and Classification

Trade/Proprietary Name: APRO® 45 Catheter
Common Name: Percutaneous Catheter
Classification Name: Percutaneous Catheter, 21 CFR 870.1250
Catheter, Percutaneous, Neurovasculature, 21 CFR 870.1250
Regulatory Class: Class II
Product Codes: DQY, QJP
Review Panel: Cardiovascular, Neurology

3) Legally Marketed Predicate Device

Predicate Device: K234115 APRO 55 Catheter

4) Device Description

The APRO 45 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 45 Catheter is designed to facilitate the insertion and guidance of interventional devices into peripheral and neuro vasculature. Using standard catheterization techniques under fluoroscopic guidance, the APRO 45 Catheter is introduced through a guide catheter or guide sheath and over a guidewire into the target vasculature. The distal segment of the catheter shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal tip of the catheter for visualization under fluoroscopy. The APRO 45 Catheter is provided with an Introducer Sheath. The APRO 45 Catheter contains animal derived materials made from tallow derivatives.

5) Indications for Use

The APRO 45 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 45 Catheter is also indicated for use as a conduit for retrieval devices.

Page 6

APRO 45 Catheter
510(k) Number: K251014
Page 2 of 5

6) Technological Characteristics Comparison

Alembic has demonstrated that the APRO 45 Catheter is substantially equivalent to the predicate device based on the similarity in materials, similarity in design concept, and the same fundamental operating principles. A comparison of the APRO 45 Catheter with the predicate is summarized in Table 1 below.

Table 1. APRO 45 Catheter Comparison with the Predicate Device

CategorySubject Device APRO 45 CatheterPredicate Device APRO 55 Catheter
510(k) NumberK251014K234115
Regulatory ClassSame as predicate. Class II, 21 CFR 870.1250, DQY, QJPClass II, 21 CFR 870.1250, DQY, QJP
Indications for UseThe APRO 45 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 45 Catheter is also indicated for use as a conduit for retrieval devices.The APRO 55 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 55 Catheter is also indicated for use as a conduit for retrieval devices.

Materials
| Hub | Same as predicate. | Nylon |
| Adhesive | Same as predicate. | Cyanoacrylate |
| Strain Relief | Same as predicate. | Thermoplastic elastomer |
| Liner | Polytetrafluoroethylene (PTFE)/Pebax composite | PTFE/Tecoflex composite |
| Shaft Coil or Braid | Same as predicate. | 304V stainless steel braid 304V stainless steel coil |
| Extrusions | Polyether block amide | Thermoplastic polyurethanes, thermoplastic elastomer |
| Marker Band | Same as predicate. | Platinum/ iridium |
| Coating | Same as predicate. | Hydrophilic coating |

Dimensions
| Proximal Outer Diameter | 0.060 inch | 0.066 inch |
| Distal Outer Diameter | 0.058 inch | 0.066 inch |
| Inner Diameter | 0.045 inch | 0.055 inch |
| Effective Lengths | 115, 125, 146, 160 cm | 125, 137 cm |
| Coated Length | 60 cm | 90, 102 cm |
| Tip Shape | Same as predicate. | Straight |

Accessories
| Introducer Sheath | Same as predicate. | Yes |

Packaging Materials
| Pouch | Same as predicate. | Nylon/polyethylene/ Tyvek |
| Packaging Tube | N/A | Polyethylene |
| Packaging Hoop Coil | Polyethylene | N/A |
| Packaging Card | Same as predicate. | Polyethylene |
| Shelf Carton | Same as predicate. | Solid bleached sulfate paperboard |

Other
| Sterilization | Same as predicate. | Ethylene oxide |
| Shelf Life | 1 year | 6 months |
| Use Conditions | Same as predicate. | Sterile, single use, disposable |

Page 7

APRO 45 Catheter
510(k) Number: K251014
Page 3 of 5

7) Performance Data

Alembic performed non-clinical bench, sterility, shelf-life, and biocompatibility testing. The results demonstrate substantial equivalence of the APRO 45 Catheter to the legally marketed predicate device.

A. Design Verification Testing – Non-Clinical Bench

Bench performance testing was conducted to support the APRO 45 Catheter submission. The results of the design verification and validation testing confirm that the APRO 45 Catheter conforms to the predefined specifications and meets the test acceptance criteria. A summary of the testing is shown in Table 2.

Table 2. Summary of Non-Clinical Bench Test Results

TestAcceptance CriteriaConclusion
Visual and Dimensional CharacteristicsCatheter meets the visual and dimensional specifications.The APRO 45 Catheter met the acceptance criteria.
ParticulateCatheter meets the acceptance criteria. Subject device was evaluated with a predicate device under the same test conditions.The APRO 45 Catheter met the acceptance criteria.
Kink ResistanceCatheter shaft shall not kink at clinically relevant radii.The APRO 45 Catheter met the acceptance criteria.
Hub Air LeakageCatheter does not leak air into hub assembly with methods specified in ISO 10555-1, Annex D.The APRO 45 Catheter met the acceptance criteria.
Hub CompatibilityCatheter meets the requirements specified in ISO 80369-7.The APRO 45 Catheter met the acceptance criteria.
Torque StrengthCatheter must withstand the minimum required number of rotations without breakage and without kinking compared to legally marketed devices.The APRO 45 Catheter met the acceptance criteria.
Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 45 Catheter met the acceptance criteria.
Liquid LeakageCatheter must withstand pressure with methods specified in ISO 10555-1, Annex C.The APRO 45 Catheter met the acceptance criteria.
Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 45 Catheter met the acceptance criteria.
Hub and Shaft Tensile StrengthCatheter hub and shaft must meet tensile strength specification.The APRO 45 Catheter met the acceptance criteria.
Tip Tensile StrengthCatheter tip must meet tip tensile strength specification.The APRO 45 Catheter met the acceptance criteria.
3-Point BendCatheter 3-Point Bend force must be acceptable. Forces were compared to the predicate.The APRO 45 Catheter met the acceptance criteria.
Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to the predicate.The APRO 45 Catheter met the acceptance criteria.
Simulated UseWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model and during simulated clot retrieval, the Catheter must meet functionality specifications including compatibility with a stent retriever.The APRO 45 Catheter met the acceptance criteria.
Corrosion ResistanceCatheter shall show no signs of corrosion when tested in accordance with ISO 10555-1, Annex A.The APRO 45 Catheter met the acceptance criteria.

Page 8

APRO 45 Catheter
510(k) Number: K251014
Page 4 of 5

B. Design Verification Testing – Animal

Substantial equivalence was established based on non-clinical bench performance data. Animal testing data were not deemed necessary.

C. Sterilization and Shelf-Life

The APRO 45 Catheter is sterilized using an ethylene oxide sterilization cycle that was verified to a sterility assurance level of 1 x 10⁻⁶ in accordance with ISO 11135.

Aging studies for the APRO 45 Catheter have established that the subject device and packaging remain functional for the labeled expiration date. Aging studies for packaging integrity, seal strength, and the device functionality were performed and met the acceptance criteria.

D. Biocompatibility

Biocompatibility testing has been completed for the APRO 45 Catheter as an externally communicating device with circulating blood contact for a limited (≤ 24 hours) duration in accordance with ISO 10993-1 and the FDA Guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process"."

Table 3. Summary of Biocompatibility Test Results

TestResultsConclusions
Sensitization (Guinea Pig Maximization)The APRO 45 Catheter did not elicit a sensitization response.Non-sensitizing
Irritation/Intracutaneous ReactivityThe APRO 45 Catheter demonstrated no evidence of irritation.Non-irritant
Cytotoxicity (MEM Elution, L929 cells)The APRO 45 Catheter did not elicit a cytotoxic response at 24 hours and 48 hours.Non-cytotoxic
Hemolysis - IndirectThere were no significant differences between the test article extract and negative control article results.Non-hemolytic
Hemolysis - DirectThere were no differences between the hemolytic index of the test article and the negative control.Non-hemolytic
Thrombogenicity – Partial Thromboplastin Time (PTT)The average clotting time of the test article was not significantly different from the vehicle and negative control.Non-thrombogenic
Thrombogenicity – Platelet Leukocyte CountThe test article performed similar to comparator and negative controls.
Thrombogenicity – Comparative Surface and Geometry AssessmentComparative surface assessment using 40x optical microscopy at representative locations of the subject device and the predicate device did not find any differences in the roughness or presence of any defects. The subject device has similar surface morphology as the predicate device APRO 55 Catheter.
SC5b9 Complement ActivationThe Sc5b9 concentration of the test article was statistically less than the positive control and was not statistically higher than the negative control.Acceptable
Acute Systemic ToxicityNo weight loss, mortality, or evidence of systemic toxicity from the extract exposure to the mice.Non-toxic

Page 9

APRO 45 Catheter
510(k) Number: K251014
Page 5 of 5

TestResultsConclusions
Material-Mediated PyrogenicityAll individual rabbits for both the test article and negative control showed a total rise in temperature of < 0.5ºC and were determined to be nonpyrogenic.Non-pyrogenic

E. Clinical Testing

Substantial equivalence was established based on non-clinical performance data. Human clinical data were not deemed necessary.

8) Conclusion

Based on the comparison of the technological characteristics and the non-clinical testing, the subject device is found to be substantially equivalent to the predicate device. The differences in technological characteristics do not raise new questions of safety and effectiveness.

Testing was conducted to demonstrate that the subject device meets the specifications and performs as intended.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).