K Number
K250831

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-04-23

(35 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Annalise Enterprise is a device designed to be used in the medical care environment to aid in triage and prioritization of studies with features suggestive of the following findings:

  • pleural effusion* [1]
  • pneumoperitoneum* [2]
  • pneumothorax
  • tension pneumothorax
  • vertebral compression fracture* [3]

*See additional information below.

The device analyzes studies using an artificial intelligence algorithm to identify findings. It makes study-level output available to an order and imaging management system for worklist prioritization or triage.

The device is not intended to direct attention to specific portions of an image and only provides notification for suspected findings.

Its results are not intended:

  • to be used on a standalone basis for clinical decision making
  • to rule out specific findings, or otherwise preclude clinical assessment of chest X-ray studies

Intended modality:
Annalise Enterprise identifies suspected findings in digitized (CR) or digital (DX) chest X-ray studies.

Intended user:
The device is intended to be used by trained clinicians who are qualified to interpret chest X-ray studies as part of their scope of practice.

Intended patient population:
The intended population is patients who are 22 years or older.

Additional information:
The following additional information relates to the findings listed above:

[1] Pleural effusion

  • specificity may be reduced in the presence of scarring and/or pleural thickening
  • standalone performance evaluation was performed on a dataset that included supine and erect positioning
  • use of this device with prone positioning may result in differences in performance

[2] Pneumoperitoneum

  • standalone performance evaluation was performed on a dataset that included supine and erect positioning where most cases were of unilateral right-sided and bilateral pneumoperitoneum
  • use of this device with prone positioning and for unilateral left-sided pneumoperitoneum may result in differences in performance

[3] Vertebral compression fracture

  • intended for prioritization or triage of worklists of Bone Health and Fracture Liaison Service program clinicians
  • standalone performance evaluation was performed on a dataset that included only erect positioning
  • use of this device with supine positioning may result in differences in performance
Device Description

Annalise Enterprise is a software workflow tool which uses an artificial intelligence (AI) algorithm to identify suspected findings on chest X-ray studies in the medical care environment. The findings identified by the device include pneumothorax, tension pneumothorax, pleural effusion, pneumoperitoneum and vertebral compression fracture.

Radiological findings are identified by the device using an AI algorithm – a convolutional neural network trained using deep-learning techniques. Images used to train the algorithm were sourced from datasets that included a range of equipment manufacturers including. This dataset, which contained over 750,000 chest X-ray imaging studies, was labelled by trained radiologists regarding the presence of the findings of interest.

The performance of the device's AI algorithm was validated in a standalone performance evaluation, in which the case-level output from the device was compared with a reference standard ('ground truth'). This was determined by two ground truthers, with a third truther used in the event of disagreement. All truthers were US board-certified radiologists.

The device interfaces with image and order management systems (such as PACS/RIS) to obtain chest X-ray studies for processing by the AI algorithm. Following processing, if any of the clinical findings of interest are identified in the study, the device provides a notification to the image and order management system for prioritization of that study in the worklist. This enables users to review the studies containing features suggestive of these clinical findings earlier than in the standard clinical workflow. It is important to note that the device will never decrease a study's existing priority in the worklist. This ensures that worklist items will never have their priorities downgraded based on AI results.

The device workflow is performed parallel to and in conjunction with the standard clinical workflow for interpretation of chest X-ray studies. The device is intended to aid in prioritization and triage of radiological medical images only.

AI/ML Overview

The Annalise Enterprise device is designed to aid in the triage and prioritization of chest X-ray studies by identifying features suggestive of several findings. The following outlines the acceptance criteria and the study conducted to prove the device meets these criteria.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for each finding are implicitly demonstrated by the reported Area Under the Curve (AUC), Sensitivity, and Specificity values, aiming for high performance in triaging positive cases while minimizing false positives. The reported device performance for each finding at various operating points is:

FindingProduct CodeAUC (95% CI)Operating Point (Threshold)Sensitivity % (Se) (95% CI)Specificity % (Sp) (95% CI)
PneumothoraxQFM0.984 (0.976, 0.990)0.20097.1 (95.5, 98.6)88.2 (85.4, 90.8)
0.25096.2 (94.3, 98.1)91.9 (89.5, 94.1)
0.30095.0 (92.8, 97.1)94.1 (91.9, 95.9)
0.35093.1 (90.7, 95.5)95.6 (93.7, 97.2)
0.40090.7 (88.0, 93.3)96.7 (95.0, 98.2)
Tension PneumothoraxQFM0.989 (0.984, 0.994)0.22596.0 (92.0, 99.2)94.0 (92.3, 95.6)
0.25095.2 (91.2, 98.4)94.6 (93.1, 96.2)
0.30093.6 (88.8, 97.6)95.6 (94.1, 96.9)
0.35089.6 (84.0, 94.4)96.6 (95.3, 97.8)
0.40087.2 (80.8, 92.8)97.5 (96.4, 98.6)
PneumoperitoneumQAS0.987 (0.976, 0.994)0.25096.2 (92.4, 99.0)87.9 (83.2, 92.1)
0.30094.3 (89.5, 98.1)90.5 (86.3, 94.2)
0.35092.4 (86.7, 97.1)93.7 (90.0, 96.8)
0.40091.4 (85.7, 96.2)95.8 (92.6, 98.4)
0.45087.6 (81.0, 93.3)98.4 (96.3, 100.0)
Pleural EffusionQFM0.977 (0.969, 0.984)0.38096.7 (95.0, 98.1)86.8 (83.6, 89.5)
0.42594.4 (92.3, 96.5)89.5 (86.8, 92.1)
0.45092.9 (90.7, 95.0)91.3 (88.6, 93.7)
0.47589.8 (87.1, 92.3)93.7 (91.5, 95.9)
0.50087.6 (84.6, 90.5)95.5 (93.5, 97.0)
Vertebral Compression FxQFM0.972 (0.960, 0.982)0.46093.4 (90.1, 96.0)85.8 (82.1, 89.6)
0.50092.6 (89.3, 95.6)90.9 (87.7, 93.7)
0.55087.1 (83.1, 90.8)94.7 (91.8, 96.9)

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The standalone performance evaluation was conducted on a total dataset of 3,252 cases.
  • Data Provenance: The data was collected retrospectively and anonymized. Cases were collected consecutively from four US hospital network sites. The datasets included a variety of patient demographics (gender, age, ethnicity, race) and technical parameters (imaging equipment make, model), indicating a diverse geographic (US) and technical (various scanner manufacturers: Agfa, Carestream, Fujifilm, GE Healthcare, Kodak, Konica Minolta, McKesson, Philips, Siemens, Varian) origin.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: At least two ABR-certified radiologists were used for each de-identified case. A third radiologist was used in the event of disagreement.
  • Qualifications: All truthers were US board-certified radiologists who interpret chest X-rays as part of their regular clinical practice and were protocol-trained.

4. Adjudication Method for the Test Set

The adjudication method used was 2+1 consensus. Each deidentified case was annotated by at least two ground truthers (radiologists), and consensus was determined by these two. In the event of disagreement between the first two, a third ground truther was used to resolve the discrepancy and establish the final ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The provided information does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done to assess how human readers improve with AI vs. without AI assistance. The study focuses on the standalone performance of the AI algorithm and its impact on triage effectiveness (turn-around time).

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

Yes, a standalone performance evaluation was done. The key results table and associated metrics (AUC, Sensitivity, Specificity) are specifically for the device's AI algorithm independent of human intervention. The study describes "case-level output from the device was compared with a reference standard ('ground truth')", confirming a standalone evaluation.

7. Type of Ground Truth Used

The ground truth used was expert consensus, established by multiple US board-certified radiologists using a 2+1 adjudication method.

8. Sample Size for the Training Set

The training dataset used to train the Convolutional Neural Network (CNN) algorithm contained over 750,000 chest X-ray imaging studies.

9. How the Ground Truth for the Training Set was Established

The studies in the training dataset were labelled by trained radiologists regarding the presence of the findings of interest. The document does not specify the exact number of radiologists or the specific consensus or adjudication method used for the training set, only that they were "trained radiologists".

FDA 510(k) Clearance Letter - Annalise Enterprise

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 23, 2025

Annalise-AI
Haylee Bosshard
Regulatory Affairs Manager
Level P, 24 Campbell St
Sydney, NSW 2000
Australia

Re: K250831
Trade/Device Name: Annalise Enterprise
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer aided triage and notification software
Regulatory Class: Class II
Product Code: QFM, QAS
Dated: April 17, 2025
Received: April 17, 2025

Dear Haylee Bosshard:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250831 - Haylee Bosshard
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250831 - Haylee Bosshard
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K250831

Device Name
Annalise Enterprise

Indications for Use (Describe)

Intended context:
Annalise Enterprise is a device designed to be used in the medical care environment to aid in triage and prioritization of studies with features suggestive of the following findings:

  • pleural effusion* [1]
  • pneumoperitoneum* [2]
  • pneumothorax
  • tension pneumothorax
  • vertebral compression fracture* [3]

*See additional information below.

The device analyzes studies using an artificial intelligence algorithm to identify findings. It makes study-level output available to an order and imaging management system for worklist prioritization or triage.

The device is not intended to direct attention to specific portions of an image and only provides notification for suspected findings.

Its results are not intended:

  • to be used on a standalone basis for clinical decision making
  • to rule out specific findings, or otherwise preclude clinical assessment of chest X-ray studies

Intended modality:
Annalise Enterprise identifies suspected findings in digitized (CR) or digital (DX) chest X-ray studies.

Intended user:
The device is intended to be used by trained clinicians who are qualified to interpret chest X-ray studies as part of their scope of practice.

Intended patient population:
The intended population is patients who are 22 years or older.

Additional information:
The following additional information relates to the findings listed above:

[1] Pleural effusion

  • specificity may be reduced in the presence of scarring and/or pleural thickening
  • standalone performance evaluation was performed on a dataset that included supine and erect positioning
  • use of this device with prone positioning may result in differences in performance

[2] Pneumoperitoneum

  • standalone performance evaluation was performed on a dataset that included supine and erect

Page 5

positioning where most cases were of unilateral right-sided and bilateral pneumoperitoneum

  • use of this device with prone positioning and for unilateral left-sided pneumoperitoneum may result in differences in performance

[3] Vertebral compression fracture

  • intended for prioritization or triage of worklists of Bone Health and Fracture Liaison Service program clinicians
  • standalone performance evaluation was performed on a dataset that included only erect positioning
  • use of this device with supine positioning may result in differences in performance

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 6

Page 1

510(k) Summary

I. SUBMITTER

Company NameAnnalise-AI
AddressLevel P, 24 Campbell StreetSydney, NSW 2000Australia
Phone Number+61 1800-958487
Contact PersonHaylee Bosshard
Date PreparedApril 23, 2025

II. SUBJECT DEVICE

Manufacturer NameAnnalise-AI
Device NameAnnalise Enterprise
Classification NameRadiological computer aided triage and notification software (21CFR892.2080)
Regulatory ClassII
Product CodeQFM, QAS*

III. PREDICATE DEVICE

Manufacturer NameAnnalise-AI
Device NameAnnalise Enterprise CXR Triage Trauma
510(k) referenceK222179
Classification NameRadiological computer aided triage and notification software (21CFR892.2080)
Regulatory ClassII
Product CodeQFM, QAS*

This predicate has not been subject to a design-related recall. No reference devices were used in this submission.

*Note that product code QAS only applies to clinical condition of interest: pneumoperitoneum. All other conditions are submitted under product code QFM.

K250831

Page 7

Page 2

IV. DEVICE DESCRIPTION

Annalise Enterprise is a software workflow tool which uses an artificial intelligence (AI) algorithm to identify suspected findings on chest X-ray studies in the medical care environment. The findings identified by the device include pneumothorax, tension pneumothorax, pleural effusion, pneumoperitoneum and vertebral compression fracture.

Radiological findings are identified by the device using an AI algorithm – a convolutional neural network trained using deep-learning techniques. Images used to train the algorithm were sourced from datasets that included a range of equipment manufacturers including. This dataset, which contained over 750,000 chest X-ray imaging studies, was labelled by trained radiologists regarding the presence of the findings of interest.

The performance of the device's AI algorithm was validated in a standalone performance evaluation, in which the case-level output from the device was compared with a reference standard ('ground truth'). This was determined by two ground truthers, with a third truther used in the event of disagreement. All truthers were US board-certified radiologists.

The device interfaces with image and order management systems (such as PACS/RIS) to obtain chest X-ray studies for processing by the AI algorithm. Following processing, if any of the clinical findings of interest are identified in the study, the device provides a notification to the image and order management system for prioritization of that study in the worklist. This enables users to review the studies containing features suggestive of these clinical findings earlier than in the standard clinical workflow. It is important to note that the device will never decrease a study's existing priority in the worklist. This ensures that worklist items will never have their priorities downgraded based on AI results.

The device workflow is performed parallel to and in conjunction with the standard clinical workflow for interpretation of chest X-ray studies. The device is intended to aid in prioritization and triage of radiological medical images only.

Page 8

Page 3

V. INDICATIONS FOR USE

The Indications for Use statement is as follows:

Intended contextAnnalise Enterprise is a device designed to be used in the medical care environment to aid in triage and prioritization of studies with features suggestive of the following findings:• pleural effusion¹• pneumoperitoneum²• pneumothorax• tension pneumothorax• vertebral compression fracture³See Additional Information below.The device analyzes studies using an artificial intelligence algorithm to identify findings. It makes study-level output available to an order and imaging management system for worklist prioritization or triage.The device is not intended to direct attention to specific portions of an image and only provides notification for suspected findings.Its results are not intended:• to be used on a standalone basis for clinical decision making• to rule out specific findings, or otherwise preclude clinical assessment of chest X-ray studies
Intended modalityAnnalise Enterprise identifies suspected findings in digitized (CR) or digital (DX) chest X-ray studies.
Intended userThe device is intended to be used by trained clinicians who are qualified to interpret chest X-ray studies as part of their scope of practice.
Intended patient populationThe intended population is patients who are 22 years or older.
Additional informationThe following additional information relates to the findings listed above:¹ Pleural Effusion• specificity may be reduced in the presence of scarring and/or pleural thickening• standalone performance evaluation was performed on a dataset that included supine and erect positioning• use of this device with prone positioning may result in differences in performance² Pneumoperitoneum

Page 9

Page 4

• standalone performance evaluation was performed on a dataset that included supine and erect positioning where most cases were of unilateral right-sided and bilateral pneumoperitoneum
• use of this device with prone positioning and for unilateral left-sided pneumoperitoneum may result in differences in performance

³ Vertebral Compression Fracture
• intended for prioritization or triage of worklists of Bone Health and Fracture Liaison Service program clinicians
• standalone performance evaluation was performed on a dataset that included only erect positioning
• use of this device with supine positioning may result in differences in performance

The Indications for Use statement of the subject device differs to the predicate device only in the clinical conditions of interest, however a standalone performance evaluation was conducted and demonstrated that the device is as safe and effective for its intended use. Both the subject and predicate device are intended for use to assist with worklist triage by providing notifications of suspected findings and their associated priority.

Page 10

Page 5

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject device was evaluated and compared to the predicate device with respect to the following characteristics:

  1. Indications for Use
  2. Target population
  3. Anatomical site and modality
  4. Intended user and clinical use environment
  5. Technical method for notification and prioritization
  6. Device input and radiological image protocol
  7. Device output and means of notification to user
  8. System components
  9. Location where results are received
  10. Prioritization relationship to standard of care workflow
  11. AI models and performance of algorithm.

The following characteristics showed a difference between the subject and predicate devices. The different characteristics include:

  1. Set of findings and algorithm
  2. AI models and performance of algorithm

All differences were technological characteristic differences that do not raise different questions of safety and effectiveness. Furthermore, we have provided a standalone performance study for this submission to evaluate these differences and establish substantial equivalence.

Page 11

Page 6

VII. PERFORMANCE DATA

The following performance data have been provided to support evaluation of substantial equivalence.

A. Software Verification and Validation Testing

Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions - Guidance for Industry and Food and Drug Administration Staff", June, 2023.

B. Performance Testing

Performance of the subject device was assessed in four performance studies to satisfy requirements set forth in the special controls per 21CFR892.2080. These included standalone performance and triage effectiveness evaluations.

Standalone performance was assessed via a retrospective, anonymized study of adult patient, DICOM-compliant chest X-ray cases. The test dataset used during the standalone performance evaluation was previously acquired (K213941, K222268 and K222179) and independent from the training dataset used in model development. The standalone performance study was conducted on four independently assessed cohorts which equated to a total dataset of 3,252 cases collected consecutively from four US hospital network sites.

The performance testing datasets included representation across subgroups for patient demographics (gender [female: 44.7-59.0%, male: 41.0-55.3], age [mean: 62.2-67.4 years, min: 22, max: 99-105], ethnicity [Hispanic: 6.5-8.3%, Unavailable: 2.6-6.5%], race [White/Caucasian: 80.3-84.9%, Other: 12.9-17.4%, Unavailable: 2.2-3.9%]), co-existing findings or abnormalities and technical parameters (imaging equipment make, model). The datasets included Agfa, Carestream, Fujifilm, GE Healthcare, Kodak, Konica Minolta, McKesson, Philips, Siemens, Varian X-ray scanners for the pivotal study.

To determine the ground truth, each deidentified case was annotated in a blinded fashion by at least two ABR-certified and protocol-trained radiologists who interpret chest X-ray as part of regular clinical practice (ground truthers), with consensus determined by two ground truthers and a third ground truther in the event of disagreement. The key results of the study are summarized in the table below.

FindingProduct CodeAUC (95% CI)
PneumothoraxQFM0.984 (0.976, 0.990)
Tension pneumothoraxQFM0.989 (0.984, 0.994)
PneumoperitoneumQAS0.987 (0.976, 0.994)
Pleural effusionQFM0.977 (0.969, 0.984)
Vertebral compression fractureQFM0.972 (0.960, 0.982)

Page 12

Page 7

FindingOperating PointSensitivity % (Se) (95% CI)Specificity % (Sp) (95% CI)
Pneumothorax0.20097.1 (95.5,98.6)88.2 (85.4,90.8)
0.25096.2 (94.3,98.1)91.9 (89.5,94.1)
0.30095.0 (92.8,97.1)94.1 (91.9,95.9)
0.35093.1 (90.7,95.5)95.6 (93.7,97.2)
0.40090.7 (88.0,93.3)96.7 (95.0,98.2)
Tension pneumothorax0.22596.0 (92.0,99.2)94.0 (92.3,95.6)
0.25095.2 (91.2,98.4)94.6 (93.1,96.2)
0.30093.6 (88.8,97.6)95.6 (94.1,96.9)
0.35089.6 (84.0,94.4)96.6 (95.3,97.8)
0.40087.2 (80.8,92.8)97.5 (96.4,98.6)
Pneumoperitoneum0.25096.2 (92.4,99.0)87.9 (83.2,92.1)
0.30094.3 (89.5,98.1)90.5 (86.3,94.2)
0.35092.4 (86.7,97.1)93.7 (90.0,96.8)
0.40091.4 (85.7,96.2)95.8 (92.6,98.4)
0.45087.6 (81.0,93.3)98.4 (96.3,100.0)
Pleural effusion0.38096.7 (95.0,98.1)86.8 (83.6,89.5)
0.42594.4 (92.3,96.5)89.5 (86.8,92.1)
0.45092.9 (90.7,95.0)91.3 (88.6,93.7)
0.47589.8 (87.1,92.3)93.7 (91.5,95.9)
0.50087.6 (84.6,90.5)95.5 (93.5,97.0)
Vertebral compression fracture0.46093.4 (90.1,96.0)85.8 (82.1,89.6)
0.50092.6 (89.3,95.6)90.9 (87.7,93.7)
0.55087.1 (83.1,90.8)94.7 (91.8,96.9)

The results demonstrate the subject device establishes effective triage within a clinician's queue based on high sensitivity and specificity. Further, these results are substantially equivalent to those of the predicate device.

Triage effectiveness (turn-around time) was assessed by an internal bench study using a dataset of n=303 cases positive for any of the findings eligible for prioritization. These cases were collected from multiple data sources spanning a variety of geographical locations, patient demographics and technical characteristics. The results demonstrated a triage turn-around time of 42.3 (95% CI: 41.2, 43.4) seconds, which is substantially equivalent to the total performance time published for the predicate device.

Therefore, the subject device has been shown to satisfy the performance requirements per 21CFR892.2080, for 'Radiological computer aided triage and notification software', by providing clinically effective triage for non-contrast brain CT studies containing features suggestive of clinical findings of interest. This data demonstrates the subject device is safe and effective for its intended use, and thereby supports substantial equivalence.

Page 13

Page 8

VIII. CONCLUSIONS

The subject device and the predicate device are both software only packages, devices intended to assist with worklist triage by providing notification of findings. The subject and predicate devices utilize the same principles of operation and work in parallel to the current standard of care workflow.

Both the subject and predicate devices use an artificial intelligence algorithm to identify findings in images and require the same inputs (DICOM image data) and provide the same outputs (prioritization for a medical worklist).

The technological differences between the subject and predicate devices do not raise new questions of safety and effectiveness.

Standalone performance testing and the comparison of technological characteristics with the predicate devices shows that the subject device:

  • performs as intended,
  • is safe and effective for its intended use, and
  • is therefore substantially equivalent to the predicate device.

§ 892.2080 Radiological computer aided triage and notification software.

(a)
Identification. Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
e.g., improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.