K Number
K222179

Validate with FDA (Live)

Date Cleared
2023-03-28

(249 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Annalise Enterprise CXR Triage Trauma is a software workflow tool designed to aid the clinical assessment of adult chest X-ray cases with features suggestive of pleural effusion and pneumoperitoneum in the medical care environment.

The device analyzes cases using an artificial intelligence algorithm to identify findings. It makes case-level output available to a PACS or RIS for worklist prioritization or triage.

The device is intended to be used by trained clinicians who are qualified to interpret chest X-rays as part of their scope of practice.

The device is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and pneumoperitoneum.

Its results are not intended to be used on a standalone basis for clinical decision making nor is it intended to rule out specific critical findings, or otherwise preclude clinical assessment of X-ray cases.

Standalone performance evaluation of the device was performed on a dataset that included supine and erect positioning. Use of this device with prone positioning may result in differences in performance.

Standalone performance evaluation of the device was performed on a dataset where most cases were of unilateral right sided and bilateral pneumoperitoneum. Use of this device for unilateral left sided pneumoperitoneum may result in differences in performance.

Specificity may be reduced for pleural effusion in the presence of scarring and and/or pleural thickening.

Device Description

Annalise Enterprise CXR Triage Trauma is a software workflow tool which uses an artificial intelligence (AI) algorithm to identify suspected findings on chest X-ray (CXR) studies in the medical care environment. The findings identified by the device include pleural effusion and pneumoperitoneum.

Radiological findings are identified by the device using an AI algorithm - a convolutional neural network trained using deep-learning techniques. Images used to train the algorithm were sourced from datasets across three continents, including a range of equipment manufacturers and models. The performance of the device's AI algorithm was validated in a standalone performance evaluation, in which the case-level output from the device was compared with a reference standard ('ground truth'). This was determined by two ground truthers, with a third truther used in the event of disagreement. All truthers were US board-certified radiologists.

The device interfaces with image and order management systems (such as PACS/RIS) to obtain CXR studies for processing by the AI algorithm. Following processing, if any of the clinical findings of interest are identified in a CXR study, the device provides a notification to the image and order management system for prioritization of that study in the worklist. This enables users to review the studies containing features suggestive of these clinical findings earlier than in the standard clinical workflow. It is important to note that the device will never decrease a study's existing priority in the worklist. This ensures that worklist items will never have their priorities downgraded based on AI results.

The device workflow is performed parallel to and in conjunction with the standard clinical workflow for interpretation of CXRs. The device is intended to aid in prioritization and triage of radiological medical images only.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

Acceptance Criteria and Device Performance

The document describes the performance of the Annalise Enterprise CXR Triage Trauma device in identifying pleural effusion and pneumoperitoneum. The acceptance criteria are implicitly demonstrated by the reported sensitivities and specificities at various operating points, which are presented as evidence of effective triage and substantial equivalence to the predicate device. While explicit "acceptance criteria" are not listed as pass/fail thresholds in the provided text, the performance metrics below are what the sponsor used to demonstrate the device met the requirements.

Table of Reported Device Performance:

FindingOperating pointSensitivity (95% CI)Specificity (95% CI)AUC (95% CI)
Pleural effusion0.230296.0 (94.2,97.7)88.3 (85.3,91.1)0.980 (0.972-0.986)
0.299093.8 (91.5,95.8)91.7 (89.3,94.1)
0.435586.3 (83.0,89.4)95.6 (93.7,97.2)
Pneumoperitoneum0.032290.1 (84.2,95.0)87.4 (82.4,92.3)0.969 (0.950-0.984)
0.048486.1 (79.2,92.1)89.6 (85.2,94.0)
0.226682.2 (75.2,89.1)96.2 (93.4, 98.9)

Additionally, the device achieved an average triage turn-around time under 30.0 seconds. This implies an implicit acceptance criterion for speed, aiming for a rapid triage capability.

Study Details

1. Sample Sizes and Data Provenance

  • Test Set Sample Size: A total of 1269 CXR cases (582 positive, 687 negative) were used across two independent cohorts.
  • Data Provenance: The data was collected retrospectively and anonymized from four US hospital network sites. The cases were collected consecutively. The text also mentions that training data was sourced from "datasets across three continents."

2. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: For ground truth determination, a minimum of two experts were used, with a third expert used in case of disagreement.
  • Qualifications: All experts (ground truthers) were ABR-certified and protocol-trained radiologists. No specific years of experience are mentioned.

3. Adjudication Method for the Test Set

  • The adjudication method for establishing ground truth was 2+1 consensus. This means two ground truthers blinded to each other's assessments first reviewed the cases, and a third ground truther was consulted to resolve any disagreements.

4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned as being performed to show how human readers improve with AI vs. without AI assistance. The performance evaluation focused on the standalone performance of the AI algorithm for triage effectiveness.

5. Standalone Performance (Algorithm Only)

  • Yes, a standalone performance evaluation was performed. The reported sensitivities, specificities, and AUC values in the table above demonstrate the algorithm's performance without integration into human workflow for diagnostic purposes, although it is evaluated for its "triage effectiveness."

6. Type of Ground Truth Used

  • The ground truth was established by expert consensus of ABR-certified and protocol-trained radiologists.

7. Sample Size for the Training Set

  • The specific sample size for the training set is not provided in the given text. It only states that the training dataset was "independent from the training dataset used in model development" relative to the test set and sourced from "datasets across three continents."

8. How the Ground Truth for the Training Set was Established

  • The text does not explicitly detail how the ground truth for the training set was established. It only states that the AI algorithm was "trained using deep-learning techniques" and that the "Images used to train the algorithm were sourced from datasets across three continents." It can be inferred that a similar expert review process, likely by radiologists, would have been used to label the training data, but this is not explicitly stated in the provided excerpt.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.

Annalise-AI Pty Ltd % Haylee Bosshard Regulatory Affairs Manager Level P, 24 Campbell St. Sydney, New South Wales 2000 AUSTRALIA

March 28, 2023

Re: K222179

Trade/Device Name: Annalise Enterprise CXR Triage Trauma Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QFM, QAS Dated: February 24, 2023 Received: February 24, 2023

Dear Minta Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222179

Device Name Annalise Enterprise CXR Triage Trauma

Indications for Use (Describe)

Annalise Enterprise CXR Triage Trauma is a software workflow tool designed to aid the clinical assessment of adult chest X-ray cases with features suggestive of pleural effusion and pneumoperitoneum in the medical care environment.

The device analyzes cases using an artificial intelligence algorithm to identify findings. It makes case-level output available to a PACS or RIS for worklist prioritization or triage.

The device is intended to be used by trained clinicians who are qualified to interpret chest X-ravs as part of their scope of practice.

The device is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and pneumoperitoneum.

Its results are not intended to be used on a standalone basis for clinical decision making nor is it intended to rule out specific critical findings, or otherwise preclude clinical assessment of X-ray cases.

Standalone performance evaluation of the device was performed on a dataset that included supine and erect positioning. Use of this device with prone positioning may result in differences in performance.

Standalone performance evaluation of the device was performed on a dataset where most cases were of unilateral right sided and bilateral pneumoperitoneum. Use of this device for unilateral left sided pneumoperitoneum may result in differences in performance.

Specificity may be reduced for pleural effusion in the presence of scarring and/or pleural thickening.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K222179

^annalise.ai

510(k) Summary

SUBMITTER I.

Company NameAnnalise-AI Pty Ltd
AddressLevel P, 24 Campbell StSydney, NSW 2000Australia
Phone Number+61 1800 958 487
Contact PersonHaylee Bosshard
Date PreparedJuly 22, 2022

II. SUBJECT DEVICE

Manufacturer NameAnnalise-AI Pty Ltd
Device NameAnnalise Enterprise CXR Triage Trauma
510(k) referenceK222179
Classification NameRadiological computer aided triage and notification software(21CFR892.2080)
Regulatory ClassII
Product CodeQFM+, QAS*
  • Note that the product code QFM applies to pleural effusion.
  • Note that the product code QAS applies to pneumoperitoneum.

PREDICATE DEVICE III.

Manufacturer NameAnnalise-AI Pty Ltd
Device NameAnnalise Enterprise CXR Triage Pneumothorax
510(k) referenceK213941
Classification NameRadiological computer aided triage and notification software(21CFR892.2080)
Regulatory ClassII
Product CodeQFM

This predicate has not been subject to a design-related recall.

{4}------------------------------------------------

Aanr

Manufacturer NameZebra Medical Vision Ltd.
Device NameHealthPPT
510(k) referenceK211803
Classification NameRadiological computer aided triage and notification software(21CFR892.2080)
Regulatory ClassII
Product CodeQFM

IV. REFERENCE DEVICE

V. DEVICE DESCRIPTION

Annalise Enterprise CXR Triage Trauma is a software workflow tool which uses an artificial intelligence (AI) algorithm to identify suspected findings on chest X-ray (CXR) studies in the medical care environment. The findings identified by the device include pleural effusion and pneumoperitoneum.

Radiological findings are identified by the device using an AI algorithm - a convolutional neural network trained using deep-learning techniques. Images used to train the algorithm were sourced from datasets across three continents, including a range of equipment manufacturers and models. The performance of the device's AI algorithm was validated in a standalone performance evaluation, in which the case-level output from the device was compared with a reference standard ('ground truth'). This was determined by two ground truthers, with a third truther used in the event of disagreement. All truthers were US board-certified radiologists.

The device interfaces with image and order management systems (such as PACS/RIS) to obtain CXR studies for processing by the AI algorithm. Following processing, if any of the clinical findings of interest are identified in a CXR study, the device provides a notification to the image and order management system for prioritization of that study in the worklist. This enables users to review the studies containing features suggestive of these clinical findings earlier than in the standard clinical workflow. It is important to note that the device will never decrease a study's existing priority in the worklist. This ensures that worklist items will never have their priorities downgraded based on AI results.

The device workflow is performed parallel to and in conjunction with the standard clinical workflow for interpretation of CXRs. The device is intended to aid in prioritization and triage of radiological medical images only.

{5}------------------------------------------------

VI. INDICATIONS FOR USE

The Indications for Use statement is as follows:

Annalise Enterprise CXR Triage Trauma is a software workflow tool designed to aid the clinical assessment of adult chest X-ray cases with features suggestive of pleural effusion and pneumoperitoneum in the medical care environment.

The device analyzes cases using an artificial intelligence algorithm to identify findings. It makes case-level output available to a PACS or RIS for worklist prioritization or triage.

The device is intended to be used by trained clinicians who are qualified to interpret chest X-rays as part of their scope of practice.

The device is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and pneumoperitoneum.

Its results are not intended to be used on a standalone basis for clinical decision making nor is it intended to rule out specific critical findings, or otherwise preclude clinical assessment of X-ray cases.

Standalone performance evaluation of the device was performed on a dataset that included supine and erect positioning. Use of this device with prone positioning may result in differences in performance.

Standalone performance evaluation of the device was performed on a dataset where most cases were of unilateral right sided and bilateral pneumoperitoneum. Use of this device for unilateral left sided pneumoperitoneum may result in differences in performance.

Specificity may be reduced for pleural effusion in the presence of scarring and/or pleural thickening.

The Indications for Use statement of the subject device differs to the predicate device only in the clinical conditions of interest, however a standalone performance evaluation was conducted and demonstrated that the device is safe and effective for its intended use. Both the subject and predicate devices are intended for use to assist with worklist triage by providing notifications of suspected findings and their associated priority.

{6}------------------------------------------------

Aannalisc

VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject device was evaluated and compared to the predicate device with respect to the following characteristics:

    1. Indications for Use
  • Target population 2.
    1. Anatomical site and modality
    1. Intended user and clinical use environment
    1. Technical method for notification and prioritization
    1. Device input and radiological image protocol
    1. Device output and means of notification to user
    1. System components
    1. Location where results are received
    1. Prioritization relationship to standard of care workflow
    1. Ability to support effective triage
    1. Set of findings

The difference between the two devices is a configuration that limits the output to the set of findings that the subject device is intended to identify. To address this difference, standalone performance data was supplied to support the conclusion that the subject device does not raise different questions of safety and effectiveness and that the subject device is as safe and effective as the predicate device.

VIII. PERFORMANCE DATA

The following performance data have been provided to support evaluation of substantial equivalence.

A. Software Verification and Validation Testing

Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", May 11, 2005.

{7}------------------------------------------------

^annalise.ai

B. Performance Testing

Performance of the subject device was assessed in two performance studies to satisfy requirements set forth in the special controls per 21CFR892.2080. These included standalone performance and triage effectiveness evaluations.

Standalone performance was assessed via a retrospective, anonymized study of adult patient, DICOM-compliant CXR cases. The test dataset used during the standalone performance evaluation was newly acquired and independent from the training dataset used in model development. The standalone performance study was conducted on two independent cohorts which equated to a total dataset of 1269 CXR cases (positive n=582 and negative n=687), collected consecutively from four US hospital network sites. The study included performance testing data obtained via an initial study protocol and additional performance testing data obtained via a study protocol addendum. Each of the two cohorts included representation across subgroups for patient demographics (gender, age, ethnicity, race), technical parameters (imaging equipment make, model), imaging parameters (positioning, projections), co-existing findings or abnormalities, finding characteristics (size and laterality) and common mimics. To determine the ground truth, each deidentified CXR case was annotated in a blinded fashion by ABR-certified and protocoltrained radiologists (ground truthers), with consensus determined by two ground truthers and a third ground truther in the event of disagreement for the primary finding. The key results of the study are summarized in the tables below.

FindingProduct CodeAUC (95% CI)
Pleural effusionQFM0.980 (0.972-0.986)
PneumoperitoneumQAS0.969 (0.950-0.984)
FindingOperating pointSensitivity (95% CI)Specificity (95% CI)
Pleural effusion0.230296.0 (94.2,97.7)88.3 (85.3,91.1)
0.299093.8 (91.5,95.8)91.7 (89.3,94.1)
0.435586.3 (83.0,89.4)95.6 (93.7,97.2)
Pneumoperitoneum0.032290.1 (84.2,95.0)87.4 (82.4,92.3)
0.048486.1 (79.2,92.1)89.6 (85.2,94.0)
0.226682.2 (75.2,89.1)96.2 (93.4, 98.9)

The results demonstrate the subject device establishes effective triage within a clinician's queue based on high sensitivity and specificity. Further, these results are substantially equivalent to those of the predicate device.

Triage effectiveness (turn-around time) was assessed as part of the standalone performance study. These cases were collected from multiple data sources spanning a variety of geographical locations, patient demographics and technical characteristics. The results demonstrated an average

{8}------------------------------------------------

triage turn-around time under 30.0 seconds, which is substantially equivalent to the total performance time published for the predicate device.

Therefore, the subject device has been shown to satisfy the performance requirements per 21CFR892.2080, for radiological triage and notification software, by providing clinically effective triage for chest X-ray studies containing features suggestive of clinical findings of interest. This data demonstrates the subject device is safe and effective for its intended use, and thereby supports substantial equivalence.

CONCLUSIONS IX.

The subject device and the predicate device are both software only packages, devices intended to assist with worklist triage by providing notification of findings in adults to trained clinicians. The subject and predicate devices utilize the same principles of operation and work in parallel to the current standard of care workflow.

Both the subject and predicate devices have the same software architecture, use the same deep learning AI principals to identify findings in images and require the same inputs and provide the equivalent outputs.

The minor differences between the subject and predicate devices do not raise new questions of safety and effectiveness.

Standalone performance testing and the comparison of technological characteristics with the predicate device shows that the subject device:

  • . performs as intended,
  • is safe and effective for its intended use, and ●
  • . is therefore substantially equivalent to the predicate device.

N/A