K Number
K250582

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-05-08

(70 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The COLONNADE Posterior Screw Fixation System is intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft and/or allograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation): spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. Fenestrated screws are not intended for use with cement.

Device Description

The COLONNADE Posterior Screw Fixation System pedicle screw is an internal fixation device for spinal surgery. Various forms and sizes of these implants are available so that constructions can be utilized to account for the unique pathology and anatomy of each patient. The subject pedicle screws and set screw of COLONNADE Posterior Screw Fixation System are manufactured from Ti-6Al-4V ELI as per ASTM F136, and are anodized The subject rods are manufactured from Ti-6Al-4V ELI as per ASTM F136 and Co-28Cr-6Mo as per ASTM F1537.

AI/ML Overview

This document, an FDA 510(k) clearance letter for the "COLONNADE Posterior Screw Fixation System," details the regulatory approval of a medical device intended for spinal surgery. However, it does not contain information about a study proving that an AI/software device meets acceptance criteria.

The 510(k) letter explicitly states, "Clinical Testing: Not Applicable" in the "Non-Clinical and/or Clinical Tests Summary & Conclusions" section on page 6. This indicates that no human clinical study was performed to assess the device's performance in a clinical setting against acceptance criteria for diagnostic or AI-driven tasks.

Instead, the provided text emphasizes the substantial equivalence of the "COLONNADE Posterior Screw Fixation System" to predicate devices based on:

  • Identical Indications for Use: The intended applications of the new device are the same as previously cleared devices.
  • Similar Technological Characteristics: The design, materials (Ti-6Al-4V ELI, Co-28Cr-6Mo), function, and fundamental technologies are comparable to predicates, with only "slight differences between the length and diameter of screws, and rod length."
  • Non-Clinical Testing: Performance was evaluated through mechanical testing standards (ASTM F1717) related to spinal implant constructs (Static Compression Bending, Static Torsion, Dynamic Compression Bending).

Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving an AI/software device meets them, as this information is not present in the provided FDA 510(k) clearance letter.

The letter pertains to a hardware device (a spinal fixation system), not a software or AI device for which performance metrics like those typically associated with diagnostic AI (e.g., sensitivity, specificity, AUC, human reader improvement) would be established and tested in a clinical trial.

FDA 510(k) Clearance Letter - COLONNADE Posterior Screw Fixation System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

May 8, 2025

CG Bio Co., Ltd.
Ms. Suwon Lee
29, Jeyakdanji-ro, Hyangnam-eup
Hwaseong-si, Gyeonggi-do 18608
Korea, South

Re: K250582
Trade/Device Name: COLONNADE Posterior Screw Fixation System
Regulation Number: 21 CFR 888.3070
Regulation Name: Thoracolumbosacral Pedicle Screw System
Regulatory Class: Class II
Product Code: NKB
Dated: February 25, 2025
Received: February 27, 2025

Dear Ms. Suwon Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K250582 - Ms. Suwon Lee Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250582 - Ms. Suwon Lee Page 3

Sincerely,

EILEEN CADEL -S

for

Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known)
K250582

Device Name
COLONNADE Posterior Screw Fixation System

Indications for Use (Describe)

The COLONNADE Posterior Screw Fixation System is intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft and/or allograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation): spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. Fenestrated screws are not intended for use with cement.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K250582 Page 1 of 2

510(k) Summary

510(k)#: K250582
Prepared on: 2025-05-07

Contact Details

21 CFR 807.92(a)(1)

FieldValue
Applicant NameCG Bio Co., Ltd.
Applicant Address29, Jeyakdanji-ro, Hyangnam-eup Hwaseong-si Gyeonggi-do 18608 Korea, South
Applicant Contact Telephone+821084661210
Applicant ContactMs. Suwon Lee
Applicant Contact Emailcgra@cgbio.co.kr

Device Name

21 CFR 807.92(a)(2)

FieldValue
Device Trade NameCOLONNADE Posterior Screw Fixation System
Common NameThoracolumbosacral pedicle screw system
Classification NameThoracolumbosacral Pedicle Screw System
Regulation Number888.3070
Product Code(s)NKB

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K222229Advanced LumFix Spinal Fixation SystemNKB
K160731LumFix Spinal Fixation SystemMNI

Device Description Summary

21 CFR 807.92(a)(4)

The COLONNADE Posterior Screw Fixation System pedicle screw is an internal fixation device for spinal surgery. Various forms and sizes of these implants are available so that constructions can be utilized to account for the unique pathology and anatomy of each patient. The subject pedicle screws and set screw of COLONNADE Posterior Screw Fixation System are manufactured from Ti-6Al-4V ELI as per ASTM F136, and are anodized The subject rods are manufactured from Ti-6Al-4V ELI as per ASTM F136 and Co-28Cr-6Mo as per ASTM F1537.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The COLONNADE Posterior Screw Fixation System is intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft and/or allograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation): spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. Fenestrated screws are not intended for use with cement.

Indications for Use Comparison

21 CFR 807.92(a)(5)

The indications for use of the subject device are identical to those of the predicate device(s).

Technological Comparison

21 CFR 807.92(a)(6)

Page 6

K250582 Page 2 of 2

The COLONNADE Posterior Screw Fixation System is substantially equivalent to the predicate devices cleared by the FDA for commercial distribution in the United States. There are slight differences between the length and diameter of screws, and rod length. But the subject device – COLONNADE Posterior Screw Fixation System – has similar technological characteristics including design, intended use, material composition, function and fundamental technologies as predicate devices.

Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

The following tests were performed in determining substantial equivalence of the COLONNADE Posterior Screw Fixation System:

ASTM F1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model

  • Static Compression Bending
  • Static Torsion
  • Dynamic Compression Bending

Clinical Testing: Not Applicable

The COLONNADE Posterior Screw Fixation System is substantially equivalent to the legally marketed predicate device identified above.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.