K Number
K160731

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-10-04

(201 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LumFix Spinal Fixation System, a posterior spinal fixation device, indicated for skeletally mature patients receiving fusion by autogenous bone graft with removal of the implants after the attainment of a solid fusion and is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

Device Description

The LumFix Spinal Fixation System is a top-loading multiple component, posterior spinal fixation system which consists of pedicle screws, set screws, rods and a crosslink linking mechanism.

The LumFix Spinal Fixation System will allow surgeons to build a spinal implant construct to stabilize and promote spinal fusion. LumFix Spinal Fixation System components are supplied non-sterile are single use and are fabricated from titanium alloy (Ti-6A1-4V ELI) that conforming to ISO 5832-3 or ASTM F136, with a cobalt-chromium-molybdenum rod option conforming to ASTM F1537 (Co-28Cr-6Mo). Various sizes of these implants are available.

AI/ML Overview

The provided text describes a medical device submission, specifically a 510(k) premarket notification for the "LumFix Spinal Fixation System." This document is focused on demonstrating substantial equivalence to previously cleared predicate devices, rather than establishing acceptance criteria or performance through a clinical study for a novel AI/software device.

Therefore, the information required to answer your request (acceptance criteria, study details, sample sizes, expert qualifications, etc.) is not present in the provided text.

The document discusses:

  • Device Name: LumFix Spinal Fixation System
  • Regulation Number: 21 CFR 888.3070
  • Regulation Name: Pedicle screw spinal system
  • Regulatory Class: Class II
  • Product Code: MNI, MNH
  • Indications for Use: Spinal fusion for various acute and chronic instabilities/deformities of the thoracic, lumbar, and sacral spine in skeletally mature patients.
  • Predicate Devices: Zenius™ Spinal System (K093104) and ANAX™ 5.5 Spinal System (K132101).
  • Non-clinical testing: Mechanical testing according to ASTM F1717 and ASTM F1798 to demonstrate equivalence. This includes static (tension, compression, torsion) and dynamic (fatigue) tests.

This is a submission for a physical medical device (spinal fixation system) and relies on mechanical testing to prove substantial equivalence to existing devices, not a clinical study to establish performance against acceptance criteria for an AI or software-based medical device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are silhouetted and appear to be looking to the right. The logo is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 4, 2016

CG Bio Co., Ltd. % Ms. April Lee Withus Group Inc. 2531 Pepperdale Drive Rowland Heights, California 91748

Re: K160731

Trade/Device Name: LumFix Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH Dated: July 5, 2016 Received: July 11, 2016

Dear Ms. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160731

Device Name LumFix Spinal Fixation System

Indications for Use (Describe)

The LumFix Spinal Fixation System, a posterior spinal fixation device, indicated for skeletally mature patients receiving fusion by autogenous bone graft with removal of the implants after the attainment of a solid fusion and is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary

1.ManufacturerCG Bio Co., Ltd.244, Galmachi-ro, Jungwon-gu,Seongnam-si, Gyeonggi-do,Korea+82-31-732-0840
2.Company ContactYuna Jeong (yuna@daewoong.co.kr)
3.Official CorrespondentApril LeeWithus Group Inc.Withus6664@gmail.com+1-909-274-9971
4.Proprietary Trade NameLumFix Spinal Fixation System
5.Common NamePedicle Screw Spinal Fixation System
6.Classification Name888.3070 – Pedicle Screw Spinal System
ClassificationClass II
Product CodesMNI, MNH
7.Date Prepared9/28/2016

General Description

The LumFix Spinal Fixation System is a top-loading multiple component, posterior spinal fixation system which consists of pedicle screws, set screws, rods and a crosslink linking mechanism.

The LumFix Spinal Fixation System will allow surgeons to build a spinal implant construct to stabilize and promote spinal fusion. LumFix Spinal Fixation System components are supplied non-sterile are single use and are fabricated from titanium alloy (Ti-6A1-4V ELI) that conforming to ISO 5832-3 or ASTM F136, with a cobalt-chromium-molybdenum rod option conforming to ASTM F1537 (Co-28Cr-6Mo). Various sizes of these implants are available.

{4}------------------------------------------------

Indications for Use

The LumFix Spinal Fixation System, a posterior spinal fixation device, indicated for skeletally mature patients receiving fusion by autogenous bone graft with removal of the implants after the attainment of a solid fusion and is intended to provide immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

Predicate Devices

The subject device is substantially equivalent to the following predicate devices:

  • Primary Predicate K093104 Zenius™ Spinal System, Medyssey Co., Ltd. .
  • . Additional Predicate - K132101 ANAX™ 5.5 Spinal System, U&I Corporation

Summary of the Technological Characteristics

The LumFix Spinal Fixation System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. There are slight differences between the length and diameter of screws and rod diameter. But the subject device -LumFix Spinal Fixation System -has similar technological characteristics including design, dimension, intended use, material composition, function and fundamental technologies as predicate devices.

Non-clinical testing

Mechanical testing that was conducted in accordance with ASTM F1717 and ASTM F1798 demonstrates equivalence to the above predicate devices.

Mechanical test reports were completed for the following test methods:

  • Static test: Tension, Compression and Torsion test report (ASTM F1717-11)
  • . Dynamic test: Fatigue test report (ASTM F1717-11)
  • . Static and fatigue properties (ASTM F1798-13)

Conclusion

Testing and other comparisons have established that the LumFix Spinal Fixation System is substantially equivalent in design, materials, indications, and performance to other predicate devices.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.