The Distalmotion Dexter L6 System is intended to assist in the accurate control of endoscopes as well as endoscopic instruments for endoscopic manipulation of tissue, including grasping, suturing, dissecting, coagulating and cutting, with or without high frequency functionality. The Distalmotion Dexter L6 System is intended for use in laparoscopic inguinal hernia repair and cholecystectomy. The system is indicated for adult use, defined as 22 years old and older. It is intended for use by trained laparoscopic or robotic surgeons in an operating room environment in accordance with the representative and specific procedures set forth in the Instructions for Use.
The Dexter L6 System is designed to enable complex surgery using a minimally invasive approach. It is composed of the Robot, the single-use accessories, fully articulated instruments, as well as reusable accessories. The Dexter L6 System is compatible with standard single-use laparoscopic trocars measuring 10 [mm] in diameter and larger.
The Robot consists of a Surgeon Console, with which the surgeon controls the movement of the instruments and of the Endoscope Arm using two Handle Grips (reusable), a Clutching Foot Pedal and an Endoscope Foot Pedal; two Patient Carts positioned at the operating room table in which the instruments are inserted and removed through the Hub during surgery; and the Dexter L6 Software installed in the Robot firmware.
The single-use, sterile instruments consist of the Needle Holder, Bipolar Johann Grasper, Bipolar Maryland Dissector, Monopolar Scissors, and Monopolar Hook.
The single-use accessories consist of the Sterile Interface, Endoscope Arm and Sterile Drapes.
The reusable accessories consist of the Accessory tray, Incision Pointer, Emergency Release Tool and Handle Grips.
This document is a 510(k) clearance letter for the Distalmotion Dexter L6 System, expanding its indications for use to include cholecystectomy procedures. The previous clearance (DEN230084) covered laparoscopic inguinal hernia repair.
Here's an analysis of the acceptance criteria and study details based on the provided text:
1. Acceptance Criteria and Reported Device Performance
The core acceptance criterion for the expanded indication of cholecystectomy is the successful completion of the Dexter-assisted procedure without conversion to an open or fully laparoscopic surgical approach. The safety is assessed by reporting adverse events.
| Acceptance Criteria Category | Specific Metric | Acceptance Criterion / Goal | Reported Device Performance (NEST Study) |
|---|---|---|---|
| Effectiveness (Primary Endpoint) | Successful completion of Dexter-assisted surgery without conversion to open or fully laparoscopic approach | High success rate | 98.0% (50/51 subjects, 95% CI: 89.6%, 100.0%) |
| Safety (Primary Endpoint) | Postoperative adverse events (Clavien-Dindo grades III-V) perioperatively up to 30 days | Low incidence of severe adverse events | 1 postoperative (Clavien-Dindo grades III-V) adverse event |
| Safety | Serious Adverse Events (SAEs) | Low incidence | 2 serious adverse events (4.0%) |
| Safety | Intraoperative Adverse Events | Low incidence | 2 intraoperative adverse events (4.0%) |
| Safety | Reoperation rate | Low incidence | None of the subjects required reoperation |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 51 subjects (mITT population)
- Data Provenance: Prospective, multicenter, open-label clinical investigation. The document mentions "patient privacy regulations in France" when discussing ethnicity, implying that at least some of the data originates from Europe (specifically France). It also compares the NEST cohort with "EU Literature" and "US Literature" for demographic and comorbidity comparison, suggesting the study itself might have collected data from multiple regions, with specific mention of France.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not explicitly state the number of experts used to establish ground truth or their qualifications. The "ground truth" in this context is the successful completion of the surgical procedure and the occurrence and grading of adverse events. These are typically determined by the operating surgeons and medical staff involved in the patient's care and follow-up, which would then be compiled and reviewed by the study investigators.
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method for the test set. For clinical trials like the NEST study, adverse events and surgical outcomes are usually documented by the clinical site and then centrally reviewed by a clinical events committee (CEC) or similar body. However, the specific method (e.g., 2+1, 3+1 consensus) is not detailed here.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The NEST study focused on the performance of the Dexter L6 System in cholecystectomy, where the "reader" would be the surgeon using the system. It was a study on the device's performance, not a comparative study against human readers (surgeons) without the device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
The Dexter L6 System is an "electromechanical surgical system" intended to "assist in the accurate control of endoscopes as well as endoscopic instruments." It is a robotic-assisted surgical system, meaning it inherently involves a human surgeon (the "operator" or "human-in-the-loop") controlling the robot. Therefore, a standalone (algorithm only) performance study would not be applicable or relevant for this type of device. The study evaluates the system's performance with the surgeon.
7. The Type of Ground Truth Used
The ground truth used for the effectiveness endpoint was surgical outcome data (successful completion of the surgery without conversion). For safety, the ground truth was clinical outcomes data (adverse events, their severity (Clavien-Dindo grades), and resolution status), which rely on clinical assessment and diagnosis.
8. The Sample Size for the Training Set
The document does not provide information on a specific training set sample size. This is a clinical study for a hardware device with an expanded indication, not typically a machine learning algorithm that requires a "training set" in the conventional sense for its "performance." The device itself (Dexter L6 System) was previously cleared, and "Extensive bench testing was conducted on the previously cleared Dexter L6 System." This bench testing likely served as a form of "training" or development/verification data, but it's not a "training set" for a diagnostic algorithm.
9. How the Ground Truth for the Training Set Was Established
As noted above, a "training set" in the context of an algorithm with associated ground truth is not explicitly mentioned or applicable in the way it would be for AI/ML software. The "ground truth" for the device's initial development and bench testing would have been established through engineering specifications, design verification, and validation activities, ensuring the device performs according to its intended design.
U.S. Food & Drug Administration 510(k) Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
May 27, 2025
Distalmotion SA
℅ Lina Kontos
Regulatory Counsel
Hogan Lovells US LLP
Columbia Square
555 Thirteenth Street, NW
Washington, District of Columbia 20004
Re: K250435
Trade/Device Name: Dexter L6 System
Regulation Number: 21 CFR 878.4965
Regulation Name: Electromechanical surgical system with transient sterile field presence of both surgeon and primary control interface
Regulatory Class: Class II
Product Code: SDD
Dated: February 14, 2025
Received: April 17, 2025
Dear Lina Kontos:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
Page 2
K250435 - Lina Kontos Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
Page 3
K250435 - Lina Kontos Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Trumbore Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
Indications for Use
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250435
Device Name
Dexter L6 System
Indications for Use (Describe)
The Distalmotion Dexter L6 System is intended to assist in the accurate control of endoscopes as well as endoscopic instruments for endoscopic manipulation of tissue, including grasping, suturing, dissecting, coagulating and cutting, with or without high frequency functionality. The Distalmotion Dexter L6 System is intended for use in laparoscopic inguinal hernia repair and cholecystectomy. The system is indicated for adult use, defined as 22 years old and older. It is intended for use by trained laparoscopic or robotic surgeons in an operating room environment in accordance with the representative and specific procedures set forth in the Instructions for Use.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
510(K) SUMMARY
DISTALMOTION DEXTER L6 SYSTEM
Page 1 of 4
K250435
Submitter: Distalmotion SA
Route de la Corniche 3b
1066 Epalinges
Switzerland
Contact: Larry Carrier
VP Regulatory Affairs & Quality Assurance
Phone: +1-707-239-6086
Email: larry.carrier@distalmotion.com
Date Prepared: May 23, 2025
Trade/Device Name: Dexter L6 System
Regulation Number: 21 CFR 878.4965
Regulation Name: Electromechanical surgical system with transient sterile field presence of both surgeon and primary control interface
Regulatory Class: Class II
Product Code: SDD
Review Panel: General & Plastic Surgery
Predicate Devices: Dexter L6 System (DEN230084)
Device Description:
The Dexter L6 System is designed to enable complex surgery using a minimally invasive approach. It is composed of the Robot, the single-use accessories, fully articulated instruments, as well as reusable accessories. The Dexter L6 System is compatible with standard single-use laparoscopic trocars measuring 10 [mm] in diameter and larger.
The Robot consists of a Surgeon Console, with which the surgeon controls the movement of the instruments and of the Endoscope Arm using two Handle Grips (reusable), a Clutching Foot Pedal and an Endoscope Foot Pedal; two Patient Carts positioned at the operating room table in which the instruments are inserted and removed through the Hub during surgery; and the Dexter L6 Software installed in the Robot firmware.
Page 6
Page 2 of 4
K250435
The single-use, sterile instruments consist of the Needle Holder, Bipolar Johann Grasper, Bipolar Maryland Dissector, Monopolar Scissors, and Monopolar Hook.
The single-use accessories consist of the Sterile Interface, Endoscope Arm and Sterile Drapes.
The reusable accessories consist of the Accessory tray, Incision Pointer, Emergency Release Tool and Handle Grips.
Intended Use / Indications for Use:
The Distalmotion Dexter L6 System is intended to assist in the accurate control of endoscopes as well as endoscopic instruments for endoscopic manipulation of tissue, including grasping, suturing, dissecting, coagulating and cutting, with or without high frequency functionality. The Distalmotion Dexter L6 System is intended for use in laparoscopic inguinal hernia repair and cholecystectomy. The system is indicated for adult use, defined as 22 years old and older. It is intended for use by trained laparoscopic or robotic surgeons in an operating room environment in accordance with the representative and specific procedures set forth in the Instructions for Use.
The Dexter L6 System is for prescription use only.
The indications for use statement are consistent with the previously cleared Dexter L6 system, except for the additional of the term "suturing" to the statement and the addition of use of the system in performing cholecystectomy procedures.
Summary of Technological Characteristics:
The subject device has the same technological characteristics as the predicate device, the Dexter L6 System (DEN230084). The labeling has been changed to address use of the device in cholecystectomy procedures, to clarify the contraindications and that the device can be used for suturing.
Performance Data:
There have been no changes to the device since the previous clearance (DEN230084), other than the modification to expand the indications for use to include cholecystectomy indication. Extensive bench testing was conducted on the previously cleared Dexter L6 System, and these data remain applicable to support the safety and effectiveness of the subject device.
Page 7
Page 3 of 4
K250435
Clinical Data:
The NEST (DEXTER-assisted Non-Emergent and acute cholecystectomy) Study was a prospective, multicenter, open-label, clinical investigation to confirm the perioperative and early postoperative safety and effectiveness of the Dexter Robotic System (Dexter) in patients undergoing cholecystectomy.
The NEST Study included 51 subjects who were enrolled and on whom robotic-assisted surgery with Dexter was at least started (mITT population). Subject data was collected at baseline, intraoperatively, discharge, and postoperatively at 30 days.
The mean age of the mITT population was 55.1 years. Female subjects accounted for 62.7% (32/51). Most subjects were classified as ASA II status 84.3% (43/51). A demographic and comorbidity comparison of the NEST cohort and the EU and US literature is included in the table below.
| Demographic/Comorbidity | NEST (n=51) | EU Literature | US Literature | Discussion |
|---|---|---|---|---|
| Age (years) | 55.1 ± 14.7 | 42.1-80 | 44.3-61.1 | NEST cohort in range |
| Gender | 64.2% (Female) | 71-76.6 (Female) | 55-77% Female | NEST cohort in range |
| Ethnicity | 58.8% (White)2% (African American)3.9% (Other)35.3% (Unknown or Not Reported) | Not reported | 53.7-89.2 (White)4-14.7% (African American)0.4-21.2 (Hispanic) | NEST cohort in range for most common White ethnicityNOTE: Due to patient privacy regulations in France, the reported ethnicity of 35.3% (Unknown/ Not Reported) also includes additional subjects with White ethnicity. |
| BMI (kg/m²) | 27.7 | 23.4-29 | 28.8-35.6 | NEST cohort similar to US population. See Obesity comorbidity below. |
| ASA Status | 92% (ASA I/II)7.8% (ASA III+) | 63 - 97% (ASA I/II)3-37% (ASA III+) | 54.4 - 80.2% (ASA I/II)19.7-54.6% (ASA III+) | NEST cohort higher proportion of ASA I/IIEU population and US population in similar range |
| Nassar Difficulty Scale Rating | 21.6% (I)45.1% (II)27.5% (III)5.9% (IV) | Not reported | Not reported | Nassar Scale only reported for NEST cohort |
| Obesity (> 30 kg/m²) | 21.6% (n=11/51) | Not reported | 5.7-54.2% | NEST cohort in range of US population |
| Hypertension | 25.5% (13/51) | Not reported | 27-59.7% | NEST cohort similar to US population |
| COPD | Not reported | 8.4% | 2.05-8.8% | EU population in range of US population |
| Cardiac Disease | 23.5% (N=12/51)* | 6.4% | 0.2-17.3 | EU population in range of US population |
| Diabetes | 13.7 (7/51) | 4.3-20.1% | 9.2-24.4% | NEST cohort in range of US population |
| Hypothyroidism | 5.8% (3.51) | Not reported | Not reported | Hypothyroidism only reported for NEST cohort |
| Smoking | Not reported | 9-35% | 13.1-28.9% | EU population in range of US population |
The operation was successfully performed in 50 out of 51 subjects of the mITT population using robotic-assisted surgery (98.0%, 50/51), (95% CI:89.6%, 100.0%).
The primary performance endpoint, defined as successful completion of the Dexter-assisted procedure without conversion to an open or fully laparoscopic surgical approach, was confirmed in 50 of 51 procedures from the mITT group (98.0%). One subject was converted to laparoscopic mode due to time constraints in which the surgeon decided to complete the procedure laparoscopically to shorten the operative time.
For the primary safety endpoint, there was one postoperative (Clavien-Dindo grades III-V) adverse
Page 8
Page 4 of 4
K250435
event perioperatively up to 30 days. There were 10 adverse events and two serious adverse events in the perioperative phase to 30 days after the procedure.
Two serious adverse events were reported (4.0%):
- Post-operative agranulocytosis and pancytopenia as an adverse reaction to metamizole therapy. The event was resolved at the time of study completion and classified as Clavien-Dindo II, not related to the procedure and not related to the device.
- Re-hospitalization due to the migration of a pre-existing choledocholithiasis. The event was resolved at the time of study completion and classified as Clavien-Dindo III, not related to the procedure and not related to the device.
Two intraoperative adverse events were reported (4.0%):
- Gallbladder perforation; the event was treated with prophylactic antibiotics. The event was resolved at the time of study completion and was classified as Clavien-Dindo II, procedure related and not related to the device.
- Injury of the posterior wall of the gallbladder; the event was treated with prophylactic antibiotics. The event was resolved at the time of study completion and was classified as Clavien-Dindo II, procedure related and not related to the device.
None of the subjects required reoperation after the index procedure. In conclusion, the results of the NEST Study demonstrate the Dexter L6 System is safe and effective for subjects undergoing cholecystectomy.
Conclusions:
The subject Dexter L6 System is as safe and effective as the predicate device. The subject Dexter L6 System has the same intended uses and similar indications, same technological characteristics, and principles of operation as its predicate device. The expanded specific indications do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. Analysis and clinical performance data of the device with cholecystectomy procedures demonstrate that the Dexter L6 System is as safe and effective as the predicate device. Thus, the subject Dexter L6 System is substantially equivalent.
N/A