DEN230084 · Distalmotion SA · SDD · Oct 25, 2024 · General, Plastic Surgery
Device Facts
Record ID
DEN230084
Device Name
Dexter L6 System
Applicant
Distalmotion SA
Product Code
SDD · General, Plastic Surgery
Decision Date
Oct 25, 2024
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 878.4965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Distalmotion Dexter L6 System is intended to assist in the accurate control of endoscopes as well as endoscopic instruments for endoscopic manipulation of tissue, including grasping, dissecting, coagulating and cutting, with or without high frequency functionality. The Distalmotion Dexter L6 System is intended for use in laparoscopic inguinal hernia repair. The system is indicated for adult uses, as defined as 22 years old and older. It is intended for use by trained laparoscopic surgeons in an operating room environment in accordance with the representative and specific procedures set forth in the Instructions for Use. The Dexter L6 System is for prescription use only.
Device Story
Dexter L6 is an electromechanical surgical system for laparoscopic inguinal hernia repair. System comprises Surgeon Console (SC) and Patient Carts (PC). Surgeon uses handle grips and pedals at SC to control instrument arms on PC; SC allows surgeon to scrub into sterile field. Master arms at SC capture surgeon's hand/arm movements; actuators at PC hub translate these into micro-movements of end effectors (scissors, hooks, graspers, needle holders). System includes endoscope arm for third-party endoscope control. Used in OR by trained surgeons. Provides precise instrument control; facilitates minimally invasive surgery; reduces surgeon fatigue/improves ergonomics. Output is mechanical manipulation of tissue. Safety features include emergency release tool, protective stop, and visual feedback via LED ring on hub.
Clinical Evidence
Prospective, multicenter, single-arm study (RAS AHEAD) of 50 adult subjects undergoing primary unilateral or bilateral inguinal hernia repair. Primary endpoint: procedural success (no conversion to open/laparoscopic) achieved in 98% (49/50). Mean operative time: 54.5 min (unilateral), 94.6 min (bilateral). No device-related serious adverse events (Clavien-Dindo III-V) reported within 30 days. 12 total AEs (18% subjects), mostly Clavien-Dindo I (16%) and II (2%). No intraoperative complications. Results comparable to literature for existing robotic systems.
Technological Characteristics
Software-controlled electromechanical system; fully positionable patient/device interfaces. Includes reusable components requiring reprocessing. Subject to electrical safety, electromagnetic compatibility, and biocompatibility standards. Requires sterilization validation for patient-contacting components.
Indications for Use
Indicated for adult patients (≥22 years) undergoing laparoscopic inguinal hernia repair. Contraindications include morbid obesity (BMI >40), bleeding diathesis, pregnancy, presence of pacemakers/ICDs, need for robotic stapling/advanced energy/ultrasound/cryoablation/microwave, and specific anatomical/surgical history (e.g., large scrotal hernia, prior major abdominal/pelvic surgery, prior preperitoneal mesh).
Regulatory Classification
Identification
An electromechanical surgical system with transient sterile field presence of both surgeon and primary control interface is a software-controlled electromechanical system with fully positionable patient/device interfaces which allows a qualified user to perform surgical techniques during minimally invasive surgical procedures. The device allows for both the surgeon and primary control interface to have presence inside the sterile field as needed within the clinical context of use.
Related Devices
K251197 — Dexter L6 System · Distalmotion SA · Sep 30, 2025
K250435 — Dexter L6 System · Distalmotion SA · May 27, 2025
K171120 — Senhance Surgical Robotic System · Transenterix, Inc. · Oct 13, 2017
K170713 — da Vinci Xi Surgical System · Intuitive Surgical, Inc. · Jun 13, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
#### DE NOVO CLASSIFICATION REQUEST FOR
#### DEXTER L6 SURGICAL SYSTEM
#### Regulatory Information
FDA identifies this generic type of device as:
Electromechanical surgical system with transient sterile field presence of both surgeon and primary control interface. An electromechanical surgical system with transient sterile field presence of both surgeon and primary control interface is a software-controlled electromechanical system with fully positionable patient/device interfaces which allows a qualified user to perform surgical techniques during minimally invasive surgical procedures. The device allows for both the surgeon and primary control interface to have presence inside the sterile field as needed within the clinical context of use.
NEW REGULATION NUMBER: 21 CFR 878.4965
CLASSIFICATION: Class II
PRODUCT CODE: SDD
#### BACKGROUND
DEVICE NAME: Dexter L6 Surgical System
SUBMISSION NUMBER: DEN230084
DATE DE NOVO RECEIVED: December 15, 2023
#### SPONSOR INFORMATION:
Distalmotion, SA % Lina Kontos Hogan Lovells US LLP 555 13th Street, NW Washington, DC 20004
#### INDICATIONS FOR USE
The Distalmotion Dexter L6 System is intended to assist in the accurate control of endoscopes as well as endoscopic instruments for endoscopic manipulation of tissue, including grasping, dissecting, coagulating and cutting, with or without high frequency functionality. The Distalmotion Dexter L6 System is intended for use in laparoscopic inguinal hernia repair. The system is indicated for adult uses, as defined as 22 years old
{1}------------------------------------------------
and older. It is intended for use by trained laparoscopic surgeons in an operating room environment in accordance with the representative and specific procedures set forth in the Instructions for Use.
The Dexter L6 System is for prescription use only.
# LIMITATIONS
The sale, distribution, and use of the Dexter L6 Surgical System are restricted to prescription use in accordance with 21 CFR 878.4964.
The device is only for distribution to facilities that implement and maintain the device specific use training program and ensure that users of the device have completed the device specific use training program. Only trained users and those who have developed adequate robotic skills to perform the tasks associated with each procedure should use the Dexter L6 System. Training provided by Distalmotion is limited to the use of the Dexter L6 System and does not replace the necessary medical training and experience required to perform laparoscopic surgery.
The demonstration of safety and effectiveness for the representative-specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition. Device usage in all surgical procedures should be guided by the clinical judgment of an adequately trained surgeon.
# PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
# DEVICE DESCRIPTION
The Dexter L6 Surgical System is intended to assist in the accurate control of its surgical endoscopic instruments across a range of surgical procedures. The system is comprised of two major elements, the Surgeon Console and the Patient Cart. The surgeon console allows the surgeon to scrub into the surgical field, and either in seated or standing position, allows control of the instrument and endoscope arms on the patient cart. The Surgeon Console provides the surgeon two handle grips and two pedals, one pedal to clutch instrument control and a second to engage endoscope control. The patient cart is a modular element of the system which is comprised of a pedestal and boom which support the instrument arm. The instrument arm on the patient cart is controlled by the Master Arms on the surgeon cart. The Master Arms are passive elements on the surgeon cart which translate the movements of the Instrument arms. A system overview is provided in Figure 1, including surgeon console, patient carts, and endoscope cart. Figure 2 shows end effectors cleared for use on the system. In addition to the major elements of the system certain reusable and disposable sterile accessories are cleared with the Dexter L6 system, shown in figures 3 and 4.
{2}------------------------------------------------
| Surgeon Console | Patient Cart 1 | Patient Cart 2 | Endoscope Cart |
|------------------------|-----------------------|-----------------------|-----------------------|
| Image: Surgeon Console | Image: Patient Cart 1 | Image: Patient Cart 2 | Image: Endoscope Cart |
Figure 1: System Overview
| Needle Holder | Bipolar Johann<br>Grasper | Bipolar<br>Maryland<br>Dissector | Monopolar<br>Scissors | Monopolar Hook |
|----------------------|-------------------------------|-----------------------------------|-----------------------------|-----------------------------|
| Image: Needle Holder | Image: Bipolar Johann Grasper | Image: Bipolar Maryland Dissector | Image: Monopolar Scissors | Image: Monopolar Hook |
| | Rated voltage of<br>770 Vp | Rated voltage of<br>770 Vp | Rated voltage of<br>4000 Vp | Rated voltage of<br>4000 Vp |
Figure 2: Instrument End Effectors and Peak Voltage rating
{3}------------------------------------------------
| Accessory Tray | Handle Grip | Incision Pointer | Emergency Release<br>Tool |
|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|----------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Image: Accessory Tray | Image: Handle Grip | Image: Incision Pointer | Image: Emergency Release Tool |
| 1x | 2x | 2x | 1x |
| To transport, store and<br>reprocess the Handle<br>Grips, the Incision<br>Pointers and the<br>Emergency Release Tool. | Connected to the<br>Surgeon Console. Held<br>by the surgeon to control<br>the Dexter L6 System. | To dock the Patient<br>Carts to the trocars. | To open instrument<br>blades when the Dexter<br>L6 System stops<br>working and instruments<br>grab tissue. |
Figure 3: Reusable Accessories
| Sterile Drapes | Sterile Interfaces |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Image: Sterile Drapes | Image: Sterile Interfaces |
| 3x Sterile Drapes for the Surgeon Console: Left Master Arm, right Master Arm and central pillar. 2x Sterile Drapes for the Patient Cart: Left Patient Cart and right Patient Cart. | 2x Inserted into the Hub. |
Figure 4: Disposable Accessories
## The Dexter L6 System and accessories:
- Dexter L6 (DM-L6): .
- · Surgeon Console (SC), which is the primary user interface for the surgeon for controlling the instrument arms. The surgeon console comprises:
- o Master Arms (SC-MA), which capture physician's arm movements.
## Page 4
De Novo Summary (DEN230084/S001)
{4}------------------------------------------------
- o Handles (SC-H), which connect to the Handle Grips and capture surgeon's hand movement.
- Wheelbase (SC-WB), which allows transport, OR positioning and console securing. o
- Patient Carts (PC), which is placed bedside for holding the laparoscopic surgical instrument. . Two patient carts are provided with the system, where each comprises:
- Hub (PC-H), which interfaces with the instrument and contains the actuators that drives O the instrument movements (micro-movements).
- o Instrument Arm (PC-IA), which supports the hub and replicates surgeon's arms movements (macro-movements).
- Boom (PC-B), which holds the instrument arms and allow its horizontal positioning. o
- · Wheelbase (PC-WB), which allows transport, bedside positioning and cart securing.
- . Software which contains safety architecture and motion control algorithms.
- Instruments, available in the following 5 models: .
- Monopolar Scissors (IN-MS), which allows performing cutting and cauterization steps. 0
- Monopolar Hook (IN-MH), which allows performing cutting and cauterization steps. 0
- Bipolar Maryland Grasper (IN-BM), which allows performing grasping and cauterization o steps.
- Bipolar Johann Grasper (IN-JG), which allows performing atraumatic grasping and o cauterization steps.
- O Needle Holder (IN-NH), which allows performing suturing steps.
- . Endoscope Arm (SOLO, third party), which holds and controls the movement of another third-party endoscope. It comes with accessories - which are used to secure the third-party endoscope to the Endoscope Arm - and drapes.
- . Endoscope Cart (EC), which supports the Endoscope Arm.
- Accessories .
- Sterile Interface (AC-SU-SI), which provides a sterile and electrical barrier between the o instrument and the hub.
- Handle Grips (AC-RU-HG), which connect to the surgeon console handles and capture O surgeon's hand movement.
- Incision Pointers (AC-RU-IP), which are used to dock the instrument arms. o
- Emergency Release Tool (ERT), which are used to release the instrument in case the o system cannot be operated.
- The Accessory Tray (AC-RU-T), which serves as a container for all accessories to be O reprocessed.
- Sterile Drape (AC-SU-SD), which serves as a sterile barrier for the surgeon console, the o patient carts
# SUMMARY OF NONCLINICAL/BENCH STUDIES
The following non-clinical test were leveraged to demonstrate safety and effectiveness for the subject device's indication for use.
The Dexter L6 Surgical System has components that are in direct patient contact. The patient contacting portion and materials were assessed in accordance with ISO 10993-1:2018 and FDA Guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. " Table 1 summarized biocompatibility testing completed for the Dexter L6.
Page 5 De Novo Summary (DEN230084/S001)
{5}------------------------------------------------
| Test | Method | Results |
|-----------------------------------|--------------------------------------------------|---------|
| Cytotoxicity | MEM Elution - ISO 10993-5 | PASS |
| Sensitization | ISO 10993-10 - Sensitization Assays | PASS |
| Irritation | ISO 10993-21 -<br>Intracutaneous Injections Test | PASS |
| Acute Systemic<br>Toxicity | ISO 10993-11 - Systemic Injection Test | PASS |
| Material-mediated<br>Pyrogenicity | ISO 10993-11 - Rabbit Pyrogen Test | PASS |
## TABLE 1. BIOCOMPATIBILITY/MATERIALS
Cleaning, disinfection, and sterilization validation testing was undertaken for all relevant elements of the Dexter L6 Surgical System (Table 2).
# Device Integrity/Use-life Testing:
Each article was processed through a total of 30 cycles. Each cycle consisted of a automated cleaning procedure followed by a sterilization cycle. Following the completion of the cleaning of each cycle, the articles were visually inspected for any sign of physical damage or corrosion, including cracking, flaking, pitting, and formation of rust. The reprocessing cycles were performed with the worst sterilization parameters along with simulated use (mechanical fatigue/wear testing). For the hand grips evaluation of test sample by minitab 19 for 2 samples for an equivalence of 30 uses with no failure and a Weibull shape factor of 3 is the following: 291,399 movements for 60000 actuations and 2916 movements for 600 releases. Knowing that 1 cycle is 2 movements (opening + closing).
| Test | Method | Results |
|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dexter L6, Customer Card | | |
| Sterilization | ISO 11135:2014 | PASS |
| Dexter L6, Sterilization Validation Protocol | | |
| Sterilization | ISO 11135:2014, ANSI / AAMI/ISO 11737-1, ANSI / AAMI / EN ISO 11737-2, ISO 11138, EN 556-1, ISO 10993-7, EN ISO 13485, EN ISO 11607, EN 1422 | PASS |
| Dexter L6, Surgical Instruments | | |
| Sterilization | SN ISO 11737-1 - European Pharmacopoeia ch. 2.6.12 and USP (61) | PASS |
| Dexter L6, Reusable Accessories (Handle Grips, Incision Pointer, and Tools) | | |
| Cleaning | ANSI/AAMI ST98:2022; ISO 10993-5:2009, AAMI TIR12:2020, AAMI TIR34:2014/(R)2021, ISO 17664-1:2021, ASTM F3208-20, AAMI ST 15883-1:2009/(R)2023 | PASS |
| Disinfection | ANSI/AAMI ST98:2022; ISO 10993-5:2009, AAMI TIR12:2020, AAMI TIR34:2014/(R)2021, ISO 17664- | PASS |
TABLE 2 SHELF LIFE/STERILITY
{6}------------------------------------------------
| Sterilization | 1:2021, ASTM F3208-20, AAMI ST 15883-<br>1:2009/(R)2023<br>ANSI/AAMI/ISO 17665-1:2006/(R)2013;<br>ANSI/AAMI/ISO 11138-1:2017, ANSI/AAMI/ISO<br>11138-3:2017, ISO 17664-1:2021, ISO 11138-7:2019,<br>ISO 11737-2:2019, ANSI/AAMI ST79:2017 and 2020,<br>EN285:2015 + A1:2021 | PASS |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
The system was evaluated to ensure basic functionality after shipping and distribution (Table 3).
| Test | Method | Results |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------|
| Environmental &<br>Distribution<br>Testing | ISTA 3A-18, ASTM D4332-22, ASTM D5276-19, ASTM D4728-17 (2022),<br>ASTM D6653/D6653M-13 (2021), D4728-17 (2022) | PASS |
TABLE 3. ENVIRONMENTAL AND DISTRIBUTION TESTING
The EMC and Electrical Safety was evaluated to mitigate the risk of electrical fault resulting in injury to patient or user. Table 4 identifies the Electrical/ Mechanical/Thermal Safety, and electromagnetic compatibility (EMC) testing that has been performed. Production batch instruments were sterilized prior to 60601-1 and collateral (including -2-2) testing.
TABLE 4. ELECTROMAGNETIC CAPABILITY & ELECTROMAGNETIC SAFETY
| Test | Method | Results |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Electrical Safety,<br>& Usability | IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC<br>60601-<br>1:2005/AMD2:2020<br>IEC 60601-1-6:2010, AMD1:2013, AMD2:2020 for use<br>in conjunction with<br>IEC 62366-1:2015, AMD1:2020, and IEC 60601-1:2005,<br>AMD1:2012,<br>AMD2:2020, (Usability/Human Factors)<br>IEC 60601-2-18:2009 for use in conjunction with IEC<br>60601-1:2005, (endoscopic equipment)<br>IEC 80601-2-77:2019 for use IEC 60601-1:2005,<br>COR1:2006,<br>COR2:2007, AMD1:2012, (RASE)<br>IEC 60601-2-2:2017 for use in conjunction with IEC<br>60601-1:2005, (HF surgical equipment) | PASS |
| EMC (Immunity) | IEC 60601-1-2:2020(Electromagnetic disturbances) | PASS |
| EMC (Emissions) | IEC 60601-1-2:2015 | PASS |
# MAGNETIC RESONANCE (MR) COMPATIBILITY
{7}------------------------------------------------
This device is not MR compatible and should not be used in or near MR equipment.
# SOFTWARE
A failure or latent flaw in the software of the Dexter L6 Surgical System could directly result in severe injury or death to the patient; therefore, the software of this device is considered to require "Enhanced Documentation".
The submission contained all the elements of "Enhanced Documentation" corresponding to a "Major" level of concern, as outlined in the FDA guidance document " Content of Premarket Submissions for Device Software Functions ", issued June 14, 2023 (https://www.fda.gov/media/73065/download). Adequate documentation describing the software. firmware, software specifications, architecture design, software development environment, traceability, revision level history, unresolved anomalies and cybersecurity provide the foundation that the software will operate in a manner as described in the specifications. A hazard analysis was performed to characterize software risks including device malfunction and measurement related errors. The submission included verification and validation (V&V) testing to address the potential hazards with satisfactory results.
# PERFORMANCE TESTING - BENCH
The following bench tests were performed to mitigate the risks of thermal, electrical, and mechanical fault which may result in injury to patient or user, tissue damage and/or injury due to svstem malfunction. user error which mav result in patient injurv. Bench testing also addressed unique risks associated with the design of the Dexter L6 surgical system.
The bench tests characterize device performance and design verification for the Dexter L6 Surgical System. All applicable testing was performed with provided and third-party devices. The descriptions and results of the bench tests are summarized in Table 5.
| Test Description | Objective | Results |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| System Testing | The intent of this section is to demonstrate that the device<br>satisfies the requirements ensuring the motion replication<br>performance, Endoscope Control essential performance, Safe<br>State Transition and Emergency Stop essential performances,<br>Protective Stop defined as "Once the RASS enters the<br>PROTECTIVE STOP, the instruments motion replication is<br>stopped, the brakes are not engaged, and the RASS continues to<br>monitor the error triggering condition, Release of the Patient,<br>Visual feedback is provided by the LED ring on the HUB | PASS |
| Accuracy and Precision | The activities in this section aim to verify that the Dexter L6<br>System meets the product requirement specifications related to<br>motion control performance. Specifically Accuracy and | PASS |
| | Precision of the position control of the instrument, as well as the<br>Minimum Controllable Motion, Minimum Measurable Motion,<br>and latency of the motion.<br>Furthermore, the integration with the 3rd Party endoscope arm is<br>verified by measuring the speed Accuracy.<br>The testing shall provide evidence of the Dexter L6 System's<br>ability replicate the surgeon's movements into corresponding<br>motion, controlling instrument's pose (position + orientation).<br>This shall be assessed systematically by verifying the<br>performance of each related subsystem ( master arm, instrument<br>arm, endoscope arm ) in terms of accuracy and precision, as well<br>as the performance of the Dexter L6 System as a whole. | |
| Workspace access | All non-clinical tests executed to evaluate the capability of the<br>robotic arm to move the instrument tip at any location of the<br>workspace maintaining the RCM, independently of the surgical<br>procedure.<br>The RCM is defined as the Remote Center of Motion is a fixed<br>point in space that is always aligned with the instrument shaft.<br>The RCM is a feature which ensures that the surgical<br>instruments rotate around their incision ports (trocar), thereby<br>minimizing the risk of harm to surrounding tissues. | PASS |
| System set-up | Verify the compatibility and combinations performance and<br>safety of the mounting hardware with the robot by testing<br>interfaces or reviewing design specifications. | PASS |
| Performance testing<br>with compatible<br>devices | Additional testing that is not fully covered by the above<br>mentioned assessment:<br>compatibility with Surgical Table and electrical connectors of<br>the Instruments to the commercially available Electrosurgical<br>Generator cords. | PASS |
| Sub-system testing,<br>encompassing for<br>the Surgeon<br>Console: Mechanicalreliability<br>Balance and Motion<br>Smoothness,<br>Functional Tests.<br>Mechanical Rigidity<br>and Resistance,<br>Mechanical Balance<br>and Motion | Non-clinical bench testing is needed on the Surgeon Console<br>subsystem to focus on mechanical aspects and data acquisition<br>process performance, as well as mechanical and electrical | PASS |
| Smoothness, Motion<br>Capture and Sensor<br>Resolution,<br>Workspace<br>Mechanical and<br>Electrical Reliability | | |
| Sub-system testing,<br>encompassing for the<br>Patient Carts:<br>Functional Tests,<br>Mechanical Rigidity<br>and<br>Resistance, Motion<br>Smoothness,<br>Workspace,<br>Mechanical and<br>Electrical Reliability | Non-clinical bench testing is needed on the Patient Cart subsystem<br>to focus on actuation and mechanical aspects, performance of<br>replicating the input motion data, as well as mechanical and<br>electrical reliability | PASS |
| Instruments,<br>encompassing for all<br>the instrument types:<br>Functional Tests,<br>Mechanical Rigidity<br>and<br>Resistance,<br>Mechanical and<br>Electrical Reliability,<br>Micro Workspace | Functional Tests section comprises all testing completed to ensure<br>stable and constant functionality of the Dexter's Instruments. The<br>Mechanical rigidity and resistance section challenges the strength of<br>the Instruments while the Mechanical and electrical reliability<br>challenges the endurance and safety. Finally, the Micro workspace<br>section gives an overview of the mechanical tip workspace<br>limitation.<br>The capture of the movements, the closing forces, the replication<br>and the accuracy of the Dexter system including the Instruments are<br>discussed in previous sections 4.3 - Tests Executed and 5 -<br>Workspace access, as part of system testing and device motion<br>accuracy testing respectively. | PASS |
| Accessories,<br>encompassing specific<br>tests for Sterile<br>Interface, Endoscope<br> | Tests to demonstrate functionality, safety and reliability of the<br>following accessories.<br>The accessories are used in the context of the surgical procedure in<br>order to make up for specific system requirements.<br>The Dexter L6 system accessories are:<br>* Sterile interface (AC-SU-SI)<br>* Endoscope cart (DM-L6-EC)<br>* Sterile drapes (AC-SU-SD)<br>Reusable accessories are assessed in specific reprocessing and<br>durability reports, in the Shelf Life and Packaging Validation<br>sections. This concerns:<br>* Accessory Tray<br>* Handle Grip<br>* Incision pointer<br>* Emergency release tool | PASS |
| Thermal effects on<br>tissue | Evaluate the thermal effects on tissue caused by the electrosurgical<br>functionalities of the Dexter L6 Electrosurgical Instruments | PASS |
# TABLE 5. BENCH TESTING
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
# TRAINING
The sponsor has developed a Dexter L6-specific use training program to ensure proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. Dexter L6 Surgical Systems are distributed only to facilities that implement and maintain the device-specific use training program, including assurance that users of the device have completed the device-specific use training program.
# HUMAN FACTORS TESTING
The Human Factors activities focused on developing the Dexter L6 Surgical System to facilitate safe and effective use by the intended users in the intended use environment. All activities were carried out in accordance with FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices", February 2016. The system was assessed through Human Factors activities in markets outside of the US, formative studies, and US Human Factors validation study.
The US validation study evaluated the training of surgical teams to complete a selection of evaluation activities (i.e., use scenarios [US] and knowledge tasks [KT]) to validate mitigations the Sponsor implemented to reduce the risk associated with critical tasks. Evaluation activities during the test session, each participant performed various tasks within representative, naturalistic use scenarios. Participants also performed knowledge tasks to evaluate critical tasks that cannot be evaluated during the hands-on use scenarios.
## SUMMARY OF CLINICAL INFORMATION
To support the De Novo request and gather evidence addressing the safety and efficacy profile of the Dexter system for use in the repair of inguinal hernias. Distalmotion conducted the the Robotic minimAlly invaSive inguinAl Hernia rEpAir with Dexter. RAS AHEAD Study. The purpose of this clinical investigation is to confirm the perioperative and early postoperative safety and clinical performance of the Dexter Robotic System, in subjects undergoing primary transperitoneal unilateral or bilateral inguinal hernia repair. The clinical investigation was conducted in accordance with Good Clinical Practice (GCP) as described in 21 CFR 812.28(A)(1).
## Study Design, Including Study Population and Investigation Period:
Single-arm, prospective, multicenter, open-label, clinical investigation. A total of 50 adult subjects across a minimum of three centers. were planned to be treated with the study device to demonstrate that the device fulfils its intended function adequately and that it was safe in doing so. The total was to ensure that at least 45 subjects completed the study for analysis.
{11}------------------------------------------------
A subject was considered "as treated" when at least one procedural step was started with the study device.
Subjects enrolled before surgery and followed up intra-operatively as well as post-operatively up to 30 days (+7 days) after the procedure. The study duration from the first procedure to the last follow-up assessment was 5 months.
## Demographics:
## Table 70 below provides a summary of the study population's demographics:
| | mITT(N=50)<br>Proportions presented as: % (n/N). (95%<br>Exact CI) |
|-----------------------------------------------------|--------------------------------------------------------------------|
| Age (years) | |
| Mean ± SD | 60.1 ± 15.2 |
| (95% CI of Mean) | (55.8, 64.4) |
| Median | 62.5 |
| Min, Max | 27.0, 89.0 |
| Race* | |
| Asian | 0.0% (0/50), (0.0%, 7.1%) |
| Black or African-American | 2.0% (1/50), (0.1%, 10.6%) |
| White/Caucasian | 58.0% (29/50), (43.2%, 71.8%) |
| Unknown or not reported | 40.0% (20/50), (26.4%, 54.8%) |
| Gender | |
| Male | 94.0% (47/50), (83.5%, 98.7%) |
| Female | 6.0% (3/50), (1.3%, 16.5%) |
| BMI (kg/m²) | |
| Mean ± SD | 25.7 ± 3.3 |
| Median | 25.1 |
| (95% CI of Mean) | (24.8, 26.6) |
| Min, Max | 18.4. 34.0 |
| ASA Status | |
| ASA I - a normal healthy subject | 26.0% (13/50), (14.6%, 40.3) |
| ASA II - a subject with mild systemic<br>disease | 66.0% (33/50), (51.2%, 78.8%) |
| ASA III - a subject with severe systemic<br>disease | 8.0% (4/50), (2.2%, 19.2%) |
## Table 70: Demographics
{12}------------------------------------------------
*Collecting information on race is considered sensitive data in certain European countries and was therefore not documented for the subjects enrolled in France (20 subjects).
## Study Endpoints:
#### Primary endpoint:
- ्तूबर-दिसंबर बाह्य Safety: Occurrence of adjudicated serious (Clavien - Dindo grades III-V), device related, adverse events perioperatively up to 30 days.
- Performance: Procedural success, defined as successful completion of the Dexter-assisted a procedure, i.e. free of any conversion to an open or fully laparoscopic surgical approach.
#### Secondary endpoints:
#### - Secondary safety endpoints:
- 1. Intraoperative Adverse Events and Complications
- 2. Postoperative complications up to 30 days post-operative (according to Clavien Dindo)
- 3. Device-related Adverse Events
- 4. Procedure related Adverse Events
- 5. Blood loss, measured in ml
- 6. Mortality
- 7. Re-hospitalization
#### Secondary performance endpoints a
- 1. Duration of docking time, measured in minutes
- 2. Operative time, measured in minutes (cut to suture)
- 3. Total time of robot in use
- 4. Instrument consumption measured in number of robotic instruments used
- 5. Use of extra trocar sites
- 6. Unplanned surgical steps
#### Other endpoints of interest -
- 1. Postoperative hospital stay, measured in days from first day after operation to discharge
- 2. Reoperation
- 3. Pain score through VAS
## Enrollment Criteria:
#### Inclusion criteria:
- Aged > 18 years ା
- = Subjects planned to undergo elective robot-assisted and laparoscopic surgery for primary unilateral or bilateral inguinal hernia repair using one camera port, two ports for the manipulating instruments, and additional ports as needed.
- Subject agrees to perform the 30-day follow-up assessment ﻴ
- Able to provide signed Informed Consent, according to local regulation -
## Exclusion criteria:
Page 13 De Novo Summary (DEN230084/S001)
{13}------------------------------------------------
## General
- As per Instructions for Use: ା
- Morbidly obese patients (BMI > 40)
- Any relative and absolute contraindications for the use of conventional endoscopes and I endoscopic surgical instruments
- I Need for robotic stapling, advanced energy delivery, ultrasound, cryoablation and microwave energy delivery
- . Bleeding diathesis
- Pregnancy 평
- Subjects with pacemakers or internal defibrillators I
- Any planned concomitant procedures . D
{14}------------------------------------------------
- Subject deprived of liberty by administrative or judicial decision or under legal guardianship. e
- ్డ Participation in another interventional clinical trial
#### Procedure-specific
- -Emergent inguinal hernia repair
- History of major* abdominal or pelvic surgery ي
- Strangulated and incarcerated inguinal hernia 测
- e Large scrotal hernia **
- Prior prostatectomy, hysterectomy and any other uterine procedures 00
- Prior radiotherapy for prostatic cancer e
- ୍ର Previous preperitoneal mesh placement on the side of the planned inguinal hernia repair. * Major is being defined as any previous surgery that involved a large abdominal incision (> 10 cm) and/or an extensive resection of organs that significantly altered the normal anatomy. ** Large scrotal hernia = when the hernia sac reaches the scrotum
# Study Investigators:
A total of three Investigators participated in this clinical study and performed the inguinal hernia procedures:
- Surgeon #1 (Centre Hospitalier de Saintes France): 20 subjects
- Surgeon #2 (Kantonspital Winterthur Switzerland): 20 subjects
- Surgeon #3 (UKSH Kiel Germany): 10 subjects
All surgeons went through a dedicated training curriculum ensuring that all successfully trained participants are capable of safe and effective usage of the Dexter System according to their roles in the OR.
All three investigators are general surgeons, their experience in robotic surgery ranged from 9 months to 5 years. Additional information on the surgeons' laparoscopic and robotic experience is provide below:
| Surgeons | Laparoscopic (LAP)<br>Experience | | Robotic Assisted Surgery<br>(RAS) Experience | | Dexter System Experience | |
|------------|----------------------------------|--------------------------------------------------------------------------------|----------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| | Years of LAP<br>experience | Estimated<br>number of LAP<br>procedures<br>completed<br>before study<br>start | Years of RAS<br>experience | Estimated<br>number of<br>RAS<br>procedures<br>completed<br>before study<br>start | Number of<br>Dexter<br>procedures<br>completed<br>before study<br>start | Number of<br>inguinal hernia<br>procedures<br>completed with<br>Dexter before<br>study start |
| Surgeon #1 | 35 | several<br>thousands | 17 months | 46 | 46 | 41 |
| Surgeon #2 | 7 | 500 | 9 months | 67 | 42 | 28 |
| Surgeon #3 | 19 | > 2600 | 5 | 290 | 38 | 7 |
Table 68: Surgeon's experience
{15}------------------------------------------------
| | Laparoscopic (LAP)<br>Experience | | Robotic Assisted Surgery<br>(RAS) Experience | | Dexter System Experience | |
|------------|----------------------------------|--------------------------------------------------------------------------------|----------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Surgeons | Years of LAP<br>experience | Estimated<br>number of LAP<br>procedures<br>completed<br>before study<br>start | Years of RAS<br>experience | Estimated<br>number of<br>RAS<br>procedures<br>completed<br>before study<br>start | Number of<br>Dexter<br>procedures<br>completed<br>before study<br>start | Number of<br>inguinal hernia<br>procedures<br>completed with<br>Dexter before<br>study start |
| Surgeon #1 | 35 | several<br>thousands | 17 months | 46 | 46 | 41 |
| Surgeon #2 | 7 | 500 | 9 months | 67 | 42 | 28 |
| Surgeon #3 | 19 | > 2600 | 5 | 290 | 38 | 7 |
Table 68: Surgeon's experience
# Clinical Study Results:
## Effectiveness Results:
Primary performance: Defined as successful completion of the Dexter-assisted procedure without conversion to an open or fully laparoscopic surgical approach, was confirmed in 49 of 50 procedures (98.0%).
In one subject, the procedure was converted to the laparoscopic mode because of work space limitation due to a non-optimal placement of the trocars. The trocars were placed too low (caudal) such that the surgeon was unable to reach the entire surgical area with the device. The surgeon safely completed the surgery in the laparoscopic mode, with no adverse event reported for this subject.
## Secondary performance:
- Duration of docking time, measured in minutes [Time Frame: Intra-operative]: The mean duration of docking time was 4.2 min (SD 2.0; range: 1.0 to 11.0 min).
- Operative time, measured in minutes (skin) [Time Frame: Intra-operative]: The mean i operative time was 54.5 min (SD 14.2; range: 28.0 to 93.0) for unilateral and 94.6 min (SD 22.2; range: 47.0 to 143.0 min) for bilateral inguinal hernia repair.
- Total time of robot in use: The mean total time of robotic use was 34.4 min (SD 11.6; range: हु 16.0 to 58.0 min) for unilateral and 68.3 min (SD 20.7; range: 22.0 to 108.0) for bilateral inguinal hernia repair.
- Instrument consumption, measured in number of robotic instruments used [Time Frame: -Intra-operative]: Procedures were performed with three instruments (96.0%) or four instruments (4.0%).
- Use of extra trocar sites: None of the procedures required the placement of extra trocars. -
- न्त Unplanned surgical steps: An unscheduled surgical step was required during three right-sided inguinal hernia procedures. All three unscheduled surgical steps referred to adhesiolysis and were performed with the robotic system.
# Safety Results:
Page 16 De Novo Summary (DEN230084/S001)
{16}------------------------------------------------
# Primary safety endpoint:
There were no postoperative serious (Clavien-Dindo grades III-V), device related, adverse events perioperatively up to 30 days. Adverse Events were reviewed by a CEC, none of the events were assessed as endpoint related.
# Secondary safety endpoints:
- Intraoperative Adverse Events and Complications [Time Frame: Intra-operative (grading according to ClassIntra)]: There were no intraoperative adverse events and complications.
{17}------------------------------------------------
- -Postoperative complications up to 30 days post-operative (according to Clavien-Dindo): There were 12 (18.0%) postoperative complications up to 30 days follow-up, thereof 11 (16.0%) Clavien-Dindo grade 1 complications and one (2.0%) Clavien-Dindo grade 2 complication (2.0%).
- · No intra-operative Adverse Event was reported. Hematoma and seroma were the most frequently reported Adverse Events, both with 6% of occurrence.
- 11 (16%) events were classified as Clavien-Dindo I, only 1 (2%) event (small o intestine occlusion) was classified as Clavien-Dindo II.
- None of the reported Adverse Events were considered as device-related.
- o Hematoma (3x):
- None of the reported hematoma required treatment/intervention' 2 out of the 3 o hematomas were resolved at study completion.
- o Seroma (3x)
- None of the reported seroma required treatment/intervention, all events were o ongoing at study completion.
- Persistent groin pain (1x) o
- o Subject received pain killer medication, event was ongoing at study completion with a VAS of 10mm, which can be considered as mild pain.
- o Trocar specific complication hyperesthesia (1x)
- o No treatment/intervention was required, event was ongoing at study completion.
- · Anesthesia complication nausea (1x)
- · No treatment/intervention was required, event was resolved at study completion.
- o Other Adverse Events (3x)
- 1x small intestine occlusion: subject was re-hospitalized and received o medications, event was resolved at study completion.
- o 1x erectile dysfunction secondary to hematoma: no treatment/intervention, event was resolved at study completion.
- 1x mild traumatic brain injury caused by a fall on the head: not procedure related, no y treatment/intervention, event was resolved with sequelae at study completion.Device-related Adverse Events [Time Frame: Intra-operative - within 30 post-operative day]: No devicerelated adverse events were reported.
- Procedure-related Adverse Events [Time Frame: intra-operative within 30 post-operative day]: There were 11 (16%) procedure-related adverse events up to 30 days follow-up, thereof 10 (16.0%)
- Mortality [Time Frame: intra-operative within 30 post-operative day]: There was no patient death.
- Blood loss, measured in ml [Time Frame: Intra-operative]: The mean intraoperative blood loss was 5.5 mL (SD 7.3: range 0.0. 20.0 mL). No intraoperative blood transfusion was needed.
- = Re-hospitalization [Time Frame: discharge - within 30 postoperative day]: One subject (2.0%) was re-hospitalized after being discharged after the index procedure
- Subject discontinuations: The study cohort included 51 enrolled subjects. The mITT population, defined as subjects who were enrolled and on whom robotic-assisted surgery
{18}------------------------------------------------
with Dexter was at least started, included 50 subject (2.0%) was withdrawn from the study for meeting an exclusion criterion (large scrotal hernia).
- Subject complaints: no subject complaints were reported. ्र
- -Device failures: no device failures were reported.
# Comparator Group:
Clinical performance and safety was compared to literature reporting for "primary unilateral or bilateral inguinal hernia via minimally invasive surgery", prepared by and independent (b)(4)(6)(6) The report discusses relevant contractor. clinical data on the safety and performance of robotic inguinal hernia repair (rIHR). For this, searches in the scientific literature were performed, including specific searches on the use of the comparative devices Da Vinci (manufacturer Intuitive Surgical, California, U.S.) and Senhance (manufacturer Asensus Surgical, North Carolina, U.S.).
The majority of studies evaluated the use of the da Vinci systems in the conduction of rTAPP (robotic transabdominal peritoneal) repair.
{19}------------------------------------------------
Systematic searches were conducted in PubMed. PubMed covers the international literature of every field of medicine and comprises more than 26 million citations for biomedical literature.
The primary component is the bibliographic database MEDLINE (Medical Literature Analysis and Retrieval System Online) which on its features approx. 5,600 international scholarly journals of the biomedicine and health sector. Furthermore, PubMed covers a wide range of life science journals, online books as well as citations that precede the date that the respective journal was selected for MEDLINE indexing.
The database search strategy was based on the definition of string searches that appropriately reflect the objectives of the present ER. Additional details are provided in the following table.
| Source | Search terms | Filters | Findings | Included |
|-------------------|--------------------------------------------------------------------------------------|---------------|----------|----------|
| MEDLINE<br>PubMed | (Robot OR Robotic) AND ("unilateral inguinal<br>hernia repair") | Last 10 years | 3 | 3 |
| | (Robot OR Robotic) AND ("bilateral inguinal hernia<br>repair") | Last 10 years | 3 | 2 |
| | (Robot OR Robotic) AND ("inguinal hernia repair")<br>AND ("safety" OR "performance") | Last 10 years | 32 | 22 |
| | ("Senhance" OR "da Vinci") AND "inguinal hernia<br>repair" | Last 10 years | 15 | 5 |
| Total | | | 53 | 32 |
Table 74: Search terms and strings defined for the systematic literature searches
The following criteria were used to select potentially relevant publications.
Table 75: Inclusion and exclusion criteria
| Inclusion<br>Criteria | The publication addresses the State-of-the-Art technology and provides relevant clinical information<br>pertaining to the safety and performance of robotic devices used for inguinal hernia repair |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The study design is appropriate to answer the proposed questions of the study (applicable when the<br>publication is derived from clinical studies) |
| | The publication presents sufficient information for a rational and objective assessment |
| Exclusion<br>Criteria | Full-text and/or publication not available |
| | The publication is a duplicate. Each publication will be considered only once. In the case of updated<br>versions of meta-analyses or systematic reviews, the most recent/updated version will be chosen |
| | Language is not English or English translation |
| | The publication is derived from a pre-clinical study (e.g., in-vitro, animal models, bench testing) or<br>phantom study |
| | The publication focus on the description of surgical technique, study protocol, and/or other similar<br>data that do not provide relevant clinical information on the safety, performance, usability, and/or off- label use<br>of the device |
{20}------------------------------------------------
Moreover, five additional publications were provided by the manufacturer Distalmotion, although one was excluded according to the above-mentioned criteria.
For further details about the search results, included and excluded and excluded articles, please see the Literature Search Report documented in the Expert [Attachment 007330 -Expert Report Literature Review (Distalmotion ER SP IHR rev01)].
This summary outlines the key elements of the Expert Report and relates them to the completed prospective inguinal hernia study with the Dexter System (RAS Ahead study #2023-01). The report is provided in Attachment 007330 - Expert Report Literature Review (Distalmotion ER SP IHR rev01).
# Medical History (Comorbidities)
Clinical data from comparative devices were extracted from a total of 19 publications.
## Demographics
## Literature Report:
The comparator population was composed mainly of men, mostly aged between 40 and 60 years. As suggested by the average BMI values, most of the subjects were either overweight or obese and although the presence of comorbidities was not reported by all the publications, individuals seem to be affected primarily by cardiovascular and metabolic complications. with hypertension (prevalence range: 29 - 46%) and diabetes (prevalence range: 7 - 15.8%) being the most frequently reported conditions. Nevertheless, some studies also reported a considerable number of patients with coronary artery disease (CAD, range: 5.1 - 16%) and chronic obstructive pulmonary (COPD. range: 5.1 - 11%).
A relevant number of individuals had a history of previous surgery, as between 21 and 48% of the patients had already undergone abdominal surgery. Regarding the American Society of Anesthesiologists (ASA) score, the majority of subjects were classified as Score 2 (moderate) and Score 3 (moderate to severe), and unilateral hernias were more frequently observed than bilateral cases. Cases of recurrent hernias were reported in different ranges, varying from 6.3 to 28.2%, and incarcerated hernias were not frequently found (1.9 - 5.1%).
# RAS Ahead Study with Dexter:
The treated population of the inguinal hernia study with Dexter is composed mainly of men (94%). The mean age of the studied population is 60.1 years. With a mean BMI of 25.7, slightly more than half of the population falls within the overweight range. The most frequently reported comorbidities are hypertension (22%), coronary artery disease (10%) and type II diabetes (8%).
Page 21 De Novo Summary (DEN230084/S001)
{21}------------------------------------------------
The majority of the subjects presented an ASA score of II (66%); for 8% of the study population de ASA score was III. More patients presented an unilateral hernia than a bilateral hernia (66% vs. 34%). Patients with a history of major abdominal or pelvic surgery were excluded from the study.
Comparison of Study group to Literature Comparator group:
Both the study and comparator cohorts consisted mostly of male subjects with unilateral hernias. The comparator cohort appears to be comprised of patients with slightly lower BMI compared to the study population. However, the rates and types of co-morbid conditions appear to be similar, with a similar distribution of ASA scores between both cohorts. Another difference between the comparator cohort and study population is the inclusion of both recurrent and incarcerated hernia (which were specifically excluded in the study population). However, considering the similarities and differences between the study and comparator cohorts, it appears the literature cohort represents a valid comparator group which may assist in the benefit risk assessment of the subject device.
| Most Frequent<br>Comorbidities | RAS Ahead (Dexter)<br>(n=50) | RWE (Expert Report) |
|--------------------------------|------------------------------|--------------------------------|
| Hypertension | 11 (22%) | prevalence range of 29 - 46% |
| Coronary artery<br>disease | 5 (10%) | prevalence range of 5.1 - 16% |
| Type II diabetes | 4 (8%) | prevalence range of 7 - 15.8%) |
Table 76: Most frequent comorbidities
# Conclusion
The prospective study on the treatment of inguinal hernia enrolled a patient population with comorbidities that are representative of the RWE population as described in the Expert Report.
{22}------------------------------------------------
Both, the Expert Report and the RAS Ahead study do report a study population primarily affected by cardiovascular and metabolic complications including hypertension, coronary artery disease and diabetes. The reported values from the RAS Ahead study are in line with the values reported from the RWE. except the hypertension rate which is lower for the RAS Ahead study. A recent study published by Hajili et al. in 2023 concluded that the TAPP method for patients with cardiovascular disease does not elevate the individual perioperative risk (Kamran Hajili, Alberto Vega Hernandez, Jakob Otten, Dana Richards, and Claudia Rudroff, "Risk factors for early and late morbidity in patients with cardiovascular disease undergoing inguinal hernia repair with tailored approach: single-center cohort study., " BMC Surgery, 2023).
# Conversion Rates and Operative Time:
# Literature Report:
Conversion rates to open surgery and total operative time are frequently used as parameters to assess the performance of the device. Conversion rates were reported by 16 out of 19 studies. Six publications reported no cases of conversion to open surgery and two publications informed no cases of conversion either to open or to laparoscopy [12] [26]. Aiolfi et al found a pooled prevalence of conversion of 0.14%, and the other publications also reported very low conversion rates, mainly below 2%. In particular, Maas and colleagues evaluated conversion rates to laparoscopy, and not to open surgery. Two cases of conversion to laparoscopy occurred among 43 patients (4.6%).
All the selected studies reported the operative time which varied widely between the studies. ranging from 18 to 240 min, with most of the studies reporting mean or median operative times between 60 and 100 min. Six publications reported an operative time >100 min. Aiolff et al., 2019 informed that operative time ranged from 45 - 180.4 min in the studies evaluated by the meta-analysis. Importantly, the different definition of operative time (e.g. docking and console time, only console time, time spent in the operating theatre, and others) contributes to a higher heterogenicity of data which may affect overall conclusions.
# RAS Ahead Study with Dexter
One procedure out of fifty (2%) was permanently converted to laparoscopy at the time of the mesh placements due to the limitation of the workspace area. The surgery was then completed laparoscopically. This workflow was applied to the right and left inguinal hernias. No intra- or post-procedural complications were reported for this patient.
During one procedure for the treatment of a unilateral inguinal hernia, the surgeon switched to the laparoscopic mode during the reduction of the hernia content due to the size of the hernia and difficulties in pulling the hernia sac. Following the mesh placement and fixation, the peritoneal closure and end of the procedure was complete robotically.
The mean reported operative times (skin incision to skin suturing) were 54.5 minutes for unilateral procedures and 94.6 minutes for bilateral procedures.
{23}------------------------------------------------
# Conclusion
With a conversion rate of 2%, the rate of permanent conversion to an open or fully laparoscopic surgical approach is comparable to what is being reported in the RWE report and previously cleared robotic systems. The reported rate of permanent conversions is also within the rate of 10% that was defined in the study protocol for the Clinical Acceptance Criteria. Regarding the operative time, results from the RAS Ahead study obtained a mean operative times of 54.5 minutes for unilateral and 94.6 minutes for bilateral hernias, which is in line with the mean operative time between of 60 to 100 minutes reported in the RWE report.
# Complications and Adverse Events
# Literature Report:
The use of da Vinci systems in the conduction of IHR was associated with low conversion rates (0 to 4.6%), in which the few reported conversion cases were related either to the experience of the surgeon, anatomical aspects, or other specific clinical characteristics. The reported complications were mainly mild and moderate (e.g. seroma, hematoma, pain) and reported at different rates (hematoma: 0.25 - 4.1%, seroma: 0.25 - 25.6%, urinary retention: 1.3 -18%). Nonetheless, the majority of publications reported low rates of intra (<1%) and postoperative (< 5%) complications.
Although the Clavien-Dindo Classification is widely used by the medical and the scientific community to better group complications, only five publications used this classification, in which the majority of the patients were classified as Clavien-Dindo Class I and II (0 - 22%). which are mainly mild and moderate events, which may require more specific medication, blood transfusion or re-hospitalization. More severe complications including Grade III and IV were observed in lower rates (< 6.7%), and no cases of Grade V were reported in patients submitted to rIHR.
Overall, the scientific literature shows that rIHR is characterized by low rates of conversion to open surgery, inguinal hernia recurrence, and complications.
## RAS Ahead Study with Dexter as compared to the above literature study.
A total of 12 Adverse Events were reported on the patient population treated within the inguinal hernia study. Adverse Events were adjudicated by a CEC:
Table 77: Perioperative and update 30 days postoperative AEs
{24}------------------------------------------------
| | mITT<br>(N= 50) | | | |
|------------------------------|-----------------|----------------|------------------------|----------------|
| | Adverse Events | | Serious Adverse Events | |
| Surgery Type | # AEs | # (%) Subjects | # AEs | # (%) Subjects |
| Anesthesia complications | 1 | 1 (2.0%) | 0 | 0 |
| Hematoma | 3 | 3 (6.0%) | 0 | 0 |
| Other | 3 | 3 (6.0%) | 1 | 1 (2.0%) |
| Persistent groin pain | 1 | 1 (2.0%) | 0 | 0 |
| Seroma | 3 | 3 (6.0%) | 0 | 0 |
| Trocar-specific complication | 1 | 1 (2.0%) | 0 | 0 |
| Total | 12 | 9 (18.0%) * | 1 | 1 (2.0%) |
* The 9 is representative of unique subjects that had a procedure-related event. Two subjects each had 2 procedure- related events which leads to the 11 events in 9 subjects.
No intra-operative Adverse Event was reported. Perioperative and postoperative, hematoma and seroma were the most frequently reported Adverse Events, both with 6% of occurrence. 11 (16%) events were classified as Clavien-Dindo I, only 1 (2%) event (small intestine occlusion) was classified as Clavien-Dindo II. None of the reported Adverse Events were considered as device-related. With respect to the reported three Hematoma:
- ા None of the reported hematoma required treatment/intervention' 2 out of the 3 hematomas were resolved at study completion.
- Seroma (3x) টি
- ﺑﺘ None of the reported seroma required treatment/intervention, all events were ongoing at study completion.
- Persistent groin pain (1x) -
- -Subject received pain killer medication, event was ongoing at study completion with a VAS of 10mm, which can be considered as mild pain.
- -Trocar specific complication - hyperesthesia (1x)
- No treatment/intervention was required, event was ongoing at study completion. -
- -Anesthesia complication - nausea (1x)
- No treatment/intervention was required, event was resolved at study completion. -
- -Other Adverse Events (3x)
- 1x small intestine occlusion: subject was re-hospitalized and received medications, event was resolved at study completion.
- . 1x erectile dysfunction secondary to hematoma: no treatment/intervention, event was resolved at study completion.
- 1 x mild traumatic brain injury caused by a fall on the head: not procedure related, no treatment/intervention, event was resolve7d with sequelae at study completion.
- -100% of the data reported in the RAS Ahead study were source data verified. This might potentially have an impact on the number of reported adverse events compared to the data reported in the RWE, where the data might not be monitored with the same scrutiny and thus have a lower number of reported complications.
# Conclusion on Safety and Performance per Literature Report
Page 25 De Novo Summary (DEN230084/S001)
{25}------------------------------------------------
Intraoperative complications and adverse events were rarely observed with comparative devices. Similar results were obtained with Dexter, where all adverse events occurred in the postoperative phase and no device-related events were reported. Adverse events based on the Clavien-Dindo classification referred to grade I and II for comparative devices, including hematoma, superficial infections, use of analgesics for pain management, or fever. Grade III and IV events only affected few patients. Similar to data from comparative devices. Grade I and II adverse events included hematoma, seroma, or groin pain. None of the study subjects experienced any Grade III and IV complications.
# Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
# LABELING
The device user manual and instructions for use include a description of the device technical parameters and instructions for use for the device. The user manual also contains relevant findings from the clinical study with the performance characteristics of the device when used as intended. The document also states the shelf life for any sterile components as well as the necessary measures to properly dispose of any single use items and clean the reusable components of the device.
Labeling also includes the following:
- (i) Exclusion parameters for use of the device.
- A detailed summary of the clinical evaluations pertinent to use of the device and (ii) accessories: and
- (iii) A statement that the device shall only be used by personnel that have been trained in its operation
## RISKS TO HEALTH
Table 12 identifies the risks to health that may be associated with use of an electromechanical surgical system with transient sterile field presence of both surgeon and primary control interface.
| Risks to Health | Mitigation Measures |
|-------------------------------------------------------------------------------------------------------------|----------------------------------|
| Electrical fault, electromagnetic interference,<br>mechanical fault, or system malfunction<br>resulting in: | Clinical performance testing |
| • Tissue injury | Postmarket surveillance |
| | Non-clinical performance testing |
| | Sterilization validation |
#### TABLE 12. RISKS TO HEALTH
{26}------------------------------------------------
| Risks to Health | Mitigation Measures |
|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Electric shock Prolonged procedure time | Reprocessing validation |
| | Annual reporting |
|…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.