Intended Use

The StaClear Syringe is intended to inject fluids into, or withdraw fluids from, the body. The StaClear Syringe is indicated for intravitreal use.

Device Description

The StaClear Syringe is a single-use piston syringe intended for use by healthcare professionals for general-purpose fluid aspiration/injection. Its operation is manual. The StaClear Syringe is single use only, non-toxic, non-pyrogenic, and sterilized by ethylene oxide gas. It uses standard syringe components with a low-particulate gas plasma crosslinked silicone coating called TriboLink-Si. The StaClear Syringe is suitable for ophthalmic use.

The StaClear Syringe is offered in the following configurations:

  • SC250AN – Consisting of a 0.25 mL graduated barrel, plunger, plunger stopper, needle, needle shield, and plunger cap.
  • SC250LS – Consisting of a 0.25 mL graduated barrel with an ISO 80369-7-compliant luer-slip connector, plunger, and plunger stopper.
  • SC250LL – Consisting of a 0.25 mL graduated barrel with an ISO 80369-7-compliant luer-lock connector, plunger, and plunger stopper.
AI/ML Overview

The provided document is an FDA 510(k) clearance letter for the StaClear Syringe, which is a medical device, not an AI/ML algorithm. Therefore, many of the requested fields, such as "number of experts used to establish the ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone algorithm performance," "type of ground truth," and information about training sets, are not applicable to this type of device and therefore not present in the document.

The document primarily focuses on demonstrating substantial equivalence to a predicate device through adherence to established performance standards for syringes.

Here's the information that can be extracted from the provided text regarding acceptance criteria and performance:

1. A table of acceptance criteria and the reported device performance

The document lists various ISO and USP standards as the basis for performance evaluation. The "Acceptance Criteria" are implied to be compliance with these standards, and the "Reported Device Performance" is stated as "Identical" or "Compliant" with these standards.

Attribute/TestAcceptance Criteria (Implied)Reported Device Performance
Syringe Performance (ISO 7886-1)Compliance with ISO 7886-1:2017Identical to predicate; Device maintains performance per ISO 7886-1 after 5-years of aging.
- CleanlinessPassIdentical
- Acidity and AlkalinityPassIdentical
- Extractable MetalsIOL limits per ISO 11979-5Identical
- LubricantPassIdentical
- Tolerance on GraduationsPassIdentical
- Stopper DetachmentPassIdentical
- Dead SpacePassIdentical
- Air and Liquid Leakage Past StopperPassIdentical
- Plunger ForcePassIdentical
- Fit of StopperPassIdentical
Needle Performance (SC250AN only) (ISO 7864 & ISO 9626)Compliance with ISO 7864:2016 & ISO 9626:2016Identical to predicate; Device maintains performance per ISO 7864, ISO 9626 after 5-years of aging.
- CleanlinessPassIdentical
- Acidity and AlkalinityPassIdentical
- Extractable MetalsIOL limits per ISO 11979-5Identical
- Tolerance on LengthPassIdentical
- Tube DefectsPassIdentical
- LubricantPassIdentical
- Point DefectsPassIdentical
- Needle Penetration ForcePassIdentical
- Bond Between Tube and HubPassIdentical
- Patency of LumenPassIdentical
- MaterialsPassIdentical
- Surface FinishPassIdentical
- Size DesignationPassIdentical
- Outside DiameterPassIdentical
- StiffnessPassIdentical
- Resistance to BreakagePassIdentical
- Resistance to CorrosionPassIdentical
Luer Connector Performance (SC250LS and SC250LL only) (ISO 80369-7)Compliance with ISO 80369-7:2021Compliant with established performance standards; Device maintains performance per ISO 80369-7 after 5-years of aging.
- MaterialsPassCompliant
- Dimensional RequirementsPassCompliant
- Fluid LeakagePassCompliant
- Air LeakagePassCompliant
- Stress CrackingPassCompliant
- Separation ForcePassCompliant
Luer-Lock only (ISO 80369-7)Compliance with ISO 80369-7:2021Compliant with established performance standards.
- Unscrewing TorquePassCompliant
- Resistance to OverridingPassCompliant
Particulate TestingCompliance with USP <788> & USP <789>Identical to predicate; Device performs as intended.
- Particulate Matter in Injections (USP <788>)PassIdentical
- Particulate Matter in Ophthalmic Solutions (USP <789>)PassIdentical
BiocompatibilityCompliance with ISO 10993-1 for external communicating device with prolonged tissue contactBiocompatible for intended use.
SterilitySterilized to SAL 10⁻⁶; Endotoxin per USP <85> with IOL limits; Sterility per USP <71>Sterilized to SAL 10⁻⁶; Endotoxin levels meet IOL limits; Device maintains sterility per USP <71> after 5-years of aging.
Shelf LifeMaintain performance and sterility for 5 yearsTesting demonstrates performance and sterility maintained over 5 years after accelerated aging.
PackagingCompliance with ISO 11607-1:2019 for sterile barrierCompliant with established requirements per ISO 11607-1:2019. (For Tyvek Pouch packaging)
- Sterile Barrier IntegrityPass (e.g., using bacterial aerosol followed by sterility tests per USP <71>)Maintained.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample sizes used for testing each attribute. The data provenance (country of origin, retrospective/prospective) is also not provided. The testing is described as "non-clinical testing" and "performance testing performed as per the design control system," suggesting in-house, prospective testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is a physical medical device, not an AI/ML algorithm requiring expert ground truth for image interpretation or diagnosis. The "ground truth" for the device's performance is objective measurements against established engineering and biocompatibility standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is a physical medical device, not an AI/ML algorithm requiring adjudication of interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a physical medical device and does not involve AI assistance for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm. The "standalone" performance refers to the device's adherence to specified standards in direct testing.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is adherence to recognized international and national standards for medical devices, specifically:

  • ISO standards (e.g., ISO 7886-1, ISO 7864, ISO 9626, ISO 80369-7, ISO 10993-1, ISO 11607-1) which define performance characteristics, materials, and testing methodologies.
  • USP standards (e.g., USP <788>, USP <789>, USP <71>, USP <85>) which define testing for particulate matter, sterility, and endotoxins.
  • ASTM standards (e.g., ASTM D4169, ASTM F1980) for packaging and accelerated aging.

These standards provide objective, measurable criteria.

8. The sample size for the training set

Not applicable. This is a physical medical device and does not involve a training set as would an AI/ML algorithm.

9. How the ground truth for the training set was established

Not applicable. This is a physical medical device and does not involve a training set.

FDA 510(k) Clearance Letter - StaClear Syringe

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 3, 2025

StaClear, Inc.
Jackson Thornton
Director of Research
7250 ACC Blvd.
Raleigh, NC 27617

Re: K243936
Trade/Device Name: StaClear Syringe (SC250AN); StaClear Syringe (SC250LS); StaClear Syringe (SC250LL)
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: QLY, FMF, FMI
Dated: November 5, 2024
Received: January 28, 2025

Dear Jackson Thornton:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K243936 - Jackson Thornton Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243936 - Jackson Thornton Page 3

Sincerely,

Shruti N. Mistry -S

Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known): K243936

Device Name: StaClear Syringe (SC250AN); StaClear Syringe (SC250LS); StaClear Syringe (SC250LL)

Indications for Use (Describe)

The StaClear Syringe is intended to inject fluids into, or withdraw fluids from, the body. The StaClear Syringe is indicated for intravitreal use.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

StaClear
7250 ACC Blvd.
Raleigh, NC 27617. USA
Phone: +1(919)838-6716
E-mail: info@staclear.com

510(k) SUMMARY

I. SUBMITTER

StaClear, Inc.
7250 ACC Blvd.
Raleigh, NC 27617
Phone: 1-919-838-6716
Email: info@staclear.com
Primary Contact: Jackson Thornton, Director of Research
Date prepared: March 26, 2025

II. DEVICE

Device Name: StaClear Syringe
Common/Usual Name: Piston Syringe
Classification Name: Ophthalmic Syringe
Regulatory Class: 2
Product Code: QLY
Subsequent Product Code: FMF; FMI

III. PREDICATE DEVICE

K200242 StaClear Syringe

The predicate device has not been subject to design-related recall.

IV. DEVICE DESCRIPTION

The StaClear Syringe is a single-use piston syringe intended for use by healthcare professionals for general-purpose fluid aspiration/injection. Its operation is manual. The StaClear Syringe is single use only, non-toxic, non-pyrogenic, and sterilized by ethylene oxide gas. It uses standard syringe components with a low-particulate gas plasma crosslinked silicone coating called TriboLink-Si. The StaClear Syringe is suitable for ophthalmic use.

The StaClear Syringe is offered in the following configurations:

  • SC250AN – Consisting of a 0.25 mL graduated barrel, plunger, plunger stopper, needle, needle shield, and plunger cap.
  • SC250LS – Consisting of a 0.25 mL graduated barrel with an ISO 80369-7-compliant luer-slip connector, plunger, and plunger stopper.
  • SC250LL – Consisting of a 0.25 mL graduated barrel with an ISO 80369-7-compliant luer-lock connector, plunger, and plunger stopper.

V. INTENDED USE / INDICATIONS FOR USE

The StaClear Syringe is intended to inject fluids into, or withdraw fluids from, the body. The StaClear Syringe is indicated for intravitreal use.

Page 1 of 6

Page 6

StaClear
7250 ACC Blvd.
Raleigh, NC 27617. USA
Phone: +1(919)838-6716
E-mail: info@staclear.com

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The technological characteristics of the StaClear Syringe are substantially equivalent to the predicate device. The intended use and indications for use of the subject device are identical to the predicate device.

Comparisons of the technological characteristics between the subject and predicate device are illustrated in the table below:

STACLEAR SYRINGE SUBSTANTIAL EQUIVALENCE COMPARISON

AttributePredicate Device K200242Subject DeviceComparison
Device NameStaClear Syringe Attached NeedleStaClear Syringe Attached NeedleStaClear Syringe Luer SlipStaClear Syringe Luer LockDifferent models based on connector type
ModelSC250SC250ANSC250LSSC250LLChange model suffix to designate: AN = Attached Needle, LS = Luer Slip, LL = Luer Lock
DeviceOphthalmic Syringe; Syringe, PistonIdentical
Device ClassClass IIIdentical
FDA Product CodeQLY; FMF; FMIQLY; FMF; FMIQLY; FMFQLY; FMFThe luer-tip models do not include a needle, so the FMI designation is not required.
Intended Use/Indication for UseThe StaClear Syringe is intended to inject fluids into, or withdraw fluids from, the body. The StaClear Syringe is indicated for intravitreal use.Identical
Mechanism of ActionManualIdentical
Single-Use OnlyYesIdentical
Provided SterileYesIdentical, endotoxin meets intraocular limits
Sterilization MethodEthylene OxideIdentical, EO/ECH residuals meet intraocular standards
SAL10⁻⁶Identical
Shelf Life1 year5 years5 years5 yearsTesting supports that performance characteristics and sterility are maintained over 5 years.
Syringe Volume0.25 mLIdentical
Connector TypeAttached NeedleAttached NeedleLuer SlipLuer LockCompliant with established performance standards: ISO 7886-1, ISO 7864, ISO 9626, ISO 80369-7. Compliant with additional testing and reduced acceptance levels for ophthalmic use.

Page 2 of 6

Page 7

StaClear
7250 ACC Blvd.
Raleigh, NC 27617. USA
Phone: +1(919)838-6716
E-mail: info@staclear.com

STACLEAR SYRINGE SUBSTANTIAL EQUIVALENCE COMPARISON (continued)

AttributePredicate Device K200242Subject DeviceComparison
Needle Gauge31 G31 GNoneNoneLuer configurations do not have an attached needle.
Needle Length5/16 in.5/16 in.NoneNoneLuer configurations do not have an attached needle.

MATERIALS

AttributePredicate Device K200242Subject DeviceComparison
BarrelPolypropyleneLuer configurations have different barrel connector types but the same barrel material/dimensions for the fluid-path contacting surfaces. Final, finished device is biocompatible for intended use per ISO 10993-1 as an external communicating device with prolonged tissue contact.
PlungerPolyethyleneIdentical
Plunger StopperPolyisopreneIdentical
Needle304 Stainless Steel304 Stainless SteelNoneNoneLuer configurations do not have an attached needle or glue.
Needle LubricantSilicone OilSilicone OilNoneNoneLuer configurations do not have an attached needle.
Needle ShieldPolyethylenePolyethyleneNoneNoneLuer configurations do not have a needle shield.
Plunger CapPolyethyleneIdentical
Barrel LubricantTriboLink-SiIdentical

PACKAGING

AttributePredicate Device K200242Subject DeviceComparison
Primary PackageLLDPE Bag (10 self-contained syringes)LLDPE Bag (20 self-contained syringes)Tyvek Pouch (1 unit)Tyvek Pouch (1 unit)LLDPE Bag for the attached needle model contains self-contained syringe units, so the sterile barrier is maintained between the needle shield and plunger cap. Tyvek Pouch packaging maintains sterility within the Tyvek Packaging. Packaging where the sterile barrier is maintained within self-contained syringe units challenged using bacterial aerosol followed by sterility tests per USP <71>.
Inner BoxSBSIdentical

Page 3 of 6

Page 8

StaClear
7250 ACC Blvd.
Raleigh, NC 27617. USA
Phone: +1(919)838-6716
E-mail: info@staclear.com

STACLEAR SYRINGE SUBSTANTIAL EQUIVALENCE COMPARISON (continued)

AttributePredicate Device K200242Subject DeviceComparison
Shipping CaseCardboardIdentical. Packaging where the sterile barrier is maintained within a Tyvek pouch complies with established requirements per ISO 11607-1:2019.

PERFORMANCE

AttributePredicate Device K200242Subject DeviceComparison
Syringe PerformanceISO 7886-1Identical
Particulate TestingUSP <788>, USP <789>Identical
Luer Connector PerformanceNoneISO 80369-7Attached needle configurations are identical so testing is not applicable. Luer configurations are compliant with established performance standards
Needle PerformanceISO 7864, ISO 9626NoneAttached needle configurations are identical. Luer configurations do not have an attached needle so testing is not applicable

VII. PERFORMANCE DATA

The following non-clinical testing was performed to confirm the safety and effectiveness of the StaClear Syringe as compared to the predicate device. Performance testing was performed as per the design control system.

  • ISO 7886-1:2017
    • Cleanliness
    • Acidity and Alkalinity
    • Extractable Metals (IOL limits per ISO 11979-5)
    • Lubricant
    • Tolerance on Graduations
    • Stopper Detachment
    • Dead Space
    • Air and Liquid Leakage Past Stopper
    • Plunger Force
    • Fit of Stopper

Page 4 of 6

Page 9

StaClear
7250 ACC Blvd.
Raleigh, NC 27617. USA
Phone: +1(919)838-6716
E-mail: info@staclear.com

  • ISO 7864:2016 (SC250AN only)

    • Cleanliness
    • Acidity and Alkalinity
    • Extractable Metals (IOL limits per ISO 11979-5)
    • Tolerance on Length
    • Tube Defects
    • Lubricant
    • Point Defects
    • Needle Penetration Force
    • Bond Between Tube and Hub
    • Patency of Lumen
  • ISO 9626:2016 (SC250AN only)

    • Materials
    • Surface Finish
    • Cleanliness
    • Limits for Acidity and Alkalinity
    • Size Designation
    • Outside Diameter
    • Stiffness
    • Resistance to Breakage
    • Resistance to Corrosion
  • ISO 80369-7:2021 (SC250LS and SC250LL only)

    • Materials
    • Dimensional Requirements
    • Fluid Leakage
    • Air Leakage
    • Stress Cracking
    • Separation Force
  • ISO 80369-7:2021 (SC250LL only)

    • Unscrewing Torque
    • Resistance to Overriding
  • Particulate Testing

    • USP <788> Particulate Matter in Injections
    • USP <789> Particulate Matter in Ophthalmic Solutions
  • Biocompatibility (ISO 10993-1)

    • StaClear Syringe is categorized as an external communicating devices with prolonged tissue contact (prolonged due to repeated/chronic treatment for intravitreal injections).
  • Sterility and Shelf Life

    • StaClear Syringe is sterilized to SAL 10⁻⁶ using a validated ethylene oxide sterilization cycle. The sterilization residual levels meet the requirements of ISO 10993-7 for IOL. Bacterial endotoxin testing per USP <85> with limits set for IOL is conducted for each lot.
    • After distribution simulation per ASTM D4169 and 5-year accelerated aging per ASTM F1980, testing demonstrates the device maintains performance per ISO 7886-1, ISO

Page 5 of 6

Page 10

StaClear
7250 ACC Blvd.
Raleigh, NC 27617. USA
Phone: +1(919)838-6716
E-mail: info@staclear.com

7864, ISO 9626, and/or ISO 80369-7, and maintains sterility per USP <71> or a sterile barrier per ISO 11607-1 after 5-years of aging.

VIII. SUBSTANTIAL EQUIVALENCE

The StaClear Syringe is substantially equivalent to the predicate when evaluating intended use and technological characteristics.

  • The subject device has the identical intended use and indications for use as the predicate device.
  • The subject device has identical materials, manufacturing process flow, sterilization process, and principle of operation as the predicate device.
  • The technological characteristics of the subject device are substantially equivalent to the predicate device with only minor differences in device design (connector type) and packaging configuration.
  • Performance testing demonstrates that the subject device is as safe and effective as the legally marketed predicate device and does not raise new or different questions of safety and effectiveness compared to the predicate device.

IX. CONCLUSIONS

StaClear Syringe meets the established safety and performance characteristics of an ophthalmic syringe. Performance testing according to ISO 7886-1, ISO 7864, ISO 9626, ISO 80369-7, and particulate testing per USP<788> and USP <789> demonstrate the StaClear Syringe is as safe and effective as the predicate device and will perform as intended.

StaClear Syringe is biocompatible for its intended use and demonstrates equivalent performance to its predicate device. The StaClear Syringe SC250AN, SC250LS, and SC250LL models are substantially equivalent to the predicate StaClear Syringe SC250 model cleared under K200242.

Page 6 of 6

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).