K Number
K243792

Validate with FDA (Live)

Date Cleared
2025-02-18

(70 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate (Black/White) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with Chemotherapy drugs and Fentanyl using ASTM D6978-05.

Device Description

Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate (Black, White) is a Class I patient examination glove and specialty chemotherapy gloves, that made from synthetic nitrile latex. They are non-sterile, powder-free, fingertip textured, ambidextrous with beaded cuff, and single use only, and come in different sizes- XS, S, M, L, XL and XXL. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs and Fentanyl Citrate.

AI/ML Overview

The provided document pertains to the 510(k) premarket notification for Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate (Black, White).

This submission is for a Class I medical device (gloves), which typically does not involve complex AI algorithms or diagnostic imaging. Therefore, many of the requested criteria related to AI/MRMC studies, expert consensus for ground truth, and training set information are not applicable to this type of device and submission.

The "acceptance criteria" here refer to the performance standards and safety requirements for medical gloves, particularly their resistance to permeation by chemotherapy drugs and fentanyl, and their physical and biological properties.

Here's a breakdown based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document presents acceptance criteria and reported performance in the "Summary of Non-Clinical Performance Data" and the "Chemotherapy Drugs Comparison Claim" tables.

Table 1: General Performance Acceptance Criteria and Reported Results (from page 10)

Test PerformedMethodologyAcceptance CriteriaResult
Freedom From HolesASTM D6319-19, ASTM D5151-19Meet requirement inspection level G-1, AQL 2.5 (ISO2859-1)Pass
Dimension-LengthASTM D6319-19Minimum 230mm for all sizesPass
Dimension-WidthASTM D6319-19XS: 70±10mm, S: 80±10mm, M: 95±10mm, L: 110±10mm, XL: 120±10mm, XXL: 130±10mmPass
Dimension-ThicknessASTM D6319-19Finger: 0.05mm (Min), Palm: 0.05mm (Min)Pass
Physical propertiesASTM D6319-19, ASTM D412-16Tensile Strength (Min 14 MPa), Elongation (Before Aging 500% and after aging 400% Min)Pass
Powder ResidueASTM D6319-19, ASTM D6124-06Not more than 2 mg per glovePass
Skin IrritationISO10993-23:2021Not a primary skin irritantPass
Skin SensitizationISO10993-10:2021Not a contact sensitizerPass
Acute Systemic ToxicityISO 10993-11:2017No acute systemic toxicityPass

Table 2: Permeation by Chemotherapy Drugs and Fentanyl Citrate Acceptance Criteria and Reported Results (from page 9)

NO.Chemotherapy DrugsMinimum Breakthrough Detection Time, minutes (Black Glove)Minimum Breakthrough Detection Time, minutes (White Glove)Predicate Device K230779 (Black)Result of Comparison (to Predicate)
1Carmustine (3.3mg/ml)14.216.711.3Different (both subject devices perform better than predicate)
2Cisplatin (1mg/ml)>240>240>240Same
3Cyclophosphamide (20mg/ml)>240>240>240Same
4Dacarbazine (10.0 mg/ml)>240>240>240Same
5Doxorubicin HCl (2 mg/ml)>240>240>240Same
6Etoposide (20mg/ml)>240>240>240Same
7Fluorouracil (50mg/ml)>240>240>240Same
8Paclitaxel (6mg/ml)>240>240>240Same
9Thiotepa (10mg/ml)14.716.224.7Different (subject devices perform worse for Thiotepa compared to predicate, but the labeling calls out the warning for these drugs)
NO.Fentanyl CitrateMinimum Breakthrough Detection Time, minutes (Black Glove)Minimum Breakthrough Detection Time, minutes (White Glove)Predicate device K230779Result of Comparison
1Fentanyl Citrate Injection, 100mcg/2ml>240>240>240Same

Note: The warning regarding Carmustine and Thiotepa for both black and white gloves is highlighted due to their low permeation times, similar to the predicate device, which also shows a low time for Carmustine and a higher time for Thiotepa. The critical aspect for these drugs is to provide clear warning on the labeling.

2. Sample size used for the test set and the data provenance

The document specifies that the tests were non-clinical, adhering to various ASTM and ISO standards. These standards prescribe the sample sizes and testing methodologies. For example:

  • Freedom from Holes (ASTM D6319-19, ASTM D5151-19): References "inspection level G-I, AQL 2.5 (ISO2859-1)," which indicates specific sampling plans and acceptance quality limits for batch inspection, not individual sample sizes like in a typical clinical study.
  • Permeation by Chemotherapy Drugs (ASTM D6978-05): This standard outlines the test method, including the number of specimens (typically 3 specimens per drug per material), continuous monitoring of breakthrough, and reporting of average breakthrough times.
  • Biocompatibility (ISO 10993 series): These standards specify the number of animals or in-vitro tests required, but precise numbers are not explicitly stated in this summary.

Data Provenance: The data comes from non-clinical laboratory testing conducted by the manufacturer (Basic Medical Technology Inc.) to demonstrate compliance with recognized industry standards (ASTM, ISO). The document does not specify the country of origin of the raw test data but implies it was generated to support a US FDA submission. The tests are prospective in the sense that they were conducted for this specific submission to demonstrate the device's performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not Applicable. This is a physical device (medical gloves) and not an AI/diagnostic software. "Ground truth" in this context is established by standardized, objective laboratory test methods (e.g., measuring dimensions, tensile strength, breakthrough time using analytical instruments), not by expert consensus on clinical images or data.

4. Adjudication method for the test set

Not Applicable. As per point 3, ground truth is established by objective laboratory measurements and adherence to standardized testing protocols, not by expert review or adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not Applicable. This is a Class I physical medical device (gloves), not an AI algorithm or diagnostic tool involving human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not Applicable. This is a physical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device is based on objective measurements performed according to recognized international and national standards (ASTM, ISO). This includes:

  • Physical properties (e.g., dimensions, tensile strength, elongation).
  • Chemical resistance (breakthrough time for chemotherapy drugs and fentanyl).
  • Biological safety (irritation, sensitization, systemic toxicity).

8. The sample size for the training set

Not Applicable. This is a physical device. There is no AI model or "training set" in the context of this submission. The "training" for the device would be the manufacturing process and quality control, which are governed by Quality System (QS) regulations.

9. How the ground truth for the training set was established

Not Applicable. As explained in point 8, there is no training set as per AI/ML terminology. The manufacturing process and material specifications are developed and validated based on established engineering principles and industry standards to ensure the gloves meet their intended performance characteristics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 18, 2025

Basic Medical Technology Inc. John Zhao General Manager 5300 Concours Street Ontario. California 91764

Re: K243792

Trade/Device Name: Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate (Black,White) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, QDO, OPJ Dated: December 10, 2024 Received: December 10, 2024

Dear John Zhao:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

ALLAN GUAN -S

For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K243792

Device Name

Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate (Black)

Indications for Use (Describe) Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentany] Citrate (Black) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with Chemotherapy drugs and Fentanyl using ASTM D6978-05. Chemotherapy Drugs Minimum Breakthrough Detection Time Carmustine (3.3mg/ml) 14.2 min Cisplatin (1mg/ml) >240 min Cyclophosphamide (20mg/ml) >240 min Dacarbazine (10.0 mg/ml) >240 min Doxorubicin HCl (2 mg/ml) >240 min Etoposide (20mg/ml) >240 min Fluorouracil (50mg/ml) >240 min Paclitaxel (6mg/ml) >240 min Thiotepa (10mg/ml) 14.7 min Fentanyl Concentration Minimum Breakthrough Detection Time Fentanyl Citrate Injection, 100mcg/2ml >240 min *Please note that the following drugs have extremely low permeation times: Carmustine: 14.2 minutes, Thiotepa:14.7 minutes Warning: Do not use with Carmustine and Thiotepa.

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K243792

Device Name

Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate (White)

Indications for Use (Describe) Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate (White) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with Chemotherapy drugs and Fentanyl using ASTM D6978-05. Chem Drug Concentration Minimum Breakthrough Detection Time Carmustine (3.3mg/ml) 16.7 min Cisplatin (1mg/ml) >240 min Cyclophosphamide (20mg/ml) >240 min Dacarbazine (10.0 mg/ml) >240 min Doxorubicin HCl (2 mg/ml) >240 min Etoposide (20mg/ml) >240 min Fluorouracil (50mg/ml) >240 min Paclitaxel (6mg/ml) >240 min Thiotepa (10mg/ml) 16.2 min Fentanyl Concentration Minimum Breakthrough Detection Time Fentanyl Citrate Injection, 100mcg/2ml >240 min *Please note that the following drugs have extremely low permeation times: Carmustine: 16.7 minutes, Thiotepa:16.2 minutes Warning: Do not use with Carmustine and Thio Tepa.

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

5300 Concours Ontario, CA 91764 K243792

510(K) Summary

1. Submission Information

Date Prepared: February 18, 2025 Submission Format: Traditional 510(k) 510(K)#: K243792

2. Submitter Information

Applicant Name: Basic Medical Technology Inc. Address: 5300 Concours Ontario, CA 91764 Contact Person: John Zhao Tel: (909) 980-1678 Email: johnzhao@basicmedical.com

3. Device Information

Trade Name: Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate (Black, White)

Common Name: Nitrile Powder-Free Exam Gloves Classification Name: Non-Powdered Patient Examination Glove Regulation: 21 CFR 880.6250 Product Code: LZA, LZC, QDO, OPJ Classification Panel: General Hospital Device Class: Class I

4. Predicate Device Information

Shanxi Hongjin Plastic Technology Co., Ltd.

Powder Free Nitrile Examination Gloves (Black), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (K230779)

Product code: LZA, LZC, QDO, OPJ.

5. Device Description:

Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentany] Citrate (Black, White) is a Class I patient examination glove and specialty chemotherapy gloves, that made from synthetic nitrile latex. They are non-sterile, powder-free, fingertip textured, ambidextrous with beaded cuff, and single use only, and come in different sizes- XS, S, M, L, XL and XXL.

The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs and Fentanyl Citrate.

1 / 6

{6}------------------------------------------------

5300 Concours Ontario, CA 91764 K243792

6. Indications for Use:

Black glove:

Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate (Black) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with Chemotherapy drugs and Fentanyl using ASTM D6978-05.

Chemotherapy DrugsMinimum Breakthrough Detection Time
Carmustine (3.3mg/ml)14.2 min
Cisplatin (1mg/ml)>240 min
Cyclophosphamide (20mg/ml)>240 min
Dacarbazine (10.0 mg/ml)>240 min
Doxorubicin HCl (2 mg/ml)>240 min
Etoposide (20mg/ml)>240 min
Fluorouracil (50mg/ml)>240 min
Paclitaxel (6mg/ml)>240 min
Thiotepa (10mg/ml)14.7 min

Fentanyl Concentration Minimum Breakthrough Detection Time Fentanyl Citrate Injection, 100mcg/2ml >240 min *Please note that the following drugs have extremely low permeation times: Carmustine: 14.2 minutes, Thiotepa:14.7 minutes Warning: Do not use with Carmustine and Thiotepa.

White glove:

Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate (White) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with Chemotherapy drugs and Fentanyl using ASTM D6978-05.

Chem Drug ConcentrationMinimum Breakthrough Detection Time
Carmustine (3.3mg/ml)16.7 min
Cisplatin (1mg/ml)>240 min
Cyclophosphamide (20mg/ml)>240 min
Dacarbazine (10.0 mg/ml)>240 min
Doxorubicin HCl (2 mg/ml)>240 min
Etoposide (20mg/ml)>240 min
Fluorouracil (50mg/ml)>240 min
Paclitaxel (6mg/ml)>240 min
Thiotepa (10mg/ml)16.2 min

Fentanyl Concentration Minimum Breakthrough Detection Time Fentanyl Citrate Injection, 100mcg/2ml >240 min *Please note that the following drugs have extremely low permeation times: Carmustine: 16.7 minutes, Thiotepa:16.2 minutes Warning: Do not use with Carmustine and Thiotepa.

{7}------------------------------------------------

5300 Concours Ontario, CA 91764 K243792

7. Comparison of Subject Device and Predicate Device:

General Comparison Table:

Subject DevicePredicate DeviceK230779Result
Trade NameNitrile Powder-Free Exam GlovesTested for Use with ChemotherapyDrugs, Opioid Fentanyl Citrate (Black,White)Powder Free Nitrile ExaminationGloves (Black), Tested for Use withChemotherapy Drugs and FentanylCitrate-
Product CodeLZA, LZC, QDO, OPJLZA, LZC, QDO, OPJSame
Regulation Number21 CFR 880.625021 CFR 880.6250Same
ClassClass IClass ISame
Indications foruse(Chemotherapydrug permeationtesting resultscompared below)Nitrile Powder-Free Exam GlovesTested for Use with ChemotherapyDrugs, Opioid Fentanyl Citrate(Black/White) is a disposable deviceintended for medical purposes that isworn on the examiner's hand toprevent contamination betweenpatient and examiner.Gloves have been tested for use withChemotherapy drugs and Fentanylusing ASTM D6978-05.Powder Free Nitrile ExaminationGloves (Black), Tested for Use withChemotherapy Drugs and FentanylCitrate is a disposable device intendedfor medical purposes that is worn on theexaminer's hand to preventcontamination between patient andexaminer.Gloves have been tested for use withchemotherapy drugs and FentanylCitrate using ASTM D6978Similar
Powder freeYesYesSame
MaterialNitrileNitrileSame
ColorBlack, WhiteBlackDifferent
SizeXS, S, M, L, XL, XXLXS, S, M, L, XL, XXLSame
SterileNon-SterileNon-SterileSame
Single UseSingle useSingle useSame
Design featureAmbidextrousAmbidextrousSame
Chemo DrugsClaimSee below comparison tableSee below comparison table/

{8}------------------------------------------------

5300 Concours Ontario, CA 91764

K243792

Analysis: * The finished subject device has been tested with performance and Biocompatibility, all the test results meet the requirements, so the difference of color does not raise questions of safety and effectiveness.

CharacteristicsSubject DevicePredicate DeviceK230779Result
DimensionASTM D6319-19LengthMin 230mmMin 230mmSame
WidthXS: 70±10 mmS: 80±10 mmM: 95±10 mmL: 110±10 mmXL: 120±10 mmXXL: 130±10 mmXS: 70±10 mmS: 80±10 mmM: 95±10 mmL: 110±10 mmXL: 120±10 mmXXL: 130±10 mmSame
ThicknessPalm: Min 0.05 mmFinger: Min 0.05 mmPalm: Min 0.05 mmFinger: Min 0.05 mmSame
PhysicalpropertiesASTM D6319-19ASTM D412-16Before agingTensile strengthMin 14MPaMin 14MPaSame
Ultimate elongationMin 500%Min 500%
After agingTensile strengthMin 14MPaMin 14MPaSame
Ultimate elongationMin 400%Min 400%
Freedom from holesASTM D6319-19ASTM D5151-19G-I, AQL2.5G-I, AQL2.5Same
Residual PowderASTM D6319-19ASTM D6124-06≤ 2 mg per glove≤ 2 mg per gloveSame
BiocompatibilityIrritationISO 10993-23Under the conditions ofthe study, not an irritantUnder the conditions of thestudy, not an irritant
SensitizationISO 10993-10Under the conditions ofthe study, not asensitizerUnder the conditions of thestudy, not a sensitizerSame
Acute SystemicToxicity TestISO 10993 -11Under the conditions ofthis study, there was noevidence of acutesystemic toxicityUnder the conditions of thisstudy, there was no evidenceof acute systemic toxicity.Same

Technological Characteristic Comparison Table:

{9}------------------------------------------------

5300 Concours Ontario, CA 91764 K243792

Chemotherapy Drugs Comparison Claim:

NO.Chemotherapy DrugsMinimum Breakthrough Detection Time, minutesResult ofcomparison
BlackWhitePredicate DeviceK230779
1Carmustine (3.3mg/ml)14.216.711.3Different
2Cisplatin (1mg/ml)>240>240>240Same
3Cyclophosphamide (20mg/ml)>240>240>240Same
4Dacarbazine (10.0 mg/ml)>240>240>240Same
5Doxorubicin HCl (2 mg/ml)>240>240>240Same
6Etoposide (20mg/ml)>240>240>240Same
7Fluorouracil (50mg/ml)>240>240>240Same
8Paclitaxel (6mg/ml)>240>240>240Same
9Thiotepa (10mg/ml)14.716.224.7Different
NO.Fentanyl CitrateMinimum Breakthrough Detection Time, minutesResult ofcomparison
BlackWhitePredicate deviceK230779
1Fentanyl Citrate Injection, 100mcg/2ml>240>240>240Same
  • Chemotherapy drugs and the minimum breakthrough time of subject device will be listed on labeling, so this different does not raise questions of safety and effectiveness.

8. Summary of Non-Clinical Performance Data

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:

Test performedMethodologyAcceptance CriteriaResult
Freedom From HolesASTM D6319-19ASTM D5151-19Meet requirement inspection level G- 1, AQL 2.5(ISO2859-1)Pass
Dimension-LengthASTM D6319-19Minimum 230mm for all the sizePass
Dimension-WidthASTM D6319-19XS: 70±10mmS: 80±10mmM: 95±10mmL: 110±10mmXL: 120±10mmXXL: 130±10mmPass
Dimension-ThicknessASTM D6319-19Finger: 0.05mm (Min)Palm: 0.05mm (Min)Pass

{10}------------------------------------------------

5300 Concours Ontario, CA 91764 K243792

Physical propertiesASTM D6319-19ASTM D412-16Tensile Strength (Min14 Mpa)Elongation (Before Aging 500% and after aging400%) MinPass
Powder ResidueASTM D6319-19ASTM D6124-06Not more than 2 mg per glovePass
Permeation byChemotherapy Drugsand Fentanyl CitrateASTM D6978-05Refer the above tablePass
Skin IrritationISO10993-23:2021Not a primary skin irritantPass
Skin SensitizationISO10993-10:2021Not a contact sensitizerPass
Acute SystemicToxicityISO 10993-11:2017No acute systemic toxicityPass
  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
  • ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
  • ASTM D6124-06 (2017), Standard Test Method for Residual Powder on Medical Gloves
  • ASTM D412-16 (2021), Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— Tension
  • · ASTM D6978-05(2023), Assessment of Reissuance of Medical Gloves to Permeation by Chemotherapy Drugs.
  • · ISO 10993-10: 2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization.
  • · ISO10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritation.
  • · ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity

9. Summary of Clinical Testing:

Not applicable.

10. Conclusion:

The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K230779.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.