K Number
K243756

Validate with FDA (Live)

Date Cleared
2025-07-17

(223 days)

Product Code
Regulation Number
870.1330
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 0.014" Willow Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.

Device Description

Like the predicate device, the 0.014" Willow Guidewire is a single-use product with a shapeable tip available in straight and pre-shaped configurations, used to gain intravascular access to facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for neuro and peripheral diagnostic and interventional procedures. The wire can be torqued to facilitate navigation through the vasculature. The 0.014" Willow Guidewire comes in three stiffness profiles: Soft, Standard and Support.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the 0.014" Willow Guidewire do not describe a study involving an AI/Machine Learning device.

Instead, this document describes a traditional medical device (a guidewire) and its clearance process through substantial equivalence to a legally marketed predicate device. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to engineering and biocompatibility bench testing, not clinical performance studies for an AI algorithm.

Therefore, many of the questions related to AI/ML device performance (like sample size for test/training sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, effect size of AI assistance) are not applicable to this document.

However, I can extract the relevant information regarding the acceptance criteria and the engineering study (bench performance testing) for the 0.014" Willow Guidewire.


Acceptance Criteria and Study for the 0.014" Willow Guidewire

The 0.014" Willow Guidewire is a traditional medical device, not an AI/ML device. Therefore, the "acceptance criteria" and "study" refer to engineering specifications and bench performance testing, along with biocompatibility testing, to demonstrate substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by "Acceptance criteria were met" or "considered acceptable" for each test. The reported device performance is that these criteria were successfully achieved.

TestTest Method SummaryReported Device Performance / Conclusions
Bench Performance Testing
Coating LubricityHydrophilic coating lubricity was assessed after multiple pull cycles through silicone pads.Acceptance criteria were met.
Coating DurabilityCoating durability was assessed after repeating multiple pull cycles through silicone pads.Acceptance criteria were met.
Coating IntegrityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Inspected pre- and post-simulated use tracking with comparison to predicate.Coating integrity was reported and considered acceptable.
Corrosion ResistancePer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019).Acceptance criteria were met.
Dimensional & Visual Insp.Per FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Verified key dimensions.Acceptance criteria were met.
Flex FatigueSubjected guidewire to multiple flexure cycles around cylindrical pins.Acceptance criteria were met.
FractureSubjected guidewire to multiple wrappings around cylinder and visually inspected for signs of fracture.Acceptance criteria were met.
Kink ResistancePer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Wrapped guidewire around test fixture with clinically relevant radii.Acceptance criteria were met.
Particulate CharacterizationPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Counted particulates after tracking through tortuous simulated use model, with comparison to predicate.Particulate counts were reported and considered acceptable.
RadiopacityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Evaluated subject and predicate guidewires under fluoroscopy.All Willow Guidewires demonstrated acceptable radiopacity; comparable to predicate.
Simulated UsePer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Tested with microcatheter, introducer, and torque device while navigating to target in a tortuous simulated use model.Acceptance criteria were met.
Tensile StrengthPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Measured force to break at each bond.Acceptance criteria were met.
Tip FlexibilityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Measured force to deflect tip at 5 mm, 10 mm, and 20 mm gauge lengths, compared to predicate.Acceptance criteria were met.
Tip ShapeabilityShaped the guidewire tip three times per labeling.Acceptance criteria were met.
Tip Shape RetentionMeasured tip shape retention after tracking through a tortuous simulated use model.Acceptance criteria were met.
TorqueabilityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Tracked through tortuous simulated use model; evaluated steerability and torque response.Acceptance criteria were met.
Torque StrengthPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Recorded rotations until failure.Acceptance criteria were met.
Biocompatibility Testing
Acute Systemic ToxicityPer ISO 10993-11.Pass: No evidence of acute systemic toxicity.
Compliment ActivationPer ISO 10993-4.Pass: Non-activator.
Cytotoxicity – MEM ElutionPer ISO 10993-5.Pass: Non-cytotoxic.
HemolysisPer ISO 10993-4.Pass: Non-hemolytic.
Intracutaneous ReactivityPer ISO 10993-10.Pass: Non-reactive.
In Vivo ThrombogenicityPer ISO 10993-4.Pass: Non-Thrombogenic.
PyrogenicityPer ISO 10993-11.Pass: Non-pyrogenic.
SensitizationPer ISO 10993-10.Pass: Non-sensitizing.

Information Not Applicable (NA) to this Device/Submission:

The following questions are specifically targeted at AI/ML medical devices and are not relevant to the 0.014" Willow Guidewire, which is a traditional physical medical device.

  1. Sample size used for the test set and the data provenance: NA (This is a physical device, not an AI/ML algorithm requiring a data test set.)
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: NA (No "ground truth" in the AI/ML sense is established for this type of device.)
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: NA (No adjudication method described for physical device testing.)
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: NA (No MRMC study or AI assistance involved.)
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: NA (Not an algorithm.)
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): NA (Ground truth is not a concept applied in this context; performance is measured against engineering specifications and predicate characteristics.)
  7. The sample size for the training set: NA (No training set for a physical device.)
  8. How the ground truth for the training set was established: NA (No training set or ground truth in this context.)

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

July 17, 2025

Arbor Endovascular, LLC
℅ Sara Toyloy
President and Principal Consultant
Fabrica Consulting, LLC
73 Lincoln Drive
Sausalito, California 94965

Re: K243756
Trade/Device Name: 0.014" Willow Guidewire
Regulation Number: 21 CFR 870.1330
Regulation Name: Catheter Guide Wire
Regulatory Class: Class II
Product Code: MOF, DQX
Dated: June 13, 2025
Received: June 13, 2025

Dear Sara Toyloy:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243756 - Sara Toyloy Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243756 - Sara Toyloy Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan -S

Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243756

Device Name: 0.014" Willow Guidewire

Indications for Use (Describe)

The 0.014" Willow Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.


This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary K243756

Submitter Name, Address and Contact:

Submitter: Arbor Endovascular, LLC
2345 Bering Drive
San Jose, CA 95131

Sponsor Contact: Kim Otto
Vice President, Operations
Phone: +1 (650) 576-9090
Email: kotto@arborendo.com

Correspondent: Sara Toyloy
Fabrica Consulting, LLC
Email: sara@fabricaconsulting.com
Phone: +1 (415) 297-8826

Date Prepared: July 8, 2025

Device Name and Classification:

Trade/Proprietary Name: 0.014" Willow™ Guidewire
Common Name: Wire, Guide, Catheter
Classification Name: Wire, Guide, Catheter, 21 CFR 870.1330 – Class II
Product Code: MOF, DQX
Legally Marketed Predicate Device: Synchro SELECT Guidewire (K202522)

Device Description:

Like the predicate device, the 0.014" Willow Guidewire is a single-use product with a shapeable tip available in straight and pre-shaped configurations, used to gain intravascular access to facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for neuro and peripheral diagnostic and interventional procedures. The wire can be torqued to facilitate navigation through the vasculature. The 0.014" Willow Guidewire comes in three stiffness profiles: Soft, Standard and Support.

Page 6

Accessories:

The 0.014" Willow Guidewire is provided with three (3) accessories, an Introducer, a Torque Device and a Shaping Mandrel.

Indications for Use:

The Indications for Use for the 0.014" Willow Guidewire are the same as the predicate Synchro SELECT Guidewire and are as follows:

The 0.014" Willow Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.

Technological Characteristics and Product Feature Comparison:

The 0.014" Willow Guidewire is substantially equivalent to the predicate device, the Synchro SELECT Guidewire (K202522), based on the following:

  • Same indications for use
  • Similar fundamental materials and manufacturing process
  • Same fundamental design and technology
  • Same operating principles
  • Similar materials and processes for packaging
  • Similar sterilization method and process for devices

A comparison of the subject device with the predicate device is summarized in Table 1 below.

Table 1. Comparison of the 0.014" Willow Guidewire and the Synchro SELECT Guidewire

CharacteristicPredicate Device Synchro SELECT™ Guidewire (K202522)Subject Device 0.014" Willow Guidewire (K243756)
Classification NameWire, Guide, Catheter, 21 CFR 870.1330, Class IISame
Product CodeMOF, DQXSame
Review PanelNeurologySame
510(k) SubmitterStryker Neurovascular 47900 Bayside Pkwy Fremont, CA 94538-6515 (USA)Arbor Endovascular, LLC 2345 Bering Drive San Jose, CA 95131

Page 7

CharacteristicPredicate Device Synchro SELECT™ Guidewire (K202522)Subject Device 0.014" Willow Guidewire (K243756)
Indications for UseThe Synchro SELECT Guidewire series is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.The 0.014" Willow Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.
Device Description/ Principle of OperationThe Synchro SELECT Guidewire is a single-use product with a shapeable tip available in straight and pre-shaped configurations, used to gain intravascular access to facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for neuro and peripheral diagnostic and interventional procedures. The wire can be torqued to facilitate navigation through the vasculature.Same
Target PopulationThe Synchro SELECT Guidewire is used in patients undergoing endovascular treatment including neuro and peripheral vasculatures.Same
AccessoriesGuidewire Introducer and Torque DeviceGuidewire Introducer, Torque Device, and Shaping Mandrel
Core Wire304 Stainless Steel with PTFE coating on the proximal sectionCobalt chromium alloy with PTFE coating on the proximal section
Core Wire Length215 cm Access Length 300 cm Exchange Length215 cm Access Length 315 cm Exchange Length
Guidewire Tip (slotted tube)Nickel-Titanium Micro-Machined NitinolNickel-Titanium Laser Cut Nitinol
Radiopaque CoilPlatinumPlatinum-Tungsten Alloy
Core Wire Length215 cm Access Length 300 cm Exchange Length215 cm Access Length 315 cm Exchange Length
Radiopaque Coil Shapes and LengthsStandard and Soft: Platinum, 10 cm Pre-shaped Support: Platinum, 7 cm Straight Support: Platinum, 6 cmStandard and Soft Straight and Pre-Shaped: Platinum/Tungsten, 10 cm Support Straight and Standard: Platinum/Tungsten, 3 cm
AdhesiveUV Curable AdhesiveSame
PrimerParylene DimerSame
Hydrophilic CoatingProprietary Hydrophilic Topcoat and BasecoatProprietary Hydrophilic Coating
Guidewire IntroducerCommercially available per K760389, B. BraunGuidewire Introducer
Shaping MandrelN/AShaping Mandrel

Page 8

CharacteristicPredicate Device Synchro SELECT™ Guidewire (K202522)Subject Device 0.014" Willow Guidewire (K243756)
Torque DeviceCommercially available per K936032, Merit MedicalTorque Device
Dispenser HoopHigh Density PolyethyleneSame
Accessory CardClay Coated Solid Bleached Sulfate (CCSBS)High Density Polyethylene (HDPE)
Sterile PouchTyvek® - PolyethyleneTyvek®/Nylon - Polyethylene
Shipping CartonSolid Bleached Sulphate (SBS)Folding Box Board (FBB)
Sterilization Method100% Ethylene OxideSame
How SuppliedSingle Use/SterileSame

The differences between the devices do not raise new questions of safety and effectiveness.

Risk Assessment:

Risk assessment of the 0.014" Willow Guidewire has been conducted in accordance with ISO 14971:2019 to show that no new risks were identified compared to the well-established commercially available predicate device. Results of non-clinical testing have demonstrated the subject device is substantially equivalent to the predicate device.

Bench Performance Testing:

The results of the bench testing conducted on the subject device demonstrate that it performs as intended and meets design specifications. Bench performance testing included the following:

Table 2: Bench Testing Summary

TestTest Method SummaryConclusions
Coating LubricityHydrophilic coating lubricity was assessed after multiple pull cycles through silicone pads.Acceptance criteria were met.
Coating DurabilityCoating durability was assessed after repeating multiple pull cycles through silicone pads.Acceptance criteria were met.
Coating IntegrityTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Coating integrity inspected pre- and post-simulated use tracking with comparison to predicate.Coating integrity was reported and considered acceptable.

Page 9

TestTest Method SummaryConclusions
Corrosion ResistanceTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019).Acceptance criteria were met.
Dimensional and Visual InspectionTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Verified key dimensions of the guidewire.Acceptance criteria were met.
Flex FatigueSubjected the guidewire to multiple flexure cycles around cylindrical pins.Acceptance criteria were met.
FractureSubjected the guidewire to multiple wrappings around cylinder and visually inspected for signs of fracture.Acceptance criteria were met.
Kink ResistanceTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Wrapped the guidewire around test fixture with clinically relevant radii.Acceptance criteria were met.
Particulate CharacterizationTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Counted particulates of various size ranges after tracking through a tortuous simulated use model, with comparison to the predicate device.Particulate counts were reported and considered acceptable.
RadiopacityTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Subject and predicate guidewires were evaluated under fluoroscopy.All Willow Guidewires demonstrated acceptable radiopacity. The radiopacity of the subject device was comparable to that of the predicate device.
Simulated UseTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Guidewires were tested for use with a microcatheter, the guidewire introducer, and torque device while navigating to target locations in a tortuous simulated use model.Acceptance criteria were met.
Tensile StrengthTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Measured the force required to break at each bond, including distal tip.Acceptance criteria were met.

Page 10

TestTest Method SummaryConclusions
Tip FlexibilityTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Measured the force required to deflect the guidewire tip when held at 5 mm, 10 mm and 20 mm gauge lengths. Forces were compared to the predicate device.Acceptance criteria were met.
Tip ShapeabilityShaped the guidewire tip for a total of three (3) times per labeling.Acceptance criteria were met.
Tip Shape RetentionMeasured the tip shape retention after tracking the guidewire through a tortuous simulated use model.Acceptance criteria were met.
TorqueabilityTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Tracked the guidewire through a tortuous simulated use model and evaluated the steerability and torque response.Acceptance criteria were met.
Torque StrengthTesting completed per FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Recorded the number of times the proximal end of the wire can be rotated until it exceeds its maximum rotations and fails.Acceptance criteria were met.

Sterilization and Shelf-Life Testing:

The device and its accessories are sterilized by 100% Ethylene Oxide and have been adopted into a validated sterilization process in accordance with the principles of ISO 11135:2014 ("Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices"). A sterility assurance level (SAL) of 10⁻⁶ has been demonstrated. The performance testing completed supports a 1-year shelf-life.

Performance Data – Animal, Clinical:

No animal or clinical studies were conducted as the indications for use and the fundamental scientific technology are the same as that of the predicate. Substantial equivalence of the subject device has been established to the predicate device through the results of non-clinical performance testing.

Biocompatibility Testing:

Biocompatibility testing was conducted on the 0.014" Willow Guidewire in accordance with FDA guidance document, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"," and

Page 11

demonstrates that the guidewire and accessories meet biological safety requirements for externally communicating medical devices with circulating blood contact for less than 24 hours.

This statement is supported by the standards assessment and the testing conducted by Arbor Endovascular.

Table 3: Summary of Biocompatibility Testing

TestTest Method SummaryConclusion
Acute Systemic ToxicityPer ISO 10993-11Pass: No evidence of acute systemic toxicity.
Compliment ActivationPer ISO 10993-4Pass: Non-activator
Cytotoxicity – MEM ElutionPer ISO 10993-5Pass: Non-cytotoxic
HemolysisPer ISO 10993-4Pass: Non-hemolytic
Intracutaneous ReactivityPer ISO 10993-10Pass: Non-reactive
In Vivo ThrombogenicityPer ISO 10993-4Pass: Non-Thrombogenic
PyrogenicityPer ISO 10993-11Pass: Non-pyrogenic
SensitizationPer ISO 10993-10Pass: Non-sensitizing

Conclusion:

The subject device is substantially equivalent to the predicate device with regards to device design, materials, intended use, and patient population. The conclusions drawn from the risk assessments and non-clinical testing conducted demonstrate that the subject device performs as intended and is substantially equivalent to the predicate device.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.