K Number
K143653

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-09-04

(255 days)

Product Code
Regulation Number
866.3980
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Verigene® Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) is a multiplexed qualitative test intended for the simultaneous detection and identification of multiple viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infection. The test is performed on the automated Verigene System utilizing reverse transcription (RT), polymerase chain reaction (PCR), and microarray hybridization to detect gene sequences of the following organism types and subtypes:

VirusesBacteria
AdenovirusBordetella parapertussis/bronchiseptica
Human MetapneumovirusBordetella holmesii
Influenza ABordetella pertussis
Influenza A (Subtype H1)
Influenza A (Subtype H3)
Influenza B
Parainfluenza 1
Parainfluenza 2
Parainfluenza 3
Parainfluenza 4
Respiratory Syncytial Virus A
Respiratory Syncytial Virus B
Rhinovirus

Detecting and identifying specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory infection with other clinical and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or patient management decisions.

Negative results in the presence of a respiratory illness do not preclude respiratory infection and may be due to infection with pathogens that are not detected by this test or lower respiratory tract infection that is not detected by an NPS specimen. Conversely, positive results do not rule-out infection with organisms not detected by RP Flex. The agent(s) detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation may be necessary to establish a final diagnosis of respiratory infection.

Clinical evaluation indicates a lower sensitivity specific to RP Flex for the detection of Rhinovirus. If infection with Rhinovirus is suspected, negative samples should be confirmed using an alternative method.

Performance characteristics for Influenza A were established when Influenza A/H1 (2009 Pandemic) and A/H3 were the predominant Influenza A viruses in circulation. RP Flex may not detect novel Influenza A strains. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection used specifically for novel virulent influenza viruses and sent to appropriate health authorities for testing. Viral culture should not be attempted in these cases unless a biosafety level (BSL) 3+ facility is available to receive and culture specimens.

Device Description

The Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) is a molecular assay that relies on detection of specific nucleic acid targets in a microarray format. For each of the bacterial or viral nucleic acid sequences detected by RP Flex, unique Capture and Mediator oligonucleotides are used, with gold nanoparticle probe-based endpoint detection. The Capture oligonucleotides are covalently bound to the microarray substrate and hybridize to a specific portion of the nucleic acid targets. The Mediator oligonucleotides have a region that binds to a different portion of the same nucleic acid targets and also have a sequence that allows binding of a gold nanoparticle probe. Specific silver enhancement of the bound gold nanoparticle probes at the capture sites results in gold-silver aggregates that scatter light with high efficiency and provide accurate detection of target capture.

The RP Flex test is performed on the Verigene System, a "sample-to-result," fully automated, bench-top molecular diagnostics workstation. The System enables automated nucleic acid extraction from nasopharyngeal swabs (NPS) and detection of analyte-specific target nucleic acids. The Verigene System consists of two components: the Verigene Reader and the Verigene Processor SP.

The Reader is the Verigene System's user interface and serves as the central control unit for all aspects of test processing, automated imaging, and result generation using a touch-screen control panel and a barcode scanner. The Verigene Processor SP executes the test procedure, automating the steps of (1) Sample Preparation and Target Amplification – cell lysis and magnetic bead-based bacterial and viral nucleic acid isolation and amplification, and (2) Hybridization- detection and identification of analyte-specific nucleic acid in a microarray format by using gold nanoparticle probe-based technology. Once the specimen is loaded by the operator, all other fluid transfer steps are performed by an automated pipette that transfers reagents between wells of the trays and finally loads the specimen into the Test Cartridge for hybridization. Single-use disposable test consumables and a self-contained Verigene Test Cartridge are used for each sample tested with the RP Flex assay.

To obtain the test results after test processing is complete, the user removes the Test Cartridge from the Processor SP, and inserts the substrate holder into the Verigene Reader for analysis. Light scatter from the capture spots is imaged by the Verigene Reader and intensities from the microarray spots are used to make a determination regarding the presence (Detected) or absence (Not Detected) of a targeted nucleic acid sequence/analyte. This determination is made by means of software-based decision algorithm resident in the Verigene Reader.

AI/ML Overview
{
  "1. A table of acceptance criteria and the reported device performance": {
    "Influenza A": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "98.3% (58/59) (91.0-99.7)",
        "Negative Percent Agreement (95% CI)": "99.4% (2121/2134) (99.0-99.6)"
      }
    },
    "Influenza A subtype H1": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "97.8% (45/46) (88.7-99.6)",
        "Negative Percent Agreement (95% CI)": "99.7% (2138/2144) (99.4-99.9)"
      }
    },
    "Influenza A subtype H3": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "100% (13/13) (77.2-100)",
        "Negative Percent Agreement (95% CI)": "99.8% (2173/2177) (99.5-99.9)"
      }
    },
    "Influenza B": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "98.0% (49/50) (89.5-99.6)",
        "Negative Percent Agreement (95% CI)": "99.6% (2139/2147) (99.3-99.8)"
      }
    },
    "RSV A": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "100% (17/17) (81.6-100)",
        "Negative Percent Agreement (95% CI)": "99.9% (2150/2153) (99.6-99.9)"
      }
    },
    "RSV B": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "100% (173/173) (97.8-100)",
        "Negative Percent Agreement (95% CI)": "98.8% (1973/1997) (98.2-99.2)"
      }
    },
    "Parainfluenza 1": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "90.0% (27/30) (74.4-96.5)",
        "Negative Percent Agreement (95% CI)": "99.9% (2165/2167) (99.7-100)"
      }
    },
    "Parainfluenza 2": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "92.3% (12/13) (66.7-98.6)",
        "Negative Percent Agreement (95% CI)": "99.9% (2181/2184) (99.6-99.9)"
      }
    },
    "Parainfluenza 3": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "82.4% (14/17) (59.0-93.8)",
        "Negative Percent Agreement (95% CI)": "99.9% (2177/2180) (99.6-99.9)"
      }
    },
    "Parainfluenza 4": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "79.2% (19/24) (59.3-90.8)",
        "Negative Percent Agreement (95% CI)": "99.8% (2169/2173) (99.5-99.9)"
      }
    },
    "Adenovirus": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "86.0% (49/57) (74.7-92.7)",
        "Negative Percent Agreement (95% CI)": "97.2% (2081/2140) (96.5-97.9)"
      }
    },
    "Human Metapneumovirus (hMPV)": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "100% (46/46) (92.3-100)",
        "Negative Percent Agreement (95% CI)": "99.7% (2145/2151) (99.4-99.9)"
      }
    },
    "Rhinovirus": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "81.9% (407/497) (78.3-85.0)",
        "Negative Percent Agreement (95% CI)": "97.1% (1578/1625) (96.2-97.8)"
      }
    },
    "Bordetella parapertussis/bronchiseptica": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "100% (2/2) (34.2-100)",
        "Negative Percent Agreement (95% CI)": "99.9% (2290/2291) (99.8-100)"
      }
    },
    "Bordetella pertussis": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "100% (8/8) (67.6-100)",
        "Negative Percent Agreement (95% CI)": "99.9% (2187/2189) (99.7-100)"
      }
    },
    "Bordetella holmesii": {
      "Performance Criteria": "Positive Agreement ≥ 90%, Negative Agreement ≥ 95%",
      "Reported Performance (All specimens)": {
        "Positive Percent Agreement (95% CI)": "100% (1/1) (20.6-100)",
        "Negative Percent Agreement (95% CI)": "100% (2305/2305) (99.8-100)"
      }
    }
  },
  "2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)": {
    "Total specimens enrolled": 3299,
    "Specimens included in performance analysis": 3266,
    "Specimen types and provenance": {
      "Prospectively-collected fresh specimens": {
        "Count": 1069,
        "Type": "Prospective"
      },
      "Prospectively-collected frozen specimens": {
        "Count": 1317,
        "Type": "Prospective"
      },
      "Retrospectively-collected frozen specimens": {
        "Count": 520,
        "Type": "Retrospective"
      },
      "Contrived frozen specimens": {
        "Count": 360,
        "Type": "Contrived"
      }
    },
    "Country of Origin": "Not explicitly stated, but implies U.S. due to FDA submission."
  },
  "3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)": "Not specified in the provided text. The ground truth was established by comparing to an FDA-cleared molecular respiratory panel and/or PCR amplification followed by confirmatory bi-directional sequencing.",
  "4. Adjudication method (e.g. 2+1, 3+1, none) for the test set": "The method of adjudication for discrepancies between the predicate and PCR/sequencing results is not explicitly detailed. The comparison was made against a 'composite of an FDA-cleared molecular respiratory panel and analytically validated PCR with bi-directional sequencing', implying a reference standard was used for ground truth, but the steps for resolving conflicting results among these methods are not described.",
  "5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance": "Not applicable. This is a standalone diagnostic device, not an AI-assisted human reader study.",
  "6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done": "Yes, a standalone performance study was conducted. The clinical performance characteristics were determined by comparing the device's results to a composite reference method (FDA-cleared molecular respiratory panel and/or PCR amplification with bi-directional sequencing). The device is described as an 'automated Verigene System' which implies a standalone algorithm making the detection calls.",
  "7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)": "A composite reference standard was used as ground truth: an FDA-cleared molecular respiratory panel and/or PCR amplification followed by confirmatory bi-directional sequencing.",
  "8. The sample size for the training set": "The document does not explicitly state a 'training set' size for a machine-learning model. This is a molecular diagnostic assay, and its development would typically involve analytical testing (LoD, inclusivity, exclusivity, etc.) rather than a 'training set' in the AI sense. The analytical inclusivity study used a comprehensive panel of 108 strains, and additional 28 strains from the LoD study, making a total of 136 strains for analytical inclusivity tests.",
  "9. How the ground truth for the training set was established": "For analytical studies (LoD, inclusivity, exclusivity), the 'ground truth' was based on known concentrations of characterized viral and bacterial strains (TCID50/mL or CFU/mL) in simulated NPS. For inclusivity, identification of the strains was confirmed from their source (e.g., ATCC, Zeptometrix, IRR) and in some cases, further confirmed by quantitative TaqMan real-time PCR or PCR/bi-directional sequencing."
}

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 4, 2015

Nanosphere, Inc. c/o Fran White MDC Associates, LLC. 180 Cabot Street Beverly, MA 01915

Re: K143653

Trade/Device Name: Verigene® Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay Regulatory Class: II Product Code: OCC, OEM, OEP, OOU, OZE, OZZ, OOI Dated: July 27, 2015 Received: July 28, 2015

Dear Ms. White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tamara V. Feldblyum -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143653

Device Name

Verigene® Respiratory Pathogens Flex Nucleic Acid Test (RP Flex)

Indications for Use (Describe)

The Verigene® Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) is a multiplexed qualitative test intended for the simultaneous detection and identification of multiple viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infection. The test is performed on the automated Verigene System utilizing reverse transcription (RT), polymerase chain reaction (PCR), and microarray hybridization to detect gene sequences of the following organism types and subtypes:

VirusesBacteria
AdenovirusBordetella parapertussis/bronchiseptica
Human MetapneumovirusBordetella holmesii
Influenza ABordetella pertussis
Influenza A (Subtype H1)
Influenza A (Subtype H3)
Influenza B
Parainfluenza 1
Parainfluenza 2
Parainfluenza 3
Parainfluenza 4
Respiratory Syncytial Virus A
Respiratory Syncytial Virus B
Rhinovirus

Detecting and identifying specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory infection with other clinical and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or patient management decisions.

Negative results in the presence of a respiratory illness do not preclude respiratory infection and may be due to infection with pathogens that are not detected by this test or lower respiratory tract infection that is not detected by an NPS specimen. Conversely, positive results do not rule-out infection with organisms not detected by RP Flex. The agent(s) detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation may be necessary to establish a final diagnosis of respiratory infection.

Clinical evaluation indicates a lower sensitivity specific to RP Flex for the detection of Rhinovirus. If infection with Rhinovirus is suspected, negative samples should be confirmed using an alternative method.

Performance characteristics for Influenza A were established when Influenza A/H1 (2009 Pandemic) and A/H3 were the predominant Influenza A viruses in circulation. RP Flex may not detect novel Influenza A strains. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection used specifically for novel virulent influenza viruses and sent to appropriate health authorities for testing. Viral culture should not be attempted in these cases unless a biosafety level (BSL) 3+ facility is available to receive and culture specimens.

{3}------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary 1.

510(k) Number:

Verigene® Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) K143653:

Summary Preparation Date:

August 18, 2015

Submitted by:

Nanosphere, Inc. 4088 Commercial Avenue Northbrook, IL 60062 Phone: 847-400-9000 Fax: 847-400-9199

Contact:

Fran White MDC Associates

Proprietary Names:

For the instrument: Verigene® System For the assay: Verigene® Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) Verigene® RP Flex

Common Names:

For the instrument:

Bench-top molecular diagnostics workstation

For the assay:

Respiratory Pathogens Nucleic Acid Test Respiratory Pathogens Flex Nucleic Acid Test Respiratory Pathogens identification and differentiation system Respiratory assay Respiratory test Verigene RP Flex RP Flex

{5}------------------------------------------------

Regulatory Information:

Regulation section:

866.3980 - Respiratory Viral Panel Multiplex Nucleic Acid Assay

Classification:

Class II

Panel:

Microbiology (83)

Product Code(s):

  • OCC Respiratory Virus Panel Nucleic Acid Assay System
  • Human Metapneumovirus (hMPV) RNA Assay System OEM
  • OEP Influenza A Virus Subtype Differentiation Nucleic Acid Assay
  • OOI Real Time Nucleic Acid Amplification System
  • OOU Parainfluenza Multiplex Nucleic Acid Assay
  • OZE Influenza A and Influenza B Multiplex Nucleic Acid Assay
  • Bordetella Pertussis DNA Assay System OZZ

Predicate Devices:

FilmArray Respiratory Panel (RP) System (K143080, K123620, K120267, K110764, and K103175) (BioFire Diagnostics, Inc.)

{6}------------------------------------------------

Intended Use:

The Verigene® Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) is a multiplexed qualitative test intended for the simultaneous detection and identification of multiple viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infection. The test is performed on the automated Verigene System utilizing reverse transcription (RT), polymerase chain reaction (PCR), and microarray hybridization to detect gene sequences of the following organism types and subtypes:

VirusesBacteria
AdenovirusBordetella parapertussis/bronchiseptica
Human MetapneumovirusBordetella holmesii
Influenza ABordetella pertussis
Influenza A (subtype H1)
Influenza A (subtype H3)
Influenza B
Parainfluenza 1
Parainfluenza 2
Parainfluenza 3
Parainfluenza 4
Respiratory Syncytial Virus A
Respiratory Syncytial Virus B
Rhinovirus

Detecting and identifying specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory infection, if used in conjunction with other clinical and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or patient management decisions.

Negative results in the presence of a respiratory illness do not preclude respiratory infection and may be due to infection with pathogens that are not detected by this test or lower respiratory tract infection that is not detected by an NPS specimen. Conversely, positive results do not rule-out infection or co-infection with organisms not detected by RP Flex. The agent(s) detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation may be necessary to establish a final diagnosis of respiratory infection.

Clinical evaluation indicates a lower sensitivity specific to RP Flex for the detection of Rhinovirus. If infection with Rhinovirus is suspected, negative samples should be confirmed using an alternative method.

Performance characteristics for Influenza A were established when Influenza A/H1 (2009 Pandemic) and A/H3 were the predominant Influenza A viruses in circulation. RP Flex may not detect novel Influenza A strains. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions used

{7}------------------------------------------------

specifically for novel virulent influenza viruses and sent to appropriate health authorities for testing. Viral culture should not be attempted in these cases unless a biosafety level (BSL) 3+ facility is available to receive and culture specimens.

Technological Characteristics:

The Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) is a molecular assay that relies on detection of specific nucleic acid targets in a microarray format. For each of the bacterial or viral nucleic acid sequences detected by RP Flex, unique Capture and Mediator oligonucleotides are used, with gold nanoparticle probe-based endpoint detection. The Capture oligonucleotides are covalently bound to the microarray substrate and hybridize to a specific portion of the nucleic acid targets. The Mediator oligonucleotides have a region that binds to a different portion of the same nucleic acid targets and also have a sequence that allows binding of a gold nanoparticle probe. Specific silver enhancement of the bound gold nanoparticle probes at the capture sites results in gold-silver aggregates that scatter light with high efficiency and provide accurate detection of target capture.

The RP Flex test is performed on the Verigene System, a "sample-to-result," fully automated, bench-top molecular diagnostics workstation. The System enables automated nucleic acid extraction from nasopharyngeal swabs (NPS) and detection of analyte-specific target nucleic acids. The Verigene System consists of two components: the Verigene Reader and the Verigene Processor SP.

The Reader is the Verigene System's user interface and serves as the central control unit for all aspects of test processing, automated imaging, and result generation using a touch-screen control panel and a barcode scanner. The Verigene Processor SP executes the test procedure, automating the steps of (1) Sample Preparation and Target Amplification – cell lysis and magnetic bead-based bacterial and viral nucleic acid isolation and amplification, and (2) Hybridization- detection and identification of analyte-specific nucleic acid in a microarray format by using gold nanoparticle probe-based technology. Once the specimen is loaded by the operator, all other fluid transfer steps are performed by an automated pipette that transfers reagents between wells of the trays and finally loads the specimen into the Test Cartridge for hybridization. Single-use disposable test consumables and a self-contained Verigene Test Cartridge are used for each sample tested with the RP Flex assay.

To obtain the test results after test processing is complete, the user removes the Test Cartridge from the Processor SP, and inserts the substrate holder into the Verigene Reader for analysis. Light scatter from the capture spots is imaged by the Verigene Reader and intensities from the microarray spots are used to make a determination regarding the presence (Detected) or absence (Not Detected) of a targeted nucleic acid sequence/analyte. This determination is made by means of software-based decision algorithm resident in the Verigene Reader.

{8}------------------------------------------------

Performance Data - Analytical Testing

Analytical Sensitivity / Limit of Detection (LoD)

Limit of Detection (LoD) of the Verigene RP Flex test was determined for twenty-eight (28) strains of respiratory pathogens, representing all sixteen (16) Verigene RP Flex reportable target analytes. The LoD was defined as the concentration at which the test produces a positive result greater than or equal to 95% of the time. Serial dilutions of the strains were tested and the initial tentative LoD confirmed with 20 replicates. To ensure the accuracy of the LoD determination, if the initial detection rate was 100%, an additional 20 replicates were performed at the next lower concentration until ≤95% was achieved. The confirmed LoDs for the twenty-eight (28) strains tested and the corresponding LoDs for the RP Flex test reportable targets are shown in the table below.

Viral Species andBacterial GenusViral Strains andBacterial SpeciesLoD
AdenovirusC (AdV-1)$1.2\times10^{1}$ TCID50/mL
AdenovirusB (AdV-3)$1.1\times10^{0}$ TCID50/mL
AdenovirusE (AdV-4)$4.1\times10^{-2}$ TCID50/mL
HumanMetapneumovirusMetapneumovirus 9 (A1)$3.0\times10^{1}$ TCID50/mL
HumanMetapneumovirusMetapneumovirus 27 (A2)$1.1\times10^{0}$ TCID50/mL
HumanMetapneumovirusMetapneumovirus 3 (B1)$1.0\times10^{1}$ TCID50/mL
HumanMetapneumovirusMetapneumovirus 8 (B2)$3.3\times10^{0}$ TCID50/mL
Influenza ABrisbane/59/2007 (H1N1)$3.0\times10^{1}$ TCID50/mL
Influenza ACalifornia/04/2009pdm09 (H1N1)$1.0\times10^{1}$ TCID50/mL
Influenza APort Chalmers/1/73 (H3N2)$3.3\times10^{0}$ TCID50/mL
Influenza AVictoria/361/2011 (H3N2)$3.7\times10^{1}$ TCID50/mL
Influenza AWisconsin/67/05 (H3N2)$3.3\times10^{0}$ TCID50/mL
Influenza BBrisbane/60/2008$1.2\times10^{1}$ TCID50/mL
Influenza BFlorida/02/2006$3.0\times10^{1}$ TCID50/mL
Influenza BMassachusetts/02/2012$1.2\times10^{1}$ TCID50/mL
ParainfluenzaParainfluenza 1$9.0\times10^{1}$ TCID50/mL
ParainfluenzaParainfluenza 2$1.0\times10^{1}$ TCID50/mL
ParainfluenzaParainfluenza 3$3.3\times10^{0}$ TCID50/mL
ParainfluenzaParainfluenza 4a$2.7\times10^{2}$ TCID50/mL
RhinovirusA (Rhinovirus 39)$1.0\times10^{1}$ TCID50/mL
RhinovirusB (Rhinovirus 14)$9.0\times10^{1}$ TCID50/mL
RhinovirusC (Rhinovirus C41)$2.4\times10^{3}$ PFU/mL
RespiratorySyncytial VirusRSV A (A2)$3.3\times10^{0}$ TCID50/mL
RespiratorySyncytial VirusRSV B (Wash/18537/62)$3.7\times10^{1}$ TCID50/mL
Bordetellaparapertussis$2.4\times10^{3}$ CFU/mL
Bordetellabronchiseptica$2.4\times10^{3}$ CFU/mL
Bordetellaholmesii$2.4\times10^{3}$ CFU/mL

Table 1: Limit of Detection (LoD)

CONFIDENTIAL AND PROPRIETARY Nanosphere, Inc.

{9}------------------------------------------------

Analytical Reactivity (Inclusivity)

The analytical reactivity (inclusivity) of the RP Flex test was demonstrated with a comprehensive panel of one-hundred and eight (108) strains representing temporal, evolutionary, and geographic diversity for each of the RP Flex panel organisms. Together with the twentyeight (28) strains evaluated as part of the Limit of Detection Study, a total of one-hundred and thirty-six (136) strains were evaluated for analytical inclusivity to RP Flex through empirical testing.

The organisms in the inclusivity panel were prepared in Simulated NPS. Thirteen (13) strains of Influenza A (subtypes H2N2, H2N3, H5N1, H5N3, H7N2, H7N7, H7N9, H9N2 & H10N7) were prepared and tested at a BSL 3 laboratory. Each sample was tested with the RP Flex in triplicate at an initial concentration 3-fold higher than the LoD determined for each analyte. In cases where the expected targets were not detected in one or more replicates, concentrations at a 3-fold higher level were evaluated.

RP Flex demonstrated analytical reactivity to all one-hundred and eight (108) strains tested, with some strains requiring higher titers for detection. The individual strains and concentrations at which positive test results were obtained for all three (3) replicates are presented by target organism in the tables below.

{10}------------------------------------------------

Adenovirus Inclusivity Results

AdenovirusSpeciesSerotypeStrain #SourceConcentration(TCID50/mL)Multiples ofLoD
A310810073CFZeptometrix$1.1\times10^0$1x
B17VR-7ATCC$3.3\times10^0$3x
B121VR-1099ATCC$3.3\times10^0$3x
B211VR-12ATCC$3.3\times10^0$3x
B2140810108CFZeptometrix$3.3\times10^0$3x
B234VR-716ATCC$3.3\times10^0$3x
B235VR-718ATCC$1.0\times10^1$9x
C2111010TriCore$3.3\times10^0$3x
C5*0810020CFZeptometrix$8.1\times10^2$729x
C6*0810111CFZeptometrix$2.7\times10^2$243x
D260810117CFZeptometrix$1.1\times10^0$1x
D370810119CFZeptometrix$1.1\times10^0$1x
F400810084CFZeptometrix$1.1\times10^0$1x
F410810085CFZeptometrix$1.1\times10^0$1x

Table 2: Adenovirus Inclusivity Results

  • Based on in silico analysis, the oligonucleotide identities of all the tested Adenovirus C subtypes have very similar ranges. Based on the investigation of viral stocks titers using a quantitative TagMan real-time PCR developed at Nanosphere that is specific for all Adenovirus species (note: the TaqMan assay are not the same primers used in the RP Flex), it appears that the amplifiable genome equivalents available in these two adenovirus viral stocks are significantly reduced comparing to that of the other adenovirus stocks tested in the study.

Influenza A Inclusivity Results

Table 3:Influenza A Inclusivity Results
-------------------------------------------
Influenza ASubtypeStrainSourceInfluenza AA/H1 or A/H3
Concentration(TCID50/mL)Multiple ofLoDConcentration(TCID50/mL)Multiples ofLoD
H1N1A/California/07/2009pdm09IRR$9.0\times10^1$3x$9.0\times10^1$9x
A/New Caledonia/20/99Zeptometrix$9.0\times10^1$3x$9.0\times10^1$9x
A/New Jersey/8/76TriCore$2.7\times10^2$9x$3.0\times10^1$3x
A/NWS/33TriCore$3.0\times10^1$1x$3.0\times10^1$3x
A/PR/8/34Charles River Labs$3.0\times10^1$1x$3.0\times10^1$3x
A1/Denver/1/57TriCore$3.0\times10^1$1x$3.0\times10^1$3x
A1/FM/1/47TriCore$3.0\times10^1$1x$3.0\times10^1$3x
A/ Solomon Islands/3/2006Zeptometrix$3.0\times10^1$1x$3.0\times10^1$3x
A/Hawaii/15/2001IRR$2.7\times10^2$9x$2.7\times10^2$27x
H3N2A/ Aichi/ 68Charles River Labs$1.0\times10^1$<1x$1.0\times10^1$3x
A/ Hong Kong/ 8/ 68Charles River Labs$3.0\times10^1$1x$1.0\times10^1$3x
A/ Victoria/ 3/ 75*Charles River Labs$2.4\times10^3$81x$2.4\times10^3$729x
A/Ohio/02/2012IRR$2.7\times10^2$9x$2.7\times10^2$81x

{11}------------------------------------------------

Influenza ASubtypeStrainSourceInfluenza AA/H1 or A/H3
Concentration(TCID50/mL)Multiple ofLoDConcentration(TCID50/mL)Multiples ofLoD
A/Indiana/08/2011IRR$1.0\times10^1$<1x$1.0\times10^1$3x
H3N2vA/Minnesota/11/2010**IRR$2.4\times10^3$81x$9.0\times10^1$27x
A/Indiana/10/2011IRR$1.0\times10^1$<1x$3.0\times10^1$9x
H2N2Japan/305/1957MRI$9.0\times10^1$3x--
H2N3Mallard/Albert79/03MRI$9.0\times10^1$3x--
H5N1A/Duck/Hunan/795/02MRI$9.0\times10^1$3x--
A/Chicken/Korea/IS/2006MRI$9.0\times10^1$3x--
A/Scaly-breasted Munia/HongKong/2006MRI$9.0\times10^1$3x--
H5N3A/Duck/Singapore/645/1997MRI$8.1\times10^2$27x--
H7N2A/New York/107/2003MRI$9.0\times10^1$3x--
H7N7A/Netherlands/219/2003MRI$2.7\times10^2$9x--
Equine-1/Prague/1956MRI$9.0\times10^1$3x--
H7N9Anhui/01/2013MRI$9.0\times10^1$3x--
H9N2Hong Kong/1073/99MRI$9.0\times10^1$3x--
Chicken/Hong Kong/G9/97MRI$9.0\times10^1$3x--
H10N7Chick/Germany/n/1949MRI$9.0\times10^1$3x--

彩 Based on in silico analysis, the oligonucleotide identities of all the tested Influenza A/H3N2 strains have very similar ranges. Based on the investigation of viral stocks titers using a quantitative TaqMan real-time PCR developed at Nanosphere that is specific for Influenza A/H3 strains (note: the primers for the TaqMan assay are not the same primers used in the RP Flex), it appears that the amplifiable genome equivalents available in this Influenza A/H3N2 viral stock are significantly reduced comparing to that of the other Influenza A/H3N2 stocks tested in the study.

** Based on in silico analysis, the oligonucleotide identities to this strain have slightly lower ranges than the other two H3N2v strains tested.

Influenza B Inclusivity Results

Table 4: Influenza B Inclusivity Results

TypeStrainSourceConcentration(TCID50/mL)Multiples ofLoD
Influenza BB/ Allen/45TriCore$9.0\times10^1$3x
B/Florida/07/2004TriCore$9.0\times10^1$3x
B/GL/1739/54TriCore$9.0\times10^1$3x
B/Hong Kong/5/72ATCC$9.0\times10^1$3x
B/Malaysia/2506/2004TriCore$9.0\times10^1$3x
B/Maryland/1/59TriCore$9.0\times10^1$3x
B/Taiwan/2/62TriCore$9.0\times10^1$3x
B/Wisconsin/01/2010IRR$9.0\times10^1$3x
B/ Lee/40Charles River Lab$9.0\times10^1$3x
B/Florida/04/2006Zeptometrix$9.0\times10^1$3x

{12}------------------------------------------------

Human Metapneumovirus Inclusivity Results

SubtypeStrainSourceConcentration(TCID50/mL)Multiples ofLoD
hMPV A116Zeptometrix 0810161CF$9.0\times10^1$3x
hMPV A220Zeptometrix 0810163CF$9.0\times10^1$3x
hMPV B15Zeptometrix 0810158CF$9.0\times10^1$3x
hMPV B24Zeptometrix 0810157CF$9.0\times10^1$3x
18Zeptometrix 0810162CF$9.0\times10^1$3x

Table 5: Human Metapneumovirus Inclusivity Results

Parainfluenza 1-4 Inclusivity Results

Parainfluenza 1-4 Inclusivity Results Table 6:

TypeSource/StrainConcentration(TCID50/mL)Multiples ofLoD
Parainfluenza 1Zeptometrix 0810014CF$2.7\times10^2$3x
Parainfluenza 2Zeptometrix 0810015CF$3.0\times10^1$3x
Parainfluenza 3ATCC VR-93*$2.7\times10^2$81x
BEI NR-3233$3.0\times10^1$9x
TriCore (ATCC VR-1782)$9.0\times10^1$27x
Parainfluenza 4Zeptometrix 0810060CF$8.1\times10^2$3x
VR-1377$8.1\times10^2$3x
Zeptometrix 0810060BCF$8.1\times10^2$3x
  • For Parainfluenza 3, the extracted eluate from the three strains tested in the inclusivity study were each evaluated with PCR/bi-directional sequencing, and the sequence information were used to assess the homology to the RP Flex oligos. Based on the in silico analysis, the three strains have the identical homology to the RP Flex oligos, indicating that the apparent difference in sensitivity was not due to sequence diversity in the gene targeted by the RP Flex. The apparent variation in the sensitivity of the RP Flex test for these strains is likely attributable to inconsistencies in the quantification of the viral stocks.

RSV Inclusivity Results

Table 7: RSV Inclusivity Results

SubtypeSource/StrainConcentration(TCID50/mL)Multiples ofLoD
Respiratory Syncytial Virus AATCC VR-26$1.0\times10^1$3x
Zeptometrix 0810040ACF$1.0\times10^1$3x
Respiratory Syncytial Virus BZeptometrix 0810040CF$1.1\times10^0$3x
ATCC VR-1400$1.1\times10^0$3x
ATCC VR-955$3.3 \times10^0$9x

{13}------------------------------------------------

Rhinovirus A and B Inclusivity Results

Rhinovirus SpeciesStrainSourceConcentration(TCID50/mL)Multiples ofLoD
Rhinovirus A1Zeptometrix 0810012CFN$2.7\times10^2$3x
2ATCC VR-482$2.7\times10^2$3x
7ATCC VR-1601$2.7\times10^2$3x
16ATCC VR-283$2.7\times10^2$3x
34ATCC VR-507$2.7\times10^2$3x
57ATCC VR-1600$2.7\times10^2$3x
77ATCC VR-1187$2.7\times10^2$3x
85ATCC VR-1195$2.7\times10^2$3x
Rhinovirus B3ATCC VR-483$2.7\times10^2$3x
17ATCC VR-1663$2.7\times10^2$3x
27ATCC VR-1137$2.7\times10^2$3x
42ATCC VR-338$2.7\times10^2$3x
83ATCC VR-1193$2.7\times10^2$3x

Rhinovirus A and B Inclusivity Results Table 8:

Rhinovirus C Inclusivity Results

Table 9: Rhinovirus C Inclusivity Results

Rhinovirus SpeciesStrainSourceConcentration(PFU/mL)*Multiples ofLoD
Rhinovirus CC2UW-Madison$7.3\times10^3$3x
Rhinovirus CC15UW-Madison$7.3\times10^3$3x
  • As there is no susceptible cell line to grow Rhinovirus C, the strains were cloned into a plasmid vector and transfected into WisL cells (primary human lung fibroblasts). All were sequenced to confirm identity. The titers were established by qPCR using serial dilutions of Rhinovirus 16 as a surrogate to provide actual PFU/mL values for the standard curve. Therefore, it has been assumed that Rhinovirus 16 has similar virulence rates to Rhinovirus C.

{14}------------------------------------------------

Bordetella Species Inclusivity Results

Bordetella SpeciesSourceRP FlexTargetConcentration(CFU/mL)Multiples ofLoD
B. pertussisATCC 51445B. pertussis$2.4\times10^3$3x
ATCC 10380$2.4\times10^3$3x
ATCC 9340$2.4\times10^3$3x
ATCC BAA-589$2.4\times10^3$3x
ATCC BAA-1335$2.4\times10^3$3x
ATCC 53894$2.4\times10^3$3x
ATCC 9306$2.4\times10^3$3x
ATCC 8467$7.3\times10^3$9x
ATCC 15237BordetellaParapertussis/bronchiseptica$7.3\times10^3$3x
ATCC 9305$7.3\times10^3$3x
B. parapertussisATCC BAA-587$7.3\times10^3$3x
ATCC 15989$7.3\times10^3$3x
Zeptometrix0801461$2.2\times10^4$9x
ATCC 4617$7.3\times10^3$
ATCC 7773$7.3\times10^3$
ATCC 785BordetellaParapertussis/bronchiseptica$7.3\times10^3$3x
B. bronchisepticaATCC 14064$7.3\times10^3$3x
ATCC 10580$7.3\times10^3$3x
ATCC 19395$7.3\times10^3$3x
Zeptometrix0801464$2.2\times10^4$9x
B. holmesiiATCC 700053B. holmesii
ATCC 700052

Table 10: Bordetella Species Inclusivity

Analytical Specificity (Exclusivity)

One hundred and seven (107) organisms (tables below), consisting of forty-six (46) bacterial/fungal strains (tested at 1×106 CFU/mL), twenty-six (26) viruses, twenty-two in-panel tested in the LoD study, and thirteen (13) additional influenza A virus strains with other hemagglutinin (HA) types were tested with RP Flex to determine analytical specificity (exclusivity).

The viral and bacterial/fungal samples were contrived in Simulated NPS at high concentrations (1×105 TCID50/mL for viral targets and at 1×106 CFU/mL for bacterial and fungal targets, except for Mumps virus which was tested at the highest available concentration of 1.60×100 TCID50/mL). Four (4) organisms which were not available as titered stocks were evaluated using genomic DNA at 1×10 copies/mL. All samples were tested in triplicate with the RP Flex.

{15}------------------------------------------------

Bacterial and Fungal Organisms Tested for RP Flex Analytical Specificity

Table 11: Bacterial and Fungal Organisms Tested for Analytical Specificity
-----------------------------------------------------------------------------------------
GenusSpeciesStrain Number
AcinetobacterbaumanniiATCC 19606
BordetellaaviumATCC 35086
BordetellahinziiATCC 51784
BordetellapetriiATCC BAA-461
BordetellatrematumATCC 700309
CandidaalbicansATCC 18804
CandidaglabrataATCC 38326
ChlamydophilapneumoniaeATCC VR-1360
Chlamydiatrachomatis Serovar DATCC VR-885
CorynebacteriumpseudodiphtheriticumATCC 10700
CorynebacteriumdiphtheriaeATCC 14779
CorynebacteriumstriatumATCC BAA-1293
EscherichiacoliATCC 25922
HaemophilusinfluenzaeATCC 49144
HaemophilusparainfluenzaeATCC 9796
Klebsiellapneumoniae subsp. pneumoniaeATCC 13883
LactobacillusacidophilusZeptometrix 0801540
LactobacillusplantarumATCC BAA-793
LegionellapneumophiliaATCC 33152
LegionellalongbechiaeATCC 33462
LegionellamicdadeiATCC 33204
ListeriainnocuaATCC 51742
Listeriamonocytogenes serotype 4bATCC 19115
Moraxella (Branhamella)catarrhalisATCC 43617
MycobacteriumtuberculosisATCC BAA-2237D-2ª
MycoplasmagenitaliumATCC 49123ª
MycoplasmahominisATCC 27545-TTR
MycoplasmapneumoniaeATCC 15531-TTR
Neisseriaelongata subsp. elongataATCC 25295
NeisseriagonorrhoeaeATCC 31426
NeisseriameningitidisATCC 53415D-5ª
NeisserialactamicaATCC 23970
NeisseriamucosaATCC 49233
NeisseriasiccaATCC 29256
PneumocystisjiroveciiErasme-Belgium-Clinical Sample*
ProteusvulgarisATCC 6380
PseudomonasaeruginosaATCC 27853
SerratiamarcescensATCC 29021
Staphylococcusaureus subsp. aureusATCC 12600
StaphylococcusepidermidisATCC 12228
StaphylococcushaemolyticusATCC 29970
StreptococcusagalactiaeATCC 12386
StreptococcuspneumoniaeATCC 6303
StreptococcuspyogenesATCC 14289
StreptococcussalivariusATCC 13419
UreaplasmaurealyticumATCC 27618ª

ª Genomic DNA tested at 1×106 copies/mL

{16}------------------------------------------------

Viral Organisms Tested for RP Flex Analytical Specificity

Virus NameTypeSource/Strain Number
Bocavirus-Clinical Sample
Coronavirus229EZeptometrix 0810229CF
CoronavirusNL63Zeptometrix 0810228CF
CoronavirusOC43Zeptometrix 0810024CF
CoronavirusHKU1LIJ-Clinical Sample
Cytomegalovirus-ATCC VR-977
Enterovirus AType 71Zeptometrix 0810047CF
Enterovirus ACoxsackievirus A2ATCC VR-1550
Enterovirus ACoxsackievirus A10Zeptometrix 0810106CF
Enterovirus BCoxsackievirus A9Zeptometrix 0810017CF
Enterovirus BCoxsackievirus B4ATCC VR-184
Enterovirus BCoxsackievirus B5ATCC VR-185
Enterovirus BEchovirus 6Zeptometrix 0810076CF
Enterovirus BEchovirus 9Zeptometrix 0810077CF
Enterovirus BEchovirus 11Zeptometrix 0810023CF
Enterovirus BEchovirus 30Zeptometrix 0810078CF
Enterovirus CCoxsackievirus A21Zeptometrix 0810235CF
Enterovirus CCoxsackievirus A24*ATCC VR-1662
Enterovirus CPoliovirus 2 (attenuated)*ATCC VR-301
Enterovirus CPoliovirus 3 (attenuated)*ATCC VR-193
Enterovirus DType 68*ATCC VR-561
Epstein Barr Virus-Zeptometrix 0810008CF
Herpes Simplex virusType 1Zeptometrix 0810005CF
Measles-ATCC VR-24
Mumps virus-ATCC VR-106
Varicella-Zoster virus-Zeptometrix 0810026CF

Table 12: Viral Organisms Tested for Analytical Specificity

In-Panel RP Flex Organisms (Viruses and Bacteria) and Additional Influenza A Virus Strains with Other Hemagglutinin (HA) Types Tested for Analytical Specificity

Table 13: In-Panel Organisms Tested for Analytical Specificity
Bacteria/Virus NameTypeSource/Strain Number
Adenovirus AType 31Zeptometrix ×810073CF
Adenovirus DType 26Zeptometrix 0810117CF
Adenovirus DType 37Zeptometrix 0810119CF
Adenovirus FType 40Zeptometrix 0810084CF
Adenovirus FType 41Zeptometrix 0810085CF
Adenovirus EType 4Zeptometrix 0810070CF
Bordetella holmesii-ATCC 51541
Bordetella pertussis-ATCC 9797
Influenza A /Brisbane/59/2007H1N1TriCore
Influenza A /Wisconsin/67/05H3N2Zeptometrix N/A
Influenza A/California/04/2009pdm09H1N1 - pandemicTriCore
Influenza A/Victoria/361/2011H3N2Zeptometrix 0810240CF
Influenza AH2N2aJapan/305/1957
Influenza AH5N1aA/Duck/Hunan/795/02

CONFIDENTIAL AND PROPRIETARY

Nanosphere, Inc.

{17}------------------------------------------------

Bacteria/Virus NameTypeSource/Strain Number
Influenza AH5N1aA/Chicken/Korea/IS/2006
Influenza AH5N1aScaly-breasted Munia/HongKong/2006
Influenza AH7N2aNew York/107/2003
Influenza AH7N7aNetherlands/219/2003
Influenza AH7N9aAnhui/01/2013
Influenza AH9N2aHong Kong/1073/99
Influenza AH2N3aMallard/Albert79/03
Influenza AH5N3aDuck/Singapore/645/1997
Influenza AH7N7aEquine-1/Prague/1956
Influenza AH9N2aChicken/Hong Kong/G9/97
Influenza AH10N7aChick/Germany/n/1949
Influenza B /Florida/02/2006-TriCore
Metapneumovirus 9Type A1TriCore
Metapneumovirus 8Type B2TriCore
Parainfluenza 1-TriCore VR-94
Parainfluenza 2-TriCore VR-92
Parainfluenza 3-Zeptometrix 0810016CF
Parainfluenza 4a-TriCore VR-1378
Respiratory Syncytial VirusType A2TriCore VR-1540
Respiratory Syncytial VirusType BTriCore VR-1580
Rhinovirus 14Type BTriCore

4 Prepared and tested at a BSL 3 laboratory.

All of the organisms tested vielded the expected "Not Detected" results at the concentrations tested with the exception of the enteroviruses marked with an asterisk and Pneumocystis jirovecii (from a clinical sample) marked with an asterisk, which gave "Rhinovirus detected" results in some of the replicates.

Based on in silico analyses, a number of Enterovirus strains have a relatively high homology to RP Flex Rhinovirus oligos, with some percent identities to Rhinovirus RP Flex oligos of 84%. As a result, some cross- reactivity at high titer was expected.

In silico analysis also determined that Pneumocystis jirovecii sequences have a maximum Oligo Identity to RP Flex targets of 67% and therefore Pneumocystis jirovecii was not predicted to be cross-reactive to RP Flex Rhinovirus probes. Extracted nucleic acids from all Rhinovirus positive tests of the Pneumocystis jirovecii positive clinical sample were evaluated with an analytically validated PCR/BDS Rhinovirus assay. PCR/BDS test results confirmed the presence of Rhinovirus in all samples, indicating that the Pneumocystis jirovecii positive clinical sample also contains Rhinovirus nucleic acids.

{18}------------------------------------------------

Interference (Exogenous and Endogenous Substances)

Microbial Interference

Three (3) representative target organisms detected by RP Flex, Adenovirus 3 (B), Influenza A (H1N1), and Bordetella pertussis were evaluated at 3x their respective LoD for potential interference in the presence of seven (7) potentially interfering microorganisms not detected by RP Flex: Staphylococcus aureus, Neisseria meningitidis, Corynebacterium diphtheria, Haemophilus influenza, Streptococcus pneumoniae, Mycoplasma pneumoniae, and cytomegalovirus. These seven (7) microorganisms represent the most prevalent microorganisms known to be present in the human upper respiratory tract and therefore the most likely to be encountered in NPS specimens. These normal flora organisms were tested at a concentration of 1×10° CFU/mL with the exception of Mycoplasma pneumoniae, which was tested at 1×10° CCU/mL, and Neisseria meningitidis, which was tested at 1×10° genomic copies/mL and cytomegalovirus, which was tested at 1×10° PFU/mL. No interference was observed with the RP Flex test for any of these samples tested.

Exogenous and Endogenous Substances

A comprehensive interfering substances study was performed to assess the potential effects of endogenous and exogenous substances that can commonly be found in clinical upper respiratory specimens. Three (3) representative target organisms detected by RP Flex, Adenovirus 3 (B), Influenza A (H1N1), and Bordetella pertussis were evaluated at 3x their respective LoD for potential interference in the presence of thirty-six (36) potentially interfering exogenous substances (table below). Two (2) endogenous substances were also included, human blood and human DNA. None of the substances at the concentrations tested showed any inhibitory effect on the detection of target respiratory pathogens using the RP Flex test.

Interfering Substances Tested
Wal-Four® Nasal SprayStaphylococcus aureus
Anefrin Nasal SprayNeisseria meningitidis
Saline Nasal SprayCorynebacterium diphtheriae
Similasan Sinus ReliefHaemophilus influenzae
Anbesol® (Anesthetic)Streptococcus pneumoniae
Qvar® (Nasal corticosteroid)Mycoplasma pneumoniae
Dexacort® (Nasal corticosteroid)Cytomegalovirus
AeroBid® (Nasal corticosteroid)Mucin, bovine submaxillary Type I-S
Triamcinolone (Nasal corticosteroid)Mucin, porcine stomach Type II
Pulmicort® (Nasal corticosteroid)Mucin, porcine stomach Type III
Flocon® (Nasal corticosteroid)BD Universal Viral Transport Media

Table 14: Interfering Substances

{19}------------------------------------------------

Interfering Substances Tested
Flonase® (Nasal corticosteroid)Remel M4®
Veramyst® (Fluticasone furoate)Remel M4-RT®
Tobramycin (systemic antibiotic)Remel M5®
Relenza™ (Anti-viral)Remel M6™
Tamiflu® (Anti-viral)BD Liquid Amies
Sulfur (Boiron®)Remel Regan Lowe Semi-Solid Transport Media
Galphimia Glauca (Boiron®)Copan ClassiqSwabs (Aluminum, rayon tipped, sterile)
Histaminum HydrochloricumCopan FloqSwabs (Nylon® ,regular, sterile)
Mupirocin (antibiotic)Ethyl Alcohol, Absolute 200 Proof
MentholAcetonitrile
Human BloodFluMist® Influenza Vaccine Live, Intranasal
Human DNA

Competitive Interference

In order to assess potential competitive inhibition for RP Flex, binary combinations of all test panel organisms representing all possible dual infections, were evaluated. Contrived samples were prepared in negative simulated NPS matrix, with one panel organism present at a Low Positive titer (3x LoD) and a second organism present at a High Positive titer (1×10) TCIDsomL for viruses, 1×10° CFU/mL for bacteria). The performance of Verigene RP Flex was evaluated with each of the one-hundred and eighty-two (182) unique sample combinations tested in replicates of three (3). No evidence of competitive inhibition was observed at the titers tested.

Carryover and Cross-Contamination

The potential for carryover and cross-contamination on the Verigene system was assessed by alternately testing three (3) high positive respiratory pathogen samples; Adenovirus 3 (B), Influenza A (H1N1) (both at 1×10° TCID50/mL), and Bordetella pertussis (at 1×10° CFU/mL), followed by testing a negative NPS sample. The high-titer sample was alternated with the negative sample five (5) times on six (6) unique Processor SPs. No carryover or crosscontamination was observed.

Specimen Stability

Fourteen (14) viral and bacterial strains in pooled Negative Clinical NPS were evaluated at Low Positive (2x LoD) and Moderate Positive (5x LoD) concentrations. Samples were stored at various temperature conditions and tested at defined timepoints in triplicate. The results of this stability study support the stability claim for RP Flex testing of clinical NPS specimens preserved in UTM at the following storage conditions: 4 hours at 20-25°C, 72 hours at 2-8°C, and 30 days at <- 70°C.

{20}------------------------------------------------

Precision

The Precision Study involved the testing of a representative test panel daily by two (2) operators for twelve (12) non-consecutive days for a total of forty-eight (48) tests per panel sample. The Precision Study used three (3) lots of each of the consumables (cartridges, extraction trays and amplification trays). All precision testing was performed at a single laboratory site with one (1) Verigene reader and twelve (12) Verigene Processor SPs. The test panel, representing all the RP Flex analytes except for B. parapertussis and B. bronchiseptica, consisted of two (2) negative samples (one negative simulated NPS matrix and one Staphylococcus aureus spiked in negative simulated NPS matrix), as well as seven (7) positive mixed samples at two (2) different concentrations for a total of sixteen (16) unique samples. Samples were prepared by spiking previously characterized and quantified organism stocks into simulated NPS matrix at Moderate Positive (5x LoD) and Low Positive (2x LoD) concentrations.

The results of the precision study are summarized below. which provides the percent agreement between the expected results and the obtained results for each sample tested.

Verigene RP Flex TargetPositive Percent Agreement (95% CI)Negative Percent Agreement*(95% CI)
LowModerate
Parainfluenza 1100%48/48(92.6-100)100%48/48(92.6-100)100%671/671(99.4-100)
Parainfluenza 2100%48/48(92.6-100)100%48/48(92.6-100)100%671/671(99.4-100)
Parainfluenza 3100%48/48(92.6-100)95.846/48(86.0-98.8)100%671/671(99.4-100)
Parainfluenza 4100%48/48(92.6-100)100%48/48(92.6-100)99.9%670/671(99.2-100)
RSV A100%48/48(92.6-100)100%48/48(92.6-100)100%671/671(99.4-100)
RSV B93.8%45/48(83.2-97.9)100%48/48(92.6-100)100%671/671(99.4-100)
Influenza A100%96/96(96.2-100)100%96/96(96.2-100)100%575/575(99.3-100)
Influenza A/H1100%48/48(92.6-100)100%48/48(92.6-100)100%671/671(92.4-100)
Influenza A/H3100%48/48(92.6-100)100%48/48(92.6-100)100%671/671(92.4-100)
Influenza B100%48/48(92.6-100)100%48/48(92.6-100)100%671/671(92.4-100)
Table 15: Precision Study
---------------------------------

CONFIDENTIAL AND PROPRIETARY Nanosphere, Inc.

Page 17 of 27

{21}------------------------------------------------

Verigene RP Flex TargetPositive Percent Agreement (95% CI)Negative Percent Agreement* (95% CI)
LowModerate
Rhinovirus97.9%47/48(89.1-99.6)100%48/48(92.6-100)99.9%670/671(99.2-100)
hMPV100%48/48(92.6-100)100%48/48(92.6-100)100%671/671(92.4-100)
Adenovirus100%48/48(92.6-100)100%48/48(92.6-100)99.7%669/671(98.9-99.9)
B. pertussis100%48/48(92.6-100)97.9%46/47(88.9-99.6)100%672/672(99.4-100)
B. holmesii100%47/48(89.1-99.6)100%47/47(92.4-100)100%672/672(99.4-100)
  • Negative Percent Agreement (NPA) was determined with all samples that did not contain the target analyte.

Reproducibility

The Reproducibility Study involved the testing of a representative test panel daily by two (2) operators for five (5) non-consecutive days at three (3) sites for a total of ninety (90) tests per sample. The test panel, representing all the RP Flex analytes except for B. parapertussis and B. bronchiseptica, consisted of two (2) negative samples (one negative simulated NPS matrix and one Staphylococcus aureus spiked in negative simulated NPS matrix), as well as seven (7) positive mixed samples at two (2) different concentrations for a total of sixteen (16) unique samples. Samples were prepared by spiking previously characterized and quantified organism stocks into simulated NPS matrix at Moderate Positive (5x LoD) and Low Positive (2x LoD) concentrations.

The results of the reproducibility study are summarized below, which provides the percent agreement between the expected results and the obtained results for each sample tested.

{22}------------------------------------------------

Table 16: Reproducibilitv Study
Verigene RP Flex TargetPositive Percent Agreement (95% CI)Negative Percent Agreement*
LowModerate(95% CI)
100%100%100%
Parainfluenza 190/9090/901258/1258
(96.2-100)(96.2-100)(99.7-100)
100%100%99.8%
Parainfluenza 289/8990/901256/1259
(95.9-100)(96.2-100)(99.3-99.9)
100%100%100%
Parainfluenza 390/9090/901258/1258
(96.2-100)(96.2-100)(99.7-100)
100%100%100%
Parainfluenza 490/9089/891259/1259
(96.2-100)(95.9-100)(99.7-100)
98.9%97.8%100%
RSV A89/9088/901258/1258
(94.0-99.8)(92.3-99.4)(99.7-100)
100%100%99.9%
RSV B90/9090/901257/1258
(96.2-100)(96.2-100)(99.6-100)
99.4100%100%
Influenza A179/179180/1801079/1079
(97.9-100)(97.9-100)(99.6-100)
100%100%99.8%
Influenza A/H190/9090/901256/1258
(96.2-100)(96.2-100)(99.4-100)
98.9%100%99.6%
Influenza A/H388/8990/901254/1259
(93.9-99.8)(96.2-100)(99.1-99.8)
100%100%99.8%
Influenza B90/9090/901255/1258
(96.2-100)(96.2-100)(99.3-99.9)
100%100%99.9%
Rhinovirus90/9090/901257/1258
(96.2-100)(96.2-100)(99.6-100)
100%100%99.9%
hMPV90/9089/891258/1259
(96.2-100)(95.9-100)(99.6-100)
100%100%99.8%
Adenovirus90/9090/901255/1258
(96.2-100)(96.2-100)(99.3 -99.9)
100%100%99.8%
Bordetella para/bronch90/9090/901256/1258
(96.2-100)(96.2-100)(99.4-100)
96.7%100%99.9%
B. pertussis87/9090/901257/1258
(90.7-98.9)(96.2-100)(99.6-100)
100%100%99.9%
B. holmesii90/9090/901257/1258
(96.2-100)(96.2-100)(99.6-100)
  • Negative Percent Agreement (NPA) was determined with all samples that did not contain the target analyte.

{23}------------------------------------------------

Performance Data - Clinical Testing

The clinical performance characteristics of the RP Flex test were determined by comparing viral and bacterial test results to an FDA-cleared molecular respiratory panel and/or PCR amplification followed by confirmatory bi-directional sequencing in a multi-site prospective investigation study at six (6) external clinical study sites. Subjects included individuals whose routine care called for respiratory pathogen testing.

A total of 3299 specimens were enrolled, of which 18 specimens were excluded from the study due to protocol violations, and 15 specimens which yielded a final "No Call" result were excluded from the performance analyses. Therefore, a total of 3266 specimens were included in the performance analyses. Enrolled specimens included 1082 prospectively-collected fresh specimens (of which 1069 were included in the performance analyses), 1330 prospectivelycollected frozen specimens (of which 1317 were included in the performance analyses), 526 retrospectively-collected frozen specimens (of which 520 were included in the performance analyses), and 361 contrived frozen specimens (of which 360 were included in the performance analyses). The following table provides a summary of demographic information for the 2412 prospectively-collected specimens in the enrolled dataset.

Prospective FreshProspective FrozenCombined
Age RangeNo. ofSpecimensPercentageNo. ofSpecimensPercentageNo. ofSpecimensPercentage
0-115114.0%16512.4%31613.1%
>1-517616.3%38228.7%55823.1%
>5-12736.7%987.4%1717.1%
>12-21746.8%675.0%1415.8%
>21-6542639.4%27520.7%70129.1%
>6516315.1%15511.7%31813.2%
Not Provided191.8%18814.1%2078.6%
Total1082100%1330100%2412100%
Table 17: Prospective Clinical Studies
------------------------------------------------

The comparator methods were a composite of an FDA-cleared molecular respiratory panel and analytically validated PCR with bi-directional sequencing. The tables below contain the clinical performance data generated during the RP Flex test clinical studies, stratified by specimen type, which, as previously described, were categorized as fresh and frozen prospectively collected specimens, frozen selected specimens and frozen contrived specimens.

{24}------------------------------------------------

Table 18: Results Stratified by Target Analyte – Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza B, Respiratory Syncytial Virus (RSV)A, Respiratory Syncytial Virus (RSV)B

Specimen Typen=% Agreement (95% CI)Specimen Typen=% Agreement (95% CI)
PositiveNegativePositiveNegative
Influenza AProspectivelyCollectedFresh1049100%12/12(75.7 - 100)99.4%1030/1037(98.6 - 99.7)Influenza A subtype H1ProspectivelyCollectedFresh1048--99.8%1046/1048(99.3 - 99.9)
Frozen114497.9%46/47(88.9 - 99.6)99.4%1091/1097(98.8 – 99.7)Frozen114497.8%45/46(88.7 - 99.6)99.6%1092/1096(99.1 - 99.9)
All219398.3%58/59(91.0-99.7)99.4%2121/2134(99.0 - 99.6)All219097.8%45/46(88.7-99.6)99.7%2138/2144(99.4-99.9)
Selected51399.2%122/123(95.5 - 99.9)99.5%387/390(97.8 - 99.7)Selected51297.6%40/41(87.4 - 99.6)99.6%469/471(98.5 - 99.9)
Contrived360--100%360/360(98.9 - 100)Contrived360--100%360/360(98.9 - 100)
Influenza A subtype H3ProspectivelyCollectedFresh1048100%12/12(75.7 - 100)99.6%1032/1036(99.0 - 99.8)Influenza BProspectivelyCollectedFresh105298.0%49/50(89.5 - 99.6)99.3%995/1002(98.6 - 99.7)
Frozen1142100%1/1(20.6 - 100)100%1141/1141(99.7 – 100)Frozen1145--99.9%1144/1145(99.5 - 100)
All2190100%13/13(77.2 - 100)99.8%2173/2177(99.5-99.9)All219798.0%49/50(89.5 - 99.6)99.6%2139/2147(99.3 - 99.8)
Selected512100%82/82(95.5-100)99.5%428/430(98.3 - 99.9)Selected516100%26/26(87.1 - 100)99.6%488/490(98.5 - 99.9)
Contrived360--100%360/360(98.9-100)Contrived360--100%360/360(98.9-100)
RSV AProspectivelyCollectedFresh1049100%11/11(74.1-100)99.8%1036/1038(99.3-99.9)RSV BProspectivelyCollectedFresh1049100%8/8(67.6 - 100)99.6%1037/1041(96.7 - 98.6)
Frozen1121100%6/6(61.0-100)99.9%1114/1115(99.5-100)Frozen1121100%165/165(97.7 - 100)97.9%936/956(96.8 - 98.6)
All2170100%17/17(81.6-100)99.9%2150/2153(99.6-99.9)All2170100%173/173(97.8 - 100)98.8%1973/1997(98.2 - 99.2)
Selected49894.8%55/58(85.9-98.2)99.3%437/440(98.0-99.8)Selected498100%23/23(85.7 - 100)98.5%468/475(97.0 - 99.3)
Contrived360--100%360/360(98.9-100)Contrived360--99.7%359/360(98.4 - 99.9)
Specimen Typen=% Agreement (95% CI)Specimen Typen=% Agreement (95% CI)
PositiveNegativePositiveNegative
ProspectivelyCollectedParainfluenza 1Fresh1052-100%1052/1052(99.6 - 100)ProspectivelyCollectedParainfluenza 2Fresh1052100%11/11(74.1 - 100)99.7%1038/1041(99.2 - 99.9)
Frozen114590.0%27/30(74.4 - 96.5)99.8%1113/1115(99.3 - 99.9)Frozen114550.0%1/2(9.5 - 90.5)100%1143/1143(99.7 - 100)
All219790.0%27/30(74.4 - 96.5)99.9%2165/2167(99.7 - 100)All219792.3%12/13(66.7 - 98.6)99.9%2181/2184(99.6 - 99.9)
Selected516100%50/50(92.9 - 100)100%466/466(99.2 - 100)Selected516100%28/28(87.9 - 100)99.8%487/488(98.8 - 100)
Contrived360-100%360/360(98.9 - 100)Contrived360-99.7%359/360(98.4 - 99.9)
ProspectivelyCollectedParainfluenza 3Fresh105283.3%10/12(55.2 - 95.3)99.7%1037/1040(99.2 - 99.9)ProspectivelyCollectedParainfluenza 4Fresh1052100%3/3(43.8 - 100)100%1049/1049(99.6 - 100)
Frozen114580.0%4/5(37.5 - 96.4)100%1140/1140(99.7 - 100)Frozen114576.2%16/21(54.9 - 89.4)99.6%1120/1124(99.1 - 99.9)
All219782.4%14/17(59.0 - 93.8)99.9%2177/2180(99.6 - 99.9)All219779.2%19/24(59.3 - 90.8)99.8%2169/2173(99.5 - 99.9)
Selected516100%31/31(89.0 - 100)100%485/485(99.2 - 100)Selected516100%41/41(91.4 - 100)99.6%473/475(98.5 - 99.9)
Contrived360-100%360/360(98.9 - 100)Contrived360-100%360/360(98.9 - 100)

{25}------------------------------------------------

Table 19: Results Stratified by Target Analyte –Parainfluenza 1, Parainfluenza 2, Parainfluenza 3, Parainfluenza 4

{26}------------------------------------------------

Specimen Typen=% Agreement (95% CI)Specimen Typen=% Agreement (95% CI)
PositiveNegativePositiveNegative
AdenovirusProspectivelyCollectedFresh105291.7%22/24(74.1 – 97.7)98.2%1009/1028(97.1 – 98.8)hMPVProspectivelyCollectedFresh1052100%10/10(72.2 – 100)99.5%1037/1042h(98.9 – 99.8)
Frozen114581.8%27/33(65.6 – 91.4)96.4%1072/1112(95.1 – 97.3)Frozen1145100%36/36(90.4 – 100)99.9%1108/1109i(99.5 – 100)
All219786.0%49/57(74.7 – 92.7)97.2%2081/2140(96.5 – 97.9)All2197100%46/46(92.3 – 100)99.7%2145/2151(99.4 – 99.9)
Selected51697.4%38/39(86.8 – 99.5)98.3%469/477(96.7 – 99.1)Selected51692.6%25/27g(76.6 – 97.9)99.8%488/489j(98.8 – 100)
Contrived36099.4%358/360(98.0 – 99.8)Contrived36099.4%358/360(98.0 – 99.8)
RhinovirusProspectivelyCollectedFresh100085.9%214/249(81.1 – 89.7)95.7%719/751(94.1 – 97.0)
Frozen112277.8%193/248(72.2 – 82.5)98.3%859/874o(97.2 – 99.0)
All212281.9%407/497(78.3 – 85.0)97.1%1578/1625(96.2 – 97.8)
Selected50980.0%28/35(64.1 – 90.0)98.3%466/474(96.7 – 99.1)
Contrived36099.7%359/360(98.4 – 99.9)

Table 20: Results Stratified by Target Analyte –Adenovirus, human Metapneumovirus (hMPV), Rhinovirus

{27}------------------------------------------------

Table 21: Results Stratified by Target Analyte –Bordetella parapertussis/bronchiseptica, Bordetella pertussis, Bordetella holmesii

SpecimenTypen=% Agreement (95% CI)SpecimenTypen=% Agreement (95% CI)
PositiveNegativePositiveNegative
Bordetella parapertussis/bronchisepticaProspectivelyCollectedFresh1041100%2/2(34.2 - 100)100%1039/1039(99.6 - 100)Bordetella pertussisProspectivelyCollectedFresh1052100%1/1(20.6 - 100)99.9%1050/1051(99.5 - 100)
Frozen1255-99.9%1254/1255(99.5 - 100)Frozen1145100%7/7(64.6 - 100)99.9%1137/1138(99.5 - 100)
All2296100%2/2(34.2 - 100)99.9%2290/2291(99.8 - 100)All2197100%8/8(67.6 - 100)99.9%2187/2189(99.7 - 100)
Selected46371.4%5/7(35.9 - 91.8)99.8%455/456(98.8 - 100)Selected51696.6%28/29(82.8 - 99.4)100%487/487(99.2 - 100)
Contrived360100%104/104(96.4 - 100)100%256/256(98.5 - 100)Contrived360-100%360/360(98.9 - 100)
Bordetella holmesiiProspectivelyCollectedFresh1043100%1/1(20.6 - 100)100%1042/1042(99.6 - 100)
Frozen1263-100%1263/1263(99.7 - 100)
All2306100%1/1(20.6 - 100)100%2305/2305(99.8 - 100)
Selected49050%1/2(9.4 - 90.1)100%488/488(99.2 - 100)
Contrived360100%56/56(93.6 - 100)100%304/304(98.6 - 100)

{28}------------------------------------------------

Substantial Equivalence

The Verigene Respiratory Pathogens Flex Nucleic Acid test (RP Flex) has been shown to be substantially equivalent to the BioFire FilmArray Respiratory Panel (RP) System. The RP Flex test has similar intended use and indications, technological characteristics, and performance characteristics. Performance data demonstrate that the RP Flex test performs comparably to the predicate device. Thus, the RP Flex test is substantially equivalent to the predicate device.

Similarities
ElementNew Device:Verigene Respiratory Pathogens Flex NucleicAcid Test (RP Flex)K143653Predicate:FilmArray® Respiratory Panel (RP)K143080/K123620/K120267/K110764/K103175
Intended UseThe Verigene® Respiratory Pathogens FlexNucleic Acid Test (RP Flex) is a multiplexedqualitative test intended for the simultaneousdetection and identification of multiple viraland bacterial nucleic acids in nasopharyngealswabs (NPS) obtained from individualssuspected of respiratory tract infection. Thetest is performed on the automated VerigeneSystem utilizing reverse transcription (RT),polymerase chain reaction (PCR), andmicroarray hybridization to detect genesequences of the following organism types andsubtypes: Adenovirus, HumanMetapneumovirus, Influenza A, Influenza A(subtype H1), Influenza A (subtype H3),Influenza B, Parainfluenza 1, Parainfluenza 2,Parainfluenza 3, Parainfluenza 4, RespiratorySyncytial Virus A, Respiratory SyncytialVirus B, Rhinovirus, Bordetellaparapertussis/bronchiseptica, Bordetellaholmesii, and Bordetella pertussis.Detecting and identifying specific viral andbacterial nucleic acids from individualsexhibiting signs and symptoms of respiratoryinfection aids in the diagnosis of respiratoryinfection, if used in conjunction with otherclinical and laboratory findings. The results ofthis test should not be used as the sole basisfor diagnosis, treatment, or patientmanagement decisions.The FilmArray® Respiratory Panel (RP) is amultiplexed nucleic acid test intended for usewith the FilmArray Instrument for thesimultaneous qualitative detection andidentification of multiple respiratory viral andbacterial nucleic acids in nasopharyngeal swabs(NPS) obtained from individuals suspected ofrespiratory tract infections. The followingorganism types and subtypes are identified usingthe FilmArray RP: Adenovirus, Coronavirus229E, Coronavirus HKU1, Coronavirus NL63,Coronavirus OC43, Human Metapneumovirus,Influenza A, Influenza A subtype H1, InfluenzaA subtype H3, Influenza A subtype H1-2009,Influenza B, Parainfluenza Virus 1,Parainfluenza Virus 2, Parainfluenza Virus 3,Parainfluenza Virus 4, HumanRhinovirus/Enterovirus, Respiratory SyncytialVirus, Bordetella pertussis, Chlamydophilapneumoniae, and Mycoplasma pneumoniae.The detection and identification of specific viraland bacterial nucleic acids from individualsexhibiting signs and symptoms of a respiratoryinfection aids in the diagnosis of respiratoryinfection if used in conjunction with otherclinical and epidemiological information. Theresults of this test should not be used as the solebasis for diagnosis, treatment, or othermanagement decisions.
Negative results in the presence of arespiratory illness do not preclude respiratoryinfection and may be due to infection withpathogens that are not detected by this test orlower respiratory tract infection that is notdetected by an NPS specimen. Conversely,Negative results in the setting of a respiratoryillness may be due to infection with pathogensthat are not detected by this test or, lowerrespiratory tract infection that is not detected bya nasopharyngeal swab specimen. Positiveresults do not rule out co-infection with otherorganisms: the agent(s) detected by the Film
Similarities
ElementNew Device:Verigene Respiratory Pathogens Flex NucleicAcid Test (RP Flex)K143653Predicate:FilmArray® Respiratory Panel (RP)K143080/K123620/K120267/K110764/K103175
infection with organisms not detected by RPFlex. The agent(s) detected may not be thedefinite cause of disease. The use ofadditional laboratory testing and clinicalpresentation may be necessary to establish afinal diagnosis of respiratory infection.Clinical evaluation indicates a lowersensitivity specific to RP Flex for thedetection of rhinovirus. If infection withRhinovirus is suspected, negative samplesshould be confirmed using an alternativemethod.Performance characteristics for Influenza Awere established when Influenza A/H1 (2009Pandemic) and A/H3 were the predominantInfluenza A viruses in circulation. RP Flexmay not detect novel Influenza A strains. Ifinfection with a novel Influenza A virus issuspected based on current clinical andepidemiological screening criteriarecommended by public health authorities,specimens should be collected withappropriate infection control precautions usedspecifically for novel virulent influenzaviruses and sent to appropriate healthauthorities for testing. Viral culture shouldnot be attempted in these cases unless abiosafety level (BSL) 3+ facility is availableto receive and culture specimens.disease. Additional laboratory testing (e.g.bacterial and viral culture, immunofluorescence,and radiography) may be necessary whenevaluating a patient with possible respiratorytract infection.Due to the small number of positive specimenscollected for certain organisms during theprospective clinical study, performancecharacteristics for Bordetella pertussis ,Coronavirus 229E, Coronavirus OC43, InfluenzaA H1, Influenza A H3, Influenza A H1-2009,Influenza B, Mycoplasma pneumoniae ,Parainfluenza Virus 1, Parainfluenza.Due to the genetic similarity between HumanRhinovirus and Enterovirus, the FilmArray RPcannot reliably differentiate them. A positiveFilmArray RP Rhinovirus/Enterovirus resultshould be followed-up using an alternate method(e.g., cell culture or sequence analysis).The FilmArray RP assay for Coronavirus OC43may cross-react with some isolates ofCoronavirus HKU1. A dual positive result maybe due to cross-reactivity or may indicate a co-infection.Performance characteristics for Influenza A wereestablished when Influenza A H1-2009, A H1,and A H3 were the predominant Influenza Aviruses in circulation. Performance of detectingInfluenza A may vary if other Influenza Astrains are circulating or a novel Influenza Avirus emerges. If infection with a novelInfluenza A virus is suspected based on currentclinical and epidemiological screening criteriarecommended by public health authorities,specimens should be collected with appropriateinfection control precautions for novel virulentInfluenza viruses and sent to state or local healthdepartments for testing. Viral culture should notbe attempted in these cases unless a BSL 3+facility is available to receive and culturespecimens.
SpecimenTypeNasopharyngeal swabs (NPS)Nasopharyngeal swabs (NPS)
Nucleic AcidAmplificationMultiplexed RT-PCRMultiplexed RT-PCR
Organisms/NAAdenovirus, Human Metapneumovirus,Adenovirus, Coronavirus 229E, Coronavirus
Similarities
ElementNew Device:Verigene Respiratory Pathogens Flex NucleicAcid Test (RP Flex)K143653Predicate:FilmArray® Respiratory Panel (RP)K143080/K123620/K120267/K110764/K103175
TargetsDetectedInfluenza A, Influenza A (subtype H1),Influenza A (subtype H3), Influenza B,Parainfluenza 1, Parainfluenza 2,Parainfluenza 3, Parainfluenza 4, RespiratorySyncytial Virus A, Respiratory SyncytialVirus B., Rhinovirus, Bordetellaparapertussis/bronchiseptica, Bordetellaholmesii, Bordetella pertussisHKU1, Coronavirus NL63, Coronavirus OC43,Human Metapneumovirus, Influenza A,Influenza A subtype H1, Influenza A subtypeH3, Influenza A subtype H1-2009, Influenza B,Parainfluenza Virus 1, Parainfluenza Virus 2,Parainfluenza Virus 3, Parainfluenza Virus 4,Human Rhinovirus/Enterovirus, RespiratorySyncytial Virus, Bordetella pertussis,Chlamydophila pneumoniae, and Mycoplasma

Table 22: Substantial Equivalence

CONFIDENTIAL AND PROPRIETARY Nanosphere, Inc.

{29}------------------------------------------------

{30}------------------------------------------------

Differences
ElementNew Device:Verigene Respiratory Pathogens Flex NucleicAcid test (RP Flex)K143653Predicate:FilmArray® Respiratory Panel (RP)K123620/K120267/K110764/K103175
Time to ResultAbout 2 hoursAbout 1 hour
DetectionMethodMultiplexed RT-PCR followed by a goldnanoparticle probe-based detection ofmicrobial-specific DNA in a microarrayformatNested multiplex RT-PCR followed by highresolution melting analysis to confirm identity ofamplified product
OpticalDetectionLight ScatteringFluorescence

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.