K Number
K243682

Validate with FDA (Live)

Date Cleared
2025-06-27

(210 days)

Product Code
Regulation Number
868.5150
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SonoPlex STIM and SonoPlex II Needles equipped with Cornerstone reflectors are used to puncture the tissue in order to gain entry and inject local anesthetics to induce regional anesthesia. An electrical stimulus may be applied to the needle via a cable and connector to assist the physician pinpoint the area of application.

Device Description

The subject device, the SonoPlex STIM and SonoPlex II nerve block needle is a single-use anesthesia conducting needle intended to gain entry or puncture the tissue and inject anesthetics to induce regional anesthesia. The SonoPlex STIM and SonoPlex II needle is intended for use under ultrasound guidance. Visibility under ultrasound is enhanced by CornerStone reflectors as cleared by FDA in several Premarket Notification Submissions. The SonoPlex STIM and SonoPlex II needle is equipped with an connecting cable, injection tube and NanoLine® coating (Parylene, no nanomaterials incorporated). The distal connection of the tube is either equipped with a LUER Connector according to ISO 80369-7 or a NRFit-Connector according to ISO 80369-6. The device in focus is a single use device and has an intended time of use up to 24 hours acc. EO-residuals acc. DIN EN ISO 10993-7. The standard time of use is less than 60 minutes. The SonoPlex STIM and SonoPlex II cannulas are not for intrathecal use. The SonoPlex STIM and SonoPlex II cannulas are produced at PAJUNK® GmbH Medizintechnologie in Geisingen, Germany.

AI/ML Overview

The provided FDA 510(k) clearance letter and associated summary pertain to the SonoPlex STIM and SonoPlex II anesthesia conduction needles. This document focuses on demonstrating substantial equivalence to a predicate device (SonoBlock/SonoBlock II) and primarily addresses changes related to sterilization and packaging, rather than the performance of an AI/ML-driven device.

Therefore, the requested information regarding AI/ML device performance (e.g., acceptance criteria for AI, sample sizes for test/training sets, expert adjudication, MRMC studies, standalone algorithm performance, ground truth definitions for AI models) cannot be extracted from this document, as it describes a traditional medical device and not an AI-powered one.

The "study" described in the document is a series of performance tests and validations for the physical properties, sterilization, and biocompatibility of the needles, and not a study validating an AI/ML diagnostic or assistive tool.

However, I can provide a summary of the acceptance criteria and the studies that prove the device (SonoPlex STIM and SonoPlex II needles) meets its acceptance criteria based on the information provided:


Acceptance Criteria and Device Performance (SonoPlex STIM and SonoPlex II Needles)

The acceptance criteria for the SonoPlex STIM and SonoPlex II needles are primarily based on established international standards for medical devices, specifically for anesthesia conduction needles. The studies performed were performance testing, sterilization validation, and shelf-life testing, demonstrating that the device meets these criteria.

1. Table of Acceptance Criteria and Reported Device Performance

The core acceptance criteria are compliance with various ISO standards, ensuring the physical and biological safety and efficacy of the needles.

Acceptance Criteria CategorySpecific Criteria (Standard/Section)Reported Device Performance
Material & Physical PropertiesISO 9626: Stainless steel needle tubing
5.2 Surface finish and visual appearancePassed
5.3 CleanlinessPassed
5.4 Limits for acidity and alkalinityPassed
5.5 Size designationPassed
5.6 DimensionsPassed
5.7 Sample sizePassed
5.8 StiffnessPassed
5.9 Resistance to breakagePassed
5.10 Resistance to corrosionPassed
Device PerformanceISO 7864: Sterile hypodermic needles for single use
4.3 CleanlinessPassed
4.4 Limits for acidity or alkalinityPassed
4.5 Limits for extractable metalsPassed
4.10 Needle TubePassed
4.11 Needle pointPassed
4.12 Bond between hub and needle tubePassed
4.13 Patency of lumenPassed
Connectivity StandardsISO 80369-6: Connectors for neuraxial applications (NRFit)
6.1 Fluid LeakagePassed
6.2 Air LeakagePassed
6.3 Stress CrackingPassed
6.4 Separation Axial LoadPassed
6.5 UnscrewingPassed
6.6 OverridingPassed
ISO 80369-7: Connectors for intravascular or hypodermic applications (LUER)
7.1 Fluid LeakagePassed
7.2 Air LeakagePassed
7.3 Stress CrackingPassed
7.4 Separation Axial LoadPassed
7.5 UnscrewingPassed
7.6 OverridingPassed
Sterilization & BiocompatibilitySterility Assurance Level (SAL)SAL = 10^-6 (Validated at Sterigenics, Wiesbaden and HA2 Medizintechnik, Halberstadt)
Ethylene Oxide (EO) Residuals (ISO 10993-7)Complies (limits: 25ppm for EO and Ethylene chlorhydrine)
Biocompatibility (ISO 10993-series)Complies with ISO 10993-1 and FDA guidance
Shelf LifeMaintenance of Sterility (after 5 years)Sterile Barrier System efficient; devices found sterile after 5 years
Maintenance of Performance (after 5 years)No decrease in performance (LUER connection, bonding, bending rigidity) after 5 years

2. Sample Sizes Used for the Test Set and Data Provenance

  • Sample Size: The document repeatedly mentions that tests were conducted, but does not specify the exact sample sizes for each of the performance tests (e.g., how many needles were tested for patency, stiffness, etc.). It mentions "process most challenging worst case devices" were used for shelf life and sterility tests.
  • Data Provenance: The devices are manufactured by PAJUNK GmbH Medizintechnologie in Geisingen, Germany. The contract sterilizers are Sterigenics Germany GmbH (Wiesbaden, Germany) and HA2 Medizintechnik GmbH (Halberstadt, Germany). The tests appear to be prospective (conducted specifically for this submission) in a laboratory/manufacturing setting.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This section is not applicable as the device is a physical medical needle, not an AI/ML diagnostic or assistive device that would require expert-established ground truth from medical images or clinical data. The acceptance criteria are based on engineering specifications and international standards.

4. Adjudication Method for the Test Set

This section is not applicable for the same reasons as above. The tests performed are objective, quantitative measurements against defined standard specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

This section is not applicable. This is a physical medical device, not an imaging or diagnostic AI/ML tool that would be subject to MRMC studies. The document specifies that the new devices are "substantially equivalent" to predicate devices, and the dual guidance (ultrasound + electrical stimulus) for localization is considered "state of the art" and "safe and effective."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This section is not applicable. The device is a needle used by a human physician, possibly with assistance from an ultrasound machine and/or an electrical stimulator. There is no AI algorithm component.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is defined by the specifications within the cited ISO standards (e.g., acceptable limits for fluid leakage, minimum breaking strength, chemical residue limits) and validated manufacturing processes (e.g., sterilization parameters, biocompatibility testing). It is not derived from patient outcomes, expert consensus on clinical findings, or pathology reports in the context of an AI/ML algorithm.

8. The Sample Size for the Training Set

This section is not applicable. There is no AI/ML component; therefore, no training set is relevant to this device submission.

9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the same reason as above.

FDA 510(k) Clearance Letter - SonoPlex STIM and SonoPlex II

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 27, 2025

PAJUNK GmbH Medizintechnologie
Chiara Meyer
Specialist Regulatory Affairs
Karl-Hall-Str. 1
Pajunkstr. 1
Geisingen, BW 78187
Germany

Re: K243682
Trade/Device Name: SonoPlex STIM; SonoPlex II
Regulation Number: 21 CFR 868.5150
Regulation Name: Anesthesia Conduction Needle
Regulatory Class: Class II
Product Code: BSP
Dated: May 27, 2025
Received: May 27, 2025

Dear Chiara Meyer:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243682 - Chiara Meyer Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243682 - Chiara Meyer Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bradley Q. Quinn -S

Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K243682

Device Name
SonoPlex STIM;
SonoPlex II

Indications for Use (Describe)
SonoPlex STIM and SonoPlex II Needles equipped with Cornerstone reflectors are used to puncture the tissue in order to gain entry and inject local anesthetics to induce regional anesthesia.
An electrical stimulus may be applied to the needle via a cable and connector to assist the physician pinpoint the area of application.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

Page 5

510(k) owner: PAJUNK® GmbH Medizintechnologie
Karl-Hall-Str. 1/ Pajunkstr. 1
78187 Geisingen
Baden-Wuerttemberg, Germany
Fon: +49(0)7704-9291-586
Fax: +49(0)7704-9291-605
Establishment Registration Number: 9611612

Submitter Information/ production site: PAJUNK® GmbH Medizintechnologie
Karl-Hall-Str. 1/ Pajunkstr. 1
78187 Geisingen
Baden-Wuerttemberg, Germany
Fon: +49(0)7704-9291-586
Fax: +49(0)7704-9291-605
Establishment Registration Number: 9611612

Manufacturer/ Submitter Contact: Christian G. H. Quass
Director Regulatory Affairs, Safety Officer
Fon: +49(0)7704-9291-586
Fax: +49(0)7704-9291-602
E-Mail: christian.quass@pajunk.com

Alternative Manufacturer/ Submitter Contact: Chiara Meyer
Regulatory Affairs Specialist
Fon: +49(0)7704-9291-627
Fax: +49(0)7704-9291-602
E-Mail: chiara.meyer@pajunk.com

USA

08 510(k) Summary

Registrations USA

510(k) Summary as required by 21 CFR 807.92(c).

Date of Preparation: 2025-05-22

Document Control Number: K243682

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
1 / 16

Page 6

USA Contact: PAJUNK MEDICAL SYSTEMS
4575 Marconi Dr.
Alpharetta, GA 30005
Registration Number: 3004076349
FEI Number*: 3004076349

US-Official Correspondent Marco Wohnig
President
PAJUNK Medical Systems
4575 Marconi Dr.
Alpharetta, GA 30005
Phone: 001-770-4936832
E-Mail: Marco.Wohnig@pajunk-usa.com

Contract Sterilizer: Sterigenics Germany GmbH
Kasteler straße 45
65203 Wiesbaden
Germany, Hessen
Establishment Registration Number: 3002807090

2nd Contract Sterilizer: HA2 MEDIZINTECHNIK GMBH
Am Bahndamm 11
Halberstadt Saxony-Anhalt, DE 38820
Registration Number: 3009039068
FEI Number*: 3009039068

Device Name and Classification

Device Name: SonoPlex STIM, SonoPlex ll
Premarket Notification Number: K243682
Classification Name: Anesthesia Conduction Needle
Classification Reference: 21 CFR § 868.5150
Product Code: BSP
Establishment Registration Number: 9611612
Regulatory Class: II
Panel: Anesthesiology

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
2 / 16

Page 7

Sterilization Method: Ethylene Oxide
disposable device, supplied sterile to the end user and non-sterile intended to be sterilized prior to use to re-packagers/ medical device manufacturers

Contract Sterilizer: Sterigenics Germany GmbH
Kasteler straße 45
65203 Wiesbaden
Germany, Hessen
Establishment Registration Number: 3002807090

Additional Contract Sterilizer: HA2 MEDIZINTECHNIK GMBH
Am Bahndamm 11
Halberstadt Saxony-Anhalt, DE 38820
Registration Number: 3009039068
FEI Number*: 3009039068

Changes to Device: Predicate Device

Device Name: SonoBlock, SonoBlock II
Premarket Notification Number: K241954
Classification Name: Anesthesia Conduction Needle
Classification Reference: 21 CFR § 868.5150
Product Code: BSP
Establishment Registration Number: 9611612
Regulatory Class: II
Panel: Anesthesiology

Sterilization Method: Ethylene Oxide
disposable device, supplied sterile to the end user and non-sterile intended to be sterilized prior to use to re-packagers/ medical device manufacturers

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
3 / 16

Page 8

Contract Sterilizer: Sterigenics Germany GmbH
Kasteler straße 45
65203 Wiesbaden
Germany, Hessen
Establishment Registration Number: 3002807090

Additional Contract Sterilizer: HA2 MEDIZINTECHNIK GMBH
Am Bahndamm 11
Halberstadt Saxony-Anhalt, DE 38820
Registration Number: 3009039068
FEI Number*: 3009039068

The subject device, the SonoPlex STIM and SonoPlex ll needle is a single-use anaesthesia conducting needle intended for use peripheral nerve blocks.

The SonoPlex STIM and SonoPlex ll needle is intended for use under ultrasound guidance. Visibility under ultrasound is enhanced by CornerStone reflectors as cleared by FDA in several Premarket Notification Submissions. Additionally and optionally an electrical stimulus may be applied to the needle via a cable and connector to assist the physician pinpoint the area of application (also referred to as Dual Guidance).

The predicate device, the SonoBlock and SonoBlock ll nerve block needle is a single-use anaesthesia conducting nerve block needle intended for performing peripheral nerve blocks.

The subject device as well as the predicate device have been cleared earlier by the Sponsor in K111374. The predicate device has also been cleared in K241954 to file a standalone 510(k) and to inform the Agency about additional options to be added to the device´s manufacturing process.

K111374 in 2011 has been filed and cleared for ultrasound enhanced Parylene coated needles (Cornerstones) optionally equipped with an electrical cable for optional application of an electrical stimulus.These needles are optionally used for Dual Guidance: Ultrasound guidance and verification via electrical stimulus (while landmark technique without ultrasound and electrical stimulus is also widespread).

Since 2011 some alterations have taken place after clearance:

  • Needles formerly named SonoPlex (Ultrasound only) have been renamed to SonoBlock
  • Hub-to-needle technology formerly cleared as gluing (2K and UV) has been optionally amended by direct injection moulding
  • For injection moulding the material of the hub has been changed
  • By direct overmoulding, the tube and where applicable the cable are directly moulded to the hub and one

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
4 / 16

Page 9

screw connection is substituted by this manufacturing step.

Each of these alterations followed FDA´s Guidance "Deciding When to Submit a 510(k) for a Change to an Existing Device". None of these alterations did result in additional hazards to the patient since the indications for use and the clinical technique have not been impacted.

Each of the alterations has been diligently verified and validated during design process and has been accompanied by risk management measures during manufacturing.

None of the alterations did impact the risk status or the labelling. Safety and effectiveness as well as efficacy of the device remain unaltered.

This Premarket Notification is intended to file a standalone 510(k) for the SonoPlex STIM and SonoPlex ll needles and to inform the Agency about additional options to be added to the device´s manufacturing process.

Subject to this 510(k) is

  • the validation of an additional optional sterilization service provider (HA2 Medizintechnik, Halberstadt, Establishment Registration number 3009039068)
  • the validation of an additional optional packaging material (medical paper).

Both, new sterilization service provider and additional packaging material, have been added to the Design History file and the Device master record in order to be able to package and sterilize the device using additional processes as an additional option.

Both, new sterilization service provider and additional packaging material, have been cleared earlier by the sponsor in K241954.

Both devices, the predicate SonoBlock and SonoBlock II as well as the subject SonoPlex STIM and SonoPlex ll are manufactured by PAJUNK GmbH Medizintechnologie, the sponsor of this submission, in Geisingen, Germany.

Both devices, the predicate SonoBlock and SonoBlock II as well as the subject SonoPlex STIM and SonoPlex II share the same indications for use. The safety and effectiveness in localization options is substantially equivalent.

Apart from stimulation option both devices, the predicate device as well as the subject device, share identical required specifications for device performance as laid down in ISO 9626, ISO 7864, ISO 80369-series, ISO 10993-series and ISO 11135-series.

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
5 / 16

Page 10

Both devices, the predicate device as well as the subject device are manufactured in the same regulatory framework, e. g. Quality Management System (compliant with 21cfr820 and ISO 13485) using the same risk management system according to ISO 14971.

The SonoPlex STIM and SonoPlex II are also available in bulk non sterile. If appropriately packed and sterilized with Ethylene oxide according to the given sterilization parameters the technological parameters remain unchanged. However, final responsibility for sterilization validation remains with the customer of SonoPlex STIM and SonoPlex II purchased bulk non-sterile.

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
6 / 16

Page 11

1 Narrative Device Description

The subject device, the SonoPlex STIM and SonoPlex II nerve block needle is a single-use anesthesia conducting needle intended to gain entry or puncture the tissue and inject anesthetics to induce regional anesthesia.

The SonoPlex STIM and SonoPlex II needle is intended for use under ultrasound guidance. Visibility under ultrasound is enhanced by CornerStone reflectors as cleared by FDA in several Premarket Notification Submissions.

The SonoPlex STIM and SonoPlex II needle is equipped with an connecting cable, injection tube and NanoLine® coating (Parylene, no nanomaterials incorporated). The distal connection of the tube is either equipped with a LUER Connector according to ISO 80369-7 or a NRFit-Connector according to ISO 80369-6.

The device in focus is a single use device and has an intended time of use up to 24 hours acc. EO-residuals acc. DIN EN ISO 10993-7. The standard time of use is less than 60 minutes.

The SonoPlex STIM and SonoPlex II cannulas are not for intrathecal use.

The SonoPlex STIM and SonoPlex II cannulas are produced at PAJUNK® GmbH Medizintechnologie in Geisingen, Germany.

2 Determination of Substantial Equivalence

Intended Use Predicate Device (510(K) Predicate)

SonoBlock and SonoBlock II Needles equipped with Cornerstone reflectors are used to puncture the tissue in order to gain entry and inject local anaesthetics to induce regional anaesthesia.

Warning:
SonoBlock and SonoBlock II Needles are not intended for RF ablation or any other type of ablation procedure.

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
7 / 16

Page 12

Intended Use Subject Device

SonoPlex STIM and SonoPlex II Needles equipped with Cornerstone reflectors are used to puncture the tissue in order to gain entry and inject local anesthetics to induce regional anesthesia.

An electrical stimulus may be applied to the needle via a cable and connector to assist the physician pinpoint the area of application.

Discussion

The Intended Use of the Predicate device and of the subject device is substantially equivalent. The Predicate device offers localization via ultrasound as it is state of the art and well established. The Subject Device additionally offers electrical stimulus for localization (Dual Guidance). Both methods are state of the art and substantially equivalent.

Conclusion: substantially equivalent

3 Technical Description

Substantial Equivalence: Tabulatory Outline

CharacteristicsSubject Device SonoPlex STIM Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenSubject Device SonoPlex II Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenPredicate Device K241954 SonoBlock Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenPredicate Device K241954 SonoBlock II Manufacturer: PAJUNK® GmbH Medizintechnologie, Geisingen
Picture
Needle TubingStainless Steel 1.4301Stainless Steel 1.4301Stainless Steel 1.4301Stainless Steel 1.4301
TipTuohy, Quincke, Chiba, FacetteTuohy, Quincke, Chiba, FacetteTuohy, Quincke, Chiba, FacetteTuohy, Quincke, Chiba, Facette

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
8 / 16

Page 13

CharacteristicsSubject Device SonoPlex STIM Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenSubject Device SonoPlex II Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenPredicate Device K241954 SonoBlock Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenPredicate Device K241954 SonoBlock II Manufacturer: PAJUNK® GmbH Medizintechnologie, Geisingen
HubCopolyester (Tritan MX731) (Indirect Patient Contact)Lustran/Guardian ABS 348 white 012002 (No Patient Contact)Copolyester (Tritan MX731) (Indirect Patient Contact)Lustran/Guardian ABS 348 white 012002 (No Patient Contact)
CoatingParylene (NanoLine®), Silicone (Dow Corning 360 Medical Fluid 12500 cSt), Silicone thinner (Dow Corning Q7-9180 Silicone Fluid 1.0 cSt)Parylene (NanoLine®), Silicone (Dow Corning 360 Medical Fluid 12500 cSt), Silicone thinner (Dow Corning Q7-9180 Silicone Fluid 1.0 cSt)Parylene (NanoLine®), Silicone (Dow Corning 360 Medical Fluid 12500 cSt), Silicone thinner (Dow Corning Q7-9180 Silicone Fluid 1.0 cSt)Parylene (NanoLine®), Silicone (Dow Corning 360 Medical Fluid 12500 cSt), Silicone thinner (Dow Corning Q7-9180 Silicone Fluid 1.0 cSt)
US enhancementCornerstonesCornerstoneCornerstoneCornerstone
GraduationTampapur TPUTampapur TPUTampapur TPUTampapur TPU
Bonding Technology Needle-to-HubgluedDirectly moldedgluedDirectly molded
Glue(Epoxy adhesive - Araldite 2011), alternatively: UV adhesive (Loctite AA 3921)n.a.(Epoxy adhesive - Araldite 2011), alternatively: UV adhesive (Loctite AA 3921)n.a.
Injection tubePVC (MED7536), PC (HP4)PVC (MED7536), PC (HP4)PVC (MED7536), PC (HP4), ABS (Terlux 2802)PVC (MED7536), PC (HP4)

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
9 / 16

Page 14

CharacteristicsSubject Device SonoPlex STIM Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenSubject Device SonoPlex II Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenPredicate Device K241954 SonoBlock Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenPredicate Device K241954 SonoBlock II Manufacturer: PAJUNK® GmbH Medizintechnologie, Geisingen
HubCopolyester (Tritan MX731) (Indirect Patient Contact)Lustran/Guardian ABS 348 white 012002 (No Patient Contact)Copolyester (Tritan MX731) (Indirect Patient Contact)Lustran/Guardian ABS 348 white 012002 (No Patient Contact)
CoatingParylene (NanoLine®), Silicone (Dow Corning 360 Medical Fluid 12500 cSt), Silicone thinner (Dow Corning Q7-9180 Silicone Fluid 1.0 cSt)Parylene (NanoLine®), Silicone (Dow Corning 360 Medical Fluid 12500 cSt), Silicone thinner (Dow Corning Q7-9180 Silicone Fluid 1.0 cSt)Parylene (NanoLine®), Silicone (Dow Corning 360 Medical Fluid 12500 cSt), Silicone thinner (Dow Corning Q7-9180 Silicone Fluid 1.0 cSt)Parylene (NanoLine®), Silicone (Dow Corning 360 Medical Fluid 12500 cSt), Silicone thinner (Dow Corning Q7-9180 Silicone Fluid 1.0 cSt)
US-enhancementCornerstonesCornerstoneCornerstoneCornerstone
GraduationTampapur TPUTampapur TPUTampapur TPUTampapur TPU
Bonding Technology Needle-to-HubgluedDirectly moldedgluedDirectly molded
Glue(Epoxy adhesive - Araldite 2011), alternatively: UV adhesive (Loctite AA 3921)n.a.(Epoxy adhesive - Araldite 2011), alternatively: UV adhesive (Loctite AA 3921)n.a.
Injection tubePVC (MED7536), PC (HP4)PVC (MED7536), PC (HP4)PVC (MED7536), PC (HP4), ABS (Terlux 2802)PVC (MED7536), PC (HP4)

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
10 / 16

Page 15

CharacteristicsSubject Device SonoPlex STIM Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenSubject Device SonoPlex II Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenPredicate Device K241954 SonoBlock Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenPredicate Device K241954 SonoBlock II Manufacturer: PAJUNK® GmbH Medizintechnologie, Geisingen
CablePVC, Copper tin-plated, brass tin-plated, ABS (Guardian 348 - RAL9003) or Brass tin-plated, Copper tin-plated, cable (tinned wire), ABS PA-757 RAL 9003 / RAL1018PVC, Copper tin-plated, brass tin-plated, ABS (Guardian 348 - RAL9003) or Brass tin-plated, Copper tin-plated, cable (tinned wire), ABS PA-757 RAL 9003 / RAL1018n.a.n.a.
Diameter20G - 25G20G - 25G20G - 25G20G - 25G
Length40mm - 120 mm40mm - 120 mm40mm - 120 mm40mm - 120 mm
ConnectivityISO 80369-7 (LUER) ISO 80369-6 (NRFit)ISO 80369-7 (LUER) ISO 80369-6 (NRFit)ISO 80369-7 (LUER) ISO 80369-6 (NRFit)ISO 80369-7 (LUER)
PackagingTyvek, Foil, chromo-duplex board Alternatively: Paper, FoilTyvek, Foil, chromo-duplex board Alternatively: Paper, FoilTyvek, Foil, chromo-duplex board Alternatively: Paper, FoilTyvek, Foil, chromo-duplex board Alternatively: Paper, Foil
BiocompatibilityComplies with ISO10993-series as required by « Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" », Docket Number : FDA-2013-D-0350
Sterilization MethodEthyleneoxideEthyleneoxideEthyleneoxideEthyleneoxide
Sterility Assurance LevelSAL=10-6SAL=10-6SAL=10-6SAL=10-6

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
11 / 16

Page 16

CharacteristicsSubject Device SonoPlex STIM Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenSubject Device SonoPlex II Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenPredicate Device K241954 SonoBlock Manufacturer: PAJUNK® GmbH Medizintechnologie, GeisingenPredicate Device K241954 SonoBlock II Manufacturer: PAJUNK® GmbH Medizintechnologie, Geisingen
Sterilization Service ProviderSterigenics, Wiesbaden HA2 Medizintechnik, HalberstadtSterigenics, Wiesbaden HA2 Medizintechnik, HalberstadtSterigenics, Wiesbaden HA2 Medizintechnik, HalberstadtSterigenics, Wiesbaden HA2 Medizintechnik, Halberstadt
Shelf Life60 month from sterilization60 month from sterilization60 month from sterilization60 month from sterilization

4 Substantial Equivalence Discussion

The Subject devices, the SonoPlex STIM and the SonoPlex II needles, are equipped with a cable for additionally pinpointing the needles position via an electrical stimulus called dual guidance (ultrasound and electrical stimulation). This is why a cable is attached to the hub in addition. The SonoBlock and SonoBlock II needles are equipped with ultrasound enhancing cornerstones without any cable since there is no stimulation.

Both, ultrasound position guidance as well as ultrasound position guidance and electrical stimulation are state of the art. Both techniques are safe and effective.

Both, the subject device as well as the predicate device can be either directly injection moulded or glued. Both manufacturing technologies are safe and the processes are running stable. Both technologies are standard technologies.

The predicate devices, the SonoPlex needles, are packed with Tyvek and foil and sterilized with Ethyleneoxide at Sterigenics, Germany.

The subject devices, the SonoPlex needles, are packed with Tyvek and foil as well as optionally with Medical Paper and foil.

Furthermore, the subject devices are sterilized at Sterigenics in Wiesbaden as well as optionally at HA2 Medizintechnik in Halberstadt.

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
12 / 16

Page 17

The main differences which are subject to this premarket submission are

a) the addition of a sterilization facility and
b) the addition of an alternative packaging material.

In order to verify substantial equivalence of the sterilization process a sterilization validation according ISO11135 has been performed and successfully accomplished. The sterility assurance level of 10-6 has been successfully validated at both facilities: Sterigenics and HA2.

In order to verify substantial equivalence of the packaging exhaustive packaging and shelf life tests have been performed. Both, medical paper and Tyvek have been validated and verified to allow sterilization using EtO and to maintain sterility after a shelf life of 60 month after sterilization.

5 Performance Testing

The needles subject to this premarket submission are subject to performance testing prior to release as well as after shelf life scenario of 5 years.

There is no impact of the change subject to this submission (additional optional sterilization provider, additional optional packaging) on the performance of the needle as defined in the user requirements specification.

Especially the hub to needle bonding requirements remain unaffacted. The processes are still valid and reliable.

The test listed below are standard tests performed on a regular basis as well as type tests performed after design transfer.

Tests are performed to comply with the international standards listed below in the respectively most current version:

  • ISO 9626 Stainless steel needle tubing for manufacture of medical devices
  • ISO 7864 Sterile hypodermic needles for single use - Requirements and test methods
  • ISO 80369-6 Small bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications
  • ISO 80369-7 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
  • ISO 80369-20 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
13 / 16

Page 18

5.1 ISO 7864

The following sections have been tested:

sectionPass / Fail
4.3 CleanlinessPassed
4.4 Limits for acidity or alkalinityPassed
4.5 Limits for extractable metalsPassed
4.10 Needle TubePassed
4.11 Needle pointPassed
4.12 Bond between hub and needle tubePassed
4.13 Patency of lumenPassed

5.2 ISO 9626

The following sections have been tested:

sectionPass / Fail
5.2 Surface finish and visual appearancePassed
5.3 CleanlinessPassed
5.4 Limits for acidity and alkalinityPassed
5.5 Size designationPassed
5.6 DimensionsPassed
5.7 Sample sizePassed
5.8 StiffnessPassed
5.9 Resistance to breakagePassed
5.10 Resistance to corrosionPassed

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
14 / 16

Page 19

5.3 ISO 80369

The following sections of the 80369-6 have been tested:

sectionPass / Fail
6.1 Fluid LeakagePassed
6.2 Air LeakagePassed
6.3 Stress CrackingPassed
6.4 Separation Axial LoadPassed
6.5 UnscrewingPassed
6.6 OverridingPassed

The following sections of the 80369-7 have been tested:

sectionPass / Fail
7.1 Fluid LeakagePassed
7.2 Air LeakagePassed
7.3 Stress CrackingPassed
7.4 Separation Axial LoadPassed
7.5 UnscrewingPassed
7.6 OverridingPassed

6 Sterilization

The contract sterilizer and the sterilizing process are identical to the process and sterilizer used for all PAJUNK® - manufactured devices which are already cleared for market or exempt.

Sterilization parameters are:

ParameterValue
SAL10-6
Type of gasEthylene Oxide 99,99%

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
15 / 16

Page 20

Sterilization has been validated according to ISO 11135-1 Overkill Approach (1 sublethal cycle, 2 half cycle, 1 full cycle)

Residuals of EO and ECH are in compliance with ISO 10993-7.

Cleaning and Sterilization method, which ensures an SAL of 10-6 as well as compliance with limits for chemical burden, bioburden, pyroburden (i.e. LAL) and EtO-residuals as well as shelf life have been validated and are safe and effective.

The limits listed below are met by each device:

Limits for Residuals: 25ppm = 25µg/(g/device) of Ehyleneoxide (EO); 25ppm = 25µg/(g/device) Ethylene chlorhydrine

The SonoPlex STIM and SonoPlex II are also available in bulk non sterile. If appropriately packed and sterilized with Ethylene oxide according to the given sterilization parameters the technological parameters remain unchanged. However, final responsibility for sterilization validation remains with the customer of spinal manometers purchased bulk non-sterile.

7 Shelf Life

Efficacy of sterile product´s lifecycle has been validated using process most challenging worst case devices.

Sterility tests have been performed using process most challenging worst case devices with similar characteristics made from the same material after 5 years. The devices were found to be sterile after 5 years, the sterile barrier system is efficient.

Performance of the device (LUER connection, stability of bonding connections, needle´s bending rigidity) has been tested. There is no decrease in performance after 5 years.

Shelf-life is set to 5 years.

8 Biocompatibility

All products comply with ISO 10993-1 and with FDA´s guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

Therefore, based on sterilization validation and residuals validation the kits also are considered to be biocompatible.

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
16 / 16

Page 21

9 Technology Characteristics

Both, the subject device and the predicate device, consist of identical components.

10 Conclusion

The comparison between the predicate device and the subject device of this submission focussed on the validated sterilization process and the shelf life testing demonstrates that the subject device can be optionally in addition packed using medical paper and can be optionally sterilized at the alternative sterilization service provider at an SAL of 10-6.

The results demonstrate that the subject device is substantially equivalent and as safe, as effective, and performs as well as the predicate device.

510(k) Premarket Notification Submission

510(k) SonoPlex STIM and SonoPlex II

R&D Resp. SCJ
Regulatory Resp. CQ/ MCH
16 / 16

§ 868.5150 Anesthesia conduction needle.

(a)
Identification. An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.(b)
Classification. Class II (performance standards).