K Number
K243473

Validate with FDA (Live)

Date Cleared
2025-07-29

(263 days)

Product Code
Regulation Number
892.1720
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PRORAD ATLAS ULTRAPORTABLE Digital X-ray system is intended to deliver high-quality, diagnostic radiographic images of the body extremities. It utilizes a portable X-ray unit, flat-panel detector and image acquisition software to produce clear digital images, enabling fast and accurate diagnosis. The portable X-ray unit is intended to be used only when stand/tripod mounted.

The PRORAD ATLAS ULTRAPORTABLE X-ray digital system is predominantly employed in various settings, including health-care centres, temporary and emergency health centres (established, especially in pandemic circumstances), outreach and field interventions (such as mobile clinics/vans, screening campaigns, and home care), and tele-radiology solutions in remote areas.

The primary users anticipated for the system include radiographers, radiological technologists, and medical professionals who are trained in safety, radiation protection, and image management.

The PRORAD ATLAS ULTRAPORTABLE PLUS Digital X-ray system is intended to deliver high-quality, diagnostic radiographic images of the body extremities. It utilizes a portable X-ray Unit, flat-panel detector and real-time image processing using software to produce clear digital images, enabling fast and accurate diagnosis. The portable X-ray unit is intended to be used only when stand/tripod mounted.

The PRORAD ATLAS ULTRAPORTABLE PLUS X-ray digital system is predominantly employed in various settings, including health-care centres, temporary and emergency health centres (established, especially in pandemic circumstances), outreach and field interventions (such as mobile clinics/vans, screening campaigns, and home care), and tele-radiology solutions in remote areas.

The primary users anticipated for the system include radiographers, radiological technologists, and medical professionals who are trained in safety, radiation protection, and image management.

Device Description

The PRORAD ATLAS X-Ray system includes the ULTRAPORTABLE and ULTRAPORTABLE PLUS, which are portable diagnostic X-ray systems with fixed 70kV and 2mA tube current. These systems are intended to produce anatomical X-rays of the body extremities in both pediatric and adult patients. The PRORAD ATLAS X-Ray system was designed, developed, and manufactured by Prognosys Medical Systems Private Limited. The model numbers are listed below.

Model NumberModel NameModel Description
A86-ATL-0001PRORAD ATLAS ULTRAPORTABLEThese are ultraportable X-ray systems designed to generate diagnostic, high-quality X-ray images
A87-ATL-0001The PRORAD ATLAS ULTRAPORTABLE PLUS

The PRORAD ATLAS X-ray system is a sophisticated, battery-powered X-ray generator offered in two versions: PRORAD ATLAS ULTRAPORTABLE and ULTRAPORTABLE PLUS. The main distinction between these models lies in their exposure time ranges and target anatomical areas. The ULTRAPORTABLE model provides exposure times ranging from 0.01 to 1.30 seconds, while the ULTRAPORTABLE PLUS model offers an extended exposure range of 0.01 to 2.5 seconds. Both models share identical internal components, software, algorithms, and operational features and are intended for imaging body extremities. The system includes a high-voltage tank with an X-ray tube mounted on an adjustable tripod stand, allowing users to adjust the height to the specific imaging area. Exposure parameters are configured through the X-ray generator's graphical user interface (GUI). After setting the parameters and positioning the patient on the detector, the X-ray is activated via an exposure switch. The detector captures the radiation, converts it into a digital signal, and transmits the data wirelessly to a computer equipped with compatible software. The images are processed and displayed on the computer for diagnostic review. The PRORAD ATLAS system is compatible with several 510(k)-cleared detectors and their associated software, listed below in Table 1. Prognosys includes one detector and its pre-configured software in the package, depending on availability. Fully battery-operated, the system does not support direct power connection but can seamlessly integrate with multiple detectors and compatible software as part of the package.

AI/ML Overview

The provided FDA 510(k) clearance letter and supporting documentation for the PRORAD ATLAS ULTRAPORTABLE X-Ray Systems do not include acceptance criteria or a detailed study that proves the device meets specific performance criteria beyond general safety and effectiveness.

The document primarily focuses on demonstrating substantial equivalence to a predicate device (Remex KA6, K212144) rather than presenting a performance study with defined acceptance criteria. The "Summary of non-clinical testing" lists a series of international standards (IEC, ISO) and FDA guidance documents that were followed for design control, risk management, verification, and validation. The "Summary of clinical testing" mentions that clinical images were collected and reviewed by a qualified radiologist, confirming they are "clinically acceptable." However, specific quantitative acceptance criteria for image quality, diagnostic accuracy, or other performance metrics, along with the study design and results against those criteria, are not detailed in this document.

Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor can I provide information about sample size, expert details, adjudication methods, MRMC studies, standalone performance, or training set specifics, as this information is not present in the provided text.

Based on the available text, here's what can be extracted:

  • Overall Conclusion: The device is deemed "safe and effective when the device is used as labelled and is substantially equivalent to the predicate device."

Here's a breakdown of why the requested information cannot be fully provided based on the input:

  1. A table of acceptance criteria and the reported device performance: This information is not explicitly stated in the document. The document confirms that "Validation of PRORAD ATLAS X-Ray System has demonstrated that the system enables optimal and quality imaging of anatomical structures" and that clinical images are "clinically acceptable," but no specific quantitative criteria or performance metrics are given.

  2. Sample size used for the test set and the data provenance: The document states that "Clinical images of body extremities were collected from patients of varying ages, weights, and BMIs." However, the exact sample size and the provenance (e.g., country of origin, retrospective/prospective nature) of this clinical image test set are not specified.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: The document mentions that images were "reviewed by a qualified radiologist." It does not specify the number of radiologists or their specific qualifications (e.g., years of experience, board certification).

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: No adjudication method for the clinical image review is described.

  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: An MRMC study is not mentioned. The device described is an X-ray system, not an AI software to assist human readers.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: This is not applicable as the device is an X-ray system, not an algorithm, and it's intended to be used by trained medical professionals.

  7. The type of ground truth used: The ground truth for the clinical images appears to be "clinical acceptability" as determined by a "qualified radiologist." This aligns with "expert consensus" in a general sense, but no more objective ground truth (e.g., pathology, outcomes data) is mentioned for the image quality assessment.

  8. The sample size for the training set: The document does not mention a training set, as it describes an X-ray hardware system, not an AI-driven software that requires a training set.

  9. How the ground truth for the training set was established: Not applicable, as no training set is mentioned.

FDA 510(k) Clearance Letter - PRORAD ATLAS ULTRAPORTABLE X-Ray Systems

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

July 29, 2025

Prognosys Medical Systems Private Limited
℅ Ankur Naik
Managing Director
IZiel Healthcare
Pentagon P1, Office No. 601 and 604, Magarpatta City,
Hadapsar
Pune, Maharashtra 411028
INDIA

Re: K243473
Trade/Device Name: Prorad Atlas Ultraportable
Prorad Atlas Ultraportable Plus
Regulation Number: 21 CFR 892.1720
Regulation Name: Mobile X-Ray System
Regulatory Class: Class II
Product Code: IZL
Dated: June 26, 2025
Received: June 26, 2025

Dear Ankur Naik:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

K243473 - Ankur Naik Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

Page 3

K243473 - Ankur Naik Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243473

Device Name:
PRORAD ATLAS ULTRAPORTABLE
PRORAD ATLAS ULTRAPORTABLE PLUS

Indications for Use (Describe)

The PRORAD ATLAS ULTRAPORTABLE Digital X-ray system is intended to deliver high-quality, diagnostic radiographic images of the body extremities. It utilizes a portable X-ray unit, flat-panel detector and image acquisition software to produce clear digital images, enabling fast and accurate diagnosis. The portable X-ray unit is intended to be used only when stand/tripod mounted.

The PRORAD ATLAS ULTRAPORTABLE X-ray digital system is predominantly employed in various settings, including health-care centres, temporary and emergency health centres (established, especially in pandemic circumstances), outreach and field interventions (such as mobile clinics/vans, screening campaigns, and home care), and tele-radiology solutions in remote areas.

The primary users anticipated for the system include radiographers, radiological technologists, and medical professionals who are trained in safety, radiation protection, and image management.

The PRORAD ATLAS ULTRAPORTABLE PLUS Digital X-ray system is intended to deliver high-quality, diagnostic radiographic images of the body extremities. It utilizes a portable X-ray Unit, flat-panel detector and real-time image processing using software to produce clear digital images, enabling fast and accurate diagnosis. The portable X-ray unit is intended to be used only when stand/tripod mounted.

The PRORAD ATLAS ULTRAPORTABLE PLUS X-ray digital system is predominantly employed in various settings, including health-care centres, temporary and emergency health centres (established, especially in pandemic circumstances), outreach and field interventions (such as mobile clinics/vans, screening campaigns, and home care), and tele-radiology solutions in remote areas.

The primary users anticipated for the system include radiographers, radiological technologists, and medical professionals who are trained in safety, radiation protection, and image management.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

510(k) summary of safety and effectiveness for PRORAD ATLAS X-Ray system is provided in accordance with 21 CFR 807.92.

Date: 29 July 2025

Submitter (Owner): Krishna Prasad Vajapeyam
Founder and President
Prognosys Medical Systems Private Limited.
#168/1, Machohalli, Dasanapura Hobli, next to Vani Vidhyalaya, Off. Magadi Road, Bangalore - 560 091, India
P: +91 9845290091
Email: krishna.prasad@prognosysmedical.com

510(k) Contact Person: Ankur Naik
Managing Director
IZiel Healthcare
Pentagon P1, Office No. 601 and 604, Magarpatta City, Hadapsar, Pune 411028, Maharashtra, India
P: +91 72762 2555 M: +91 7069553814
Email: ankur.naik@izielhealthcare.com

Device Trade Name: PRORAD ATLAS ULTRAPORTABLE
PRORAD ATLAS ULTRAPORTABLE PLUS

Regulation Number: 21 CFR 892.1720
Review Panel: Radiology
Device Class: Class II
Product Code: IZL
Predicate Device: Remex KA6 (K212144)

Device Description

The PRORAD ATLAS X-Ray system includes the ULTRAPORTABLE and ULTRAPORTABLE PLUS, which are portable diagnostic X-ray systems with fixed 70kV and 2mA tube current. These systems are intended to produce anatomical X-rays of the body extremities in both pediatric and adult patients. The PRORAD ATLAS X-Ray system was designed, developed, and manufactured by Prognosys Medical Systems Private Limited. The model numbers are listed below.

Model NumberModel NameModel Description
A86-ATL-0001PRORAD ATLAS ULTRAPORTABLEThese are ultraportable X-ray systems designed to generate diagnostic, high-quality X-ray images
A87-ATL-0001The PRORAD ATLAS ULTRAPORTABLE PLUS

K243473

Page 6

The PRORAD ATLAS X-ray system is a sophisticated, battery-powered X-ray generator offered in two versions: PRORAD ATLAS ULTRAPORTABLE and ULTRAPORTABLE PLUS. The main distinction between these models lies in their exposure time ranges and target anatomical areas. The ULTRAPORTABLE model provides exposure times ranging from 0.01 to 1.30 seconds, while the ULTRAPORTABLE PLUS model offers an extended exposure range of 0.01 to 2.5 seconds. Both models share identical internal components, software, algorithms, and operational features and are intended for imaging body extremities. The system includes a high-voltage tank with an X-ray tube mounted on an adjustable tripod stand, allowing users to adjust the height to the specific imaging area. Exposure parameters are configured through the X-ray generator's graphical user interface (GUI). After setting the parameters and positioning the patient on the detector, the X-ray is activated via an exposure switch. The detector captures the radiation, converts it into a digital signal, and transmits the data wirelessly to a computer equipped with compatible software. The images are processed and displayed on the computer for diagnostic review. The PRORAD ATLAS system is compatible with several 510(k)-cleared detectors and their associated software, listed below in Table 1. Prognosys includes one detector and its pre-configured software in the package, depending on availability. Fully battery-operated, the system does not support direct power connection but can seamlessly integrate with multiple detectors and compatible software as part of the package.

Table 1: List of flat-panel detectors and compatible software

S. noFlat panel DetectorsSoftwareManufacturer/Supplier Name510(k) Number
1.Yushan X-ray Flat Panel detector with DROC Model name: V17CeDROCInnoCare Optoelectronics Corp.K220510
2.Yushan X-ray Flat Panel detector with DROC Model name: F14C, F14GDROCInnoCare Optoelectronics Corp.K210988
3.Yushan X-ray Flat Panel detector with DROC Model Name: V14C, V14G, V17C, V17G, V17GeDROCInnoCare Optoelectronics Corp.K201528
4.DIGITAL RADIOGRAPHY CXDI Control SoftwareCanon Inc.K213780

Page 7

S. noFlat panel DetectorsSoftwareManufacturer/Supplier Name510(k) Number
5.DIGITAL RADIOGRAPHY CXDI-CS01CXDI Control SoftwareCanon Inc.K230175
6.Flat panel Digital X-ray detector 14HQ721G-B14HQ721G-B SoftwareLG Electronics Inc.K221394
7.Flat panel Digital X-ray Detector 17HQ901G-B17HQ901G-B SoftwareLG Electronics Inc.K214044
8.Flat panel Digital X-ray Detector 14HK701G-W14HK701G-W SoftwareLG Electronics Inc.K182348
9.Mars1717X Wireless Digital Flat Panel DetectorMars1717X iRay SDKiRay Technology Taicang Ltd.K210314
10.Wireless Digital Flat Panel Detector Mars1417XiRay SDKiRay Technology Taicang Ltd.K210316
11.Nexus DR TM Digital X-ray Imaging System (with PaxScan 4343RC and PaxScan 4343Rv3)Nexus DR TM Digital X-ray Imaging System -Digital Imaging softwareVarex Imaging CorporationK172951
12.XRpad2 4343 HWC-M Flat Panel DetectorXRpad2 4343 HWC-M Flat Panel DetectorVarex Imaging CorporationK181526
13PRORAD X-ray Flat Panel detector with DROC Model Name: V14 Clarity ,V14 HC ,V17 Clarity ,V17 HC,W17 HC ,W17 Clarity ,F14 Clarity ,F14 HCDROCPRORAD.K240771

Page 8

Intended Use / Indications for Use

PRORAD ATLAS ULTRAPORTABLE

The PRORAD ATLAS ULTRAPORTABLE Digital X-ray system is intended to deliver high-quality, diagnostic radiographic images of the body extremities. It utilizes a portable X-ray unit, flat-panel detector and image acquisition software to produce clear digital images, enabling fast and accurate diagnosis. The portable X-ray unit is intended to be used only when stand/tripod mounted.

The PRORAD ATLAS ULTRAPORTABLE X-ray digital system is predominantly employed in various settings, including health-care centres, temporary and emergency health centres (established, especially in pandemic circumstances), outreach and field interventions (such as mobile clinics/vans, screening campaigns, and home care), and tele-radiology solutions in remote areas.

The primary users anticipated for the system include radiographers, radiological technologists, and medical professionals who are trained in safety, radiation protection, and image management.

PRORAD ATLAS ULTRAPORTABLE PLUS

The PRORAD ATLAS ULTRAPORTABLE PLUS Digital X-ray system is intended to deliver high-quality, diagnostic radiographic images of the body extremities. It utilizes a portable X-ray Unit, flat-panel detector and real-time image processing using software to produce clear digital images, enabling fast and accurate diagnosis. The portable X-ray unit is intended to be used only when stand/tripod mounted.

The PRORAD ATLAS ULTRAPORTABLE PLUS X-ray digital system is predominantly employed in various settings, including health-care centres, temporary and emergency health centres (established, especially in pandemic circumstances), outreach and field interventions (such as mobile clinics/vans, screening campaigns, and home care), and tele-radiology solutions in remote areas.

The primary users anticipated for the system include radiographers, radiological technologists, and medical professionals who are trained in safety, radiation protection, and image management.

Substantial Equivalence Discussion

Remex KA6 (K212144) has been selected as the primary predicate device for the substantial equivalence discussion for PRORAD ATLAS X-Ray System. The details regarding the substantial equivalence between the subject device and the predicate device are explained below:

Page 9

Technological Characteristics Comparison

Table 2: Technological Characteristics Comparison

Comparable PropertiesSubject DevicePredicate Device (K212144)Comparison Results
Product NamePRORAD ATLAS X-Ray SystemRemex KA6Not applicable
ManufacturerPrognosys Medical Systems Private LimitedRemedi Co., Ltd.Not applicable
Regulation Number21 CFR 892.172021 CFR 892.1720Identical
Product CodeIZLIZLIdentical
Product ClassIIIIIdentical
Indication for UseThe PRORAD ATLAS ULTRAPORTABLE Digital X-ray system is intended to deliver high-quality, diagnostic radiographic images of the body extremities. It utilizes a portable X-ray unit, flat-panel detector and image acquisition software to produce clear digital images, enabling fast and accurate diagnosis. The portable X-ray unit is intended to be used only when stand/tripod mounted. The PRORAD ATLAS ULTRAPORTABLE X-ray digital system is predominantly employed in various settings, including health-care centres, temporary and emergency health centres (established,The KA6 is a handheld portable X-ray system for diagnostic imaging of body extremities.Identical The indications for use statement of the subject device is elaborated to include the use environment and the primary users of the device. The intended use of both the subject device and the predicate device remains the same and this is demonstrated through the various testing data conducted to support the safety and performance of the subject device.

Page 10

Comparable PropertiesSubject DevicePredicate Device (K212144)Comparison Results
especially in pandemic circumstances), outreach and field interventions (such as mobile clinics/vans, screening campaigns, and home care), and tele-radiology solutions in remote areas. The primary users anticipated for the system include radiographers, radiological technologists, and medical professionals who are trained in safety, radiation protection, and image management. The PRORAD ATLAS ULTRAPORTABLE PLUS Digital X-ray system is intended to deliver high-quality, diagnostic radiographic images of the body extremities. It utilizes a portable X-ray Unit, flat-panel detector and real-time image processing using software to produce clear digital images, enabling fast and accurate diagnosis. The portable X-ray unit is intended to be used only when stand/tripod mounted. The PRORAD ATLAS ULTRAPORTABLE PLUS X-ray digital system is

Page 11

Comparable PropertiesSubject DevicePredicate Device (K212144)Comparison Results
predominantly employed in various settings, including health-care centres, temporary and emergency health centres (established, especially in pandemic circumstances), outreach and field interventions (such as mobile clinics/vans, screening campaigns, and home care), and tele-radiology solutions in remote areas. The primary users anticipated for the system include radiographers, radiological technologists, and medical professionals who are trained in safety, radiation protection, and image management.
Multiple UseYesYesIdentical
Prescription (Rx only)YesYesIdentical
Anatomical SiteBody extremitiesBody extremitiesIdentical
Intended UserThe main users of the system are expected to be radiographers, radiological technologists and medical professionals trained on safety, radiation protection and image management.Intended for use by a qualified/trained physician or technicianIdentical
Use EnvironmentThe PRORAD ATLAS X-Ray System is primarily used in:Remex KA6 is intended for use in emergency room,Identical

Page 12

Comparable PropertiesSubject DevicePredicate Device (K212144)Comparison Results
Size: BodyHealth-Care Centers; temporary and emergency Health-Centers (also established in Pandemic circumstances); outreach and field interventions (i.e.: mobile clinics/ vans, screening campaigns, home care); tele-radiology solutions in remote areas. 220 x 170 x 135mm (LxWxH)nursing home, NICU, operating room, ICU, negative pressure isolation room (NPIR) 165 x 176 x 255 mmIdentical Different The subject device has been verified and validated. Therefore, the difference in size does not raise any concerns regarding the safety or performance of the device.
Weight2.8 kg (Including Battery)2.4 kg (including cone 190g)Different The subject device has been verified and validated. Therefore, the difference in weight does not raise any concerns regarding the safety or performance of the device.
Energy SourcePRORAD ATLAS ULTRAPORTABLE: Lithium-Ion Polymer Rechargeable Batteries (11.1VDC), 100 exposures per chargeLithium Polymer Rechargeable Battery (22.2 VDC), 300 exposures per chargeDifferent Due to the difference in battery capacity between the subject and predicate devices,

Page 13

Comparable PropertiesSubject DevicePredicate Device (K212144)Comparison Results
PRORAD ATLAS ULTRAPORTABLE PLUS: Lithium-Ion Polymer Rechargeable Batteries (22.2VDC), 100 exposures per chargethere is a variation in the number of exposures. However, this does not impact the safety or performance of the device.
Mounting methodUnit is usually mounted on tripod/standUnit is usually handheld, or mounted on optional arms or tripodIdentical
User InterfaceUp-Down pushbuttons for selection of exposure time, increase and decrease with LED indicators, laser ON/OFFUp-Down pushbuttons for selection of exposure time, kVp, and mA, with LED indicatorsDifferent The differences in the user interface do not raise any new concerns of safety or performance and this has been demonstrated through relevant validation.
Exposure switchDual-Stage, Exposure Hand switchDual-stage, Deadman typeIdentical
SoftwarePRORAD ATLAS ULTRAPORTABLE SoftwareBuilt-in Software - S/W name: RPG-F-0706 - S/W version: 1.00Different The PRORAD ATLAS X-Ray System uses its own software, which has been verified and validated. Therefore, this does not raise any concerns regarding the safety and performance of the device.
Exposure timesPRORAD ATLAS ULTRAPORTABLE: 0.01-0.1 sec (0.01 sec steps) 0.1-0.5 sec(0.02 sec steps) 0.5-1.0 sec(0.05 sec steps) 1.00-1.30 sec(0.1 sec steps)0.06 -2.0 sec (0.01 sec. steps)Different The difference in the exposure does not change the intended application. As per the standard of care and

Page 14

Comparable PropertiesSubject DevicePredicate Device (K212144)Comparison Results
PRORAD ATLAS ULTRAPORTABLE PLUS: 0.05-0.10 sec(0.01 sec steps) 0.10-0.50 sec(0.02 sec steps) 0.50-1.00 sec(0.05 sec steps) 1.00-1.30 sec(0.1 sec steps) 1.30-2.00 sec(0.2 sec steps)guidelines, the user can set the appropriate exposure time required for the examination. Therefore, this does not raise any concerns regarding the safety and performance of the device.
Tube potential (kV)70 kVp40 - 70 kVp (1 kVp steps)Similar
Tube current (mA)PRORAD ATLAS ULTRAPORTABLE: 2 mA PRORAD ATLAS ULTRAPORTABLE PLUS: 6 mA2 mA - 6 mA (1 mA steps)Similar
Alerts and AlarmsAt least two specific alerts and alarms indicate equipment state of readiness to useAt least three specific alerts and alarms indicate equipment state of readiness to use.Different The devices are designed to ensure safe operation for the operator and the patient. Verification and validation activities have been carried out to demonstrate that the subject performs as intended and is safe.
X-ray TubePRORAD ATLAS ULTRAPORTABLE: RADII KL11-0.4-70 Series PRORAD ATLAS ULTRAPORTABLE PLUS: OX-80Canon D-041SBDifferent The subject device utilizes a different x-ray tube when compared to the predicate device. The device has been tested, and it has been demonstrated that the device performs as intended and is safe. Therefore, this does not

Page 15

Comparable PropertiesSubject DevicePredicate Device (K212144)Comparison Results
raise any concerns regarding the safety and performance of the device
Focal spot sizePRORAD ATLAS ULTRAPORTABLE: 0.4 mm PRORAD ATLAS ULTRAPORTABLE PLUS: 0.8 mm0.4 mmDifferent The focal spot size of the PRORAD ATLAS ULTRAPORTABLE is identical to that of the predicate device, and the focal spot size of the PRORAD ATLAS ULTRAPORTABLE PLUS falls outside of the predicate device. The device has been tested, and it has been demonstrated that the device performs as intended and is safe. Therefore, this does not raise any concern regarding the safety and performance of the device.
Total filtration1.5 mm AL Eq1.6 mm Al equivalentDifferent The subject device has been validated and tested to demonstrate that the difference in total filtration does not raise any new questions of safety or performance of the device.
Image receptorDigital radiographyDigital radiographyIdentical
CollimatorBeam limiting device with crosshair laser lightSquare collimator with LED Light Field Center IndicatorDifferent

Page 16

Comparable PropertiesSubject DevicePredicate Device (K212144)Comparison Results
The principle of operation and usage of the collimator in the subject device and predicate device remains the same. The subject device has been validated and tested to demonstrate that the different collimator does not raise any new questions of safety or performance of the device.
Triggering mechanismTwo stage triggeringTwo stage triggeringIdentical
Performance standardIEC 60601-1-3:2008+AMD1:2013 +AMD2:2021 IEC 60601-2-54:2022IEC 60601-1-3:2013 IEC 60601-2-28:2017 IEC 60601-2-54:2009Identical
Electrical safetyIEC 60601-1:2005 12+AMD1:2012-07+AMD2:20 2008 IEC 60601-1-2:2014+AMD1:2020 IEC 60601-1-3:2008+AMD1:2013 +AMD2:2021 IEC 60601-2-54:2022 IEC 62304:2015 IEC 62366-1:2020IEC 60601-1:2012 IEC 60601-1-2:2014 IEC 60601-1-6:2013 IEC 62304:2006 IEC 62366:2007Identical

Page 17

Discussion of similarities and differences

PRORAD ATLAS X-Ray System is intended for X-ray imaging of human anatomy in adult and paediatric populations. The primary users anticipated for the system include radiographers, radiological technologists, and medical professionals who are trained in safety, radiation protection, and image management. The subject device and predicate device are intended for use on body extremities. The subject device utilizes FDA-cleared X-ray flat panel detectors. The other components of the imaging system, such as the X-Ray generator and X-Ray tube, are either identical or similar to the predicate device. Both the subject and predicate device are powered by a rechargeable battery and operated by push buttons and/or a touch screen on the device itself. There are minor differences in the technical specifications of the subject and predicate devices; however, these differences do not raise any new issues of safety or performance.

Summary of non-clinical testing

The following performance data were provided in support of the substantial equivalence determination:

  1. IEC 60601-1:2005+AMD1:2012+AMD2:2020 CSV Consolidated version Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.

  2. IEC 60601-1-2:2014+AMD1:2020 CSV Consolidated Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

  3. IEC 60601-2-54 Edition 2.0 2022-09 Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy

  4. IEC 60601-1-3 Edition 2.2 2021-01 CONSOLIDATED VERSION Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment

  5. IEC 62366-1:2015 Medical devices — Part 1: Application of usability engineering to medical devices.

  6. IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device software - Software life cycle processes.

  7. ISO 14971:2019 Medical devices — Application of risk management to medical devices.

  8. ISO 15223-1 Fourth edition 2021-07 Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements.

  9. ISO 20417 First edition 2021-04 Corrected version 2021-12 Medical devices — Information to be supplied by the manufacturer.

Page 18

  1. ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

  2. Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions | FDA (Document issued on June 27, 2025).

  3. Content of Premarket Submissions for Device Software Functions | FDA (Document issued on June 14, 2023).

The devices have been developed as per 21 CFR 820 and have completed design control activities, including risk management, verification, and validation.

Validation of PRORAD ATLAS X-Ray System has demonstrated that the system enables optimal and quality imaging of anatomical structures.

Summary of clinical testing

Clinical images of body extremities were collected from patients of varying ages, weights, and BMIs. These images were reviewed by a qualified radiologist, who confirmed that the images obtained using the PRORAD ATLAS ULTRAPORTABLE and PRORAD ATLAS ULTRAPORTABLE PLUS are clinically acceptable.

Conclusion

All the above details collectively demonstrate that PRORAD ATLAS X-Ray System is safe and effective when the device is used as labelled and is substantially equivalent to the predicate device.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.