K Number
K181526

Validate with FDA (Live)

Date Cleared
2018-07-03

(22 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The XRpad2 4343 HWC-M, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.

Device Description

The XRpad2 4343 HWC-M is a wireless, lightweight, cassette-sized, flat panel X-ray detector for digital radiography. It fits into a conventional table or wall-stand Bucky, just like a film-screen cassette. The X-ray detector consists of an amorphous silicon flat panel with a directly deposited Csl:Tl scintillator and dedicated read-out, scan, and control electronics, all packaged in a carbon-fiber and aluminum enclosure. The outside dimensions of the detector are 460 mm × 460 mm × 15.5 mm, which fits into a standard X-ray cassette Bucky.

The detector can be integrated with an X-ray system to enable digital radiography.

The XRpad2 4343 HWC-M device, like the predicate, relies on the X-ray Imaging Software Library (XISL), to control the detector from the host computer. The XISL is a C/C++-coded software library that provides parameterized functions to interface the device to a host work station via an IP Network connection. By use of the application programming interface (API) of XISL one can acquire images and configure all use functions of the XRpad2 4343 HWC-M detector.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the device meets the acceptance criteria for the XRpad2 4343 HWC-M Flat Panel Detector.

The acceptance criteria are established by demonstrating substantial equivalence to a predicate device (XRpad2 4336 HWC-M) through non-clinical testing. The key acceptance criteria hinge on the comparability of critical image quality performance attributes and compliance with international and FDA recognized consensus standards.

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance (XRpad2 4343 HWC-M)
Intended UseSame as predicate (diagnostic X-ray procedures, non-mammographic)Same as predicate: "generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use."
Technology/MaterialsSimilar to predicateUtilizes similar technology and materials (amorphous silicon flat panel with CsI:Tl scintillator).
Design/ConstructionSimilar to predicate (slight dimensional change)Similar in design and construction. Primary difference is a 76 mm increase in width, which "does not impact image quality."
Image QualityComparable to predicate (MTF, DQE)"This testing demonstrated that critical image quality performance attributes, such as MTF and DQE, of the XRpad2 4343 HWC-M are comparable to the predicate device."
Operational StandardsCompliance with international and FDA recognized consensus standardsComplies with IEC and ISO standards and regulations related to medical safety, EMC, and bio-compatibility. Followed guidance for 510(k) for Solid State X-ray Imaging Devices.
Safety and EffectivenessSubstantially equivalent to predicate"Based on the information supplied in this 510(k) Varex Imaging concludes, the XRpad2 4343 HWC-M is substantially equivalent to the currently marketed device, XRpad2 4336 HWC-M (K161966) in terms of safety and effectiveness."
Risk MitigationIdentified risks mitigated and acceptedRisks and hazardous impacts analyzed by FMEA methodology. Specific risk control and protective measures reviewed. "Overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "A clinical study was not required for the XRpad2 4343 HWC-M device." Therefore, there is no test set of clinical images nor data provenance in the traditional sense for evaluation. The "test set" for proving substantial equivalence was the device itself undergoing non-clinical technical testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical study or human expert evaluation was used to establish ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical study or human expert evaluation was conducted.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable, as no clinical study, MRMC study, or AI assistance is mentioned. This device is an X-ray detector, not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The study was a non-clinical technical evaluation of the flat panel detector's physical and technical performance attributes like MTF and DQE, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" for the non-clinical testing was established by technical measurements and compliance with recognized industry standards (IEC, ISO). For image quality attributes like MTF (Modulation Transfer Function) and DQE (Detective Quantum Efficiency), the 'ground truth' is the quantitative measurement of these physical properties as defined by relevant standards.

8. The sample size for the training set

Not applicable, as this is a physical medical device (X-ray detector), not a machine learning model. There is no concept of a "training set" for this type of device submission.

9. How the ground truth for the training set was established

Not applicable, as there is no training set mentioned.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 3, 2018

Varex Imaging Corporation % Mr. Paul Sawicki Director, Quality Assurance and Regulatory Affairs 2175 Mission College Blvd. SANTA CLARA CA 95054

Re: K181526

Trade/Device Name: XRpad2 4343 HWC-M Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: June 7, 2018 Received: June 11, 2018

Dear Mr. Sawicki:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Mr. Paul Sawicki

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Hsl 2. Mils

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K181526

Device Name

XRpad2 4343 HWC-M

Indications for Use (Describe)

The XRpad2 4343 HWC-M, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a logo for Varex Imaging. The logo consists of the letter V, which is made up of several vertical lines in different shades of green and blue. To the right of the V is the word "VAREX" in blue, and below that is the word "IMAGING" in a smaller font.

510(K) SUMMARY

XRpad2 4343 HWC-M Flat Panel Detector

This summary of 510(k) safety and effectiveness information is supplied in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K181526

Date: June 27, 2018

Submitted By:Varex Imaging Corporation2175 Mission College Blvd.Santa Clara, CA 95054U.S.A.
Contact Person:Paul SawickiDirector, Quality Assurance and Regulatory AffairsTel. 408-469-6462Fax. 408-969-6493E-mail: paul.sawicki@vareximaging.com
Device Name:XRpad2 4343 HWC-M Flat Panel Detector
Classification:Product Code: MQBClassification Name: Solid state X-ray imager (flat panel/digital imager)Regulation Name: Stationary X-ray systemClassification Regulation: 21 CFR 892.1680
Predicate Device:XRpad 4336 HWC-M Flat Panel Detector510(k) Clearance: K161966; September 8, 2016Product Code: MQBClassification Name: Solid state X-ray imager (flat panel/digital imager)Regulation Name: Stationary X-ray system

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a logo for Varex Imaging. The logo consists of the word "VAREX" in blue, with the word "IMAGING" in smaller letters below it. To the left of the word "VAREX" are several vertical lines in different colors, including blue, green, and yellow. The logo is simple and modern, and the colors are bright and eye-catching.

Device Description:

The XRpad2 4343 HWC-M is a wireless, lightweight, cassette-sized, flat panel X-ray detector for digital radiography. It fits into a conventional table or wall-stand Bucky, just like a film-screen cassette. The X-ray detector consists of an amorphous silicon flat panel with a directly deposited Csl:Tl scintillator and dedicated read-out, scan, and control electronics, all packaged in a carbon-fiber and aluminum enclosure. The outside dimensions of the detector are 460 mm × 460 mm × 15.5 mm, which fits into a standard X-ray cassette Bucky.

The detector can be integrated with an X-ray system to enable digital radiography. The following accessories are available for the XRpad2 4343 HWC-M:

  • XRpad LBC-2 (Lithium Battery Charger)
  • . XRpad IPU-2 (Interface and Power Unit)
  • . XRpad LPT2 Detector Cable
  • . XRpad Protective Insert
  • AC Cable IEC 60320 C13 DE
  • AC Cable IEC 60320 C13 US
  • Trigger Cable 5 m/16.5 ft.
  • . Trigger Cable 20 m/65.5 ft.
  • GigE Interface Cable 7.6 m/25 ft.
  • . GigE Interface Cable 15.25 m/50 ft.

Device Software:

The XRpad2 4343 HWC-M device, like the predicate, relies on the X-ray Imaging Software Library (XISL), to control the detector from the host computer. The XISL is a C/C++-coded software library that provides parameterized functions to interface the device to a host work station via an IP Network connection. By use of the application programming interface (API) of XISL one can acquire images and configure all use functions of the XRpad2 4343 HWC-M detector.

Indications for Use:

The XRpad2 4343 HWC-M, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is the logo for Varex Imaging. The logo has a stylized V made of four lines, with the leftmost line being dark blue, the next two lines being green, and the rightmost line being light green. To the right of the stylized V is the text "VAREX" in dark blue, and below that is the text "IMAGING" in a smaller font, also in dark blue.

Comparison Chart:

ModelProposed Model
CharacteristicsXRpad2 4336 HWC-M(K161966)XRpad2 4343 HWC-M
Intended Use /Indications for UseThe XRpad 4336 HWC-M when usedwith a radiographic imaging system, isindicated for use in generatingradiographic images of humananatomy for diagnostic X-rayprocedures, wherever conventionalscreen-film (SF), digital radiography(DR), or computed radiography (CR)systems may be used. It is notintended for mammographic use.Same
PanelSingle substrate amorphous siliconactive TFT/diode arraySame
ScintillatorDirect deposition CsI:TlSame
Pixel pitch100 μmSame
Pixel matrix3524 × 42884288 × 4288
Limiting resolution5 lp/mmSame
Binning capable2 × 2 binning for 200 µmSame
Data transmit area352 mm × 429 mm429 mm × 429 mm
External dimensions(w × l × h)384 mm × 460 mm × 15.5 mm460 mm × 460 mm × 15.5 mm
Weight3.2 kg3.8 kg
Housing materialAluminum with carbon-fiberSame
FinishMatte carbon-fiber front with whitesilk-screen active area markersSame
Communication interfaceGb Ethernet or 802.11n WiFiSame
WiFi band5.15 - 5.85 GHzSame
X-ray synchronizationinterfaceDedicated trigger in/out signal lines orAutomatic Exposure DetectionSame
PowerExternal power supply or batterySame
Software libraryX-ray Imaging Software Library (XISL),which provides control of detectorconfiguration and image acquisitionsfrom a host computerSame
Software operating systemWindows and Linux OSSame

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Varex Imaging. The logo consists of the letter V, which is made up of four vertical lines in different shades of blue and green. To the right of the V is the word "VAREX" in blue, and below that is the word "IMAGING" in smaller, gray letters. The logo is simple and modern, and the colors are calming and professional.

Summary of Studies:

The proposed new device (XRpad2 4343 HWC-M) and the predicate device (XRpad2 4336 HWC-M) utilize similar technology and materials, are similar in design and construction, and have the same intended use. The construction and physical characteristics of the two devices are similar. The difference is related to a 76 mm increase in the width dimension of the proposed device, which does not impact image quality.

The XRpad2 4343 HWC-M flat panel detector has successfully completed internal non-clinical testing, and complies with IEC and ISO standards and regulations related to medical safety, EMC, and bio-compatibility. A clinical study was not required for the XRpad2 4343 HWC-M device. Taking into consideration the similarities in construction and physical characteristics of the XRpad2 4343 HWC-M to the predicate device, substantial equivalence of imaging performance between the two XRpad2 flat panel detectors can be demonstrated by a comparison of the non-clinical testing results only.

Summary of Design Control and Risk Management:

The XRpad2 4343 HWC-M flat panel X-ray detector is a modification of the XRpad2 4336 HWC-M (K161966). The modifications are related to an increase in the width dimension of the device, and do not impact image quality.

The risks and hazardous impacts of the device modification were analyzed by FMEA methodology. The specific risk control and protective measures to mitigate the risks from the modification were reviewed as part of product design. The overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted.

Summary of Non-Clinical Data:

To demonstrate the equivalence of the XRpad2 4343 HWC-M to the predicate device, Varex Imaging has performed internal non-clinical testing and demonstrated compliance with accepted IEC and ISO standards and regulations. In addition, Varex Imaging has followed the non-clinical considerations outlined in the Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, September, 2016. This testing demonstrated that critical image quality performance attributes, such as MTF and DQE, of the XRpad2 4343 HWC-M are comparable to the predicate device. The resulting data from the conducted non-clinical tests demonstrate substantial equivalence.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a logo for Varex Imaging. The logo has the word "VAREX" in a sans-serif font, with the letters in a dark blue color, except for the "X" which is in green. Below the word "VAREX" is the word "IMAGING" in a smaller, sans-serif font, also in a dark blue color. To the left of the word "VAREX" is a stylized "V" made up of four vertical lines, with the first line in dark blue, the second line in green, the third line in yellow, and the fourth line in dark blue.

Substantial Equivalency:

The proposed device and predicate device (XRpad2 4336 HWC-M flat panel detector) both utilize similar technology and materials, are similar in design and construction, and have been shown to produce images of equivalent diagnostic quality. Both devices are intended for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. The devices are not intended for mammographic use. Both devices produce digital images which can be transmitted to imaging software of the X-ray unit.

Conclusion:

Similar to the predicate device, the XRpad2 4343 HWC-M has comparable performance and comply with similar operational standards. Potential hazards have been studied by a Risk Management Plan. The nonclinical verification and validation test results demonstrate that the XRpad2 4343 HWC-M complies with international and FDA recognized consensus standards and meets the acceptance criteria and is adequate for its intended use.

Based on the information supplied in this 510(k) Varex Imaging concludes, the XRpad2 4343 HWC-M is substantially equivalent to the currently marketed device, XRpad2 4336 HWC-M (K161966) in terms of safety and effectiveness.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.