K Number
K243437

Validate with FDA (Live)

Date Cleared
2025-08-14

(282 days)

Product Code
Regulation Number
880.5400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The mOm Essential Incubator is indicated for thermoregulation for both premature and full-term infants up to a maximum of 6 kg (13 lbs).

The mOm Essential Incubator is not intended for home use.

Device Description

The mOm Essential Incubator, ME1, is a neonatal incubator used for thermoregulation for both premature and full-term infants up to a maximum of 6 kg (13 lbs). Power is sourced either from an external AC supply or an internal rechargeable battery.

The mOm Essential Incubator (ME1) is intended to provide a heated environment with a stable temperature for infants who require assistance with thermoregulation but do not require a humidity-controlled environment. The ME1 should only be used within healthcare settings providing neonatal and infant care, by properly trained personnel, as directed by an appropriately qualified attending physician familiar with current known risks and benefits of infant incubator use.

The mOm Essential Incubator provides a filtered air system creating a thermally controlled environment based on air temperature control. It provides incubation as well as a degree of isolation for infants who may be near others or being cared for in clinical areas. The incubator is designed to allow adequate visibility and access to the infant to deliver care.

The Infant Compartment is designed to be disassembled and packed into a compact unit for storage and transportation and must be assembled before use. It has been designed as a cost-effective alternative to traditional incubators.

The mOm Essential Incubator is made up of several key components and sub-assemblies, which are assembled by the user prior to use:

  • The incubator base unit, which includes the device's electronics, fan, heating system, display and controls.
  • A rigid front panel which contains a door and two portholes.
  • A flexible panel that makes up the top and rear of the infant compartment – this includes an inflatable section and is supplied with a manual inflation bulb.
  • A rigid top and back U-shaped panel which provide the structure for the top and rear of the infant compartment.
  • Two side panels which form the ends of the infant compartment.

An infant mattress is supplied, which is inserted into the infant compartment. A skin temperature probe is also supplied. An optional cart (trolley) is available.

AI/ML Overview

The provided 510(k) clearance letter pertains to the mOm Essential Incubator (ME1), a Class II medical device. It focuses on the device's technical specifications and comparison to a predicate device (Isolette® 8000 plus), as well as a range of non-clinical bench testing to demonstrate safety and effectiveness.

*Crucially, this document does not describe a study involving a test set, expert ground truth establishment, a multi-reader multi-case (MRMC) study, or a standalone algorithm performance study.

The "study" referenced in the provided text is a collection of non-clinical performance data and bench testing to demonstrate the device meets various safety and performance standards for neonatal incubators. It explicitly states: "Bench testing of the mOm Essential Incubator met all relevant requirements for neonatal incubators. Testing was performed in accordance with applicable standards, to characterize and evaluate the device, including performance testing, functional/operational testing, verification and validation, biocompatibility, human factors, risk analysis and verification of risk control measures. Testing included accessories and optional components."

Therefore, I cannot populate the requested sections related to traditional "studies" with human readers, expert ground truth, or AI performance metrics, as this device is a physical medical device (an incubator) not a software-driven diagnostic or AI-powered system that would require such studies.

However, I can extract the acceptance criteria and the results of the non-clinical testing for critical device features based on the provided document.


Acceptance Criteria and Reported Device Performance (Based on Non-Clinical Bench Testing)

The acceptance criteria for the mOm Essential Incubator (ME1) are largely aligned with international standards for medical electrical equipment and infant incubators (e.g., IEC 60601 series) and are demonstrated through direct comparison to the predicate device and specific non-clinical tests.

Feature / Acceptance Criteria CategorySpecific Acceptance Criteria (Implied by Standards/Predicate)Reported Device Performance (from non-clinical testing)
Intended UseNeonatal incubator for thermoregulation for premature and full-term infants up to 6 kg (13 lbs). Not for home use.Device functions as intended for thermoregulation for premature and full-term infants up to 6 kg (13 lbs). Not intended for home use. (Identical to predicate's core function, with a defined max weight for ME1)
Safety & Essential PerformanceCompliance with IEC 60601-1 (General requirements for basic safety and essential performance).Complies with IEC 60601-1:2005 + A1:2012 + A2:2020 (Edition 3.2 2020-08 consolidated version).
Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2.Complies with IEC 60601-1-2:2014+A1:2020 (Edition 4.1 2020-09 consolidated version).
Usability (Human Factors)Compliance with IEC 60601-1-6 and IEC 62366-1/2.Complies with IEC 60601-1-6:2010 + A1:2013 + A2:2020 and IEC 62366-1:2015/A1:2020. Summative Human Factors testing conducted.
Alarm SystemsCompliance with IEC 60601-1-8.Complies with IEC 60601-1-8:2006+A1:2012+A2:2020.
Infant Incubator Specific RequirementsCompliance with IEC 60601-2-19 (Particular requirements for basic safety and essential performance of infant incubators).Complies with IEC 60601-2-19:2020 (Edition 3.0 2020-09).
Software Level of ConcernSoftware verified and validated appropriately for its risk level.Software deemed "major" level of concern. Software verification and validation testing conducted as per FDA guidance (May 2005).
BiocompatibilityBiologically safe for intended use, compliance with ISO 10993 series and ISO 18562 series (especially for breathing gas pathways).Complies with ISO 10993-1:2018 (corrected 2018-10, Edition 2020-12), ISO 10993-5:2009, ISO 10993-10:2021, ISO 10993-23:2021, ISO 18562-1:2017, ISO 18562-2:2017, ISO 18562-3:2017. Evaluation of results from material investigations, chemical characterization, and literature shows biological safety.
Cleaning and Disinfection IntegrityMaintain integrity and usability after repeated cleaning/disinfection and assembly/disassembly.No issues encountered after 3 years of simulated deep cleans and disassembly/re-assembly.
Earth Leakage Current< 500 µA (Predicate value)< 350 µA (ME1's reported performance, smaller than predicate).
Parameters ControlledTemperature control.Temperature control only. (Predicate offers optional humidity/oxygen). Verified as safe for intended population not requiring these.
Air Temperature Set Point RangeClinically appropriate range.28 to 37°C in 0.5°C increments (82.4 – 98.6 in 0.9°F increments).
Air Temperature Deviation Limit (Up/Down)Within acceptable medical tolerance.± 1.5°C (±2.7°F).
Warm-up TimeTimely provision of thermal care.40 minutes (Predicate: < 35 minutes). Deemed clinically acceptable as incubators are typically kept ready.
Variability/Stability< 0.5°C (<0.9°F)< 0.5°C (<0.9°F) (Identical to predicate).
OvershootAcceptable during warm-up.2°C (3.6°F) (Predicate: < 0.5°C). Meets IEC 60601-2-19 requirements.
Uniformity (level mattress)< 0.8°C (<1.4°F)< 0.8°C (<1.4°F) (Identical to predicate).
Noise Level (inside infant compartment)Clinically acceptable for infant environment.≤ 60 dB(A) (Meets IEC 60601-2-19 requirements).
Air Velocity (inside infant compartment)<0.10 m/sec (Predicate value)<0.07 m/sec (Maximum measured), 0.054 m/sec average (Meets IEC 60601-2-19 requirements).
Maximum CO2 Level< 0.8% (Predicate value)≤ 0.3% (Lower than predicate, deemed safer for infant).
Acoustical Level (all alarms on)75 dB(A) maximum (Predicate value)70 dB(A) (Lower than predicate, still acceptable for caregiver notification).
In-Use Battery LifeMaintain set temperature above ambient for a specific duration.Can maintain set temperature of 11°C (51.8°F) above ambient (20-30°C / 68-86°F) for at least one hour.
Battery Life (Cycles)Durable battery for repeated use.1500 charge cycles.

Regarding the Absence of Certain Study Types:

This 510(k) submission is for a physical device (neonatal incubator) whose primary function is thermoregulation, and its safety and effectiveness are established through engineering design, material selection, adherence to electrical and performance standards, and rigorous bench testing. It is not an AI/ML-driven diagnostic or treatment device, nor does it involve image analysis, therefore:

  • Sample size used for the test set and data provenance: Not applicable in the context of clinical "test sets" as described for AI/diagnostic devices. The "test set" here refers to the actual physical device prototypes subjected to various non-clinical bench tests in a lab environment.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. "Ground truth" for an incubator is established by physical measurement against calibrated standards (e.g., temperature probes for thermal regulation, sound meters for noise).
  • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Performance is measured against engineering specifications and international standards.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, because this device does not involve human "readers" interpreting data or cases. It's an incubator.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, there is no discrete AI algorithm whose performance is being evaluated in isolation. The "algorithm" here refers to the software controlling the physical properties of the incubator, which is verified through functional testing of the device itself.
  • The type of ground truth used: For performance metrics like temperature, noise, CO2 levels, etc., the "ground truth" is derived from measurements taken with calibrated instruments against established physical and engineering standards (e.g., IEC standards for incubators). For biocompatibility, it's based on laboratory testing of materials as per ISO standards.
  • The sample size for the training set: Not applicable. This is not an AI/ML model that undergoes "training."
  • How the ground truth for the training set was established: Not applicable for the same reason.

In summary, the provided document details the regulatory clearance of a physical medical device (an incubator) based on a comprehensive set of non-clinical bench tests and engineering evaluations, rather than clinical efficacy studies involving patient data or AI performance metrics.

FDA 510(k) Clearance Letter - mOm Essential Incubator (ME1)

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 14, 2025

Mom Incubators Limited
Rosalyn Archer
Director of Clinical Affairs
Newstead House
Pelham Road
Nottingham, Nottinghamshire NG5 1AP
United Kingdom

Re: K243437
Trade/Device Name: mOm Essential Incubator (ME1)
Regulation Number: 21 CFR 880.5400
Regulation Name: Neonatal incubator
Regulatory Class: Class II
Product Code: FMZ
Dated: July 16, 2025
Received: July 16, 2025

Dear Rosalyn Archer:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K243437 - Rosalyn Archer Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243437 - Rosalyn Archer Page 3

Sincerely,

David Wolloscheck -S

David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known)
K243437

Device Name
mOm Essential Incubator (ME1)

Indications for Use (Describe)
The mOm Essential Incubator is indicated for thermoregulation for both premature and full-term infants up to a maximum of 6 kg (13 lbs).

The mOm Essential Incubator is not intended for home use.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary

[as required by 21 CFR §807.92(c)]

mOm Essential Incubator (ME1)

510(k) Number: K243437
DATE PREPARED: August 15, 2025
APPLICANT: Mom Incubators Ltd
Newstead House Pelham Road
Nottingham NG5 1AP
United Kingdom

CONTACT: Rosalyn Archer
Director of Clinical Affairs Newstead
House Pelham Road Nottingham
NG5 1AP United Kingdom
Telephone: +44 7853 633 079
Email: rarcher@momincubators.com

TRADE NAME: mOm Essential Incubator (ME1)
COMMON NAME: Neonatal incubator
CLASSIFICATION REGULATION: Neonatal incubator
21 CFR §880.5400, Product Code FMZ
DEVICE CLASS: II
PANEL CODE: General hospital / 880
PREDICATE DEVICE: Isolette® 8000 plus, K172154

Page 6

INTENDED USE/INDICATIONS FOR USE:

The mOm Essential Incubator is indicated for thermoregulation for both premature and full-term infants up to a maximum of 6 kg (13 lbs).

The mOm Essential Incubator is not intended for home use.

DEVICE DESCRIPTION:

The mOm Essential Incubator, ME1, is a neonatal incubator used for thermoregulation for both premature and full-term infants up to a maximum of 6 kg (13 lbs). Power is sourced either from an external AC supply or an internal rechargeable battery.

The mOm Essential Incubator (ME1) is intended to provide a heated environment with a stable temperature for infants who require assistance with thermoregulation but do not require a humidity-controlled environment. The ME1 should only be used within healthcare settings providing neonatal and infant care, by properly trained personnel, as directed by an appropriately qualified attending physician familiar with current known risks and benefits of infant incubator use.

The mOm Essential Incubator provides a filtered air system creating a thermally controlled environment based on air temperature control. It provides incubation as well as a degree of isolation for infants who may be near others or being cared for in clinical areas. The incubator is designed to allow adequate visibility and access to the infant to deliver care.

The Infant Compartment is designed to be disassembled and packed into a compact unit for storage and transportation and must be assembled before use. It has been designed as a cost-effective alternative to traditional incubators.

The mOm Essential Incubator is made up of several key components and sub-assemblies, which are assembled by the user prior to use:

  • The incubator base unit, which includes the device's electronics, fan, heating system, display and controls.
  • A rigid front panel which contains a door and two portholes.
  • A flexible panel that makes up the top and rear of the infant compartment – this includes an inflatable section and is supplied with a manual inflation bulb.
  • A rigid top and back U-shaped panel which provide the structure for the top and rear of the infant compartment.
  • Two side panels which form the ends of the infant compartment.

An infant mattress is supplied, which is inserted into the infant compartment. A skin temperature probe is also supplied. An optional cart (trolley) is available.

510(k) Summary_K243437 Page 2 of 18

Page 7

COMPARISON WITH PREDICATE DEVICE:

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
510(k)K243437K172154--
ManufacturerMom Incubators LtdDraeger Medical Systems Inc--
Regulation number880.5400880.5400IdenticalN/A
Product CodeFMZFMZIdenticalN/A
ClassificationClass IIClass IIIdenticalN/A
Intended UseNeonatal incubatorNeonatal incubatorIdenticalN/A
Indication for UseThe mOm Essential Incubator is indicated for thermoregulation for both premature and full-term infants up to a maximum of 6 kg (13 lbs). The mOm Essential Incubator is not intended for home use.The Isolette® 8000 plus incubator is indicated for thermoregulation and controlling oxygen (optional), and humidity (optional) for both premature and full-term infants up to a maximum of 10 kg (22 lbs). The Isolette® 8000 plus is not intended for home use.Similar – while the wording is different, the intended use of the ME1 has not changed from that of the predicate. Both devices are incubators, both provide a controlled environment, both are used for the same patient population, and both devices are not intended for home use.The subset of infants the ME1 is intended for are those who are more well and who are near-term rather than the most premature infants who require humidity, which is not provided or facilitated by the ME1. For infants who are ≤60cm long and fit in our device, the maximum weight of 6 kg was considered to be a more accurate indication of the maximum weight of intended users, which would be the 98.9th percentile for a 30 day old infant (the maximum length of 60cm would be the 99.7th percentile for a 30 day old infant according to WHO growth charts.) 1,2,3 There is no negative impact to the
Target populationPre-term and term infants up to 6kg (13lbs) in weight, who do not require a humidity-controlled environmentNewly born infants up to 10 kg. (22 lbs)Similar – the subject has a lower allowable infant weight; both devices can provided a non-humidity controlled environment

¹ WHO Weight for boys 0-13 weeks age. https://cdn.who.int/media/docs/default-source/child-growth/child-growth-standards/indicators/weight-for-age/wfa-boys-0-13-percentiles.pdf?sfvrsn=292d03f1_11
² WHO Infant Growth Chart Calculator: Weight Age WHO 0-2 Year. https://www.infantchart.com/infantweightage.php
³ WHO Infant Growth Chart Calculator: Length Age WHO 0-2 Year. https://www.infantchart.com/infantlengthage.php

Page 8

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
safety or effectiveness of the device
Environment of UseWithin healthcare setting providing neonatal and infant careFor use in any department of the hospital that provides neonatal and infant care, including NICU/Special Care Baby Unit and the Step Down Nursery, Newborn Nursery and pediatrics.IdenticalN/A

System Specifications

| Principle of Operation | Controller-based incubator that enables control of temperature affecting the infant | Controller-based incubator that enables simultaneous control of temperature, oxygen, and humidity parameters affecting the infant | Similar | The intended population of the ME1 are pre-term and term infants who do not require a humidity-controlled environment. There is no negative impact to the safety or effectiveness of the device. |
| Protection class | Class I, Type BF, continuous operation | Class I, Type BF, continuous operation, not AP | Identical | N/A |
| Ingress of liquids and particulate matter (IEC 60601-1) | IPX0 | IPX0 | Identical | N/A |
| Intended for use in Transport vehicles | No | No | Identical | N/A |

510(k) Summary_K243437 Page 4 of 18

Page 9

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
Electrical Safety and PerformanceIEC 60601-1 IEC 60601-1-2 IEC 60601-2-19IEC 60601-1 IEC 60601-1-2 IEC 60601-2-19IdenticalN/A

Physical Attributes

| Height | 44.2 cm (17.4 inches) 129 cm (51 inches) with cart | 133.3 to 153.7 cm (52.5 to 60.5 inches) | Different | The difference in height has no impact on the patient population to be treated or the environment in which it is to be used; difference also has no impact on overall safety and effectiveness based on non-clinical testing performed |
| Width | 44.5 cm (17.5 in) | < 104cm (41 in) | Different | The difference in width has no impact on the patient population to be treated based on WHO data;1,2,3 there is no clinical relevance for the overall width of the device |
| Weight | 20 kg (44 lb) without options/accessories 73 kg (161lb) with cart | <98.5 kg (217 lb.) without options/accessories | Different | The difference in weight has no clinical relevance to the patient population to be treated or to the clinical environment where it will be used |
| Depth | 76.4 cm (30.1 inches) | < 76.2 cm (30 inches) | Same | N/A |
| Mattress size | 30 x 62 x 2.5 cm | ≥ 38 x 74 x 3 cm (15 x 29.1 x 1.2 inches) | Similar | The difference in the mattress size has no impact on the patient population to be treated as it is sized to fit the device; the mattress size has no relevance on the clinical environment. For infants who are ≤60cm long and |

510(k) Summary_K243437 Page 5 of 18

Page 10

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
fit in our device, the maximum weight of 6 kg was considered to be a more accurate indication of the maximum weight of intended users, which would be the 98.9th percentile for a 30 day old infant (the maximum length of 60cm would be the 99.7th percentile for a 30 day old infant according to WHO growth charts.)1,2,3
TiltNoneContinuously variable to 12 deg. +1 deg.DifferentThe absence of tilt has no impact on the patient population to be treated or relevance to the clinical environment where the device is to be used
Infant weightMaximum 6 kg (13.2 lbs)Maximum 10 kg (22 lbs)SimilarThe subset of infants the ME1 is intended for are those who are more well and who are near-term. For infants who are ≤60cm long and fit in our device, the maximum weight of 6 kg was considered to be a more accurate indication of the maximum weight of intended users

Environment

| Operating temperature | 20-30°C (68-86 °F) | 20-30°C (68-86 °F) | Identical | N/A |

510(k) Summary_K243437 Page 6 of 18

Page 11

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
Operating humidity10-80% RH non-condensing5-95% RH non-condensingSimilarThe incubator is for use within a hospital setting. As standard NICU healthcare facilities require a minimum of 30% humidity and a maximum of 60%.4 The operating range of the ME1 fall within this range and non-clinical testing supports that the device can function in this range as intended. This operating range has no impact on the patient population to be treated.
Operating ambient air pressure520-760mmHg (70-101 kPa)70-110 kPaSimilarThe ME1 can operate at normal ambient conditions; the minimal difference does not impact the patient population to be treated nor the environment in which the incubator is to be used
Storage temperature0 to 40°C (32 – 104°F)-20 to 60°C (-4 to 140°F)DifferentThe storage conditions for the ME1 are within the range cleared for the predicate device; thus the conditions are suitable for a hospital/healthcare setting; non-clinical testing supports the device functions as intended when stored under these conditions; the difference in storage temperature has no impact on the

4 NICU Standard 11: Ambient Temperature and Ventilation. https://nicudesign.nd.edu/nicu-standards/nicu-standard-12-ambient-temperature-and-ventilation/#:~:text=The%20NICU%20shall%20be%20designed,two%20changes%20being%20outside%20air.

510(k) Summary_K243437 Page 7 of 18

Page 12

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
patient population to be treated

Electrical Requirements

| Power requirements | 100-240V, 50-60 Hz at 2.5-9.0A | 100/120V, 50/60 Hz, 1900 W max, 9.9 A max | Similar | The ME1 can operate in a larger voltage range, which is supported by the non-clinical testing performed; the broadened range affords more flexibility when operating the unit; the difference in allowable voltage has no impact on the patient population to be treated |
| Auxiliary power sockets | n/a | All 50/60 Hz 100 V, 100W max 120 & 230 V 300W max | Different | The absence of auxiliary power sockets does not affect the function of the subject device as a neonatal incubator, nor the patient population to be treated |
| In-Use Battery Life | On Battery Power, ME1 can maintain a set temperature of 11°C (51.8°F) above ambient (20 to 30°C (68 - 86°F)) for at least one hour | n/a | Different | ME1 can continue to provide thermal care in the event of a power outage or disconnection from wall power either by accident or for transfer within the hospital on the ME1 trolley. This feature is supported by non-clinical testing reported in the Verification Test Report for the Pilot Production Build Essential Incubator, CHH-1493-TR (page 62). This feature has no impact on the patient population to be treated. |
| Battery Life | 1500 charge cycles | n/a | Different – predicate does not include | This feature is supported by non- |

510(k) Summary_K243437 Page 8 of 18

Page 13

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
a batteryclinical testing provided on the Technical Datasheet by the battery manufacturer (URB1270 Technical Datasheet). This feature has no impact on the patient population to be treated or the clinical environment in which it is to be used.
Earth leakage<350 μA< 500 μASimilarThe ME1 has a smaller residual current than the predicate device as determined via non-clinical testing; the smaller residual current had no impact on the patient population to be treated or clinical environment in which it is to be used

General Performance

| Parameters controlled | Temperature | Temperature Humidity (optional) Oxygen (optional) | Similar | The intended population of the ME1 are pre-term and term infants who do not require a humidity-controlled environment. There is no negative impact to the safety or effectiveness of the device as a result of not having humidity and/or oxygen functions |
| Operations | Controller based – air temperature mode | Controller based – air temperature mode and skin temperature mode | Similar | ME1 does not provide skin temperature mode control to avoid the risk of masking clinical conditions; difference does not impact the patient population to be treated nor the environment in which it is to be used |
| Skin temperature set point range | n/a | 34-37°C (93.2-98.6 °F) | Different | |

510(k) Summary_K243437 Page 9 of 18

Page 14

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
Air temperature set point range28 to 37°C in 0.5°C increments (82.4 – 98.6 in 0.9°F increments)20 to 39°C (68-102.2°F)SimilarThe temperature range provided by the ME1 was identified during formative user requirement investigation, and has been adjusted based on our Post Market Surveillance to meet the full range required by clinicians using our device. There is no impact to the patient population to be treated due to the tightened air temperature ranges
Air temperature control override range-37-39°C (98.6-102.2°F)Different
Air setpoint display range28-37°C (82.4 – 98.6°F)15 to 45°C (59-113°F)DifferentIt is important for the clinicians to see the temperature that has been set for the incubator to reach. By only displaying the allowed range of set temperatures on the ME1 it avoids the impression that the device can provide a wider range of temperatures. The difference has no impact on the patient population to be treated
Air temperature upward deviation limit+ 1.5°C (+2.7°F)+ 1.5 to +2.5°C (+2.7 to 4.5°F)SimilarThe ME1 ensures that the accuracy of the air temperature is within the tolerance of + 1.5°C (+2.7°F) for a given Set Temperature; ensuring that the incubator provides medically appropriate thermal care, as decided by the clinician. The difference in the tolerance has no impact on the patient population to be treated
Air temperature downward deviation limit range- 1.5°C (-2.7°F)-1.5 to -2.5°C (-2.7 to -4.5°F)Similar

510(k) Summary_K243437 Page 10 of 18

Page 15

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
Warm up time at ambient temperature40 minutes< 35 minutesSimilarTimely provision of thermal care is important for cold infants. However, in normal practice incubators are kept on and ready during births or in relevant hospital departments and will not be impacted by a difference in warm-up time. This difference does not impact the patient population to be treated
Variability/Stability< 0.5°C (<0.9°F)< 0.5°C (<0.9°F)IdenticalN/A
Overshoot2°C (3.6°F)< 0.5°C (<0.9°F)SimilarThe overshoot during warm-up meets the requirements of the non-clinical testing performed in accordance with IEC 60601-2-19; this difference has no impact on the patient population to be treated nor on the environment in which it is to be used
Uniformity with a level mattress< 0.8°C (<1.4°F)<0.8°C (<1.4°F)IdenticalN/A

Misc. Specifications

| Noise Level inside infant compartment | ≤ 60 dB(A) | ≤ 47 dB(A) with 37dB(A) or less ambient (without servo O2 control) | Similar | The noise level of the subject device in normal conditions meets the requirements of IEC 60601-2-19; this difference has no impact on the patient population to be treated nor on the environment in which it is to be used |
| Air Velocity inside infant | <0.07 m/sec (Maximum measured) | <0.10m/sec (4in/sec); average of 5 points at 10cm(4 in) above | Similar | The air velocity meets the requirements of IEC 60601-2-19; this |

510(k) Summary_K243437 Page 11 of 18

Page 16

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
compartment0.054m/sec average of 5 points at 10cm(4 in) above mattressmattressdifference has no impact on the patient population to be treated nor on the environment in which it is to be used
Maximum CO2 level≤ 0.3%< 0.8%SimilarCO2 can be dangerous for an infant if it builds up within the enclosed incubator environment. The ME1 CO2 level remains lower than the predicate; this difference has no impact on the patient population to be treated nor on the environment in which it is to be used

Components

| Acoustical level (all alarms on) | 70 dB(A) | 75 dB(A) maximum | Similar | The lower noise level in alarm conditions meets the requirements of IEC 60601-2-19; Loud noises can impact infant sleep and intra cranial pressure and recommendations point to low infant environmental sound levels. While alarm sound levels are higher than is desirable for the infant, they are required to let a care giver know that there may be a higher risk situation and therefore the benefit outweighs the risk; this difference has no impact on the patient population to be treated nor on the environment in which it is to be used |

510(k) Summary_K243437 Page 12 of 18

Page 17

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
Hand ports (Porthole Doors)Yes – 2 on the front of the incubatorYes – 4 total 2 on either side of the incubatorSimilar – both configurations allow for access to the infant for care and/or medical emergenciesInfants in the incubator need to be cared for or may need to be accessed for medical procedures. Users of the ME1 have found the 2 port doors and incubator door adequate for their needs based on non-clinical testing performed; this difference has no impact on the patient population to be treated nor on the environment in which it is to be used
Iris PortsNoYes – 2 One on either end of the incubatorDifferent – lack of iris ports does not impact the safety or effectiveness of the device
Access panelYesYesIdenticalN/A
Air filterYesYesIdenticalN/A
Oxygen filtern/aYesDifferentThe design of the ME1 device allows for oxygen to be delivered via an external source to the infant while inside of the incubator via the IV-ports; this difference has no impact on the patient population to be treated nor on the environment in which it is to be used

510(k) Summary_K243437 Page 13 of 18

Page 18

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
Probes / SensorsSkin temperature (optional)Skin temperature Humidity (optional) Oxygen (optional)Similar – both devices allow for the measurement of skin temperature; the optional features of the predicate device are not required for the functionality of the ME1 deviceInfants in an incubator require frequent monitoring. An embedded skin temperature probe enables this without any additional equipment; the optional features of the predicate device are not required for the functionality of the ME1 device and do not impact the patient population intended to be treated by the ME1 or the environment in which it is to be used
Serial portNoYesDifferentThe absence of a serial port and COM Port does not affect the function of the ME1 device as a neonatal incubator, the patient population to be treated or the environment in which it is to be used
COM Port (output only)No output COM PortOnly connects to devices that fulfill the requirements of the standard IEC 60950-1 on unearthed SELV circuits or the requirements of the standard IEC 60601-1 on accessible secondary circuits with max. 60V DC nominal voltage.Different
AccessoriesCart (Trolley), castor wheels with brakesTrolley, castor wheels with brakesIdenticalN/A
Optional ComponentsN/AIV pole, drawers, water reservoir, collection bottle, cylinder mount litter bag holder, basket for gloves, breathing hose holder kit, tray, high frequency ventilation (HFV) door with grommets, positioning aids, incubator cover, and monitor shelfDifferentThe optional accessories available for the predicate do not impact the basic function of the ME1 as a neonatal incubator, the patient population it intends to treat or the environment in which it is to be used

510(k) Summary_K243437 Page 14 of 18

Page 19

Biocompatibility and Sterility

SubjectmOm Essential Incubator (ME1)Isolette® 8000 plusComparisonImpact of Differences on patient population, clinical use environment and testing to support differences
BiocompatibilityAccording to ISO 10993 and ISO 18562: • ISO 10993-1 • ISO 10993-5 • ISO 10993-10 • ISO 10993-23 • ISO 18562-1 • ISO 18562-2 • ISO 18562-3According to ISO 10993SimilarThe ME1 was subjected to additional biocompatibility testing to confirm the lack of particulates, volatile organic compounds, or leachables present within the air pathway; the difference has no impact on the environment in which it is to be used
SterileNoNoIdenticalN/A

510(k) Summary_K243437 Page 15 of 18

Page 20

The mOm Essential Incubator and the predicate device have the same intended use, as neonatal incubators. Both the subject and predicate are controller-based incubators that can control temperature.

The subject device and the predicate device have similar technological characteristics. The mOm Essential Incubator has a subset of the indications and the same risks, basic technology, and potential adverse events as the predicate (K172154). The primary technological difference between the predicate and the subject devices is that the subject device includes an air temperature control mode, whereas the predicate includes both air temperature and skin temperature modes. Also, the subject device does not include the control of oxygen and humidity, which are options in the predicate device.

NON-CLINICAL TESTING/PERFORMANCE DATA:

Bench testing of the mOm Essential Incubator met all relevant requirements for neonatal incubators. Testing was performed in accordance with applicable standards, to characterize and evaluate the device, including performance testing, functional/operational testing, verification and validation, biocompatibility, human factors, risk analysis and verification of risk control measures. Testing included accessories and optional components.

Electrical safety, electromagnetic compatibility (EMC) and Performance:

Electrical safety testing and performance testing were conducted on the mOm Essential Incubator. Testing shows that the device complies with the requirements of IEC 60601-1, IEC 60601-1-6, IEC 60601-1-8, and IEC 60601-2-19. EMC testing was also conducted on the mOm Essential Incubator. Testing shows that the device complies with the requirements of IEC 60601-1-2:

  • IEC 60601-1:2005 + IEC 60601-1:2005/A1:2012 + IEC 60601-1:2005/A2:2020. Edition 3.2 2020-08 consolidated version. Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
  • IEC 60601-1-2:2014+IEC 60601-1-2:2014/A1:2020. Edition 4.1 2020-09 consolidated version. Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance. Collateral Standard. Electromagnetic disturbances. Requirements and tests.
  • IEC 60601-1-6:2010 + IEC 60601-1-6:2010/A1:2013 + IEC 60601-1-6:2010/A2:2020 for use in conjunction with IEC 62366:2015+A1:2020. Edition 3.2 2020-07 consolidate version. Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance. Collateral standard. Usability.
  • IEC 60601-1-8:2006+IEC 60601-1-8:2006/A1:2012+IEC 60601-1-8:2006/A2:2020. Edition 2.2 2020-07 consolidated version. Medical electrical equipment – Part 1-8: General requirements for basic safety and essential performance – Collateral Standard: General requirements. Tests and guidance for alarm systems in medical electrical equipment and medical electrical systems.
  • IEC 60601-2-19:2020. Edition 3.0 2020-09. Medical electrical equipment – Part 2-19: Particular requirements for the basic safety and essential performance of infant incubators.

510(k) Summary_K243437 Page 16 of 18

Page 21

Software verification and validation:

The software controlling operations in the mOm Essential Incubator was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator. Software verification and validation testing were conducted as outlined in the FDA guidance document entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," May 2005.

Performance data demonstrate that the device functions as intended and has a safety and effectiveness profile that is similar to the predicate device.

Biocompatibility:

To evaluate the biocompatibility of the mOm Essential Incubator, a risk management approach, in accordance with ISO 10993-1 and ISO 18562 series, was chosen when considering the requirements for the chemical, physical and biological characteristics of the device. Testing was also conducted in accordance with the FDA guidance entitled "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'," September 2020.

The following tests were performed on the mOm Essential Incubator:

  • Test for volatile organic compounds
  • Emission of particles and Inorganic gases
  • Material characterization
  • Cytotoxicity evaluation
  • Irritation evaluation
  • Sensitization evaluation

These tests were performed in compliance to the following standards:

  • ISO 10993-1:2018, including corrected version 2018-10. Edition 2020-12. Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process.
  • ISO 10993-5:2009. Edition 2009-06. Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity.
  • ISO 10993-10:2021. Edition 2023-02. Biological evaluation of medical devices — Part 10: Tests for skin sensitization.
  • ISO 10993-23:2021. Edition 2021-03. Biological evaluation of medical devices — Part 23: Tests for irritation.
  • ISO 18562-1:2017. Edition 2020-02. Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 1: Evaluation and testing within a risk management process.
  • ISO 18562-2:2017. Edition 2020-02. Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter.
  • ISO 18562-3:2017. Edition 2020-02. Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds (VOCs).

510(k) Summary_K243437 Page 17 of 18

Page 22

The evaluation of the results from material investigations, chemical characterization, and evaluation of literature, show that the mOm Essential Incubator is biologically safe for its intended use and application.

Human Factors:

Summative Human Factors testing was conducted in accordance with IEC 62366-1 and IEC 62366-2:

  • IEC 62366-1:2015/A1:2020, incorporating amendment 1:2020 and corrigendum July 2016, Medical devices - Part 1. Application of usability engineering to medical devices. Published 2020-06.
  • IEC 62366-2:2016, Medical devices-Part 2. Guidance on the application of usability engineering to medical devices. Published 2016-04.

Cleaning and Disinfection:

Cleaning and Life testing has been conducted on the mOm Essential Incubator to evaluate the integrity of the incubator and its components after repeated cleaning and use, in accordance with the IFU. No issues were encountered that would affect the safety or usability of the incubator after three (3) years of simulated deep cleans and disassembly/re-assembly.

CONCLUSION:

The mOm Essential Incubator (ME1) has the same intended use and the same or similar technological characteristics as the predicate. Performance data demonstrates that the device functions as intended, and is as safe and effective as the predicate device.

510(k) Summary_K243437 Page 18 of 18

§ 880.5400 Neonatal incubator.

(a)
Identification. A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.(b)
Classification. Class II (performance standards).