K Number
K243322

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-05-09

(198 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Altaviz Intravitreal Syringe is intended for use by heath care professionals for general purpose fluid aspiration/injection. It is indicated for general and intravitreal use in adult patients.

Device Description

The Altaviz Intravitreal Syringe is a three-piece sterile, single use, hypodermic syringe with male 6% (Luer) conical lock fittings, which are connectable to a compatible female 6% (Luer) connector. The syringe assembly consists of a lubricated polycarbonate barrel with a graduated scale, a lubricated synthetic rubber stopper and a polypropylene plunger rod. The plunger rod is pulled back to aspirate fluids or depressed to inject or expel fluids. The barrel scale of the Altaviz Intravitreal Syringe incorporates a scale graduated in units of milliliters and is provided sterile by an irradiation sterilization method. The only difference between the Altaviz Intravitreal Syringe and the BD 1mL Luer-Lok Syringe is that the Altaviz Intravitreal Syringe is routinely tested to, and must meet, additional requirements to support intraocular use including endotoxin and particulate matter. The Altaviz Intravitreal Syringe is provided sterile by irradiation sterilization.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the Altaviz Intravitreal Syringe does not describe acceptance criteria, a study that proves the device meets specific acceptance criteria, or any information related to AI or ground truth as requested in the prompt.

The document states that "performance testing was also conducted and results confirmed that the Altaviz Intravitreal Syringe complies with the requirements of ISO 80369-7 (...) and ISO 7886-1 (...)". These ISO standards contain technical specifications and test methods for the performance of syringes. However, the document does not provide the specific acceptance criteria (e.g., maximum force, leakage rates, etc.) derived from these standards or the actual performance results of the Altaviz Intravitreal Syringe against those criteria.

Therefore, most of the requested information cannot be extracted from the provided text.

Here is the information that can be extracted or inferred to the best of my ability, highlighting where information is missing:


1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Material/Design Conformance:
Barrel (Polycarbonate)Identical to predicate
Barrel Lubricant (Silicone)Identical to predicate
Plunger Rod (Polypropylene)Identical to predicate
Stopper (Polyisoprene Rubber)Identical to predicate
Stopper Lubricant (Silicone)Identical to predicate
Sterilization:
Sterilization Method (Gamma Irradiation)Identical to predicate
Sterility Assurance Level (SAL) 10⁻⁶Identical to predicate
Shelf Life:
5 YearsIdentical to predicate
Particulate Matter (for intravitreal use):
USP <788> Particulate Matter in Injections and USP <789> Particulate Matter in Ophthalmic Solutions requirements"Passed" (Specific values not provided)
Performance (ISO 80369-7 & ISO 7886-1):
Requirements of ISO 80369-7 (Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications)"Complies" (Specific acceptance criteria/values not provided)
Requirements of ISO 7886-1 (Sterile hypodermic syringes for single use - Part 1: Syringes for manual use)"Complies" (Specific acceptance criteria/values not provided)
Biocompatibility:
Cytotoxicity, sensitization, irritation, acute systemic toxicity, material mediated pyrogenicity, hemolysis.Conducted on predicate device (K941562)
Subacute toxicity (Chemical characterization and toxicological risk assessment)."Performed" (Specific results not detailed beyond "ocular study...demonstrated that the device was not considered inflammatory to intraocular tissues of the rabbit.")

Information not available in the provided text:

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not specified. The document only mentions "Particulate testing", "Performance testing", and "Non-Clinical Biocompatibility Testing" without detailing sample sizes or provenance of the test subjects/samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. This device is not an AI/imaging device requiring expert adjudication for ground truth. The testing relates to physical and biological performance characteristics of a syringe. The ocular study was performed in rabbits, not human experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is not an AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For physical and chemical tests (e.g., particulate matter, ISO standards compliance), the "ground truth" is defined by the specific parameters and thresholds set within the respective standards (USP <788/789>, ISO 80369-7, ISO 7886-1).
  • For biological tests (biocompatibility, ocular study), the "ground truth" is determined by established biological safety standards, material characterization, and the observed responses in the animal model.

8. The sample size for the training set

  • Not applicable. This is not an AI device needing a training set.

9. How the ground truth for the training set was established

  • Not applicable. This is not an AI device needing a training set.

FDA 510(k) Clearance Letter - Altaviz Intravitreal Syringe

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 9, 2025

Altaviz, LLC
℅ Sean Griffin
President
Allied Regulatory Consulting
1540 Keller Parkway, Suite 108 #170
Keller, Texas 76248

Re: K243322
Trade/Device Name: Altaviz Intravitreal Syringe
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston syringe
Regulatory Class: Class II
Product Code: QLY, FMF
Dated: April 4, 2025
Received: April 7, 2025

Dear Sean Griffin:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K243322 - Sean Griffin Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243322 - Sean Griffin Page 3

Sincerely,

Shruti N. Mistry -S

Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K243322

Device Name
Altaviz Intravitreal Syringe

Indications for Use (Describe)
The Altaviz Intravitreal Syringe is intended for use by heath care professionals for general purpose fluid aspiration/injection. It is indicated for general and intravitreal use in adult patients.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

K243322 510(k) Summary

This 510(k) Summary has been prepared in accordance with 21 CFR 807.92.

Applicant

The name and address of the Applicant is:

Altaviz LLC
13766 Irvine Blvd, Suite 143
Irvine, CA 92618

Application Correspondent:
Sean Griffin
President
Allied Regulatory Consulting
1540 Keller Parkway, Suite 108 #170
Keller, TX 76248
Phone: (817) 994-4306

Date Prepared: May 9, 2025

Device

Device Subject to this 510(k):

Trade Name: Altaviz Intravitreal Syringe
Common Name: Piston Syringe
Product Code: QLY (Ophthalmic Syringe), FMF (Syringe, Piston); 21 CFR 880.5860
Classification: II

Predicate Device

510(k) Number: K941562
Device: 1mL Luer-Lok Syringe

Page 6

Device Description

The Altaviz Intravitreal Syringe is identical to the BD (Becton Dickenson) 1mL Luer-Lok Syringe cleared in K941562 with the exception of the Indications for Use. The Altaviz Intravitreal Syringe is a three-piece sterile, single use, hypodermic syringe with male 6% (Luer) conical lock fittings, which are connectable to a compatible female 6% (Luer) connector. The syringe assembly consists of a lubricated polycarbonate barrel with a graduated scale, a lubricated synthetic rubber stopper and a polypropylene plunger rod. The plunger rod is pulled back to aspirate fluids or depressed to inject or expel fluids. The barrel scale of the Altaviz Intravitreal Syringe incorporates a scale graduated in units of milliliters and is provided sterile by an irradiation sterilization method. The only difference between the Altaviz Intravitreal Syringe and the BD 1mL Luer-Lok Syringe is that the Altaviz Intravitreal Syringe is routinely tested to, and must meet, additional requirements to support intraocular use including endotoxin and particulate matter.

The Altaviz Intravitreal Syringe is provided sterile by irradiation sterilization.

Indications for Use

The Altaviz Intravitreal Syringe is intended for use by heath care professionals for general purpose fluid aspiration/injection. It is indicated for general and intravitreal use in adult patients.

Technological Characteristics

The subject Altaviz Intravitreal Syringe is identical to the predicate BD 1mL Luer-Lok Syringe in terms of materials and performance characteristics. The technological characteristics of the subject device and predicate device are substantially equivalent as they are identical devices.

Substantial Equivalence Discussion

The subject device is substantially equivalent to the predicate device when evaluating intended use and technological characteristics. The subject device is additionally indicated for intravitreal use. This difference in indications for use between the subject and predicate device is addressed through additional biocompatibility, sterility and performance testing, and does not raise new or different questions of safety and effectiveness.

Page 7

Features and CharacteristicsSubject Device Altaviz Intravitreal SyringePredicate Device BD 1mL Luer-Lok SyringeComparison of Subject Device and Predicate
510(k) NumberK243322K941562
Device CodeQLY, FMFFMFIdentical
Indications for Use/Intended UseThe Altaviz Intravitreal Syringe is intended for use by heath care professionals for general purpose fluid aspiration/injection. It is indicated for general and intravitreal use in adult patients.The BD 1mL LuerLok™ Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection.Same as predicate with exception of additional indication for intravitreal use.
Syringe Materials
BarrelPolycarbonatePolycarbonateIdentical
Barrel LubricantSiliconeSiliconeIdentical
Plunger RodPolypropylenePolypropyleneIdentical
StopperPolyisoprene RubberPolyisoprene RubberIdentical
Stopper LubricantSiliconeSiliconeIdentical
Mechanism of ActionManualManualIdentical
Sterilization MethodGamma IrradiationGamma IrradiationIdentical
SAL10⁻⁶10⁻⁶Identical
Shelf Life5 Years5 YearsIdentical

Non-Clinical Performance Testing

Particulate testing of the subject device according to USP <788> Particulate Matter in Injections and USP <789> Particulate Matter in Ophthalmic Solutions was performed and passed.

Performance testing was also conducted and results confirmed that the Altaviz Intravitreal Syringe complies with the requirements of ISO 80369-7 (Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications) and ISO 7886-1 (Sterile hypodermic syringes for single use - Part 1: Syringes for manual use).

Non-Clinical Biocompatibility Testing

In addition to the biocompatibility testing conducted on the predicate (cytotoxicity, sensitization, irritation, acute systemic toxicity, material mediated pyrogenicity, hemolysis), chemical characterization and toxicological risk assessment of the device was performed to evaluate the subacute toxicity. An ocular study was conducted following intravitreal injection in rabbits and demonstrated that the device was not considered inflammatory to intraocular tissues of the rabbit.

Page 8

Conclusion

The Altaviz Intravitreal Syringe is substantially equivalent to the predicate device.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).