K Number
K243254

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-08-08

(297 days)

Product Code
Regulation Number
872.3765
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Parkell Pit and Fissure Sealant is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations.

Device Description

The Device is a light-cured, flowable, resin-based sealant for prophylactic sealing of pits and fissures as well as for micro-restorative procedures. The Device is light-curable under standard power (600 mW/cm²) and contains filling agents which cause the Device to release and recharge beneficial ions (fluoride and calcium).

AI/ML Overview

The provided text describes the 510(k) clearance for the "Parkell Pit and Fissure Sealant." However, the clearance letter and accompanying 510(k) summary pertain to a physical dental device (a sealant), not an AI/software medical device.

Therefore, many of the requested criteria regarding AI/software performance studies (such as sample size for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable and are not mentioned in this document.

The document focuses on the physical and chemical properties of the sealant and compares them to a predicate device. The performance data is based on non-clinical, in-vitro testing of these physical properties.

Below is the information that can be extracted from the provided text, adapted as best as possible to the requested format, with clear indications where information is not applicable (N/A) due to the nature of the device.


Acceptance Criteria and Device Performance for Parkell Pit and Fissure Sealant

This device is a physical dental sealant, not an AI/software medical device. Therefore, the performance evaluation focuses on physical and chemical properties and biocompatibility, not algorithmic performance on image data.

1. Table of Acceptance Criteria and Reported Device Performance

PropertyAcceptance Criteria (Pass/Fail)Reported Device Performance (Parkell Pit and Fissure Sealant)
Bond Strength to Dry Enamel≥10 MPa≥10 MPa
Compressive Strength≥200 MPa≥200 MPa
Flexural Strength≥90 MPa≥90 MPa
Depth of Cure1.5-2.5 mm1.5-2.5 mm
Shrinkage≤8%≤8%
Viscosity (initial)1-5 Pa.s1-5 Pa.s
Shelf LifeNot explicitly stated2 years (established by testing)
Material BiocompatibilitySubstantially equivalent to predicate deviceSubstantially equivalent to predicate device
Beneficial Ion ReleaseFluoride & CalciumFluoride & Calcium (vs. Predicate's Fluoride, Calcium, Phosphate)

2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified in the document. The testing was non-clinical (in-vitro).
  • Data Provenance: Non-clinical physical property testing. Not applicable for geographical or retrospective/prospective distinctions.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • N/A. This is a physical dental material. "Ground truth" in this context typically refers to objective measurements from laboratory testing equipment (e.g., universal testing machines for strength, viscometers for viscosity, etc.), not expert interpretation of data like in AI/imaging studies.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • N/A. As above, this is laboratory-based physical property testing, not subjective assessment requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • N/A. This is not an AI-assisted device. No MRMC study was performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • N/A. This is not an algorithm. Standalone performance is not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Laboratory Standard Measurements / In-vitro physical property testing. The performance criteria (e.g., strength in MPa, shrinkage in %, viscosity in Pa.s) are based on standardized tests for dental materials.

8. The sample size for the training set

  • N/A. This is not an AI/machine learning device. No training set was used.

9. How the ground truth for the training set was established

  • N/A. As above, no training set was used.

FDA 510(k) Clearance Letter - Parkell Pit and Fissure Sealant

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 8, 2025

Parkell, Inc.
℅ Keith Barritt
Attorney
Fish & Richardson P.C.
1000 Maine Avenue, S.W.
Suite 1000
Washington, District of Columbia 20024

Re: K243254
Trade/Device Name: Parkell Pit and Fissure Sealant
Regulation Number: 21 CFR 872.3765
Regulation Name: Pit And Fissure Sealant And Conditioner
Regulatory Class: Class II
Product Code: EBC
Dated: July 8, 2025
Received: July 8, 2025

Dear Keith Barritt:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243254 - Keith Barritt Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

Page 3

K243254 - Keith Barritt Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bobak Shirmohammadi -S

For Michael E. Adjodha, M.ChE., RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K243254

Device Name
Parkell Pit and Fissure Sealant

Indications for Use (Describe)
The Parkell Pit and Fissure Sealant is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary -K243254

Parkell, Inc.

Parkell Pit and Fissure Sealant

Submitter

(i) 510(k) Submitter

Parkell, Inc.
300 Executive Drive
Edgewood, NJ 11717
FDA Registration No. 2411797

(ii) 510(k) Submitter Contact

Keith A. Barritt
Fish & Richardson P.C.
1000 Maine Avenue, S.W.
Suite 1000
Washington, DC 20024
Phone: (202) 626-6433
Facsimile: (202) 783-2331
Email: barritt@fr.com

(iii) Preparation Date

August 8, 2025

Device

Trade or Proprietary Name: Parkell Pit and Fissure Sealant
Common Name: pit and fissure sealant and conditioner
Classification Name: sealant, pit and fissure, and conditioner
Product Code: EBC, 21 CFR 872.3765
Class: 2

Page 6

Predicate Device

The Parkell Pit and Fissure Sealant device (the "Device") is substantially equivalent to Premier Dental Company Products' "Premier Sealant" (K#161580, the "Predicate Device").

The Device is indicated for "prophylactic sealing of pits and fissures as well as for micro-restorative procedures." The Indications for Use statement for the Predicate Device is nearly identical, namely "prophylactic sealing of pits and fissures. It may also be used for micro-restorative or initial layer of composite restorations." The minor difference in wording is irrelevant and at most is narrowing, as the Indications for Use for the Device does not refer to "or initial layer" as does the Indications for Use for the Predicate Device.

Device Description

The Device is a light-cured, flowable, resin-based sealant for prophylactic sealing of pits and fissures as well as for micro-restorative procedures. The Device is light-curable under standard power (600 mW/cm²) and contains filling agents which cause the Device to release and recharge beneficial ions (fluoride and calcium).

The device is intended for use by licensed healthcare professionals only. The device does not come sterilized and is not intended to be sterilized prior to use.

Indications for Use

The Device is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations.

Comparison of Technological Characteristics

The Device has the same basic technological characteristics in terms of design, material, and chemical composition as the Predicate Device and is substantially equivalent thereto as shown below:

Page 7

Shelf Life Testing

The Device has been tested to establish a shelf life of two years.

PropertyParkell Device Pit and Fissure Sealant (Parkell, Inc.) K243254Predicate Device 510(k) no. K161580, BioCoat, filed as "Premier Sealant" (Premier Dental Company)
Intended usesParkell Pit and Fissure Sealant is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations.Premier Sealant, a prescription only medical device used for prophylactic sealing of pits and fissures. It may also be used for micro-restorative or "initial layer" of composite restorations
Classification Product CodeEBCEBC
Regulation Number872.3765872.3765
Principle of operationProphylactic sealing of pits and fissuresProphylactic sealing of pits and fissures
Material formLow viscosity pasteLow viscosity paste
Polymerization MethodLight-CureLight-Cure
Curing Time20 sec20 sec
Delivery System1.2 mL syringe1.2 mL syringe
Bond Strength to Dry Enamel (MPa) Pass/Fail Criteria: ≥10 MPa≥10 MPa≥10 MPa
Compressive Strength (MPa) Pass/Fail Criteria: ≥200MPa≥200MPa≥200MPa
Flexural Strength (MPa) Pass/Fail Criteria: ≥90MPa≥90MPa≥90MPa
Depth of cure (mm) Pass/Fail Criteria: 1.5-2.5mm1.5-2.5mm1.5-2.5mm
Shrinkage (%) Pass/Fail Criteria: ≤8%≤8%≤8%
Viscosity (initial) (Pa.S) Pass/Fail Criteria: 1-5Pa.s.1-5Pa.s1-5Pa.s
Beneficial Ion Release and RechargeFluoride CalciumFluoride Calcium Phosphate
AccessoriesApplicator tipApplicator tip

Page 8

Material And Chemical Composition

The Device does come into direct contact with the patient. Though the precise composition is confidential, the Device contains monomers, fillers, photoinitiator/coactivators, a calcium donor, inhibitor, fluorescence agent, and colorant, and is substantially equivalent to the Predicate Device in terms of ingredients and biocompatibility.

Performance Data Summary

Non-clinical testing of the physical properties of the Device was conducted. There were no clinical tests performed for the Device.

A comparison of the Device with the Predicate Device is shown in the table below:

Page 9

PropertyParkell Device Pit and Fissure Sealant (Parkell, Inc.) K243254Predicate Device 510(k) no. K161580, BioCoat, filed as "Premier Sealant" (Premier Dental Company)
Intended usesParkell Pit and Fissure Sealant is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations.Premier Sealant, a prescription only medical device used for prophylactic sealing of pits and fissures. It may also be used for micro-restorative or "initial layer" of composite restorations
Classification Product CodeEBCEBC
Regulation Number872.3765872.3765
Principle of operationProphylactic sealing of pits and fissuresProphylactic sealing of pits and fissures
Material formLow viscosity pasteLow viscosity paste
Polymerization MethodLight-CureLight-Cure
Curing Time20 sec20 sec
Delivery System1.2 mL syringe1.2 mL syringe
Bond Strength to Dry Enamel (MPa) Pass/Fail Criteria: ≥10 MPa≥10 MPa≥10 MPa
Compressive Strength (MPa) Pass/Fail Criteria: ≥200MPa≥200MPa≥200MPa
Flexural Strength (MPa) Pass/Fail Criteria: ≥90MPa≥90MPa≥90MPa
Depth of cure (mm) Pass/Fail Criteria: 1.5-2.5mm1.5-2.5mm1.5-2.5mm
Shrinkage (%) Pass/Fail Criteria: ≤8%≤8%≤8%
Viscosity (initial) (Pa.S) Pass/Fail Criteria 1-5Pa.s.1-5Pa.s1-5Pa.s
Beneficial Ion Release and RechargeFluoride CalciumFluoride Calcium Phosphate
AccessoriesApplicator tipApplicator tip

Conclusion

Based on the non-clinical testing conducted of the physical properties of the Device in comparison to the Predicate Device identified above, and based on the biocompatibility of ingredients for the Device, Parkell concludes that the Device is substantially equivalent to the Predicate Device.

§ 872.3765 Pit and fissure sealant and conditioner.

(a)
Identification. A pit and fissure sealant and conditioner is a device composed of resin, such as polymethylmethacrylate, intended for use primarily in young children to seal pit and fissure depressions (faults in the enamel) in the biting surfaces of teeth to prevent cavities.(b)
Classification. Class II.