K Number
K243172

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-06-18

(261 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Halyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs or Sterile Singles, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:

  • Azacitidine (25 mg/ml)
  • Bendamustine HCl (5 mg/ml)
  • Bleomycin Sulfate (15 mg/ml)
  • Bortezomib (1 mg/ml)
  • Busulfan (6 mg/ml)
  • Capecitabine (26 mg/ml)
  • Carboplatin (10 mg/ml)
  • Carfilzomib (2 mg/ml)
  • Cetuximab (2 mg/ml)
  • Cisplatin (1 mg/ml)
  • Cladribine (1 mg/ml)
  • Cyclophosphamide (20 mg/ml)
  • Cytarabine HCl (100 mg/ml)
  • Dacarbazine (10 mg/ml)
  • Dactinomycin (0.5 mg/ml)
  • Daunorubicin HCl (5 mg/ml)
  • Decitabine (5 mg/ml)
  • Docetaxel (10 mg/ml)
  • Doxorubicin HCl (2 mg/ml)
  • Epirubicin HCl (2 mg/ml)
  • Etoposide (20 mg/ml)
  • Fludarabine Phosphate (25 mg/ml)
  • Fluorouracil (50 mg/ml)
  • Fulvestrant (50 mg/ml)
  • Gemcitabine HCl (38 mg/ml)
  • Idarubicin HCl (1 mg/ml)
  • Ifosfamide (50 mg/ml)
  • Irinotecan HCl (20 mg/ml)
  • Leuprolide Acetate (5 mg/ml)
  • Mechlorethamine HCl (1 mg/ml)
  • Melphalan HCl (5 mg/ml)
  • Methotrexate (25 mg/ml)
  • Mitomycin C (0.5 mg/ml)
  • Mitoxantrone HCl (2 mg/ml)
  • Oxaliplatin (5 mg/ml)
  • Paclitaxel (6 mg/ml)
  • Pemetrexed (25 mg/ml)
  • Raltitrexed (0.5 mg/ml)
  • Rituximab (10 mg/ml)
  • Temsirolimus (25 mg/ml)
  • Topotecan HCl (1 mg/ml)
  • Trisenox (Arsenic Trioxide) (1 mg/ml)
  • Vinblastine Sulfate (1 mg/ml)
  • Vincristine (1 mg/ml)
  • Vinorelbine Tartrate (10 mg/ml)

The following chemotherapy drugs and concentration showed breakthrough detected in less than 100 minutes:

  • Carmustine (3.3 mg/ml) No breakthrough up to 44.5 minutes.
  • Thiotepa (10 mg/ml) No breakthrough up to 99.1 minutes.
  • Warning- Not for use with Carmustine and ThioTEPA

The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes:

  • Fentanyl Citrate Injection (100 mcg/2 ml)
  • Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution

The following hazardous drugs and concentration had NO breakthrough detected up to 240 minutes:

  • Chloroquine (50 mg/ml)
  • Cyclosporin A (100 mg/ml)
  • Cytovene (10 mg/ml)
  • Retrovir (10 mg/ml)
  • Triclosan (2 mg/ml)
  • Zoledronic Acid (0.8 mg/ml)
Device Description

The Halyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs or Sterile Singles, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid are disposable, 9.5" purple‐colored, chlorinated, nitrile, powder‐free, textured fingertip, ambidextrous, sterile patient examination gloves.

AI/ML Overview

The provided text is a 510(k) clearance letter and summary for medical examination gloves, not an AI/ML device. Therefore, it does not contain information related to AI/ML specific acceptance criteria or study details such as:

  • Sample size used for the test set and data provenance (e.g., country of origin, retrospective/prospective).
  • Number of experts used to establish ground truth and their qualifications.
  • Adjudication method for the test set.
  • Multi-Reader Multi-Case (MRMC) comparative effectiveness study.
  • Standalone (algorithm-only) performance.
  • Type of ground truth used (expert consensus, pathology, outcomes data).
  • Sample size for the training set.
  • How ground truth for the training set was established.

Instead, the document focuses on the physical and chemical properties of the gloves and their resistance to various chemicals, using established ASTM and ISO standards for medical gloves.

However, I can extract the acceptance criteria and reported device performance for these medical gloves based on the provided document.


Analysis of Acceptance Criteria and Device Performance for Medical Gloves

The document details the non-clinical testing performed on the Halyard Purple Nitrile* Powder-Free Exam Gloves to demonstrate their safety and effectiveness. The acceptance criteria and the results of these tests are clearly outlined.

1. Table of Acceptance Criteria and Reported Device Performance

TestStandardAcceptance CriteriaReported Device Performance
DimensionsASTM D 6319Length ≥230 mm Palm Width Size Small: 70 - 90 mm Med: 85–105 mm Large: 100 - 120 mm Finger thickness ≥0.05 mm Palm thickness ≥0.05 mm Cuff thickness ≥0.05 mmMeets requirements
Physical Properties (Tensile Strength & Elongation)ASTM D 6319AQL 4.0 Before Aging: Tensile Strength: ≥14 MPa Ultimate elongation: ≥500% After Aging: Tensile Strength: ≥14 MPa Ultimate elongation: ≥400%Meets requirements. The physical dimensions and properties (tensile strength before and after aging, and elongation before and after aging) met the requirements of the standard.
Freedom from PinholesASTM D 6319, ASTM D 5151AQL 2.5% No leakageMeets requirements. Testing of the subject device shows it meets the 2.5% AQL requirement in the standards for leakage.
SterilityANSI/AAMI/ISO 1113710⁻⁶10⁻⁶
Powder FreeASTM D 6124, ASTM D 6319≤ 2 mg / gloveMeets requirements. Residual powder on the subject device is an average of 0.4 mg/glove, which is within the powder-free limit of < 2 mg maximum powder per glove.
Test for IrritationISO 10993, Part 23Grade 1 (Implies non-irritant)Under the conditions of the study, the device is not an irritant.
Test for Acute Systemic ToxicityISO 10993, Part 11No animals treated with test extracts exhibit greater reaction than control animals.Under the conditions of the study, no evidence of acute systemic toxicity.
Test for Skin SensitizationISO 10993, Part 10Grade < 1.0 (Implies non-sensitizer)Under the conditions of the study, the device is not a sensitizer.
Resistance to Permeation by Chemotherapy DrugsASTM D6978‐05No breakthrough for up to 240 minutes for listed drugs.51 Drugs tested showed minimum breakthrough detection time up to 240 minutes, including various concentrations of: Azacitidine, Bendamustine HCl, Bleomycin Sulfate, Bortezomib, Busulfan, Capecitabine, Carboplatin, Carfilzomib, Cetuximab, Cisplatin, Cladribine, Cyclophosphamide, Cytarabine HCl, Dacarbazine, Dactinomycin, Daunorubicin HCl, Decitabine, Docetaxel, Doxorubicin HCl, Epirubicin HCl, Etoposide, Fludarabine Phosphate, Fluorouracil, Fulvestrant, Gemcitabine HCl, Idarubicin HCl, Ifosfamide, Irinotecan HCl, Leuprolide Acetate, Mechlorethamine HCl, Melphalan HCl, Methotrexate, Mitomycin C, Mitoxantrone HCl, Oxaliplatin, Paclitaxel, Pemetrexed, Raltitrexed, Rituximab, Temsirolimus, Topotecan HCl, Trisenox (Arsenic Trioxide), Vinblastine Sulfate, Vincristine, Vinorelbine Tartrate.Exceptions (breakthrough detected in less than 100 minutes):- Carmustine (3.3 mg/ml): No breakthrough up to 44.5 minutes (Warning: Not for use).- Thiotepa (10 mg/ml): No breakthrough up to 99.1 minutes (Warning: Not for use).
Resistance to Permeation by Hazardous Drugs (Opioids)ASTM D6978‐05No breakthrough for up to 240 minutes for listed drugs.No breakthrough detected up to 240 minutes for:- Fentanyl Citrate Injection (100 mcg/2 ml)- Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution
Resistance to Permeation by Other Hazardous DrugsASTM D6978‐05No breakthrough for up to 240 minutes for listed drugs.No breakthrough detected up to 240 minutes for:- Chloroquine (50 mg/ml)- Cyclosporin A (100 mg/ml)- Cytovene (10 mg/ml)- Retrovir (10 mg/ml)- Triclosan (2 mg/ml)- Zoledronic Acid (0.8 mg/ml)

2. Sample size used for the test set and the data provenance

The document indicates that testing was performed according to specific ASTM and ISO standards (e.g., ASTM D6978-05, ASTM D5151, ASTM D6124, ASTM D6319, ISO 10993 Parts 10, 11, 23, ANSI/AAMI/ISO 11137). These standards inherently define the sample sizes and methodologies for testing. For instance, AQL (Acceptable Quality Level) values like 2.5% and 4.0% are mentioned for pinholes and physical properties, which dictates sampling plans for quality control.

The data provenance is not explicitly stated as "country of origin" or "retrospective/prospective" in the context of an AI/ML study. It is implied to be laboratory testing data generated prospectively for the purpose of this 510(k) submission, confirming product specifications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This is not applicable to the evaluation of medical gloves. The "ground truth" for glove performance is established by adherence to recognized international standards (ASTM, ISO, ANSI/AAMI) for physical, chemical permeation, and biocompatibility testing, rather than expert human interpretation of medical images or data. The testing is objective and quantitative.

4. Adjudication method for the test set

Not applicable. Testing involves standardized objective measurements, not subjective human evaluations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done

Not applicable. MRMC studies are relevant for imaging diagnostics or other AI/ML applications where human interpretation is a key component. This document describes the testing of a physical medical device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm, but a physical product.

7. The type of ground truth used

The "ground truth" used is defined by the objective measurement criteria and methodologies specified in the cited international standards (ASTM D6978-05, ASTM D5151, ASTM D6319, ASTM D6124, ISO 10993, ANSI/AAMI/ISO 11137). For example, for chemical permeation, the ground truth is the chemical breakthrough time measured according to ASTM D6978-05. For biocompatibility, it's the observed biological reactions in animal models or in vitro tests compared to established toxicity/irritation/sensitization thresholds.

8. The sample size for the training set

Not applicable. This is a physical device, not an AI/ML model that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As above, this document describes a physical medical device, not an AI/ML system.

FDA 510(k) Clearance Letter - K243172

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 18, 2025

O&M Halyard, Inc.
Caitlin Senter
Director, Global Regulatory Affairs
9120 Lockwood Blvd
Mechanicsville, Virginia 23116

Re: K243172
Trade/Device Name: Halyard Purple Nitrile* Powder-Free Exam Gloves, Sterile Pairs; Halyard Purple Nitrile* Powder-Free Exam Gloves, Sterile Singles
Regulation Number: 21 CFR 880.6250
Regulation Name: Non-powdered patient examination glove
Regulatory Class: Class I, reserved
Product Code: LZA, LZC, OPJ, QDO
Dated: September 30, 2024
Received: May 9, 2025

Dear Caitlin Senter:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K243172 - Caitlin Senter Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243172 - Caitlin Senter Page 3

Sincerely,

ALLAN GUAN -S

For Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026

Indications for Use

See PRA Statement below.

510(k) Number (if known)
K243172

Device Name
Halyard Purple Nitrile* Powder-Free Exam Gloves, Sterile Pairs;
Halyard Purple Nitrile* Powder-Free Exam Gloves, Sterile Singles

Indications for Use (Describe)
The Halyard Purple Nitrile* Powder-Free Exam Gloves, Sterile Pairs or Sterile Singles, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:

  • Azacitidine (25 mg/ml)
  • Bendamustine HCl (5 mg/ml)
  • Bleomycin Sulfate (15 mg/ml)
  • Bortezomib (1 mg/ml)
  • Busulfan (6 mg/ml)
  • Capecitabine (26 mg/ml)
  • Carboplatin (10 mg/ml)
  • Carfilzomib (2 mg/ml)
  • Cetuximab (2 mg/ml)
  • Cisplatin (1 mg/ml)
  • Cladribine (1 mg/ml)
  • Cyclophosphamide (20 mg/ml)
  • Cytarabine HCl (100 mg/ml)
  • Dacarbazine (10 mg/ml)
  • Dactinomycin (0.5 mg/ml)
  • Daunorubicin HCl (5 mg/ml)
  • Decitabine (5 mg/ml)
  • Docetaxel (10 mg/ml)
  • Doxorubicin HCl (2 mg/ml)
  • Epirubicin HCl (2 mg/ml)
  • Etoposide (20 mg/ml)
  • Fludarabine Phosphate (25 mg/ml)
  • Fluorouracil (50 mg/ml)
  • Fulvestrant (50 mg/ml)
  • Gemcitabine HCl (38 mg/ml)
  • Idarubicin HCl (1 mg/ml)
  • Ifosfamide (50 mg/ml)
  • Irinotecan HCl (20 mg/ml)
  • Leuprolide Acetate (5 mg/ml)
  • Mechlorethamine HCl (1 mg/ml)
  • Melphalan HCl (5 mg/ml)
  • Methotrexate (25 mg/ml)
  • Mitomycin C (0.5 mg/ml)
  • Mitoxantrone HCl (2 mg/ml)
  • Oxaliplatin (5 mg/ml)
  • Paclitaxel (6 mg/ml)
  • Pemetrexed (25 mg/ml)

FORM FDA 3881 (8/23) Page 1 of 2

Page 5

  • Raltitrexed (0.5 mg/ml)
  • Rituximab (10 mg/ml)
  • Temsirolimus (25 mg/ml)
  • Topotecan HCl (1 mg/ml)
  • Trisenox (Arsenic Trioxide) (1 mg/ml)
  • Vinblastine Sulfate (1 mg/ml)
  • Vincristine (1 mg/ml)
  • Vinorelbine Tartrate (10 mg/ml)

The following chemotherapy drugs and concentration showed breakthrough detected in less than 100 minutes:

  • Carmustine (3.3 mg/ml) No breakthrough up to 44.5 minutes.
  • Thiotepa (10 mg/ml) No breakthrough up to 99.1 minutes.
  • Warning- Not for use with Carmustine and ThioTEPA

The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes:

  • Fentanyl Citrate Injection (100 mcg/2 ml)
  • Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution

The following hazardous drugs and concentration had NO breakthrough detected up to 240 minutes:

  • Chloroquine (50 mg/ml)
  • Cyclosporin A (100 mg/ml)
  • Cytovene (10 mg/ml)
  • Retrovir (10 mg/ml)
  • Triclosan (2 mg/ml)
  • Zoledronic Acid (0.8 mg/ml)

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 2 of 2

Page 6

510(k) Summary for K243172

This summary of 510(k) K243172 is being submitted in accordance with 21 CFR 807.92.

Date Summary was PreparedJune 18, 2025
510(k) SubmitterO & M Halyard, Inc.1220 Old Alpharetta Rd., Ste. 320Alpharetta, GA 30005
Primary Contact for this 510(k) SubmissionCaitlin Senter, MS, RACTel: 678‐221‐7330Email: caitlin.senter@owens‐minor.com
Marketed Device Trade NameHalyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs;Halyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Singles
Device Submission Trade name and DescriptionHalyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs;Halyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Singles
Device Common NameMedical Exam Gloves
Device Product Code and Classification NameLZA Class I, 21 CFR §880.6250 Polymer Patient Examination GloveLZC Class I, 21 CFR §880.6250 Medical Glove, SpecialtyOPJ Class I, 21 CFR §880.6250 Medical Gloves with Chemotherapy Labeling Claims ‐ Test For Use with Chemotherapy DrugsQDO Class I, 21 CFR §880.6250 Fentanyl and Other Opioid Protection Glove
Predicate DeviceKimberly‐Clark Purple Nitrile XTRA* Sterile Powder‐Free Exam Glove (Chemotherapy Glove) ‐ 12 Sterile Pairs (K102032)
Reference DevicesHalyard Purple Nitrile, Powder‐Free Exam Gloves Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid (K213929)
Subject Device DescriptionThe Halyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs or Sterile Singles, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid are disposable, 9.5" purple‐colored, chlorinated, nitrile, powder‐free, textured fingertip, ambidextrous, sterile patient examination gloves.

Page 7

| Indications for Use | Halyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs or Sterile Singles, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:- Azacitidine (25 mg/ml)- Bendamustine HCl (5 mg/ml)- Bleomycin Sulfate (15 mg/ml)- Bortezomib (1 mg/ml)- Busulfan (6 mg/ml)- Capecitabine (26 mg/ml)- Carboplatin (10 mg/ml)- Carfilzomib (2 mg/ml)- Cetuximab (2 mg/ml)- Cisplatin (1 mg/ml)- Cladribine (1 mg/ml)- Cyclophosphamide (20 mg/ml)- Cytarabine HCl (100 mg/ml)- Dacarbazine (10 mg/ml)- Dactinomycin (0.5 mg/ml)- Daunorubicin HCl (5 mg/ml)- Decitabine (5 mg/ml)- Docetaxel (10 mg/ml)- Doxorubicin HCl (2 mg/ml)- Epirubicin HCl (2 mg/ml)- Etoposide (20 mg/ml)- Fludarabine Phosphate (25 mg/ml)- Fluorouracil (50 mg/ml)- Fulvestrant (50 mg/ml)- Gemcitabine HCl (38 mg/ml)- Idarubicin HCl (1 mg/ml)- Ifosfamide (50 mg/ml)- Irinotecan HCl (20 mg/ml)- Leuprolide Acetate (5 mg/ml)- Mechlorethamine HCl (1 mg/ml)- Melphalan HCl (5 mg/ml)- Methotrexate (25 mg/ml)- Mitomycin C (0.5 mg/ml)- Mitoxantrone HCl (2 mg/ml)- Oxaliplatin (5 mg/ml)- Paclitaxel (6 mg/ml)- Pemetrexed (25 mg/ml)- Raltitrexed (0.5 mg/ml)- Rituximab (10 mg/ml)- Temsirolimus (25 mg/ml)- Topotecan HCl (1 mg/ml) |

Page 8

  • Trisenox (Arsenic Trioxide) (1 mg/ml)
  • Vinblastine Sulfate (1 mg/ml)
  • Vincristine (1 mg/ml)
  • Vinorelbine Tartrate (10 mg/ml)

The following chemotherapy drugs and concentration showed breakthrough detected in less than 100 minutes:

  • Carmustine (3.3 mg/ml) No breakthrough up to 44.5 minutes.
  • Thiotepa (10 mg/ml) No breakthrough up to 99.1 minutes.
  • Warning- Not for use with Carmustine and ThioTEPA

The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes:

  • Fentanyl Citrate Injection (100 mcg/2 ml)
  • Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution

The following hazardous drugs and concentration had NO breakthrough detected up to 240 minutes:

  • Chloroquine (50 mg/ml)
  • Cyclosporin A (100 mg/ml)
  • Cytovene (10 mg/ml)
  • Retrovir (10 mg/ml)
  • Triclosan (2 mg/ml)
  • Zoledronic Acid (0.8 mg/ml)

Page 9

Subject DevicePredicate Device (K102032)Reference Device (K213929)Comparison
FDA Product CodeLZA, LZC, OPJ, QDOLZA, LZCLZA, OPJ, QDO
FDA ClassificationClass IClass IClass I
Regulation Number880.6250880.6250880.6250
Common NameMedical Exam GloveMedical Exam GloveMedical Exam Glove
Device Trade NameHalyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs; Halyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile SinglesKimberly‐Clark Purple Nitrile XTRA* Sterile Powder‐Free Exam Glove (Chemotherapy Glove) ‐ 12 Sterile PairsHalyard Purple Nitrile, Powder‐Free Exam Gloves Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid
Intended Use/Indications for UseHalyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs or Sterile Singles, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Azacitidine (25 mg/ml) Bendamustine HCl (5 mg/ml) Bleomycin Sulfate (15A powder‐free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978‐05: The following drugs had NO breakthrough detected up to 240 minutes: Bleomycin Sulfate (15 mg/ml) Busulfan (6 mg/ml) Carboplatin (10 mg/ml) Cisplatin (1 mg/ml) Cyclophosphamide (20 mg/ml) Cytarabine HCl (100 mg/ml) Dacarbazine (10 mg/ml) Daunorubicin HCl (5 mg/ml) Docetaxel (10 mg/ml) Doxorubicin HCl (Adriamycin) (2 mg/ml)Halyard Purple Nitrile* Powder‐Free Exam Gloves, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Azacitidine (25

Page 10

Subject DevicePredicate Device (K102032)Reference Device (K213929)Comparison
mg/ml) Bortezomib (1 mg/ml) Busulfan (6 mg/ml) Capecitabine (26 mg/ml) Carboplatin (10 mg/ml) Carfilzomib (2 mg/ml) Cetuximab (2 mg/ml) Cisplatin (1 mg/ml) Cladribine (1 mg/ml) Cyclophosphamide (20 mg/ml) Cytarabine HCl (100 mg/ml) Dacarbazine (10 mg/ml) Dactinomycin (0.5 mg/ml) Daunorubicin HCl (5 mg/ml) Decitabine (5 mg/ml) Docetaxel (10 mg/ml) Doxorubicin HCl (2 mg/ml) Epirubicin HCl (2 mg/ml) Etoposide (20 mg/ml) Fludarabine Phosphate (25 mg/ml) Fluorouracil (50 mg/ml) Fulvestrant (50 mg/ml) Gemcitabine HCl (38 mg/ml) Idarubicin HCl (1 mg/ml) Ifosfamide (50 mg/ml) Irinotecan HCl (20 mg/ml) Leuprolide Acetate (5 mg/ml) Mechlorethamine HCl (1 mg/ml) Melphalan HCl (5 mg/ml) Methotrexate (25 mg/ml) Mitomycin C (0.5Ellence (Epirubicin) (2 mg/ml) Etoposide (20 mg/ml) Fludarabine (25 mg/ml) Fluorouracil (adrucil) (50 mg/ml) Gemcitabine (38.0 mg/ml)mg/ml) Bendamustine HCl (5 mg/ml) Bleomycin Sulfate (15 mg/ml) Bortezomib (1 mg/ml) Busulfan (6 mg/ml) Capecitabine (26 mg/ml) Carboplatin (10 mg/ml) Carlzomib (2 mg/ml) Cetuximab (2 mg/ml) Chloroquine (50 mg/ml) Cisplatin (1 mg/ml) Cladribine (1 mg/ml) Cyclophosphamide (20 mg/ml) Cyclosporin A (100 mg/ml) Cytarabine (Cytosine) (100 mg/ml) Cytovene (Ganciclovir) (10 mg/ml) Dacarbazine (DTIC) (10 mg/ml) Dactinomycin (0.5 mg/ml) Daunorubicin HCl (5 mg/ml) Decitabine (5 mg/ml) Docetaxel (10 mg/ml) Doxorubicin HCl (2 mg/ml) Epirubicin HCl (Ellence) (2 mg/ml) Etoposide (Toposar) (20 mg/ml) Fludarabine (25 mg/ml) 5‐Fluorouracil (50 mg/ml) Fulvestrant (50 mg/ml) Gemcitabine (38 mg/ml)

Page 11

Subject DevicePredicate Device (K102032)Reference Device (K213929)Comparison
mg/ml) Mitoxantrone HCl (2 mg/ml) Oxaliplatin (5 mg/ml) Paclitaxel (6 mg/ml) Pemetrexed (25 mg/ml) Raltitrexed (0.5 mg/ml) Rituximab (10 mg/ml) Temsirolimus (25 mg/ml) Topotecan HCl (1 mg/ml) Trisenox (Arsenic Trioxide) (1 mg/ml) Vinblastine Sulfate (1 mg/ml) Vincristine (1 mg/ml) Vinorelbine Tartrate (10 mg/ml) The following chemotherapy drugs and concentration showed breakthrough detected in less than 100 minutes: Carmustine (3.3 mg/ml) No breakthrough up to 44.5 minutes. Thiotepa (10 mg/ml) No breakthrough up to 99.1 minutes. Warning‐ Not for use with Carmustine and ThioTEPA The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2 ml) Simulated Gastric AcidIdarubicin (1 mg/ml) Ifosfamide (50 mg/ml) Irinotecan HCl (20 mg/ml) Leuprolide Acetate Salt (5 mg/ml) Mechlorethamine HCl (1 mg/ml) Melphalan (5 mg/ml) Methotrexate (25 mg/ml) MitomycinC (0.5 mg/ml) Mitoxantrone (2 mg/ml) Oxaliplatin (5 mg/ml) Paclitaxel (6 mg/ml) Pemetrexed (25 mg/ml) Raltitrexed (0.5 mg/ml) Retrovir (10 mg/ml) Rituximab (10 mg/ml) Temsirolimus (25 mg/ml) Topotecan HCl (1 mg/ml) Triclosan (2 mg/ml) Trisenox (1 mg/ml) Vinblastine Sulfate (1 mg/ml) Vincristine (1 mg/ml) Vinorelbine (10 mg/ml) Zoledronic Acid (0.8 mg/ml) The following chemotherapy drugs and concentration showed breakthrough detected in less than 90 minutes: Carmustine (3.3 mg/ml) No breakthrough up toIdarubicin (1 mg/ml) Ifosfamide (50 mg/ml) Irinotecan HCl (20 mg/ml) Leuprolide Acetate Salt (5 mg/ml) Mechlorethamine HCl (1 mg/ml) Melphalan (5 mg/ml) Methotrexate (25 mg/ml) Mitomycin C (0.5 mg/ml) Mitoxantrone (2 mg/ml) Oxaliplatin (5 mg/ml) Paclitaxel (6 mg/ml) Pemetrexed (25 mg/ml) Raltitrexed (0.5 mg/ml) Retrovir (10 mg/ml) Rituximab (10 mg/ml) Temsirolimus (25 mg/ml) Topotecan HCl (1 mg/ml) Triclosan (2 mg/ml) Trisenox (1 mg/ml) Vinblastine Sulfate (1 mg/ml) Vincristine (1 mg/ml) Vinorelbine (10 mg/ml) Zoledronic Acid (0.8 mg/ml) The following chemotherapy drugs and concentration showed breakthrough detected in less than 90 minutes: Carmustine (3.3 mg/ml) No

Page 12

Subject DevicePredicate Device (K102032)Reference Device (K213929)Comparison
Fluid/Fentanyl Citrate Injection Mix 50/50 Solution The following hazardous drugs and concentration had NO breakthrough detected up to 240 minutes: Chloroquine (50 mg/ml) Cyclosporin A (100 mg/ml) Cytovene (10 mg/ml) Retrovir (10 mg/ml) Triclosan (2 mg/ml) Zoledronic Acid (0.8 mg/ml)55.3 minutes. Thiotepa (10 mg/ml) No breakthrough up to 78.8 minutes. Warning‐ Not for use with Carmustine and ThioTEPA No breakthrough was detected up to 240 minutes for Fentanyl Citrate Injection (100 mcg/2 ml) and Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solutionbreakthrough up to 55.3 minutes. Thiotepa (10 mg/ml) No breakthrough up to 78.8 minutes. Warning‐ Not for use with Carmustine and ThioTEPA No breakthrough was detected up to 240 minutes for Fentanyl Citrate Injection (100 mcg/2 ml) and Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution
Technological CharacteristicsColored, 9.5 inch, chlorinated, nitrile, powder‐free, textured fingertip, ambidextrous, sterile patient examination gloveColored, 9.5 inch, chlorinated, nitrile, powder‐free, textured fingertip, ambidextrous, sterile patient examination gloveColored, 9.5 inch, chlorinated, nitrile, powder‐free, textured fingertips, ambidextrous, non‐sterile patient examination glove
Sizes of glovesS, M, LS, M, LXS, S, M, L, XL
ColorPurplePurplePurple

Page 13

Subject DevicePredicate Device (K102032)Reference Device (K213929)Comparison
TextureTextured fingertipsTextured fingertipsTextured fingertips
SterilitySterileSterileNon‐Sterile
BiocompatibilityBased on ISO 10993, Part 11 Biological Evaluation of Medical Devices – Test for systemic toxicity, the test article was considered non‐toxic. Meets the acceptance criteria. Based on ISO 10993, Part 23‐ Biological Evaluation of Medical Devices – Test for irritation, the test article was considered non‐irritant. Meets the acceptance criteria. Based on ISO 10993, Part 10 ‐ Biological Evaluation of Medical Devices – Test for skin sensitization, the test article was considered a non‐sensitizer. Meets the acceptance criteria.Based ISO 10993 Biological Evaluation of Medical devices – Test for systemic toxicity, the test article was considered non‐toxic. Meets the acceptance criteria. Based on ISO 10993, Part 10‐ Biological Evaluation of Medical Devices – Test for irritation, the test article was considered non‐irritant. Meets the acceptance criteria. Based on ISO 10993, Part 10 ‐ Biological Evaluation of Medical Devices – Test for skin sensitization, the test article was considered non‐sensitizer. Meets the acceptance criteria.Based ISO 10993 Biological Evaluation of Medical devices – Test for systemic toxicity, the test article was considered non‐toxic. Meets the acceptance criteria. Based on ISO 10993, Part 10‐ Biological Evaluation of Medical Devices – Test for irritation, the test article was considered non‐irritant. Meets the acceptance criteria. Based on ISO 10993, Part 10 ‐ Biological Evaluation of Medical Devices – Test for skin sensitization, the test article was considered non‐sensitizer. Meets the acceptance criteria.
StandardSubject DevicePredicate Device (K102032)Reference Device (K213929)Comparison
ASTM D6978‐05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy DrugsThe following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Azacitidine (25 mg/ml)These chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978‐05: The following drugs had NO breakthrough detected up to 240 minutes: Bleomycin Sulfate (15 mg/ml)The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Azacitidine (25 mg/ml)Similar

Page 14

StandardSubject DevicePredicate Device (K102032)Reference Device (K213929)Comparison
Bendamustine HCl (5 mg/ml) Bleomycin Sulfate (15 mg/ml) Bortezomib (1 mg/ml) Busulfan (6 mg/ml) Capecitabine (26 mg/ml) Carboplatin (10 mg/ml) Carfilzomib (2 mg/ml) Cetuximab (2 mg/ml) Cisplatin (1 mg/ml) Cladribine (1 mg/ml) Cyclophosphamide (20 mg/ml) Cytarabine HCl (100 mg/ml) Dacarbazine (10 mg/ml) Dactinomycin (0.5 mg/ml) Daunorubicin HCl (5 mg/ml) Decitabine (5 mg/ml) Docetaxel (10 mg/ml) Doxorubicin HCl (2 mg/ml) Epirubicin HCl (2 mg/ml) Etoposide (20 mg/ml) Fludarabine Phosphate (25 mg/ml) Fluorouracil (50 mg/ml) Fulvestrant (50 mg/ml) Gemcitabine HCl (38 mg/ml) Idarubicin HCl (1 mg/ml) Ifosfamide (50 mg/ml) Irinotecan HCl (20 mg/ml) Leuprolide Acetate (5 mg/ml) Mechlorethamine HCl (1 mg/ml) Melphalan HCl (5 mg/ml)Busulfan (6 mg/ml) Carboplatin (10 mg/ml) Cisplatin (1 mg/ml) Cyclophosphamide (20 mg/ml) Cytarabine HCl (100 mg/ml) Dacarbazine (10 mg/ml) Daunorubicin HCl (5 mg/ml) Docetaxel (10 mg/ml) Doxorubicin HCl (Adriamycin) (2 mg/ml) Ellence (Epirubicin) (2 mg/ml) Etoposide (20 mg/ml) Fludarabine (25 mg/ml) Fluorouracil (adrucil) (50 mg/ml) Gemcitabine (38.0 mg/ml)Bendamustine HCl (5 mg/ml) Bleomycin Sulfate (15 mg/ml) Bortezomib (1 mg/ml) Busulfan (6 mg/ml) Capecitabine (26 mg/ml) Carboplatin (10 mg/ml) Carlzomib (2 mg/ml) Cetuximab (2 mg/ml) Chloroquine (50 mg/ml) Cisplatin (1 mg/ml) Cladribine (1 mg/ml) Cyclophosphamide (20 mg/ml) Cyclosporin A (100 mg/ml) Cytarabine (Cytosine) (100 mg/ml) Cytovene (Ganciclovir) (10 mg/ml) Dacarbazine (DTIC) (10 mg/ml) Dactinomycin (0.5 mg/ml) Daunorubicin HCl (5 mg/ml) Decitabine (5 mg/ml) Docetaxel (10 mg/ml) Doxorubicin HCl (2 mg/ml) Epirubicin HCl (Ellence) (2 mg/ml) Etoposide (Toposar) (20 mg/ml) Fludarabine (25 mg/ml) 5‐Fluorouracil (50 mg/ml) Fulvestrant (50 mg/ml) Gemcitabine (38 mg/ml)

Page 15

StandardSubject DevicePredicate Device (K102032)Reference Device (K213929)Comparison
Methotrexate (25 mg/ml) Mitomycin C (0.5 mg/ml) Mitoxantrone HCl (2 mg/ml) Oxaliplatin (5 mg/ml) Paclitaxel (6 mg/ml) Pemetrexed (25 mg/ml) Raltitrexed (0.5 mg/ml) Rituximab (10 mg/ml) Temsirolimus (25 mg/ml) Topotecan HCl (1 mg/ml) Trisenox (Arsenic Trioxide) (1 mg/ml) Vinblastine Sulfate (1 mg/ml) Vincristine (1 mg/ml) Vinorelbine Tartrate (10 mg/ml) The following chemotherapy drugs and concentration showed breakthrough detected in less than 100 minutes: Carmustine (3.3 mg/ml) No breakthrough up to 44.5 minutes. Thiotepa (10 mg/ml) No breakthrough up to 99.1 minutes. Warning‐ Not for use with Carmustine and ThioTEPA The following hazardous drugs and concentration had NO breakthrough detected up to 240 minutes: Chloroquine (50 mg/ml) Cyclosporin A (100Idarubicin (1 mg/ml) Ifosfamide (50 mg/ml) Irinotecan HCl (20 mg/ml) Leuprolide Acetate Salt (5 mg/ml) Mechlorethamine HCl (1 mg/ml) Melphalan (5 mg/ml) Methotrexate (25 mg/ml) MitomycinC (0.5 mg/ml) Mitoxantrone (2 mg/ml) Oxaliplatin (5 mg/ml) Paclitaxel (6 mg/ml) Pemetrexed (25 mg/ml) Raltitrexed (0.5 mg/ml) Retrovir (10 mg/ml) Rituximab (10 mg/ml) Temsirolimus (25 mg/ml) Topotecan HCl (1 mg/ml) Triclosan (2 mg/ml) Trisenox (1 mg/ml) Vinblastine Sulfate (1 mg/ml) Vincristine (1 mg/ml) Vinorelbine (10 mg/ml) Zoledronic Acid (0.8 mg/ml) The following chemotherapy drugs and concentration showed breakthrough detected in less than 90 minutes: Carmustine (3.3 mg/ml) NoIdarubicin (1 mg/ml) Ifosfamide (50 mg/ml) Irinotecan HCl (20 mg/ml) Leuprolide Acetate Salt (5 mg/ml) Mechlorethamine HCl (1 mg/ml) Melphalan (5 mg/ml) Methotrexate (25 mg/ml) Mitomycin C (0.5 mg/ml) Mitoxantrone (2 mg/ml) Oxaliplatin (5 mg/ml) Paclitaxel (6 mg/ml) Pemetrexed (25 mg/ml) Raltitrexed (0.5 mg/ml) Retrovir (10 mg/ml) Rituximab (10 mg/ml) Temsirolimus (25 mg/ml) Topotecan HCl (1 mg/ml) Triclosan (2 mg/ml) Trisenox (1 mg/ml) Vinblastine Sulfate (1 mg/ml) Vincristine (1 mg/ml) Vinorelbine (10 mg/ml) Zoledronic Acid (0.8 mg/ml) The following chemotherapy drugs and concentration showed breakthrough detected in less than 90 minutes: Carmustine (3.3 mg/ml) No

Page 16

StandardSubject DevicePredicate Device (K102032)Reference Device (K213929)Comparison
mg/ml) Cytovene (10 mg/ml) Retrovir (10 mg/ml) Triclosan (2 mg/ml) Zoledronic Acid (0.8 mg/ml)breakthrough up to 55.3 minutes. Thiotepa (10 mg/ml) No breakthrough up to 78.8 minutes. Warning‐ Not for use with Carmustine and ThioTEPAbreakthrough up to 55.3 minutes. Thiotepa (10 mg/ml) No breakthrough up to 78.8 minutes. Warning‐ Not for use with Carmustine and ThioTEPA
ASTM D6978‐05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy DrugsThe following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2 ml) Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 SolutionNot Previously TestedNo breakthrough was detected up to 240 minutes for Fentanyl Citrate Injection (100 mcg/2 ml) and Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 SolutionDifferent
ASTM D5151‐06 Standard Test Method for Detection of Holes in Medical GlovesTesting of the subject device shows it meets the 2.5% AQL requirement in the standards for leakage. The device meets the acceptance criteria of the standard.Testing of the predicate device shows it meets the 2.5% AQL requirement in the standards for leakage. The device meets the acceptance criteria of the standard.Testing of the reference device shows it meets the 2.5% AQL requirement in the standards for leakage. The device meets the acceptance criteria of the standard.Same
ASTM D6124‐06 Standard Test Method for Residual Powder on Medical GlovesResidual powder on the subject device is an average of 0.4 mg/glove within the powder‐free limit of < 2 mg maximum powder per glove andResidual powder on the predicate device is an average of 0.4 mg/glove within the powder‐free limit of < 2 mg maximum powder per glove and meets the acceptance criteriaResidual powder on the reference device is an average of 0.4 mg/glove within the powder‐free limit of < 2 mg maximum powder per glove and meets the acceptance criteriaSame

Page 17

StandardSubject DevicePredicate Device (K102032)Reference Device (K213929)Comparison
meets the acceptance criteria for powder‐free.for powder‐freefor powder‐free.
ASTM D6319‐10 Standard Specification for Nitrile Examination Gloves for Medical ApplicationsThe physical dimensions of the subject device are within the limits of the standard and the physical properties of the subject device met the requirements for tensile strength before and after aging. The subject device also met the requirement for elongation before and after aging.The physical dimensions of the predicate device are within the limits of the standard and the physical properties of the predicate device met the requirements for tensile strength before and after aging. The predicate device also met the requirement for elongation before and after aging.The physical dimensions of the reference device are within the limits of the standard and the physical properties of the predicate device met the requirements for tensile strength before and after aging. The predicate device also met the requirement for elongation before and after aging.Same

SUMMARY OF NON‐CLINICAL TESTING

Brief description of non‐clinical tests:TestStandardAcceptance CriteriaResults
DimensionsASTM D 6319Length ≥230 mmPalm Width Size Small: 70 ‐ 90 mmMed: 85–105 mmLarge: 100 ‐ 120 mmFinger thickness ≥0.05 mmPalm thickness ≥0.05 mmCuff thickness ≥0.05 mmMeets requirements

Page 18

Brief description of non‐clinical tests:TestStandardAcceptance CriteriaResults
Physical PropertiesASTM D 6319AQL 4.0Before AgingTensile Strength: ≥14 MPaUltimate elongation: ≥500%After AgingTensile Strength: ≥14 MPaUltimate elongation: ≥400%Meets requirements
Freedom from PinholesASTM D 6319ASTM D 5151AQL 2.5%No leakageMeets requirements
SterilityANSI/AAMI/ISO 1113710⁻⁶10⁻⁶
Powder FreeASTM D 6124ASTM D 6319≤ 2 mg / gloveMeets requirements
Test for irritationISO 10993, Part 23Grade 1Under the conditions of the study, the device is not an irritant.
Test for acute systemic toxicityISO 10993, Part 11No animals treated with test extracts exhibit greater reaction than control animals.Under the conditions of the study, no evidence of acute systemic toxicity.
Test for skin sensitizationISO 10993, Part 10Grade < 1.0Under the conditions of the study, the device is not a sensitizer.
Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy DrugsASTM D6978‐05No breakthrough for up to 240 minutes51 Drugs tested showed minimum breakthrough detection time up to 240 minutesCarmustine 3.3mg/ml minimum breakthrough detection time is 44.5 minutes

Page 19

Brief description of non‐clinical tests:TestStandardAcceptance CriteriaResults
Thiotepa 10mg/ml minimum breakthrough detection time is 99.1 minutes.

Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the subject devices (Halyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs or Sterile Singles) are as safe, as effective, and performs as well as the legally marketed devices cleared under K102032.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.