K Number
K243074

Validate with FDA (Live)

Date Cleared
2025-04-04

(189 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Spectrum Spine Lumbar Cage System devices with roughened surface textured features, are indicated for use in skeletally mature patients with lumbar degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 – S1 whose condition requires the use of interbody fusion, confirmed by imaging studies (radiographs, CT, MRI). DDD is defined as discogenic pain with degeneration of the disc. DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had at least six months of nonoperative treatment. The Spectrum Spine Lumbar Cage System is indicated to be used with supplemental fixation cleared by the FDA for use in the lumbar spine and autograft bone, allograft bone comprised of cancellous and/or corticocancellous bone, demineralized allograft with bone marrow aspirate, or a combination thereof.

Device Description

The Spectrum Spine Lumbar Cage System is composed of lumbar interbody fusion devices. The Spectrum Spine Lumbar Cage System is offered in multiple footprints with varying heights and lordotic angles designed to accommodate patient anatomy. Additionally, all interbody cages are offered with a smooth finish and a rough finish. The interbody devices also contain a large graft window through the body of the device to allow for placement of bone graft and facilitate fusion. All implant components are manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136. The Spectrum Spine Lumbar Cage System also includes instruments to facilitate implantation of the subject implant devices. Instruments are manufactured from medical grade stainless steel.

AI/ML Overview

The provided FDA 510(k) clearance letter for the Spectrum Spine Lumbar Cage System does not contain any information about a study proving the device meets acceptance criteria related to AI/ML device performance, human reader improvement with AI assistance, or the establishment of ground truth for such studies.

The document pertains to the clearance of an intervertebral body fusion device, which is a medical implant made of titanium alloy. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" described in the document relate solely to the mechanical and physical performance of the implant itself, not to the performance of an AI/ML algorithm.

Therefore, I cannot fulfill your request to describe acceptance criteria and a study proving device performance in the context of AI/ML, as the provided text does not contain this information.

Based on the provided text, here's what can be extracted regarding the "acceptance criteria" and "study" for the physical device:

1. Table of Acceptance Criteria and Reported Device Performance (Mechanical/Physical Device):

Acceptance Criterion (Test Type)Standard AppliedReported Device Performance Statement
Static Axial CompressionASTM F2077-22"results have shown them to be substantially equivalent to the predicate interbody devices."
Dynamic Axial CompressionASTM F2077-22"results have shown them to be substantially equivalent to the predicate interbody devices."
Static Compression ShearASTM F2077-22"results have shown them to be substantially equivalent to the predicate interbody devices."
Dynamic Compression ShearASTM F2077-22"results have shown them to be substantially equivalent to the predicate interbody devices."
SubsidenceASTM F2267-22"results have shown them to be substantially equivalent to the predicate interbody devices."
ExpulsionNot explicitly stated"results have shown them to be substantially equivalent to the predicate interbody devices."

Note: The specific numerical performance values are not provided in this regulatory summary; only the conclusion of "substantial equivalence" is stated.

2. Sample Size and Data Provenance (Mechanical/Physical Device):

  • Sample Size: Not specified in the document.
  • Data Provenance: The tests are described as "Mechanical testing... performed on the subject Spectrum Spine Lumbar Cage devices." This implies laboratory testing on physical samples, not human or patient data.

Regarding your other points (relevant to AI/ML devices), the document provides NO information:

  • Number of experts used to establish ground truth: Not applicable/not mentioned.
  • Adjudication method: Not applicable/not mentioned.
  • MRMC comparative effectiveness study: Not applicable/not mentioned.
  • Standalone (algorithm-only) performance: Not applicable/not mentioned.
  • Type of ground truth used: Not applicable/not mentioned (for AI/ML devices).
  • Sample size for training set: Not applicable/not mentioned.
  • How ground truth for training set was established: Not applicable/not mentioned.

In summary, the provided FDA clearance letter is for a physical medical device (spinal implant) and its mechanical properties, not an AI/ML diagnostic or therapeutic device. Therefore, the details requested for AI/ML device validation are not present in this document.

FDA 510(k) Clearance Letter - Spectrum Spine Lumbar Cage System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.03

April 4, 2025

Spectrum Spine, Inc.
℅ Ms. Christine Scifert
Partner
MRC Global, LLC
9085 East Mineral Cir., Suite 110
Centennial, Colorado 80112

Re: K243074
Trade/Device Name: Spectrum Spine Lumbar Cage System
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: MAX
Dated: September 27, 2024
Received: February 4, 2025

Dear Ms. Scifert:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K243074 - Christine Scifert Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243074 - Christine Scifert Page 3

Sincerely,

Katherine D. Kavlock -S

for
Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Food and Drug Administration
Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Submission Number (if known)
K243074

Device Name
Spectrum Spine Lumbar Cage System

Indications for Use (Describe)

The Spectrum Spine Lumbar Cage System devices with roughened surface textured features, are indicated for use in skeletally mature patients with lumbar degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion, confirmed by imaging studies (radiographs, CT, MRI). DDD is defined as discogenic pain with degeneration of the disc. DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had at least six months of nonoperative treatment. The Spectrum Spine Lumbar Cage System is indicated to be used with supplemental fixation cleared by the FDA for use in the lumbar spine and autograft bone, allograft bone comprised of cancellous and/or corticocancellous bone, demineralized allograft with bone marrow aspirate, or a combination thereof.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K243074 - 510(k) Summary

Spectrum Spine Lumbar Cage System

March 28, 2025

Company: Spectrum Spine, Inc.
20 Harmony Ct.
Jasper, GA 30143

Primary/Secondary Contact: Christine Scifert – Partner
MRC Global
9085 E. Mineral Cir., Suite 110
Centennial, CO 80112
Phone: (901) 831-8053
Email: christine.scifert@AskMRCGlobal.com

Jen McBride – Regulatory Consultant
(901) 481-5902
Email: jen.mcbride@askmcrglobal.com

Company Contact: Scott Baltsen
COO
Spectrum Spine, Inc.
Phone: (888) 377-7328
sbaltsen@spectrumspine.com

Trade Name: Spectrum Spine Lumbar Cage System
Common Name: Intervertebral Fusion Device With Bone Graft, Lumbar
Classification: Class II
Regulation: 21 CFR 888.3080 (Intervertebral Fusion Device)
Panel: Orthopedic
Product Code: MAX
Primary Predicate: Medtronic Sofamor Danek USA Inc. Endoskeleton TA Interbody System, Endoskeleton TAS & TAS Hyp Interbody System, Endoskeleton TL Interbody System, Endoskeleton TL Hyp. Interbody System, Endoskeleton TO Interbody System, Endoskeleton TT Interbody System – K211258

K243074
1 of 3

Page 6

Device Description:

The Spectrum Spine Lumbar Cage System is composed of lumbar interbody fusion devices. The Spectrum Spine Lumbar Cage System is offered in multiple footprints with varying heights and lordotic angles designed to accommodate patient anatomy. Additionally, all interbody cages are offered with a smooth finish and a rough finish. The interbody devices also contain a large graft window through the body of the device to allow for placement of bone graft and facilitate fusion. All implant components are manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136.

The Spectrum Spine Lumbar Cage System also includes instruments to facilitate implantation of the subject implant devices. Instruments are manufactured from medical grade stainless steel.

Indications for Use:

The Spectrum Spine Lumbar Cage System devices with roughened surface textured features, are indicated for use in skeletally mature patients with lumbar degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 – S1 whose condition requires the use of interbody fusion, confirmed by imaging studies (radiographs, CT, MRI). DDD is defined as discogenic pain with degeneration of the disc. DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had at least six months of nonoperative treatment. The Spectrum Spine Lumbar Cage System is indicated to be used with supplemental fixation cleared by the FDA for use in the lumbar spine and autograft bone, allograft bone comprised of cancellous and/or corticocancellous bone, demineralized allograft with bone marrow aspirate, or a combination thereof.

Substantial Equivalence:

The subject Spectrum Spine Lumbar Cage System is substantially equivalent to the following predicate devices:

Primary Predicate:
Medtronic Sofamor Danek USA Endoskeleton TA Interbody System, Endoskeleton TAS & TAS Hyp Interbody System, Endoskeleton TL Interbody System, Endoskeleton TL Hyp. Interbody System, Endoskeleton TO Interbody System, Endoskeleton TT Interbody System – K211258

Additional Predicates:

  • Eminent Spine 3D Lumbar Interbody Fusion Systems – K230219
  • Spectrum Cervical Cage System – K240838
  • Anjon Bremer Molded Halo Crown – K193256
  • GSMedical Co, Ltd, AnyPlus T-PLIF PEEK Cages – K131612
  • Spectrum Spine Expandable Titanium PLIF/TLIF System – K201024; K173518

There are insignificant differences between the subject Spectrum Spine Lumbar Cage System and the predicates. The Indications for Use and Materials for predicate devices are all inclusive of the subject device. Geometry and design of the subject device is similar to that of the predicates. Testing shows that

K243074
2 of 3

Page 7

the subject Spectrum Spine Lumbar Cage performs equivalent to or better than previously cleared devices. Thus, it can be concluded that the subject does not raise new questions about safety and effectiveness.

Performance Testing:

Mechanical testing, including static and dynamic axial compression and static and dynamic compression shear per ASTM F2077-22, as well as subsidence per ASTM F2267-22 and expulsion have been performed on the subject Spectrum Spine Lumbar Cage devices and the results have shown them to be substantially equivalent to the predicate interbody devices.

Conclusion:

The subject device is determined to be substantially equivalent to the predicate devices.

K243074
3 of 3

N/A