K Number
K131612

Validate with FDA (Live)

Date Cleared
2014-04-18

(319 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AnyPlus® PEEK Lumbar Fusion Cage device is intended for use with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The AnyPlus PEEK Lumbar Cage is to be combined with internal supplemental fixation cleared for use in the lumbar spine.

Device Description

The GS Medical AnyPlus PEEK Lumbar Cage device consists of implants available in various heights and lordotic configurations with an open architecture to accept packing of bone graft material. The implants are made of polyether-ether-ketone (PEEK OPTIMA LTI) (Manufacturer - INVIBIO) body (PEEK conforming to ASTM F2026) with the x-ray markers made of Tantulum (conforming to ASTM F560). This submission includes the addition of sizes to the model PLIF devices, a revision of the shapes of the ALIF, PLIF and TLIF model and the addition of the TPLIF model based on the PLIF design to accommodate surgeons using a Transforaminal Posterior Lateral Interbody Fusion (TPLIF) surgical approach.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (from ASTM F2077 and ASTM F2267)Reported Device Performance
Static Axial CompressionEquivalent to or higher than legally marketed predicate device
Dynamic Axial CompressionEquivalent to or higher than legally marketed predicate device
Static Compression-Shear TestingEquivalent to or higher than legally marketed predicate device
Dynamic Compression-Shear TestingEquivalent to or higher than legally marketed predicate device
Static Torsion TestingEquivalent to or higher than legally marketed predicate device
Dynamic Torsion TestingEquivalent to or higher than legally marketed predicate device
Expulsion TestingEquivalent to or higher than legally marketed predicate device
Static Subsidence Testing under Axial CompressionEquivalent to or higher than legally marketed predicate device

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: Not explicitly stated in the document. The testing was non-clinical (mechanical bench testing), so it would involve a certain number of devices tested for each parameter. The document indicates "All performance test results" suggests that a sufficient number was tested to make a comparative claim, but the exact count isn't provided.
  • Data Provenance: Non-clinical (bench testing) performed by the manufacturer, GS Medical Co., Ltd. The country of origin for the device manufacturing is Korea.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Number of Experts: Not applicable. This was non-clinical mechanical testing, not a study involving human experts establishing ground truth for diagnostic or clinical outcomes.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set:

  • Adjudication Method: Not applicable. As non-clinical mechanical testing, there is no "adjudication method" in the traditional sense. The performance was measured directly against established standards (ASTM F2077, ASTM F2267) and compared to a predicate device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • Was an MRMC study done? No. The document explicitly states: "No clinical testing was performed."

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Was a standalone study done? This is a medical device (implantable cage) and not an AI algorithm. Therefore, the concept of a "standalone algorithm" study does not apply. The performance evaluated was the standalone mechanical integrity of the device itself.

7. Type of Ground Truth Used:

  • Type of Ground Truth: The "ground truth" for this non-clinical testing was defined by the performance specifications outlined in the ASTM standards (ASTM F2077, ASTM F2267) and the measured performance of a legally marketed predicate device. The device was deemed "substantially equivalent" if its performance met or exceeded these established benchmarks.

8. Sample Size for the Training Set:

  • Training Set Sample Size: Not applicable. This refers to the training of an AI algorithm. This device is a physical medical implant, not a software algorithm, and therefore does not have a "training set."

9. How the Ground Truth for the Training Set Was Established:

  • How Ground Truth Was Established: Not applicable, as this is not an AI algorithm.

{0}------------------------------------------------

510(k) SUMMARY

Manufacturer:

GS Medical Co., Ltd. 90, Osongsaengmyeong 4-ro Osong-eup Cheongwon-gun Chungcheongbuk-do 363-951 Korea

Date:

April 18, 2014

Submitted by:

GS Medical Co., Ltd

Company Contact

J. W. Park +82 43 2377393 +82 43 2377403(FAX) jaewoo@gsmedi.cafe24.com

US Agent Information Official Correspondent Orgenix LLC Mr. Donald W. Guthner 111 Hill Road Douglassville, PA 19518 +1-646-460-2984 +1-484-363-5879 (FAX) dg@orgenix.com

Intervertebral Body Fusion Device

Classification Name:

Common/Usual Name:

Performance standards:

Proprietary Name:

AnyPlus® PEEK ALIF PEEK Lumbar Fusion Cage AnyPlus® PEEK PLIF PEEK Lumbar Fusion Cage AnyPlus® PEEK TLIF PEEK Lumbar Fusion Cage AnyPlus® PEEK TPLIF PEEK Lumbar Fusion Cage

Intervertebral Body Fusion Device, IBF Device

The GS Medical AnyPlus PEEK Lumbar Cage was non-clinically tested according to the ASTM 2077-03 and ASTM F2267-04 performance standards.

Classification no .:

21 CFR 888.3080 MAX - Intervertebral body fusion device Class II

Substantial Equivalence:

Substantial equivalence for the GS Medical AnyPlus PEEK Lumbar Cage is based on its similarities in indications for use, design features, operational principles and material composition when compared to the

APR 1 8 2014

{1}------------------------------------------------

predicate devices cleared under the following submissions:

  • K111354GS Medical AnyPlus® PEEK Lumbar Cage .
    Predicate Devices:

The subject device is substantially equivalent to similar previously cleared devices.

Device Description: The GS Medical AnyPlus PEEK Lumbar Cage device consists of implants available in various heights and lordotic configurations with an open architecture to accept packing of bone graft material. The implants are made of polyether-ether-ketone (PEEK OPTIMA LTI) (Manufacturer - INVIBIO) body (PEEK conforming to ASTM F2026) with the x-ray markers made of Tantulum (conforming to ASTM F560). This submission includes the addition of sizes to the model PLIF devices, a revision of the shapes of the ALIF, PLIF and TLIF model and the addition of the TPLIF model based on the PLIF design to accommodate surgeons using a Transforaminal Posterior Lateral Interbody Fusion (TPLIF) surgical approach.

The AnyPlus PEEK Lumbar Cage is indicated for use with autogenous Intended Use: bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The AnyPlus PEEK Lumbar Cage is to be combined with internal supplemental fixation cleared for use in the lumbar spine.

Summary of Technological The GS Medical AnyPlus PEEK Lumbar Cage devices are designed for Characteristics restoring the height of the intervertebral space after resection of the disc. The AnyPlus PEEK Lumbar Cage devices consist of implants available in various heights and lordotic configurations with an open architecture to accept packing of bone graft material. The implants are made of polyether-ether-ketone (PEEK) body with the x-ray markers made of The intended use, technological characteristics, mode of Tantulum action and materials of construction are the same as those of the referenced predicate devices

The GS Medical AnyPlus PEEK Lumbar Cage devices were tested Non-Clinical Testing according to the ASTM 2077, specifically, Static and Dynamic Axial Compression, Static and Dynamic Compression-Shear Testing, Static and Dynamic Torsion Testing Expulsion Testing and Static Subsidence testing under Axial Compression, per ASTM 2267.

All performance test results were equivalent to or higher than a legally marketed predicate device.

{2}------------------------------------------------

Clinical Testing

Conclusion

No clinical testing was performed.

The GS Medical AnyPlus® ALIF PEEK Cage, AnyPlus® PLIF PEEK Cage, AnyPlus® T-PLIF PEEK Cage and AnyPlus® TLIF PEEK Cage have the same intended use and similar indications, principles of operation, and technological characteristics as the GS Medical AnyPlus® ALIF PEEK Cage, AnyPlus® PLIF PEEK Cage and AnyPlus® TLIF PEEK Cage. The minor differences in the designs do not raise any new questions of safety or effectiveness. Performance data demonstrates that the AnyPlus® ALIF PEEK Cage, AnyPlus® PLIF PEEK Cage, AnyPlus® T-PLIF PEEK Cage and AnyPlus® TLIF PEEK Cage are as safe and effective as the AnyPlus® ALIF PEEK Cage, AnyPlus® PLIF PEEK Cage and AnyPlus® TLIF PEEK Cage].

Thus, the AnyPlus® ALIF PEEK Cage, AnyPlus® PLIF PEEK Cage, AnyPlus® T-PLIF PEEK Cage and AnyPlus® TLIF PEEK Cage are substantially equivalent to its predicate devices

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 18, 2014

GS Medical Company, Limited % Orgenix, LLC Mr. Donald W. Guthner Regulatory Consultant 111 Hill Road Douglassville, Pennsylvania 19518

Re: K131612

Trade/Device Name: AnyPlus® ALIF, PLIF, TLIF and TPLIF PEEK Lumbar Fusion Cages Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: March 19, 2014 Received: March 20, 2014

Dear Mr. Guthner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Donald W. Guthner

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Ronald P. Jean -S for

  • Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{5}------------------------------------------------

Indications for Use Statement

K131612 510(k) Number (if known):

Device Name: AnyPlus® ALIF PEEK Cage, AnyPlus® PLIF PEEK Cage, AnyPlus® T-PLIF PEEK Cage and AnyPlus® TLIF PEEK Cage

Indications for Use:

AnyPlus® PEEK Lumbar Fusion Cage device is intended for use with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The AnyPlus PEEK Lumbar Cage is to be combined with internal supplemental fixation cleared for use in the lumbar spine.

Prescription Use X (Per 21 C.F.R. 801.109) AND/OR

Over-The-Counter Use (Per 21 C.F.R. 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anton E. Dimitriev, PhD Division of Orthopedic Devices

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.