K Number
K243027

Validate with FDA (Live)

Date Cleared
2025-04-08

(193 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is intended for use by health care professionals for puncture, drainage and fluid collection of the anterior chamber of the eye. Its operation is manual. The Disposable Aqueous Humor Collector is single use only. The Disposable Aqueous Humor Collector is suitable for ophthalmic use.

Device Description

Disposable aqueous humor collector is composed of a liquid collecting needle sheath, a liquid collecting needle, a needle seat, a liquid collecting needle rubber sheath, a rubber plug, a liquid collecting tube, a cylinder body, a rubber ring, push rod and buckle structure.
Models: SnovoDAHC I-50, SnovoDAHC II-50, SnovoDAHC II-50-27, SnovoDAHC II-100-27
Volume: 50μL, 100μL
Needle Gauge: 25G, 27G

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the "Disposable Aqueous Humor Collector" does not contain acceptance criteria or a detailed study description of device performance, beyond general statements about compliance with ISO standards and a brief mention of an OUS study.

The document primarily focuses on demonstrating substantial equivalence to a predicate device (StaClear Syringe) by comparing specifications and stating that differences do not raise new safety or effectiveness concerns, supported by various non-clinical tests (biocompatibility, sterility, bench functional performance).

Therefore, I cannot fulfill the request to provide the acceptance criteria and the study that proves the device meets them based solely on the provided text. The document refers to testing being performed, but does not present the specific acceptance criteria or the detailed results of those tests.

However, I can extract the information that is present and highlight what is missing based on your request:


Missing Information: The provided document states that "required testing was conducted to validate the cumulative modifications made to the subject devices" and that "Substantial Equivalence is being supported with full performance testing," including bench functional performance tests. It also mentions "An outside of the United States (OUS) study was performed demonstrating the effectiveness of the subject device to collect 50 uL of liquid with the 50 uL model." However, it does not detail the specific acceptance criteria for these tests, nor does it provide a full study report with reported device performance against those criteria.

Here is what can be extracted and what is missing:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Stated or Implied)Reported Device Performance
Bench Functional Performance:
ISO 7886-1:2017: Sterile hypodermic syringes for single use. Part 1: Syringes for manual use"Complied"
ISO 9626:2016: Stainless steel needle tubing for the manufacture of medical devices — Requirements and test methods"Complied"
ISO 7864:2016: Sterile hypodermic needles for single use — Requirements and test methods"Complied"
Needle point penetration force: $\le$ 0.70N(Not explicitly reported, but implied to be met for compliance with ISO 9626)
Liquid production volume and precision: 50 µL ± 20%, 100 µL ± 20%"Effectiveness... to collect 50 uL of liquid with the 50 uL model" (from OUS study, specific performance values not given)
Resistance to corrosion of the liquid collecting needle: (Specific criteria not provided)(Not explicitly reported, but implied to be met for compliance with ISO 9626)
Total content of heavy metals: not exceed 5 µg/mL(Not explicitly reported)
Biocompatibility: Compliant with ISO 10993 series"Differences do not raise new or different questions of safety and effectiveness"
Irritation, Ocular: Not considered irritants to the ocular tissue"Not considered irritants to the ocular tissue"
Irritation; Intracameral Injection: Not considered inflammatory to intraocular tissues"Not considered inflammatory to intraocular tissues"
Sterility: SAL 10^-6"SAL: 10^-6" (achieved via Irradiation sterilization)
Shelf Life: 2 years"Confirms the 2 year shelf life"
USP <788> Particulate Matter in Injection: (Specific criteria not provided)(Not explicitly reported, but implied compliance)
USP <789> Particulate Matter in ophthalmic solutions: (Specific criteria not provided)(Not explicitly reported, but implied compliance)
Bacterial Endotoxins (ANSI AAMI ST72:2019, USP <85>): (Specific criteria not provided)(Not explicitly reported, but implied compliance)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified for any of the tests.
  • Data Provenance: The document mentions "An outside of the United States (OUS) study." This indicates a prospective clinical study conducted outside the US. No further details on the country of origin are provided. Bench and biocompatibility testing locations are not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Not Applicable. This device is a medical instrument (collector/syringe) for fluid collection, not an AI/diagnostic device that typically requires expert-established ground truth for its performance evaluation (e.g., image interpretation). Its performance is evaluated through physical and chemical property testing (e.g., volume accuracy, needle strength, sterility).

4. Adjudication Method for the Test Set

  • Not Applicable. As noted above, this type of device does not involve expert adjudication for ground truth establishment.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC study is typically for evaluating the performance of diagnostic devices, especially those involving human interpretation of data (e.g., radiology images) with and without AI assistance. This device is a manual fluid collector, so such a study is not relevant.

6. If a Standalone Performance Study Was Done

  • Yes, implicitly. The bench functional performance tests (ISO 7886-1, ISO 9626, ISO 7864), biocompatibility tests, sterility tests, and the OUS study (which demonstrated effectiveness to collect 50 µL) all represent standalone performance evaluations of the device itself, without human interpretation as a variable.

7. The Type of Ground Truth Used

  • Metrological and Biological Standards: The ground truth for this device's performance is established by reference to international standards (e.g., ISO for syringes and needles), specific quantitative measurements (e.g., needle force, volume, heavy metal limits, SAL for sterility), and established biocompatibility principles (ISO 10993 series). In the case of the OUS study, the "ground truth" for liquid collection would be the actual measured volume collected by the device.

8. The Sample Size for the Training Set

  • Not Applicable. This is a hardware medical device, not an AI/software device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As it's not an AI/software device, a training set and its associated ground truth are not relevant.

FDA 510(k) Clearance Letter - Disposable Aqueous Humor Collector

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 8, 2025

Hangzhou Sightnovo Medical Technology Co., Ltd.
Eric Shi, QA&QC
101M, Building 7, No. 22, Xinyan Road, Yuhang Economic and
Technological Development Zone, Yuhang District, Hangzhou
Hangzhou City, 311100, China

Re: K243027
Trade/Device Name: Disposable Aqueous Humor Collector (SnovoDAHC I-50, SnovoDAHC II-50, SnovoDAHC II-50-27, SnovoDAHC II-100-27)
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: QLY, FMI
Dated: September 23, 2024
Received: March 24, 2025

Dear Eric Shi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K243027 - Eric Shi Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243027 - Eric Shi Page 3

Sincerely,

Shruti N. Mistry -S

Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243027

Device Name: Disposable Aqueous Humor Collector (SnovoDAHC I-50, SnovoDAHC II-50, SnovoDAHC II-50-27, SnovoDAHC II-100-27)

Indications for Use (Describe):
It is intended for use by health care professionals for puncture, drainage and fluid collection of the anterior chamber of the eye.

Its operation is manual. The Disposable Aqueous Humor Collector is single use only.

The Disposable Aqueous Humor Collector is suitable for ophthalmic use.

Type of Use (Select one or both, as applicable):
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Sponsor: Hangzhou Sightnovo Medical Technology Co., Ltd.
Subject Device: Disposable Aqueous Humor Collector

K243027 510(k) Summary

1. Submitter Information

Company Name: Hangzhou Sightnovo Medical Technology Co., Ltd.

Address: 101M, Building 7, No. 22, Xinyan Road, Yuhang Economic and Technological Development Zone, Yuhang District, Hangzhou City, Zhejiang Province, China.

Phone: +86-571-88776936

Contact Person (including title): Eric Shi (QA&QC Manager)

E-mail: Eric-shi@isosh.com

Date Prepared: April 8, 2025

2. Subject Device Information

Type of 510(k) submission: Traditional

Common Name: Ophthalmic Syringe; Needle, Hypodermic;Single Lumen

Trade Name: Disposable Aqueous Humor Collector (SnovoDAHC I-50, SnovoDAHC II-50, SnovoDAHC II-50-27, SnovoDAHC II-100-27)

Regulation Name: Piston syringe; Hypodermic single lumen needle

Review Panel: General Hospital

Product Code: QLY;FMI

Regulation Number: 21 CFR 880.5860;21 CFR 880.5570

Regulation Class: 2

3. Predicate Device Information

K200242

Type of 510(k) submission: Traditional

Common Name: Ophthalmic Syringe; Syringe,Piston; Needle,Hypodermic,Single Lumen

Trade Name: StaClear Syringe

Regulation Name: Piston syringe; Hypodermic single lumen needle

Review Panel: General Hospital

Product Code: QLY; FMF; FMI,

Page 6

Sponsor: Hangzhou Sightnovo Medical Technology Co., Ltd.
Subject Device: Disposable Aqueous Humor Collector

Regulation Number: 21 CFR 880.5860; 21 CFR 880.5570

Regulation Class: 2

4. Device Description

Model: SnovoDAHC Ⅰ-50, SnovoDAHC Ⅱ-50, SnovoDAHC Ⅱ-50-27, SnovoDAHC Ⅱ-100-27

Volume: 50μL,100μL

Needle Gauge: 25G, 27G

The liquid collecting needle of this product is 25G,27G, the Needle point penetration force is ≤ 0.70N, and the liquid production volume and precision is 50 μL±20%, 100μL±20%, resistance to corrosion of the liquid collecting needle, total content of heavy metals not exceed 5μg/mL.

Disposable aqueous humor collector is composed of a liquid collecting needle sheath, a liquid collecting needle, a needle seat, a liquid collecting needle rubber sheath, a rubber plug, a liquid collecting tube, a cylinder body, a rubber ring, push rod and buckle structure.

5. Intended Use/Indications For Use

It is intended for use by health care professionals for puncture, drainage and fluid collection of the anterior chamber of the eye.

Its operation is manual. The Disposable Aqueous Humor Collector is single use only.

The Disposable Aqueous Humor Collector is suitable for ophthalmic use.

6. Summary of Comparison and Technological Characteristics

Table 1-General Comparison

Elements of ComparisonSubject DevicePredicate DeviceComment
Trade NameDisposable Aqueous Humor CollectorStaClear Syringe--
Common NameOphthalmic Syringe; Needle,Hypodermic; Single LumenOphthalmic Syringe; Syringe,Piston; Needle,Hypodermic,Single Lumen--

Page 7

Sponsor: Hangzhou Sightnovo Medical Technology Co., Ltd.
Subject Device: Disposable Aqueous Humor Collector

Elements of ComparisonSubject DevicePredicate DeviceComment
510(k) NumberK243027K200242N/A
Regulation NameOphthalmic Syringe;Needle, Hypodermic,Single LumenPiston syringe; Hypodermic single lumen needleSame
Regulation Number21 CFR 880.5860; 21 CFR 880.557021 CFR 880.5860; 21 CFR 880.5570Same
Product CodeQLY;FMIQLY;FMF;FMISame
Indications For UseIt is intended for use by health care professionals for puncture, drainage and fluid collection of the anterior chamber of the eye. Its operation is manual. The Disposable Aqueous Humor Collector is single use only. The Disposable Aqueous Humor Collector is suitable for ophthalmic use.The StaClear Syringe is intended to inject fluids into, or withdraw fluids from, the body. The StaClear Syringe is indicated for intravitreal use.Note 1
Mechanism of ActionManualManualSame
Volume50μL,100μL0.25 mL(250μL)Note 2
Single Use OnlyYESYESSame
Sterilization InformationProvided Sterile, single-use Sterilization Method: Irradiation sterilization SAL: 10-6Provided Sterile, single-use Sterilization Method: Ethylene Oxide SAL: 10-6Note 3
Shelf Life2 years1 yearNote 4
Device Materials1. liquid collecting needle sheath-MABS; 2. liquid collecting needle-ASTM 304 Stainless; 3.needle seat-Polypropylene; 4. liquid collecting needle rubber sheath-Natural rubber; 5.rubber plug-Butyl Rubber; 6. liquid collecting tube-polyethylene terephthalate; 7. cylinder body-High transparency MABS; 8.rubber ring-silica gel;Barrel-Polypropylene Plunger-Polyethylene Plunger Stopper-Polyisoprene (Nipol IR2200, Zeon Chemicals) Needle-ASTM 304 Stainless Steel Needle Shield-Polyethylene Plunger Cap-Polyethylene Barrel Lubricant-Silicone Oil, crosslinked with inert argon gas plasmaNote 5

Page 8

Sponsor: Hangzhou Sightnovo Medical Technology Co., Ltd.
Subject Device: Disposable Aqueous Humor Collector

Elements of ComparisonSubject DevicePredicate DeviceComment
push rod-Porcelain white ABS; buckle-ABS
Connector TypeAttached needleAttached needleSame
Needle Gauge25G,27G31GNote 2
Needle Length5(+0.5,-1mm)5/16 in (7.937mm)Note 2
Biocompatibility Endpoint For Ophthalmic UseIrritation, Ocular: Not considered irritants to the ocular tissue Irritation; Intracameral Injection: Not considered inflammatory to intraocular tissuesIrritation,Ocular: Not considered irritants to the ocular tissue Irritation Intravitreal Injection: Not considered inflammatory to intraocular tissuesNote 5
Performance testComplied with ISO 7886-1, ISO 9626, ISO 7864Complied with ISO 7886-1, ISO 9626, ISO 7864Same
LabelingMeet the requirements of 21 CFR Part 801Meet the requirements of 21 CFR Part 801Same

Note 1: The subject device, which is intended to withdraw fluids from the body, is appropriate for marketing under the regulations for piston syringe, 21 CFR 880.5860, and hypodermic single lumen needle, 21 CFR 880.5570.

However, after collecting aqueous humor from the eye, this device does not have the function of injection.

This difference in indications for use between the subject and predicate device does not raise new or different questions of safety and effectiveness.

Note 2: The volume of the subject device is smaller than that of the Predicate Device.

The parameters of the subject device's Needle Gauge and Needle Length are smaller than the Predicate Device. The technological characteristics of the subject device are substantially equivalent to the predicate device with differences in the device materials, syringe volume, needle gauge, and needle length. The differences in volume and needle gauge/length do not raise new or different questions of safety and effectiveness.

Note 3: The sterilization method of the subject device is different as the sterilization of the predicate device. However, they all have SAL of 10-6. The difference in sterilization method does not raise new or different questions of safety and effectiveness.

Note 4: Sterilization and shelf life validation for this product confirms the 2 year shelf life.

Page 9

Sponsor: Hangzhou Sightnovo Medical Technology Co., Ltd.
Subject Device: Disposable Aqueous Humor Collector

Note 5: Material differences are addressed through biocompatibility testing per the ISO 10993 series.

7. Non-Clinical Tests Performed On The Proposed Device

The company has performed non-clinical performance testing based on its risk assessment utilizing Failure Mode Effect Analysis (FMEA).

Following Quality System processes, required testing was conducted to validate the cumulative modifications made to the subject devices.

Substantial Equivalence is being supported with full performance testing.

7.1 Bench Functional Performance

The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device.

These tests include:

  • ISO 7886-1:2017, Sterile hypodermic syringes for single use. Part 1: Syringes for manual use
  • ISO 9626:2016, Stainless steel needle tubing for the manufacture of medical devices — Requirements and test methods
  • ISO 7864:2016, Sterile hypodermic needles for single use — Requirements and test methods

7.2 Clinical Testing

An outside of the United States (OUS) study was performed demonstrating the effectiveness of the subject device to collect 50 uL of liquid with the 50 uL model.

8. Biocompatibility Testing

The Disposable Aqueous Humor Collector is categorized as an externally communicating device with limited (≤24 hrs) tissue contact. Testing was conducted according to relevant sections of the ISO 10993 series.

Additionally,

  • USP <788> Particulate Matter in Injection.
  • USP <789> Particulate Matter in ophthalmic solutions

Page 10

Sponsor: Hangzhou Sightnovo Medical Technology Co., Ltd.
Subject Device: Disposable Aqueous Humor Collector

9. Reprocessing, Sterility and Shelf-Life

  • ISO 11137-2:2022 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose
  • ANSI AAMI ST72:2019 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing
  • USP <85> Bacterial Endotoxin Determination
  • Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile - Guidance for Industry and Food and Drug Administration Staff
  • Endotoxin Testing Recommendations for Single-Use Intraocular Ophthalmic Devices - Guidance for Industry and Food and Drug Administration Staff

10. Conclusion

The technological characteristics of the subject device are substantially equivalent to the predicate device with differences in the indications for use, device materials, syringe volume, needle gauge, needle length. Biocompatibility testing and performance testing demonstrate these differences do not raise new or different questions of safety and effectiveness when compared to the predicate device. Based on the information provided within this 510(k) submission, the proposed subject device is substantially equivalent to the predicate device and is as safe and as effective as the legally marketed predicate device.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).