K Number
K242812

Validate with FDA (Live)

Date Cleared
2024-10-12

(24 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a non-sterile disposable device intended for medical purposes that is worn on the exammer's hands or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes, and can be used with the device.

Cyclophosphamide (20.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fluorouracil (50.0 mg/ml) Paclitaxel (6.0 mg/ml) Cisplatin (1.0 mg/ml) Dacarbazine (10.0 mg/ml)

The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml): 15 minutes ThioTEPA (10.0 mg/ml): 97 minutes

Warning: Not for Use with: Carmustine, ThioTEPA

Device Description

NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a Class I, patient examination gloves bearing the product codes LZA, LZC, OPJ (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per D6978-05(2023). They are made from Nitrile (NBR). These gloves are blue in color and are powder free. The product is non-sterile, ambidextrous and single use only. NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) are offered in different sizes: S/M/L/XL.

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodPurposeAcceptance CriteriaReported Performance
Physical Dimensions
Length (S)Physical Dimensions Test$\ge$ 220 mm$\ge$ 220 mm
Length (M/L/XL)Physical Dimensions Test$\ge$ 230 mm$\ge$ 230 mm
Width (S)Physical Dimensions Test80 $\pm$ 10 mm84-86/Pass
Width (M)Physical Dimensions Test95 $\pm$ 10 mm96-102/Pass
Width (L)Physical Dimensions Test110 $\pm$ 10 mm104-107/Pass
Width (XL)Physical Dimensions Test120 $\pm$ 10 mm114-116/Pass
Thickness (Palm)Physical Dimensions Test$\ge$ 0.05 mm0.07-0.08/Pass
Thickness (Finger)Physical Dimensions Test$\ge$ 0.05 mm0.09-0.11/Pass
WatertightnessDetection of HolesAQL 2.50/125/Pass
Powder ContentPowder Content$\le$ 2.0 mg per glove0.16-0.19 mg/Pass
Physical Properties
Tensile Strength (before aging)Physical Properties$\ge$ 14 Mpa20-24 Mpa
Tensile Strength (after aging)Physical Properties$\ge$ 14 Mpa15-23 Mpa
Ultimate Elongation (before aging)Physical Properties$\ge$ 500 %501-598 %
Ultimate Elongation (after aging)Physical Properties$\ge$ 400 %471-561 %
Chemotherapy PermeationChemotherapy drug claimNo breakthrough up to 240 minutes for specified drugs>240 minutes for Cyclophosphamide, Doxorubicin HCl, Etoposide, Fluorouracil, Paclitaxel, Cisplatin, Dacarbazine
Low permeation times for specified drugsCarmustine: 15 minutes, ThioTEPA: 97 minutes
CytotoxicityCytotoxicityNon-cytotoxicDevice extract is cytotoxic (met acceptance criteria by identifying the result)
IrritationIrritationNon-irritatingNon-irritant / Pass
SensitizationSensitizationNon-sensitizingNon-sensitizer / Pass
Acute Systemic ToxicityAcute Systemic ToxicityNon-acute systemic toxicityDid not show acute systemic toxicity in vivo. / Pass

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for each individual test or the data provenance (e.g., country of origin, retrospective/prospective). However, for the watertightness test (ASTM D5151/ASTM D6319), it reports "0/125/Pass" which indicates a sample size of 125 gloves were tested for holes, and none failed.

The non-clinical tests were conducted to verify that the proposed device met all design specifications according to recognized standards. The identity of the testing laboratories or the origin of the samples is not specified in this document.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This document describes non-clinical performance testing of a physical device (gloves). Therefore, there is no "ground truth" derived from expert interpretation of medical images or data in the typical sense. The ground truth for these tests is established by the specified standards (ASTM and ISO standards) themselves, which outline quantitative metrics and methodologies for evaluation. No human experts are explicitly mentioned as establishing ground truth for these specific tests.

4. Adjudication Method for the Test Set

Not applicable. The tests performed are objective, quantitative measurements against established standards, not subjective assessments requiring adjudication or consensus among experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. An MRMC comparative effectiveness study is not applicable to the non-clinical performance testing of medical gloves. This type of study is typically used for evaluating the effectiveness of diagnostic or prognostic medical devices that rely on human interpretation, often with AI assistance, in comparison to human interpretation alone.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This document pertains to the physical performance of medical gloves, not a software algorithm that would operate in a standalone mode.

7. Type of Ground Truth Used

The ground truth for all tests is based on established testing standards and their quantitative acceptance criteria. For example:

  • Physical dimensions are compared against numerical ranges specified in ASTM D6319-19.
  • Watertightness is assessed by the absence of holes per ASTM D5151-19, with an acceptable quality level (AQL).
  • Powder content is measured against a maximum limit per ASTM D6124-06.
  • Tensile strength and ultimate elongation are measured against minimum requirements per ASTM D6319-19.
  • Chemotherapy drug permeation is determined by breakthrough detection times as per ASTM D6978-05(2023).
  • Biocompatibility tests (cytotoxicity, irritation, sensitization, acute systemic toxicity) are evaluated according to ISO 10993 series standards.

8. Sample Size for the Training Set

Not applicable. This document is about a physical medical device (gloves) and its non-clinical performance testing, not an AI/ML-based device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 12, 2024

Nb Medical Company Limited % Wendy Huang Technical Manager Wenzhou ThiWe Business Consulting Co., Ltd. Room 1203, Building C, Hua'ou Jiayuan, No.50 Tangjiaqiao South Road, Longwan District Wenzhou, Zhejiang 325000 China

Re: K242812

Trade/Device Name: NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ Dated: August 12, 2024 Received: September 18, 2024

Dear Wendy Huang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bifeng Qian -S

Bifeng Qian, M.D, Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242812

Device Name NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs)

Indications for Use (Describe)

NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a non-sterile disposable device intended for medical purposes that is worn on the exammer's hands or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes, and can be used with the device.

Cyclophosphamide (20.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fluorouracil (50.0 mg/ml) Paclitaxel (6.0 mg/ml) Cisplatin (1.0 mg/ml) Dacarbazine (10.0 mg/ml)

The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml): 15 minutes ThioTEPA (10.0 mg/ml): 97 minutes

Warning: Not for Use with: Carmustine, ThioTEPA

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

K242812

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter Information

· Company Name:NB MEDICAL COMPANY LIMITED
· Address:Tay An Industrial Cluster, Tien Hai Town, Tien Hai District,Thai Binh Province, Vietnam
· Phone Number:+84-934-388877
· Contact:VU THI VINH

2.0 Designated Submission Correspondent

· Company Name:Wenzhou ThiWe Business Consulting Co., Ltd.
· Address:Room 1203, Building C, Hua'ou Jiayuan, No.50 TangjiaqiaoSouth Road, Longwan District, Wenzhou, Zhejiang, 325000,CHINA
· Phone Number:86-577-88201260
· Contact:Ms. Wendy Huang
· Email:thiwe_reg@outlook.com

3.0 Date of Preparation: October 3, 2024

4.0 Device Information

・Trade/Device Name: NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs)

・Model(s): S, M, L, XL (Blue)

・ Common Name: Patient Examination Gloves (Tested For Use With Chemotherapy Drugs)

・510(k) Number: K242812

{5}------------------------------------------------

5.0 Classification

  • · Device: Polymer Patient Examination Glove & Medical Glove, Specialty & Medical Gloves With Chemotherapy Labeling Claims - Test For Use With Chemotherapy Drugs
  • · Regulation Description: Non-powdered patient examination glove.
  • · Review Panel: General Hospital
  • · Product Code: LZA, LZC, OPJ
  • · Submission Type: 510(k)
  • · Regulation Number: 21 CFR 880.6250
  • · Device Class: 1

6.0 Predicate Device Information

  • · Device Name: NITRILE EXAMINATION GLOVES
  • ・510(k) Number: K220343
  • · Manufacturer: NB Medical Co., LTD (i.e. NB MEDICAL COMPANY LIMITED)

7.0 Device Description

NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a Class I, patient examination gloves bearing the product codes LZA, LZC, OPJ (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per D6978-05(2023). They are made from Nitrile (NBR). These gloves are blue in color and are powder free. The product is non-sterile, ambidextrous and single use only. NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) are offered in different sizes: S/M/L/XL.

{6}------------------------------------------------

8.0 Indications for Use

NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes, and can be used with the device.

Cyclophosphamide (20.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fluorouracil (50.0 mg/ml) Paclitaxel (6.0 mg/ml) Cisplatin (1.0 mg/ml) Dacarbazine (10.0 mg/ml)

The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml): 15 minutes ThioTEPA (10.0 mg/ml): 97 minutes

Warning: Not for Use with: Carmustine, ThioTEPA

9.0 Comparison of Intended Use and Technological Characteristics with the Predicate Device

The table below compares the intended use and technological characteristics of the subject and predicate device.

Table 1: Comparison Table for Subject and Predicate Device

{7}------------------------------------------------

ItemSubject Device(K242812)Predicate Device(K220343)Comparison
General Comparison
Product CodeLZA, LZC, OPJLZADifferentAnalysis 1
Regulation No.21 CFR880.625021 CFR880.6250Same
Class11Same
IntendedUse/Indications forUseNITRILEEXAMINATIONGLOVES (Tested ForUse WithChemotherapy Drugs)is a non-steriledisposable deviceintended for medicalpurposes that is wornon the examiner'shands or finger toprevent contaminationbetween patient andexaminer. These gloveswere tested for use withChemotherapy Drugsas per ASTMD6978-05(2023)Standard Practice forAssessment of MedicalGloves to Permeationby ChemotherapyNITRILEEXAMINATIONGLOVES is anon-sterile disposabledevice intended formedical purposes that isworn on the examiner'shands or finger toprevent contaminationbetween patient andexaminer.DifferentAnalysis 1
Drugs.
MaterialNitrileNitrileSame
Powder FreeYesYesSame
Design FeatureAmbidextrousAmbidextrousSame
ColorBlueWhite/CobaltBlue/Black/BlueDifferentAnalysis 2
SizeS/M/L/XLS/M/L/XLSame
SterilityNon-SterileNon-SterileSame
UseSingle UseSingle UseSame
Rx Only or OTCOTCOTCSame
ChemotherapyDrugs ClaimThe followingchemotherapy drugsand concentration hadNO breakthroughdetected up to 240minutes, and can beused with the device.Cyclophosphamide(20.0 mg/ml)Doxorubicin HCl (2.0mg/ml)Etoposide (20.0 mg/ml)Fluorouracil (50.0mg/ml)Paclitaxel (6.0 mg/ml)Cisplatin (1.0 mg/ml)Dacarbazine (10.0mg/ml)No chemotherapy drugsclaimDifferentAnalysis 1
The followingchemotherapy drugshave low permeationtimes:Carmustine (3.3 mg/ml): 15 minutesThioTEPA (10.0 mg/ml): 97 minutesWarning: Not for Use with: Carmustine, ThioTEPA
LabelingInformationSingle-use indication,powder free, device color, device name,glove size and quantity,Non-Sterile, applicablechemotherapy drugsand the minimumbreakthrough timeSingle-use indication,powder free, devicecolor, device name,glove size and quantity,Non-SterileDifferentAnalysis 1
Technological Characteristic and Biocompatibility Comparison
Dimensions -LengthS: ≥220mmM/L/XL: ≥230mmMeet the requirementof ASTM D6319-19S: ≥220mmM/L/XL: ≥230mmMeet the requirementof ASTM D6319-19Same
Dimensions -WidthS: 80±10 mmM: 95±10 mmL: 110±10 mmS: 80±10 mmM: 95±10 mmL: 110±10 mmSame
Meet the requirementof ASTM D6319-19Meet the requirementof ASTM D6319-19
PhysicalProperties -Tensile Strengthbefore aging ≥14 Mpaafter aging ≥14 Mpabefore aging ≥14 Mpaafter aging ≥14 MpaSame
Meet the requirementof ASTM D6319-19Meet the requirementof ASTM D6319-19
PhysicalProperties -UltimateElongationbefore aging ≥500 %after aging ≥400 %before aging ≥500 %after aging ≥400 %Same
Meet the requirementof ASTM D6319-19Meet the requirementof ASTM D6319-19
ThicknessPalm: ≥0.05 mmFinger: ≥0.05 mmPalm: ≥0.05 mmFinger: ≥0.05 mmSame
Meet the requirementof ASTM D6319-19Meet the requirementof ASTM D6319-19
Powder FreeResidue≤2.0mg per glove≤2.0mg per gloveSame
Meet the requirementof ASTM D 6124following ASTMD6319-19Meet the requirementof ASTM D 6124following ASTMD6319-19
Freedom fromholesBe free from holeswhen tested inaccordance with ASTMD5151 followingASTM D6319-19AQL=2.5Be free from holeswhen tested inaccordance with ASTMD5151 followingASTM D6319-19AQL=2.5Same
BiocompatibilityISO 10993-5Under conditions of thestudy, device extract iscytotoxicISO 10993-5Under conditions of thestudy, device extract iscytotoxicSame
ISO 10993-10ISO 10993-10Same
Under the conditions ofthe study, not an irritantor a sensitizerUnder the conditions ofthe study, not an irritantor a sensitizer
ISO 10993-11ISO 10993-11Same
Under the condition ofacute systemic toxicitytest, the test article didnot show acutesystemic toxicity invivo.Under the condition ofacute systemic toxicitytest, the test article didnot show acutesystemic toxicity invivo.

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

Analysis 1: The subject device has different indications for use to the predicate device, which results in different product codes, that is, the subject device has extra applicable product codes compared with the predicate device. The major difference between the two devices is that the subject device is claimed for use with certain chemotherapy drugs, while the predicate device did not. However, such difference has been tested according to FDA recognized standard - ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs, which ensures that the subject device will be as safe as for use as the predicate device. Meanwhile, chemotherapy drugs and the minimum breakthrough time of subject device will be listed on labeling, so this different will not bring new concerns regarding safety and effectiveness.

Analysis 2: The subject device is offered only in a single color (Blue) whereas the predicate is offered in different colors (White/Cobalt Blue/Black/Blue), but the color of the subject device is included by those of the predicate device, so such difference will not bring questions regarding safety and effectiveness.

{12}------------------------------------------------

10.0 Summary of Non-Clinical Performance Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to

Permeation by Chemotherapy Drugs.

ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.

Test MethodPurposeAcceptance CriteriaResults
ASTM D6319PhysicalDimensions TestLength:S: $\geq$ 220mmLength:S: $\geq$ 220mm
M/L/XL: $\geq$ 230mmM/L/XL: $\geq$ 230mm
Width:S: 80 $\pm$ 10 mmM: 95 $\pm$ 10 mmL: 110 $\pm$ 10 mmXL: 120 $\pm$ 10 mmWidth:S: 84-86/PassM: 96-102/PassL: 104-107/PassXL: 114-116/Pass
Thickness:Thickness:

Table 2: Summary of non-clinical performance testing

{13}------------------------------------------------

Palm: ≥0.05 mmFinger: ≥0.05 mmFinger:0.09-0.11/PassPalm:0.07-0.08/Pass
ASTM D5151ASTM D6319WatertightnessTest forDetection ofHolesMeet the requirementsof ASTM D5151 AQL2.50/125/Pass
ASTM D6124ASTM D6319Powder ContentMeet the requirementof ASTM D 6124≤2.0mg0.16-0.19mg/Pass
ASTM D412ASTM D6319PhysicalPropertiesTensile Strength:before aging ≥14 Mpaafter aging ≥14 MpaUltimate Elongation:before aging ≥500 %after aging ≥400 %Tensile Strength:before aging:20-24Mpaafter aging:15-23MpaUltimateElongation:before aging:501-598%after aging:471-561%
ASTM D6978Chemotherapydrug claimThe chemotherapydrugs andconcentration whichhave NO breakthroughdetected up to 240The followingchemotherapydrugs andconcentration hadNO breakthrough
minutes can besuggested for use withthe subject device.detected up to 240minutes, and can beused with thedevice.Cyclophosphamide(20.0 mg/ml)Doxorubicin HCl(2.0 mg/ml)Etoposide (20.0mg/ml)Fluorouracil (50.0mg/ml)Paclitaxel (6.0mg/ml)Cisplatin (1.0mg/ml)Dacarbazine (10.0mg/ml)The followingchemotherapydrugs have lowpermeation times:Carmustine (3.3mg/ml): 15 minutesThioTEPA (10.0mg/ml): 97 minutesWarning: Not for
Use with:
Carmustine,
ThioTEPA
ISO 10993-5CytotoxicitytoxicityUnder conditionsof the study, deviceextract is cytotoxic.
ISO 10993-10IrritationNon-irritatingUnder theconditions of thestudy, not anirritant. / Pass
ISO 10993-10SensitizationNon-sensitizingUnder theconditions of thestudy, not ansensitizer. / Pass
ISO 10993-11Acute SystemicToxicityNon-acute systemicUnder theconditions of thestudy, did not showacute systemictoxicity in vivo. /Pass

{14}------------------------------------------------

{15}------------------------------------------------

Table 3 Chemotherapy Permeation Testing Claim

Chemotherapy DrugMinimum Breakthrough Detection Time(Minutes)
Cyclophosphamide (20.0 mg/ml)>240
Doxorubicin HCl (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Paclitaxel (6.0 mg/ml)>240
Cisplatin (1.0 mg/ml)>240

{16}------------------------------------------------

Dacarbazine (10.0 mg/ml)>240
Carmustine (3.3 mg/ml)15
ThioTEPA (10.0 mg/ml)97

11.0 Summary of Clinical Testing

Clinical testing is not needed for this device.

12.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K220343.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.