K Number
K242770

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-03-20

(188 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EXPD 114, EXPD 114P, EXPD 114G, EXPD 114PG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications. It is intended for both adult and pediatric populations.

Device Description

EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG are flat-panel type digital X-ray detector that captures projection radiographic images in digital format within seconds, eliminating the need for an entire x-ray film or an image plate as an image capture medium. EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG differs from traditional X-ray systems in that, instead of exposing a film and chemically processing it to create a hard copy image, a device called a Detector is used to capture the image in electronic form.

EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG are indirect conversion devices in the form of a square plate in which converts the incoming X-rays into visible light. This visible light is then collected by an optical sensor, which generates an electric charges representation of the spatial distribution of the incoming X-ray quanta.

The charges are converted to a modulated electrical signal thin film transistors. The amplified signal is converted to a voltage signal and is then converted from an analog to digital signal which can be transmitted to a viewed image print out, transmitted to remote viewing or stored as an electronic data file for later viewing.

AI/ML Overview

The DRTECH Corporation's EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG Digital X-ray detectors were assessed for substantial equivalence to a predicate device (K223124). The company conducted non-clinical performance testing (bench tests) and a "Concurrence Study" for image quality to demonstrate this.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a numerical or pass/fail format for the Concurrence Study beyond broad equivalence. Instead, it compares the performance of the subject device to the predicate device. The performance data mentioned primarily relates to technical specifications and general image quality assessment.

ParameterAcceptance Criteria (Implied: Equivalent to or comparable to predicate)Reported Device Performance (Subject Device)Reported Predicate Device Performance (K223124)
DQEEquivalent to or comparable to predicateEXPD 114: 45% @0.5lp/mm EXPD 114G: 25% @0.5lp/mm EXPD 114P: 45% @0.5lp/mm EXPD 114PG: 25% @0.5lp/mmEXPD 129P, EXPD 86P: 50.0 % at 0.5 lp/mm EXPD 129PG, EXPD 86PG: 25.0 % at 0.5 lp/mm
MTFEquivalent to or comparable to predicateEXPD 114: 40% @2.0lp/mm EXPD 114G: 40% @2.0lp/mm EXPD 114P: 40% @2.0lp/mm EXPD 114PG: 40% @2.0lp/mmEXPD 129P, EXPD 86P: 45.0 % at 2.0 lp/mm EXPD 129PG, EXPD 86PG: 45.0 % at 2.0 lp/mm
ResolutionEquivalent to or comparable to predicate3.5 lp/mm3.5 lp/mm
Image Quality (Clinical Assessment)Equivalent to predicate device."the image quality of the subject device is equivalent to that of the predicate device"Standard established by predicate device.

Note on DQE and MTF: The subject device's DQE for EXPD 114/P is slightly lower than the predicate EXPD 129P/86P (45% vs 50%). Similarly, the MTF for all subject devices is lower than the predicate (40% vs 45%). Despite these numerical differences, the overall conclusion states "basically equal or worth the predicate device" and that the device meets acceptance criteria. This suggests that the measured differences were considered clinically acceptable within the context of substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

The document states: "Our Concurrence Study for Image Quality was based on body parts (Chest, C-spine AP, L-spine AP, Shoulder AP, Pelvis AP, Extremity) to compare subject device and predicate device(K223124)."

  • Sample Size: The exact number of images or cases analyzed in the Concurrence Study is not specified in the provided text. It only lists the anatomical sites included.
  • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It is a "Concurrence Study" which implies a direct comparison, likely of newly acquired images, but this is not explicitly stated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: "a qualified clinical expert" (singular) is mentioned.
  • Qualifications of Experts: The expert is described as "qualified clinical expert." No further details on their specific qualifications (e.g., radiologist, years of experience, board certification) are provided.

4. Adjudication Method for the Test Set

  • Adjudication Method: The document only mentions "a qualified clinical expert confirmed" the image quality. This strongly suggests a single-reader assessment without any explicit adjudication method (e.g., 2+1, 3+1 consensus).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: Based on the description of "a qualified clinical expert confirmed," it appears that a formal MRMC comparative effectiveness study was not conducted. The assessment seems to be a qualitative comparison of image quality by a single expert.
  • Effect Size of Human Reader Improvement: As an MRMC study was not indicated, there is no information on the effect size of how much human readers improve with AI vs. without AI assistance. The device is a digital X-ray detector, not an AI software intended for interpretation assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • The device is a digital X-ray detector, not an AI algorithm. Therefore, the concept of "standalone performance" of an AI algorithm is not applicable in this context. The "performance" described relates to the physical characteristics of the detector (DQE, MTF, Resolution) and its ability to produce images of diagnostic quality.

7. Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for the Concurrence Study was based on the expert's subjective assessment of image quality compared to the predicate device, stating that "the image quality of the subject device is equivalent to that of the predicate device." This is a form of expert consensus, albeit with only one expert explicitly mentioned. It's not based on pathology or outcomes data.

8. Sample Size for the Training Set

  • The document describes a device (digital X-ray detector), not a machine learning model. Therefore, the concept of a "training set" in the context of an AI/ML algorithm is not applicable.

9. How the Ground Truth for the Training Set Was Established

  • As the device is not an AI/ML algorithm, the concept of "training set ground truth" is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 20, 2025

DRTECH Corporation % Yeongsuk An Assistant Manager Suite No. 1, 2 Floor/Suite No. 2, 3 Floor 29, Dunchon-daero 541beon-gil Jungwon-gu, Seongnam-si, Gyeonggi-do, 13216 SOUTH KOREA

Re: K242770

Trade/Device Name: EXPD 114; EXPD 114G; EXPD 114P; EXPD 114PG Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MQB Dated: September 13, 2024 Received: February 18, 2025

Dear Yeongsuk An:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242770

Device Name

EXPD 114: EXPD 114G;

EXPD 114P; EXPD 114PG

Indications for Use (Describe)

EXPD 114, EXPD 114P, EXPD 114G, EXPD 114PG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications. It is intended for both adult and pediatric populations.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K242770

510(k) Summary

[As required by 21 CFR 807.92]

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92

1. Date Prepared [21 CFR 807.92(a) (1)]

February/18/2025

2. Submitter's Information [21 CFR 807.92(a) (1)]

  • Name of Sponsor: ● DRTECH Corporation ● Address: Suite No.1, 2 Floor / Suite No. 2, 3 Floor, 29, Dunchon-daero541 beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, 13216, Republic of Korea ● Contact Name: Yeongsuk, An
    • 82-31-779-7787 Telephone No.: ●
  • Fax No .: + 82-31-779-7790
  • Email Address : drtechra@drtech.com
  • Registration Number: 3005172103 ●
  • Name of Manufacturer: Same as Sponsor

3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]

  • Trade Name: ● EXPD 114; EXPD 114G; EXPD 114P; EXPD 114PG
  • Common Name: Digital Flat Panel X-ray Detector
  • Classification Name: Stationary X-ray System
  • Classification Panel: Radiology ●
  • Classification Regulation: 21 CFR 892.1680
  • Product Code: MOB
  • . Device Class: II

4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

510(k) Number:K223124
Applicant:DRTECH Corporation
Trade Name:EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PC
Classification Name:Radiology
  • Classification Panel: ● Stationary X-ray System
  • Classification Regulation: 21 CFR 892.1680
  • Product Code: MQB
  • Device Class: II ●

{5}------------------------------------------------

5. Description of the Modified Device [21 CFR 807.92(a) (4)]

Addition of EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG : The differences between the subject devices and the predicate devices are Dimension , TFT panel material, and Components(optional). The Dimension of Subject devices is shorter than predicate devices.

6. Indication for Use [21 CFR 807.92(a)(5)]

EXPD 114, EXPD 114P, EXPD 114G, EXPD 114PG is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications. It is intended for both adult and pediatric populations.

7. Technological Characteristics [21 CFR 807.92(a)(6)]

EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG are flat-panel type digital X-ray detector that captures projection radiographic images in digital format within seconds, eliminating the need for an entire x-ray film or an image plate as an image capture medium. EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG differs from traditional X-ray systems in that, instead of exposing a film and chemically processing it to create a hard copy image, a device called a Detector is used to capture the image in electronic form.

EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG are indirect conversion devices in the form of a square plate in which converts the incoming X-rays into visible light. This visible light is then collected by an optical sensor, which generates an electric charges representation of the spatial distribution of the incoming X-ray quanta.

The charges are converted to a modulated electrical signal thin film transistors. The amplified signal is converted to a voltage signal and is then converted from an analog to digital signal which can be transmitted to a viewed image print out, transmitted to remote viewing or stored as an electronic data file for later viewing.

{6}------------------------------------------------

DRTECH

8. Integrated Configuration for X-Ray system

EXPD 114, EXPD 114G, EXPD 114PG are connected with General diagnostic X-ray system.

Image /page/6/Figure/4 description: This image shows a diagram of an X-ray system. The diagram includes a detector, an X-ray system, a console, and a hospital network. The detector is connected to the power source via a POE cable, and it is also connected to the X-ray system via a SYNC cable. The console is connected to the detector via a LAN cable, and it is connected to the hospital network via an Ethernet cable.

No.ComponentDescription
POE Cable1) DC 24V voltage2) It is supplied from Adaptor to Detector.
LAN Cable1) Gigabit Ethernet Communications2) It is transmitted from the detector to the CONSOLE PC.
SYNC Cable1) two-way TTL communication2) It is communicated between the detector and the X-ray System Generator.
Ethernet Cable1) DICOM communication2) It is transmitted from the CONSOLE PC to the hospital network.

{7}------------------------------------------------

For successful integration of detector with X-ray system, we designed the hardware of EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG as below function. The Energy of generator is controlled by Radiography imaging software(Econsole2, K240343), not firmware of Detector.

ModeDescription
AED/AWC ModeThe detector detects actual amount of X-rays without any connection to the X-raygenerator.No signal used (No need of connector interface cable)
Sync. Trigger ModeThe detector receives EXP_REQ signal that X-ray generator is prepared togenerate X-rays.The detector prepares image acquiring and then responds EXP_OK signal to theX-ray generator.The X-ray generator confirms EXP_OK signal and generates X-ray, then thedetector performs image acquiring, according to image acquisition time andtransmits the image data.EXP_REQ (Generator→ Detector), EXP_OK (Detector →Generator)

{8}------------------------------------------------

9. Substantial Equivalence [21 CFR 807.92(b)]

ParameterSubject DevicePredicate Device
510(K) NumberUnknownK223124
ManufacturerDRTECH CorporationDRTECH Corporation
Model NameEXPD 114, EXPD 114G, EXPD 114P,EXPD 114PGEXPD 86P, EXPD 86PG,EXPD 129P, EXPD 129PG
Classification NameStationary X-ray SystemStationary X-ray System
Classification PanelRadiologyRadiology
ClassificationRegulation21 CFR 892.168021 CFR 892.1680
Product CodeMQBMQB
Device ClassClass IIClass II
Intended UseEXPD 114, EXPD 114P, EXPD 114G,EXPD 114PG is indicated for digitalimaging solution designed for providinggeneral radiographic diagnosis of humananatomy. This device is intended toreplace film or screen based radiographicsystems in all general purpose diagnosticprocedures. This device is not intendedfor mammography applications.The EXPD 86P/ EXPD 86PG/ EXPD129P/ EXPD 129PG Digital X-raydetector is indicated for digital imagingsolution designed for providing generalradiographic diagnosis of humananatomy. This device is intended toreplace film or screen based radiographicsystems in all general purpose diagnosticprocedures. This device is not intendedfor mammography applications.
PanelShapeRectangular PanelEXPD 129P, EXPD 129PG, EXPD 86P,EXPD 86PG : Rectangular Panel
DetectorSize17" × 46"EXPD 129P, EXPD 129PG: 17" × 51"EXPD 86P, EXPD 86PG: 17" × 34"
DesignDimensions460mm (W) × 1,178mm (L) × 20mm (H)[±0.5mm]EXPD 129P, EXPD 129PG: 460mm (W) × 1324mm (L) × 20mm(H) [±0.5mm]EXPD 86P, EXPD 86PG: 460mm (W) × 894.14mm (L) × 15.5mm(H) [±0.5mm]
Pixel Pitch140μm140μm
Image Size3,072 × 8,192EXPD 129P, EXPD 129PG: 3,072 × 9,216EXPD 86P, EXPD 86PG: 3,072 × 6,144
ParameterSubject DevicePredicate Device
Panel & ScintillatorMaterialsPanel Material- EXPD 114, EXPD 114G: TFT -amorphous Silicon- EXPD 114P, EXPD 114PG : TFT -IGZOTFT - amorphous Silicon
Scintillator MaterialEXPD 114, EXPD 114P: CsIEXPD 114G, EXPD 114PG: Gadox,Scintillator Material- EXPD 129P, EXPD 86P: CsI- EXPD 129PG, EXPD 86PG: Gadox
Perfor-manceDQEEXPD 114: 45% @0.5lp/mmEXPD 114G: 25% @0.5lp/mmEXPD 114P : 45% @0.5lp/mmEXPD 114PG: 25% @0.5lp/mmEXPD 129P, EXPD 86P: 50.0 % at 0.5 lp/mmEXPD 129PG, EXPD 86PG: 25.0 % at 0.5 lp/mm
MTFEXPD 114: 40% @2.0lp/mmEXPD 114G: 40% @2.0lp/mmEXPD 114P : 40% @2.0lp/mmEXPD 114PG: 40% @2.0lp/mmEXPD 129P, EXPD 86P: 45.0 % at 2.0 lp/mmEXPD 129PG, EXPD 86PG: 45.0 % at 2.0 lp/mm
Resolution3.5 lp/mm3.5 lp/mm
Anatomical SitesGeneral RadiographyGeneral Radiography
Power Supply100-240V~, 50-60 Hz100240V, 50/60 Hz
ComponentsAdaptor & AC Power CordPOE CableSYNC CableLAN CableAdaptor & AC Power CordLLD POE CableLLD Sync Cable
CommunicationInterfaceWired: EthernetWired: Ethernet
DiagramImage: Diagram of the subject deviceImage: Diagram of the predicate device

{9}------------------------------------------------

DRTECH


{10}------------------------------------------------

DRTECH

When compared to the predicate devices (K223124), the EXPD 114G, EXPD 114G, EXPD 114PG presented in this submission have similar:

  • Intended Use •
  • Technological characteristics
  • . Operating principle
  • Performance (Resolution)
  • . Communication Method

A few differences are as follows

  • Image Size •
  • . Active area
  • . TFT panel
  • MTF and DQE
  • Components (optional)

{11}------------------------------------------------

10. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]

For Performance

The non-clinical performance testing(Bench Test) and Concurrence Study constrains the main physical values for comparison of X-ray devices like DQE and MTF are basically equal or worth the predicate device as the following table:

ParameterSubject DevicePredicate Device
DQEEXPD 114: 45% @0.5lp/mmEXPD 114G: 25% @0.5lp/mmEXPD 114P: 45% @0.5lp/mmEXPD 114PG: 25% @0.5lp/mmEXPD 129P, EXPD 86P: 50.0 % at 0.5 lp/mmEXPD 129PG, EXPD 86PG: 25.0 % at 0.5 lp/mm
MTFEXPD 114: 40% @2.0lp/mmEXPD 114G: 40% @2.0lp/mmEXPD 114P: 40% @2.0lp/mmEXPD 114PG: 40% @2.0lp/mmEXPD 129P, EXPD 86P: 45.0 % at 2.0 lp/mmEXPD 129PG, EXPD 86PG: 45.0 % at 2.0 lp/mm

For Firmware

Firmware Validation and Verification has been conducted according to the FDA guidance document "Content of Premarket Submissions for Device Software Functions".

For Cybersecurity

Cybersecurity Testing was performed in accordance with the guidance of "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions ".

RecognitionNo.Standard No.Title of StandardDate ofRecognitionRemark
5-125ISO 14971 Third Edition2019-12Medical devices - Application of riskmanagement to medical devices12/23/2019
19-46ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012,C1:2009/(R)2012 &A2:2010/(R)2012 (Cons.Text) [Incl. AMD2:2021]Medical electrical equipment - Part 1:General requirements for basic safety andessential performance (IEC 60601-1:2005, MOD) [Including Amendment 2(2021)]05/30/2022
RecognitionNo.Standard No.Title of StandardDate ofRecognitionRemark
19-36IEC 60601-1-2 Edition 4.12020-09 CONSOLIDATEDVERSIONMedical electrical equipment - Part 1-2:General requirements for basic safety andessential performance - CollateralStandard: Electromagnetic disturbances -Requirements and tests12/21/2020
5-132IEC 60601-1-6 Edition 3.22020-07 CONSOLIDATEDVERSIONMedical electrical equipment - Part 1-6:General requirements for basic safety andessential performance - Collateralstandard: Usability12/21/2020
5-129IEC 62366-1 Edition 1.12020-06 CONSOLIDATEDVERSIONMedical devices - Part 1: Application ofusability engineering to medical devices07/06/2020
13-79IEC 62304 Edition 1.1 2015-06 CONSOLIDATEDVERSIONMedical device software - Software lifecycle processes01/14/2019
12-349NEMA PS 3.1 - 3.20 2022dDigital Imaging and Communications inMedicine (DICOM) Set12/19/2022
12-289IEC 62220-1-1 Edition 1.02015-03Medical electrical equipment-Characteristics of digital X-ray imagingdevices Part 1-1: Determination of thedetective quantum efficiency Detectorsused in radiographic imaging08/14/2015

And The EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG comply with the following international and FDA-recognized consensus standards:

{12}------------------------------------------------

DRTECH

11. Summary of Clinical Data[21 CFR 807.92(b)(2)]

Our Concurrence Study for Image Quality was based on body parts (Chest, C-spine AP, L-spine AP, Shoulder AP, Pelvis AP, Extremity) to compare subject device and predicate device(K223124). In this Concurrence Study, a qualified clinical expert confirmed that the image quality of the subject device is equivalent to that of the predicate device, demonstrating that the subject device meets the diagnostic quality of the predicate device.

9/10 510(k) Summary

{13}------------------------------------------------

12. Related FDA Guidance documents

We reviewed the below FDA guidance documents and it was reflected in the development of the modified EXPD 114, EXPD 114G, EXPD 114P, EXPD 114PG.

Title of GuidanceRemark
Guidance for the Submission of 510(k)s forSolid State X-ray Imaging Devices
Cybersecurity in Medical Devices: Quality System Considerations and Content ofPremarket Submissions
Content of Premarket Submissions for Device Software Functions
Pediatric Information for X-ray Imaging Device Premarket Notifications

{14}------------------------------------------------

13. Conclusion [21 CFR 807.92(b)(3)]

The modified EXPD 114, EXPD 114G, EXPD 114PG are substantially equivalent to the currently marketed and predicate device (K223124) in terms of fundamental scientific technology, indications for use, and safety and effectiveness.

Additionally, Substantial equivalence was demonstrated through the non-clinical performance such as Bench Test, Software Validation and Verification, Cybersecurity Testing, which complied with the requirements specified in the international and FDA-recognized consensus standards, ANSI AAMI ES60601-1:2005/AMD2:2021, IEC 60601-1-2:2020, IEC 60601-1-6 Edition 3.2, IEC 62304:2015, IEC 6220-1-1, IEC 81001-5-1 and ANSI UL 2900-1. which complied with the requirements specified in the CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices.

Plus, A Concurrence Study was conducted to confirm the clinical equivalence of the subject device with the predicate device. Through the clinical study, it was confirmed that the image quality of the subject device is equivalent to that of the predicate device.

The results of these data demonstrate that EXPD 114G, EXPD 114P, EXPD 114PG meet the acceptance criteria and are adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, and safety testing demonstrates that the device is as safe, as effective, and performs as well or better than the predicate devices.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.