K Number
K242582

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-09-27

(28 days)

Product Code
Regulation Number
882.5950
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.

Device Description

The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called the Optima Coil System (OptiBlock Line Extension). This submission focuses on demonstrating substantial equivalence to a predicate device (Optima Coil System, K223386) for which it is a modification or line extension.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of acceptance criteria and the reported device performance:

TestAcceptance CriteriaReported Device Performance
Visual Inspection, Dimensional Inspection, and Resistance CheckThe test samples shall meet established test acceptance criteria for visual physical damage, secondary diameter and length, and resistance.Pass
Simulated UseThe test samples shall be prepared in accordance with the instructions for use and meet established test acceptance criteria for device performance in a clinically relevant model.Pass

2. Sample size used for the test set and the data provenance:

The document explicitly states "The following non-clinical bench testing was performed to evaluate the device changes and to demonstrate substantial equivalence...". It mentions "test samples" but does not specify the sample size for either of the tests performed. The data provenance is from bench testing for a medical device line extension, not involving human subjects or clinical data in this submission. Therefore, country of origin is not applicable in the context of clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. Given that the studies are non-clinical bench tests, the "ground truth" would be established by engineering specifications and objective measurements rather than expert consensus on clinical cases.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not provided and is generally not applicable to non-clinical bench testing. Adjudication methods are typically used in clinical studies where expert consensus is needed to establish ground truth for ambiguous cases.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable as the described device is a neurovascular embolization coil system, not an AI software or a device that assists human readers in interpreting medical images. There is no mention of AI in the document.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not applicable for the same reasons as point 5. The device is a physical medical implant, not an algorithm.

7. The type of ground truth used:

For the "Visual Inspection, Dimensional Inspection, and Resistance Check," the ground truth would be engineering specifications and manufacturing tolerances. For "Simulated Use," the ground truth would be pre-defined performance criteria within a clinically relevant model, also based on engineering and functional specifications for the device's intended operation. There is no mention of expert consensus, pathology, or outcomes data being used to establish ground truth for these non-clinical tests.

8. The sample size for the training set:

This information is not provided and is not applicable given that this is not an AI/machine learning device. The concepts of "training set" and "ground truth for the training set" are relevant to machine learning model development, not for a physical medical device.

9. How the ground truth for the training set was established:

This information is not provided and is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

September 27, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

Balt USA, LLC Kavita Chandrashekar Senior Regulatory Affairs Specialist 29 Parker Irvine, California 92618

Re: K242582

Trade/Device Name: Optima Coil System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: August 29, 2024 Received: August 30, 2024

Dear Kavita Chandrashekar:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Sara S. Thompson -S

Sara Thompson, D.V.M. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242582

Device Name Optima Coil System

Indications for Use (Describe)

The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Applicant:Balt USA, LLC29 ParkerIrvine, CA 92618Registration No.: 3014162263
Primary Contact Person:Kavita Chandrashekar
Senior Regulatory Affairs Specialist
Telephone: (858) 405-1276
Email: kavita.chandrashekar@baltgroup.com
Secondary ContactBrandon Shepard
Director, Regulatory Affairs
Telephone: (949) 788-1443
Email: brandon.shepard@baltgroup.com
Date Summary Prepared:09/25/2024
Trade Name:Optima Coil System
Common Name:Neurovascular embolization device; Vascular embolization device
Review Panel:Neurology; Cardiovascular
Product Code:HCG, KRD
Regulation Number:21 CFR 882.5950 (HCG), 21 CFR 870.3300 (KRD)
Regulation Name:Device, Neurovascular Embolization;Device, Vascular, for Promoting Embolization
Device Classification:Class II
Predicate Device:Optima Coil System, 510(k): K223386

Device Description:

The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded.

Indications for Use:

The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.

Intended Use:

The Optima Coil System is intended for use in the peripheral and neuro-vasculature to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

{5}------------------------------------------------

Comparison of Technological Characteristics:

CategoryOptima Coil System (predicate deviceK223386)Modified Optima Coil System(OptiBlock Line Extension,subject device K242582)
Intended UseThe Optima Coil System is intended for use inthe peripheral and neuro-vasculature toendovascularly obstruct or occlude blood flowin vascular abnormalities of the neurovascularand peripheral vessels.Same
Indications for UseThe Optima Coil System is intended for theendovascular embolization of intracranialaneurysms and other neurovascularabnormalities such as arteriovenousmalformations and arteriovenous fistulae. TheOptima Coil System is also intended forvascular occlusion of blood vessels within theneurovascular system to permanently obstructblood flow to an aneurysm or other vascularmalformation and for arterial and venousembolizations in the peripheral vasculature.Same
Risk ClassClass IISame
Regulation NameNeurovascular Embolization DeviceSame
Generic NameNeurovascular Embolization DeviceSame
Anatomical LocationNeuro Vasculature & Peripheral VasculatureSame
VisualizationVisible under radiographic imagingSame
Method of SupplySterile; single-useSame
Coil Delivery MechanismPusher assemblySame
System ComponentsCoil (implant)Delivery SystemDetachment ControllerSame
Sterilization Method, SALGamma irradiation, 10-6Same
CategoryOptima Coil System (predicate device K223386)Modified Optima Coil System (OptiBlock Line Extension, subject device K242582)
Shelf-Life5 yearsSame
Coil (Implant features)
Main Coil Material92/8 Platinum/Tungsten (Pt/W) AlloySame
Coil Length3 cm – 65 cmSame
Primary Coil Wind Diameter0.012"-0.014"0.010" - 0.014"
Coil Secondary Shape Diameter1mm-14mmSame
Coil Wire Diameter0.00125" - .0030"Same
Secondary ShapesComplex, HelicalSame
Stretch-resistance Thread.0022" Polyolefin Engage thread (Annealed).0018" Polyolefin Engage thread (Annealed)Same
Coupler90/10 Platinum/Iridium (Pt/I) alloy MarkerbandSame
Primary Wind & Filar Combinations.012" x .00125".014" x .0015"0.014 x .00275" (gap) + 0.014 x 0.002"0.014 x .00275" (gap) + 0.014 x 0.0025"0.014" x .003" (gap) + 0.014" x .003".012" x .00125".014" x .0015".010" x .00125".014" x .00275" (gap) + .014" x .002".014" x .00275" (gap) + .014" x .0025"0.014 x 0.003" (gap) + 0.014 x 0.003"
No. & Size of Distal Helical Loops(Complex Implant)3 loops of 100% nominal diameter (with gaps of 0.0015")Same
CategoryOptima Coil System (predicate deviceK223386)Modified Optima Coil System(OptiBlock Line Extension, subjectdevice K242582)
Body Coil Components4-part coil:1) Heater coil (92/8 Platinum/Tungsten)2) Distal Coil (Stainless Steel)3) Radio-opaque (RO) Coil (92/8Platinum/Tungsten)4) Proximal Coil (Stainless Steel)Same
HypotubeStainless Steel HypotubeSame
Connector (to Detachment Controller)Gold plated, stainless steel hypotubeSame
Fluorosafe MarkersPad printed PET Shrink tubeSame
EpoxyEpoxy 353NDSame
Lead WiresPolyimide coated silver lead wiresSame
Pusher Working Length185 cmSame
Delivery Wire Diameter0.0125" Stainless Steel Coil0.0125" Platinum/Tungsten Coil0.0135" Stainless Steel hypotubeSame
PET Jacket.016" ID x .00025" Wall PETSame
SolderGold solderSame
Bonding AdhesiveDymax 1128A-M-VTSame
Attachment Zone Length0.045"Same
Detachment Equipment
Detachment Equipment ComponentsXcel Detachment ControllerSame
User InterfaceHandheldSame
Power SourceBattery, n=3 12V, size A23Same
Detachment MechanismThermal via detachment controllerSame
Controller SterilizationEthylene oxide gasSame
Method of supplySterile; single useSame

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

The modified Optima Coil System (OptiBlock Line Extension) and the predicate Optima Coil System differ in the following:

  • Addition of coil models with new combinations of diameter and length.
  • Reduction of the coil's helical proximal loops by 1mm or 2mm in diameter, for coil models ● with nominal loop diameters of 3-5mm and 6-14mm, respectively.
  • Implementation of 0.0015-inch gaps added to the primary wind of distal helical loops for ● secondary diameters ranging from 3 mm to 3.5 mm.
  • Implementation of a smaller .010" Primary wind OD. ●

The differences between the subject and predicate devices do not raise new questions of safety and effectiveness.

Performance Data - Bench:

The following non-clinical bench testing was performed to evaluate the device changes and to demonstrate substantial equivalence of the modified Optima Coil System (OptiBlock Line Extension) to the predicate device:

TestAcceptance CriteriaResults
Visual Inspection, DimensionalInspection, and Resistance CheckThe test samples shall meet established testacceptance criteria for visual physicaldamage, secondary diameter and length,and resistance.Pass
Simulated UseThe test samples shall be prepared inaccordance with the instructions for use andmeet established test acceptance criteria fordevice performance in a clinically relevantmodel.Pass

Conclusion:

There is no change to the intended use, materials, principles of operation or performance requirements of the modified Optima Coil System (OptiBlock Line Extension) in comparison to the predicate device (K223386). The successful completion of non-clinical bench testing demonstrates that the modified Optima Coil System (OptiBlock Line Extension) is substantially equivalent to the predicate device.

§ 882.5950 Neurovascular embolization device.

(a)
Identification. A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).