K Number
K223386

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-12-06

(29 days)

Product Code
Regulation Number
882.5950
Age Range
18 - 65
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.

Device Description

The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device called the "Optima Coil System." It describes the device, its intended use, and its comparison to a predicate device. It also includes a table of performance data from bench testing. However, this document does not contain information about a study involving human readers or AI assistance, or any data related to diagnostic accuracy, sensitivity, or specificity.

The "acceptance criteria" and "device performance" described are for bench testing only, which evaluates the physical and functional characteristics of the device itself, not its performance in a clinical scenario involving human interpretation or AI.

Therefore, I cannot fulfill the request for information related to:

  • A table of acceptance criteria and reported device performance in a human-AI study. The provided table only covers bench tests.
  • Sample size used for the test set and data provenance for a human-AI study.
  • Number of experts used to establish ground truth and their qualifications for a human-AI study.
  • Adjudication method for a human-AI study.
  • Multi-reader multi-case (MRMC) comparative effectiveness study or effect size.
  • Standalone (algorithm only) performance.
  • Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for a human-AI study.
  • Sample size for the training set for an AI study.
  • How ground truth for the training set was established for an AI study.

The document's "Performance Data - Bench" section is directly relevant to the acceptance criteria of the device's physical properties and functionality, not an AI or diagnostic performance study.

Here's the information that can be extracted from the provided text regarding device acceptance criteria and performance (bench testing):

Device Acceptance Criteria and Reported Performance (Bench Testing)

TestAcceptance CriteriaReported Device Performance
Visual Inspection, Dimensional Inspection, and Resistance CheckThe test samples shall meet established test acceptance criteria for visual physical damage and secondary diameter and length.Pass
Simulated UseThe test samples shall be prepared in accordance with the instructions for use and meet established test acceptance criteria for device performance in a clinically relevant model.Pass
Stretch ResistanceThe samples shall meet established test acceptance criteria for tensile strength.Pass

Explanation of the Bench Study (based on the provided text):

The performance data presented is from non-clinical bench testing. Its purpose was to evaluate changes made to the Optima Coil System (e.g., new coil sizes and combinations of dimensions) and to demonstrate its substantial equivalence to the previously cleared predicate device (K200030).

  • Sample size: Not explicitly stated for each bench test, but implied to be "test samples."
  • Data provenance: Bench testing results, implying in-house lab testing, not clinical data from patients or specific countries. It is retrospective in the sense that it evaluates a final product design.
  • Experts for ground truth: Not applicable for this type of bench testing. Ground truth is based on engineering specifications, physical measurements, and performance standards.
  • Adjudication method: Not applicable. Results are based on physical measurements and functional checks against pre-defined engineering criteria.
  • MRMC Comparative Effectiveness Study: No, this document does not describe such a study.
  • Standalone Performance: Not applicable in the context of AI; this refers to the physical device's performance.
  • Type of Ground Truth: Engineering specifications, physical dimensions, material properties, and functional performance in simulated environments.
  • Training Set Sample Size/Ground Truth for Training Set: Not applicable, as this document describes bench testing of a physical medical device, not the development or training of an AI algorithm.

In summary, the provided document focuses solely on the substantial equivalence of a physical neurovascular embolization device based on bench testing of its modifications. It does not contain any information related to AI performance, human-in-the-loop studies, or diagnostic efficacy.

{0}------------------------------------------------

December 6, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo includes a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Balt USA, LLC Kavita Chandrashekar Senior Regulatory Affairs Specialist 29 Parker Irvine, California 92618

Re: K223386

Trade/Device Name: Optima Coil System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: November 4, 2022 Received: November 7, 2022

Dear Kavita Chandrashekar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan -S

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223386

Device Name Optima Coil System

Indications for Use (Describe)

The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

X Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summarv K223386

Applicant:Balt USA, LLC29 ParkerIrvine, CA 92618Registration No.: 3014162263
Primary Contact Person:Kavita ChandrashekarSenior Regulatory Affairs SpecialistTelephone: (858) 405-1276Email : kavita.chandrashekar@baltgroup.com
Secondary ContactBrandon ShepardDirector, Regulatory AffairsTelephone: (949) 788-1443Email : brandon.shepard@baltgroup.com
Date Summary Prepared:December 02, 2022
Trade Name:Optima Coil System
Common Name:Neurovascular embolization device; Vascular embolization device
Review Panel:Neurology; Cardiovascular
Product Code:HCG, KRD
Regulation Number:21 CFR 882.5950 (HCG), 21 CFR 870.3300 (KRD)
Regulation Name:Device, Neurovascular Embolization;Device, Vascular, for Promoting Embolization
Device Classification:Class II
Predicate Device:Optima Coil System, 510(k): K200030

Device Description:

The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded.

Indications for Use:

The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.

Intended Use:

The Optima Coil System is intended for use in the peripheral and neuro-vasculature to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

{4}------------------------------------------------

Comparison of Technological Characteristics:

Table I Device Comparison
CategoryOptima Coil System (predicate device K200030)Modified Optima CoilSystem (subject deviceK223386)
Intended UseThe Optima Coil System is intended for use in theperipheral and neuro-vasculature to endovascularlyobstruct or occlude blood flow in vascular abnormalitiesof the neurovascular and peripheral vessels.Same
Indications for UseThe Optima Coil System is intended for the endovascularembolization of intracranial aneurysms and otherneurovascular abnormalities such as arteriovenousmalformations and arteriovenous fistulae. The OptimaCoil System is also intended for vascular occlusion ofblood vessels within the neurovascular system topermanently obstruct blood flow to an aneurysm or othervascular malformation and for arterial and venousembolizations in the peripheral vasculature.Same
Risk ClassClass IISame
Product CodeHCG, KRDSame
Common NameNeurovascular Embolization DeviceSame
Anatomical LocationNeuro Vasculature and Peripheral VasculatureSame
VisualizationVisible under radiographic imagingSame
Method of SupplySterile; single-useSame
Coil Delivery MechanismPusher assemblySame
System ComponentsCoil (implant)Delivery SystemDetachment ControllerSame
Main Coil Material92/8 Platinum/Tungsten (Pt/W) AlloySame
Coil Length1 cm - 65 cm3 cm - 65 cm
Coil Secondary Diameter1 mm – 24 mm1 mm – 14 mm
Coil Wire Diameter0.00125" - 0.0035"Same
Primary Coil Wind Diameter0.010" - 0.014"0.012" - 0.014"
CategoryOptima Coil System (predicate device K200030)Modified Optima Coil System(subject device K223386)
No. & Size of Helical Loopson Distalmost End(Complex only)1.5 loops of 70% nominal diameter3 loops of 70% nominal secondary diameter3 loops of 100% nominal secondary diameter (with gaps of 0.0015")
Primary Wind and FilarCombinationsComplexHelicalComplexHelical
.010" x .00125".011" x .0015".010" x .0015".011" x .00175".012" x .002".013" x .00225".014" x .003".014" x .0035".012" x .00125".014" x .0015".014" x .00175".014" x .002".012" x .002".011" x .0015".010" x .00125".012" x .00125".014" x .0015".014" x .00175".014" x .002".012" x .00125".014" x .0015".014" x .0020".014" x .0025".014" x .00275"(gap).014" x .0030".014" x .0030"(gap)There are no new dimensions that are subject to this submission.
No. of sizes offered209239 (30 additional models as part of this line extension, all new sizes fall within existing ranges)
DetachmentThermal via Detachment ControllerSame
PouchTyvek®Same
Shelf Life5 YearsSame
Sterilization Method,Sterility Assurance Level(SAL)Gamma irradiation, 10-6Same
Delivery System (pusher)
Construction/DesignBody coil laser welded to hypotubeSame
Body coil4-part coil:A. Heater Coil (92/8 Pt/W)B. Distal Coil (SSTL)Same

Table 1 Device Co .

{5}------------------------------------------------

{6}------------------------------------------------

C. Radio-opaque (RO, 92/8 Pt/W) Coil
D. Proximal Coil (SSTL)
HypotubeSSTL HypotubeSame
ConnectorGold plated, SSTL hypotubeSame
AdhesiveDymax 1128A-M-VTSame
JacketPETSame
Fluoro safe markersPad Printed PET Shrink tubeSame
EpoxyEpoxy 353 NDSame
Lead wiresPolyimide coated silver lead wiresSame
Detachment Controller
Coil detachmentThermal via detachment controllerSame

{7}------------------------------------------------

The modified Optima Coil System and the predicate Optimal Coil System differ in the following:

  • Addition of new sizes of the coil implant. ●
  • Addition of new combinations of coil OD, filar diameter, secondary shape diameter and ● lengths.

The differences between the subject and predicate devices do not raise new questions of safety and effectiveness.

Performance Data - Bench:

The following non-clinical bench testing was performed to evaluate the device changes and to demonstrate substantial equivalence of the modified Optima Coil System to the predicate device:

TestAcceptance CriteriaResults
Visual Inspection, DimensionalInspection, and Resistance CheckThe test samples shall meet established testacceptance criteria for visual physicaldamage and secondary diameter and length.Pass
Simulated UseThe test samples shall be prepared inaccordance with the instructions for use andmeet established test acceptance criteria fordevice performance in a clinically relevantmodel.Pass
Stretch ResistanceThe samples shall meet established testacceptance criteria for tensile strength.Pass

Conclusion:

There is no change to the intended use, materials, principles of operation or performance requirements of the modified Optima Coil System in comparison to the predicate device (K200030). The successful completion of non-clinical bench testing demonstrates that the modified Optima Coil System is substantially equivalent to the predicate device.

§ 882.5950 Neurovascular embolization device.

(a)
Identification. A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).