K Number
K242532

Validate with FDA (Live)

Device Name
BTL-785BNF-E
Date Cleared
2024-12-13

(109 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is indicated for aesthetic use including facial and neck or body skin stimulation.

Device Description

The BTL-785BNF-E is a state-of-the-art electrostimulation device that enables the application of therapy by electrical stimulation to patient's skin for aesthetic use. The main control unit is equipped with a large color touch screen that facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are set using the touch screen of the device. During the therapy, the device displays information about the applied therapy type, the remaining therapy time and the main therapy parameters on the screen. The generated electrical stimulation is intenact with the tissues of the human body in order to achieve skin stimulation in the face. The device consists of the following main components: BTL-785 main unit, BTL-785-7 Handpiece, Single-use application electrodes, Holding arm, Connection cables, Therapy discomfort button.

AI/ML Overview

The provided document is a 510(k) summary for the BTL-785BNF-E device. It describes the device's indications for use, its substantial equivalence to a predicate device, and the non-clinical testing performed.

Crucially, the document explicitly states under the "Clinical performance data" section: "A clinical study was not conducted."

This means that the information typically required to answer your questions regarding acceptance criteria and performance studies (such as sample size, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth establishment) is not available within this document because no clinical study was performed.

The substantial equivalence determination for this device is based on its similarity to a legally marketed predicate device (BTL-785BNF Handpiece, K233604) and non-clinical testing (electrical safety, EMC, usability, risk management, and biocompatibility standards compliance), not on a new clinical performance study. The key difference in the subject device is the introduction of a new disposable applicator (BTL-785-7-11), which the manufacturer asserts does not significantly influence safety or effectiveness given its similar materials, manufacturing, and technology.

Therefore, I cannot provide a table of acceptance criteria and proven performance based on a clinical study for this device, nor can I answer the specific questions about clinical study design parameters.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a human figure. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.

December 13, 2024

BTL Industries, Inc. David Chmel CEO North America 362 Elm Street Marlborough, Massachusetts 01752

Re: K242532

Trade/Device Name: BTL-785BNF-E Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: November 20, 2024 Received: November 20, 2024

Dear David Chmel:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Heather L. Dean -S

Heather Dean, PhD

Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K242532

Device Name

BTL-785BNF-E

Indications for Use (Describe)

The BTL-785BNF-E device has the following indications for use: The device is indicated for aesthetic use including facial and neck or body skin stimulation. Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary K242532

General Information

Sponsor:BTL Industries, Inc.362 Elm StreetMarlborough, MA 01752Tel: +1-866-285-1656Fax: +1-888-499-2502
Applicant:BTL Industries, Inc.362 Elm StreetMarlborough, MA 01752Tel: +1-866-285-1656Fax: +1-888-499-2502
Contact Person:David ChmelBTL Industries, Inc.chmel@btlnet.com
Summary PreparationDate:November 20, 2024

Device

Trade/Proprietary Name:BTL-785BNF-E
Primary Classification Name:Transcutaneous Electrical Nerve Stimulator for Pain Relief
Classification Regulation:21 CFR 882.5890, Class II
Classification Product Code:NFO

{5}------------------------------------------------

Legally Marketed Predicate Device

The BTL-785BNF-E is a state-of-the-art device with accessories intended for electrostimulation therapy and RF warming, and is substantially equivalent to the following product that is already cleared for distribution in the USA under the following 510(k) Premarket Notification number:

  • BTL-785BNF Handpiece (K233604) ●

Product Description

The BTL-785BNF-E is a state-of-the-art electrostimulation device that enables the application of therapy by electrical stimulation to patient's skin for aesthetic use.

The main control unit is equipped with a large color touch screen that facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are set using the touch screen of the device. During the therapy, the device displays information about the applied therapy type, the remaining therapy time and the main therapy parameters on the screen.

The generated electrical stimulation is intenact with the tissues of the human body in order to achieve skin stimulation in the face.

The device consists of the following main components:

  • · BTL-785 main unit
  • BTL-785-7 Handpiece
  • Single-use application electrodes
  • Holding arm
  • · Connection cables
  • Therapy discomfort button

{6}------------------------------------------------

Indications for Use

The BTL-785BNF-E device has the following indications for use:

The device is indicated for aesthetic use including facial and neck or body skin stimulation.

Non-clinical Testing (Performance Data)

The BTL-785BNF-E device has been thoroughly evaluated for electrical safety. The device has been found to comply with applicable medical device safety standards:

IEC 60601-1Medical electrical equipment – Part 1: General requirements for basic safety andessential performance
IEC 60601-1-2Medical electrical equipment – Part 1-2: General requirements for basic safetyand essential performance - Collateral standard: Electromagnetic disturbances– Requirements and tests
IEC 60601-2-10Medical electrical equipment – Part 2-10: Particular requirements for the basicsafety and essential performance of nerve and muscle stimulators
IEC 60601-1-6Medical electrical equipment – Part 1-6: General requirements for basic safetyand essential performance - Collateral standard: Usability
IEC 62304Medical device software – Software life cycle processes
ISO 14971Medical devices – Application of risk management to medical devices
ISO 10993-1Biological evaluation of medical devices – Part 1: Evaluation and testing withina risk management process
ISO 10993-5Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
ISO 10993-10Biological evaluation of medical devices – Part 10: Tests for irritation and skinsensitization

Clinical performance data

A clinical study was not conducted.

{7}------------------------------------------------

Comparison with the Predicate Device

510(k) numberNot assignedK232172
Device nameBTL-785BNF-EBTL-785BNF Handpiece
Company nameBTL Industries, Inc.BTL Industries, Inc.
TypeSubject devicePredicate device
Product Codeand RegulationNeurology21 CFR 882.5890NFO - Stimulator, TranscutaneousElectrical, Aesthetic PurposesNeurology21 CFR 882.5890NFO - Stimulator, TranscutaneousElectrical, Aesthetic Purposes
Clinical useRx OnlyRx Only
Indications forUseThe device is indicated for aesthetic useincluding facial and neck or body skinstimulation.The device is indicated for aesthetic useincluding facial and neck or body skinstimulation.
Anatomic SitesFace, neck, and bodyFace, neck, and body
Compliance withVoluntaryStandardsIEC 60601-1IEC 60601-1-2IEC 60601-2-10ISO 14971IEC 62366IEC 60601-1IEC 60601-1-2IEC 60601-2-10ISO 14971IEC 62366
Handpiece Weight2500 g2500 g
HandpieceDimensions[W x H x D]38.98" x 53.94" x 34.65"38.98" x 53.94" x 34.65"
Housing Materialsand ConstructionPCBs and leads inside plastic, or foamcase housingPCBs and leads inside plastic, or foamcase housing
User InterfaceDisplayYes on the main controll unit.Yes on the main controll unit.
Energy typeElectric stimulationElectric stimulation
Power sourceMains power100-240V, 50-60 HzMains power100-240V, 50-60 Hz
Waveform ShapeBiphasic rectangular pulse modulated bytrapezoidalBiphasic rectangular pulse modulated bytrapezoidal

{8}------------------------------------------------

BTL-785-7-2, -8, -9BTL-785-7-2, -8, -9
48,5V @ 500 Ω48,5V @ 500 Ω
48,5V @ 2000 Ω48,5V @ 2000 Ω
0 @ 10 000 Ω0 @ 10 000 Ω
Max outputvoltage [V](+/-20%)
- at 500 ΩBTL-785-7-1, -7BTL-785-7-1, -7
- at 2,000 Ω48,5V @ 500 Ω48,5V @ 500 Ω
- at 10,000 Ω48,5V @ 2000 Ω48,5V @ 2000 Ω
0 @ 10 000 Ω0 @ 10 000 Ω
BTL-785-7-11
48,5V @ 500 Ω
48,5V @ 2000 Ω
0 @ 10 000 Ω
BTL-785-7-2, -8, -9BTL-785-7-2, -8, -9
Max. 130 mA @ 10 - 373 OhmMax. 130 mA @ 10 - 373 Ohm
97 mA @ 500 Ω97 mA @ 500 Ω
24 mA @ 2000 Ω24 mA @ 2000 Ω
Max outputcurrent(+/-20%)0 mA @ 10000 Ω0 mA @ 10000 Ω
- at 500 ΩBTL-785-7-1, -7BTL-785-7-1, -7
Max. 64 mA @ 10 - 757 ΩMax. 64 mA @ 10 - 757 Ω
- at 2,000 Ω64 mA @ 500 Ω64 mA @ 500 Ω
24 mA @ 2000 Ω24 mA @ 2000 Ω
- at 10,000 Ω0 mA @ 10000 Ω0 mA @ 10000 Ω
BTL-785-7-11
Max. 30 mA @ 10 - 1900 Ω
30 mA @ 500 Ω
26 mA @ 2000 Ω
0 mA @ 10000 Ω

{9}------------------------------------------------

Max currentdensity at 500 Ω[µA/mm²]BTL-785-7-2 – 243 µA/mm²BTL-785-7-2 – 243 µA/mm²
BTL-785-7-8 – 335 µA/mm²BTL-785-7-8 – 335 µA/mm²
BTL-785-7-9 – 277 µA/mm²BTL-785-7-9 – 277 µA/mm²
BTL-785-7-1 – 168 µA/mm²BTL-785-7-1 – 168 µA/mm²
BTL-785-7-7 – 337 µA/mm²BTL-785-7-7 – 337 µA/mm²
BTL-785-7-11 – 164 µA/mm²
Max averagepowerdensity at 500 Ω[W/cm²]BTL-785-7-2 – 0,047 W/cm²BTL-785-7-2 – 0,047 W/cm²
BTL-785-7-8 – 0,065 W/cm²BTL-785-7-8 – 0,065 W/cm²
BTL-785-7-9 – 0,054 W/cm²BTL-785-7-9 – 0,054 W/cm²
BTL-785-7-1 – 0,033 W/cm²BTL-785-7-1 – 0,033 W/cm²
BTL-785-7-7 – 0,065 W/cm²BTL-785-7-7 – 0,065 W/cm²
BTL-785-7-11 – 0,032 W/cm²
Net Charge perpulseZero (pulse is biphasic symmetrical)Zero (pulse is biphasic symmetrical)
Max phase chargeper pulse at 500 Ω[µC]BTL-785-7-2, -8, -9BTL-785-7-2, -8, -9
15 µC15 µC
BTL-785-7-1, -7BTL-785-7-1, -7
10 µC10 µC
BTL-785-7-115 µC
Duration ofprimary [µs](depolarizingphase)80 µs80 µs
Pulse Duration[µs]160 µs160 µs

{10}------------------------------------------------

(a) ElectrodesmaterialsConductive ink (Silver)Conductive ink (Silver)
(b)electroconductivemediaConductive hydrogelConductive hydrogel
(c) electrode-to-skin impedancerange10-5000 Ohm10-5000 Ohm
(d) max durationof use20 min20 min
(e) conductivesurface areaBTL-785-7-1 - 3.8 cm²BTL-785-7-2 - 4 cm²BTL-785-7-7 - 1.9 cm²BTL-785-7-8 - 2.9 cm²BTL-785-7-9 - 3.5 cm²BTL-785-7-11 - 1.8 cm²BTL-785-7-1 - 3.8 cm²BTL-785-7-2 - 4 cm²BTL-785-7-7 - 1.9 cm²BTL-785-7-8 - 2.9 cm²BTL-785-7-9 - 3.5 cm²

Substantial Equivalence

The difference between the subject device and the predicate device (K232172) is the introduction of a new disposable applicator, the BTL-785-7-11. This applicator features a shape specifically designed to better fit the treatment area - the upper cheek and the periocular region outside the orbital rim. Otherwise, it remains substantially equivalent to the applicators previously cleared for the predicate device (K232172) in terms of materials used, manufacturing processes, technology used, method of use, and intended purpose.

The BTL-785BNF-E device has the same technological characteristics and intended use compared to the predicate device. Differences between the predicate device and the BTL-785BNF-E have no significant influence on safety or effectiveness of the BTL-785BNF-E device.

Therefore, the BTL-785BNF-E device is substantially equivalent to the predicate device.

Conclusion

Based upon the intended use, comparison of technical characteristics and performance testing provided in this premarket notification, the BTL-785BNF-E device has been shown to be substantially equivalent to the currently cleared predicate device for requested intended use.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).