K Number
K232172

Validate with FDA (Live)

Date Cleared
2023-09-18

(59 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BTL-785BNF Handpiece device has the following indications for use: The device is indicated for aesthetic use including facial and neck stimulation or body skin stimulation.

Device Description

The BTL-785BNF Handpiece is a state-of-the-art electrostimulation device that enables the application of therapy by an electromagnetic field. The subjected device is intended to be used together with a main control unit that is equipped with a large color touch screen that significantly facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen. The BTL-785BNF Handpiece is a specially developed device that can be used with the applicators and the main control unit. The generated electromagnetic field is interact with the tissues of the human body in order to achieve neuro-muscle stimulation. The radiofrequency is providing a low degree of warming to enhance the patient's comfort during the treatment. The BTL-785BNF Handpiece consists of the following main components: BTL-785BNF Handpiece, Single-use application electrodes, Holding arm, Connection cables, Therapy discomfort button.

AI/ML Overview

This document, primarily a 510(k) summary for the BTL-785BNF Handpiece, does not contain information about acceptance criteria for a device's performance in a clinical study, nor does it describe a study that explicitly proves the device meets specific acceptance criteria. This type of FDA submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting detailed performance metrics against pre-defined acceptance criteria from a clinical trial.

However, it does provide information about non-clinical testing performed to establish safety and compliance with relevant standards, which can be thought of as a form of "acceptance criteria" for safety.

Here's a breakdown of the requested information based on the provided text, noting where information is explicitly stated or absent:

  1. A table of acceptance criteria and the reported device performance

    • Not explicitly provided in the context of clinical performance. The document states, "A clinical study was not conducted."
    • The document does list compliance with voluntary standards as a form of performance or acceptance criteria for safety. The reported device performance is that it "has been found to comply with applicable medical device safety standards."
    Acceptance Criteria (Safety Standards)Reported Device Performance (Compliance)
    IEC 60601-1 (Basic safety & essential performance)Complies
    IEC 60601-1-2 (EMC)Complies
    IEC 60601-2-10 (Nerve & muscle stimulators)Complies
    IEC 60601-1-6 (Usability)Complies
    IEC 62304 (Software life cycle)Complies
    ISO 14971 (Risk management)Complies
    ISO 10993-1 (Biological evaluation)Complies
    ISO 10993-5 (In vitro cytotoxicity)Complies
    ISO 10993-10 (Irritation & sensitization)Complies
  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. The document explicitly states, "A clinical study was not conducted." Therefore, there is no test set in the traditional sense of a clinical trial. The "test set" for substantiating safety and performance was the device itself subjected to non-clinical laboratory testing against international standards.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No clinical study was conducted. The "ground truth" for the non-clinical tests would be the specifications and requirements of the referenced international safety standards. These do not involve human experts establishing ground truth in the context of interpreting clinical outcomes.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical study was conducted. Adjudication methods are typically relevant in clinical trials for resolving disagreements among readers or evaluators of clinical data.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This document is for a physical medical device (Transcutaneous Electrical Nerve Stimulator) and not an AI-assisted diagnostic or predictive device. Therefore, an MRMC study and AI assistance are not relevant here.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is not an algorithm; it is a physical electrostimulation device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the non-clinical safety and performance aspects, the "ground truth" would be the requirements and specifications outlined in the international consensus standards (e.g., IEC 60601-series, ISO 14971, ISO 10993-series). The device was tested to demonstrate compliance with these standards.
  8. The sample size for the training set

    • Not applicable. No clinical study was conducted, and this is not an AI/algorithm-based device that would typically involve a "training set."
  9. How the ground truth for the training set was established

    • Not applicable. As above, no training set for an AI/algorithm was used or established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The seal features a stylized depiction of a human figure, while the text is in a clear, sans-serif font, with "FDA" in a blue square.

September 18, 2023

BTL Industries Inc. David Chmel VP of Operations 362 Elm Street Marlborough, Massachusetts 01752

Re: K232172

Trade/Device Name: BTL-785BNF Handpiece Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NFO Dated: July 3, 2023 Received: July 21, 2023

Dear David Chmel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tushar Bansal -S

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K232172

Device Name BTL-785BNF Handpiece

Indications for Use (Describe)

The BTL-785BNF Handpiece device has the following indications for use:

The device is indicated for aesthetic use including facial and neck stimulation or body skin stimulation.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

General Information

Sponsor:BTL Industries, Inc.362 Elm StreetMarlborough, MA 01752Tel: +1-866-285-1656Fax: +1-888-499-2502
Applicant:BTL Industries, Inc.362 Elm StreetMarlborough, MA 01752Tel: +1-866-285-1656Fax: +1-888-499-2502
Contact Person:David ChmelBTL Industries, Inc.chmel@btlnet.com
Summary PreparationDate:June 23, 2023

Device

Trade/Proprietary Name:BTL-785BNF Handpiece
Primary Classification Name:Transcutaneous Electrical Nerve Stimulator for Pain Relief
Classification Regulation:21 CFR 882.5890, Class II
Classification Product Code:NFO

{4}------------------------------------------------

Legally Marketed Predicate Device

The BTL-785BNF Handpiece is a state-of-the-art device with accessories intended for electrostimulation therapy, and is substantially equivalent to the following products that are already cleared for distribution in the USA under the following 510(k) Premarket Notification numbers:

  • Avazzia OTC TENS for Aesthetics, model BEST-AV1: EZZI-LIFT Device ( K191951) ●

Product Description

The BTL-785BNF Handpiece is a state-of-the-art electrostimulation device that enables the application of therapy by an electromagnetic field.

The subjected device is intended to be used together with a main control unit that is equipped with a large color touch screen that significantly facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.

The BTL-785BNF Handpiece is a specially developed device that can be used with the applicators and the main control unit. The generated electromagnetic field is interact with the tissues of the human body in order to achieve neuro-muscle stimulation. The radiofrequency is providing a low degree of warming to enhance the patient's comfort during the treatment.

The BTL-785BNF Handpiece consists of the following main components:

  • BTL-785BNF Handpiece
  • · Single-use application electrodes
  • Holding arm
  • · Connection cables
  • Therapy discomfort button

Indications for Use

The BTL-785BNF Handpiece device has the following indications for use:

The device is indicated for aesthetic use including facial and neck or body skin stimulation.

Non-clinical Testing (Performance Data)

The BTL-785BNF Handpiece device has been thoroughly evaluated for electrical safety. The device has been found to comply with applicable medical device safety standards:

{5}------------------------------------------------

IEC 60601-1Medical electrical equipment - Part 1: General requirements for basic safety andessential performance
IEC 60601-1-2Medical electrical equipment – Part 1-2: General requirements for basic safetyand essential performance - Collateral standard: Electromagnetic disturbances- Requirements and tests
IEC 60601-2-10Medical electrical equipment – Part 2-10: Particular requirements for the basicsafety and essential performance of nerve and muscle stimulators
IEC 60601-1-6Medical electrical equipment - Part 1-6: General requirements for basic safetyand essential performance – Collateral standard: Usability
IEC 62304Medical device software - Software life cycle processes
ISO 14971Medical devices – Application of risk management to medical devices
ISO 10993-1Biological evaluation of medical devices – Part 1: Evaluation and testing withina risk management process
ISO 10993-5Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10Biological evaluation of medical devices - Part 10: Tests for irritation and skinsensitization

Clinical performance data

A clinical study was not conducted.

Comparison with the Predicate Device

510(k) numberK232172K191951
Device nameCompany nameBTL-785BNF HandpieceBTL Industries, Inc.Avazzia OTC TENS for Aesthetics,model BEST-AV1: EZZI-LIFTDevice Avazzia, Inc
TypeSubject devicePredicate device
Product Codeand RegulationNeurology21 CFR 882.5890NFO– Stimulator, TranscutaneousElectrical, Aesthetic PurposesNeurology21 CFR 882.5890NFO– Stimulator, TranscutaneousElectrical, Aesthetic Purposes

{6}------------------------------------------------

Indications for UseThe device is indicated for aesthetic use including facial and neck or body skin stimulation.Device is indicated for over-the counter aesthetic use including facial and neck stimulation or body skin stimulation.
Anatomic SitesFace, neck, and bodyFace, neck, and body
Number of Output Modes14
Low Battery IndicatorN/AYes
Automatic shut offN/A60 min
Compliance with Voluntary StandardsIEC 60601-1IEC 60601-1-2IEC 60601-2-10ISO 14971IEC 62366IEC 60601-1IEC 60601-1-2IEC 60601-2-10ISO 14971IEC 62366
Compliance with 21 CFR 898YesYes
Weight2500 g150 g
Dimensions [W x H x D] inches38.98" x 53.94" x 34.65"2.6" X 4.7" X 1.35"
Housing Materials and ConstructionPCBs and leads inside plastic, or foam case housingPCBs inside plastic case housing
User Interface DisplayYes on the main controll unit.LEDs and switches
Energy typeElectric stimulationElectric stimulation
Additional energyRF warmingN/A
Number, Size, and Type of BatteriesMains power100-240V, 50-60 HzTwo 1.5 V AA batteries
Waveform ShapeBiphasic rectangular pulse modulated by trapezoidalpositive square wave followed by a damped sinusoidal waveform of variable duration depending on damping and body loading
Max output voltage [V] (+/-20%)- at 500 Ω- at 2,000 Ω- at 10,000 ΩBTL-785-2, -8, -9Max. 48,5V @ 10 - 373 Ω48,5V @ 500 Ω48,5V @ 2000 Ω0 @ 10 000 Ω-42 @ 500 Ω-122 @ 2000 Ω-348 @ 10 000 Ω
BTL-785-1, -7Max. 48,5V @ 10 - 757 Ω48,5V @ 500 Ω48,5V @ 2000 Ω0 @ 10 000 Ω
BTL-785-2, -8, -9Max. 130 mA @ 10 - 373 Ohm
Max output current(+/-20%)- at 500 Ω- at 2,000 Ω- at 10,000 Ω97 mA @ 500 Ω24 mA @ 2000 Ω0 mA @ 10000 Ω
BTL-785-1, -7Max. 64 mA @ 10 - 757 Ω64 mA @ 500 Ω24 mA @ 2000 Ω0 mA @ 10000 Ω
Max current density at500 Ω [μA/mm²]BTL-785-2 - 243 μA/mm²BTL-785-8 - 335 μA/mm²BTL-785-9 - 277 μA/mm²BTL-785-1 - 168 μA/mm²BTL-785-7 - 337 μA/mm²Built-in, Y,Brush: 800 μA/mm²Pencil: 19,000 μA/mm²
Max average powerdensity at 500 Ω[μW/mm²]BTL-785-2 - 0,047 W/cm²BTL-785-8 - 0,065 W/cm²BTL-785-9 - 0,054 W/cm²BTL-785-1 - 0,033 W/cm²BTL-785-7 - 0,065 W/cm²Built-in, Y,Brush: 500 μW/mm²Pencil: 3,500 μW/mm²
Net Charge per pulseZero (pulse is biphasic symmetrical)4 μC
Max phase charge perpulse at 500 Ω [μC]BTL-785-2, -8, -915 μCBTL-785-1, -710 μC10 μC
10 μC
Duration of primary[μs](depolarizing phase)80 μs500 μs
Pulse Duration [μs]160 μs1 100 μs
Frequency [Hz]25015 to 121
Electrodes(a) materialsConductive ink (Silver)Stainless steel 316
(b) electroconductivemediaConductive hydrogeln/a
(c) electrode-to-skinimpedance range10-5000 Ohmless than 15
(d) max duration ofuse (same as deviceshut off)20 min60 min
(e) conductivesurface areaBTL-785-1 - 3.8 cm²BTL-785-2 - 4 cm²BTL-785-7 - 1.9 cm²BTL-785-8 - 2.9 cm²BTL-785-9 - 3.5 cm²built in rectangular - 1.94 cm²spherical - 5.60 cm²brush/comb - 0.56 cm²small circular,pencil-like - 0.023 cm²

{7}------------------------------------------------

{8}------------------------------------------------

Substantial Equivalence

The BTL-785BNF Handpiece device has equivalent technological characteristics and similar intended use compared to the predicate device.

The predicate device utilizes electric stimulation in order to achieve therapeutic effect. The subjected device BTL-785BNF Handpiece also utilizes electromagnetic stimulation for its main therapeutic effect; however, the device is also equipped with a monopolar radiofrequency generator. The RF warms the patient at the application site to increase patient convenience during therapy.

Any differences between the predicate devices and BTL-785BNF Handpiece have no significant influence on safety or effectiveness of the BTL-785BNF Handpiece.

Therefore, the BTL-785BNF Handpiece device is substantially equivalent to the predicate device.

Conclusion

Based upon the intended use, comparison of technical characteristics and performance testing provided in this premarket notification, the BTL-785BNF Handpiece device has been shown to be substantially equivalent to the currently cleared predicate device and secondary predicate device for requested intended use.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).