Intended Use

Endo Motor with E-LITE MAX, E-LITE PRO and ELITE-INO is cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used for preparation and enlargement of root canals, or measuring the canal length. And it can be used to enlarge the canals while monitoring the position of the file tip inside the canal.

Device Description

E-LITE MAX & ELITE-INO:

The Endo Motor with E-LITE MAX and ELITE-INO is a low-speed rotating oral equipment mainly used for root canal preparation and root canal measurement. It is designed to enlarge the root canal, facilitate dentin filling, or measuring the root canal length. The product is a portable device consist of a cordless hand-piece and a charge base, it is powered by built-in lithium batteries and charged by the adapter. LCD displays parameters such as speed, torque, file system, Apex etc. Users can also set and modify by keys, and provide design of factory initialization, and calibration of the apex location. It can be used as a low-speed motorized handpiece and device for measuring canal length. The rotary speed and torque range are adjustable.

There are three function features for E-LIT MAX and ELITE-INO:

  • Root Canal enlargement: Prepare the root canal with apex locator function.
  • Apex Location: Measure the length of the root canal.
  • Multi-function: Measuring the length while root canal preparation.

The device can save 9 programs, each program can set different parameters according to the user.

E-LITE PRO:

The Endo Motor with E-LITE PRO is a low-speed rotating oral equipment mainly used for root canal preparation. It is designed to enlarge the root canal or facilitate dentin filling. The product is a portable device having a cordless hand-piece with a charging adapter, it is powered by built-in lithium batteries and charged by the adapter. LCD displays parameters such as speed, torque, file system, Apex etc. Users can also set and modify by keys, and provide design of factory initialization, and calibration of the apex location. It can be used as a low-speed motorized handpiece and device for measuring canal length. The rotary speed and torque range are adjustable.

There is only one function features for E-LITE PRO:

  • Root Canal enlargement: Prepare the root canal, without apex locator function.

The device can save 9 programs, each program can set different parameters according to the user.

The Endo Motor must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

The Endo Motor is intended to be sterilized prior to use.

The compatible device can be used with Endo Motor is rotary instruments such as rotary root canal files, the working part of rotary instruments normally made of Nickel titanium (NiTi). These files are not included in the submission.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the "Endo Motor" (K242514) indicate that this is a dental handpiece and accessories, specifically an endodontic motor with root canal measurement capabilities. This type of device is not an AI/ML-enabled medical device and therefore the provided document does not contain the information typically found in an FDA submission for AI/ML devices regarding acceptance criteria, training data, ground truth establishment, or clinical study design (e.g., MRMC studies).

The document states: "No clinical test data was used to support the decision of substantial equivalence." This further confirms that a study validating an AI/ML component was not part of this clearance.

The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the performance of a traditional medical device (an endomotor) against established electrical safety, EMC, usability, and functional standards, as opposed to the performance of an AI/ML algorithm.

Therefore, I cannot provide the requested information related to AI/ML device validation because it is not present in the provided document. The document describes compliance with recognized standards for traditional medical devices (e.g., IEC 60601-1 for electrical safety, ISO 10993-1 for biocompatibility).

However, I can extract the general "acceptance criteria" and "device performance" as understood for this type of non-AI device from the provided tables, focusing on the comparative aspects:


Acceptance Criteria and Device Performance (Based on traditional device comparison):

The acceptance criteria for the Endo Motor (E-LITE MAX, E-LITE PRO, ELITE-INO) are primarily established through a direct comparison to a predicate device (Guilin Woodpecker Medical Instrument Co., Ltd.'s Endo Motor, K203320) and compliance with relevant international standards for medical devices. The objective is to demonstrate that the proposed device is "substantially equivalent" in terms of safety and effectiveness.

1. Table of Acceptance Criteria and Reported Device Performance

Feature/CriterionAcceptance Criteria (from Predicate Device or Standards)Reported Device Performance (Proposed Device)Judgment
General
Device ClassClass 1Class 1SE
Prescription UsePrescription UsePrescription UseSE
Anatomical SitesRoot canal, Softened dentinRoot canal, Softened dentinSE
Target PopulationPatients in need of root canal surgeryPatients in need of root canal surgerySE
Where UsedDental Clinic, University, Hospital, other clinical settingsDental Clinic, University, Hospital, other clinical settingsSE
Intended UseCordless endodontic treatment motorized handpiece with root canal measurement capability. Can be used for preparation and enlargement of root canals, or measuring canal length, and enlarging while monitoring file tip position.Identical to predicate device.SE
Safety & Performance
Electrical SafetyCompliance with IEC 60601-1Compliance with IEC 60601-1SE
EMCCompliance with IEC 60601-1-2Compliance with IEC 60601-1-2SE
Usability EngineeringCompliance with IEC 62366-1Compliance with IEC 62366-1SE
SoftwareCompliance with IEC 62304Compliance with IEC 62304SE
BiocompatibilityCompliance with ISO 10993-1Compliance with ISO 10993-1SE
Power SupplyLithium battery (DC3.7)Lithium battery (DC3.7)SE
Charger Power SupplyAC100V-240VAC100V-240VSE
Frequency of Supply Voltage50Hz ~ 60Hz50Hz ~ 60HzSE
ComponentsMeasuring wire, File clip, Lip hook, Touch probe, Tester, Spray NozzleMeasuring wire, File clip, Lip hook, Touch probe, Tester, Spray NozzleSE
Safety MechanismAutomatically stops or reversely rotates as soon as the file reaches the apical stop, to prevent perforation.Automatically stops or reversely rotates as soon as the file reaches the apical stop, to prevent perforation.SE
Micro MotorBrushlessBrushlessSE
USB PortsYesYesSE
Operation ModeContinuous operationContinuous operationSE
Key Performance Parameters (with noted differences)
Dimension (Handpiece)Ai-Motor: 28x33.7x199.2mm; Endo Radar Plus: 27.8x29x212mmE-LITE MAX: φ32x150mm; E-LITE PRO: φ30x151mm; ELITE-INO: φ30x200mmNote 1 (ergonomic design doesn't compromise safety/performance)
Weight (Handpiece)Ai-Motor: 150g; Endo Radar Plus: 151gE-LITE MAX: 110g; E-LITE PRO: 120g; ELITE-INO: 175gNote 1 (weight reduction doesn't compromise structural integrity/performance)
Torque Range0.4 Ncm ~ 5 NcmE-LITE MAX: 0.63.9N·cm; ELITE-INO: 0.64.0N·cm; E-LITE PRO: 0.6~3.9N·cmNote 2 (adjustable by operator; device complies with IEC80601-2-60 and has auto-reverse/stop)
Rotate Speed100 rpm~1200 rpmE-LITE MAX: 150800 rpm; ELITE-INO: 1201200 rpm; E-LITE PRO: 150~800 rpmNote 3 (within predicate range; does not affect safety/effectiveness)
Primary Contra Angle Gear Ratio6:1E-LITE MAX: 16:1; ELITE-INO: 6:1; E-LITE PRO: 1:1Note 4 (tested; does not affect safety/effectiveness)

2. Sample Size Used for the Test Set and Data Provenance

This document does not describe a "test set" in the context of an AI/ML model's performance on a dataset of patient cases. Instead, "tests" refer to engineering and biocompatibility validations of the physical device. Therefore, typical sample size metrics for AI/ML validation are not applicable here. The data provenance is implied by the testing standards (e.g., in-house testing by the manufacturer based on international standards). The nature of this submission indicates a retrospective comparison to an already cleared predicate device, rather than a prospective clinical trial.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This is not an AI/ML device and does not involve human expert adjudication of medical image data for ground truth establishment.

4. Adjudication Method for the Test Set

Not applicable for the reasons stated above.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. The document explicitly states: "No clinical test data was used to support the decision of substantial equivalence." This type of study is specifically relevant to AI/ML devices where the goal is to show the impact of the AI on human reviewer performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical endodontic handpiece, not a software algorithm.

7. The type of ground truth used

For this conventional medical device, "ground truth" refers to the established performance requirements defined by recognized standards (e.g., IEC, ISO) and the functional specifications of the predicate device. Compliance is demonstrated through laboratory testing of the physical device, rather than through expert consensus on medical data or pathology results.

8. The sample size for the training set

Not applicable, as this is not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable, as this is not an AI/ML device.

FDA 510(k) Clearance Letter - Endo Motor

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 24, 2025

Shenzhen Rogin Medical Co., Ltd.
℅ Salon Chen
System Engineer
IMD Medical & Drug Technology Service Institutions
Room 308, Building 11, No. 23 Jinqu Road, Wanjiang District
Dongguan, Guangdong 523039
CHINA

Re: K242514
Trade/Device Name: Endo Motor (E-LITE MAX, E-LITE PRO, E-LITE INO)
Regulation Number: 21 CFR 872.4200
Regulation Name: Dental Handpiece And Accessories
Regulatory Class: Class I, reserved
Product Code: EKX
Dated: September 25, 2024
Received: March 26, 2025

Dear Salon Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K242514 - Salon Chen
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K242514 - Salon Chen
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K242514

Device Name
Endo Motor (E-LITE MAX, E-LITE PRO, ELITE-INO)

Indications for Use (Describe)
Endo Motor with E-LITE MAX, E-LITE PRO and ELITE-INO is cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used for preparation and enlargement of root canals, or measuring the canal length. And it can be used to enlarge the canals while monitoring the position of the file tip inside the canal.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

K242514

1. Sponsor Identification:

  • Company Name: Shenzhen Rogin Medical Co., Ltd.
  • Address: 5th Floor, Block Block I, Jinchangda Science Park, Shangwei Industry District, Shangkeng Community, Guanhu Street, Longhua District, Shenzhen Guangdong, China 518110.
  • Phone: +86-755-26758272
  • Fax: +86-755-26758272
  • Contact Person (Title): Nature Xu (General Manager)
  • E-mail: admin@rogindental.com
  • Date of Preparation: April 23, 2025

2. Name of the Device:

  • Trade name: Endo Motor
  • Model: E-LITE MAX, E-LITE PRO, ELITE-INO

3. Common Name and Classification:

  • Trade/Proprietary Name: End Motor
  • Device Classification Name: Handpiece, Direct Drive, Ac-powered
  • Regulation Classification: Dental Handpiece and Accessories
  • Product Code: EKX
  • Regulation Number: 872.4200
  • Class: Class 1
  • Review Panel: Dental

4. Predicate Device Information:

  • 510(k) Number: K203320
  • Sponsor: Guilin Woodpecker Medical Instrument Co., Ltd.

Page 6

  • Trade/Proprietary Name: Endo Motor
  • Device Classification Name: Handpiece, Direct Drive, Ac-powered
  • Device Description: Dental Handpiece and accessories
  • Classification Product Code: EKX
  • Regulation Number: 21 CFR 872.4200
  • Class: 1
  • Review Panel: Dental

5. Application Correspondent

  • Company Name: IMD Medical & Drug technology service institutions
  • Phone: +86-18613190779
  • Fax: +86-755-62809168
  • Contact Person (Title): Salon Chen (System engineer)
  • E-mail: 33999439@qq.com
  • Address: Room 308, Building 11, No. 23 Jinqu Road, Wanjiang District, Dongguan City, Guangdong Province, China

6. Device Description

E-LITE MAX & ELITE-INO:

The Endo Motor with E-LITE MAX and ELITE-INO is a low-speed rotating oral equipment mainly used for root canal preparation and root canal measurement. It is designed to enlarge the root canal, facilitate dentin filling, or measuring the root canal length. The product is a portable device consist of a cordless hand-piece and a charge base, it is powered by built-in lithium batteries and charged by the adapter. LCD displays parameters such as speed, torque, file system, Apex etc. Users can also set and modify by keys, and provide design of factory initialization, and calibration of the apex location. It can be used as a low-speed motorized handpiece and device for measuring canal length. The rotary speed and torque range are adjustable.

There are three function features for E-LIT MAX and ELITE-INO:

Page 7

  • Root Canal enlargement: Prepare the root canal with apex locator function.
  • Apex Location: Measure the length of the root canal.
  • Multi-function: Measuring the length while root canal preparation.

The device can save 9 programs, each program can set different parameters according to the user.

E-LITE PRO:

The Endo Motor with E-LITE PRO is a low-speed rotating oral equipment mainly used for root canal preparation. It is designed to enlarge the root canal or facilitate dentin filling. The product is a portable device having a cordless hand-piece with a charging adapter, it is powered by built-in lithium batteries and charged by the adapter. LCD displays parameters such as speed, torque, file system, Apex etc. Users can also set and modify by keys, and provide design of factory initialization, and calibration of the apex location. It can be used as a low-speed motorized handpiece and device for measuring canal length. The rotary speed and torque range are adjustable.

There is only one function features for E-LITE PRO:

  • Root Canal enlargement: Prepare the root canal, without apex locator function.

The device can save 9 programs, each program can set different parameters according to the user.

The Endo Motor must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

The Endo Motor is intended to be sterilized prior to use.

The compatible device can be used with Endo Motor is rotary instruments such as rotary root canal files, the working part of rotary instruments normally made of Nickel titanium (NiTi). These files are not included in the submission.

7. Indications for Use

Endo Motor with E-LITE MAX, E-LITE PRO and ELITE-INO is cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used for

Page 8

preparation and enlargement of root canals, or measuring the canal length. And it can be used to enlarge the canals while monitoring the position of the file tip inside the canal.

8. Comparison to the predicate device

8.1. Table 1 General Comparison

Elements of ComparisonProposed DevicePredicate DeviceJudgment
Company NameShenzhen Rogin Medical Co., Ltd.Guilin Woodpecker Medical Instrument Co., Ltd./
Device Trade NameEndo MotorEndo Motor/
Device Common NameHandpiece, Direct Drive, Ac-poweredHandpiece, Direct Drive, Ac-poweredSE
Classification Product CodeEKXEKXSE
Regulation21 CFR 872.420021 CFR 872.4200SE
Regulation DescriptionDental handpiece and accessoriesDental handpiece and accessoriesSE
Review PanelDentalDentalSE
Class11SE
Prescription or OTCPrescription UsePrescription UseSE

Page 9

Anatomical SitesRoot canal Softened dentinRoot canal Softened dentinSE
Target PopulationPatients in need of root canal surgeryPatients in need of root canal surgerySE
Where usedDental ClinicUniversityHospitalThe other clinical settingsDental ClinicUniversityHospitalThe other clinical settingsSE
Intended UseEndo Motor with E-LITE MAX, E-LITE PRO and ELITE-INO is cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used for preparation and enlargement of root canals, or measuring the canal length. And it can be used to enlarge the canals while monitoring the position of the file tip inside the canal.Endo Motor, Ai-Motor and Endo Radar Plus are cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used for preparation and enlargement of root canals, or measuring the canal length. And it can be used to enlarge the canals while monitoring the position of the file tip inside the canal.SE

8.2. Table 2 Safety factor & Performance Comparison

Safety factor & PerformanceProposed DevicePredicate DeviceJudgment
Electrical SafetyCompliance with IEC 60601-1Compliance with IEC 60601-1SE
EMCCompliance with IEC 60601-1-2Compliance with IEC 60601-1-2SE
Usability EngineeringCompliance with IEC 62366-1Compliance with IEC 62366-1SE
SoftwareCompliance with IEC 62304Compliance with IEC 62304SE
BiocompatibilityCompliance with ISO 10993-1Compliance with ISO 10993-1SE

Page 10

Safety factor & PerformanceProposed DevicePredicate DeviceJudgment
DimensionE-LITE MAX: φ32 (Biggest Diameter) * 150mm (Length)E-LITE PRO: φ30 (Biggest Diameter) * 151mm (Length)ELITE-INO: 30(Biggest Diameter) * 200mm (Length)Motor Handpiece:Ai- Motor: 28mm x 33.7mm x 199.2mEndo Radar Plus: 27.8mm x 29mm x 212mmNote 1
WeightE-LITE MAX 110gE-LITE PRO 120gELITE-INO: 175gAi- Motor: 150gEndo Radar Plus: 151gNote 1
power SupplyLithium battery (DC3.7)Lithium battery (DC3.7)SE
Charger Power SupplyAC100V-240VAC100V-240VSE
Frequency of Supply Voltage50Hz ~ 60Hz50Hz ~ 60HzSE
ComponentsMeasuring wireFile clip Lip hookTouch probeTesterSpray NozzleMeasuring wireFile clip Lip hookTouch probeTesterSpray NozzleSE
Safety MechanismAutomatically stops or reversely rotates as soon as the file reaches the apical stop, so as to prevent perforation.Automatically stops or reversely rotates as soon as the file reaches the apical stop, so as to prevent perforation.SE

Page 11

Safety factor & PerformanceProposed DevicePredicate DeviceJudgment
Torque RangeE-LITE MAX:0.63.9N·cmELITE-INO: 0.64.0N·cmE-LITE PRO: 0.6~3.9N·cm0.4 Ncm ~ 5 NcmNote 2
Rotate SpeedE-LITE MAX:150800 rpmELITE-INO:120 rpm1200 rpmE-LITE PRO: 150~800 rpm(File in rotary mode)100 rpm~1200 rpm(File in rotary mode)Note 3
Adjustability of Speed and Torque0.5 Torque setting (Ncm)50 speed steps0.5 Torque setting (Ncm)50 speed stepsSE
Micro MotorBrushlessBrushlessSE
USB PortsYesYesSE
Primary Contra Angle Gear RatioE-LITE MAX :16:1ELITE-INO: 6:1E-LITE PRO: 1:16:1Note 4
Operation ModeContinuous operationContinuous operationSE

As shown in the above comparison Table, the proposed device is in similarity to the predicate device. Accordingly, the proposed devices are substantially equivalent to the predicate Endo Motor (K203320).

Design and Technology – The basic design and technology of providing Endo Motor is the same or similar.

Performance and Specifications – The subject device has similar handpiece system and specifications to the predicate device.

Page 12

Indications – The indication is identical with the predicate device.

8.3. Review of Differences:

Note1:

There are differences in the handpiece dimensions and weight between the proposed and predicate devices. The proposed devices have a cylindrical handle design with a slightly shorter length compared to the predicates. The proposed devices have passed usability compliance testing for their intended use. The weight reduction, combined with the ergonomic design, does not compromise the device's structural integrity or performance. Therefore, these differences do not introduce any new safety or effectiveness concerns when compared to the predicate devices.

Note2:

There is the difference in the torque range as compared to the predicate, however it is adjustable in the main unit by the operator, the device provides auto reverse and stop function when motor bear the torque resistance higher than the setting torque and the device comply with IEC80601-2-60 requirements, this difference would not affect its safety and effectiveness.

Note3:

The rotation speed of the prosed device is within the speed of the predicate device, thus substantially equivalent, the difference would not affect its safety and effectiveness.

Note4:

There is the difference in Primary Contra Angle Gear Ratio as compared to the predicate, Contra Angle Gear Ratio is tested, the difference would not affect its safety and

Page 13

effectiveness.

9. Discussion of Non-clinical Tests Performed for Determination of Substantial Equivalence are as follows:

The proposed device is the same as the predicate device but the two devices are different. A series of safety and performance tests were conducted on the subject device:

  • Biocompatibility Assessment per ISO 10993-1
  • Reprocessing Validation (cleaning, disinfection, sterilization)
  • Software Validation
  • Electromagnetic compatibility and electrical safety
  • Function test

All the test results demonstrate the Endo Motor meets the requirements of its pre-defined acceptance criteria and intended uses, and is substantially equivalent to the predicate device.

10. Clinical Tests Performed

No clinical test data was used to support the decision of substantial equivalence.

11. Conclusion

The subject devices have all features of the predicate devices. The few differences do not affect the safety and effectiveness of the subject devices.

Thus, the subject devices are substantially equivalent to the predicate devices.

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.