K Number
K242300

Validate with FDA (Live)

Device Name
MI View&GO
Date Cleared
2024-08-30

(28 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).

MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.

Device Description

MI View&GO is a software-only medical device which will be delivered in conjunction with Siemens SPECT/CT and PET/CT scanners. MI View&GO software provides additional specific capabilities for handling of PET and SPECT as well as CT and MR data directly at the acquisition console.

The MI View&GO software integrates molecular imaging more efficiently in the clinical environment by providing an interface for its users to review, post-process and read medical images immediately after acquisition. The purpose of the MI View&GO is to allow the technologist and reading physician to:

  • Review acquired and reconstructed images at the scanner console
  • Determine that the acquired data is of sufficient quality for reading, so the patient can be released.
  • Prepare images for reading
  • Perform a basic read
AI/ML Overview

The provided text is a 510(k) summary for the device MI View&GO VA30. It states that "The device under application, MI View&GO, did not conduct any additional performance testing" for determining substantial equivalence to the predicate device. Therefore, a study proving the device meets acceptance criteria as described in the prompt was not conducted and thus, the requested information (table of acceptance criteria, sample sizes, expert qualifications, etc.) is not available in the provided text.

The substantial equivalence determination was based on:

  1. Software Verification and Validation: This involved demonstrating continued conformance with special controls for medical devices containing software. The testing supported that all software specifications met predetermined acceptance criteria and substantiated all requirements and functional specifications, including those related to device hazards.
  2. Risk Analysis: A risk analysis was completed, and risk control was implemented to mitigate identified hazards.
  3. Comparison to Predicate Device: The device was deemed substantially equivalent to the predicate device (MI View&GO VA20, K201202) because there are no differences in the Indications for Use, Intended Use, or Fundamental Technological Characteristics. The new features implemented (e.g., O Brain AC-PC, PERCIST, VOI Isocontour) did not raise any new issues of safety and effectiveness.

In summary, the document indicates that the device's performance was not evaluated through a specific comparative study against acceptance criteria of the kind typically seen for novel diagnostic algorithms. Instead, its substantial equivalence was primarily established through software verification and validation, risk analysis, and a direct comparison of its technological characteristics and intended use to a previously cleared predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 30, 2024

Siemens Medical Solutions USA, Inc. Brian Wui Official Correspondent 2501 North Barrington Road Hoffman Estates, Illinois 60192

Re: K242300

Trade/Device Name: MI View&GO Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: August 2, 2024 Received: August 2, 2024

Dear Brian Wui:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

FDA

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K242300

Device Name

MI View&GO

Indications for Use (Describe)

MI View&GO is a medical diagnostic application for viewing, manipulation, guantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).

MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K242300

Image /page/3/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.

510(k) Summary

1. Identification of the Submitter

Submitter / Primary ContactPersonBrian WuiRegulatory Affairshansong.wui@siemens-healthineers.com+1 (865) 367-4337
Secondary Contact PersonClayton GinnRegulatory Affairsclayton.ginn@siemens-healthineers.com+1 (865) 898-2692
Applicant Name and AddressSiemens Medical Solutions, Inc. USA2501 North Barrington RoadHoffman Estates IL, 60192, USAEstablishment Registration Number: 1423253

2. Device Name and Classification

Product Trace Name:MI View&GO VA30
Common Name:Medical image management and processing system
Classification Name:Automated Radiological Image Processing Software
Classification Panel:Radiology
CFR Section:21 CFR §892.2050
Device Class:Class II
Product Code:QIH

3. Predicate Devices

Primary Predicate Device:

Product Trace Name:MI View&GO VA20
510(k) Number:K201202
Clearance Date:07/21/2022
Common Name:Medical image management and processing system
Classification Name:Automated Radiological Image Processing Software
Classification Panel:Radiology
CFR Section:21 CFR §892.2050
Device Class:Class II
Product Code:QIH

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.

Reference Device

Trade Name:syngo.via MI Workflows VB80
510(k) Number:K232000
Clearance Date:11/28/2023
Common Name:Medical image management and processing system
Classification Name:Automated Radiological Image Processing Software
Classification Panel:Radiology
CFR Section:21 CFR §892.5050
Device Class:Class II
Product Code:QIH

4. Device Description

MI View&GO is a software-only medical device which will be delivered in conjunction with Siemens SPECT/CT and PET/CT scanners. MI View&GO software provides additional specific capabilities for handling of PET and SPECT as well as CT and MR data directly at the acquisition console.

The MI View&GO software integrates molecular imaging more efficiently in the clinical environment by providing an interface for its users to review, post-process and read medical images immediately after acquisition. The purpose of the MI View&GO is to allow the technologist and reading physician to:

  • Review acquired and reconstructed images at the scanner console
  • Determine that the acquired data is of sufficient quality for reading, so the patient can be released.
  • Prepare images for reading
  • Perform a basic read

5. Indications for Use

MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).

MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.

6. Indications for Use Comparison to the Predicate Device

The indications for use are the same between the subject device and the primary predicate device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.

7. Comparison of Technological Characteristics with the Pro

MI View&GO with software version VA30 software provides the same technological characteristics in terms of materials, energy source, and control mechanisms when compared to the legally marketed predicate device since all devices are software only devices.

The software features have been modified in comparison to the predicate device to support enhanced device functionality.

The intended use, indications for use, and fundamental scientific technology for the subject device remains unchanged from the predicate device. No features present from the predicate device have been de-scoped.

At a high level, the subject and predicate devices are based on the following same technological elements:

  • . Data Supported (PET, SPECT, CT, MR)
  • . Server/Client architecture
  • . Workflow Activities (preprocessing, evaluation and reading, reporting and storage)
  • . Feature Licensing Structure
  • SUV values calculated

The following technological differences exist between the subject device and predicate devices.

MI View&GO VA30:

  • MI Features: ●
    • O Brain AC-PC(Anterior Commissure – Posterior Commissure)
    • O PERCIST
    • VOI Isocontour O
  • Shared Software Common Tools and CT Features: .
    • Layout Editor O
    • O Average
    • Stroke Layout O
  • . Layout Improvements:
    • MIP (Maximum Intensity Projection) Rotation O
    • o Volume Stripe Image Order

Any differences in technological characteristics do not raise different questions of safety and effectiveness. Testing and validation are completed. Test results show that the subject devices are comparable to the predicate devices in terms of technological characteristics and safety and effectiveness and therefore are substantially equivalent to the predicate devices.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a cluster of orange dots arranged in a circular pattern.

8. Non-Clinical and/or Clinical Test Summary & Conclusion

The following performance data were provided in support of the substantial equivalence determination.

Software Verification and Validation

'Enhanced' software documentation per FDA's guidance document "Content of Premarket Submissions for Device Software Functions" issued on June, 2023 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. The testing supports that all software specifications have met the predetermined acceptance criteria. Verification and validation testing substantiates all requirement and functional specifications, including specifications related to device hazards, and supports the claim of substantial equivalence.

Performance Testing

The device under application, MI View&GO, did not conduct any additional performance testing.

Risk Analysis

The risk analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the predetermined acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.

Standards

Siemens hereby certifies that MI View&GO meets the following FDA Recognized Consensus standards listed below:

DesignationNumber andEdition/DateTitleStandardsDevelopmentOrganizationRecognitionNumber
PS 3.1 - 3.202023eDigital Imaging and Communications in Medicine(DICOM) SetNEMA12-349
62304Edition 1.12015-06CONSOLIDATEDVERSIONMedical Device Software – Software Life CycleProcessesAAMI, ANSI,IEC13-79
14971 ThirdEdition2019-12Medical devices – Application of risk management tomedical devicesISO5-125

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words are a series of orange dots arranged in a circular pattern.

62366-1Edition 1.12020-06CONSOLIDATEDVERSIONMedical devices - Part 1: Application of usabilityengineering to medical devicesAAMI, ANSI,IEC5-129
15223-1Fourthedition2021-07Medical devices - Symbols to be used with medicaldevice labels, labelling, and information to besupplied - Part 1: General requirementsISO5-134
20417 Firstedition2021-04Correctedversion2021-12Medical devices - Information to be supplied by themanufacturerISO5-135

Conclusion

There are no differences in the Indications for Use, Intended Use, or Fundamental Technological Characteristics of the MI View&GO VA30 software as compared to the currently commercially available MI View&GO VA20 software (K222172).

Both the current and predicate devices are used for viewing, manipulation, quantification, analysis, and comparison of medical images from single or multiple imaging modalities with one or more time-points.

Additionally, the new features implemented within this release do not raise any new issues of safety and effectiveness as compared to the predicate device. Based on this information, as well as the documentation in support of the modifications, it is Siemens' opinion that the MI View&GO software with the modifications outlined in this application—is substantially equivalent to the predicate device.

N/A